Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review
Open accessAbstract
Background Wearable consumer technology allows for the collection of a growing amount of personal health data. Through the analysis of reflected LED light on the skin, heart rate measurement and arrhythmia detection can be performed. Given that melanin alters skin light absorption, this study seeks to summarize the accuracy of cardiac data from wrist-worn wearable devices for participants of varying skin tones. Methods We conducted a systematic review, searching Embase, MEDLINE, CINAHL, and Cochrane for original studies that stratified heart rate and rhythm data for consumer wearable technology according to participant race and/or skin tone. Results A total of 10 studies involving 469 participants met inclusion criteria. The frequency-weighted Fitzpatrick score for skin tone was reported in six studies ( n = 293), with a mean participant score of 3.5 (range 1–6). Overall, four of the ten studies reported a significant reduction in accuracy of heart rate measurement with wearable devices in darker-skinned individuals, compared to participants with lighter skin tones. Four studies noted no effect of user skin tone on accuracy. The remaining two studies showed mixed results. Conclusions Preliminary evidence is inconclusive, but some studies suggest that wearable devices may be less accurate for detecting heart rate in participants with darker skin tones. Higher quality evidence is necessary, with larger sample sizes and more objective stratification of participants by skin tone, in order to characterize potential racial bias in consumer devices.
AI snapshot
Written from the paper’s full text.
TL;DR
This systematic review of 10 observational studies and 469 participants found mixed evidence on whether wrist-worn wearable heart-rate accuracy differs by skin tone. Some studies reported reduced accuracy in darker-skinned individuals, but the authors conclude the evidence is preliminary and inconclusive.
Key findings
- Of 10 included studies, 4 reported statistically significant reductions in heart-rate accuracy in participants with darker skin tones, 4 found no effect of skin tone, and 2 showed mixed results; no study reported higher accuracy in darker-skinned participants.
- Six studies reported a frequency-weighted mean Fitzpatrick skin-tone score of 3.5 (range 1–6) across 293 participants.
- Specific reported effects included lower accuracy in darker skin in Hermand et al. (p < 0.001), Pasadyn et al. (p = 0.01), Hochstadt et al. (linear regression coefficient = 0.98, p < 0.001), and Spierer et al., where Mio Alpha mean average error was 16 bpm in Fitzpatrick 6 vs 3 bpm in Fitzpatrick 1.
- In all five studies comparing manufacturers, the Apple Watch was found more accurate than other commercial devices; Etiwy et al. reported a correlation coefficient of 0.80 for Apple Watch vs 0.52 for Garmin devices.
Methods
- Systematic review conducted per PRISMA guidelines, searching Embase, MEDLINE, CINAHL, and Cochrane CENTRAL from database inception to Nov 5, 2021 for original studies stratifying heart-rate or rhythm data from consumer wearables by participant race and/or skin tone.
- The search yielded 656 articles; after duplicate removal, 581 records remained, 25 full texts were reviewed, 6 met inclusion criteria, and 4 additional studies were found by hand and backwards searching, yielding 10 observational cohort studies with 469 participants (mean age 34 years, range 20–62; 38% women).
- Data extraction and quality assessment used the RoBANS tool, with findings regarding race outcomes extracted and qualitatively analyzed; heterogeneity precluded meta-analysis.
Limitations
- The authors state that included studies were not blinded, had small sample sizes, evaluated primarily young subjects, and used different wearable device brands and models.
- The authors also state that many studies reporting significant skin-tone effects did not quantify the magnitude of error, skin-tone reporting was inconsistent, and heterogeneity precluded meta-analysis; chest straps used as a gold standard in 3 of 10 studies have never been validated across different skin tones.
- Evident from the text: only 3 of 10 studies included participants with Fitzpatrick 6 skin tones, and none of the included studies compared atrial fibrillation or other arrhythmia detection, or oxygen saturation, across skin tones.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 48 cited sources checked · 0 supported · 0 flagged
- Findingown data
The identified studies in this review included small sample sizes.
“The identified studies in this review were not blinded, included small sample sizes, evaluated primarily young subjects, and used different wearable device brands and models.”
Quote in the paperSubstantial claim - Findingown data
The identified studies in this review used different wearable device brands and models.
“The identified studies in this review were not blinded, included small sample sizes, evaluated primarily young subjects, and used different wearable device brands and models.”
Quote in the paperSubstantial claim - Findingcitation
Hermand et al. reported lower heart rate accuracy in patients with darker skin (p < 0.001).
“Hermand et al. reported lower heart rate accuracy in patients with darker skin (p < 0.001) [15].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] Hermand E, Cassirame J, Ennequin G, Hue O. Validation of a photoplethysmographic heart rate monitor: polar OH1. Int J Sport Med [Internet]. 2019/06/13. 2019;40:462–7. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31189190 source
Not checked in the free run.
- Findingcitation
As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used.
“As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used, two modalities that are known to correlate well [13].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [13] Schaffarczyk M, Rogers B, Reer R, Gronwald T. Validity of the polar H10 sensor for heart rate variability analysis during resting state and incremental exercise in recreational men and women. Sensors. 2022;22:6536. source
Not checked in the free run.
- Findingcitation
In three of the included studies, skin tone was the primary variable considered in the accuracy of wrist-worn heart rate monitors.
“In three of the included studies, skin tone was the primary variable considered in the accuracy of wrist-worn heart rate monitors [1, 2, 14].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [1] Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019;43. source
Not checked in the free run.
- cites [2] Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med [Internet]. 2020/02/13. 2020;3:18. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32047863 source
Not checked in the free run.
- cites [14] Ray I, Liaqat D, Gabel M, E. de L. Skin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches. IEEE Int Conf Pervasive Comput Commun Work other Affil Events. 2021;213–9. source
Not checked in the free run.
- Findingnone given
The identified studies in this review evaluated primarily young subjects.
“The identified studies in this review were not blinded, included small sample sizes, evaluated primarily young subjects, and used different wearable device brands and models.”
Quote in the paperNot a substantial claim - Findingreasoning
Some studies grouped participants by Fitzpatrick scale ranges because of a shortage of participants at the extremes of skin tone (Fitzpatrick 1 and 6).
“Some studies grouped participants by Fitzpatrick scale ranges given a shortage of participants at the extremes of skin tone (i.e., Fitzpatrick scale 1 and 6).”
Quote in the paperNot a substantial claim - Findingown data
Discrepancies existed in how skin tone was reported across included studies, such as categorizations as light or dark, or by race, ethnicity, or Fitzpatrick scale subgroups.
“Furthermore, discrepancies existed in how skin tone was reported, such as categorizations of skin tones as either light or dark, or by race, ethnicity, or Fitzpatrick scale subgroups.”
Quote in the paperSubstantial claim - Findingreasoning
Many studies that reported a significant interaction between skin tone and cardiac data accuracy did not quantify the magnitude of error.
“Many of the studies that reported a significant interaction between skin tone and cardiac data accuracy did not quantify the magnitude of error.”
- Also stated in: § Results · plain
“However, the magnitude of measurement inaccuracy was not quantified in either study.”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Findingnone given
The identified studies in this review were not blinded.
“The identified studies in this review were not blinded, included small sample sizes, evaluated primarily young subjects, and used different wearable device brands and models.”
Quote in the paperNot a substantial claim - Findingown data
Only three of the ten included studies included participants with Fitzpatrick 6 skin tones.
“Only three of the ten studies included participants with Fitzpatrick 6 skin tones.”
Quote in the paperNot a substantial claim - Findingcitation
The remaining two of the ten included studies observed mixed results across different wearable devices.
“The remaining two studies showed mixed results.”
- Also stated in: § Results · plain
“The remaining two studies observed mixed results across different wearable devices [12, 16].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [12] Wallen MP, Gomersall SR, Keating SE, Wisloff U, Coombes JS. Accuracy of heart rate watches: implications for weight management. PLoS One [Internet]. 2016/05/28. 2016;11:e0154420. Available from: https://www.ncbi.nlm.nih.gov/pubmed/27232714 source
Not checked in the free run.
- cites [16] Spierer DK, Rosen Z, Litman LL, Fujii K. Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol [Internet]. 2015/06/27. 2015;39:264–71. Available from: https://www.ncbi.nlm.nih.gov/pubmed/26112379 source
Not checked in the free run.
- Also stated in: § Results · plain
- Findingcitation
All five studies that compared manufacturers found the Apple Watch to be more accurate than the other commercial devices.
“Of those, all five found the Apple Watch to be more accurate than the other commercial devices [2, 9–12].”
Quote in the paperNot a substantial claim0 of 5 sources checked- cites [2] Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med [Internet]. 2020/02/13. 2020;3:18. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32047863 source
Not checked in the free run.
- cites [9] Etiwy M, Akhrass Z, Gillinov L, Alashi A, Wang R, Blackburn G, et al. Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther [Internet]. 2019/07/06. 2019;9:262–71. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31275816 source
Not checked in the free run.
- cites [10] Pasadyn SR, Soudan M, Gillinov M, Houghtaling P, Phelan D, Gillinov N, et al. Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther [Internet]. 2019/09/27. 2019;9:379–85. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31555543 source
Not checked in the free run.
- cites [11] Shcherbina A, Mattsson CM, Waggott D, Salisbury H, Christle JW, Hastie T, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med [Internet]. 2017/05/26. 2017;7. Available from: https://www.ncbi.nlm.nih.gov/pubmed/28538708 source
Not checked in the free run.
- cites [12] Wallen MP, Gomersall SR, Keating SE, Wisloff U, Coombes JS. Accuracy of heart rate watches: implications for weight management. PLoS One [Internet]. 2016/05/28. 2016;11:e0154420. Available from: https://www.ncbi.nlm.nih.gov/pubmed/27232714 source
Not checked in the free run.
- Findingnone given
This is the first systematic review of the accuracy of cardiac data by wearable devices based on race and/or skin tone.
“Overall, this is the first systematic review of the accuracy of cardiac data by wearable devices based on race and/or skin tone.”
Quote in the paperSubstantial claim - Findingcitation
Four of the ten included studies found no difference in heart rate accuracy between participants of different skin tones.
“Four studies noted no effect of user skin tone on accuracy.”
- Also stated in: § Results · plain
“Four studies found no difference in accuracy between participants of different skin tones [1, 2, 9, 14].”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [1] Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019;43. source
Not checked in the free run.
- cites [2] Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med [Internet]. 2020/02/13. 2020;3:18. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32047863 source
Not checked in the free run.
- cites [9] Etiwy M, Akhrass Z, Gillinov L, Alashi A, Wang R, Blackburn G, et al. Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther [Internet]. 2019/07/06. 2019;9:262–71. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31275816 source
Not checked in the free run.
- cites [14] Ray I, Liaqat D, Gabel M, E. de L. Skin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches. IEEE Int Conf Pervasive Comput Commun Work other Affil Events. 2021;213–9. source
Not checked in the free run.
- Also stated in: § Results · plain
- Findingown data
Across six studies (n = 293), the frequency-weighted mean Fitzpatrick score for skin tone was 3.5 (range 1-6).
“The frequency-weighted Fitzpatrick score for skin tone was reported in six studies (n = 293), with a mean participant score of 3.5 (range 1–6).”
- Also stated in: § Results · plain
“Six studies reported an average Fitzpatrick score, and across these studies the frequency-weighted mean score was 3.5 (range 1–6) (n = 293).”
Quote in the paperNot a substantial claim - Also stated in: § Results · plain
- Findingown data
No included studies reported higher heart rate measurement accuracy in participants with darker skin tones.
“No studies reported a higher accuracy of heart rate measurement in participants with darker skin tones.”
Quote in the paperSubstantial claim - Findingown data
Four of the ten included studies reported statistically significant reductions in heart rate measurement accuracy in participants with darker skin tones compared to participants with lighter skin tones.
“Overall, four of the ten studies reported a significant reduction in accuracy of heart rate measurement with wearable devices in darker-skinned individuals, compared to participants with lighter skin tones.”
- Also stated in: § Results · plain
“Four of the ten studies reported statistically significant reductions in accuracy of heart rate data in participants with darker skin tones [3, 10, 11, 15].”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [3] Hochstadt A, Havakuk O, Chorin E, Schwartz AL, Merdler I, Laufer M, et al. Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients. J Electrocardiol [Internet]. 2020/05/04. 2020;60:138–41. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32361522 source
Not checked in the free run.
- cites [10] Pasadyn SR, Soudan M, Gillinov M, Houghtaling P, Phelan D, Gillinov N, et al. Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther [Internet]. 2019/09/27. 2019;9:379–85. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31555543 source
Not checked in the free run.
- cites [11] Shcherbina A, Mattsson CM, Waggott D, Salisbury H, Christle JW, Hastie T, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med [Internet]. 2017/05/26. 2017;7. Available from: https://www.ncbi.nlm.nih.gov/pubmed/28538708 source
Not checked in the free run.
- cites [15] Hermand E, Cassirame J, Ennequin G, Hue O. Validation of a photoplethysmographic heart rate monitor: polar OH1. Int J Sport Med [Internet]. 2019/06/13. 2019;40:462–7. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31189190 source
Not checked in the free run.
- Also stated in: § Results · plain
- Findingcitation
Pasadyn et al. found that the accuracy of heart rate detection was slightly lower in non-white participants (p = 0.01).
“Pasadyn et al. similarly found that the accuracy of heart rate detection was slightly lower in non-white participants (p = 0.01) [10].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Pasadyn SR, Soudan M, Gillinov M, Houghtaling P, Phelan D, Gillinov N, et al. Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther [Internet]. 2019/09/27. 2019;9:379–85. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31555543 source
Not checked in the free run.
- Findingown data
None of the included studies compared the detection of atrial fibrillation or other arrhythmias in participants with various skin tones.
“None of the included studies compared the detection of atrial fibrillation or other arrhythmias in participants with various skin tones.”
Quote in the paperNot a substantial claim - Findingcitation
Hochstadt et al. noted a linear regression coefficient of 0.98 (p < 0.001) when comparing PPG to ECG data in patients with darker skin (Fitzpatrick 5 or 6).
“Hochstadt et al. noted a linear regression coefficient of 0.98 (p < 0.001) when comparing PPG to ECG data in patients with darker skin (Fitzpatrick 5 or 6), suggesting that darker skin tone reduced accuracy, albeit with a relatively small impact [3].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [3] Hochstadt A, Havakuk O, Chorin E, Schwartz AL, Merdler I, Laufer M, et al. Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients. J Electrocardiol [Internet]. 2020/05/04. 2020;60:138–41. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32361522 source
Not checked in the free run.
- Findingcitation
Spierer et al. found that the Mio Alpha device had a higher mean average error in Fitzpatrick 6 patients compared to Fitzpatrick 1 patients (16 beats/min compared to 3 beats/min).
“Spierer et al. found that one of the two devices tested (Mio Alpha) had a higher mean average error in Fitzpatrick 6 patients compared to Fitzpatrick 1 patients (16 beats/min compared to 3 beats/min, respectively) [16].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [16] Spierer DK, Rosen Z, Litman LL, Fujii K. Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol [Internet]. 2015/06/27. 2015;39:264–71. Available from: https://www.ncbi.nlm.nih.gov/pubmed/26112379 source
Not checked in the free run.
- Findingcitation
Bent et al. found equivalent accuracy for all devices tested across various skin tones.
“Bent et al. found equivalent accuracy for all devices tested across various skin tones [2].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2] Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med [Internet]. 2020/02/13. 2020;3:18. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32047863 source
Not checked in the free run.
- Interpretationreasoning
The relationship between darker skin tone and reduced accuracy is less evident in studies of reflectance PPG used in wearable devices than in transmission PPG used in pulse oximetry.
“While the racial limitations of transmission PPG have been consistently demonstrated in pulse oximetry, this systematic review demonstrated that this relationship is less evident in studies performed to date on the reflectance PPG used in wearable devices.”
Quote in the paperSubstantial claim - Premisecitation
The accuracy of PPG is impaired by factors that impede light transmission, such as elevated body mass index, tattoos, and higher hair follicle density.
“While safe and inexpensive, the accuracy of PPG is impaired by factors that impede light transmission, such as elevated body mass index, tattoos, and higher hair follicle density [4].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [4] Nelson BW, Low CA, Jacobson N, Arean P, Torous J, Allen NB. Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research. NPJ Digit Med [Internet]. 2020/07/03. 2020;3:90. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32613085 source
Not checked in the free run.
- Premisecitation
Electrical sensors require the user to remain still and touch the device with their opposite hand throughout the detection period.
“The associated hardware in modern devices relies on two components: photoplethysmography (PPG), which passively monitors heart rate and rhythm, and electrical sensors, which require the user to remain still and touch the device with their opposite hand throughout the detection period [1].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [1] Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019;43. source
Not checked in the free run.
- Premisecitation
In Sjoding et al.'s first cohort (n = 10,789), 11.4% of Black patients had undetected hypoxemia compared to 3.6% of white patients.
“Specifically, in the first cohort (n = 10,789), 11.4% Black patients had undetected hypoxemia (pulse oximetry > 92% and arterial oxygen saturation < 88%), compared to 3.6% white patients, and in the second cohort (n = 37,308), 17.0% Black patients had undetected hypoxemia, compared to 6.2% white patients.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [6] Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383:2477–8. source
Not checked in the free run.
- Premisecitation
Sjoding et al. found that compared to white patients, Black patients had a threefold increased frequency of undetected hypoxemia when using pulse oximeters.
“A recent study by Sjoding et al. found that when compared to white patients, Black patients had a threefold increased frequency of undetected hypoxemia when using pulse oximeters [6].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [6] Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383:2477–8. source
Not checked in the free run.
- Premisecitation
A racial discrepancy has been established in pulse oximeters, which rely on transmission PPG.
“A racial discrepancy has been established in pulse oximeters, which rely on transmission PPG, a slightly different form of the technology.”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [5] Khan Y, Han D, Pierre A, Ting J, Wang X, Lochner CM, et al. A flexible organic reflectance oximeter array. Proc Natl Acad Sci U S A. 2018;115:E11015–24. source
Not checked in the free run.
- cites [6] Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383:2477–8. source
Not checked in the free run.
- cites [7] Fawzy A, Wu TD, Wang K, Robinson ML, Farha J, Bradke A, et al. Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19. JAMA Intern Med. 2022;21224. source
Not checked in the free run.
- Interpretationreasoning
Because electrical sensors require active user engagement, the majority of wearable heart rate and rhythm data is collected using PPG.
“Given that the electrical sensors require active user engagement to measure heart rate and rhythm, the majority of wearable heart rate and rhythm data is collected using PPG.”
Quote in the paperNot a substantial claim - Premisecitation
In Sjoding et al.'s second cohort (n = 37,308), 17.0% of Black patients had undetected hypoxemia compared to 6.2% of white patients.
“Specifically, in the first cohort (n = 10,789), 11.4% Black patients had undetected hypoxemia (pulse oximetry > 92% and arterial oxygen saturation < 88%), compared to 3.6% white patients, and in the second cohort (n = 37,308), 17.0% Black patients had undetected hypoxemia, compared to 6.2% white patients.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [6] Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383:2477–8. source
Not checked in the free run.
- Premisecitation
Fallow et al. identified that green wavelengths (520 nm) displayed greater accuracy in heart rate measurement regardless of skin type when compared to other wavelengths, both at rest and during exercise.
“Fallow et al. identified that green wavelengths (520 nm) displayed greater accuracy in heart rate measurement regardless of skin type when compared to other wavelengths, both at rest and during exercise [20].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [20] Fallow BA, Tarumi T, Tanaka H. Influence of skin type and wavelength on light wave reflectance. J Clin Monit Comput [Internet]. 2013/02/12. 2013;27:313–7. Available from: https://www.ncbi.nlm.nih.gov/pubmed/23397431 source
Not checked in the free run.
- Classified as probably not a substantial claim. 17
- May combine two claims. 19
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataOf the ten included studies, five compared the relative accuracy of different manufacturers.
“Of the ten studies, five compared the relative accuracy of different manufacturers.”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationEtiwy et al. found the Apple Watch to follow ECG data with a correlation coefficient of 0.80, compared to 0.52 in Garmin devices.
“For instance, Etiwy et al. found the Apple Watch to follow ECG data with a correlation coefficient of 0.80, compared to 0.52 in Garmin devices [9].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Etiwy M, Akhrass Z, Gillinov L, Alashi A, Wang R, Blackburn G, et al. Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther [Internet]. 2019/07/06. 2019;9:262–71. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31275816 source
Not checked in the free run.
May combine two claims.
Findingown dataThe review identified a relative scarcity of studies considering skin tone interactions in smart watch cardiac accuracy, resulting in inconclusive findings.
“This review identified a relative scarcity of studies that consider the interactions of skin tone when characterizing smart watch device accuracy in cardiac outcomes, resulting in inconclusive findings.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the seven other included studies, skin tone was one of several covariates considered in the accuracy analysis.
“For the seven other studies, skin tone was one of several covariates considered in the accuracy analysis.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe systematic review included 10 studies involving 469 participants.
“A total of 10 studies involving 469 participants met inclusion criteria.”
- Also stated in: § Results · plain
“The 10 studies included a total of 469 participants, with a mean age of 34 years (range 20–62); a total of 38% (range 0–67%) of participants were women.”
Quote in the paperNot a substantial claim- Also stated in: § Results · plain
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe Fitzpatrick scale was used to classify participant skin tones in eight of the ten included studies.
“The Fitzpatrick scale was used to classify participant skin tones in eight of the ten studies.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataNone of the included studies assessed oxygen saturation or other vital signs.
“Similarly, none of the studies assessed oxygen saturation or other vital signs.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataThe most common wearable device brands studied were Apple Watch (n = 6), Fitbit (n = 5), Mio Alpha (n = 3), and Garmin (n = 3).
“The most common wearable device brands that were studied included the Apple Watch (n = 6), Fitbit (n = 5), Mio Alpha (n = 3), and Garmin (n = 3).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn the two studies that did not use the Fitzpatrick scale, one classified participants as white, Black, or other, and another classified participants by ethnic background including Asian, Black, and Hispanic, precluding aggregate analysis of skin tone outcomes.
“In the two studies that did not use the Fitzpatrick scale, one classified participants as white, Black, or other, and another classified participants based on ethnic background, including Asian, Black, and Hispanic — precluding aggregate analysis of skin tone outcomes.”
Quote in the paperSubstantial claimMay combine two claims.
FindingcitationWallen et al. noted lower heart rate accuracy in participants with a Fitzpatrick scale score > 4 with the Apple Watch, but not in the other studied smartwatches.
“Wallen et al. noted lower heart rate accuracy in participants with a Fitzpatrick scale score > 4 with the Apple Watch, but not in the other studied smartwatches [12].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [12] Wallen MP, Gomersall SR, Keating SE, Wisloff U, Coombes JS. Accuracy of heart rate watches: implications for weight management. PLoS One [Internet]. 2016/05/28. 2016;11:e0154420. Available from: https://www.ncbi.nlm.nih.gov/pubmed/27232714 source
Not checked in the free run.
May combine two claims.
FindingcitationRay et al. found that various WearOS watches systematically underestimate the reliability of HR readings taken from dark skin, despite no substantial differences in error, leading to significantly fewer recorded data points in patients with dark skin.
“Ray et al. found that various WearOS watches systematically underestimate the reliability of HR readings taken from dark skin, despite no substantial differences in error, leading to significantly fewer recorded data points in patients with dark skin [14].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] Ray I, Liaqat D, Gabel M, E. de L. Skin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches. IEEE Int Conf Pervasive Comput Commun Work other Affil Events. 2021;213–9. source
Not checked in the free run.
May combine two claims.
FindingcitationShcherbina et al. noted that smart watch device error was higher for darker skin tones, but the degree of this effect was not reported.
“Shcherbina et al. noted that smart watch device error was higher for darker skin tones, but the degree of this effect was not reported [11].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [11] Shcherbina A, Mattsson CM, Waggott D, Salisbury H, Christle JW, Hastie T, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med [Internet]. 2017/05/26. 2017;7. Available from: https://www.ncbi.nlm.nih.gov/pubmed/28538708 source
Not checked in the free run.
May combine two claims.
FindingcitationEtiwy et al. and Sanudo et al. found that skin color did not influence heart rate accuracy.
“Etiwy et al. and Sanudo et al. similarly found that skin color did not influence heart rate accuracy [1, 9].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [1] Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019;43. source
Not checked in the free run.
- cites [9] Etiwy M, Akhrass Z, Gillinov L, Alashi A, Wang R, Blackburn G, et al. Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther [Internet]. 2019/07/06. 2019;9:262–71. Available from: https://www.ncbi.nlm.nih.gov/pubmed/31275816 source
Not checked in the free run.
May combine two claims.
InterpretationreasoningHeterogeneity across included studies precluded meta-analysis and may contribute to variability in reported results.
“This heterogeneity precluded meta-analysis and may contribute to the variability in reported results.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationECG and chest strap are two modalities that are known to correlate well.
“As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used, two modalities that are known to correlate well [13].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] Schaffarczyk M, Rogers B, Reer R, Gronwald T. Validity of the polar H10 sensor for heart rate variability analysis during resting state and incremental exercise in recreational men and women. Sensors. 2022;22:6536. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe evidence is inconclusive, but some studies suggest that wearable devices may be less accurate for detecting heart rate in participants with darker skin tones.
“Preliminary evidence is inconclusive, but some studies suggest that wearable devices may be less accurate for detecting heart rate in participants with darker skin tones.”
- Also stated in: § Discussion · hedged
“However, the included studies described mixed results and as such, it remains unclear whether wearable devices have reduced accuracy in heart rate and arrhythmia detection in people with darker skin tones.”
- Also stated in: § Conclusion · boosted
“Early evidence of racial bias in wrist-worn wearables is mixed, but some studies demonstrate reduced accuracy of heart rate data in users with darker skin tones.”
Quote in the paperSubstantial claim- Also stated in: § Discussion · hedged
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenNo exclusions were applied on the basis of language or country of origin.
“No exclusions were applied on the basis of language or country of origin.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenThe study used rigorous research methodology including the search of multiple research databases and screening of publications by two independent reviewers in duplicate.
“The study used rigorous research methodology including the search of multiple research databases and screening of publications by two independent reviewers in duplicate.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenTitle and abstract screening, full-text review, and data extraction were performed independently by two investigators, with a third reviewer resolving discrepancies.
“Title and abstract screening, full-text review, and data extraction were performed independently by two investigators (SK and DK), with a third reviewer (AK) resolving discrepancies.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationModern wearable devices rely on two components: photoplethysmography (PPG) and electrical sensors.
“The associated hardware in modern devices relies on two components: photoplethysmography (PPG), which passively monitors heart rate and rhythm, and electrical sensors, which require the user to remain still and touch the device with their opposite hand throughout the detection period [1].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [1] Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019;43. source
Not checked in the free run.
May combine two claims.
PremisecitationPulse oximeters measure LED light that shines through the tissue to the opposite side, restricting them to locations that can be transilluminated, such as the earlobes and fingers.
“Instead of photodetectors measuring reflected light, they measure LED light that shines through the tissue to the opposite side, restricting it to locations that can be transilluminated, such as the earlobes and fingers [5].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [5] Khan Y, Han D, Pierre A, Ting J, Wang X, Lochner CM, et al. A flexible organic reflectance oximeter array. Proc Natl Acad Sci U S A. 2018;115:E11015–24. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataThe Risk of Bias Assessment Tool for Nonrandomized Studies (RoBANS) was used to assess methodological quality of included studies, appraised independently by two reviewers.
“The Risk of Bias Assessment Tool for Nonrandomized Studies (RoBANS) tool was used to assess the methodological quality of included studies. Studies were appraised independently by two parallel reviewers (SK, DK).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe authors systematically searched four databases from database inception to Nov 5, 2021.
“The authors systematically searched four databases from database inception to Nov 5, 2021: (1) the Ovid versions of MEDLINE and MEDLINE Daily including e-publications, in progress, and non-indexed citations; (2) Embase Classic and Embase; (3) CINAHL; and (4) Cochrane CENTRAL.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe review was conducted according to PRISMA guidelines.
“This review was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationcitationDarker skin tone reduced accuracy in the Hochstadt et al. comparison, albeit with a relatively small impact.
“Hochstadt et al. noted a linear regression coefficient of 0.98 (p < 0.001) when comparing PPG to ECG data in patients with darker skin (Fitzpatrick 5 or 6), suggesting that darker skin tone reduced accuracy, albeit with a relatively small impact [3].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [3] Hochstadt A, Havakuk O, Chorin E, Schwartz AL, Merdler I, Laufer M, et al. Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients. J Electrocardiol [Internet]. 2020/05/04. 2020;60:138–41. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32361522 source
Not checked in the free run.
Classified as probably not a substantial claim.
PremisereasoningIt remains unclear whether the inaccuracy of transmission PPG in darker skin tones holds true for the reflectance PPG used in consumer wrist-worn devices.
“While the inaccuracy of transmission PPG used in pulse oximeters is well established in darker skin tones, it remains unclear whether this holds true for the reflectance PPG used in consumer wrist-worn devices.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationHemoglobin in dermal blood vessels drives the bulk of light absorption, and changes in blood pressure and flow rate affect light absorption, which is detected by PPG optical sensors.
“Hemoglobin in dermal blood vessels drives the bulk of light absorption, and thus changes in blood pressure and flow rate affect light absorption, which is then detected by PPG optical sensors [2, 3].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [2] Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med [Internet]. 2020/02/13. 2020;3:18. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32047863 source
Not checked in the free run.
- cites [3] Hochstadt A, Havakuk O, Chorin E, Schwartz AL, Merdler I, Laufer M, et al. Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients. J Electrocardiol [Internet]. 2020/05/04. 2020;60:138–41. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32361522 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBackwards searching was performed by screening the reference lists of all included studies to identify relevant articles.
“Backwards searching was also performed by screening the reference lists of all included studies to identify relevant articles.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPPG relies on LEDs projecting light onto the skin and photodetectors measuring the quantity of reflected light.
“This technology relies on LEDs projecting light onto the skin, and photodetectors measuring the quantity of reflected light [1, 2].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [1] Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019;43. source
Not checked in the free run.
- cites [2] Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med [Internet]. 2020/02/13. 2020;3:18. Available from: https://www.ncbi.nlm.nih.gov/pubmed/32047863 source
Not checked in the free run.
Classified as probably not a substantial claim.
InterpretationreasoningThere are limitations to this systematic review, largely related to the preliminary nature of the evidence base.
“There are limitations to this systematic review, largely related to the preliminary nature of the evidence base.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationA retrospective study of 7126 patients with COVID-19 found that pulse oximetry overestimated arterial oxygen saturation among Asian, Black, and Hispanic patients compared to white patients, leading to a delay in guideline-based therapies.
“Similarly, a retrospective study of 7126 patients with COVID-19 found that pulse oximetry overestimated arterial oxygen saturation among Asian, Black, and Hispanic patients compared to white patients, leading to a delay in the initiation of guideline-based therapies [7].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [7] Fawzy A, Wu TD, Wang K, Robinson ML, Farha J, Bradke A, et al. Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19. JAMA Intern Med. 2022;21224. source
Not checked in the free run.
The paper relies on this without support.
Melanin is the main absorbent of light in the epidermis.
“Given that melanin is the main absorbent of light in the epidermis, it is theorized that this might similarly interfere with the PPG signal estimation process.”
The paper relies on this without support.
Chest straps have never been validated across different skin tones.
“Further, despite being used as a gold standard in three of the ten included studies, chest straps have never been validated across different skin tones.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Melanin absorption of green light is a primary factor interfering with optical signal detection in PPG-based measurements.
“In PPG-based measurements, melanin absorption of green light was identified as a primary factor interfering with optical signal detection [9].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [9]
The Introduction says melanin absorbs light in the epidermis and is theorized to interfere with PPG signal estimation. It does not identify absorption of green light as a primary interfering factor. The Discussion instead reports that green wavelengths showed greater heart-rate measurement accuracy across skin types than other wavelengths. (LLM, conf=0.88)
Experimental — accuracy not yet measuredApple devices tended to perform better during exercise but still slightly underestimated peak values, a discrepancy deemed clinically negligible.
“By contrast, Apple devices tended to perform better during exercise but still slightly underestimated peak values, a discrepancy deemed clinically negligible [9].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [9]
The Results (page not available in our copy) report that all five studies comparing manufacturers found the Apple Watch more accurate than other commercial devices. The review does not say that this advantage was specific to exercise, or that Apple devices slightly underestimated peak heart rates or that such a discrepancy was clinically negligible. (LLM, conf=0.86)
Experimental — accuracy not yet measuredNo single brand was consistently superior across all skin tones and activities.
“Koerber et al. (2023), in a systematic review, emphasized this heterogeneity, concluding that no single brand was consistently superior across all skin tones and activities [9].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [9]
The Results (page unavailable) report mixed findings across skin tones and some differences between devices, supporting the claim of heterogeneity. But they also say that all five studies comparing manufacturers found the Apple Watch more accurate than the other devices; the review does not establish a brand ranking across all skin tones and activities. (LLM, conf=0.84)
Experimental — accuracy not yet measuredApple and Fitbit's proprietary algorithms outperform smaller-brand or open-source counterparts, largely because of superior noise filtering and adaptive calibration.
“Comparative evaluations indicate that Apple and Fitbit’s proprietary algorithms outperform smaller-brand or open-source counterparts, largely because of superior noise filtering and adaptive calibration [9,11].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [9]
The Results (page not available in our copy) report that all five studies comparing manufacturers found the Apple Watch more accurate than other commercial devices. The Introduction says proprietary algorithms convert optical measurements into heart-rate data, but the paper does not establish that Fitbit also outperforms other brands, compare either brand with open-source algorithms, or attribute an advantage to noise filtering or adaptive calibration. (LLM, conf=0.88)
Experimental — accuracy not yet measuredErrors during exercise monitoring may result in inappropriate training intensities or misleading recovery assessments.
“Errors during exercise monitoring may result in inappropriate training intensities or misleading recovery assessments [7,8,9].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [9]
The Results (page unavailable) report heart-rate measurement errors in some wrist-worn devices, which supports the premise that exercise-monitoring readings can be inaccurate. The paper does not address whether those errors lead to inappropriate training intensities or misleading recovery assessments. (LLM, conf=0.79)
Experimental — accuracy not yet measured
Reduced accuracy in darker skin tones has been linked to missed detection of atrial fibrillation when irregular pulse notifications fail to capture abnormal rhythms.
“Reduced accuracy in darker skin tones has been linked to missed detection of atrial fibrillation when irregular pulse notifications fail to capture abnormal rhythms.”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [9]
The Results and Conclusion report that some studies found less accurate wearable heart-rate measurements in people with darker skin tones, although the evidence was mixed. The Discussion says it remains unclear whether skin tone affects arrhythmia detection; it does not link reduced accuracy to missed atrial fibrillation or failed irregular-pulse notifications. Page numbers were not available in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measuredThere have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR or pulse oximeter SpO2.
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO2 [36-38].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [35]
Studies have shown that heart rate sensors on some wearables are less accurate for individuals with darker skin tones, reflecting a racial design bias that may affect health outcomes.
“For example, studies have shown that heart rate sensors on some wearables are less accurate for individuals with darker skin tones 57 , reflecting a racial design bias that may affect health outcomes.”
From Privacy in consumer wearable technologies: a living systematic analysis of data policies across leading manufacturers (2025) · cites this paper as [57]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Spierer results describe skin types the table says were absent
MajorTable 1 says the Spierer study had no Fitzpatrick 1 or 6 participants, but the Results section compares measurement errors in Fitzpatrick 1 and 6 patients from that study.
“Mean FP = 3.0, no FP 1 or 6 participants”
“Spierer et al. found that one of the two devices tested (Mio Alpha) had a higher mean average error in Fitzpatrick 6 patients compared to Fitzpatrick 1 patients (16 beats/min compared to 3 beats/min, respectively) [16].”
Hochstadt study year differs between table and reference
MinorTable 1 labels the Hochstadt study as 2021, while its reference gives a 2020 publication year.
“Hochstadt 20214”
“Hochstadt A, Havakuk O, Chorin E, Schwartz AL, Merdler I, Laufer M, et al. Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients. J Electrocardiol [Internet]. 2020/05/04. 2020;60:138–41.”
Wallen study year differs between table and reference
MinorTable 1 labels the Wallen study as 2015, while its reference gives a 2016 publication year.
“Wallen 201510”
“Wallen MP, Gomersall SR, Keating SE, Wisloff U, Coombes JS. Accuracy of heart rate watches: implications for weight management. PLoS One [Internet]. 2016/05/28. 2016;11:e0154420.”
Cites
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339 · see citing works
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · cited 5× · see citing works
“This technology relies on LEDs projecting light onto the skin, and photodetectors measuring the quantity of reflected light [1, 2].”
BackgroundInfluentialShcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625 · see citing works
Ly, B. C. K., Dyer, E. B., Feig, J. L., Chien, A. L., & Del Bino, S. (2020) · Journal of Investigative Dermatology · doi:10.1016/j.jid.2019.11.003 · Cited by 453 · see citing works
Schaffarczyk, M., Rogers, B., Reer, R., & Gronwald, T. (2022) · Sensors · doi:10.3390/s22176536 · Cited by 349 · see citing works
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · see citing works
Fawzy, A., Wu, T. D., Wang, K., Robinson, M. L., Farha, J., Bradke, A., Golden, S. H., Xu, Y., & Garibaldi, B. T. (2022) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2022.1906 · Cited by 326 · see citing works
A flexible organic reflectance oximeter array
Khan, Y., Han, D., Pierre, A., Ting, J., Wang, X., Lochner, C. M., Bovo, G., Yaacobi-Gross, N., Newsome, C., Wilson, R., et al. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1813053115 · Cited by 278 · see citing works
Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
Nelson, B. W., Low, C. A., Jacobson, N., Areán, P., Torous, J., & Allen, N. B. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0297-4 · Cited by 241 · see citing works
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162 · see citing works
Accuracy of wearable heart rate monitors in cardiac rehabilitation
Etiwy, M., Akhrass, Z., Gillinov, L., Alashi, A., Wang, R., Blackburn, G., Gillinov, S. M., Phelan, D., Gillinov, A. M., Houghtaling, P. L., et al. (2019) · Cardiovascular Diagnosis and Therapy · doi:10.21037/cdt.2019.04.08 · Cited by 143 · see citing works
Validation of photoplethysmography as a method to detect heart rate during rest and exercise
Spierer, D. K., Rosen, Z., Litman, L. L., & Fujii, K. (2015) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2015.1047536 · Cited by 139 · see citing works
Accuracy of commercially available heart rate monitors in athletes: a prospective study
Pasadyn, S. R., Soudan, M., Gillinov, M., Houghtaling, P., Phelan, D., Gillinov, N., Bittel, B., & Desai, M. Y. (2019) · Cardiovascular Diagnosis and Therapy · doi:10.21037/cdt.2019.06.05 · Cited by 135 · see citing works
Artificial intelligence in emergency medicine: A scoping review
Kirubarajan, A., Taher, A., Khan, S., & Masood, S. (2020) · JACEP Open · doi:10.1002/emp2.12277 · Cited by 127 · see citing works
Self-citationReflectance spectrophotometer: the dermatologists’ sphygmomanometer for skin phototyping?
Pershing, L. K., Tirumala, V. P., Nelson, J. L., Corlett, J. L., Lin, A. G., Meyer, L. J., & Leachman, S. A. (2008) · Journal of Investigative Dermatology · doi:10.1038/sj.jid.5701238 · Cited by 72 · see citing works
Use of wearable devices for post-discharge monitoring of ICU patients: a feasibility study
for the WEARable Information Technology for hospital INpatients (WEARIT-IN) study group, Kroll, R. R., McKenzie, E. D., Boyd, J. G., Sheth, P., Howes, D., Wood, M., & Maslove, D. M. (2017) · Journal of Intensive Care · doi:10.1186/s40560-017-0261-9 · Cited by 69 · see citing works
Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH1
Hermand, E., Cassirame, J., Ennequin, G., & Hue, O. (2019) · International Journal of Sports Medicine · doi:10.1055/a-0875-4033 · Cited by 59 · see citing works
Sañudo, B., De Hoyo, M., Muñoz-López, A., Perry, J., & Abt, G. (2019) · Journal of Medical Systems · doi:10.1007/s10916-019-1325-2 · Cited by 48 · cited 5× · see citing works
“The associated hardware in modern devices relies on two components: photoplethysmography (PPG), which passively monitors heart rate and rhythm, and electrical sensors, which require the user to remain still and touch the device with their opposite hand throughout the detection period [1].”
BackgroundInfluentialSkin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches
Ray, I., Liaqat, D., Gabel, M., & de Lara, E. (2021) · 2021 IEEE International Conference on Pervasive Computing and Communications Workshops and other Affiliated Events (PerCom Workshops) · doi:10.1109/percomworkshops51409.2021.9431120 · Cited by 23 · see citing works
Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients
Hochstadt, A., Havakuk, O., Chorin, E., Schwartz, A. L., Merdler, I., Laufer, M., Lubman, N., Ghantous, E., Viskin, S., & Rosso, R. (2020) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2020.04.017 · Cited by 18 · see citing works
WRIST WORNE WEARABLE FOR DETECTING ATRIAL FIBRILLATION
Banerjee, S., Banerjee, R. K., Zhuang, L., et al. (2020) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(20)34188-7 · Cited by 2 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Setchfield, K., Gorman, A., Simpson, A. H. R. W., Somekh, M. G., & Wright, A. J. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.1.010901 · Cited by 72 · see citing works
Doherty, C., Baldwin, M., Lambe, R., Altini, M., & Caulfield, B. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01757-1 · Cited by 55 · see citing works
Setchfield, K., Gorman, A., Simpson, A. H. R. W., Somekh, M. G., & Wright, A. J. (2023) · Biomedical Optics Express · doi:10.1364/boe.493588 · Cited by 49 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO 2 [36-38].”
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring
Asif, S., AlSaafeen, A., Nadar, S., Nambiar, S., Dannawi, J., Korrapati, H., & Wilkhoo, H. S. (2025) · Cureus · doi:10.7759/cureus.94074 · Cited by 9 · see citing works
- Authors
- Koerber, D., Khan, S., Shamsheri, T., Kirubarajan, A., & Mehta, S.
- Affiliations
- University of Toronto (CA)
- McMaster University (CA)
- Mount Sinai Hospital (CA)
- Corresponding author
- Mehta, Sangeeta
- Venue
- Journal of Racial and Ethnic Health Disparities
- Published
- 2022-11-14
- Volume
- 10
- Issue
- 6
- Pages
- 2676-2684
- ISSN
- 2196-8837 (ISSN-L)
- 2197-3792
- PMC
- PMC9662769
- OpenAlex
- W4308963722
- Semantic Scholar
- 1c5e53690208ac097e5604fd23564e3cf10dfb99
- Cited by
- 91
- References
- 22
- Citations by year
- 2026: 26 · 2025: 37 · 2024: 21 · 2023: 7
- FWCI
- 1.45
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- Locations
- Journal of Racial and Ethnic Health Disparities (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
- Keywords
- heart rate measurement, wearable devices, skin tone, Fitzpatrick skin type, melanin, racial bias, consumer health technology, arrhythmia detection, accuracy
- MeSH terms
- Wearable Electronic Devices, Heart Rate, Humans, Skin Pigmentation, Wrist
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review