Normal variation of magnetic resonance T1 relaxation times in the human population at 1.5 T using ShMOLLI
Open accessThe DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Quantitative T1-mapping is rapidly becoming a clinical tool in cardiovascular magnetic resonance (CMR) to objectively distinguish normal from diseased myocardium. The usefulness of any quantitative technique to identify disease lies in its ability to detect significant differences from an established range of normal values. We aimed to assess the variability of myocardial T1 relaxation times in the normal human population estimated with recently proposed Shortened Modified Look-Locker Inversion recovery (ShMOLLI) T1 mapping technique. METHODS: A large cohort of healthy volunteers (n = 342, 50% females, age 11-69 years) from 3 clinical centres across two countries underwent CMR at 1.5T. Each examination provided a single average myocardial ShMOLLI T1 estimate using manually drawn myocardial contours on typically 3 short axis slices (average 3.4 ± 1.4), taking care not to include any blood pool in the myocardial contours. We established the normal reference range of myocardial and blood T1 values, and assessed the effect of potential confounding factors, including artefacts, partial volume, repeated measurements, age, gender, body size, hematocrit and heart rate. RESULTS: Native myocardial ShMOLLI T1 was 962 ± 25 ms. We identify the partial volume as primary source of potential error in the analysis of respective T1 maps and use 1 pixel erosion to represent "midwall myocardial" T1, resulting in a 0.9% decrease to 953 ± 23 ms. Midwall myocardial ShMOLLI T1 was reproducible with an intra-individual, intra- and inter-scanner variability of ≤2%. The principle biological parameter influencing myocardial ShMOLLI T1 was the female gender, with female T1 longer by 24 ms up to the age of 45 years, after which there was no significant difference from males. After correction for age and gender dependencies, heart rate was the only other physiologic factor with a small effect on myocardial ShMOLLI T1 (6ms/10bpm). Left and right ventricular blood ShMOLLI T1 correlated strongly with each other and also with myocardial T1 with the slope of 0.1 that is justifiable by the resting partition of blood volume in myocardial tissue. Overall, the effect of all variables on myocardial ShMOLLI T1 was within 2% of relative changes from the average. CONCLUSION: Native T1-mapping using ShMOLLI generates reproducible and consistent results in normal individuals within 2% of relative changes from the average, well below the effects of most acute forms of myocardial disease. The main potential confounder is the partial volume effect arising from over-inclusion of neighbouring tissue at the manual stages of image analysis. In the study of cardiac conditions such as diffuse fibrosis or small focal changes, the use of "myocardial midwall" T1, age and gender matching, and compensation for heart rate differences may all help to improve the method sensitivity in detecting subtle changes. As the accuracy of current T1 measurement methods remains to be established, this study does not claim to report an accurate measure of T1, but that ShMOLLI is a stable and reproducible method for T1-mapping.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heart
Messroghli, D. R., Radjenovic, A., Kozerke, S., Higgins, D. M., Sivananthan, M. U., & Ridgway, J. P. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20110 · Cited by 1,349
Bottomley, P. A., Foster, T. H., Argersinger, R. E., & Pfeifer, L. M. (1984) · Medical Physics · doi:10.1118/1.595535 · Cited by 1,192
Flett, A. S., Hayward, M. P., Ashworth, M. T., Hansen, M. S., Taylor, A. M., Elliott, P. M., McGregor, C., & Moon, J. C. (2010) · Circulation · doi:10.1161/circulationaha.109.930636 · Cited by 915
Self-citationIles, L., Pfluger, H., Phrommintikul, A., Cherayath, J., Aksit, P., Gupta, S. N., Kaye, D. M., & Taylor, A. J. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.06.049 · Cited by 736
Piechnik, S. K., Ferreira, V. M., Dall'Armellina, E., Cochlin, L. E., Greiser, A., Neubauer, S., & Robson, M. D. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-69 · Cited by 670
Self-citationFerreira, V. M., Piechnik, S. K., Dall'Armellina, E., Karamitsos, T. D., Francis, J. M., Choudhury, R. P., Friedrich, M. G., Robson, M. D., & Neubauer, S. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-42 · Cited by 460
Self-citationMyocardial T1 mapping: Application to patients with acute and chronic myocardial infarction
Messroghli, D. R., Walters, K., Plein, S., Sparrow, P., Friedrich, M. G., Ridgway, J. P., & Sivananthan, M. U. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21272 · Cited by 351
Dass, S., Suttie, J. J., Piechnik, S. K., Ferreira, V. M., Holloway, C. J., Banerjee, R., Mahmod, M., Cochlin, L., Karamitsos, T. D., Robson, M. D., et al. (2012) · Circulation: Cardiovascular Imaging · doi:10.1161/circimaging.112.976738 · Cited by 348
Self-citation
Cited by
Messroghli, D. R., Moon, J. C., Ferreira, V. M., Grosse-Wortmann, L., He, T., Kellman, P., Mascherbauer, J., Nezafat, R., Salerno, M., Schelbert, E. B., et al. (2016) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-017-0389-8 · Cited by 1,727
Self-citationMoon, J. C., Messroghli, D. R., Kellman, P., Piechnik, S. K., Robson, M. D., Ugander, M., Gatehouse, P. D., Arai, A. E., Friedrich, M. G., Neubauer, S., et al. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-92 · Cited by 1,136
Self-citationT1-mapping in the heart: accuracy and precision
Kellman, P., & Hansen, M. S. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-2 · Cited by 764
Normal values for cardiovascular magnetic resonance in adults and children
Kawel-Boehm, N., Maceira, A., Valsangiacomo-Buechel, E. R., Vogel-Claussen, J., Turkbey, E. B., Williams, R., Plein, S., Tee, M., Eng, J., & Bluemke, D. A. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-015-0111-7 · Cited by 746
Kawel-Boehm, N., Hetzel, S. J., Ambale-Venkatesh, B., Captur, G., Francois, C. J., Jerosch-Herold, M., Salerno, M., Teague, S. D., Valsangiacomo-Buechel, E., van der Geest, R. J., et al. (2020) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-020-00683-3 · Cited by 610
Native T1 Mapping in Transthyretin Amyloidosis
Fontana, M., Banypersad, S. M., Treibel, T. A., Maestrini, V., Sado, D. M., White, S. K., Pica, S., Castelletti, S., Piechnik, S. K., Robson, M. D., et al. (2014) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2013.10.008 · Cited by 432
Self-citationT1 Mapping for the Diagnosis of Acute Myocarditis Using CMR
Ferreira, V. M., Piechnik, S. K., Dall'Armellina, E., Karamitsos, T. D., Francis, J. M., Ntusi, N., Holloway, C., Choudhury, R. P., Kardos, A., Robson, M. D., et al. (2013) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2013.03.008 · Cited by 383
Self-citationDabir, D., Child, N., Kalra, A., Rogers, T., Gebker, R., Jabbour, A., Plein, S., Yu, C.-Y., Otton, J., Kidambi, A., et al. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-014-0069-x · Cited by 347
T1 Mapping in Characterizing Myocardial Disease: A Comprehensive Review
Puntmann, V. O., Peker, E., Chandrashekhar, Y., & Nagel, E. (2016) · Circulation Research · doi:10.1161/circresaha.116.307974 · Cited by 337
T1 Mapping for Myocardial Extracellular Volume Measurement by CMR
White, S. K., Sado, D. M., Fontana, M., Banypersad, S. M., Maestrini, V., Flett, A. S., Piechnik, S. K., Robson, M. D., Hausenloy, D. J., Sheikh, A. M., et al. (2013) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2013.01.011 · Cited by 293
Self-citationNakamori, S., Dohi, K., Ishida, M., Goto, Y., Imanaka-Yoshida, K., Omori, T., Goto, I., Kumagai, N., Fujimoto, N., Ichikawa, Y., et al. (2018) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2017.04.006 · Cited by 260
Tham, E. B., Haykowsky, M. J., Chow, K., Spavor, M., Kaneko, S., Khoo, N. S., Pagano, J. J., Mackie, A. S., & Thompson, R. B. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-48 · Cited by 229
Ferreira, V. M., Piechnik, S. K., Dall'Armellina, E., Karamitsos, T. D., Francis, J. M., Ntusi, N., Holloway, C., Choudhury, R. P., Kardos, A., Robson, M. D., et al. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-36 · Cited by 226
Self-citationMyocardial T1 and T2 Mapping: Techniques and Clinical Applications
Kim, P. K., Hong, Y. J., Im, D. J., Suh, Y. J., Park, C. H., Kim, J. Y., Chang, S., Lee, H.-J., Hur, J., Kim, Y. J., et al. (2017) · Korean Journal of Radiology · doi:10.3348/kjr.2017.18.1.113 · Cited by 226
“T1 and T2 mapping CMR sequences are rapidly becoming the gold standard and have been integrated into routine CMR imaging protocols (2).”
BackgroundThompson, R. B., Chow, K., Khan, A., Chan, A., Shanks, M., Paterson, I., & Oudit, G. Y. (2013) · Circulation: Cardiovascular Imaging · doi:10.1161/circimaging.113.000482 · Cited by 194
Kellman, P., Arai, A. E., & Xue, H. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-56 · Cited by 177
“Although absolute accuracy of in-vivo measurements is an important and open subject, these methods have been shown to be highly reproducible [18] in practice despite bias errors.”
MethodNoncontrast myocardial T1 mapping using cardiovascular magnetic resonance for iron overload
Sado, D. M., Maestrini, V., Piechnik, S. K., Banypersad, S. M., White, S. K., Flett, A. S., Robson, M. D., Neubauer, S., Ariti, C., Arai, A., et al. (2015) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24727 · Cited by 175
“5T in which only three were found to have a T1 of <904 msec, with none having a value <901 msec (19); and 6) there is a narrower normal range for T1 compared to T2*, with normal T1 being 968 6 32 msec and T2* 31 6 4 msec in this study.”
BackgroundMethodInfluentialSelf-citationFerreira, V. M., Piechnik, S. K., Robson, M. D., Neubauer, S., & Karamitsos, T. D. (2014) · Journal of Thoracic Imaging · doi:10.1097/rti.0000000000000077 · Cited by 166
Self-citationRobson, M. D., Piechnik, S. K., Tunnicliffe, E. M., & Neubauer, S. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24867 · Cited by 161
“3-(3)-3-(3)-5 (MOLLI) HS1 135 267 142 1.”
MethodInfluentialSelf-citation
Metadata
- Authors
- Piechnik, S. K., Ferreira, V. M., Lewandowski, A. J., Ntusi, N. A., Banerjee, R., Holloway, C., Hofman, M. B., Sado, D. M., Maestrini, V., White, S. K., Lazdam, M., Karamitsos, T., Moon, J. C., Neubauer, S., Leeson, P., & Robson, M. D.
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2013-01
- Volume
- 15
- Issue
- 1
- Pages
- 13
- ISSN
- 1097-6647 (ISSN-L)
- OpenAlex
- W2143064593
- Cited by
- 273
- References
- 21
- FWCI
- 21.39
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.