Student or educator? Sign up with your school email and get 6 months of Pro free.
CoChat

The DOI matches its Crossref record; 38 of the 41 references it cites were found in a registry or an index.

Restoration of motor control and proprioceptive and cutaneous sensation in humans with prior upper-limb amputation via multiple Utah Slanted Electrode Arrays (USEAs) implanted in residual peripheral arm nerves

Open access

Abstract

BACKGROUND: Despite advances in sophisticated robotic hands, intuitive control of and sensory feedback from these prostheses has been limited to only 3-degrees-of-freedom (DOF) with 2 sensory percepts in closed-loop control. A Utah Slanted Electrode Array (USEA) has been used in the past to provide up to 81 sensory percepts for human amputees. Here, we report on the advanced capabilities of multiple USEAs implanted in the residual peripheral arm nerves of human amputees for restoring control of 5 DOF and sensation of up to 131 proprioceptive and cutaneous hand sensory percepts. We also demonstrate that USEA-restored sensory percepts provide a useful source of feedback during closed-loop virtual prosthetic hand control. METHODS: Two 100-channel USEAs were implanted for 4-5 weeks, one each in the median and ulnar arm nerves of two human subjects with prior long-duration upper-arm amputations. Intended finger and wrist positions were decoded from neuronal firing patterns via a modified Kalman filter, allowing subjects to control many movements of a virtual prosthetic hand. Additionally, USEA microstimulation was used to evoke numerous sensory percepts spanning the phantom hand. Closed-loop control was achieved by stimulating via an electrode of the ulnar-nerve USEA while recording and decoding movement via the median-nerve USEA. RESULTS: Subjects controlled up to 12 degrees-of-freedom during informal, 'freeform' online movement decode sessions, and experienced up to 131 USEA-evoked proprioceptive and cutaneous sensations spanning the phantom hand. Independent control was achieved for a 5-DOF real-time decode that included flexion/extension of the thumb, index, middle, and ring fingers, and the wrist. Proportional control was achieved for a 4-DOF real-time decode. One subject used a USEA-evoked hand sensation as feedback to complete a 1-DOF closed-loop virtual-hand movement task. There were no observed long-term functional deficits due to the USEA implants. CONCLUSIONS: Implantation of high-channel-count USEAs enables multi-degree-of-freedom control of virtual prosthetic hand movement and restoration of a rich selection of both proprioceptive and cutaneous sensory percepts spanning the hand during the short 4-5 week post-implant period. Future USEA use in longer-term implants and in closed-loop may enable restoration of many of the capabilities of an intact hand while contributing to a meaningful embodiment of the prosthesis.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/s12984-017-0320-4

Cites

38 of 38 references verified0 citation errors · 0 not checkable

Statistics

Figures

Fig. 1 USEAs implanted in human peripheral arm nerves were used to provide amputees with multi-DOF control of virtual prosthetic hand movement and restore numerous hand sensations. a Scanning electron microscope image of a USEA [21]. b Two USEAs were implanted in each subject (S4, shown here), one in each of the median and ulnar arm nerves. An organic nerve wrap, fastened with vascular clips, enclosed each USEA. c USEA lead wires and ground and reference wires were connected to external connectors via a percutaneous incision (S3, shown here). d USEA recordings were used to provide subjects with control of movement of a virtual prosthetic hand (S3, shown here). e USEA stimulation was used to provide subjects with numerous sensations on the phantom hand. Subjects documented the nature of each sensation (location, quality, and intensity/size) using custom software
Fig. 1 USEAs implanted in human peripheral arm nerves were used to provide amputees with multi-DOF control of virtual prosthetic hand movement and restore numerous hand sensations. a Scanning electron microscope image of a USEA [21]. b Two USEAs were implanted in each subject (S4, shown here), one in each of the median and ulnar arm nerves. An organic nerve wrap, fastened with vascular clips, enclosed each USEA. c USEA lead wires and ground and reference wires were connected to external connectors via a percutaneous incision (S3, shown here). d USEA recordings were used to provide subjects with control of movement of a virtual prosthetic hand (S3, shown here). e USEA stimulation was used to provide subjects with numerous sensations on the phantom hand. Subjects documented the nature of each sensation (location, quality, and intensity/size) using custom softwareCC BY · PubMed Central
Fig. 2 Number of working electrodes (impedance <500 kΩ) of 96 total recording electrodes per array over time for S3 and S4. The number of working electrodes for the ulnar array of S3 and median array of S4 is relatively stable, whereas the number of working electrodes steadily declines for the median array of S3 and ulnar array of S4
Fig. 2 Number of working electrodes (impedance <500 kΩ) of 96 total recording electrodes per array over time for S3 and S4. The number of working electrodes for the ulnar array of S3 and median array of S4 is relatively stable, whereas the number of working electrodes steadily declines for the median array of S3 and ulnar array of S4CC BY · PubMed Central
Fig. 3 USEA recordings were collected during a training session and used to train a decode (either a standard Kalman filter or a ReFIT Kalman filter). Subjects were then given online control of the virtual hand via real-time output from the trained decode. a Training data was collected by recording via USEAs while the subject imitated pre-programmed, single-DOF or multi-DOF virtual hand movements with their phantom hand. b Neuronal spiking was observed during intended movements (inset shows neuronal action potential waveforms). c For each trial, the firing rate on a given electrode was computed and compared to the movement cue position via correlation. d The correlation between firing-rate and movement cue was determined for each USEA electrode across many trials of a given movement, and electrodes with high correlations and independent activation for specific movements were selected as input for an online decode
Fig. 3 USEA recordings were collected during a training session and used to train a decode (either a standard Kalman filter or a ReFIT Kalman filter). Subjects were then given online control of the virtual hand via real-time output from the trained decode. a Training data was collected by recording via USEAs while the subject imitated pre-programmed, single-DOF or multi-DOF virtual hand movements with their phantom hand. b Neuronal spiking was observed during intended movements (inset shows neuronal action potential waveforms). c For each trial, the firing rate on a given electrode was computed and compared to the movement cue position via correlation. d The correlation between firing-rate and movement cue was determined for each USEA electrode across many trials of a given movement, and electrodes with high correlations and independent activation for specific movements were selected as input for an online decodeCC BY · PubMed Central
Fig. 4 Distinct patterns of USEA electrodes with firing rates correlated to movement position (across two electrode arrays, median array (m) and ulnar array (u)) are apparent for different movements (i.e. unique sub-populations of axons fire with specific movement efforts). Shown here are the patterns of the firing rates during movement cues for 2 DOFs (middle finger and wrist pitch) for S4. To see this figure in color, go online
Fig. 4 Distinct patterns of USEA electrodes with firing rates correlated to movement position (across two electrode arrays, median array (m) and ulnar array (u)) are apparent for different movements (i.e. unique sub-populations of axons fire with specific movement efforts). Shown here are the patterns of the firing rates during movement cues for 2 DOFs (middle finger and wrist pitch) for S4. To see this figure in color, go onlineCC BY · PubMed Central
Fig. 5 Number of electrodes with driven units for S4. Electrodes with activity correlated to volitional movement (r ≥ 0.5) were tabulated periodically throughout the experiment duration. The total number of driven electrodes peaks on post implant day 13, then steadily decreases over time
Fig. 5 Number of electrodes with driven units for S4. Electrodes with activity correlated to volitional movement (r ≥ 0.5) were tabulated periodically throughout the experiment duration. The total number of driven electrodes peaks on post implant day 13, then steadily decreases over timeCC BY · PubMed Central
Fig. 6 Decode output and raster plot during a 5-DOF target touching task for S4. Each figure shows the target position (solid black line) and decode output (dashed and colored line) on the top 5 lines. DOFs displayed in order from the top are thumb flex, index flex, middle flex, ring flex, and wrist flex/extend. Displayed below in black hash marks is the raster plot from 19 electrodes shown (15 from the median nerve array and 4 from the ulnar array). a thumb flexion targets b index finger flexion targets, c middle finger flexion targets, d ring finger targets, e thumb-index pinch target. Note that the subject was able to perform this combination movement despite training only single single-DOF training movements. f wrist flexion and extension targets
Fig. 6 Decode output and raster plot during a 5-DOF target touching task for S4. Each figure shows the target position (solid black line) and decode output (dashed and colored line) on the top 5 lines. DOFs displayed in order from the top are thumb flex, index flex, middle flex, ring flex, and wrist flex/extend. Displayed below in black hash marks is the raster plot from 19 electrodes shown (15 from the median nerve array and 4 from the ulnar array). a thumb flexion targets b index finger flexion targets, c middle finger flexion targets, d ring finger targets, e thumb-index pinch target. Note that the subject was able to perform this combination movement despite training only single single-DOF training movements. f wrist flexion and extension targetsCC BY · PubMed Central
Fig. 7 S4 tracked the position of three different moving virtual targets with the thumb, index, and middle fingers of a virtual prosthetic hand and then tracked the combined movement of all three targets with at least the middle finger and index finger. The top three traces depict the target location (solid line) and the subject-controlled, decoded virtual finger location (dashed line) for the thumb, index, and middle fingers. The subject independently tracked single thumb, index, and middle finger targets (depicted from left to right, respectively), and simultaneously tracked at least the middle and index finger targets in combination even though training data included no combination movements. The lower portion of the figure shows a raster indicating the times of recorded spike events from 8 selected median or ulnar nerve USEA electrodes during this task. The firing rates of spike events on these electrodes are uniquely tuned to different movements. To see this figure in color, go online
Fig. 7 S4 tracked the position of three different moving virtual targets with the thumb, index, and middle fingers of a virtual prosthetic hand and then tracked the combined movement of all three targets with at least the middle finger and index finger. The top three traces depict the target location (solid line) and the subject-controlled, decoded virtual finger location (dashed line) for the thumb, index, and middle fingers. The subject independently tracked single thumb, index, and middle finger targets (depicted from left to right, respectively), and simultaneously tracked at least the middle and index finger targets in combination even though training data included no combination movements. The lower portion of the figure shows a raster indicating the times of recorded spike events from 8 selected median or ulnar nerve USEA electrodes during this task. The firing rates of spike events on these electrodes are uniquely tuned to different movements. To see this figure in color, go onlineCC BY · PubMed Central
Fig. 8 “Best case scenario” for multi-DOF offline decode. Bar chart depicts the correlation coefficients between predicted and attempted training movements of the most correlated movements in a multi-DOF offline decode of a single data set. The data set contained 8 trials of each DOF where 4 trials were used for training the decoder and 4 trials were used for testing. The solid bar represents the mean correlation coefficient of the highest correlated N movements, where N is the number of simultaneous DOF tested (whiskers are standard deviations)
Fig. 8 “Best case scenario” for multi-DOF offline decode. Bar chart depicts the correlation coefficients between predicted and attempted training movements of the most correlated movements in a multi-DOF offline decode of a single data set. The data set contained 8 trials of each DOF where 4 trials were used for training the decoder and 4 trials were used for testing. The solid bar represents the mean correlation coefficient of the highest correlated N movements, where N is the number of simultaneous DOF tested (whiskers are standard deviations)CC BY · PubMed Central
Fig. 9 USEA microstimulation provided a rich selection of percepts of various qualities and locations spanning the phantom hand (S4 shown here). a Stimulation of individual electrodes via two USEAs restored 131 percepts across the phantom hand, including both proprioceptive and cutaneous percepts (collected over a 2-day period). Numerous cutaneous percepts were evoked on each digit and the palm, and proprioceptive percepts were restored for 17 different movements, including flexion and extension of each finger and flexion of the thumb. For proprioceptive percepts, upward arrows indicate extension, whereas downward arrows indicate flexion. b 131 electrodes across the 10 × 10 USEAs evoked the percepts shown in part A, with no apparent somatotopic arrangement across the nerve cross-section. c Evoked percepts were of various qualities, with 26% of evoked percepts described as proprioceptive, and 74% of evoked percepts being cutaneous (including ‘tingle’, ‘vibration’, and ‘pressure’). To see this figure in color, go online
Fig. 9 USEA microstimulation provided a rich selection of percepts of various qualities and locations spanning the phantom hand (S4 shown here). a Stimulation of individual electrodes via two USEAs restored 131 percepts across the phantom hand, including both proprioceptive and cutaneous percepts (collected over a 2-day period). Numerous cutaneous percepts were evoked on each digit and the palm, and proprioceptive percepts were restored for 17 different movements, including flexion and extension of each finger and flexion of the thumb. For proprioceptive percepts, upward arrows indicate extension, whereas downward arrows indicate flexion. b 131 electrodes across the 10 × 10 USEAs evoked the percepts shown in part A, with no apparent somatotopic arrangement across the nerve cross-section. c Evoked percepts were of various qualities, with 26% of evoked percepts described as proprioceptive, and 74% of evoked percepts being cutaneous (including ‘tingle’, ‘vibration’, and ‘pressure’). To see this figure in color, go onlineCC BY · PubMed Central
Fig. 10 Percepts evoked by median and ulnar nerve USEAs are generally within the established intact-hand innervation regions for each nerve. For the example shown (S3, week 2), 92% and 89% of median-nerve-USEA- and ulnar–nerve-USEA-evoked percepts are within their expected distributions, respectively. To see this figure in color, go online
Fig. 10 Percepts evoked by median and ulnar nerve USEAs are generally within the established intact-hand innervation regions for each nerve. For the example shown (S3, week 2), 92% and 89% of median-nerve-USEA- and ulnar–nerve-USEA-evoked percepts are within their expected distributions, respectively. To see this figure in color, go onlineCC BY · PubMed Central
Fig. 11 a The number of electrodes which evoked a sensory percept for each USEA across the implant duration (max amplitude, 120 μA, pulse duration of 200 μs). Note that perceptual thresholds for S4 were not tested on weeks 1, 3, and 4. Also, on week 5, S4’s ulnar n. USEA did not evoke any sensory percepts. b Weekly median and interquartile range boundaries threshold amplitudes across electrodes on each USEA. Outliers not shown
Fig. 11 a The number of electrodes which evoked a sensory percept for each USEA across the implant duration (max amplitude, 120 μA, pulse duration of 200 μs). Note that perceptual thresholds for S4 were not tested on weeks 1, 3, and 4. Also, on week 5, S4’s ulnar n. USEA did not evoke any sensory percepts. b Weekly median and interquartile range boundaries threshold amplitudes across electrodes on each USEA. Outliers not shownCC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 41 references

Research works printed but not linked: 4

  1. Clark GA, et al. Restoring cutaneous and proprioceptive somatosensory function with nerve stimulation after hand amputations. In: 2016 4th International Conference on Medical Bionics hosted by the Bionics Institute; Brisbane; 2016.
  2. G. A. Clark, D. J. Warren, N. M. Ledbetter, M. Lloyd, and R. A. Normann, “Microelectrode array system with integrated reference microelectrodes to reduce detected electrical noise and improve selectivity of activation,” US8359083 B2, 2013.
  3. Gescheider G. Chapter 3. The classical psychophysical methods. In: Psychophys. Fundam. 3rd Ed Mahwah Lawrence Erlbaum Assoc.; 1997.
  4. Wendelken SM, et al. Motor decoding and sensory stimulation for upper-limb prostheses using implanted neural and muscular electrode arrays. In: NANS2-Neural Interface Conference (NIC). Baltimore; 2016.

Cited by

See all 237

Metadata

Authors
Wendelken, S., Page, D. M., Davis, T., Wark, H. A. C., Kluger, D. T., Duncan, C., Warren, D. J., Hutchinson, D. T., & Clark, G. A.
Affiliations
University of Utah (US)
Corresponding author
Clark, Gregory A.
Venue
Journal of NeuroEngineering and Rehabilitation
Published
2017-11-25
Volume
14
Issue
1
Pages
121-121
ISSN
1743-0003 (ISSN-L)
OpenAlex
W2769345133
Cited by
216
References
41
Citations by year
2026: 7 · 2025: 26 · 2024: 20 · 2023: 22 · 2022: 25 · 2021: 42 · 2020: 30 · 2019: 33 · 2018: 11
FWCI
9.90
Type
journal article
Language
en
Open access
Licence
Locations
  • Journal of NeuroEngineering and Rehabilitation (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • Defense Advanced Research Projects Agency (N66001-12-C-4042, N66001-15-C-4017)
  • National Center for Advancing Translational Sciences (1ULTR001067)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
  • Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
Keywords
upper limb amputation, peripheral nerve stimulation, sensory feedback, multi-degree-of-freedom control, proprioception, skin sensation, median nerve, ulnar nerve, feedback control, phantom hand, neuroprosthetics, microstimulation, prosthetic hand control
MeSH terms
Adult, Amputees, Arm, Artificial Limbs, Electrodes, Implanted, Electric Stimulation Therapy, Humans, Male, Middle Aged, Proprioception, Prosthesis Design, Utah, Feedback, Sensory

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Restoration of motor control and proprioceptive and cutaneous sensation in humans with prior upper-limb amputation via multiple Utah Slanted Electrode Arrays (USEAs) implanted in residual peripheral arm nerves. https://app.cochat.ai/papers/10.1186/s12984-017-0320-4
The paper
Wendelken, S., Page, D. M., Davis, T., Wark, H. A. C., Kluger, D. T., Duncan, C., Warren, D. J., Hutchinson, D. T., & Clark, G. A. (2017). Restoration of motor control and proprioceptive and cutaneous sensation in humans with prior upper-limb amputation via multiple Utah Slanted Electrode Arrays (USEAs) implanted in residual peripheral arm nerves. Journal of NeuroEngineering and Rehabilitation. https://doi.org/10.1186/s12984-017-0320-4

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.