CoChat

Multicenter performance evaluation of a second generation cortisol assay

Open access

The DOI matches its Crossref record; 34 of the 34 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Untreated disorders of the adrenocortical system, such as Cushing's or Addison's disease, can be fatal, and accurate quantification of a patient's cortisol levels is vital for diagnosis. The objective of this study was to assess the analytical performance of a new fully-automated Elecsys® Cortisol II assay (second generation) to measure cortisol levels in serum and saliva. METHODS: Four European investigational sites assessed the intermediate precision and reproducibility of the Cortisol II assay (Roche Diagnostics) under routine conditions. Method comparisons of the Cortisol II assay vs. liquid chromatography-tandem mass spectrometry (LC-MS/MS), the gold standard for cortisol measurement, were performed. Cortisol reference ranges from three US sites were determined using samples from self-reported healthy individuals. RESULTS: The coefficients of variation (CVs) for repeatability, intermediate precision, and reproducibility for serum samples were ≤2.6%, ≤5.8%, and ≤9.5%, respectively, and for saliva were ≤4.4% and ≤10.9%, and ≤11.4%, respectively. Agreement between the Cortisol II assay and LC-MS/MS in serum samples was close, with a slope of 1.02 and an intercept of 4.473 nmol/L. Reference range samples were collected from healthy individuals (n=300) and serum morning cortisol concentrations (5-95th percentile) were 166.1-507 nmol/L and afternoon concentrations were 73.8-291 nmol/L. Morning, afternoon, and midnight saliva concentrations (95th percentile) were 20.3, 6.94, and 7.56 nmol/L, respectively. CONCLUSIONS: The Cortisol II assay had good precision over the entire measuring range and had excellent agreement with LC-MS/MS. This test was found suitable for routine diagnostic application and will be valuable for the diagnosis of adrenocortical diseases.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1515/cclm-2016-0400

Cites

34 of 34 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 Method comparison results and Bland-Altman plots of the Cortisol assay vs. the Cortisol II assay in serum or plasma samples (A) and (B), and saliva samples (C) and (D).
Figure 1 Method comparison results and Bland-Altman plots of the Cortisol assay vs. the Cortisol II assay in serum or plasma samples (A) and (B), and saliva samples (C) and (D).CC-BY-NC-ND · From the PDF
Figure 2 Method comparison results and Bland-Altman plots of the Cortisol II assay vs. LC-MS/MS in serum or plasma samples (A) and (B) and saliva samples (C) and (D).
Figure 2 Method comparison results and Bland-Altman plots of the Cortisol II assay vs. LC-MS/MS in serum or plasma samples (A) and (B) and saliva samples (C) and (D).CC-BY-NC-ND · From the PDF

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 34 references

Research works printed but not linked: 40

  1. 2005, Wincaev – a new program supporting evaluations of reagents and analysers, Clin Chim Acta, 355, S361
  2. 32. Fiers T, Delanghe J, T'Sjoen G, Van Caenegem E, Wierckx K, Kaufman JM. A critical evaluation of salivary testosterone as a method for the assessment of serum testosterone. Steroids 2014;86:5-9. · DOI
  3. 16. Leung KS, Fong BM. LC-MS/MS in the routine clinical laboratory: has its time come? Anal Bioanal Chem 2014;406:2289-301. · DOI
  4. 12. Briegel J, Sprung CL, Annane D, Singer M, Keh D, Moreno R, et al. Multicenter comparison of cortisol as measured by different methods in samples of patients with septic shock. Intensive Care Med 2009;35:2151-6. · DOI · Paper page
  5. 2000, Clinical diagnosis of hyper- and hypocortisolism, Noise Health, 2, 39
  6. 36. Ju Bae Y, Gaudl A, Jaeger S, Stadelmann S, Hiemisch A, Kiess W, et al. Immunoassay or LC-MS/MS for the measurement of salivary cortisol in children? Clin Chem Lab Med 2016;54:811-22.
  7. 2007, Saliva collection method affects predictability of serum cortisol, Clin Chim Acta, 382, 15, 10.1016/j.cca.2007.03.009 · DOI · Paper page
  8. 24. Chiu SK, Collier CP, Clark AF, Wynn-Edwards KE. Salivary cortisol on ROCHE Elecsys immunoassay system: pilot biological variation studies. Clin Biochem 2003;36:211-4. · DOI
  9. 1999, Is the prevalence of Addison’s disease underestimated?, J Clin Endocrinol Metab, 84, 1762
  10. 2014, Cross-reactivity of steroid hormone immunoassays: clinical significance and two-dimensional molecular similarity prediction, BMC Clin Pathol, 14, 33, 10.1186/1472-6890-14-33 · DOI · Paper page
  11. 35. Gaudl A, Kratzsch J, Bae YJ, Kiess W, Thiery J, Ceglarek U. Liquid chromatography quadrupole linear ion trap mass spectrometry for quantitative steroid hormone analysis in plasma, urine, saliva and hair. J Chrom A 2016;1464:64-71. · DOI
  12. 2016, Mass spectrometry methods in clinical diagnostics – state of the art and perspectives, Trends Analyt Chem, 84, 1, 10.1016/j.trac.2016.04.018 · DOI
  13. 2013, Method-specific serum cortisol responses to the adrenocorticotrophin test: comparison of gas chromatography-mass spectrometry and five automated immunoassays, Clin Endocrinol (Oxf), 78, 673, 10.1111/cen.12039 · DOI · Paper page
  14. 28. Thienpont LM, De Leenheer AP, Siekmann L, Kristiansen N, Linsinger T, Schimmel H, et al. The characterisation of cortisol concentrations in a serum reference panel (IRMM/IFCC No. 451). EUR 19007 EN. IRMM/IFCC Information, Reference Materials. Brussels, Belgium. 1999.
  15. 33. Janssens H, Clays E, Fiers T, Verstraete AG, de Bacquer D, ­ Braeckman L. Hair cortisol in relation to job stress and depressive symptoms. Occup Med (Lond). 2016; pii: kqw114. [Epub ahead of print].
  16. 38. Clinical and Laboratory Standards Institute. Mass spectrometry for androgen and estrogen measurements in serum, 1st ed. C57 2015;35:4. · DOI
  17. 39. Clinical and Laboratory Standards Institute. Liquid chromatography-mass spectrometry methods; Approved Guideline. C62-A 2014;34:16.
  18. 41. Official Journal of the European Communities. Commission Decision of 12 August 2002; Implementing Council Directive 96/23/ EC concerning the performance of analytical methods and the interpretation of results. Available at: http:/ /eur-lex.europa.eu/ legal-content/EN/TXT/PDF/?uri=CELEX:32002D0657&from=EN. Accessed: 23 June 2015.
  19. 2009, Utility of salivary cortisol measurements in Cushing’s syndrome and adrenal insufficiency, J Clin Endocrinol Metab, 94, 3647, 10.1210/jc.2009-1166 · DOI · Paper page
  20. 2004, Midnight salivary cortisol for the initial diagnosis of Cushing’s syndrome of various causes, J Clin Endocrinol Metab, 89, 3345, 10.1210/jc.2003-031790 · DOI · Paper page
  21. 2009, Cortisol assays and diagnostic laboratory procedures in human biological fluids, Clin Biochem, 42, 1205, 10.1016/j.clinbiochem.2009.04.011 · DOI · Paper page
  22. 29. Garde AH, Hansen AM. Long-term stability of salivary cortisol. Scand J Clin Lab Invest 2005;65:433-6. · DOI · Paper page
  23. 30. Clinical and Laboratory Standards Institute. Interference testing in clinical chemistry; approved guideline - second edition. EP07-A2 2005;25:27.
  24. 31. Clinical and Laboratory Standards Institute. Evaluation of precision of quantitative measurement methods: approved guideline - third edition. EP05-A3 2014;34:13.
  25. 1994, Eighty-six cases of Addison’s disease, Clin Endocrinol (Oxf), 41, 757, 10.1111/j.1365-2265.1994.tb02790.x · DOI · Paper page
  26. 2004, Interferences in hormone immunoassays, Clin Lab Med, 24, 1, 10.1016/j.cll.2004.01.003 · DOI
  27. 2013, Determination of cortisol in serum, saliva and urine, Best Pract Res Clin Endocrinol Metab, 27, 795, 10.1016/j.beem.2013.10.008 · DOI · Paper page
  28. 43. Kunst A, Busse Grawitz A, Engeldinger W, Koch W, Luthe H, Stockmann W. Wincaev - a new program supporting evaluations of reagents and analysers. Clin Chim Acta 2005;355:S361.
  29. 44. Clinical and Laboratory Standards Institute. Measurement procedure comparison and bias estimation using patient samples; approved guideline - third edition. EP09-A3 2013 33:11.
  30. 2009, Rapid quantification of steroid patterns in human serum by on-line solid phase extraction combined with liquid chromatography-triple quadrupole linear ion trap mass spectrometry, Clin Chim Acta, 401, 114, 10.1016/j.cca.2008.11.022 · DOI
  31. 2003, Automated measurement of salivary cortisol, Clin Chem, 49, 1408, 10.1373/49.8.1408 · DOI · Paper page
  32. 37. Suhr AC, Vogeser M, Grimm SH. Isotope inversion experiment evaluating the suitability of calibration in surrogate matrix for quantification via LC-MS/MS-Exemplary application for a steroid multi-method. J Pharm Biomed Anal 2016;124:309-18. · DOI
  33. 45. Clinical and Laboratory Standards Institute. Evaluation of detection capability for clinical laboratory measurement procedures; approved guideline - second edition. EP17-A2 2012;32:8. · DOI · Paper page
  34. 1999, The characterisation of cortisol concentrations in a serum reference panel (IRMM/IFCC No. 451). EUR 19007 EN
  35. 46. Guignat L, Bertherat J. The diagnosis of Cushing's syndrome: an Endocrine Society Clinical Practice Guideline: commentary from a European perspective. Eur J Endocrinol 2010;163:9-13. · DOI · Paper page
  36. 47. UK NEQAS for Steroid Hormones. Review for UK NEQAS Clinical Chemistry Specialist Advisory Group for Endocrinology and Immunoassay 2011. Available at: http:/ /www.birminghamquality.org.uk/%5b1%5dSteroidAnnualReview\_2011.pdf. Accessed: 26 Apr 2016.
  37. 1983, Salivary cortisol: a better measure of adrenal cortical function than serum cortisol, Ann Clin Biochem, 20, 329, 10.1177/000456328302000601 · DOI · Paper page
  38. 2011, LC-MS/MS in the clinical laboratory – where to from here?, Clin Biochem Rev, 32, 5
  39. 2016, Clinical mass spectrometry-achieving prominence in laboratory medicine, Clin Chem, 62, 1, 10.1373/clinchem.2015.251272 · DOI
  40. 2002, Nighttime salivary cortisol: a useful test for the diagnosis of Cushing’s syndrome, J Clin Endocrinol Metab, 87, 4515, 10.1210/jc.2002-020534 · DOI · Paper page

Cited by

Metadata

Authors
Vogeser, M., Kratzsch, J., Ju Bae, Y., Bruegel, M., Ceglarek, U., Fiers, T., Gaudl, A., Kurka, H., Milczynski, C., Prat Knoll, C., Suhr, A. C., Teupser, D., Zahn, I., & Ostlund, R. E.
Affiliations
  • LMU Klinikum (DE)
  • University Hospital Leipzig (DE)
  • Institut für Laboratoriumsmedizin (DE)
  • Ghent University Hospital (BE)
  • Washington University in St. Louis (US)
Corresponding author
Vogeser, Michael
Venue
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2017-06-01
Volume
55
Issue
6
Pages
826-835
ISSN
  • 1434-6621 (ISSN-L)
  • 1437-4331
OpenAlex
W2557579377
Cited by
65
References
34
Citations by year
2026: 8 · 2025: 11 · 2024: 7 · 2023: 9 · 2022: 4 · 2021: 9 · 2020: 10 · 2019: 1 · 2018: 2 · 2017: 4
FWCI
1.88
Type
journal article
Language
en
Open access
Locations
  • Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion, cc-by-nc-nd)
  • PubMed (publishedVersion)
  • Open access LMU (Ludwid Maxmilian's Universitat Munchen) (submittedVersion, cc-by-nc-nd)
Funders
National Institute of Standards and Technology
Topics
  • Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Adrenal Hormones and Disorders
  • Life Sciences › Neuroscience › Behavioral Neuroscience › Stress Responses and Cortisol
Keywords
cortisol assay, automated immunoassay, analytical performance, method comparison, LC-MS/MS, salivary cortisol, serum cortisol, reference values, adrenocortical disease, Cushing's disease, Addison's disease, assay precision
MeSH terms
Blood Chemical Analysis, Cross Reactions, Humans, Hydrocortisone, Reference Values, Saliva, Limit of Detection

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.