Estimation of physiological tremor from accelerometers for real-time applications

Open access

The DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index; 2 of the 19 sources its claims cite were checked for support.

Abstract

Accurate filtering of physiological tremor is extremely important in robotics assisted surgical instruments and procedures. This paper focuses on developing single stage robust algorithms for accurate tremor filtering with accelerometers for real-time applications. Existing methods rely on estimating the tremor under the assumption that it has a single dominant frequency. Our time-frequency analysis on physiological tremor data revealed that tremor contains multiple dominant frequencies over the entire duration rather than a single dominant frequency. In this paper, the existing methods for tremor filtering are reviewed and two improved algorithms are presented. A comparative study is conducted on all the estimation methods with tremor data from microsurgeons and novice subjects under different conditions. Our results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation. A procedure to separate the intended motion/drift from the tremor component is formulated.

Overview

CoChat Summary

TL;DR

The paper argues that physiological tremor is not well modeled as a single dominant frequency and presents two improved BMFLC-based methods for real-time tremor estimation from accelerometers. On data from 6 healthy subjects and 6 microsurgeons, BMFLC-Kalman performed best, with an average RMS error of 0.003 ± 0.002 μm.

Key findings

  • Time-frequency analysis of tremor data from 6 microsurgeons and 6 healthy subjects showed multiple dominant frequencies throughout the tremor period, with a bandwidth of 2–5 Hz and multiple peaks in the amplitude spectrum.
  • In the comparative evaluation of five algorithms, BMFLC-based methods outperformed WFLC-based methods; reported accuracies were 90% for WFLC, 93% for WFLC-Kalman, 94% for BMFLC-LMS, 99% for BMFLC-RLS, and 99.5% for BMFLC-Kalman.
  • BMFLC-Kalman achieved the lowest average RMS error, reported as 0.003 ± 0.002 μm, and the paper states it was more suited for real-time estimation.
  • For voluntary-motion separation on raw tracing data, the proposed method had lower delay and error than zero-phase low-pass filtering: 0.082 s delay and RMSE 4.61 μm versus 0.2 s and 17.8 μm; for a pointing task, 0.072 s and 6.63 μm versus 0.186 s and 26.5 μm.

Methods

  • Comparative algorithm study on tremor estimation using five methods: WFLC, WFLC-Kalman, BMFLC-LMS, BMFLC-RLS, and BMFLC-Kalman.
  • Data/sample: tremor recordings from 6 healthy subjects and 6 microsurgeons; each performed a stationary task and a tracing task, with two trials per task.
  • Main analysis: time-frequency analysis with BMFLC, bandpass-filtered comparison data for algorithm evaluation, and error/delay assessment using RMSE and reported accuracy measures.

Limitations

  • The paper states that BMFLC requires an initial period of 1 s for weight stabilization.
  • The text makes evident that the reported comparison relies on bandpass-filtered signals for WFLC and WFLC-Kalman “for sake of ideal comparison,” which is not ideal for real-time use because it introduces delay; this limitation is noted in the text as affecting practical applicability.
  • The paper notes that further research is needed for complex gestures and sudden jerks in microsurgery; this is stated as a remaining challenge.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.3390/s110303020

Cites

36 of 36 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1. Micro Motion Sensing System (M2S2) setup.
Figure 1. Micro Motion Sensing System (M2S2) setup.CC BY · PubMed Central
Figure 2. Time-Frequency mapping of surgeon #1 and novice subject #1 in the band of 7–14 Hz.
Figure 2. Time-Frequency mapping of surgeon #1 and novice subject #1 in the band of 7–14 Hz.CC BY · PubMed Central
Figure 3. BMFLC Architecture.
Figure 3. BMFLC Architecture.CC BY · PubMed Central
Figure 4. Block diagram for BMFLC-Kalman tremor filtering.
Figure 4. Block diagram for BMFLC-Kalman tremor filtering.CC BY · PubMed Central
Figure 5. (a) Raw data recorded from Subject #4 with tracing task and the identified voluntary motion; (b) Comparison with zero-phase lowpass filter and Lowpass filter.
Figure 5. (a) Raw data recorded from Subject #4 with tracing task and the identified voluntary motion; (b) Comparison with zero-phase lowpass filter and Lowpass filter.CC BY · PubMed Central
Figure 6. (a) Raw data recorded from Surgeon #1 with pointing task and the identified voluntary motion; (b) Comparison with Lowpass filter.
Figure 6. (a) Raw data recorded from Surgeon #1 with pointing task and the identified voluntary motion; (b) Comparison with Lowpass filter.CC BY · PubMed Central
Figure 7. Performance of all algorithms with Surgeon #1 (pointing task).
Figure 7. Performance of all algorithms with Surgeon #1 (pointing task).CC BY · PubMed Central
Figure 8. Average RMS tracking error (μm) for all algorithms with tracing and pointing tasks; error bars represent standard deviation around mean.
Figure 8. Average RMS tracking error (μm) for all algorithms with tracing and pointing tasks; error bars represent standard deviation around mean.CC BY · PubMed Central

Verification

Claims the paper makes

1 Valid · 1 Not valid · 9 Not checked · 19 Own results and interpretations

11 more claims

Create a free account to see all 30 claims and to verify the 2 references still open.

See all claims — free account
  • It may state two separate claims as one. 12
  • Probably not a substantial claim. 3

Create a free account to see these claims.

Citations

Cites

Show all 36 references

Research works printed but not linked: 1

  1. Elble, RJ, and Koller, WC (1985). Tremor, John Hopkins University Press.

Cited by

See all 117

Metadata

Authors
Veluvolu, K. C., & Ang, W. T.
Affiliations
  • Kyungpook National University (KR)
  • Nanyang Technological University (SG)
Corresponding author
Veluvolu, Kalyana C.
Venue
Sensors
Publisher
MDPI AG
Published
2011-03-07
Volume
11
Issue
3
Pages
3020-3036
ISSN
1424-8220 (ISSN-L)
OpenAlex
W2071890760
Cited by
117
References
36
Citations by year
2026: 5 · 2025: 8 · 2024: 9 · 2023: 5 · 2022: 5 · 2021: 7 · 2020: 9 · 2019: 5 · 2018: 8 · 2017: 6 · 2016: 15 · 2015: 5 · 2014: 7 · 2013: 14 · 2012: 5
FWCI
4.87
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
  • Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
  • Health Sciences › Medicine › Surgery › Surgical Simulation and Training
Keywords
physiological tremor, tremor filtering, accelerometer, real-time signal processing, time-frequency analysis, surgical robotics, drift estimation
MeSH terms
Acceleration, Algorithms, Fourier Analysis, Humans, Physicians, Physiology, Time Factors, Tremor, Least-Squares Analysis

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Estimation of physiological tremor from accelerometers for real-time applications. https://app.cochat.ai/papers/10.3390/s110303020
The paper
Veluvolu, K. C., & Ang, W. T. (2011). Estimation of physiological tremor from accelerometers for real-time applications. Sensors. https://doi.org/10.3390/s110303020

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.