Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women
Open accessAbstract
Heart rate variability (HRV) is frequently applied in sport-specific settings. The rising use of freely accessible applications for its recording requires validation processes to ensure accurate data. It is the aim of this study to compare the HRV data obtained by the Polar H10 sensor chest strap device and an electrocardiogram (ECG) with the focus on RR intervals and short-term scaling exponent alpha 1 of Detrended Fluctuation Analysis (DFA a1) as non-linear metric of HRV analysis. A group of 25 participants performed an exhaustive cycling ramp with measurements of HRV with both recording systems. Average time between heartbeats (RR), heart rate (HR) and DFA a1 were recorded before (PRE), during, and after (POST) the exercise test. High correlations were found for the resting conditions (PRE: r = 0.95, rc = 0.95, ICC3,1 = 0.95, POST: r = 0.86, rc = 0.84, ICC3,1 = 0.85) and for the incremental exercise (r > 0.93, rc > 0.93, ICC3,1 > 0.93). While PRE and POST comparisons revealed no differences, significant bias could be found during the exercise test for all variables (p < 0.001). For RR and HR, bias and limits of agreement (LoA) in the Bland–Altman analysis were minimal (RR: bias of 0.7 to 0.4 ms with LoA of 4.3 to −2.8 ms during low intensity and 1.3 to −0.5 ms during high intensity, HR: bias of −0.1 to −0.2 ms with LoA of 0.3 to −0.5 ms during low intensity and 0.4 to −0.7 ms during high intensity). DFA a1 showed wider bias and LoAs (bias of 0.9 to 8.6% with LoA of 11.6 to −9.9% during low intensity and 58.1 to −40.9% during high intensity). Linear HRV measurements derived from the Polar H10 chest strap device show strong agreement and small bias compared with ECG recordings and can be recommended for practitioners. However, with respect to DFA a1, values in the uncorrelated range and during higher exercise intensities tend to elicit higher bias and wider LoA.
Overview
AI snapshot
Written from the paper’s full text.
TL;DR
In 25 recreational men and women, the Polar H10 chest strap with Elite HRV showed near-perfect agreement with a 12-channel ECG for RR and HR at rest and during incremental cycling, with small bias despite significant exercise differences. DFA a1 agreement was high but weaker, with wider bias and limits of agreement, especially at higher exercise intensities.
Key findings
- Linear HRV measures showed near-perfect agreement: PRE/POST RR and HR had r = 1.00, rc = 1.00, ICC3,1 = 1.00, and incremental exercise also had r = 1.00, rc = 1.00, ICC3,1 = 1.00. Resting PRE/POST comparisons showed no significant differences, while exercise comparisons showed significant bias for all variables (p < 0.001).
- DFA a1 correlations were high but comparatively weaker: PRE/POST r > 0.86, rc > 0.84, ICC3,1 > 0.85; incremental exercise r > 0.93, rc > 0.93, ICC3,1 > 0.93, with significant differences during exercise (p < 0.001).
- DFA a1 bias and limits of agreement were wider than for linear measures: PRE bias −2.1%, POST bias 2.5%; during exercise, low intensity bias 0.9% with LoA 11.6 to −9.9%, and high intensity bias 8.6% with LoA 58.1 to −40.9%.
- DFA a1 index values were approximately 6% and 7.5% higher for Polar H10 at fixed values of 0.75 and 0.5 compared with ECG, so threshold boundaries were crossed at higher HRs when using Polar H10.
Methods
- Cross-sectional validation study: 25 participants (men n = 14, age 40 ± 14 years; women n = 11, age 34 ± 10 years) performed an exhaustive cycling ramp with simultaneous Polar H10/Elite HRV and 12-channel ECG recordings before (PRE), during, and after (POST) exercise.
- RR, HR, and DFA a1 were analyzed in Kubios HRV Premium 3.5.0; resting analyses used a 2-min window, exercise used a 2-min moving window with 20-s grid, and datasets with artefacts >5% were excluded.
- Agreement was assessed with linear regression, Pearson’s r, Lin’s concordance correlation coefficient, ICC3,1, R², SEE, Bland–Altman LoA, paired t-test or Wilcoxon signed ranks test, and Cohen’s d, with significance at p ≤ 0.05.
Limitations
- Stated by the paper: the study did not group samples by individual personal characteristics; future studies could examine whether body composition, gender, age, or ventricular characteristics influence outcomes.
- Stated by the paper: the Polar H10 stores RR intervals but not ECG waveforms, so excessive artefacts cannot be identified or corrected, and noise cannot be differentiated from cardiac arrhythmia; five participants were excluded from exercise analysis and two from resting POST measurement due to artefacts >5%.
- Stated by the paper: raw data were analyzed with Kubios HRV Premium using automatic artefact correction, and the free standard Kubios version with threshold-based correction could yield slightly different results; the text also makes evident that device and lead placement differences prevent attributing discrepancies to a single variable.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 of 18 cited sources checked · 0 supported · 0 flagged
The paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“However, the amount of deviation was variable, with some participants showing very close agreement between devices.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesThe paper states this without a source or data behind it here, so there is nothing to check it against. The quote shows it is in the paper.
In the paper
“Both threshold boundaries were crossed at higher HRs when using the Polar H10”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“and a bias of 0.2 to 0.0 ms with upper and lower LoA of 1.6 to 0.5 ms and −1.2 to −0.6 ms in POST”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The comparisons during resting state conditions PRE and POST showed nearly perfect correlations for the linear parameters RR and HR (r = 1.00, rc = 1.00, ICC3,1 = 1.00).”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“With respect to DFA a1 there was bias of −2.1% (upper LoA of 14.7 to 16.4% and lower LoA of −18.9 to −20.5%)”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Bland–Altman analysis revealed a mean difference of 0.2 to 0.1 ms for RR with upper LoA of 2.6 to 1.3 ms and lower LoA of −2.2 to −1.1 ms in PRE”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Index values tended to be approximately 6% and 7.5% higher at the fixed values of 0.75 and 0.5 for Polar H10 in comparison to the ECG (Figure 5C)”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There were no significant differences for both conditions and the three variables of interest (p > 0.05) (Table 1).”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesA result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“and 2.5% for POST (upper LoA of 15.6 to 23.7% and lower LoA of −10.2 to −19.1%)”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIn the paper
“There were no differences observed in the results of the Polar H10 in comparison to the ECG when Kubios HRV software was used to obtain HRV data.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [10] Zabala, 2022, Validity and Reliability of Different Smartphones Applications to Measure HRV during Short and Ultra-Short Measurements in Elite Athletes, Comput. Methods Programs Biomed. source
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passes
- It may state two separate claims as one. 11
- Probably not a substantial claim. 22
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“DFA a1 showed wider bias (0.9 to 8.6%) and LoAs of 11.6 to −9.9% during low intensity and 58.1 to −40.9% during high intensity.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.2. Incremental Exercise Analysis
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Figure 7 illustrates this with an example of a female participant from the present study whose data during the incremental cycling ramp was excluded from validation analysis since artifact exceeded >5%. On closer examination of the parallel recorded ECG, these “artifacts” were classified as frequent atrial premature complexes (APCs)”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Tests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Although correlations were also high for DFA a1 (r > 0.86, rc > 0.84, ICC3,1 > 0.85), they were comparatively weaker”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.1. PRE and POST Analysis
Strength:plain· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Bland–Altman plot for RR showed a mean difference of 0.7 to 0.4 ms with upper and lower LoA of 4.3 to −2.8 ms during low intensity and 1.3 to −0.5 ms during high intensity exercise.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.2. Incremental Exercise Analysis
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“High correlations could be found during the incremental exercise test for the linear parameters (r = 1.00. rc = 1.00, ICC3,1 = 1.00), and DFA a1 (r > 0.93. rc > 0.93, ICC3,1 > 0.93), although with a statistically significant difference (p < 0.001)”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results > 3.2. Incremental Exercise Analysis
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The effects of high degrees of artifact correction are demonstrated, showing either data loss or HRV bias.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Tests✓ Asserted: passes✓ Checkable: passes? Load-bearing: unclear✓ Based: passesIt may state two separate claims as one.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Five participants were excluded from exercise analysis and two participants from resting POST measurement due to artefacts >5%.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 3. Results
Tests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
None of the sources cited for this claim have been checked yet.
In the paper
“recent studies show the potential of this parameter to demarcate physiological threshold boundaries by means of fixed DFA a1 values at the aerobic (~0.75) and anaerobic threshold (~0.5) for usage in performance monitoring and exercise intensity distribution in endurance sports [16,17].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [16] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Detection of the Anaerobic Threshold in Endurance Sports: Validation of a New Method Using Correlation Properties of Heart Rate Variability. J. Funct. Morphol. Kinesiol., 6. source
- Not checked yetcites [17] Mateo-March, M., Moya-Ramón, M., Javaloyes, A., Sánchez-Muñoz, C., and Clemente-Suárez, V.J. (2022). Validity of Detrended Fluctuation Analysis of Heart Rate Variability to Determine Intensity Thresholds in Professional Cyclists. Eur. J. Sport Sci., 1–20. source
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
In the paper
“The gold standard measure for the quantification of RR intervals and assessment of HRV is the electrocardiogram (ECG) [7].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [7] Sammito, 2016, Reference Values for Time- and Frequency-Domain Heart Rate Variability Measures, Heart Rhythm. source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
In the paper
“In this report, there was a substantially better signal quality during high-intensity activities for the Polar H10 in comparison to a 3-lead ECG Holter monitor”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [13] Schweizer, 2019, RR Interval Signal Quality of a Heart Rate Monitor and an ECG Holter at Rest and during Exercise, Eur. J. Appl. Physiol. source
Tests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“However, it has become common practice to use mobile systems (e.g., apps, wearables) and chest straps for data recording, storage, analysis and/or export which offers superior practicability with respect to cost, ease to use, portability and interpretation [8,9,10].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [8] Perrotta, 2017, Validity of the Elite HRV Smartphone Application for Examining Heart Rate Variability in a Field-Based Setting, J. Strength Cond. Res. source
- Not checked yetcites [9] Dobbs, 2019, The Accuracy of Acquiring Heart Rate Variability from Portable Devices: A Systematic Review and Meta-Analysis, Sports Med. source
- Not checked yetcites [10] Zabala, 2022, Validity and Reliability of Different Smartphones Applications to Measure HRV during Short and Ultra-Short Measurements in Elite Athletes, Comput. Methods Programs Biomed. source
Strength:plain· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. It may be too vague to check against evidence. The paper’s conclusions may not depend on it.
None of the sources cited for this claim have been checked yet.
In the paper
“The non-linear short-term scaling exponent alpha 1 of Detrended Fluctuation Analysis (DFA a1) has shown its suitability to describe the complex cardiac autonomic regulation during various exercise intensities, modalities, and environmental conditions while possessing a wide dynamic range throughout all intensity zones in contrast to standard linear parameters [14,15].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [14] Gronwald, 2020, Correlation Properties of Heart Rate Variability during Endurance Exercise: A Systematic Review, Ann. Noninvasive Electrocardiol. source
- Not checked yetcites [15] Rogers, 2022, Fractal Correlation Properties of Heart Rate Variability as a Biomarker for Intensity Distribution and Training Prescription in Endurance Exercise: An Update, Front. Physiol. source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may state two separate claims as one.
In the paper
“Additionally, exercise associated DFA a1 dynamics may also give insights about physiological status in terms of monitoring of fatigue during low-intensity exercise [18].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 1. Introduction
What the sources say
- Not checked yetcites [18] Rogers, 2021, Fractal Correlation Properties of Heart Rate Variability as a Biomarker of Endurance Exercise Fatigue in Ultramarathon Runners, Physiol. Rep. source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“The basic, cost-free “standard” version of Kubios HRV with an alternate threshold-based artifact correction method could lead to slightly different results than presented [24].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
What the sources say
- Not checked yetcites [24] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Influence of Artefact Correction and Recording Device Type on the Practical Application of a Non-Linear Heart Rate Variability Biomarker for Aerobic Threshold Determination. Sensors, 21. source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Hence, it is certainly plausible that some participants had either different or sub optimal signal strength using a chest strap device as opposed to ECG lead 2.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“It can be considered as a possible limitation that this study did not group the samples by their individual personal characteristics.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Strength:hedged· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
In the paper
“Data sets with artefacts >5% were excluded from analysis based on [24].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.3. Data Processing
What the sources say
- Not checked yetcites [24] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Influence of Artefact Correction and Recording Device Type on the Practical Application of a Non-Linear Heart Rate Variability Biomarker for Aerobic Threshold Determination. Sensors, 21. source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“Prior to testing, the gas analyzers were calibrated according to the manufacturer’s instructions.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.2. Exercise Protocol and Data Acquisition
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The potential effect of lead placement (chest strap electrodes vs. ECG lead 2) is also notable in terms of DFA a1 determination.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
Strength:hedged· not yet calibratedCausal:correlational· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✗ Based: failsProbably not a substantial claim. The paper’s conclusions may not depend on it. The paper may give no reason why it holds.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“Participants were asked to abstain from caffeine, alcohol, tobacco, and vigorous exercise 24 h before testing”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.1. Participants
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Nevertheless, since this mobile-based HRV recording setup displays superior practicability with generally comparable results, its commercial use for the monitoring of HRV data during resting and endurance exercise conditions could be justified.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 5. Conclusions
Load-bearingStrength:hedged· not yet calibratedTests✓ Asserted: passes✗ Checkable: fails✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim. It may be too vague to check against evidence.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Linear HRV measurements derived from the Polar H10 chest strap device show strong agreement and small bias compared with ECG recordings and can be recommended for practitioners.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > Abstract
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesIt may state two separate claims as one.
A conclusion the paper draws from its own results — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“This may partly be related to the expected differences in non-linear HRV associated with ECG lead placement.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 5. Conclusions
Tests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✗ Based: failsProbably not a substantial claim. The paper may give no reason why it holds.
In the paper
“For the 12-channel ECG, lead 2 was used as the comparable lead to the chest strap device [21].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.3. Data Processing
What the sources say
- Not checked yetcites [21] Jeyhani, V., Mantysalo, M., Noponen, K., Seppanen, T., and Vehkaoja, A. (2019, January 23–27). Effect of Different ECG Leads on Estimated R–R Intervals and Heart Rate Variability Parameters. Proceedings of the 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany. source
Load-bearingStrength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✓ Load-bearing: passes✓ Based: passesProbably not a substantial claim.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“provided written informed consent.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.1. Participants
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“Entry criteria included adults (>18 years) of either sex and of any fitness level without previous medical history, current medications or recent illness.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.1. Participants
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
What the paper did — it cites no source for this, so there is none to check. The quote shows it is in the paper.
In the paper
“Ethical approval for the study was obtained by the University of Hamburg, Department of Psychology and Movement Science, Germany (reference no.: 2021_400) and is in line with the principles of the Declaration of Helsinki.”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 2. Materials and Methods > 2.1. Participants
Strength:plain· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it. It may state two separate claims as one.
In the paper
“The degree of bias may simply relate to the discordance between actual ECG waveform signal strength and morphology [21].”
§ Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women > 4. Discussions
What the sources say
- Not checked yetcites [21] Jeyhani, V., Mantysalo, M., Noponen, K., Seppanen, T., and Vehkaoja, A. (2019, January 23–27). Effect of Different ECG Leads on Estimated R–R Intervals and Heart Rate Variability Parameters. Proceedings of the 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany. source
Strength:hedged· not yet calibratedCausal:conditional· not yet calibratedTests✓ Asserted: passes✓ Checkable: passes✗ Load-bearing: fails✓ Based: passesProbably not a substantial claim. The paper’s conclusions may not depend on it.
The paper relies on this without support.
To date no validation has been conducted for non-linear measures of HRV.
“In addition to the requirement to perform population-based validations, to date no validation has been conducted for non-linear measures of HRV.”
The paper relies on this without support.
Researchers and recreational athletes frequently use the Polar H10 device for data collection and rely on appropriate measurements.
“since researchers and recreational athletes frequently use this device for data collection and therefore rely on appropriate measurements.”
The paper relies on this without support.
The Polar H10 only stores RR intervals and not ECG waveform recordings.
“Since the Polar H10 only stores RR intervals and not ECG waveform recordings”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used, two modalities that are known to correlate well [13].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [13]
In the citing paper
“As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used, two modalities that are known to correlate well [13].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [13]
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Quantification of scaling exponents and crossover phenomena in nonstationary heartbeat time series
Peng, C., Havlin, S., Stanley, E., & Goldberger, A. L. (1995) · Chaos An Interdisciplinary Journal of Nonlinear Science · doi:10.1063/1.166141 · Cited by 3,865
Age-predicted maximal heart rate revisited
Tanaka, H., Monahan, K. D., & Seals, D. R. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01054-8 · Cited by 3,775 · cited 1×
“The following criteria served for the assumption of exhaustion: (A) heart rate >90% of the maximum predicted heart rate (prediction model according to [20]: 208 − (0.”
BackgroundKubios HRV–heart rate variability analysis software.
Tarvainen, M. P., Niskanen, J.-P., Lipponen, J. A., Ranta-aho, P. O., & Karjalainen, P. A. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2013.07.024 · Cited by 2,720 · cited 4×
“0 (Biosignal Analysis and Medical Imaging Group, Department of Physics, University of Kuopio, Kuopio, Finland, [19]) was conducted.”
BackgroundMethodInfluentialEstimates of Location Based on Rank Tests
Hodges, J. L., & Lehmann, E. L. (1963) · The Annals of Mathematical Statistics · doi:10.1214/aoms/1177704172 · Cited by 1,276
Biostatistics 104: correlational analysis.
Chan, Y. H. (2003) · PubMed · Cited by 1,223
Monitoring training status with HR measures: do all roads lead to Rome?
Buchheit, M. (2014) · Frontiers in Physiology · doi:10.3389/fphys.2014.00073 · Cited by 1,016
Cardiac Parasympathetic Reactivation Following Exercise: Implications for Training Prescription
Stanley, J., Peake, J. M., & Buchheit, M. (2013) · Sports Medicine · doi:10.1007/s40279-013-0083-4 · Cited by 551
Flagged first, then by citations
1 of the 38 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Items 201–300 of 373
Accuracy of Polar Wrist-Worn Devices Compared to Chest-Worn Devices: A Bland-Altman Meta-Analysis
Choe, J.-P., Oh, S.-E., & Kang, M. (2026) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2026.2658553 · Cited by 1
Diaw, M., Connes, P., Charlot, K., Diop, S., Gallou‐Guyot, M., Coly, M. S., Carin, R., Ducray, M., Gadji, M., Miyachi, M., et al. (2026) · European Journal Of Haematology · doi:10.1111/ejh.70201 · Cited by 1
Rho, G., Carbonaro, N., Laurino, M., Tognetti, A., & Greco, A. (2026) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2026.110531 · Cited by 1
Cross-Cultural Adaptation and Validation of the Arabic Rating-of-Fatigue Scale
Baccouche, M. A., Trabelsi, K., Masmoudi, L., Jahrami, H., Ammar, A., & Chtourou, H. (2026) · Sports Medicine - Open · doi:10.1186/s40798-026-01052-7 · Cited by 1
Kamiya, Y., Saita, K., Kaneko, F., & Okamura, H. (2026) · Mental Health & Prevention · doi:10.1016/j.mhp.2026.200529 · Cited by 1
Kerkeni, M., Ammar, A., Boukhris, O. F., Hbaieb, E., Kerkeni, M., Miśkiewicz, P., Chtourou, H., Aziz, A. R., Jahrami, H., Turki, M., et al. (2026) · Journal of Sports Sciences · doi:10.1080/02640414.2026.2702738 · Cited by 1
Hagner–Derengowska, M., Michalska, A., Ardelean, V. P., Geantă, V. A., & Trybulski, R. (2026) · European Journal of Applied Physiology · doi:10.1007/s00421-026-06379-3 · Cited by 1
Čabarkapa, D., Yang, Y., Montalvo, S., Ríos-Gallardo, P. T., Camacho, A. A., & Munce, T. A. (2026) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000005616 · Cited by 1
DARE to wear: digital health promotion and disease prevention using wearable devices
D’Ascanio, I., Moscato, S., Sanabria, J. L. A., Bacalini, M. G., Bellafiore, M., Belli, L., Berti, L., Bhuyan, M. J., Cainelli, E., Cappon, G., et al. (2026) · npj Digital Medicine · doi:10.1038/s41746-026-03230-z · Cited by 1
Diciolla, N. S., Rebelo, P. F. S., Rodrigues, G., Grave, A. S., Dias, C., Gomes, M. D. C. V. A., Santos, E., Pereira, Z., Pereira, L., & Marques, A. S. P. D. D. (2024) · Heart & Lung · doi:10.1016/j.hrtlng.2024.04.014 · Cited by 1
Alessandria, M., Angilletta, S., Pivetta, I., Annone, B., Cravanzola, S., & Giorgio, A. D. (2024) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2024.1424756 · Cited by 2
Yuan, K., OKUYAMA, T., Hortobágyi, T., Nagatomi, R., & Négyesi, J. (2024) · Scientific Reports · doi:10.1038/s41598-024-67467-y · Cited by 1
Jeevanandam, S., Tabango, M. W., Wang, X., & Jain, N. (2024) · Proceedings of the Human Factors and Ergonomics Society Annual Meeting · doi:10.1177/10711813241261345 · Cited by 1
Theobald, P., Herold, F., Gronwald, T., & Müller, N. G. (2024) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-024-00985-4 · Cited by 1
Self-citationBesson, C., Stefani, G. D., Baggish, A. L., Schmitt, L., Millet, G. P., & Grémeaux, V. (2024) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2024-002292 · Cited by 1
Design of IoT-Based Digital Medical Record Devices in Medically Underserved Areas
Wirayudha, I. A., Taqwa, A., & Handayani, A. S. (2025) · Jurnal Teknologi Informasi dan Pendidikan · doi:10.24036/jtip.v17i2.875 · Cited by 1
The Heart of Effort: Revealing Heart Rate Patterns in Real-Effort Tasks
Greif-Winzrieth, A., Dorner, V., Könemann, J., & Fellner-Röhling, G. (2025) · Lecture notes in information systems and organisation · doi:10.1007/978-3-031-71385-9_25 · Cited by 1
de Almeida‐Neto, P. F., Alves, J. C. M., de Assis, G. G., Rocha, R. F. C., de Oliveira, F. C. S., da Silva, I. M., Ferreira, A. B. D. M., Rocha, M. D. L., Aidar, F. J., Dantas, P. M. S., et al. (2025) · Human Movement · doi:10.5114/hm/199539 · Cited by 1
Boot, M., Jutte, A., Ulak, M. B., & Geurs, K. (2025) · SSRN Electronic Journal · doi:10.2139/ssrn.5344791 · Cited by 1
Zhang, Y. (2026) · European Journal of Preventive Cardiology · doi:10.1093/eurjpc/zwag300 · Cited by 1
Torres-Herrera, C., Silva-Tetlak, M., Zarate-Vidal, L., Burrows-Rojo, D., Vega-Zambrano, M. I., & Gutiérrez-Arias, R. E. (2023) · Current Respiratory Medicine Reviews · doi:10.2174/1573398x19666230117123608
VidyaMedic: Smart Alert System For Continous Vital Sign Monitoring
Ikawati, V., & Prihatmanto, A. S. (2023) · doi:10.1109/iceei59426.2023.10346844
DECODIFICAÇÃO E SISTEMA IHM DE PROCESSAMENTO DE BIOSSINAIS UTILIZANDO MICROCONTROLADORES
BOTTON, L. F. D. S. D. (2023) · doi:10.17771/pucrio.acad.65481
Analysis of heart rate and scores in student archery athletes
Oktafiranda, N. D., Ihsani, S. I., & Kuswahyudi, K. (2023) · Jurnal MAENPO Jurnal Pendidikan Jasmani Kesehatan dan Rekreasi · doi:10.35194/jm.v13i2.3728
Iani, C., Bori, L., & Rubichi, S. (2024) · Proceedings of the Human Factors and Ergonomics Society Annual Meeting · doi:10.1177/10711813241277537
Dias, K. J., Rhyne, R. S., Callahan, D. M., Durbin, K. N., Kasten, A. R., Heitzman, J., & Nadler, D. R. (2024) · International journal of exercise science · doi:10.70252/pyaw2914
GASTEL, A. V., HERMANS, K., Vleugels, J., & Verwulgen, S. (2024) · 3DBODY.TECH journal · doi:10.15221/24.53
Pilot Psychophysiological Study of the Co-watcher Effect
Nazarov, N., Silantieva, O., Brueva, A., Zaladina, A., & Демарева, В. (2024) · Studies in computational intelligence · doi:10.1007/978-3-031-76516-2_28
Norton, J., Geraghty, A., Howie, E. E., Boyle, C., Yule, S., & Tambyraja, A. L. (2024) · Journal of Surgical Protocols and Research Methodologies · doi:10.1093/jsprm/snae016
Akyıldız, Z., Şentürk, D., KİRİŞÇİ, İ., Kabadayi, Ş., & Dişçeken, O. (2024) · Spor ve Performans Araştırmaları Dergisi · doi:10.17155/omuspd.1525108
Jackson, J. E., Cox, A. M., Windmill, C., & Safari, R. (2025) · International Journal of Mental Health Nursing · doi:10.1111/inm.13499
Vecchio, L. D., Climstein, M., & Beavers, R. (2025) · Movement & Sport Sciences - Science & Motricité · doi:10.1051/sm/2024033
Latino, F., Limone, P., Cassese, F. P., Martínez-Roig, R., Amato, G., & Tafuri, F. (2025) · Retos · doi:10.47197/retos.v63.111368
Suggestions for APA Research and Translational Perspective
Cascella, M. (2024) · Springer eBooks · doi:10.1007/978-3-031-78833-8_5
Jauvin, F., Massaro, S., Coulombe, P., Brouillette, J., & Forest, J. (2025) · Motivation and Emotion · doi:10.1007/s11031-025-10120-6 · Cited by 1
THE RELATIONSHIP BETWEEN HEART RATE VARIABILITY AND ATHLETIC PERFORMANCE IN BASKETBALL PLAYERS
Karahan, B., Aslan, H., Uysal, E., & Kılıç, R. T. (2025) · Türk fizyoterapi ve rehabilitasyon dergisi · doi:10.21653/tjpr.1471373
Dunton, G. F., & Belcher, B. R. (2025) · British Journal of Sports Medicine · doi:10.1136/bjsports-2024-108363
Assessment of Heart Rate Variability Among Plant Nursery Workers
Thamsuwan, O., & Cherik, K. (2025) · Journal of Agromedicine · doi:10.1080/1059924x.2025.2517845
MCI detection using transformers for EEG-HRV fusionPreprint
Boudaya, A., Chaabene, S., Bouaziz, B., & Chaâri, L. (2025) · Research Square · doi:10.21203/rs.3.rs-6893576/v1
WiCG: Heartbeat Sensing Using COTS WiFi Devices with Common Antenna
Zhang, Y., Liu, Z., Wu, C., Li, J., & Tang, S. (2025) · ACM Transactions on Sensor Networks · doi:10.1145/3748330
de Medeiros, J. A., Borges, U., Wilde, P., de Medeiros, R. C. D. S. C., Alves, J. C. M., Neto, A. G. D. M., Jaggers, J. R., Machado, D. G. D. S., dos Santos, R. V. T., & Dantas, P. M. S. (2025) · PLoS ONE · doi:10.1371/journal.pone.0326793
Kuchapin, M., Шевченко, О. О., Shyrokopetleva, M. S., Kotelnikov, R. V., & Filipov, O. O. (2025) · Scientific Works of Vinnytsia National Technical University · doi:10.31649/2307-5376-2025-2-83-92
Agarao, K. A., Mesa, E. D. D., Gomez, I. N., Mendinueto, A. P. R., Ng, A. M., Agustin, B. T., Kim, M., Baetiong, S. A., Rafael, R. C., & Taguibao, J. G. (2025) · doi:10.36413/pjahs.0901.005
Ramos, R. K., Cosep, L. S., Gomez, I. N., Pesayco, E. E., Guzman, L. L. D., Goco, G. A., Velasquez, V. N., Penafiel, R. L., Lava, Y. I. C., & Villasis, J. (2025) · doi:10.36413/pjahs.0901.006
Khushhal, A. A. (2025) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.25.16270-1
Compère, V. N., Balci, S., Heinz, C., Schade‐Brittinger, C., Rastan, A. J., Salzmann-Djufri, M., Niemann, B., Boening, A., Choi, Y., Zarski, A., et al. (2025) · BMJ Open · doi:10.1136/bmjopen-2024-092482
Ferreira, V. N., Bell-Mizori, A. R., Domogalla, S. A., Serpeloni, F., Tokovská, M., Koebach, A., Abreu, L., & Oliveira, N. C. D. S. (2025) · Research Square · doi:10.21203/rs.3.rs-6779026/v1
Jiu-jitsu-specific performance test induces neuromuscular fatigue in experienced fighters
Correia, B. P., de Oliveira, D. C. X., Bertochi, G. F. A., Souza, L. M. V., da Mota, G. R., & dos Santos, I. A. (2025) · journal of physical education · doi:10.4025/jphyseduc.v36i1.3617
Güzel, H. Ç., Türkmen, O. B., Mutlu, E. K., Türkmen, G., Şakar, O., & Örmeci, G. Ö. (2025) · Journal of Oral Rehabilitation · doi:10.1111/joor.70096
Laulan, P., & Rimmele, U. (2025) · doi:10.22541/au.176492367.71651719/v1
Sathirapanya, P., Limotai, N., Suanprasert, N., Rujirussawarawong, S., Kongkamol, C., Cheetanom, N., & Limotai, C. (2025) · Parkinsonism & Related Disorders · doi:10.1016/j.parkreldis.2025.108146
Boyle, C., Norton, J., Kelleher, M., Howie, E. E., Wigmore, S. J., Skipworth, R. J. E., & Yule, S. (2025) · Journal of Surgical Protocols and Research Methodologies · doi:10.1093/jsprm/snaf011
Syari'ah, K. S. N., Pitriani, P., & Novian, G. (2025) · Journal of Sport Science and Fitness · doi:10.15294/jssf.v11i1.26238
Lopes, T. D. S., Lopes, L. C. D., Antunes, T. P. C., Fernandes, I. R., Morimoto, M., Pinheiro, G., de Camargo, E. D. L. B., Nonato, B. F., Cantisani, M. S., Casali, K. R., et al. (2025) · European Journal of Neuroscience · doi:10.1111/ejn.70370
When Effort Feels Different: Physiological Evidence of Affective Variation in Real Effort Tasks
Dorner, V., Fellner-Röhling, G., Ferlini, G. A., Fink, N. S., Goldmann, J., Greif-Winzrieth, A., Schranzova, S., Tatar, C.-I., & Tomek, R. (2025) · Lecture notes in information systems and organisation · doi:10.1007/978-3-032-00815-2_25
Heart Matters: The Influence of Foreign Biofeedback on Decision-Making in Economic Games
Angerer, T. A., Brawenz, S., Dorner, V., Fellner-Röhling, G., Greif-Winzrieth, A., Müller, M., & Pluy, V. (2025) · Lecture notes in information systems and organisation · doi:10.1007/978-3-032-00815-2_8
Biofeedback-Enhanced Virtual Reality Training for Stress Regulation: Concept and Pilot Study Results
Schwarz, J., Kelter, M., & Baljan, M. (2026) · Lecture notes in computer science · doi:10.1007/978-3-032-13174-4_18
Composite Filtering Methods for Millimeter-Wave Radar-Based Heart Rate Monitoring
Wang, C.-H., Lin, C.-H., & Chen, C.-W. (2025) · International Conference on Intelligent Informatics and Biomedical Sciences (ICIIBMS) · doi:10.1109/iciibms66230.2025.11316706
ECG-based People Identification Across Emotions, Experiments and Days
Meng, Y., Huang, Z., Guo, A., & Ma, J. (2025) · IEEE Intl Conf on Dependable, Autonomic and Secure Computing, Intl Conf on Pervasive Intelligence and Computing, Intl Conf on Cloud and Big Data Computing, Intl Conf on Cyber Science and Technology Congress (DASC/PiCom/CBDCom/CyberSciTech) · doi:10.1109/cyberscitech68397.2025.00088
Wilde, P., de Queirós, V. S., Jaggers, J. R., Sabag, A., Alves, J. M., Fernandes, E., Rocha, R. F. C., de Almeida‐Neto, P. F., dos Santos, R. V. T., & Dantas, P. M. S. (2026) · HIV Medicine · doi:10.1111/hiv.70193
Laiti, J., Javadpour, S., Mullen, J. M., Dunne, S. F., Mundadan, M., O’Conghaile, C., Conway, C., Byrne, E., & Dunne, P. J. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.6123413
Edwards, A. M., Wong, A. L. Y., Fenghua, S. U. N., Chow, C. C., & Kuo, C.-H. (2026) · doi:10.22541/au.176967223.31854792/v1
Implementation of a Gamified Mechatronics System for Human Locomotion Exercises
Koçak, M., Köprülüoğlu, M., Gürşan, İ. N., & Gezgin, E. (2026) · Mechanisms and machine science · doi:10.1007/978-3-032-14023-4_28
Khan, J. A., Ali, U., Ahsan, M. T., Roy, R. C., Alamu, O. S., & Longa, F. A. (2026) · Cureus · doi:10.7759/cureus.102943
Maeseneer, J. D., Olieslagers, A., Gronwald, T., & de Beukelaar, T. T. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.02.02.703217
Self-citationOu, Z., Yang, L., Zhu, H., Weng, X., & Xu, G. (2026) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-026-01602-2
Thakkar, K., & Shah, S. (2025) · Parul University Journal of Health Sciences and Research · doi:10.4103/pujhsr.pujhsr_20_25
The Effects of Acute Exercise on Memory: Considerations of the Backward Testing Effect
Patrick, Z., DuBois, S. L., Etnier, J. L., & Loprinzi, P. D. (2026) · Journal of Cognitive Enhancement · doi:10.1007/s41465-026-00349-5
Arayshi, M. E., Vitarelli, M., Laterza, F., Marziale, A., Iellamo, F., Manzi, V., Caminiti, G., Verrelli, C. M., & Volterrani, M. (2026) · Discover Applied Sciences · doi:10.1007/s42452-026-08459-7
Doliny, A. E. F., Paladini, L. H., Siega, J., Filho, C. A. R., & Israel, V. L. (2026) · Physical & Occupational Therapy In Geriatrics · doi:10.1080/02703181.2026.2637742
Bonnabesse, A. L. F., Exposto, F. G., Bodere, C., & Svensson, P. (2026) · European Journal of Pain · doi:10.1002/ejp.70241
Karakaş, H., Ertekin, Ö., Yavaş, İ., Çelik, R. G. G., Kazdağlı, H., Seebacher, B., & Kahraman, T. (2026) · Physiotherapy Research International · doi:10.1002/pri.70186
Cutting weight, gaining stress: the hidden battle that outweighs fight day tension in kickboxing
Karababa, B., Ulupınar, S., Gençoglu, C., Asan, S., Bedir, D., İnce, İ., Çabuk, S., Canyurt, F., Özbay, S., & Kaya, K. (2026) · Frontiers in Psychology · doi:10.3389/fpsyg.2026.1762659
Dehydration reduces heart rate variability in recreationally resistance-trained men
Hernandez, M. I., Jiwan, N. C., Appell, C. R., Dunn, R. P., KY, A. T., Keefe, M. S., Luk, H., & Sekiguchi, Y. (2026) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.26.17774-3
Pinto, M., Crisóstomo, J., & Monteiro, L. (2026) · Applied Sciences · doi:10.3390/app16062875
Xie, Y., Eungpinichpong, W., & Panyaek, N. (2026) · Journal of Human Sport and Exercise · doi:10.55860/6k972x29
Jones, M., Smith, G., Acevedo, A. M., & Melton, B. F. (2026) · Journal of Exercise and Nutrition · doi:10.53520/jen2026.103232
Quiblier, S. P. H., García-Pérez-de-Sevilla, G., Santiago, C., Iturriaga, T., Benítez-Valero, R., Revuelta, C., Carrasco, A. F., & Domínguez-Balmaseda, D. (2026) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2026.1768915
de la Torre-Gómez, S., Clemente‐Suárez, V. J., & Bustamante-Sánchez, Á. (2026) · Research Square · doi:10.21203/rs.3.rs-9061140/v1
Xiao, X., Moon, S., Kim, Y. C., & Choi, Y. (2026) · Bioengineering · doi:10.3390/bioengineering13040429
Chandra, S., Pasupuleti, D., Castaneda, G. C., Zheng, C., Avijeet, P., Dixon, M. J., & Dautenhahn, K. (2026) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3772318.3790706 · Cited by 2
Salagean, A., Ledger, D. A., Potts, D., Sehgal, T., Binesmael, S., Kalkanel, K., Karim, H., Wu, M., Fletcher, G., O'Neill, E. J., et al. (2026) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3772318.3791423
Helscel, K., Spears, N., Oberyszyn, A., Marquardt, M. D., Leahy, N. A., Emerson, A. M., Quatman-Yates, C., & Crick, J. (2026) · Hawaii International Conference on System Sciences (HICSS) · doi:10.24251/hicss.2026.445
Ulupınar, S., Özbay, S., Gençoglu, C., İnce, İ., Çabuk, S., Bakar, Ö., Demirli, A., & Kaya, K. (2026) · Sensors · doi:10.3390/s26082448
Chen, S., Zhang, Y., Li, X., Liu, Y., Yang, S., Zhang, X., & Liang, W. (2026) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-026-01716-7
Sholihat, A., Halonen, R., Möttönen, R., & Pesonen, A. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.05.12.724548
Darmová, B., Fajnerová, I., & Appel, L. (2026) · Technologies · doi:10.3390/technologies14050305
Mocrousov, E., & Olga, T. (2026) · Higher Education Research · doi:10.11648/j.her.20261103.11
Navarro-Gómez, P. (2026) · European Journal of Applied Physiology · doi:10.1007/s00421-026-06280-z
THE ROLE OF THE AUTONOMIC NERVOUS SYSTEM IN THE PATHOGENESIS OF ASTHMA EXACERBATION
Vysochyna, I. L., & Березуцький, І. (2026) · Здобутки клінічної і експериментальної медицини · doi:10.11603/1811-2471.2026.v.i2.16394
Dias, R. D., Michael, S. H., Harari, R. E., Yule, S., McGinnis, L. J., Skipworth, R., Mann-Salinas, E. A., & Davis, W. T. (2026) · Sensors · doi:10.3390/s26113596
Ji, W., Zheng, Z., Li, Z., Qiu, L., & Li, H. (2026) · Journal of Intelligent & Fuzzy Systems · doi:10.1177/18758967261457274
Palanisamy, S., Janakiraman, B., Balasubramaniam, A., Kandasamy, M., Parthasarathy, R., .Nalini, D. D., & Devi, K. S. (2026) · International Journal of Drug Delivery Technology · doi:10.25258/ijddt.16.50s.113
Analyzing VR-based group discussions for timely speaking-intention feedback to leaders
Gu, C., Chen, J., Yang, T., Xu, F., Ma, B., & Konomi, S. (2026) · Frontiers in Computer Science · doi:10.3389/fcomp.2026.1794972
Utsumi, A., & Yakoh, T. (2026) · SAE technical papers on CD-ROM/SAE technical paper series · doi:10.4271/2026-01-0690
Penna, G., Jelaska, I. D., Raiola, G., & Esposito, G. (2026) · Sci · doi:10.3390/sci8060141
Arık, M. I., Delibay, A. A., Tekin, E., & Küçük, A. (2026) · Perceptual and Motor Skills · doi:10.1177/00315125261462637
Evaluating Stress Monitoring Pipelines at the Ultra-Edge: A Mobile Device-Based Study
Gamero, V. J., & Kofuji, S. T. (2026) · doi:10.5753/sbcas.2026.21682
Metadata
- Authors
- Schaffarczyk, M., Rogers, B., Reer, R., & Gronwald, T.
- Affiliations
- Universität Hamburg (DE)
- University of Central Florida (US)
- MSH Medical School Hamburg – University of Applied Sciences and Medical University (DE)
- Corresponding author
- Schaffarczyk, Marcelle
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2022-08-30
- Volume
- 22
- Issue
- 17
- Pages
- 6536
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC9459793
- OpenAlex
- W4293715556
- Semantic Scholar
- d503adc11de24a90cadc9e05705b05b448ea1071
- Cited by
- 349
- References
- 38
- Citations by year
- 2026: 132 · 2025: 108 · 2024: 75 · 2023: 32 · 2022: 2
- FWCI
- 37.75
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Keywords
- heart rate variability, electrocardiogram comparison, RR interval, incremental exercise, measurement bias, limits of agreement, recreational athletes, sex differences
- MeSH terms
- Bicycling, Electrocardiography, Exercise Test, Female, Heart Rate, Humans, Male, Exercise
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-09, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-09 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review