Evaluation of EMG pattern recognition for upper limb prosthesis control: a case study in comparison with direct myoelectric control
Open accessThe DOI matches its Crossref record; 49 of the 49 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Although electromyogram (EMG) pattern recognition (PR) for multifunctional upper limb prosthesis control has been reported for decades, the clinical benefits have rarely been examined. The study purposes were to: 1) compare self-report and performance outcomes of a transradial amputee immediately after training and one week after training of direct myoelectric control and EMG pattern recognition (PR) for a two-degree-of-freedom (DOF) prosthesis, and 2) examine the change in outcomes one week after pattern recognition training and the rate of skill acquisition in two subjects with transradial amputations. METHODS: In this cross-over study, participants were randomized to receive either PR control or direct control (DC) training of a 2 DOF myoelectric prosthesis first. Participants were 2 persons with traumatic transradial (TR) amputations who were 1 DOF myoelectric users. Outcomes, including measures of dexterity with and without cognitive load, activity performance, self-reported function, and prosthetic satisfaction were administered immediately and 1 week after training. Speed of skill acquisition was assessed hourly. One subject completed training under both PR control and DC conditions. Both subjects completed PR training and testing. Outcomes of test metrics were analyzed descriptively. RESULTS: Comparison of the two control strategies in one subject who completed training in both conditions showed better scores in 2 (18%) dexterity measures, 1 (50%) dexterity measure with cognitive load, and 1 (50%) self-report functional measure using DC, as compared to PR. Scores of all other metrics were comparable. Both subjects showed decline in dexterity after training. Findings related to rate of skill acquisition varied considerably by subject. CONCLUSIONS: Outcomes of PR and DC for operating a 2-DOF prosthesis in a single subject cross-over study were similar for 74% of metrics, and favored DC in 26% of metrics. The two subjects who completed PR training showed decline in dexterity one week after training ended. Findings related to rate of skill acquisition varied considerably by subject. This study, despite its small sample size, highlights a need for additional research quantifying the functional and clinical benefits of PR control for upper limb prostheses.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Motor Control and Learning: A Behavioral Emphasis
Schmidt, R. A. (1982) · Cited by 4,770
Estimating the Prevalence of Limb Loss in the United States: 2005 to 2050
Ziegler-Graham, K., MacKenzie, E. J., Ephraim, P. L., Travison, T. G., & Brookmeyer, R. (2008) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2007.11.005 · Cited by 2,816 · cited 1×
“Background In the United States, there are approximately 41,000 persons who have major upper limb amputations [1].”
BackgroundA new strategy for multifunction myoelectric control
Hudgins, B., Parker, P., & Scott, R. N. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.204774 · Cited by 2,149 · cited 1×
“The detailed calculation of these features can be found in a previous publication [32].”
BackgroundAdult Norms for the Box and Block Test of Manual Dexterity
Mathiowetz, V., Volland, G., Kashman, N., & Weber, K. (1985) · The American Journal of Occupational Therapy · doi:10.5014/ajot.39.6.386 · Cited by 2,105 · cited 1×
“Performance metrics included the Box and Block test [42], the modified Jebsen-Taylor”
MethodA robust, real-time control scheme for multifunction myoelectric control
Englehart, K., & Hudgins, B. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.813539 · Cited by 1,743
Motor Learning and Control : Concepts and Applications
Magill, R. A. (2020) · LSU Scholarly Repository (Louisiana State University) · Cited by 1,306
Electromyography: Physiology, Engineering, and Noninvasive Applications
(2004) · Wiley eBooks · doi:10.1002/0471678384 · Cited by 1,151
Upper limb prosthesis use and abandonment: A survey of the last 25 years
Biddiss, E. A., & Chau, T. T. (2007) · Prosthetics & Orthotics International · doi:10.1080/03093640600994581 · Cited by 1,111
Scheme, E., & Englehart, K. (2011) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2010.09.0177 · Cited by 969 · cited 1×
“The EMG PR was also evaluated while the users performed virtual tasks or selected physical tasks [22].”
MethodAssessing Disability and Change on Individual Patients: A Report of a Patient Specific Measure
Stratford, P. (1995) · Physiotherapy Canada · doi:10.3138/ptc.47.4.258 · Cited by 928 · cited 5×
“Further, the subject reported less difficulty in task performance (PSFS) when reporting on DC as compared to PR control (MDC 90 unknown).”
MethodInfluentialLiterature Review on Needs of Upper Limb Prosthesis Users
Cordella, F., Ciancio, A. L., Sacchetti, R., Davalli, A., Cutti, A. G., Guglielmelli, E., & Zollo, L. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00209 · Cited by 752
Kuiken, T. A., Dumanian, G. A., Lipschutz, R. D., Miller, L. A., & Stubblefield, K. A. (2004) · Prosthetics & Orthotics International · doi:10.3109/03093640409167756 · Cited by 583 · cited 2×
“reinnervation [20] or intramuscular EMG recordings [51]) might be combined with EMG PR to further improve intuitive prosthesis control.”
BackgroundMethodStudy of stability of time-domain features for electromyographic pattern recognition
Tkach, D., Huang, H., & Kuiken, T. A. (2010) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-7-21 · Cited by 582 · cited 1×
“This was because EMG PR lacked robustness against EMG variations caused by electrode shift, electrode-skin impedance change, muscle fatigue, or other environmental or physiological factors [33, 35], which sometimes led to decreased performance over time.”
BackgroundSelf-citationA Strategy for Identifying Locomotion Modes Using Surface Electromyography
Huang, H., Kuiken, T. A., & Lipschutz, R. D. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.2003293 · Cited by 512 · cited 1×
“Note that since EMG crosstalk has little influence on the performance of EMG PR [31], targeting specific muscles for EMG electrode placement was not necessary.”
BackgroundSelf-citationKuiken, T. A., Miller, L. A., Lipschutz, R. D., Lock, B. A., Stubblefield, K., Marasco, P. D., Zhou, P., & Dumanian, G. A. (2007) · The Lancet · doi:10.1016/s0140-6736(07)60193-7 · Cited by 482
Resolving the Limb Position Effect in Myoelectric Pattern Recognition
Fougner, A., Scheme, E., Chan, A. D. C., Englehart, K., & Stavdahl, Ø. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2163529 · Cited by 402 · cited 1×
“Different postures were tested because previous studies showed that EMG PR performance decreased with changes of upper limb posture [38].”
MethodZhai, X., Jelfs, B., Chan, R. H. M., & Tin, C. (2017) · Frontiers in Neuroscience · doi:10.3389/fnins.2017.00379 · Cited by 357
Design of a cybernetic hand for perception and action
Carrozza, M. C., Cappiello, G., Micera, S., Edin, B. B., Beccai, L., & Cipriani, C. (2006) · Biological Cybernetics · doi:10.1007/s00422-006-0124-2 · Cited by 341
Classification of Simultaneous Movements Using Surface EMG Pattern Recognition
Young, A. J., Smith, L. H., Rouse, E. J., & Hargrove, L. J. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2232293 · Cited by 320
Heinemann, A. W., Bode, R. K., & O'Reilly, C. (2003) · Prosthetics & Orthotics International · doi:10.1080/03093640308726682 · Cited by 314 · cited 3×
“Heinemann AW, Bode RK, O'Reilly C. Development and measurement properties of the orthotics and prosthetics Users’ survey (OPUS): a comprehensive set of clinical outcome instruments.”
MethodProsthesis use in persons with lower- and upper-limb amputation
Raichle, K. A. (2008) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2007.09.0151 · Cited by 288
Implantable Myoelectric Sensors (IMESs) for Intramuscular Electromyogram Recording
Weir, R. F., Troyk, P. R., DeMichele, G. A., Kerns, D. A., Schorsch, J. F., & Maas, H. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.2005942 · Cited by 266 · cited 1×
“reinnervation [20] or intramuscular EMG recordings [51]) might be combined with EMG PR to further improve intuitive prosthesis control.”
MethodHefford, C., Abbott, J. H., Arnold, R., & Baxter, G. D. (2012) · Journal of Orthopaedic & Sports Physical Therapy · doi:10.2519/jospt.2012.3953 · Cited by 241 · cited 5×
“Further, the subject reported less difficulty in task performance (PSFS) when reporting on DC as compared to PR control (MDC 90 unknown).”
MethodInfluentialSensinger, J. W., Lock, B. A., & Kuiken, T. A. (2009) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2023282 · Cited by 239 · cited 1×
“This was because EMG PR lacked robustness against EMG variations caused by electrode shift, electrode-skin impedance change, muscle fatigue, or other environmental or physiological factors [33, 35], which sometimes led to decreased performance over time.”
BackgroundIson, M., & Artemiadis, P. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/5/051001 · Cited by 225 · cited 1×
“Nevertheless, almost all commercial electric prostheses, including modern dexterous hands, still use a ‘direct myoelectric control’ approach, first offered commercially in the 1970s [13].”
BackgroundMyoelectric control of prostheses.
Parker, P. A., & Scott, R. N. (1986) · PubMed · Cited by 203
Hargrove, L. J., Miller, L. A., Turner, K., & Kuiken, T. A. (2017) · Scientific Reports · doi:10.1038/s41598-017-14386-w · Cited by 190 · cited 1×
“compared EMG PR and direct control of a 3 DOF prosthesis after 6–8 weeks of home use for transhumeral amputees who had targeted muscle reinnervation [52].”
BackgroundHuang, H., Zhou, P., Li, G., & Kuiken, T. A. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.910282 · Cited by 187 · cited 1×
“Finally, the feature vector in each window was fed into a linear discriminant analysis (LDA)-based classifier [37].”
MethodSelf-citationGeng, Y., Zhou, P., & Li, G. (2012) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-9-74 · Cited by 183
Vidovic, M. M.-C., Hwang, H.-J., Amsuss, S., Hahne, J. M., Farina, D., & Muller, K.-R. (2016) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2015.2492619 · Cited by 171
Gallagher, P., Franchignoni, F. P., Giordano, A., & MacLachlan, M. (2010) · American Journal of Physical Medicine & Rehabilitation · doi:10.1097/phm.0b013e3181dd8cf1 · Cited by 157 · cited 3×
“Self-report measures included the Upper Extremity Functional Scale (UEFS) from the Orthotics and Prosthetics Users Survey (OPUS) [46, 47], the Patient Specific Functional Scale (PSFS) [48, 49], which has been used in the VA studies of the DEKA Arm, and the satisfaction scale from the Trinity Amputations and Prosthetic…”
MethodDevelopment and Evaluation of the Activities Measure for Upper Limb Amputees
Resnik, L. J., Adams, L., Borgia, M. L., Delikat, J., Disla, R., Ebner, C., & Walters, L. S. (2012) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2012.10.004 · Cited by 155 · cited 6×
“Hand Function (JTHF) test [43], the Activities Measure for Upper Limb Amputees (AM-ULA) [44], the Clothes Pin Relocation Task [45], and the Clothes Pin Relocation Task with dual-task performance where the subject repeats the clothes pin task while simultaneously performing a cognitively demanding task: 1) counting bac…”
MethodInfluentialSelf-citationKhushaba, R. N., Al-Timemy, A. H., Al-Ani, A., & Al-Jumaily, A. (2017) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2017.2687520 · Cited by 144
Kuiken, T. A., Miller, L. A., Turner, K., & Hargrove, L. J. (2016) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2016.2616123 · Cited by 135 · cited 5×
“compared control of a 3 DOF prosthesis (hand open/close, wrist pronation/supination and wrist flexion/extension) controlled by EMG PR and EMG DC in the laboratory, and their study included 4 weeks of use in a home setting [30].”
BackgroundMethodResultInfluentialDesign and evaluation of a low-cost instrumented glove for hand function assessment
Oess, N. P., Wanek, J., & Curt, A. E. P. (2012) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-9-2 · Cited by 122
The design of anthropomorphic prosthetic hands: A study of the Southampton Hand
Kyberd, P., Light, C. M., Chappell, P. H., Nightingale, J. M., Whatley, D. R., & Evans, M. (2001) · Robotica · doi:10.1017/s0263574701003538 · Cited by 114
Burger, H., Franchignoni, F. P., Heinemann, A. W., Kotnik, S., & Giordano, A. (2008) · Journal of Rehabilitation Medicine · doi:10.2340/16501977-0183 · Cited by 99
Orthotics and Prosthetics in Rehabilitation
Nielsen, C. C., & Jorge, M. (2013) · Elsevier eBooks · doi:10.1016/b978-1-4377-1936-9.00001-1 · Cited by 93
Powered upper limb prostheses : control, implementation and clinical application
Muzumdar, A. (2004) · Springer eBooks · Cited by 87
A real-time, practical sensor fault-tolerant module for robust EMG pattern recognition
Zhang, X., & Huang, H. (2015) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-015-0011-y · Cited by 77 · cited 3×
“ments OMAP3503 600 MHz processor based on the ARM Cortex-A8 architecture) [25].”
MethodInfluentialSelf-citationPractical Methods for Controlling Powered Upper-Extremity Prostheses
Williams, T. W. (1990) · Assistive Technology · doi:10.1080/10400435.1990.10132142 · Cited by 71 · cited 1×
“The motor speed is proportional to the magnitude of the EMG signal [14, 15].”
BackgroundDevelopment and testing of new upper-limb prosthetic devices: Research designs for usability testing
Resnik, L. J. (2011) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2010.03.0050 · Cited by 65 · cited 1×
“The DEKA arm, an advanced prosthetic arm recently approved by the FDA, has a fully powered 3 degree-offreedom (DOF) shoulder, elbow, 2-DOF wrist, and 6 hand gripping patterns [12].”
BackgroundSelf-citationReliability and Validity of Outcome Measures for Upper Limb Amputation
Resnik, L. J., & Borgia, M. L. (2012) · JPO Journal of Prosthetics and Orthotics · doi:10.1097/jpo.0b013e31826ff91c · Cited by 63 · cited 1×
“Once training with the actual prosthesis commenced, subjects were administered the Jebsen Taylor Hand Function (JTHF) page- turning test [39] at 1 h intervals to assess the ease of learning and speed of mastery of the new EMG PR control system as compared to EMG DC.”
MethodSelf-citationHuang, Q., Yang, D., Jiang, L., Zhang, H., Liu, H., & Kotani, K. (2017) · Sensors · doi:10.3390/s17061370 · Cited by 57
Reliability, Validity, and Responsiveness of the QuickDASH in Patients With Upper Limb Amputation
Resnik, L. J., & Borgia, M. L. (2015) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2015.03.023 · Cited by 49 · cited 1×
“Hand Function (JTHF) test [43], the Activities Measure for Upper Limb Amputees (AM-ULA) [44], the Clothes Pin Relocation Task [45], and the Clothes Pin Relocation Task with dual-task performance where the subject repeats the clothes pin task while simultaneously performing a cognitively demanding task: 1) counting bac…”
MethodSelf-citationClassification complexity in myoelectric pattern recognition
Nilsson, N., Håkansson, B. E. V., & Ortiz-Catalan, M. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0283-5 · Cited by 39
Resolving the effect of wrist position on myoelectric pattern recognition control
Adewuyi, A. A., Hargrove, L. J., & Kuiken, T. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0246-x · Cited by 33
Zaki, G., Plishker, W. L., Li, W., Lee, J., Quon, H., Wong, J., & Shekhar, R. (2016) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2016.2597838 · Cited by 22
Graupe, D., Aa, B., Wj, M., & Magnussen, I. (1977) · PubMed · Cited by 20
Metadata
- Authors
- Resnik, L., Huang, H. (., Winslow, A., Crouch, D. L., Zhang, F., & Wolk, N.
- Venue
- Journal of NeuroEngineering and Rehabilitation
- Published
- 2018-03-15
- Volume
- 15
- Issue
- 1
- Pages
- 23-23
- ISSN
- 1743-0003 (ISSN-L)
- OpenAlex
- W2801921491
- Cited by
- 174
- References
- 49
- FWCI
- 6.42
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.