Validity of (ultra-)short recordings for heart rate variability measurements.
Open accessThe DOI matches its Crossref record; 26 of the 27 references it cites were found in a registry or an index.
Abstract
OBJECTIVES: In order to investigate the applicability of routine 10s electrocardiogram (ECG) recordings for time-domain heart rate variability (HRV) calculation we explored to what extent these (ultra-)short recordings capture the "actual" HRV. METHODS: The standard deviation of normal-to-normal intervals (SDNN) and the root mean square of successive differences (RMSSD) were measured in 3,387 adults. SDNN and RMSSD were assessed from (ultra)short recordings of 10s(3x), 30s, and 120s and compared to 240s-300s (gold standard) measurements. Pearson's correlation coefficients (r), Bland-Altman 95% limits of agreement and Cohen's d statistics were used as agreement analysis techniques. RESULTS: Agreement between the separate 10s recordings and the 240s-300s recording was already substantial (r = 0.758-0.764/Bias = 0.398-0.416/d = 0.855-0.894 for SDNN; r = 0.853-0.862/Bias = 0.079-0.096/d = 0.150-0.171 for RMSSD), and improved further when three 10s periods were averaged (r = 0.863/Bias = 0.406/d = 0.874 for SDNN; r = 0.941/Bias = 0.088/d = 0.167 for RMSSD). Agreement increased with recording length and reached near perfect agreement at 120s (r = 0.956/Bias = 0.064/d = 0.137 for SDNN; r = 0.986/Bias = 0.014/d = 0.027 for RMSSD). For all recording lengths and agreement measures, RMSSD outperformed SDNN. CONCLUSIONS: Our results confirm that it is unnecessary to use recordings longer than 120s to obtain accurate measures of RMSSD and SDNN in the time domain. Even a single 10s (standard ECG) recording yields a valid RMSSD measurement, although an average over multiple 10s ECGs is preferable. For SDNN we would recommend either 30s or multiple 10s ECGs. Future research projects using time-domain HRV parameters, e.g. genetic epidemiological studies, could calculate HRV from (ultra-)short ECGs enabling such projects to be performed at a large scale.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Fig 2 Biases and 95% LoAs milliseconds(ms) of the log-transformed (a) SDNN [in blue] and (b) RMSSD [in red] measured from the recordings with different time lengths compared to the total recording. The diamonds are the observed bias and the lines represent the intervals defined by the observed 95%LoAs. The bands show the 95% reference ranges of the simulated biases and the dark gray bands show the 95% reference ranges of the upper and lower 95%LoAs from the simulation. The dotted lines represent the intervals defined by the simulated 95%LoAs. For both SDNN and RMSSD for all recording lengths the observed biases did not differ from the expectation, but the observed 95%LoAs were much narrower than expected.](https://figures.cochat.ai/paper-figures/9165b3621f0449e2b57f1940787871b9/validity-of-ultra-short-recordings-for-heart-rate-variability-measurements-figure-2.jpg)



Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Those 10 s periods don’t need to be contiguous (i.e., not in succession), so it is possible to obtain a good estimate from several 10 s measurements spread over a trial or the experiment.”
From Heart Rate Variability and Cardiac Vagal Tone in Psychophysiological Research – Recommendations for Experiment Planning, Data Analysis, and Data Reporting (2017) · cites this paper as Munoz et al., 2015
The cited source says what the claim attributes to it.
In the citing paper
“In addition, even a single 10 s (standard ECG) was found to yield a valid RMSSD measurement, although an average over multiple 10 s ECGs is preferable.”
From Heart Rate Variability and Cardiac Vagal Tone in Psychophysiological Research – Recommendations for Experiment Planning, Data Analysis, and Data Reporting (2017) · cites this paper as Munoz et al., 2015
The cited source says what the claim attributes to it.
In the citing paper
“Very recent work by Munoz et al. (2015) investigated the validity of ultra-short and short recordings on HRV measurements in a very large adult sample (N = 3,387). They found that it is unnecessary to use recordings longer than 120 s to obtain accurate measures of RMSSD.”
From Heart Rate Variability and Cardiac Vagal Tone in Psychophysiological Research – Recommendations for Experiment Planning, Data Analysis, and Data Reporting (2017) · cites this paper as Munoz et al. (2015)
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Statistical Power Analysis for the Behavioral Sciences
Muller, K. (1989) · Technometrics · doi:10.1080/00401706.1989.10488618 · Cited by 99,167
Statistical power analysis for the behavioral sciences
(1990) · Computers, Environment and Urban Systems · doi:10.1016/0198-9715(90)90050-4 · Cited by 65,269
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Heart rate variability: A review
Rajendra Acharya, U., Paul Joseph, K., Kannathal, N., Lim, C. M., & Suri, J. S. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0119-0 · Cited by 2,820 · cited 2×
“components [1–3].”
BackgroundApplying the right statistics: analyses of measurement studies
Bland, J. M., & Altman, D. G. (2003) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.122 · Cited by 1,311 · cited 18×
“However, the variation in those paired differences between the observed measurements of the (ultra-)short segments and those of the total recording is smaller than from the respective paired differences of the simulated measurements, explaining the much higher correlation and tighter 95%LoAs.”
BackgroundMethodInfluentialMalik, M. (1996) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.1996.tb00275.x · Cited by 850
Heart Rate Variability ? A Historical Perspective
Billman, G. E. (2011) · Frontiers in Physiology · doi:10.3389/fphys.2011.00086 · Cited by 786 · cited 1×
“Both are easy to calculate and among the most widely used indices of HRV [7].”
How are we doing in soft psychology?
Rosenthal, R. (1990) · American Psychologist · doi:10.1037//0003-066x.45.6.775 · Cited by 515 · cited 1×
“80 is a large difference) [14,20].”
Dekker, J. M., Schouten, E. G., Klootwijk, P., Pool, J., Swenne, C. A., & Kromhout, D. (1997) · American Journal of Epidemiology · doi:10.1093/oxfordjournals.aje.a009049 · Cited by 512
Cited by
Items 401–422 of 422
Shah, S. T. A., Pereira, A. B., Santos, J. P., Fernandes, J. M., & Constantinescu, G. (2026) · International Conference in Electronic Engineering & Information Technology (EEITE) · doi:10.1109/eeite69609.2026.11679620
Multimodal Physiological Assessment of Attentional States and Driving Styles in Simulated Driving
Zhu, W., & Lin, Y. (2026) · Sensors · doi:10.3390/s26185883
“This study reported that the neural precursors of mind wandering have been reported to emerge 4–10 min before a subjective report [23].”
Candela-Leal, M. O., Calderón-Gurubel, J. E., Ortiz, A., Elizondo-García, J., Vélez-Saboyá, C. S., Lozoya-Santos, J. D. J., Rodríguez-Sánchez, I., Cebral-Loureda, M., & Ramírez-Moreno, M. A. (2026) · Frontiers in Systems Neuroscience · doi:10.3389/fnsys.2026.1908489
“…HRV estimates derived from 90 s segments should not be regarded as interchangeable with those obtained from conventional 5 min recordings, particularly for metrics that require longer observation periods (Shaffer and Ginsberg, 2017; Baek et al., 2015; Munoz et al., 2015; Shaffer et al., 2020).”
Choi, I., Čukić, M., & Choi, Y. (2026) · Research Square · doi:10.21203/rs.3.rs-11054648/v1
Ali, S. A., Demeter, D. V., Baim, A. R., Ahern, J., Ahn, J. H. B., Chang, S. E., Koithan, E. M., Park, S., Smarr, B. L., & Greene, D. J. (2026) · Frontiers in Network Physiology · doi:10.3389/fnetp.2026.1881914
“No additional interpolation-based ectopic-beat correction was applied to IBIs falling within these physiological bounds.”
InfluentialDevelopment of a Low-Cost Portable ECG Device for Predicting Cardiovascular Diseases
Hatratchai, N., Sasithong, P., & Swatdipisal, K. (2026) · International Symposium on Communications and Information Technologies · doi:10.1109/iscit70453.2026.11695330
“Finally, a 30-second segment is extracted from the middle of each record as the input to feature extraction; this duration is consistent with short-term HRV standards [12] and supports reliable RMSSD and SDNN computation [13].”
Izhaki, N., Perek, S., Safadi, S., & Raz‐Pasteur, A. (2026) · Medicina · doi:10.3390/medicina62101845
“Assessment of usHRV time-domain indices revealed that even a single 10 s RMSSD measurement provides valid and accurate HRV parameters [20].”
Qu, Q., Tang, Y., Guo, F., Duffy, V. G., & Man, S. S. (2026) · Safety Science · doi:10.1016/j.ssci.2026.107471
Cabrera-Ontiveros, T. I., García-Cárdenas, D. E., Ramírez, C. I. L., Abarca-Castro, E. A., Talavera-Peña, A. K., & Lagos, J. J. R. (2026) · Cognitive Processing · doi:10.1007/s10339-026-01392-y
“Therefore, this study mainly focused on these indices when analyzing ultra-short HRV data, while frequency-domain measures were explored more tentatively due to the shorter window length (Munoz et al. 2015).”
Ryu, G., Choi, J. M., Choi, B., & Shin, H. (2026) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-026-01506-z
The Associations of Heart Rate Variability With Inflammatory Biomarkers
Moloney, B. M., Correa, S., Song, E. H., Chertow, G. M., & Causland, F. R. M. (2026) · Kidney International Reports · doi:10.1016/j.ekir.2026.107133
Zordan, M., Lei, Z., & Pei, J. (2026) · Developments in the Built Environment · doi:10.1016/j.dibe.2026.101070
Estimating heart rate variability using facial video photoplethysmography: a pilot validation study
Pstras, L., Okupnik, T., Ponikowska, B., & Paleczny, B. (2026) · Scientific Reports · doi:10.1038/s41598-026-70292-0
“For RMSSD, it has been shown that even measurements shorter than 60 s should be sufficient to obtain results similar to standard 5-min measurements 35–37 , whereas for SDNN the recommendations vary from 30 s 35 , through 60 s 38 , up to 4 min 36 .”
Jiang, W., Sun, W., Li, C., & Li, S. (2026) · Expert Systems with Applications · doi:10.1016/j.eswa.2026.134727
Gupta, Y., Suttkus, S., Altmann, U., de la Cruz, F., Geisler, M., Schumann, A., & Bär, K. (2026) · Psychophysiology · doi:10.1111/psyp.70427
“…analyses commonly use longer recording segments, 60 s windows have been widely adopted in dynamic and ultra- short-term HRV analyses as a practical compromise between estimation stability and temporal resolution (Munoz et al. 2015; Shaffer and Ginsberg 2017; Laborde et al. 2017; Burma et al. 2021).”
Application of an Alert-Based Data Quality Model to ECG Signals in the MIMIC-IV Database
Gyulgyulyan, E., Koteska, B., Mishev, A., & Astsatryan, H. V. (2026) · International Conference on Informatics and Information Technologies (CIIT) · doi:10.1109/ciit68512.2026.11731574
“This window-based approach is supported by previous studies showing that short ECG segments can provide reliable HR estimates for the short window of the signal [16].”
Atmojo, U. D., Kolesnikov, M. V., Vyatkin, V. V., & Chakraborty, A. (2026) · IEEE International Conference on Emerging Technologies and Factory Automation (ETFA) · doi:10.1109/etfa68377.2026.11725088
“This calculation is performed using the last 10 seconds of RR interval data to ensure a more stable RMSSD value, which has been shown to be valid by academic research [13].”
Kafaei, M., Burry, J., Latifi, M., Ciorciari, J., Aminitabar, A., & Gao, N. (2026) · Indoor Air · doi:10.1155/ina/7968077
Mousavi, M. A., Hendrickson, A. T., Powell, W. A., & Louwerse, M. M. (2026) · Cognitive Systems Research · doi:10.1016/j.cogsys.2026.101524
Discriminative Capability of Classical HRV Features for Ultra-Short-ECG Arrhythmia Classification
Zavaleta-Cavero, J. A., Flores-Perez, M. A., Espinoza-Quispe, A. L., Moura, J. M. D., & Bazán-Lavanda, N. B. (2026) · International Conference on Electrical Engineering, Computing Science and Automatic Control (CCE) · doi:10.1109/cce72160.2026.11734966
“…these results indicate that classical time-domain HRV descriptors preserve relatively stable marginal distributions across independent databases, consistent with previous studies demonstrating the robustness of linear HRV indices for arrhythmia classification using PhysioNet databases [20]–[22].”
Pretty, E. J., Guarese, R., Hamari, J., Fayek, H. M., & Zambetta, F. (2026) · Annual Symposium on Computer-Human Interaction in Play · doi:10.1145/3800965.3834327
Kling, J., Ekblom, Ö. B., Asplund, R. P., Hallman, D., Jiang, Y., & Blom, V. (2026) · European Journal of Applied Physiology · doi:10.1007/s00421-026-06467-4
Metadata
- Authors
- Munoz, M. L., van Roon, A., Riese, H., Thio, C., Oostenbroek, E., Westrik, I., de Geus, E. J. C., Gansevoort, R., Lefrandt, J., Nolte, I. M., & Snieder, H.
- Affiliations
- University Medical Center Groningen (NL)
- Corresponding author
- Nolte, Ilja M.
- Snieder, Harold
- ORCID
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2015-09-28
- Volume
- 10
- Issue
- 9
- Pages
- e0138921
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4586373
- OpenAlex
- W2298185395
- Cited by
- 423
- References
- 27
- Citations by year
- 2026: 80 · 2025: 55 · 2024: 47 · 2023: 33 · 2022: 45 · 2021: 51 · 2020: 32 · 2019: 26 · 2018: 26 · 2017: 17 · 2016: 8
- FWCI
- 12.36
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- University of Groningen research database (University of Groningen / Centre for Information Technology) (publishedVersion, cc-by)
- VU Research Portal (publishedVersion, other-oa)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- University of Groningen research database (University of Groningen / Centre for Information Technology) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- heart rate variability, time-domain HRV, SDNN, RMSSD, measurement agreement, method validation, autonomic nervous system, adults, limits of agreement
- MeSH terms
- Adult, Demography, Electrocardiography, Female, Heart Rate, Humans, Male, Middle Aged, Statistics as Topic, Time Factors, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.