Motion artifact reduction in wearable photoplethysmography based on multi-channel sensors with multiple wavelengths.
Open accessAbstract
Photoplethysmography (PPG) is an easy and convenient method by which to measure heart rate (HR). However, PPG signals that optically measure volumetric changes in blood are not robust to motion artifacts. In this paper, we develop a PPG measuring system based on multi-channel sensors with multiple wavelengths and propose a motion artifact reduction algorithm using independent component analysis (ICA). We also propose a truncated singular value decomposition for 12-channel PPG signals, which contain direction and depth information measured using the developed multi-channel PPG measurement system. The performance of the proposed method is evaluated against the R-peaks of an electrocardiogram in terms of sensitivity (Se), positive predictive value (PPV), and failed detection rate (FDR). The experimental results show that Se, PPV, and FDR were 99%, 99.55%, and 0.45% for walking, 96.28%, 99.24%, and 0.77% for fast walking, and 82.49%, 99.83%, and 0.17% for running, respectively. The evaluation shows that the proposed method is effective in reducing errors in HR estimation from PPG signals with motion artifacts in intensive motion situations such as fast walking and running.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Our approximation of wear time was based on available per-minute HR values, which may be an imperfect proxy because wearable device data are affected by motion artifacts.
“First, our approximation of wear time was based on available per-minute HR values, which have been used in previous studies22; however, wearable device data are affected by motion artifacts.64,65 Thus, our measure of wear time may be an imperfect proxy.”
From Association of Demographic and Socioeconomic Indicators With the Use of Wearable Devices Among Children (2023) · cites this paper as [65]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · see citing works
Photophysical processes in recent medical laser developments: A review
Boulnois, J.-L. (1986) · Lasers in Medical Science · doi:10.1007/bf02030737 · Cited by 663 · see citing works
Multiparameter Respiratory Rate Estimation From the Photoplethysmogram
Karlen, W., Raman, S., Ansermino, J. M., & Dumont, G. A. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2246160 · Cited by 486 · see citing works
RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345 · see citing works
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226 · see citing works
Motion-tolerant magnetic earring sensor and wireless earpiece for wearable photoplethysmography
Poh, M., Swenson, N. C., & Picard, R. W. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2010.2042607 · Cited by 189 · see citing works
Reddy, K. A., George, B., & Kumar, V. J. (2008) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2008.2009136 · Cited by 165 · see citing works
A Real Time Analysis of PPG Signal for Measurement of SpO2 and Pulse Rate
Bagha, S., & Shaw, L. (2011) · International Journal of Computer Applications · Cited by 143 · see citing works
Motion Artifact Reduction for Wrist-Worn Photoplethysmograph Sensors Based on Different Wavelengths
Zhang, Y., Song, S., Vullings, R., Biswas, D., Simões-Capela, N., van Helleputte, N., van Hoof, C., & Groenendaal, W. (2019) · Sensors · doi:10.3390/s19030673 · Cited by 140 · see citing works
Chawla, M. P. S. (2010) · Applied Soft Computing · doi:10.1016/j.asoc.2010.08.001 · Cited by 127 · see citing works
Noninvasive cardiac output estimation using a novel photoplethysmogram index
Wang, L., Pickwell‐MacPherson, E., Liang, Y. P., & Zhang, Y. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5333091 · Cited by 117 · see citing works
The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals
Lee, H.-W., Lee, J., Jung, W.-G., & Lee, G.-K. (2007) · International Journal of Control Automation and Systems · Cited by 114 · see citing works
Self-citationSpīgulis, J., Gailite, L., Lihachev, A., & Erts, R. (2007) · Applied Optics · doi:10.1364/ao.46.001754 · Cited by 98 · see citing works
Design and development of non invasive glucose measurement system
Paul, B., Manuel, M. P., & Alex, Z. C. (2012) · International Symposium on Physics and Technology of Sensors · doi:10.1109/ispts.2012.6260873 · Cited by 63 · see citing works
Lee, H., Chung, H., Ko, H., & Lee, J. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2801385 · Cited by 61 · see citing works
Self-citationRam, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., Sivani, K., & Reddy, K. A. (2013) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2013.2259114 · Cited by 52 · see citing works
Stress measurement from wearable photoplethysmographic sensor using heart rate variability data
Mohan, P. M., Nagarajan, V., & Das, S. R. (2016) · doi:10.1109/iccsp.2016.7754331 · Cited by 52 · see citing works
Warren, K., Harvey, J. S., Chon, K. H., & Mendelson, Y. (2016) · Sensors · doi:10.3390/s16030342 · Cited by 47 · see citing works
Comparison of Green, Blue and Infrared Light in Wrist and Forehead Photoplethysmography
Vizbara, V. (2013) · Cited by 40 · see citing works
Wang, L., Poon, C. C. Y., & Zhang, Y. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/5/008 · Cited by 33 · see citing works
Effects of motion artifact on the blood oxygen saturation estimate in pulse oximetry
Clarke, G. W. J., Chan, A. D. C., & Adler, A. (2014) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2014.6860071 · Cited by 31 · see citing works
A Pulse Rate Detection Method for Mouse Application Based on Multi-PPG Sensors
Lin, S.-T., Chen, W., & Lin, Y. (2017) · Sensors · doi:10.3390/s17071628 · Cited by 31 · see citing works
Validation du Sport Tester PE 3000 en fonction de l'enregistrement Holter
Vogelaere, P. P., Meyer, F. D., Duquet, W., & Vandevelde, P. (1987) · Science & Sports · doi:10.1016/s0765-1597(87)80024-2 · Cited by 21 · see citing works
Use of Multi-Scale Principal Component Analysis for motion artifact reduction of PPG signals
Ram, M. R., Madhav, K. V., Krishna, E. H., Reddy, K. N., & Reddy, K. A. (2011) · IEEE Recent Advances in Intelligent Computational Systems · doi:10.1109/raics.2011.6069348 · Cited by 17 · see citing works
SISTEMIC, F. D. I., Rojano, J. F., & Isaza, C. (2016) · International Journal of Signal Processing Systems · doi:10.18178/ijsps.4.6.475-482 · Cited by 15 · see citing works
Design of Filter to Reject Motion Artifacts of PPG Signal Using Multiwave Optical Source
Hee-Jung, P., Nam, J., & Lee, J. (2014) · Journal of the Korea Society of Computer and Information · doi:10.9708/jksci.2014.19.2.101 · Cited by 2 · see citing works
Self-citationMulti-modal Wearable Device for Cardiac Arrest Detection
Jun, A. H., You, S. M., Cho, K., Park, H.-K., & Kim, J. Y. (2017) · Journal of Biomedical Engineering Research · Cited by 1 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Photoplethysmogram Analysis and Applications: An Integrative Review
Park, J., Seok, H. S., Kim, S.-S., & Shin, H. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2021.808451 · Cited by 450 · see citing works
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Kim, K. B., & Baek, H. J. (2023) · Electronics · doi:10.3390/electronics12132923 · Cited by 238 · see citing works
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 210 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Martinez-Ríos, E. A., Montesinos, L., Alfaro-Ponce, M., & Pecchia, L. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102813 · Cited by 147 · see citing works
Scardulla, F., Cosoli, G., Spinsante, S., Poli, A., Iadarola, G., Pernice, R., Busacca, A., Pasta, S., Scalise, L., & D’Acquisto, L. (2023) · Measurement · doi:10.1016/j.measurement.2023.113150 · Cited by 120 · see citing works
The Current State of Optical Sensors in Medical Wearables
Vavrinsky, E., Esfahani, N. E., Hausner, M., Kuzma, A., Rezo, V., Donoval, M., & Kosnacova, H. (2022) · Biosensors · doi:10.3390/bios12040217 · Cited by 103 · see citing works
“Wrist-worn reflectance LED Motion artifact reduction PP: Four x OSRAM SFH7050, Motion Sensor: InvenSense MPU9250 Wavelengths 530, 660, 940 nm [75]”
BackgroundStretchable PPG sensor with light polarization for physical activity–permissible monitoring
Lee, G. H., Kang, H., Chung, J. W., Lee, Y., Yoo, H., Jeong, S., Cho, H. K., Kim, J., Kang, S., Jung, J. Y., et al. (2022) · Science Advances · doi:10.1126/sciadv.abm3622 · Cited by 95 · see citing works
Self-citationTrovato, V., Sfameni, S., Rando, G., Rosace, G., Libertino, S., Ferri, A., & Plutino, M. R. (2022) · Molecules · doi:10.3390/molecules27175709 · Cited by 91 · see citing works
Pankaj, Kumar, A., Komaragiri, R., & Kumar, M. (2021) · Archives of Computational Methods in Engineering · doi:10.1007/s11831-021-09597-4 · Cited by 84 · see citing works
Ultralow-Power Photoplethysmography (PPG) Sensors: A Methodological Review
Ebrahimi, Z., & Gosselin, B. J. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3284818 · Cited by 78 · see citing works
Ueafuea, K., Boonnag, C., Sudhawiyangkul, T., Leelaarporn, P., Gulistan, A., Chen, W., Mukhopadhyay, S. C., Wilaiprasitporn, T., & Piyayotai, S. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3046259 · Cited by 72 · see citing works
Saganowski, S. (2022) · Electronics · doi:10.3390/electronics11030496 · Cited by 65 · see citing works
Kim, E. H., Jenness, J. L., Miller, A. B., Halabi, R., de Zambotti, M., Bagot, K. S., Baker, F. C., & Pratap, A. (2023) · JAMA Network Open · doi:10.1001/jamanetworkopen.2023.5681 · Cited by 53 · see citing works
Li, J., Chu, H., Chen, Z., Yiu, C. K., Qu, Q., Li, Z.-Y., & Yu, X. (2024) · ACS Nano · doi:10.1021/acsnano.4c04291 · Cited by 50 · see citing works
Moscato, S., Giudice, S. L., Massaro, G., & Chiari, L. (2022) · Sensors · doi:10.3390/s22155831 · Cited by 51 · see citing works
Shi, Z., Gu, T., Zhang, Y., & Zhang, X. (2022) · ACM Conference on Embedded Networked Sensor Systems (SenSys) · doi:10.1145/3560905.3568506 · Cited by 48 · see citing works
Pankaj, Kumar, A., Komaragiri, R., & Kumar, M. (2021) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2021.105081 · Cited by 42 · see citing works
Detection and Monitoring of Viral Infections via Wearable Devices and Biometric Data
Goergen, C. J., Tweardy, M. J., Steinhubl, S. R., Wegerich, S., Singh, K., Mieloszyk, R. J., & Dunn, J. (2021) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-103020-040136 · Cited by 41 · see citing works
Zeng, T., Wu, Y., & Lei, M. (2024) · Advanced Composites and Hybrid Materials · doi:10.1007/s42114-024-00949-9 · Cited by 40 · see citing works
State-of-the-art wearable sensors for cardiovascular health: a review
Xie, H., Yang, L., Jiang, B., Huang, Z., & Lin, Y. (2025) · npj Cardiovascular Health · doi:10.1038/s44325-025-00090-6 · Cited by 44 · see citing works
Choi, S., Gao, Y., Jin, Y., Kim, S. J., Li, J., Xu, W., & Jin, Z. (2022) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3534597 · Cited by 34 · see citing works
A Dual-Channel PPG Readout System With Motion-Tolerant Adaptability for OLED-OPD Sensors
Pandey, R. K., & Chao, P. C.-P. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3138996 · Cited by 32 · see citing works
Reddy, G. N. K., Manikandan, M. S., & Murty, N. V. L. N. (2022) · IEEE Access · doi:10.1109/access.2022.3148256 · Cited by 34 · see citing works
A review: Blood pressure monitoring based on PPG and circadian rhythm
Chen, G., Zou, L., & Ji, Z. (2024) · APL Bioengineering · doi:10.1063/5.0206980 · Cited by 34 · see citing works
Application of Modern Multi-Sensor Holter in Diagnosis and Treatment
Vavrinský, E., Subjak, J., Donoval, M., Wagner, A., Zavodnik, T., & Svobodová, H. (2020) · Sensors · doi:10.3390/s20092663 · Cited by 32 · see citing works
Iqbal, S. M. A., Mahgoub, I., Du, E., Leavitt, M. A., & Asghar, W. (2022) · Scientific Reports · doi:10.1038/s41598-022-23680-1 · Cited by 30 · see citing works
Singh, S., Kozłowski, M., García-López, I., Jiang, Z., & Rodriguez–Villegas, E. (2021) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2021.3083415 · Cited by 28 · see citing works
Gautam, Y., & Jebelli, H. (2024) · Automation in Construction · doi:10.1016/j.autcon.2024.105537 · Cited by 27 · see citing works
PPG and Bioimpedance-Based Wearable Applications in Heart Rate Monitoring—A Comprehensive Review
Lapsa, D., Janeliukštis, R., Metshein, M., & Selavo, L. (2024) · Applied Sciences · doi:10.3390/app14177451 · Cited by 30 · see citing works
Uncertainties in the Analysis of Heart Rate Variability: A Systematic Review
Lü, L., Zhu, T., Morelli, D., Creagh, A. P., Liu, Z., Yang, J., Liu, F., Zhang, Y., & Clifton, D. A. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2023.3271595 · Cited by 28 · see citing works
Park, P., Lee, W., & Cho, S. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3315297 · Cited by 23 · see citing works
Fazio, R. D., Vittorio, M. D., & Visconti, P. (2022) · Future Internet · doi:10.3390/fi14060183 · Cited by 19 · see citing works
Reali, P., Lolatto, R., Coelli, S., Tartaglia, G., & Bianchi, A. M. (2022) · Sensors · doi:10.3390/s22041428 · Cited by 18 · see citing works
Seamlessly Integrable Optoelectronics for Clinical Grade Wearables
Polat, E. O. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202000853 · Cited by 18 · see citing works
Multi-point sensing organic light-emitting diode display based mobile cardiovascular monitor
Kim, C., Lee, K. J., Kim, J., Yang, D., Lee, H., Moon, G., Kim, Y., Cho, D., Bae, K. S., Kim, G.-H., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-56915-6 · Cited by 17 · see citing works
Huang, K., Ma, Z., & Khoo, B. L. (2025) · Advanced Science · doi:10.1002/advs.202415215 · Cited by 23 · see citing works
Two-Stage Motion Artifact Reduction Algorithm for rPPG Signals Obtained from Facial Video Recordings
Abdulrahaman, L. Q. (2023) · Arabian Journal for Science and Engineering · doi:10.1007/s13369-023-07845-2 · Cited by 17 · see citing works
An ML-Based Approach to Reconstruct Heart Rate from PPG in Presence of Motion Artifacts
Vicente-Samper, J. M., Tamantini, C., Ávila-Navarro, E., de la Casa, M. A., Zollo, L., Sabater-Navarro, J. M., & Cordella, F. (2023) · Biosensors · doi:10.3390/bios13070718 · Cited by 17 · see citing works
Thakur, S., Chao, P. C.-P., & Tsai, C.-H. (2023) · Sensors · doi:10.3390/s23136180 · Cited by 16 · see citing works
Ahn, J., Ra, H.-K., Yoon, H. J., Son, S. H., & Ko, J. (2020) · IEEE Access · doi:10.1109/access.2020.3025776 · Cited by 16 · see citing works
Pankaj, Kumar, A., Ashdhir, A., Komaragiri, R., & Kumar, M. (2022) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2022.107294 · Cited by 16 · see citing works
Sawangjai, P., Seesawad, N., & Wilaiprasitporn, T. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3529546 · Cited by 12 · see citing works
Tsiakaka, O., Gosselin, B. J., & Feruglio, S. (2020) · Sensors · doi:10.3390/s20113302 · Cited by 12 · see citing works
Measuring Arterial Oxygen Saturation Using Wearable Devices Under Varying Conditions
Hearn, E. L., Byford, J., Wolfe, C., Agyei, C., Hodkinson, P. D., Pollock, R. D., & Smith, T. G. (2023) · Aerospace Medicine and Human Performance · doi:10.3357/amhp.6078.2023 · Cited by 12 · see citing works
vital_sqi: A Python package for physiological signal quality control
Khoa, L. D. V., Ho, H. B., Karolcik, S., Hernandez, B., Greeff, H., Nguyen, V. H., Phan, N. Q. K., Le, T. P., Thwaites, L., Georgiou, P., et al. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.1020458 · Cited by 11 · see citing works
Morresi, N., Casaccia, S., & Revel, G. M. (2021) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea52024.2021.9478713 · Cited by 11 · see citing works
Tunable Multi-Channels Bandpass InGaAsP Plasmonic Filter Using Coupled Arrow Shape Cavities
Abd-Elsamee, S., Areed, N. F. F., El-Mikati, H. A., & Obayya, S. S. A. (2022) · Photonics · doi:10.3390/photonics9100720 · Cited by 11 · see citing works
de Pedro-Carracedo, J., Ugena, A. M., & Gonzalez-Marcos, A. P. (2021) · Applied Sciences · doi:10.3390/app11146508 · Cited by 11 · see citing works
Machine Learning-Based VO2 Estimation Using a Wearable Multiwavelength Photoplethysmography Device
Hsiao, C., Tong, C. W.-C., & Coté, G. L. (2025) · Biosensors · doi:10.3390/bios15040208 · Cited by 10 · see citing works
PPG Earring: Wireless Smart Earring for Heart Health Monitoring
Xue, Q. S., Nissanka, D., Yan, J. T., Wang, R., Patel, S., & Iyer, V. (2025) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3706598.3713856 · Cited by 10 · see citing works
Generative adversarial networks with fully connected layers to denoise PPG signals
Castro, I. A. A., Oliveira, H. P., Correia, R. G., Hayes‐Gill, B., Morgan, S. P., Korposh, S., Gomez, D., & Pereira, T. (2025) · Physiological Measurement · doi:10.1088/1361-6579/ada9c1 · Cited by 9 · see citing works
Merschel, S., & Reinhardt, L. (2022) · JMIR Formative Research · doi:10.2196/29479 · Cited by 8 · see citing works
Bradke, B. S., Miller, T. A., & Everman, B. (2021) · Sensors · doi:10.3390/s21134543 · Cited by 9 · see citing works
Posada-Quintero, H. F., Derrick, B. J., Winstead‐Derlega, C., Gonzalez, S. I., Ellis, M. C., Freiberger, J. J., & Chon, K. H. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9629924 · Cited by 9 · see citing works
The importance of physiological data variability in wearable devices for digital health applications
Cosoli, G., Poli, A., Spinsante, S., & Scalise, L. (2022) · ACTA IMEKO · doi:10.21014/acta_imeko.v11i2.1135 · Cited by 9 · see citing works
Performance Analysis for Human Crowd Monitoring to Control COVID-19 disease by Drone Surveillance
Conte, C., de Alteriis, G., de Pandi, F., Caputo, E., Moriello, R. S. L., Rufino, G., & Accardo, D. (2021) · IEEE International Workshop on Metrology for AeroSpace (MetroAeroSpace) · doi:10.1109/metroaerospace51421.2021.9511671 · Cited by 9 · see citing works
Chen, K., Tseng, C.-H., Lee, H., Liu, W., Chou, K., & Wu, H. (2025) · SLEEP · doi:10.1093/sleep/zsae317 · Cited by 10 · see citing works
Zhang, Y., Zhu, H., Liu, H., Zheng, D., Zhang, S., & Pan, Y. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871902 · Cited by 8 · see citing works
Deep Learning Model for Blood Pressure Estimation from PPG Signal
Kim, M., Lee, H., Kim, K., & Kim, K. H. (2022) · IEEE International Conference on Metrology for eXtended Reality, Artificial Intelligence and Neural Engineering (MetroXRAINE) · doi:10.1109/metroxraine54828.2022.9967606 · Cited by 8 · see citing works
Self-citationZhu, W., Xiang, L., Long, Y., Xun, Q., Kuang, J.-L., & He, L. (2022) · Sleep Medicine · doi:10.1016/j.sleep.2022.12.023 · Cited by 8 · see citing works
Mascret, Q., Gurve, D., Abdou, A., Bhadra, S., Lasry, N., Mai, K., Krishnan, S., & Gosselin, B. J. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3388950 · Cited by 8 · see citing works
A Power-Efficient Sensing Approach for Pulse Wave Palpation-Based Heart Rate Measurement
Bravo, G., Silva, J. M., Morales, S. A. N., Gómez, E. A. M., Aguilera, F. J. E., & Sifuentes, E. (2021) · Sensors · doi:10.3390/s21227549 · Cited by 8 · see citing works
He, Q., Sun, Z., Li, Y., Wang, W., & Wang, R. K. (2021) · Biomedical Optics Express · doi:10.1364/boe.423160 · Cited by 9 · see citing works
Wearable Sensors for Monitoring Abnormalities in Sleep-Related Breathing
Wang, J., Xu, J., Ma, J., Gong, J., Ji, L., Guo, W., Yue, H., Qiao, Y., & Zhou, J. (2026) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c22504 · Cited by 7 · see citing works
Nantume, A., Kiwanuka, N., Muyinda, A., Cauvel, T., & Shah, S. (2022) · Digital Health · doi:10.1177/20552076221102262 · Cited by 9 · see citing works
Pandey, R. K., & Chao, P. C.-P. (2021) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas51556.2021.9401265 · Cited by 7 · see citing works
An Effective Photoplethysmography Denosing Method Based on Diffusion Probabilistic Model
Xia, Z., Luo, Z., Chen, C., & Shen, X. (2025) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2025.3530517 · Cited by 6 · see citing works
mmBP+: Contact-Free Blood Pressure Measurement Using Millimeter-Wave Radar
Shi, Z., Gu, T., Zhang, Y., & Zhang, X. (2025) · IEEE Transactions on Mobile Computing · doi:10.1109/tmc.2025.3594316 · Cited by 6 · see citing works
Artificial Intelligence-Driven Wearable Sensors for Cardiovascular Health Monitoring
Deng, H. T., Ma, B., Chen, G., Chi, J., Tong, J., Yu, F., Lü, B., Chen, Y., Zhao, C., Zhang, C., et al. (2026) · ACS Sensors · doi:10.1021/acssensors.5c04840 · Cited by 6 · see citing works
Principal component analysis of photoplethysmography signals for improved gesture recognition
Ruan, Y., Chen, X. '., Zhang, X., & Chen, X. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.1047070 · Cited by 6 · see citing works
Hwang, C., Kim, Y. H., Hyun, J. K., Kim, J. H., Lee, S. R., Kim, C. M., Nam, J., & Kim, E. Y. (2023) · Bioengineering · doi:10.3390/bioengineering10101222 · Cited by 6 · see citing works
Development and Feasibility Assessment of Virtual Reality-Based Relaxation Self-Training Program
Jeong, H., Oh, J., Paik, M., Kim, H., Jang, S., Kim, B. S., & Kim, J.-J. (2022) · Frontiers in Virtual Reality · doi:10.3389/frvir.2021.722558 · Cited by 6 · see citing works
Continuous Blood Pressure Estimation in Wearable Devices Using Photoplethysmography: A Review
Medina, A., Colque, N. L., Galdos, J., Supo, E., Rendulich, J., & Espinoza, E. S. (2022) · International Journal of Emerging Technology and Advanced Engineering · doi:10.46338/ijetae1022_12 · Cited by 9 · see citing works
Artifacts classification and apnea events detection in neck photoplethysmography signals
García-López, I., Pramono, R. X. A., & Rodriguez–Villegas, E. (2022) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02666-1 · Cited by 6 · see citing works
Micromotion Artefact Reduction of a Wrist Worn PPG Sensor Using Green Light PPG and Surface EMG
Silverio, A. A., Asilo, J. E., Balagat, S. L. D., Pineda, A. A., Tangi, J. A., Tuazon, D. T. P., & Piedra, B. M. B. (2020) · IEEE Region Humanitarian Technology Conference · doi:10.1109/r10-htc49770.2020.9356956 · Cited by 6 · see citing works
Nawar, A., Kilic, O. S., Rahman, F. N., Chen, C., Berkebile, J. A., Chan, M., Shah, A. J., & Inan, O. T. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3596522 · Cited by 5 · see citing works
PPGANet: Removal of Motion Artifacts from the PPG Signal Using Generative Adversarial Networks
Sawangjai, P., & Wilaiprasitporn, T. (2023) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas58349.2023.10388620 · Cited by 5 · see citing works
Intraoperative Optical Monitoring of Spinal Cord Hemodynamics Using Multiwavelength Imaging System
Mainard, N., Tsiakaka, O., Li, S., Denoulet, J., Messaoudene, K., Vialle, R., & Feruglio, S. (2022) · Sensors · doi:10.3390/s22103840 · Cited by 5 · see citing works
Regalia, G., Resnati, D., & Tognetti, S. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-822548-6.00130-8 · Cited by 5 · see citing works
Posada-Quintero, H. F., Derrick, B. J., Ellis, M. C., Natoli, M. J., Winstead‐Derlega, C., Gonzalez, S. I., Allen, C. M., Makowski, M. S., Keuski, B. M., Moon, R. E., et al. (2024) · Communications Medicine · doi:10.1038/s43856-024-00688-4 · Cited by 7 · see citing works
Mobile and Wireless Autofluorescence Detection Systems and Their Application for Skin Tissues
Wang, Y., Zhang, Y., Li, Y., & Cai, F. (2025) · Biosensors · doi:10.3390/bios15080501 · Cited by 4 · see citing works
Mbihi, J., Otam, U. S., Lonla, B. M., & Ngounou, C. E. R. G. (2020) · Journal of Electrical Engineering, Electronics, Control and Computer Science · Cited by 4 · see citing works
Photoplethysmogram analysis and applications: An Integrative Review (Preprint)
Shin, H., Park, J., Seok, H. S., & Kim, S. (2020) · JMIR Biomedical Engineering · doi:10.2196/25567 · Cited by 4 · see citing works
Ahmed, M. M., Othman, Z. K., Mahamed, S. M., Okesanya, O. J., Musa, S. S., Branda, F., & Lucero‐Prisno, D. E. (2025) · Digital Health · doi:10.1177/20552076251396563 · Cited by 3 · see citing works
Advancing entropy analysis for heart rate variability: clinical insights for aging and diabetes
Yang, S., Wu, H., Liao, X., Lin, Y., & Liou, J. J. (2026) · Frontiers in Physiology · doi:10.3389/fphys.2026.1795118 · Cited by 3 · see citing works
Time Series Analysis of Muscle Deformation During Physiotherapy Using Optical Wearable Sensors
Shimabukuro, S., Miyake, T., & Tamaki, E. (2025) · Sensors · doi:10.3390/s25113507 · Cited by 3 · see citing works
Edge computing design space exploration for heart rate monitoring
Miranda, J., Canabal, M. F., Gutiérrez-Martín, L., Lanza-Gutiérrez, J. M., & López-Ongil, C. (2022) · Integration · doi:10.1016/j.vlsi.2022.02.003 · Cited by 3 · see citing works
A Robust Beat-to-Beat Artifact Detection Algorithm for Pulse Wave
Hu, Q., Deng, X., Liu, X., Wang, A., & Yang, C. (2020) · Mathematical Problems in Engineering · doi:10.1155/2020/5691805 · Cited by 3 · see citing works
Research progress on emotion recognition based on electroencephalogram signals
Li, X., Yan, Z., Gong, P., Lin, D., & Chen, F. (2025) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2025.109188 · Cited by 2 · see citing works
Moghaddam, M., Collins, J. P., Gardner, C. E., & Rabel, M. C. (2025) · Sensors · doi:10.3390/s26010176 · Cited by 3 · see citing works
Fotiadou, E., Melaet, R., & Vullings, R. (2022) · Frontiers in Signal Processing · doi:10.3389/frsip.2022.981453 · Cited by 2 · see citing works
Blood Oxygen (SpO2) and Pulse Rate(PR) Wearable Sensor for Hemodialysis Patients
Liou, J., Lin, Z., & Hsieh, Z.-X. (2023) · International Conference on Applied System Innovation · doi:10.1109/icasi57738.2023.10179568 · Cited by 2 · see citing works
Botrugno, C., & Dell’Olio, F. (2024) · IEEE Sensors Applications Symposium (SAS) · doi:10.1109/sas60918.2024.10636647 · Cited by 2 · see citing works
Special Issue “Advanced Signal Processing in Wearable Sensors for Health Monitoring”
Abbod, M., & Shieh, J.-S. (2022) · Sensors · doi:10.3390/s22062189 · Cited by 2 · see citing works
Low-Noise and Cost-Effective Active Electrodes for Dry Contact ECG Applications
Voropai, A., & Sarana, V. (2022) · Science and innovation · doi:10.15407/scine18.01.112 · Cited by 1 · see citing works
80‐1: Sensor OLED Display‐Based Mobile Cardiovascular Health Monitor
Kim, C., Yang, D., Moon, G., Kim, G., Kim, Y., & Lee, C. (2025) · SID Symposium Digest of Technical Papers · doi:10.1002/sdtp.18365 · Cited by 1 · see citing works
Representation biases in remote photoplethysmography datasets – a narrative review
Khan, S. S., & Mostafa, N. (2025) · Applied Computing and Informatics · doi:10.1108/aci-06-2025-0270 · Cited by 1 · see citing works
FlexiMA: Robust Motion Artifact Removal in PPG Signals Using an Adversarial Variational Autoencoder
Liu, Q., Zou, H., Wu, Y., Chen, Y., Ma, H., & Wang, B. (2026) · IEEE Internet of Things Journal · doi:10.1109/jiot.2025.3650539 · Cited by 1 · see citing works
- Authors
- Lee, J., Kim, M., Park, H.-K., & Kim, I. Y.
- Affiliations
- Hanyang University (KR)
- Corresponding author
- Park, Hoon-Ki
- Kim, In Young
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-03-09
- Volume
- 20
- Issue
- 5
- Pages
- 1493
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC7085621
- OpenAlex
- W3009321689
- Cited by
- 153
- References
- 29
- Citations by year
- 2026: 34 · 2025: 27 · 2024: 13 · 2023: 15 · 2022: 35 · 2021: 19 · 2020: 10
- FWCI
- 6.20
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Research Foundation of Korea (2018R1A5A7025522)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- photoplethysmography, motion artifact reduction, wearable sensors, multi-channel sensor, multiple wavelengths, independent component analysis, truncated singular value decomposition, heart rate estimation, R-peak detection, electrocardiogram, walking and running
- MeSH terms
- Wearable Electronic Devices, Adult, Electrocardiography, Equipment Design, Female, Heart Rate, Humans, Male, Signal Processing, Computer-Assisted, Walking, Artifacts, Photoplethysmography, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review