Smart Textile Based on Fiber Bragg Grating Sensors for Respiratory Monitoring: Design and Preliminary Trials
Open accessAbstract
Continuous respiratory monitoring is important to assess adequate ventilation. We present a fiber optic-based smart textile for respiratory monitoring able to work during Magnetic Resonance (MR) examinations. The system is based on the conversion of chest wall movements into strain of two fiber Bragg grating (FBG) sensors, placed on the upper thorax (UT). FBGs are glued on the textile by an adhesive silicon rubber. To increase the system sensitivity, the FBGs positioning was led by preliminary experiments performed using an optoelectronic system: FBGs placed on the chest surface experienced the largest strain during breathing. System performances, in terms of respiratory period (TR), duration of inspiratory (TI) and expiratory (TE) phases, as well as left and right UT volumes, were assessed on four healthy volunteers. The comparison of results obtained by the proposed system and an optoelectronic plethysmography highlights the high accuracy in the estimation of TR, TI, and TE: Bland-Altman analysis shows mean of difference values lower than 0.045 s, 0.33 s, and 0.35 s for TR, TI, and TE, respectively. The mean difference of UT volumes between the two systems is about 8.3%. The promising results foster further development of the system to allow routine use during MR examinations.Continuous respiratory monitoring is important to assess adequate ventilation. We present a fiber optic-based smart textile for respiratory monitoring able to work during Magnetic Resonance (MR) examinations. The system is based on the conversion of chest wall movements into strain of two fiber Bragg grating (FBG) sensors, placed on the upper thorax (UT). FBGs are glued on the textile by an adhesive silicon rubber. To increase the system sensitivity, the FBGs positioning was led by preliminary experiments performed using an optoelectronic system: FBGs placed on the chest surface experienced the largest strain during breathing. System performances, in terms of respiratory period (TR), duration of inspiratory (TI) and expiratory (TE) phases, as well as left and right UT volumes, were assessed on four healthy volunteers. The comparison of results obtained by the proposed system and an optoelectronic plethysmography highlights the high accuracy in the estimation of TR, TI, and TE: Bland-Altman analysis shows mean of difference values lower than 0.045 s, 0.33 s, and 0.35 s for TR, TI, and TE, respectively. The mean difference of UT volumes between the two systems is about 8.3%. The promising results foster further development of the system to allow routine use during MR examinations.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Commercial T-shirts equipped with a single fiber optic housing two FBGs glued on the textile with silicone rubber can monitor RR and HR waves.
“Furthermore, the commercial T-shirts equipped with a single fiber optic housing two FBGs for RR and HR waves’ monitoring, where the FBGs were glued on the textile with silicone rubber were reported [249,250,251].”
From The Current State of Optical Sensors in Medical Wearables (2022) · cites this paper as [250]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300 · see citing works
Review of the present status of optical fiber sensors
Lee, B. (2003) · Optical Fiber Technology · doi:10.1016/s1068-5200(02)00527-8 · Cited by 1,755 · see citing works
Fiber optic sensor technology: an overview
Grattan, K. T. V., & Sun, T. S. (2000) · Sensors and Actuators A Physical · doi:10.1016/s0924-4247(99)00368-4 · Cited by 906 · see citing works
(2011) · Wiley eBooks · doi:10.1002/9781118014103 · Cited by 783 · see citing works
Chest wall and lung volume estimation by optical reflectance motion analysis
Cala, S. J., Kenyon, C. M., Ferrigno, G., Carnevali, P., Aliverti, A., Pedotti, A., Macklem, P. T., & Rochester, D. F. (1996) · Journal of Applied Physiology · doi:10.1152/jappl.1996.81.6.2680 · Cited by 377 · see citing works
Fiber grating sensors in medicine: Current and emerging applications
Mishra, V., Singh, N., Tiwari, U. K., et al. (2011) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2011.02.045 · Cited by 275 · see citing works
Robust ballistocardiogram acquisition for home monitoring
Inan, O. T., Etemadi, M., Wiard, R. M., Giovangrandi, L., & Kovacs, G. T. A. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/2/005 · Cited by 234 · see citing works
Optical Fiber Sensors Embedded Into Medical Textiles for Healthcare Monitoring
Grillet, A., Kinet, D., Witt, J., Schukar, M., Krebber, K., Pirotte, F., & Depré, A. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2008.926518 · Cited by 234 · see citing works
Optical Fiber-Based MR-Compatible Sensors for Medical Applications: An Overview
Taffoni, F., Formica, D., Saccomandi, P., Pino, G. D., & Schena, E. (2013) · Sensors · doi:10.3390/s131014105 · Cited by 221 · see citing works
Self-citationOptical Fibre Pressure Sensors in Medical Applications
Poeggel, S., Tosi, D., Duraibabu, D. B., et al. (2015) · Sensors · doi:10.3390/s150717115 · Cited by 209 · see citing works
Monitoring Respiration and Cardiac Activity Using Fiber Bragg Grating-Based Sensor
Dziuda, Ł., Skibniewski, F. W., Krej, M., & Lewandowski, J. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2194145 · Cited by 186 · see citing works
Medical Smart Textiles Based on Fiber Optic Technology: An Overview
Massaroni, C., Saccomandi, P., & Schena, E. (2015) · Journal of Functional Biomaterials · doi:10.3390/jfb6020204 · Cited by 172 · see citing works
Self-citationMedical Textiles With Embedded Fiber Optic Sensors for Monitoring of Respiratory Movement
Witt, J., Narbonneau, F., Schukar, M., Krebber, K., Jonckheere, J. D., Jeanne, M., Kinet, D., Paquet, B., Depré, A., D’Angelo, L., et al. (2011) · IEEE Sensors Journal · doi:10.1109/jsen.2011.2158416 · Cited by 167 · see citing works
Quandt, B. M., Scherer, L. J., Boesel, L. F., Wolf, M., Bona, G. L., & Rossi, R. M. (2014) · Advanced Healthcare Materials · doi:10.1002/adhm.201400463 · Cited by 148 · see citing works
Saccomandi, P., Schena, E., Caponero, M., et al. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2210895 · Cited by 147 · see citing works
Self-citationAn Optical Fibre-Based Sensor for Respiratory Monitoring
Krehel, M., Schmid, M., Rossi, R. M., Boesel, L. F., Bona, G. L., & Scherer, L. J. (2014) · Sensors · doi:10.3390/s140713088 · Cited by 127 · see citing works
FBG Sensing Glove for Monitoring Hand Posture
da Silva, A. F., Gonçalves, A., Mendes, P. M., et al. (2011) · IEEE Sensors Journal · doi:10.1109/jsen.2011.2138132 · Cited by 122 · see citing works
ECG patch monitors for assessment of cardiac rhythm abnormalities
Lobodzinski, S. S. (2013) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2013.08.006 · Cited by 119 · see citing works
Fiber Bragg Grating Strain Sensor Incorporated to Monitor Patient Vital Signs During MRI
Dziuda, Ł., Krej, M., & Skibniewski, F. W. (2013) · IEEE Sensors Journal · doi:10.1109/jsen.2013.2279160 · Cited by 107 · see citing works
Dziuda, Ł. (2015) · Journal of Biomedical Optics · doi:10.1117/1.jbo.20.1.010901 · Cited by 106 · see citing works
New devices for very long-term ECG monitoring
Lobodzinski, S. S., & Laks, M. M. (2012) · Cardiology Journal · doi:10.5603/cj.2012.0039 · Cited by 106 · see citing works
Application of Fiber Bragg Gratings to Wearable Garments
Carmo, J. P., da Silva, A. F., Rocha, R. P., & Correia, J. H. (2011) · IEEE Sensors Journal · doi:10.1109/jsen.2011.2161281 · Cited by 51 · see citing works
Technological Solutions and Main Indices for the Assessment of Newborns’ Nutritive Sucking: A Review
Tamilia, E., Taffoni, F., Formica, D., et al. (2014) · Sensors · doi:10.3390/s140100634 · Cited by 48 · see citing works
Self-citationRomagnoli, P., Lanini, B., Binazzi, B., Bianchi, R., Coli, C., Stendardi, L., Gigliotti, F., & Scano, G. (2008) · Sensors · doi:10.3390/s8127951 · Cited by 36 · see citing works
Allsop, T., Bhamber, R., Lloyd, G., et al. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.11.117001 · Cited by 36 · see citing works
Krej, M., Dziuda, Ł., & Skibniewski, F. W. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2015.2392796 · Cited by 33 · see citing works
Cleutjens, F. A. H. M., Wouters, E. F. �., Dijkstra, J. B., Spruit, M. A., Franssen, F. M. E., Vanfleteren, L. E. G. W., Ponds, R., & Janssen, D. J. A. (2014) · BMJ Open · doi:10.1136/bmjopen-2013-004495 · Cited by 17 · see citing works
Smart Textile Embedding Optical Fibre Sensors for Healthcare Monitoring during MRI
Narbonneau, F., Kinet, D., Paquet, B., et al. (2008) · Advances in science and technology · doi:10.4028/www.scientific.net/ast.60.134 · Cited by 14 · see citing works
Massaroni, C., Schena, E., Bastianini, F., Scorza, A., Saccomandi, P., Lupi, G., Botta, F., Sciuto, S. A., & Silvestri, S. (2014) · The Open Biomedical Engineering Journal · doi:10.2174/1874120701408010120 · Cited by 12 · see citing works
Self-citationMassaroni, C., Schena, E., Saccomandi, P., et al. (2015) · doi:10.1109/embc.2015.7318604 · Cited by 12 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Smart textile for respiratory monitoring and thoraco-abdominal motion pattern evaluation
Massaroni, C., Venanzi, C., Silvatti, A. P., Lo Presti, D., Saccomandi, P., Formica, D., Giurazza, F., Caponero, M. A., & Schena, E. (2018) · Journal of Biophotonics · doi:10.1002/jbio.201700263 · Cited by 139 · see citing works
Self-citationFiber Bragg Grating Sensors for Cardiorespiratory Monitoring: A Review
Massaroni, C., Zaltieri, M., Lo Presti, D., Nicolo, A., Tosi, D., & Schena, E. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2988692 · Cited by 128 · see citing works
Self-citationSmart textile using hetero-core optical fiber for heartbeat and respiration monitoring
Koyama, Y., Nishiyama, M., & Watanabe, K. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2847333 · Cited by 126 · see citing works
The Current State of Optical Sensors in Medical Wearables
Vavrinsky, E., Esfahani, N. E., Hausner, M., Kuzma, A., Rezo, V., Donoval, M., & Kosnacova, H. (2022) · Biosensors · doi:10.3390/bios12040217 · Cited by 103 · see citing works
- Authors
- Ciocchetti, M., Massaroni, C., Saccomandi, P., Caponero, M., Polimadei, A., Formica, D., & Schena, E.
- Affiliations
- Università Campus Bio-Medico (IT)
- National Agency for New Technologies, Energy and Sustainable Economic Development (IT)
- Corresponding author
- Schena, Emiliano
- ORCID
- Venue
- Biosensors
- Publisher
- MDPI AG
- Published
- 2015-09-14
- Volume
- 5
- Issue
- 3
- Pages
- 602-615
- ISSN
- 2079-6374 (ISSN-L)
- PMC
- PMC4600174
- OpenAlex
- W1867009124
- Cited by
- 142
- References
- 30
- Citations by year
- 2026: 3 · 2025: 7 · 2024: 15 · 2023: 20 · 2022: 17 · 2021: 9 · 2020: 16 · 2019: 16 · 2018: 16 · 2017: 15 · 2016: 8
- FWCI
- 6.47
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Biosensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Virtual Community of Pathological Anatomy (University of Castilla La Mancha) (submittedVersion, other-oa)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- ENEA Open Archive (National Agency for New Technologies, Energy and Sustainable Economic Development) (submittedVersion)
- Funders
- Ministero dell’Istruzione, dell’Università e della Ricerca (PRIN2012, 20127XJX57)
- Università Campus Bio-Medico di Roma
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced Fiber Optic Sensors
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Keywords
- fiber Bragg grating sensors, smart textiles, respiratory monitoring, magnetic resonance imaging, ventilation assessment, plethysmography, optoelectronic plethysmography, wearable sensors, non-invasive monitoring, Bland-Altman agreement
- MeSH terms
- Fiber Optic Technology, Humans, Monitoring, Physiologic, Pulmonary Ventilation, Biosensing Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review