Low-Cost, 3-Dimension, Office-Based Inertial Sensors for Automated Tremor Assessment: Technical Development and Experimental Verification

Open access
Bhidayasiri, Roongroj, Petchrutchatachart, Sitthi, Pongthornseri, Ronachai, Anan, Chanawat, Dumnin, Songphon, Thanawattano, Chusak · 2014 · Journal of Parkinson’s Disease
Cited by 26

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

Background: Tremors are common clinical complaints among the elderly and non-specialist physicians frequently are challenged by the need to provide an accurate diagnosis of various tremor syndromes, particularly Parkinson's disease and essential tremor in their busy practices. Objective: We sought to develop an easy-to-use, mobile robust, accurate, and cost-effective instrument that can objectively quantify tremors. Method: The low-cost, 3-dimension, inertial sensors were developed for automated tremor assessment. The main sensor unit consists of a 3-axis accelerometer and a 3-axis gyroscope for the purpose of measuring the tilting angle relative to the gravity, linear acceleration, and angular velocity of the body segments affected by tremors. The transmitter consists of five main modules, including a microcontroller, power management module, sensor module, external memory interface module, and Bluetooth™ communication interface module, which connects to the sensors by a thin wire. The signal processing utilized fast Fourier transform analysis to include RMS angular rate, RMS angle, RMS rate, RMS velocity, peak frequency, peak frequency magnitude, and dispersion of frequency as variables. Result: The prototype was tested with a tremor simulator at programmable angular rates of 2-, 4-, and 8-Hz confirming its accuracy. Twenty subjects (10 PD and 10 age-matched ET patients) participated as part of the experimental verification to perform three tremor tasks, including rest, postural, and kinetic tremor according to the teaching videotape of the motor section of the UPDRS. The mean peak frequency was significantly lower in PD than ET patients at rest on the x- (p < 0.01) and z-axis (p < 0.01). In PD patients, the RMS angular rate, RMS angle, RMS rate, RMS velocity, and peak magnitude were all significantly higher than those values in ET patient at rest while the data was not significantly difference during postural and kinetic actions. ET patients had significantly higher peak frequency during postural action in the y-axis than PD patients (p < 0.05). Conclusion: The study provides the technical development of an accurate, inexpensive, and simple method to measure the kinematics of tremor in humans. Further studies are warranted to confirm the validity of each parameter and the diagnostic accuracy in each tremor syndrome.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3233/jpd-130311

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Cited by

See all 26

Metadata

Authors
Bhidayasiri, R., Petchrutchatachart, S., Pongthornseri, R., Anan, C., Dumnin, S., & Thanawattano, C.
Venue
Journal of Parkinson’s Disease
Published
2014-05
Volume
4
Issue
2
Pages
273-282
ISSN
  • 1877-7171 (ISSN-L)
  • 1877-718X
OpenAlex
W1817415608
Cited by
26
References
10
FWCI
2.33
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Low-Cost, 3-Dimension, Office-Based Inertial Sensors for Automated Tremor Assessment: Technical Development and Experimental Verification. https://app.cochat.ai/papers/10.3233/jpd-130311
The paper
Bhidayasiri, R., Petchrutchatachart, S., Pongthornseri, R., Anan, C., Dumnin, S., & Thanawattano, C. (2014). Low-Cost, 3-Dimension, Office-Based Inertial Sensors for Automated Tremor Assessment: Technical Development and Experimental Verification. Journal of Parkinson’s Disease. https://doi.org/10.3233/jpd-130311

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.