CoChat

Clinical Validation of Heart Rate Apps: Mixed-Methods Evaluation Study

Open access
Vandenberk, Thijs, Stans, Jelle, Mortelmans, Christophe, Van Haelst, Ruth, Van Schelvergem, Gertjan, Pelckmans, Caroline, Smeets, Christophe JP, Lanssens, Dorien, De Cannière, Hélène, Storms, Valerie, Thijs, Inge M, Vaes, Bert, Vandervoort, Pieter M · 2017 · JMIR mHealth and uHealth
Cited by 58

The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Photoplethysmography (PPG) is a proven way to measure heart rate (HR). This technology is already available in smartphones, which allows measuring HR only by using the smartphone. Given the widespread availability of smartphones, this creates a scalable way to enable mobile HR monitoring. An essential precondition is that these technologies are as reliable and accurate as the current clinical (gold) standards. At this moment, there is no consensus on a gold standard method for the validation of HR apps. This results in different validation processes that do not always reflect the veracious outcome of comparison. OBJECTIVE: The aim of this paper was to investigate and describe the necessary elements in validating and comparing HR apps versus standard technology. METHODS: The FibriCheck (Qompium) app was used in two separate prospective nonrandomized studies. In the first study, the HR of the FibriCheck app was consecutively compared with 2 different Food and Drug Administration (FDA)-cleared HR devices: the Nonin oximeter and the AliveCor Mobile ECG. In the second study, a next step in validation was performed by comparing the beat-to-beat intervals of the FibriCheck app to a synchronized ECG recording. RESULTS: In the first study, the HR (BPM, beats per minute) of 88 random subjects consecutively measured with the 3 devices showed a correlation coefficient of .834 between FibriCheck and Nonin, .88 between FibriCheck and AliveCor, and .897 between Nonin and AliveCor. A single way analysis of variance (ANOVA; P=.61 was executed to test the hypothesis that there were no significant differences between the HRs as measured by the 3 devices. In the second study, 20,298 (ms) R-R intervals (RRI)-peak-to-peak intervals (PPI) from 229 subjects were analyzed. This resulted in a positive correlation (rs=.993, root mean square deviation [RMSE]=23.04 ms, and normalized root mean square error [NRMSE]=0.012) between the PPI from FibriCheck and the RRI from the wearable ECG. There was no significant difference (P=.92) between these intervals. CONCLUSIONS: Our findings suggest that the most suitable method for the validation of an HR app is a simultaneous measurement of the HR by the smartphone app and an ECG system, compared on the basis of beat-to-beat analysis. This approach could lead to more correct assessments of the accuracy of HR apps.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2196/mhealth.7254

Cites

6 of 6 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 6 references

Cited by

See all 58

Metadata

Authors
Vandenberk, T., Stans, J., Mortelmans, C., Van Haelst, R., Van Schelvergem, G., Pelckmans, C., Smeets, C. J., Lanssens, D., De Cannière, H., Storms, V., Thijs, I. M., Vaes, B., & Vandervoort, P. M.
Venue
JMIR mHealth and uHealth
Published
2017-08-25
Volume
5
Issue
8
Pages
e129
ISSN
2291-5222 (ISSN-L)
OpenAlex
W2747585374
Cited by
58
References
6
FWCI
2.44
Type
journal article
Language
en
Open access
gold

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Clinical Validation of Heart Rate Apps: Mixed-Methods Evaluation Study. https://app.cochat.ai/papers/10.2196/mhealth.7254
The paper
Vandenberk, T., Stans, J., Mortelmans, C., Van Haelst, R., Van Schelvergem, G., Pelckmans, C., Smeets, C. J., Lanssens, D., De Cannière, H., Storms, V., Thijs, I. M., Vaes, B., & Vandervoort, P. M. (2017). Clinical Validation of Heart Rate Apps: Mixed-Methods Evaluation Study. JMIR mHealth and uHealth. https://doi.org/10.2196/mhealth.7254

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.