Estimation of physiological tremor from accelerometers for real-time applications
Open accessThe DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index; 2 of the 19 sources its claims cite were checked for support.
Abstract
Accurate filtering of physiological tremor is extremely important in robotics assisted surgical instruments and procedures. This paper focuses on developing single stage robust algorithms for accurate tremor filtering with accelerometers for real-time applications. Existing methods rely on estimating the tremor under the assumption that it has a single dominant frequency. Our time-frequency analysis on physiological tremor data revealed that tremor contains multiple dominant frequencies over the entire duration rather than a single dominant frequency. In this paper, the existing methods for tremor filtering are reviewed and two improved algorithms are presented. A comparative study is conducted on all the estimation methods with tremor data from microsurgeons and novice subjects under different conditions. Our results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation. A procedure to separate the intended motion/drift from the tremor component is formulated.
Overview
CoChat Summary
TL;DR
The paper argues that physiological tremor is not well modeled as a single dominant frequency and presents two improved BMFLC-based methods for real-time tremor estimation from accelerometers. On data from 6 healthy subjects and 6 microsurgeons, BMFLC-Kalman performed best, with an average RMS error of 0.003 ± 0.002 μm.
Key findings
- Time-frequency analysis of tremor data from 6 microsurgeons and 6 healthy subjects showed multiple dominant frequencies throughout the tremor period, with a bandwidth of 2–5 Hz and multiple peaks in the amplitude spectrum.
- In the comparative evaluation of five algorithms, BMFLC-based methods outperformed WFLC-based methods; reported accuracies were 90% for WFLC, 93% for WFLC-Kalman, 94% for BMFLC-LMS, 99% for BMFLC-RLS, and 99.5% for BMFLC-Kalman.
- BMFLC-Kalman achieved the lowest average RMS error, reported as 0.003 ± 0.002 μm, and the paper states it was more suited for real-time estimation.
- For voluntary-motion separation on raw tracing data, the proposed method had lower delay and error than zero-phase low-pass filtering: 0.082 s delay and RMSE 4.61 μm versus 0.2 s and 17.8 μm; for a pointing task, 0.072 s and 6.63 μm versus 0.186 s and 26.5 μm.
Methods
- Comparative algorithm study on tremor estimation using five methods: WFLC, WFLC-Kalman, BMFLC-LMS, BMFLC-RLS, and BMFLC-Kalman.
- Data/sample: tremor recordings from 6 healthy subjects and 6 microsurgeons; each performed a stationary task and a tracing task, with two trials per task.
- Main analysis: time-frequency analysis with BMFLC, bandpass-filtered comparison data for algorithm evaluation, and error/delay assessment using RMSE and reported accuracy measures.
Limitations
- The paper states that BMFLC requires an initial period of 1 s for weight stabilization.
- The text makes evident that the reported comparison relies on bandpass-filtered signals for WFLC and WFLC-Kalman “for sake of ideal comparison,” which is not ideal for real-time use because it introduces delay; this limitation is noted in the text as affecting practical applicability.
- The paper notes that further research is needed for complex gestures and sudden jerks in microsurgery; this is stated as a remaining challenge.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures








Verification
Claims the paper makes
1 Valid · 1 Not valid · 9 Not checked · 19 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Among the five algorithms BMFLC-Kalman performed better producing an accurate estimate of tremor with an average RMS error of 0.003 ± 0.002 (μm) and is more suited for real-time estimation.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Conclusions
plainn = 6- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Our results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation.”
- Also stated in: § Results · boosted
“It is clearly evident that BMFLC-RLS and BMFLC-Kalman perform better than the rest of the algorithms.”
- Also stated in: § Conclusions · plain
“The proposed methods BMFLC-Kalman and BMFLC RLS performed better than the existing methods WFLC, WFLC-Kalman and BMFLC-LMS.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Abstract
boostedstates a causen = 6- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Results · boosted
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Our time-frequency analysis on physiological tremor data revealed that tremor contains multiple dominant frequencies over the entire duration rather than a single dominant frequency.”
- Also stated in: § Time-Frequency Characteristics of Tremor · plain
“Observation of tremor amplitude spectrum and BMFLC time-frequency map reveals that existence of multiple dominant frequencies through out the tremor period.”
- Also stated in: § Time-Frequency Characteristics of Tremor · plain
“In all the subjects involved in our study there existed a bandwidth of 2–5 Hz with several peaks in the amplitude spectrum and several dominant time-frequency components in the time-frequency map.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Abstract
plainn = 6- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Time-Frequency Characteristics of Tremor · plain
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As pointing task is less complex than the tracing task, the average RMS estimation errors for tracing tasks are higher than that of pointing tasks.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Discussion
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Employing WFLC-Kalman improved the accuracy 93%.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Results
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
11 more claims
Create a free account to see all 30 claims and to verify the 2 references still open.
- It may state two separate claims as one. 12
- Probably not a substantial claim. 3
Citations
Cites
Haykin, S. (1986) · Cited by 12,603
New Results in Linear Filtering and Prediction Theory
Kalman, R. E., & Bucy, R. S. (1961) · Journal of Basic Engineering · doi:10.1115/1.3658902 · Cited by 6,387
(1991) · Springer eBooks · doi:10.1007/978-3-7091-2840-4 · Cited by 5,095
Anderson, B. D. O., Moore, J. B., & Eslami, M. (1982) · IEEE Transactions on Systems Man and Cybernetics · Cited by 3,195
Time Series Analysis and Its Applications
Shumway, R. H., & Stoffer, D. S. (2011) · Springer Texts in Statistics · doi:10.1007/978-1-4419-7865-3 · Cited by 2,012
Consensus Statement of the Movement Disorder Society on Tremor
Deuschl, G., Bain, P., & Brin, M. (2008) · Movement Disorders · doi:10.1002/mds.870131303 · Cited by 1,981
Time Series Analysis and Its Applications
Shumway, R. H., & Stoffer, D. S. (2000) · Springer Texts in Statistics · doi:10.1007/978-1-4757-3261-0 · Cited by 1,039
Deuschl, G., Raethjen, J., Lindemann, M., & Krack, P. (2001) · Muscle & Nerve · doi:10.1002/mus.1063 · Cited by 508 · cited 2×
“Physiological hand tremor lies in the band of 8–12 Hz with an amplitude of 50 μm and can be approximated by a sinusoidal movement [1,3].”
BackgroundA state-space approach to adaptive RLS filtering
Sayed, A. H., & Kailath, T. (1994) · IEEE Signal Processing Magazine · doi:10.1109/79.295229 · Cited by 449 · cited 11×
“The recursive least squares algorithm (RLS) [28,29] is an adaptive algorithm that uses successive corrections for the filter coefficients and minimizes weighted least squares error function related to the input signal.”
MethodInfluentialDesign and Validation of a Rehabilitation Robotic Exoskeleton for Tremor Assessment and Suppression
Rocon, E., Belda-Lois, J. M., Ruiz, A. F., Manto, M., Moreno, J. C., & Pons, J. L. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.903917 · Cited by 349
Adaptive cancelling of physiological tremor for improved precision in microsurgery
Riviere, C. N., Rader, R. S., & Thakor, N. V. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.686791 · Cited by 260 · cited 3×
“As part of our continuing research in developing smart surgical device such as Micron [12,26], with accelerometers, the algorithms are developed for cancellation of tremor in real-time with accelerometers.”
BackgroundMethodToward active tremor canceling in handheld microsurgical instruments
Riviere, C. N., Ang, W. T., & Khosla, P. K. (2003) · IEEE Transactions on Robotics and Automation · doi:10.1109/tra.2003.817506 · Cited by 245 · cited 1×
“In [11,12] a robotic handheld instrument to cancel physiological tremor of surgeon in vitreoretinal microsurgery was implemented.”
MethodSelf-citationComputer-integrated surgery. Technology and clinical applications. 1996.
Taylor, R. H., Lavelleé, S., Burdea, G., & Mösges, R. (1998) · PubMed · Cited by 218
Numerical double integration of acceleration measurements in noise
Thong, Y. K., Woolfson, M. S., Crowe, J. A., Hayes-Gill, B. R., & Jones, D. A. (2004) · Measurement · doi:10.1016/j.measurement.2004.04.005 · Cited by 206 · cited 2×
“MEMS accelerometers constitutes the most common for tremor sensing in robotics assisted procedures [14,15].”
BackgroundMethodTime-frequency analysis of tremors
O'Suilleabhain, P. (1998) · Brain · doi:10.1093/brain/121.11.2127 · Cited by 157 · cited 2×
“The frequency of the pathological tremor tends to remain constant with slight variations [5].”
BackgroundComputer-Integrated Surgery: Technology and Clinical Applications
Taylor, R. H., Lavall??e, S., Burdea, G. C., & M??sges, R. (1998) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-199809000-00002 · Cited by 150
Balakrishnan, A. V. '. (1984) · Cited by 146
Tarvainen, M. P., Hiltunen, J. K., Ranta-aho, P. O., & Karjalainen, P. A. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.821029 · Cited by 133 · cited 1×
“Given the measurement sequence s1:k−1, the estimated state ŵk and the estimated state error covariance Pk can be obtained by Kalman filter recursively [34]: • Compute Kalman gain Kk Kk = Pkx T k ( xk Pkxk + R )−1 (14)”
Modeling and canceling tremor in human-machine interfaces
Riviere, C. N., & Thakor, N. V. (1996) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.499755 · Cited by 114 · cited 3×
“Weighted Fourier Linear Combiner (WFLC) [20] and WFLC-Kalman filters [7] rely on single frequency estimation, whereas Bandlimited Multiple-Fourier Linear Combiner (BMFLC) rely on multiple frequency components estimation.”
MethodAn adaptive estimation of periodic signals using a Fourier linear combiner
Vaz, C., Kong, X., & Thakor, N. (1994) · IEEE Transactions on Signal Processing · doi:10.1109/78.258116 · Cited by 108 · cited 3×
“The time constant for convergence can be shown to be 1 2μ [21].”
BackgroundMethodReal-Time Estimation of Pathological Tremor Parameters from Gyroscope Data
Gallego, J., Rocon, E., Roa, J. O., Moreno, J., & Pons, J. L. (2010) · Sensors · doi:10.3390/s100302129 · Cited by 107
Bandlimited Multiple Fourier Linear Combiner for Real-time Tremor Compensation
Veluvolu, K. C., Tan, U. X., Latt, W. T., Shee, C. Y., & Ang, W. T. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4352922 · Cited by 83 · cited 2×
“Band limited multiple Fourier linear combiner (BMFLC) [9,22] is also an adaptive algorithm developed to track multiple dominant frequencies in tremor for accurate tremor filtering.”
MethodSelf-citationAdaptive Fourier modeling for quantification of tremor
Riviere, C. N., Reich, S. G., & Thakor, N. V. (1997) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(97)02263-2 · Cited by 83 · cited 1×
“For the tremor signal consisting of multiple dominant frequencies closely, the accuracy in tremor estimation decreases with WFLC [6].”
Veluvolu, K. C., & Ang, W. T. (2010) · The International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.340 · Cited by 83
Self-citationADAPTIVE SIGNAL PROCESSING FOR ADAPTIVE CONTROL
Widrow, B., & Walach, E. (1984) · Adaptive Systems in Control and Signal Processing 1983 · doi:10.1016/b978-0-08-030565-3.50007-1 · Cited by 76
Estimating displacement of periodic motion with inertial sensors
Tan, U.-X., Veluvolu, K. C., Latt, W. T., Shee, C. Y., Riviere, C. N., & Ang, W. T. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2008.917488 · Cited by 67 · cited 1×
“To overcome the drift, BMFLC can be employed in the acceleration domain to directly obtain the displacement of the tremor signal [35].”
MethodSelf-citation(1992) · doi:10.1201/b12577 · Cited by 57
Optimal digital filtering for tremor suppression
Gonzalez, J. G., Heredia, E. A., Rahman, T., Barner, K. E., & Arce, G. R. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.841338 · Cited by 56 · cited 1×
“Although linear filters [16,17] are successful in compensating tremor, the inherent time delay [18] is a major drawback where zero-phase filtering is required.”
BackgroundAn Active Hand-Held Instrument for Enhanced Microsurgical Accuracy
Ang, W. T., Riviere, C. N., & Khosla, P. K. (2000) · Lecture Notes in Computer Science · doi:10.1007/978-3-540-40899-4_91 · Cited by 52 · cited 3×
“As part of our continuing research in developing smart surgical device such as Micron [12,26], with accelerometers, the algorithms are developed for cancellation of tremor in real-time with accelerometers.”
BackgroundMethodSelf-citationReal-Time Modeling and Prediction of Physiological Hand Tremor
Zhang, J., & Chu, F. (2006) · Proceedings. (ICASSP '05). IEEE International Conference on Acoustics, Speech, and Signal Processing, 2005. · doi:10.1109/icassp.2005.1416386 · Cited by 43 · cited 1×
“Although linear filters [16,17] are successful in compensating tremor, the inherent time delay [18] is a major drawback where zero-phase filtering is required.”
BackgroundSaberi, A., Stoorvogel, A. A., & Sannuti, P. (2007) · Cited by 38
Design and Calibration of an Optical Micro Motion Sensing System for Micromanipulation Tasks
Win, T. L., Tan, U. X., Shee, C. Y., & Ang, W. T. (2007) · Proceedings 2007 IEEE International Conference on Robotics and Automation · doi:10.1109/robot.2007.363995 · Cited by 25
Self-citationExtraction of Sources of Tremor in Hand Movements of Patients With Movement Disorders
Vinjamuri, R., Crammond, D. J., Kondziolka, D., Lee, H.-N., & Mao, Z.-H. (2009) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2008.2006403 · Cited by 23
Physiological tremor sensing using only accelerometers for real-time compensation
Latt, W. T., Tan, U.-X., Veluvolu, K. C., Shee, C. Y., & Ang, W. T. (2009) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio.2009.4913049 · Cited by 21 · cited 1×
“MEMS accelerometers constitutes the most common for tremor sensing in robotics assisted procedures [14,15].”
MethodSelf-citationDSP controller based signal processing of physiological hand tremor
Zhang, J., Chu, F., & Mohammed, N. (2005) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2005.1470191 · Cited by 18
System to Assess Accuracy of Micromanipulation
Latt, W. T., Tan, U. X., Veluvolu, K. C., Lin, J. K. D., Shee, C. Y., & Ang, W. T. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4353651 · Cited by 13
Self-citation
Research works printed but not linked: 1
- Elble, RJ, and Koller, WC (1985). Tremor, John Hopkins University Press.
Cited by
Telesurgery: Past, Present, and Future
Choi, P. J., Oskouian, R. J., & Tubbs, R. S. (2018) · Cureus · doi:10.7759/cureus.2716 · Cited by 204
Quantitative Assessment of Parkinsonian Tremor Based on an Inertial Measurement Unit
Dai, H., Zhang, P., & Lueth, T. (2015) · Sensors · doi:10.3390/s151025055 · Cited by 159
“Tremor Quantification has been performed by signal processing methods such as time-domain analysis, time-frequency analysis, spectral analysis, and nonlinear analysis [31,32].”
MethodSimultaneous Estimation of Road Profile and Tire Road Friction for Automotive Vehicle
Rath, J. J., Veluvolu, K. C., & Defoort, M. (2014) · IEEE Transactions on Vehicular Technology · doi:10.1109/tvt.2014.2373434 · Cited by 88
“Recently, new methods have been developed to provide driftfree displacement and velocity estimation from acceleration for various applications [24].”
MethodSelf-citationTatinati, S., Veluvolu, K. C., & Ang, W. T. (2015) · IEEE Transactions on Cybernetics · doi:10.1109/tcyb.2014.2381495 · Cited by 77
“The time-frequency characteristics of surgeons and novice subjects were analyzed in [15].”
BackgroundMethodSelf-citationReview of Advanced Medical Telerobots
Mehrdad, S., Liu, F., Pham, M. T., Lelevé, A., & Atashzar, S. F. (2020) · Applied Sciences · doi:10.3390/app11010209 · Cited by 66
“…it can be mentioned that through advanced computer vision and image processing modules (most of which have a Kalman filter core algorithm [11,192,193,195–198]) it is now possible to measure motions of moving organs such as heart and lung during the operation and compensate automatically for…”
Metadata
- Authors
- Veluvolu, K. C., & Ang, W. T.
- Affiliations
- Kyungpook National University (KR)
- Nanyang Technological University (SG)
- Corresponding author
- Veluvolu, Kalyana C.
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2011-03-07
- Volume
- 11
- Issue
- 3
- Pages
- 3020-3036
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC3231635
- OpenAlex
- W2071890760
- Cited by
- 117
- References
- 36
- Citations by year
- 2026: 5 · 2025: 8 · 2024: 9 · 2023: 5 · 2022: 5 · 2021: 7 · 2020: 9 · 2019: 5 · 2018: 8 · 2017: 6 · 2016: 15 · 2015: 5 · 2014: 7 · 2013: 14 · 2012: 5
- FWCI
- 4.87
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
- Health Sciences › Medicine › Surgery › Surgical Simulation and Training
- Keywords
- physiological tremor, tremor filtering, accelerometer, real-time signal processing, time-frequency analysis, surgical robotics, drift estimation
- MeSH terms
- Acceleration, Algorithms, Fourier Analysis, Humans, Physicians, Physiology, Time Factors, Tremor, Least-Squares Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.