Textile Electrocardiogram (ECG) Electrodes for Wearable Health Monitoring
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Wearable health-monitoring systems should be comfortable, non-stigmatizing, and able to achieve high data quality. Smart textiles with electronic elements integrated directly into fabrics offer a way to embed sensors into clothing seamlessly to serve these purposes. In this work, we demonstrate the feasibility of electrocardiogram (ECG) monitoring with sewn textile electrodes instead of traditional gel electrodes in a 3-lead, chest-mounted configuration. The textile electrodes are sewn with silver-coated thread in an overlapping zig zag pattern into an inextensible fabric. Sensor validation included ECG monitoring and comfort surveys with human subjects, stretch testing, and wash cycling. The electrodes were tested with the BIOPAC MP160 ECG data acquisition module. Sensors were placed on 8 subjects (5 males and 3 females) with double-sided tape. To detect differences in R peak detectability between traditional and sewn sensors, effect size was set at 10% of a sample mean for heart rate (HR) and R-R interval. Paired student’s t-tests were run between adhesive and sewn electrode data for R-R interval and average HR, and a Wilcoxon signed-rank test was run for comfort. No statistically significant difference was found between the traditional and textile electrodes (R-R interval: t = 1.43, p > 0.1; HR: t = −0.70, p > 0.5; comfort: V = 15, p > 0.5).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Heart rate variability: origins, methods, and interpretive caveats
BERNTSON, G. G., THOMAS BIGGER, J., ECKBERG, D. L., GROSSMAN, P., KAUFMANN, P. G., MALIK, M., NAGARAJA, H. N., PORGES, S. W., SAUL, J. P., STONE, P. H., et al. (1997) · Psychophysiology · doi:10.1111/j.1469-8986.1997.tb02140.x · Cited by 3,962 · cited 2×
“R-R interval is the time between each R peak (the largest positive peak in the signal), and HR is the number of beats per minute (bpm) [6].”
BackgroundDry-Contact and Noncontact Biopotential Electrodes: Methodological Review
Chi, Y. M., Jung, T.-P., & Cauwenberghs, G. (2010) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2010.2084078 · Cited by 1,061 · cited 2×
“…this type of electrode: (1) adhesives can cause discomfort when left on the skin for an extended period of time; (2) the monitoring system is not well concealed and can be obvious to peers; and (3) the conductive gel within the electrodes can dry up over time, impacting the signal quality [8,9].”
BackgroundA Wearable Sensor for Unobtrusive, Long-Term Assessment of Electrodermal Activity
Poh, M.-Z., Swenson, N. C., & Picard, R. W. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2038487 · Cited by 693
Detecting Vital Signs with Wearable Wireless Sensors
Yilmaz, T., Foster, R., & Hao, Y. (2010) · Sensors · doi:10.3390/s101210837 · Cited by 344
Flexible polymeric dry electrodes for the long-term monitoring of ECG
Baek, J.-Y., An, J.-H., Choi, J.-M., Park, K.-S., & Lee, S.-H. (2008) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2007.11.019 · Cited by 340 · cited 2×
“These electrodes can be fabric-based or can utilize other substrates, such as flexible elastomers [10].”
BackgroundMethodEnhancing the Performance of Stretchable Conductors for E‐Textiles by Controlled Ink Permeation
Jin, H., Matsuhisa, N., Lee, S., Abbas, M., Yokota, T., & Someya, T. (2017) · Advanced Materials · doi:10.1002/adma.201605848 · Cited by 283
Performance Evaluation of Sensing Fabrics for Monitoring Physiological and Biomechanical Variables
Scilingo, E. P., Gemignani, A., Paradiso, R., Taccini, N., Ghelarducci, B., & DeRossi, D. (2005) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2005.854506 · Cited by 247 · cited 1×
“Health monitoring using wearable sensor systems is a rapidly expanding area of research that promises an increase in the availability of health data [1–5].”
BackgroundDevendorf, L., Lo, J., Howell, N., Lee, J. L., Gong, N.-W., Karagozler, M. E., Fukuhara, S., Poupyrev, I., Paulos, E., & Ryokai, K. (2016) · Proceedings of the 2016 CHI Conference on Human Factors in Computing Systems · doi:10.1145/2858036.2858192 · Cited by 235 · cited 1×
“The development of “smart” textiles (textiles fabricated with conductive elements directly integrated) is on the rise, creating new opportunities for integrating advanced monitoring technologies into everyday items [11,12].”
BackgroundFabric-Based Wearable Dry Electrodes for Body Surface Biopotential Recording
Yokus, M. A., & Jur, J. S. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2462312 · Cited by 214 · cited 2×
“There are three key challenges that arise with this type of electrode: (1) adhesives can cause discomfort when left on the skin for an extended period of time; (2) the monitoring system is not well concealed and can be obvious to peers; and (3) the conductive gel within the electrodes can dry up over time, impacting t…”
BackgroundMethodA Wearable ECG Acquisition System With Compact Planar-Fashionable Circuit Board-Based Shirt
Yoo, J., Yan, L., Lee, S., Kim, H., & Yoo, H.-J. (2009) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2033053 · Cited by 181
Sensing Fabrics for Monitoring Physiological and Biomechanical Variables: E-textile solutions
Pacelli, M., Loriga, G., Taccini, N., & Paradiso, R. (2006) · 2006 3rd IEEE/EMBS International Summer School on Medical Devices and Biosensors · doi:10.1109/issmdbs.2006.360082 · Cited by 173
ELECTROACTIVE FABRICS AND WEARABLE BIOMONITORING DEVICES
Rossi, D. D., Carpi, F., Lorussi, F., Mazzoldi, A., Paradiso, R., Pasquale Scilingo, E., & Tognetti, A. (2003) · AUTEX Research Journal · doi:10.1515/aut-2003-030404 · Cited by 160 · cited 1×
“The low resistance of these textile electrodes is essential due to the lack of conductive gel to reduce skin/electrode impedance [21].”
BackgroundPerformance Evaluation of Textile-Based Electrodes and Motion Sensors for Smart Clothing
Cho, G., Jeong, K., Paik, M. J., Kwun, Y., & Sung, M. (2011) · IEEE Sensors Journal · doi:10.1109/jsen.2011.2167508 · Cited by 158
Two‐Layered and Stretchable e‐Textile Patches for Wearable Healthcare Electronics
La, T., Qiu, S., Scott, D. K., Bakhtiari, R., Kuziek, J. W. P., Mathewson, K. E., Rieger, J., & Chung, H. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201801033 · Cited by 124 · cited 1×
“Only a handful of the groups listed in Table 1 conducted wash testing, and only one conducted stretch testing [20]; none of these groups have performed all three evaluations (stretch, bend, and wash testing) together.”
BackgroundSkin-Electrode Impedance and Its Effect on Recording Cardiac Potentials
Spach, M. S., Barr, R. C., Havstad, J. W., & Long, E. C. (1966) · Circulation · doi:10.1161/01.cir.34.4.649 · Cited by 120 · cited 1×
“ECG monitoring traditionally relies upon adhesive electrodes stuck on the skin for an extended period of time [7].”
BackgroundTextile Electrodes in ECG Measurement
Pola, T., & Vanhala, J. J. (2007) · International Conference on Intelligent Sensors, Sensor Networks and Information Processing · doi:10.1109/issnip.2007.4496917 · Cited by 119 · cited 1×
“A challenge for all textile electrodes is maintaining skin contact during movement, as shown in the work of Pola and Vanhala [25].”
Kannaian, T., Neelaveni, R., & Thilagavathi, G. (2012) · Journal of Industrial Textiles · doi:10.1177/1528083712438069 · Cited by 96 · cited 1×
“The ultimate goal of these textile electrodes is broad use, and while many previous studies have conducted human subject testing, it is often performed with only one subject [24–26].”
BackgroundAdaCAD: Crafting Software For Smart Textiles Design
Friske, M., Wu, S., & Devendorf, L. (2019) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3290605.3300575 · Cited by 92 · cited 1×
“The development of “smart” textiles (textiles fabricated with conductive elements directly integrated) is on the rise, creating new opportunities for integrating advanced monitoring technologies into everyday items [11,12].”
BackgroundAdvances in textile technologies for unobtrusive monitoring of vital parameters and movements
Paradiso, R., & Rossi, D. D. (2006) · Conference proceedings · doi:10.1109/iembs.2006.259307 · Cited by 83 · cited 1×
“Health monitoring using wearable sensor systems is a rapidly expanding area of research that promises an increase in the availability of health data [1–5].”
BackgroundWearable Measurement of ECG Signals Based on Smart Clothing
Li, M., Xiong, W., & Li, Y. (2020) · International Journal of Telemedicine and Applications · doi:10.1155/2020/6329360 · Cited by 59 · cited 1×
“The ultimate goal of these textile electrodes is broad use, and while many previous studies have conducted human subject testing, it is often performed with only one subject [24–26].”
BackgroundA Wearable Fabric Computer by Planar-Fashionable Circuit Board Technique
Kim, H. J., Kim, Y., Kim, B., & Yoo, H. (2009) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.51 · Cited by 58 · cited 2×
“Conductive inks and pastes tend to be expensive to purchase or manufacture, and the curing process takes more time to be test-ready than sewing [15–18].”
BackgroundPreliminary study of dry knitted fabric electrodes for physiological monitoring
Mestrovic, M. A., Helmer, R. J. N., Kyratzis, I. L., & Kumar, D. (2007) · International Conference on Intelligent Sensors, Sensor Networks and Information Processing · doi:10.1109/issnip.2007.4496911 · Cited by 56 · cited 1×
“They found success with these electrodes, but only tested them on one subject [13].”
BackgroundDevelopment of the Polipo Pressure Sensing System for Dynamic Space-Suited Motion
Anderson, A. P., Mengüç, Y., Wood, R. J., & Newman, D. J. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2449304 · Cited by 30 · cited 1×
“With many methods of flexible electrode fabrication, bending and stretching can cause small breaks in the conductive substrate (whether it be thread or ink), so it is encouraging to see that resistance of the sewn electrodes is impacted neither by these motions nor by the strain of washing [29].”
MethodSelf-citation
Cited by
Advances in healthcare wearable devices
Iqbal, S. M. A., Mahgoub, I., Du, E., Leavitt, M. A., & Asghar, W. (2021) · npj Flexible Electronics · doi:10.1038/s41528-021-00107-x · Cited by 714
“Reproduced with permission from Arquilla et al.25 Copyright © Sensors, 2020. e E-TeCS wearable with sensors on the shirt.”
MethodInfluentialCICIoMT2024: A benchmark dataset for multi-protocol security assessment in IoMT
Dadkhah, S., Neto, E. C. P., Ferreira, R., Molokwu, R. C., Sadeghi, S., & Ghorbani, A. A. (2024) · Internet of Things · doi:10.1016/j.iot.2024.101351 · Cited by 234
Textronics—A Review of Textile‐Based Wearable Electronics
Choudhry, N. A., Arnold, L., Rasheed, A., Khan, I. A., & Wang, L. (2021) · Advanced Engineering Materials · doi:10.1002/adem.202100469 · Cited by 181
Metadata
- Authors
- Arquilla, K., Webb, A., & Anderson, A.
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-02-13
- Volume
- 20
- Issue
- 4
- Pages
- 1013
- ISSN
- 1424-8220 (ISSN-L)
- OpenAlex
- W3006120342
- Cited by
- 165
- References
- 23
- FWCI
- 6.86
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.