Estimation of physiological tremor from accelerometers for real-time applications
Open accessThe DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index; 2 of the 19 sources its claims cite were checked for support.
Abstract
Accurate filtering of physiological tremor is extremely important in robotics assisted surgical instruments and procedures. This paper focuses on developing single stage robust algorithms for accurate tremor filtering with accelerometers for real-time applications. Existing methods rely on estimating the tremor under the assumption that it has a single dominant frequency. Our time-frequency analysis on physiological tremor data revealed that tremor contains multiple dominant frequencies over the entire duration rather than a single dominant frequency. In this paper, the existing methods for tremor filtering are reviewed and two improved algorithms are presented. A comparative study is conducted on all the estimation methods with tremor data from microsurgeons and novice subjects under different conditions. Our results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation. A procedure to separate the intended motion/drift from the tremor component is formulated.
Overview
CoChat Summary
TL;DR
The paper argues that physiological tremor is not well modeled as a single dominant frequency and presents two improved BMFLC-based methods for real-time tremor estimation from accelerometers. On data from 6 healthy subjects and 6 microsurgeons, BMFLC-Kalman performed best, with an average RMS error of 0.003 ± 0.002 μm.
Key findings
- Time-frequency analysis of tremor data from 6 microsurgeons and 6 healthy subjects showed multiple dominant frequencies throughout the tremor period, with a bandwidth of 2–5 Hz and multiple peaks in the amplitude spectrum.
- In the comparative evaluation of five algorithms, BMFLC-based methods outperformed WFLC-based methods; reported accuracies were 90% for WFLC, 93% for WFLC-Kalman, 94% for BMFLC-LMS, 99% for BMFLC-RLS, and 99.5% for BMFLC-Kalman.
- BMFLC-Kalman achieved the lowest average RMS error, reported as 0.003 ± 0.002 μm, and the paper states it was more suited for real-time estimation.
- For voluntary-motion separation on raw tracing data, the proposed method had lower delay and error than zero-phase low-pass filtering: 0.082 s delay and RMSE 4.61 μm versus 0.2 s and 17.8 μm; for a pointing task, 0.072 s and 6.63 μm versus 0.186 s and 26.5 μm.
Methods
- Comparative algorithm study on tremor estimation using five methods: WFLC, WFLC-Kalman, BMFLC-LMS, BMFLC-RLS, and BMFLC-Kalman.
- Data/sample: tremor recordings from 6 healthy subjects and 6 microsurgeons; each performed a stationary task and a tracing task, with two trials per task.
- Main analysis: time-frequency analysis with BMFLC, bandpass-filtered comparison data for algorithm evaluation, and error/delay assessment using RMSE and reported accuracy measures.
Limitations
- The paper states that BMFLC requires an initial period of 1 s for weight stabilization.
- The text makes evident that the reported comparison relies on bandpass-filtered signals for WFLC and WFLC-Kalman “for sake of ideal comparison,” which is not ideal for real-time use because it introduces delay; this limitation is noted in the text as affecting practical applicability.
- The paper notes that further research is needed for complex gestures and sudden jerks in microsurgery; this is stated as a remaining challenge.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures








Verification
Claims the paper makes
1 Valid · 1 Not valid · 9 Not checked · 19 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Among the five algorithms BMFLC-Kalman performed better producing an accurate estimate of tremor with an average RMS error of 0.003 ± 0.002 (μm) and is more suited for real-time estimation.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Conclusions
plainn = 6- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Our results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation.”
- Also stated in: § Results · boosted
“It is clearly evident that BMFLC-RLS and BMFLC-Kalman perform better than the rest of the algorithms.”
- Also stated in: § Conclusions · plain
“The proposed methods BMFLC-Kalman and BMFLC RLS performed better than the existing methods WFLC, WFLC-Kalman and BMFLC-LMS.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Abstract
boostedstates a causen = 6- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Results · boosted
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Our time-frequency analysis on physiological tremor data revealed that tremor contains multiple dominant frequencies over the entire duration rather than a single dominant frequency.”
- Also stated in: § Time-Frequency Characteristics of Tremor · plain
“Observation of tremor amplitude spectrum and BMFLC time-frequency map reveals that existence of multiple dominant frequencies through out the tremor period.”
- Also stated in: § Time-Frequency Characteristics of Tremor · plain
“In all the subjects involved in our study there existed a bandwidth of 2–5 Hz with several peaks in the amplitude spectrum and several dominant time-frequency components in the time-frequency map.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Abstract
plainn = 6- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § Time-Frequency Characteristics of Tremor · plain
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“As pointing task is less complex than the tracing task, the average RMS estimation errors for tracing tasks are higher than that of pointing tasks.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Discussion
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Employing WFLC-Kalman improved the accuracy 93%.”
§ Estimation of Physiological Tremor from Accelerometers for Real-Time Applications > Results
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
11 more claims
Create a free account to see all 30 claims and to verify the 2 references still open.
- It may state two separate claims as one. 12
- Probably not a substantial claim. 3
Citations
Cites
Haykin, S. (1986) · Cited by 12,603
New Results in Linear Filtering and Prediction Theory
Kalman, R. E., & Bucy, R. S. (1961) · Journal of Basic Engineering · doi:10.1115/1.3658902 · Cited by 6,387
(1991) · Springer eBooks · doi:10.1007/978-3-7091-2840-4 · Cited by 5,095
Anderson, B. D. O., Moore, J. B., & Eslami, M. (1982) · IEEE Transactions on Systems Man and Cybernetics · Cited by 3,195
Time Series Analysis and Its Applications
Shumway, R. H., & Stoffer, D. S. (2011) · Springer Texts in Statistics · doi:10.1007/978-1-4419-7865-3 · Cited by 2,012
Consensus Statement of the Movement Disorder Society on Tremor
Deuschl, G., Bain, P., & Brin, M. (2008) · Movement Disorders · doi:10.1002/mds.870131303 · Cited by 1,981
Time Series Analysis and Its Applications
Shumway, R. H., & Stoffer, D. S. (2000) · Springer Texts in Statistics · doi:10.1007/978-1-4757-3261-0 · Cited by 1,039
Deuschl, G., Raethjen, J., Lindemann, M., & Krack, P. (2001) · Muscle & Nerve · doi:10.1002/mus.1063 · Cited by 508 · cited 2×
“Physiological hand tremor lies in the band of 8–12 Hz with an amplitude of 50 μm and can be approximated by a sinusoidal movement [1,3].”
BackgroundA state-space approach to adaptive RLS filtering
Sayed, A. H., & Kailath, T. (1994) · IEEE Signal Processing Magazine · doi:10.1109/79.295229 · Cited by 449 · cited 11×
“The recursive least squares algorithm (RLS) [28,29] is an adaptive algorithm that uses successive corrections for the filter coefficients and minimizes weighted least squares error function related to the input signal.”
MethodInfluentialDesign and Validation of a Rehabilitation Robotic Exoskeleton for Tremor Assessment and Suppression
Rocon, E., Belda-Lois, J. M., Ruiz, A. F., Manto, M., Moreno, J. C., & Pons, J. L. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.903917 · Cited by 349
Research works printed but not linked: 2
- Elble RJ Koller WC Tremor John Hopkins University Press Baltimore, MD, USA 1985
- Veluvolu KC Wang Y Kavuri S BMFLC based Time-frequency decomposition of EEG for event-related desynchronization detection Proceedings of International Conference on Computational Intelligence and Vehicular System Seoul, Korea 22–23 September 2010 117 120
Cited by
Telesurgery: Past, Present, and Future
Choi, P. J., Oskouian, R. J., & Tubbs, R. S. (2018) · Cureus · doi:10.7759/cureus.2716 · Cited by 204
Quantitative Assessment of Parkinsonian Tremor Based on an Inertial Measurement Unit
Dai, H., Zhang, P., & Lueth, T. (2015) · Sensors · doi:10.3390/s151025055 · Cited by 159
“Tremor Quantification has been performed by signal processing methods such as time-domain analysis, time-frequency analysis, spectral analysis, and nonlinear analysis [31,32].”
MethodSimultaneous Estimation of Road Profile and Tire Road Friction for Automotive Vehicle
Rath, J. J., Veluvolu, K. C., & Defoort, M. (2014) · IEEE Transactions on Vehicular Technology · doi:10.1109/tvt.2014.2373434 · Cited by 88
“Recently, new methods have been developed to provide driftfree displacement and velocity estimation from acceleration for various applications [24].”
MethodSelf-citationTatinati, S., Veluvolu, K. C., & Ang, W. T. (2015) · IEEE Transactions on Cybernetics · doi:10.1109/tcyb.2014.2381495 · Cited by 77
“The time-frequency characteristics of surgeons and novice subjects were analyzed in [15].”
BackgroundMethodSelf-citationReview of Advanced Medical Telerobots
Mehrdad, S., Liu, F., Pham, M. T., Lelevé, A., & Atashzar, S. F. (2020) · Applied Sciences · doi:10.3390/app11010209 · Cited by 67
“…it can be mentioned that through advanced computer vision and image processing modules (most of which have a Kalman filter core algorithm [11,192,193,195–198]) it is now possible to measure motions of moving organs such as heart and lung during the operation and compensate automatically for…”
Drift-Free Position Estimation of Periodic or Quasi-Periodic Motion Using Inertial Sensors
Latt, W. T., Veluvolu, K. C., & Ang, W. T. (2011) · Sensors · doi:10.3390/s110605931 · Cited by 53
“In physiological tremor sensing for real-time compensation [9,10], zero-phase adaptive filtering algorithms based on truncated Fourier series such as weighted-frequency Fourier linear combiner (WFLC) [11-13] or band-limited multiple Fourier linear combiner (BMFLC) [14-16], which can detect periodic or quasi-periodic s…”
BackgroundMethodSelf-citationPractical Applications in Biomedical Engineering
Andrade, A. O., Pereira, A. A., Naves, E. L. M., & Soares, A. B. (2013) · InTech eBooks · doi:10.5772/3331 · Cited by 54
Multistep Prediction of Physiological Tremor for Surgical Robotics Applications
Veluvolu, K. C., Tatinati, S., Hong, S.-M., & Ang, W. T. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2264546 · Cited by 51
“After this stage, to separate the tremulous motion from the voluntary motion and to remove the unwanted integration drift and noise [30], a fifth-order Butterworth filter with passband 2–20 Hz is employed as shown in the block diagram.”
MethodSelf-citationVoluntary-Driven Elbow Orthosis with Speed-Controlled Tremor Suppression
Herrnstadt, G., & Menon, C. (2016) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2016.00029 · Cited by 49
A Quaternion Weighted Fourier Linear Combiner for Modeling Physiological Tremor
Adhikari, K., Tatinati, S., Ang, W. T., Veluvolu, K. C., & Nazarpour, K. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2530564 · Cited by 48
Self-citationA Multimodal Human–Robot Interface to Drive a Neuroprosthesis for Tremor Management
Gallego, J. Á., Ibanez, J., Dideriksen, J. L., Serrano, J. I., del Castillo, M. D., Farina, D., & Rocon, E. (2012) · IEEE Transactions on Systems, Man, and Cybernetics, Part C (Applications and Reviews) · doi:10.1109/tsmcc.2012.2200101 · Cited by 46
“sors, either accelerometers [18] or gyroscopes [19], or electromyographic recordings [20], although research works on fusion of both technologies have just started appearing [21].”
BackgroundAtashzar, S. F., Shahbazi, M., Samotus, O., Tavakoli, M., Jog, M. S., & Patel, R. V. (2016) · IEEE Journal of Selected Topics in Signal Processing · doi:10.1109/jstsp.2016.2530632 · Cited by 45
Adaptive sliding bandlimited multiple fourier linear combiner for estimation of pathological tremor
Wang, S., Gao, Y., Zhao, J., & Cai, H. (2014) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2013.10.004 · Cited by 42
Characterization of Parkinsonian Hand Tremor and Validation of a High-Order Tremor Estimator
Zhou, Y., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2018) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2018.2859793 · Cited by 42
“Other WFLC derivatives [56], [57] use multiple frequency bands to increase the estimation accuracy.”
BackgroundMachine Learning in Tremor Analysis: Critique and Directions
De, A., Bhatia, K. P., Volkmann, J., Peach, R. L., & Schreglmann, S. R. (2023) · Movement Disorders · doi:10.1002/mds.29376 · Cited by 47
Tatinati, S., Veluvolu, K. C., Hong, S.-M., Latt, W. T., & Ang, W. T. (2013) · IEEE Sensors Journal · doi:10.1109/jsen.2013.2271737 · Cited by 38
“To validate the approach, we compared the estimation accuracy of AR-KF with BMFLC-KF and WFLC-KF methods.”
BackgroundMethodInfluentialSelf-citationShahtalebi, S., Atashzar, S. F., Samotus, O., Patel, R. V., Jog, M. S., & Mohammadi, A. (2020) · Scientific Reports · doi:10.1038/s41598-020-58912-9 · Cited by 32
Sabatini, A. M., Ligorio, G., & Mannini, A. (2015) · BioMedical Engineering OnLine · doi:10.1186/s12938-015-0103-8 · Cited by 33
“In particular, methods to obtain position information from periodic acceleration have been based on analytical integration so as to avoid drift caused by numerical integration [12, 13].”
MethodBonnet, V., Mazzà, C., McCamley, J., & Cappozzo, A. (2013) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-10-29 · Cited by 32
“wbk is introduced in the computation of εk to estimate the bias present in the signal [9-11], due to possible low frequency components and/or drift, and remove it.”
BackgroundMethodShahtalebi, S., Atashzar, S. F., Patel, R. V., & Mohammadi, A. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2018.10.007 · Cited by 29
Real-Time Voluntary Motion Prediction and Parkinson’s Tremor Reduction Using Deep Neural Networks
Ibrahim, A. H., Zhou, Y., Jenkins, M. E., Trejos, A. L., & Naish, M. D. (2021) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2021.3097007 · Cited by 34
Menguc, E. C., cnar, S., Xiang, M., & P. Mandic, D. (2021) · IEEE Signal Processing Letters · doi:10.1109/lsp.2021.3097279 · Cited by 25
Liu, Z., Luo, J., Wang, L., Zhang, Y., Chen, C. L. P., & Chen, X. (2013) · International Journal of Systems Science · doi:10.1080/00207721.2013.821718 · Cited by 26
“…1: Tremor signal with small amplitude and low frequency According to the characteristics of tremor (Elble and Koller 1990; Riviere et al. 2006; Veluvolu and Ang 2011), the desired input signal d(t) and tremor signal n(t) are defined as follows: d(t) = 3.5 sin(12πt) + 1.5 cos(9πt) + 3t4 −…”
BackgroundInfluentialNoncontact Tremor Characterization Using Low-Power Wideband Radar Technology
Blumrosen, G., Uziel, M., Rubinsky, B., & Porrat, D. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2177977 · Cited by 28
“inertial sensors (INS) [5]–[9] optical tracking systems [10]–[14] tracking devices that exploit electromagnetic (EM) radiation properties like [15]–[20].”
BackgroundMethodWang, J., & Barry, O. (2021) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2021.3061361 · Cited by 24
“For active tremor suppression, we designed a band-limited multi-frequency Fourier linear combiner (BMFLC) model, which is directly implemented as an adaptive control reference model for IO-RAC to compensate for the synthetic tremor [9], [30].”
MethodHuman Tremor: Origins, Detection and Quantification
Adriano, O., Alves, A., de Almeida, M. F. S., Lopes, G., da Silva, A. P. S. P. B., Bernardino, S., & Carlos, V. (2013) · InTech eBooks · doi:10.5772/54524 · Cited by 34
“The signal processing and analysis of tremor often involves the spectral analysis, based on Fast Fourier Transform – FFT [35; 39; 41], but this technique as modified weighted Fourier Linear Combiner –WFLC [44] is suitable for periodic or quasi-periodic estimation of motion with single dominant frequency, whereas Band-…”
MethodTremor Assessment during Virtual Reality Brain Tumor Resection
Siyar, S., Azarnoush, H., Rashidi, S., & Maestro, R. F. D. (2019) · Journal of surgical education · doi:10.1016/j.jsurg.2019.11.011 · Cited by 22
Wang, Y., & Veluvolu, K. C. (2017) · Sensors · doi:10.3390/s17061386 · Cited by 26
“Model of Sparse-BMFLC The band-limited multiple Fourier linear combiner (BMFLC) divides the frequency band of interest, [ f1, fM], into M divisions with a fixed frequency gap ∆ f and estimates the signal amplitude according to the formation of the Fourier series [4,21].”
MethodSelf-citationFan, Y., He, Y., & Tan, U.-X. (2020) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2020.2975836 · Cited by 21
“In [18]–[22], weighted-frequency Fourier linear combiner (WFLC) and band-limited multiple linear Fourier combiner (BMFLC) are proposed.”
MethodShah, A., Coste, J., Lemaire, J.-J., Taub, E., Schüpbach, W. M. M., Pollo, C., Schkommodau, E., Guzman, R., & Hemm-Ode, S. (2017) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1559-9 · Cited by 21
Physiological Tremor Filtering Without Phase Distortion for Robotic Microsurgery
Adhikari, K., Tatinati, S., Veluvolu, K. C., & Chambers, J. A. (2020) · IEEE Transactions on Automation Science and Engineering · doi:10.1109/tase.2020.3041427 · Cited by 22
Self-citationHejrati, B., Merryweather, A., & Abbott, J. J. (2017) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2017.2740060 · Cited by 17
“to estimate pathological tremor parameters from gyroscope and accelerometer sensor data [23].”
BackgroundDrift-Free Position Estimation for Periodic Movements Using Inertial Units
Millor, N., Lecumberri, P., Gómez, M., Martínez-Ramírez, A., & Izquierdo, M. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2286697 · Cited by 18
“Through the literature, different methods can be found to partially solve this problem: aided sensors or sensing systems data fusion [11]–[14], wavelet analysis [15], [16], Fourier-based filters [17]–[21], band-pass filtering [22], and polynomial data fitting [23].”
MethodHMFP-DBRNN: Real-Time Hand Motion Filtering and Prediction via Deep Bidirectional RNN
Shahtalebi, S., Atashzar, S. F., Patel, R. V., & Mohammadi, A. (2019) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2019.2894005 · Cited by 17
Classification of Fatigue Phases in Healthy and Diabetic Adults Using Wearable Sensor
Aljihmani, L., Kerdjidj, O., Zhu, Y., Mehta, R. K., Erraguntla, M., Sasangohar, F., & Qaraqe, K. A. (2020) · Sensors · doi:10.3390/s20236897 · Cited by 20
“presented band-limited multiple Fourier linear combiner with Kalman filter and recursive least squares algorithms for recognizing the voluntary and involuntary motion from the accelerometer data in real time [23]; multistep prediction of physiological tremor in real time to reduce the time-varying delay applying a ker…”
BackgroundMethodTremor Estimation and Removal in Robot-Assisted Surgery Using Lie Groups and EKF
Rana, R., Gaur, P., Agarwal, V., & Parthasarathy, H. (2019) · Robotica · doi:10.1017/s0263574719000341 · Cited by 17
Estimation of road profile for suspension systems using adaptive super-twisting observer
Rath, J. J., Veluvolu, K. C., & Defoort, M. (2014) · doi:10.1109/ecc.2014.6862248 · Cited by 17
“There are several works in the literature that employed accelerometers to obtain displacement and velocity for various applications [19].”
BackgroundSelf-citationSang, H., Yang, C., Liu, F., Yun, J., Jin, G., & Chen, F. (2016) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.1741 · Cited by 19
“To verify the performance of each component in the ZPAFKF, simulations were performed using artificial physiological tremor signals n(t) and artificial voluntary movement d(t) and they can be described as follows (10,13):”
BackgroundMethodA novel control architecture for physiological tremor compensation in teleoperated systems
Ghorbanian, A., Zareinejad, M., Rezaei, S. M., Sheikhzadeh, H., & Baghestan, K. (2012) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.1436 · Cited by 17
“5 for Δ fand the band 7–14Hz are optimum for tremor estimation (21).”
BackgroundMethodReal time estimation and suppression of hand tremor for surgical robotic applications
Kumar, A., Kumar, S., Kaushik, A. K., Kumar, A., & Saini, J. S. (2020) · Microsystem Technologies · doi:10.1007/s00542-019-04736-1 · Cited by 17
“For instance, during eye surgery, the tool tip oscillations of the order of 50 lm peak-to-peak (p–p) have been observed during manual vitreoretinal microsurgery, whereas the required accuracy is 10 lm (Veluvolu and Ang 2011; Carignan et al. 2012).”
MethodAtashzar, S. F., Jafari, N., Shahbazi, M., Janz, H. L., Tavakoli, M., Patel, R. V., & Adams, K. D. (2017) · Journal of Medical Robotics Research · doi:10.1142/s2424905x17400013 · Cited by 15
“In the literature, BMFLC fi ltering has also been used for extracting physiological hand tremor of surgeons [36 – 38] and recently has shown good performance in extracting pathological involuntary motions [33].”
MethodEnhancing Human–Robot Collaboration: Supernumerary Robotic Limbs for Object Balance
Luo, J., Liu, S., Si, W., et al. (2024) · IEEE Transactions on Systems Man and Cybernetics Systems · doi:10.1109/tsmc.2024.3501389 · Cited by 13
“It is noted that human tremor brings high-frequency noise into p and θ [54].”
Berbakov, L., Jovanović, Č., Svetel, M. V., Vasiljević, J., Dimić, G., & Radulović, N. (2019) · Sensors · doi:10.3390/s19194246 · Cited by 17
“On the other hand, filtering of tremor signals acquired by accelerometers for real-time applications is presented in [9], where the authors propose two improved algorithms by considering that the tremor contains multiple dominant frequencies over the entire signal duration.”
BackgroundPrediction of Voluntary Motion Using Decomposition-and-Ensemble Framework With Deep Neural Networks
Yoon, D., Kim, E., Choi, I., Han, S. W., & Yang, S. (2020) · IEEE Access · doi:10.1109/access.2020.3035498 · Cited by 12
“The learning model either adaptively updates the time-varying frequencies [15] and amplitudes or update the amplitudes of a fixed set of multiple frequencies [16].”
BackgroundMethodHilbert–Huang transform based instrumental assessment of intention tremor in multiple sclerosis
Carpinella, I., Cattaneo, D., & Ferrarin, M. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/4/046011 · Cited by 12
Understanding Tremor in Rapid Upper Limb Movements Using 3D Accelerometers Data
Pascoal‐Faria, P., Patrício, M., Leal, A., Caramelo, F. J., Freire, A., Januário, C., & Castelo-Branco, M. S. (2014) · Neuroscience & Medicine · doi:10.4236/nm.2014.55024 · Cited by 12
“Each of these can be characterized by different amplitudes and frequencies and may occur in various body parts, where the position of the limbs also plays an important role [2] [6] [7].”
BackgroundPrediction of pathological tremor using adaptive multiple oscillators linear combiner
Xiao, F., Gao, Y., Wang, S., & Zhao, J. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.01.006 · Cited by 11
Online LS-SVM based multi-step prediction of physiological tremor for surgical robotics
Tatinati, S., Wang, Y., Shafiq, G., & Veluvolu, K. C. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610930 · Cited by 11
“To separate the tremulous motion from the sensed motion by accelerometers, bandpass filter was employed in [2], [3].”
MethodInfluentialSelf-citationQuantitative assessment of tremor during deep brain stimulation using a wearable glove system
Dai, H., & D’Angelo, L. (2013) · doi:10.1109/iot-nc.2013.6694054 · Cited by 14
“indicated that the tremor amplitude calculated with PSD correlated well with clinical ratings even for a broad power spectra distribution [24], the power distribution in other frequency points still influences the tremor amplitude calculation in these measurements [20].”
Multibody Analysis and Control of a Full-Wrist Exoskeleton for Tremor Alleviation
Wang, J., & Barry, O. (2020) · Journal of Biomechanical Engineering · doi:10.1115/1.4047424 · Cited by 9
Font, D., Tresánchez, M., Pallejà, T., et al. (2011) · Sensors · doi:10.3390/s111211856 · Cited by 9
“The lowest operational frequency range defined for micro-electro mechanical accelerometers is usually 1 Hz [23-29] so this proposal opens a new range of applications that require the measurement of low frequency oscillations with a low-cost measurement setup.”
BackgroundMulti-step prediction of physiological tremor for robotics applications
Veluvolu, K. C., Tatinati, S., Hong, S. M., & Ang, W. T. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610689 · Cited by 7
“To estimate the filtered tremor signal in real-time, adaptive algorithms based on Fourier series (weighted frequency Fourier linear combiner (WFLC) [3] and band limited multiple Fourier linear combiner (BMFLC)[4]) and Autoregressive (AR) method [5] were developed.”
BackgroundMethodSelf-citationPrediction of Physiological Tremor Based on Deep Learning for Vascular Interventional Surgery Robot
Zhang, L., Guo, S., & Yang, C. W. (2021) · IEEE International Conference on Mechatronics and Automation (ICMA) · doi:10.1109/icma52036.2021.9512713 · Cited by 11
“In order to improve the efficiency of the WFLC algorithm, Veluvolu of Nanyang Technological University and others put forward a new band-limited multiple linear Fourier linear combiner (BMFLC) algorithm after lucubrate the problem about physiological tremor [22][23].”
MethodExploring data-driven modeling and analysis of nonlinear pathological tremors
Wang, J., & Barry, O. (2021) · Mechanical Systems and Signal Processing · doi:10.1016/j.ymssp.2021.107659 · Cited by 7
Tensometric tremorography in high-precision medical diagnostic systems
Aleksanyan, Z. A., Игоревна, Б. О., & Safyannikov, N. M. (2018) · Medical Devices Evidence and Research · doi:10.2147/mder.s168831 · Cited by 7
“Under various pathologies, the picture can change.(24) For example, under tremor form of parkinsonism, the higher level of activity in the whole frequency range is registered, with”
Study on physiological tremor recognition algorithm in the vascular interventional surgical robot
Guo, S., Shen, R., Xiao, N., Bao, X., Yang, C. W., & Cui, J. (2018) · IEEE International Conference on Mechatronics and Automation (ICMA) · doi:10.1109/icma.2018.8484567 · Cited by 9
“The estimation accuracy of MWLSSVMAF, BMFLC-KF [36], WFLC-KF [36] and LS-SVM are (a)”
MethodResultDynamic estimation strategy for E-BMFLC filters in analyzing pathological hand tremors
Ghassab, V. K., Mohammadi, A., Atashzar, S. F., & Patel, R. V. (2017) · IEEE Global Conference on Signal and Information Processing · doi:10.1109/globalsip.2017.8308681 · Cited by 8
Wang, S., Gao, Y., Zhu, Y., & Zhao, J. (2016) · Advances in Mechanical Engineering · doi:10.1177/1687814016654872 · Cited by 6
“In the limitation block, the dominant frequency is derived from the estimated peak frequency as expressed in equation (15)”
Modeling 3D tremor signals with a quaternion weighted Fourier Linear Combiner
Adhikari, K., Tatinati, S., Veluvolu, K. C., & Nazarpour, K. (2015) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2015.7146744 · Cited by 6
“Tremor recordings were performed with the Micro Motion Sensing System [3].”
MethodSelf-citationQuantitative assessment of tremor during deep brain stimulation using a wearable glove system
Dai, H., & D'Angelo, L. T. (2013) · IEEE International Conference on Sensing, Communication, and Networking (SECON) · doi:10.1109/sahcn.2013.6644965 · Cited by 7
“indicated that the tremor amplitude calculated with PSD correlated well with clinical ratings even for a broad power spectra distribution [24], the power distribution in other frequency points still influences the tremor amplitude calculation in these measurements [20].”
Tatinati, S., Veluvolu, K. C., & Ang, W. T. (2011) · International Conference on Control, Automation and Systems (ICCAS) · Cited by 6
Self-citationA hybrid approach for real-time respiratory motion prediction for radiotherapy applications
Rasheed, A., & Veluvolu, K. C. (2025) · Measurement · doi:10.1016/j.measurement.2025.117819 · Cited by 5
Self-citationRasheed, A., Kim, J., Ang, W. T., & Veluvolu, K. C. (2025) · ISA Transactions · doi:10.1016/j.isatra.2024.12.040 · Cited by 6
Self-citationState of the Art and Development Trend of Laparoscopic Surgical Robot and Master Manipulator
Han, R., Sang, H., Liu, F., & Huang, F. (2024) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.70020 · Cited by 8
Wang, W., Jia, B., Ma, J., Wang, X., & Song, H. (2024) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2666 · Cited by 5
“In their experiments, they could basically achieve filtering of the tremor signal, but with a delay of 0.2 s, which is not suitable for real‐time systems such as minimally invasive surgery [16].”
InfluentialFractal measures of video-recorded trajectories can classify motor subtypes in Parkinson’s Disease
Figueiredo, T. C., Costa, J. V., Peña, N., & Miranda, J. G. V. (2016) · Physica A Statistical Mechanics and its Applications · doi:10.1016/j.physa.2016.05.050 · Cited by 5
Aye, Y. N., Zhao, S., & Ang, W. T. (2013) · International Journal of Advanced Robotic Systems · doi:10.5772/56571 · Cited by 5
“where k y denotes the estimated signal at sampling instant k, and rk a and rk b represent the adaptive weights corresponding to the frequency r ω at instant k [25].”
BackgroundSelf-citationTan, G. Y., Bakibillah, A. S. M., Chan, P. Y., Tan, C. P., & Nurzaman, S. G. (2025) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2025.109814 · Cited by 8
Liu, X., Schauer, T., Goldack, A., Bleicher, A., & Schlaich, M. (2016) · Journal of Physics Conference Series · doi:10.1088/1742-6596/744/1/012041 · Cited by 4
Beretta, E., Ferrigno, G., & Momi, E. D. (2016) · Journal of Medical Robotics Research · doi:10.1142/s2424905x1650001x · Cited by 5
Acquisition And Analysis Of Tremor Signals Using Force And Accelerometer Sensors
Hagan, M. (2012) · Zenodo (CERN European Organization for Nuclear Research) · doi:10.5281/zenodo.43222 · Cited by 4
Multiple Order Fourier Linear Combiner: Estimating Phase Shift for Real-Time Vibration Compensation
He, Y., Fan, Y., & Tan, U.-X. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3179703 · Cited by 3
Wang, J., & Barry, O. R. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3111878 · Cited by 3
Bremm, R. P., Werle, A., Auer, C., Hertel, F., Gonçalves, J. M., & Koch, K. P. (2021) · arXiv (Cornell University) · doi:10.48550/arxiv.2108.01736 · Cited by 9
“Moreover, the filtering of physiological tremor is more challenging with the presence of drift, noise, and gravity in acceleration measurements [61], [62].”
BackgroundShahtalebi, S., Mohammadi, A., Atashzar, S. F., & Patel, R. V. (2017) · IEEE Global Conference on Signal and Information Processing · doi:10.1109/globalsip.2017.8308679 · Cited by 3
“Among different solutions [14–20] designed to address the need for online tremor estimation and extraction, adaptive filters developed based on Fourier linear combiners (FLC) and wavelet transforms have shown superior performance.”
BackgroundEstrada, A., Efimov, D. V., & Perruquetti, W. (2016) · International Journal of Control · doi:10.1080/00207179.2016.1175670 · Cited by 3
“While not requiring S1 or S2, methods like those in Peeters (2006) and Guangyu and Shize (2009) are devoted to identify a sinusoidal dominant frequency and the results in Veluvolu and Ang (2011) assumes band limited signals consisting of multiple sinusoidal components.”
MethodShanghavi, A. A., Larrañaga, D., Patil, R., Frazier, E. M., Ambike, S., Duerstock, B. S., & Sereno, A. B. (2024) · Scientific Reports · doi:10.1038/s41598-024-60286-1 · Cited by 3
“Tremor is defined as a rhythmic, involuntary movement of a body part 10 .”
Yang, D., Jiang, L., & Xu, K. (2023) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio58561.2023.10354895 · Cited by 2
“However, the main defect of a low-pass filter is the time delay introduced to the control system, which degrades the performance of the robotic system[8].”
BackgroundXiao, F., Zhong, B., Wu, Y., & Wang, Y. (2022) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2022.3222299 · Cited by 2
“In [37], the authors have discussed the side effects of phase delay on tremor suppression.”
BackgroundModeling of physiological tremor with quaternion variant of extreme learning machines
Tatinati, S., Wang, Y., & Veluvolu, K. C. (2016) · International Conference on Communication and Information Processing (ICCIP) · doi:10.1145/3018009.3018053 · Cited by 2
“Adaptive modeling of physiological tremor based on truncated Fourier series (weighted Fourier linear combiner (WFLC) and Band-limited Fourier linear combiner (BMFLC)) has been developed for hand-held instruments [3].”
BackgroundSelf-citationA frame work for classifying physiological tremor variants employing principal component analysis
Krishna Prasad, S. J., Priyanka, D., & Talasila, V. (2014) · International Conference on Circuits, Control, Communication and Computing (I4C) · doi:10.1109/cimca.2014.7057784 · Cited by 2
“Results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation [2].”
MethodTatinati, S., & Veluvolu, K. C. (2013) · International Conference on Power, Energy and Control · doi:10.1109/icpec.2013.6527703 · Cited by 2
“The time-frequency characteristics of surgeons and novice subjects are analysed in [6].”
BackgroundMethodInfluentialSelf-citationOnline learning for vibration suppression in physical robot interaction using power tools
Solak, G., & Ajoudani, A. (2026) · Autonomous Robots · doi:10.1007/s10514-026-10252-9 · Cited by 4
“We also find multiple methods in FLC-related works that employ RLS [26, 29] or Kalman filter [25, 29–31].”
InfluentialTremor estimation and filtering in robotic-assisted surgery
Jia, B., Wang, W., Tian, X., & Wang, X. (2025) · Cognitive Neurodynamics · doi:10.1007/s11571-025-10387-y · Cited by 1
“As noted in Veluvolu and Ang (2011), a delay of less than 200 ms is sufficient, and in practice, a delay within the range of 100 ms is safer.”
Niță, E. I., Comeaga, D. C., Apostol, D. A., & Duma, V.-F. (2025) · Applied Sciences · doi:10.3390/app15073686 · Cited by 2
Real-Time Vibration Estimation and Compensation With Long Short-Term Memory Recurrent Neural Network
He, Y., Zhang, Y., Fan, Y., & Tan, U.-X. (2024) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2024.3496533 · Cited by 2
“Sharing this idea, the Fourier linear combiners (FLC)-based methods [8], [9] are widely used for signal estimation; while the dynamic harmonics regression method works for delay correction [10].”
Assessment and Elimination of Surgeon Hand Tremor During Robotic Surgery
Kumar, A., Kumar, S., & Kumar, A. (2023) · Springer eBooks · doi:10.1007/978-981-99-2074-7_97 · Cited by 3
Mechatronic Devices for Upper Limb Tremor
Zhou, Y., Daemi, P., Edmonds, B. P. R., Habibollahi, Z., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2023) · Contemporary clinical neuroscience · doi:10.1007/978-3-031-26128-2_22 · Cited by 1
Model Predictive Control for Tremor Suppressing Exoskeleton
Samal, S., & Barry, O. (2023) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2023.12.047 · Cited by 1
Machine Learning for Fatigue Estimation and Prediction “An Introduction Study”
Aljihmani, L., Ursuțiu, D., Samoilă, C., & Qaraqe, K. A. (2022) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-030-93564-1_25 · Cited by 1
Assessment and Elimination of Surgeon Hand Tremor During Robotic Surgery
Kumar, A., Kumar, S., & Kumar, A. (2022) · Springer eBooks · doi:10.1007/978-981-19-1550-5_97-1 · Cited by 4
He, Y., Fan, Y., Zhang, D., & Tan, U.-X. (2018) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio.2018.8665174 · Cited by 1
“[12] proposed band limited multiple linear Fourier combiner (BMFLC) and in [13] the algorithm is developed with Kalman filter (BMFLC-Kalman) and recursive least square (BMFLC-RLS) as updating algorithm.”
MethodWide bandwidth sensing of micro angular motion
Fan, Y., Li, H.-Y., Yang, L., & Tan, U.-X. (2016) · International Conference on Manipulation, Automation and Robotics at Small Scales (MARSS) · doi:10.1109/marss.2016.7561743 · Cited by 1
“To address the drift, WFLC is utilized in [5]–[8].”
MethodDesign and realization of measuring device for tremor evaluation
Havlík, J., Tomáš, S., Nemeckova, M., Vavrova, D., Řasová, K., Zeman, J., & Sovka, P. (2015) · International Workshop on Computational Intelligence for Multimedia Understanding (IWCIM) · doi:10.1109/iwcim.2015.7347079 · Cited by 3
“The similar values are showed also in [13] for physiological tremor or in [10] for Parkinsonian tremor and physiological tremor.”
BackgroundPhysiological Tremor Estimation Methods : An Overview
Naik, N. S., & Dube, R. R. (2014) · Cited by 1
Tremor acquisition and reduction for robotic surgical applications
Gopal, P., Kumar, S., Bachhal, S., et al. (2013) · doi:10.1109/icaes.2013.6659416 · Cited by 1
Fusion of Inertial Measurements and Vision Feedback for Microsurgery
Aye, Y. N., Zhao, S., Shee, C. Y., & Ang, W. T. (2012) · Advances in intelligent systems and computing · doi:10.1007/978-3-642-33932-5_4 · Cited by 2
“Several techniques for online modeling and estimation of tremor have been implemented [6,7,8].”
MethodSelf-citationZhang, T., Sun, C., Zhou, X., Chen, Y., Yue, X. H., & Bi, J. (2026) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.70159 · Cited by 1
A dual-axis MEMS stage for microlens positioning toward optical image stabilization
Jafarsadeghipournaki, I., Michael, A., & Pota, H. R. (2026) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2026.117784
A Dual-Axis Micro-Stage for Optical Image StabilizationPreprint
Jafarsadeghipournaki, I., Michael, A., & Pota, H. R. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.6066103
Metadata
- Authors
- Veluvolu, K. C., & Ang, W. T.
- Affiliations
- Kyungpook National University (KR)
- Nanyang Technological University (SG)
- Corresponding author
- Veluvolu, Kalyana C.
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2011-03-07
- Volume
- 11
- Issue
- 3
- Pages
- 3020-3036
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC3231635
- OpenAlex
- W2071890760
- Cited by
- 117
- References
- 36
- Citations by year
- 2026: 5 · 2025: 8 · 2024: 9 · 2023: 5 · 2022: 5 · 2021: 7 · 2020: 9 · 2019: 5 · 2018: 8 · 2017: 6 · 2016: 15 · 2015: 5 · 2014: 7 · 2013: 14 · 2012: 5
- FWCI
- 4.87
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
- Health Sciences › Medicine › Surgery › Surgical Simulation and Training
- Keywords
- physiological tremor, tremor filtering, accelerometer, real-time signal processing, time-frequency analysis, surgical robotics, drift estimation
- MeSH terms
- Acceleration, Algorithms, Fourier Analysis, Humans, Physicians, Physiology, Time Factors, Tremor, Least-Squares Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.