An Evaluation of a Zero-Heat-Flux Cutaneous Thermometer in Cardiac Surgical Patients
The DOI matches its Crossref record; 40 of the 40 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Although core temperature can be measured invasively, there are currently no widely available, reliable, noninvasive thermometers for its measurement. We thus compared a prototype zero-heat-flux thermometer with simultaneous measurements from a pulmonary artery catheter. Specifically, we tested the hypothesis that zero-heat-flux temperatures are sufficiently accurate for routine clinical use. METHODS: Core temperature was measured from the thermistor of a standard pulmonary artery catheter and with a prototype zero-heat-flux deep-tissue thermometer in 105 patients having nonemergent cardiac surgery. Zero-heat-flux probes were positioned on the lateral forehead and lateral neck. Skin surface temperature probes were attached to the forehead just adjacent to the zero-heat-flux probe. Temperatures were recorded at 1-minute intervals, excluding the period of cardiopulmonary bypass, and for the first 4 postoperative hours. Zero-heat-flux and pulmonary artery temperatures were compared with bias analysis; differences exceeding 0.5°C were considered to be potentially clinically important. RESULTS: The mean duration in the operating room was 279 ± 75 minutes, and the mean cross-clamp time was 118 ± 50 minutes. All subjects were monitored for an additional 4 hours in the intensive care unit. The average overall difference between forehead zero-heat-flux and pulmonary artery temperatures (i.e., forehead minus pulmonary artery) was -0.23°C (95% limits of agreement of ±0.82); 78% of the differences were ≤0.5°C. The average intraoperative temperature difference was -0.08°C (95% limits of agreement of ±0.88); 84% of the differences were ≤0.5°C. The average postoperative difference was -0.32°C (95% limits of agreement of ±0.75); 84% of the differences were ≤0.5°C. Bias and precision values for neck site were similar to the forehead values. Uncorrected forehead skin temperature showed an increasing negative bias as core temperature decreased. CONCLUSIONS: Core temperature can be noninvasively measured using the zero-heat-flux method. Bias was small, but precision was slightly worse than our designated 0.5°C limits compared with measurements from a pulmonary artery catheter.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
Kurz, A., Sessler, D. I., & Lenhardt, R. (1996) · New England Journal of Medicine · doi:10.1056/nejm199605093341901 · Cited by 2,760
Self-citationAgreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 2,065
Frank, S. M. (1997) · JAMA · doi:10.1001/jama.1997.03540380041029 · Cited by 1,261
The Effects of Mild Perioperative Hypothermia on Blood Loss and Transfusion Requirement
Rajagopalan, S., Mascha, E., Na, J., & Sessler, D. I. (2008) · Anesthesiology · doi:10.1097/01.anes.0000296719.73450.52 · Cited by 905
Self-citationTemperature Monitoring and Perioperative Thermoregulation
Sessler, D. I., Warner, D. S., & Warner, M. A. (2008) · Anesthesiology · doi:10.1097/aln.0b013e31817f6d76 · Cited by 758
Self-citationMelling, A. C., Ali, B., Scott, E. M., & Leaper, D. J. (2001) · The Lancet · doi:10.1016/s0140-6736(01)06071-8 · Cited by 714
Mild Intraoperative Hypothermia Prolongs Postanesthetic Recovery
Lenhardt, R., Marker, E., Goll, V., Tschernich, H., Kurz, A., Sessler, D. I., Narzt, E., & Lackner, F. (1997) · Anesthesiology · doi:10.1097/00000542-199712000-00009 · Cited by 560
Self-citation
Cited by
Mäkinen, M.-T., Pesonen, A., Jousela, I., Päivärinta, J., Poikajärvi, S., Albäck, A., Salminen, U.-S., & Pesonen, E. (2016) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2016.03.141 · Cited by 69
“471C, 95% limits of agreement; see Fig 4) was approximately as good as the agreement with the esophageal temperature in vascular surgery, but it was better than the recently reported corresponding agreement of SpotOn temperature with PA temperature for which a prototype of the SpotOn system was evaluated.(30) The resu…”
Evaluation of the Temple Touch Pro, a Novel Noninvasive Core-Temperature Monitoring System
Evron, S., Weissman, A., Toivis, V., Shahaf, D. B., You, J., Sessler, D. I., & Ezri, T. (2017) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001695 · Cited by 38
Self-citationAccuracy of Zero-Heat-Flux Cutaneous Temperature in Intensive Care Adults
Dahyot-Fizelier, C., Lamarche, S., Kerforne, T., Bénard, T., Giraud, B., Bellier, R., Carise, E., Frasca, D., & Mimoz, O. (2017) · Critical Care Medicine · doi:10.1097/ccm.0000000000002317 · Cited by 34
Metadata
- Authors
- Eshraghi, Y., Nasr, V., Parra-Sanchez, I., Van Duren, A., Botham, M., Santoscoy, T., & Sessler, D. I.
- Venue
- Anesthesia & Analgesia
- Published
- 2014-09
- Volume
- 119
- Issue
- 3
- Pages
- 543-549
- ISSN
- 0003-2999 (ISSN-L)
- OpenAlex
- W2028080150
- Cited by
- 161
- References
- 40
- FWCI
- 8.66
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: An Evaluation of a Zero-Heat-Flux Cutaneous Thermometer in Cardiac Surgical Patients. https://app.cochat.ai/papers/10.1213/ane.0000000000000319
- The paper
- Eshraghi, Y., Nasr, V., Parra-Sanchez, I., Van Duren, A., Botham, M., Santoscoy, T., & Sessler, D. I. (2014). An Evaluation of a Zero-Heat-Flux Cutaneous Thermometer in Cardiac Surgical Patients. Anesthesia & Analgesia. https://doi.org/10.1213/ane.0000000000000319
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.