Photoplethysmography
Abstract
The photoplethysmographic (PPG) waveform, also known as the pulse oximeter waveform, is one of the most commonly displayed clinical waveforms. First described in the 1930s, the technology behind the waveform is simple. The waveform, as displayed on the modern pulse oximeter, is an amplified and highly filtered measurement of light absorption by the local tissue over time. It is optimized by medical device manufacturers to accentuate its pulsatile components. Physiologically, it is the result of a complex, and not well understood, interaction between the cardiovascular, respiratory, and autonomic systems. All modern pulse oximeters extract and display the heart rate and oxygen saturation derived from the PPG measurements at multiple wavelengths. "As is," the PPG is an excellent monitor for cardiac arrhythmia, particularly when used in conjunction with the electrocardiogram (ECG). With slight modifications in the display of the PPG (either to a strip chart recorder or slowed down on the monitor screen), the PPG can be used to measure the ventilator-induced modulations which have been associated with hypovolemia. Research efforts are under way to analyze the PPG using improved digital signal processing methods to develop new physiologic parameters. It is hoped that when these new physiologic parameters are combined with a more modern understanding of cardiovascular physiology (functional hemodynamics) the potential utility of the PPG will be expanded. The clinical researcher's objective is the use of the PPG to guide early goal-directed therapeutic interventions (fluid, vasopressors, and inotropes), in effect to extract from the simple PPG the information and therapeutic guidance that was previously only obtainable from an arterial pressure line and the pulmonary artery catheter.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“It is reported that the lower-frequency wave manifestation in the total PPG signal can be attributed to the venous vascular system [119].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [119]
Citations
Cites
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Non-contact, automated cardiac pulse measurements using video imaging and blind source separation
Poh, M., McDuff, D., & Picard, R. W. (2010) · Optics Express · doi:10.1364/oe.18.010762 · Cited by 1,652
Gan, T. J., Soppitt, A. J., Maroof, M., El-Moalem, H. E., Robertson, K. M., Moretti, E., Dwane, P., & Glass, P. S. A. (2002) · Anesthesiology · doi:10.1097/00000542-200210000-00012 · Cited by 950
Hamilton, M. A., Cecconi, M., & Rhodes, A. (2010) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181eeaae5 · Cited by 897
Changes in Arterial Pressure during Mechanical Ventilation
Michard, F. (2005) · Anesthesiology · doi:10.1097/00000542-200508000-00026 · Cited by 702
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589
Cannesson, M. P., Manach, Y. L., Hofer, C. K., Goarin, J., Lehot, J.-J., Vallet, B., & Tavernier, B. (2011) · Anesthesiology · doi:10.1097/aln.0b013e318225b80a · Cited by 549
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 460
Photoplethysmography: Beyond the Calculation of Arterial Oxygen Saturation and Heart Rate
Shelley, K. H. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269512.82836.c9 · Cited by 450
Functional hemodynamic monitoring.
Pinsky, M. R., & Payen, D. M. (2005) · Critical Care · doi:10.1186/cc3927 · Cited by 369
Research works printed but not linked: 3
- Hertzman, 1937, Observations on the finger volume pulse recorded photoelectrically, Am J Physiol, 119, 334
- August, 1851, Versuch der absorptions-verhaltnisse des cordierites fur rothes licht zu bestimmen, Ann Physik Chem Ger, 84, 37
- Shelley, 2005, The effect of venous pulsation on the forehead pulse oximeter wave form as a possible source of error in SpO2 calculation, Anesth Analg, 100, 743, 10.1213/01.ANE.0000145063.01043.4B · DOI · Paper page
Cited by
Sorted by citations
Photoplethysmogram Analysis and Applications: An Integrative Review
Park, J., Seok, H. S., Kim, S.-S., & Shin, H. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2021.808451 · Cited by 514
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348
Advances in Photopletysmography Signal Analysis for Biomedical Applications
Moraes, J. L., Rocha, M. X., Vasconcelos, G. G., Vasconcelos Filho, J. E., De Albuquerque, V. H. C., & Alexandria, A. R. (2018) · Sensors · doi:10.3390/s18061894 · Cited by 292
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263
Heart rate variability with photoplethysmography in 8 million individuals: a cross-sectional study
Natarajan, A., Pantelopoulos, A., Emir-Farinas, H., & Natarajan, P. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(20)30246-6 · Cited by 229
A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China
Liang, Y., Chen, Z., Liu, G., & Elgendi, M. (2018) · Scientific Data · doi:10.1038/sdata.2018.20 · Cited by 215
Diagnostic Features and Potential Applications of PPG Signal in Healthcare: A Systematic Review
Almarshad, M. A., Islam, M. S., Al-Ahmadi, S. A., & BaHammam, A. S. (2022) · Healthcare · doi:10.3390/healthcare10030547 · Cited by 215
Halbig, A., & Latoschik, M. E. (2021) · Frontiers in Virtual Reality · doi:10.3389/frvir.2021.694567 · Cited by 166
Li, K., Cardoso, C., Moctezuma-Ramirez, A., Elgalad, A., & Perin, E. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20247146 · Cited by 160
Pulse oximeter perfusion index as an early indicator of sympathectomy after epidural anesthesia
Ginosar, Y., Weiniger, C. F., Meroz, Y., Kurz, V., Bdolah‐Abram, T., Babchenko, A., Nitzan, M., & Davidson, E. M. (2009) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2009.01968.x · Cited by 157
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124
Coutrot, M., Dudoignon, E., Joachim, J., Gayat, É., Vallée, F., & Dépret, F. (2021) · Anaesthesia Critical Care & Pain Medicine · doi:10.1016/j.accpm.2021.100964 · Cited by 119
A Review of Deep Learning-Based Contactless Heart Rate Measurement Methods
Ni, A., Azarang, A., & Kehtarnavaz, N. (2021) · Sensors · doi:10.3390/s21113719 · Cited by 109
Measurement of respiratory rate using wearable devices and applications to COVID-19 detection
Natarajan, A., Su, H.-W., Heneghan, C., Blunt, L., O’Connor, C., & Niehaus, L. (2021) · npj Digital Medicine · doi:10.1038/s41746-021-00493-6 · Cited by 99
Hypertension Assessment Using Photoplethysmography: A Risk Stratification Approach
Liang, Y., Chen, Z., Ward, R. K., & Elgendi, M. (2018) · Journal of Clinical Medicine · doi:10.3390/jcm8010012 · Cited by 99
Mansi, S. A., Pigliautile, I., Arnesano, M., & Pisello, A. L. (2022) · Building and Environment · doi:10.1016/j.buildenv.2022.109385 · Cited by 90
Pulse Wave Amplitude Drops Index: A Biomarker of Cardiovascular Risk in Obstructive Sleep Apnea
Solelhac, G., Sánchez‐de‐la‐Torre, M., Blanchard, M., Berger, M., Hirotsu, C., Imler, T., Sánchez-de-la-Torre, A., Haba‐Rubio, J., Marchi, N. A., Bayon, V., et al. (2023) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.202206-1223oc · Cited by 94
Origin of Photoplethysmographic Waveform at Green Light
Kamshilin, A. A., & Борисович, М. Н. (2017) · Physics Procedia · doi:10.1016/j.phpro.2017.01.024 · Cited by 90
Deep Learning on 1-D Biosignals: a Taxonomy-based Survey
Ganapathy, N., Swaminathan, R., & Deserno, T. (2018) · Yearbook of Medical Informatics · doi:10.1055/s-0038-1667083 · Cited by 83
Lin, Q., Xu, J., Song, S., Breeschoten, A., Konijnenburg, M., Hoof, C. A. V., Tavernier, F., & Helleputte, N. V. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.3001449 · Cited by 79
An Overview of the Sensors for Heart Rate Monitoring Used in Extramural Applications
Galli, A., Montree, R. J. H., Que, S., Peri, E., & Vullings, R. (2022) · Sensors · doi:10.3390/s22114035 · Cited by 87
Approaches, Applications, and Challenges in Physiological Emotion Recognition—A Tutorial Overview
Can, Y. S., Mahesh, B., & André, E. (2023) · Proceedings of the IEEE · doi:10.1109/jproc.2023.3286445 · Cited by 81
Wearable Multiple Modality Bio-Signal Recording and Processing on Chip: A Review
Lin, Q., Song, S., Castro, I. D., Jiang, H., Konijnenburg, M., van Wegberg, R., Biswas, D., Stanzione, S., Sijbers, W., Hoof, C. A. V., et al. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3016115 · Cited by 72
Ramesh, J., Solatidehkordi, Z., Aburukba, R., & Sagahyroon, A. A. (2021) · Sensors · doi:10.3390/s21217233 · Cited by 73
Aldughayfiq, B., Ashfaq, F., Jhanjhi, N. Z., & Humayun, M. (2023) · Diagnostics · doi:10.3390/diagnostics13142442 · Cited by 75
Susana, E., Ramli, K., Murfi, H., & Apriantoro, N. H. (2022) · Information · doi:10.3390/info13020059 · Cited by 59
Chang, X., Li, G., Xing, G., Zhu, K., & Tu, L. (2021) · ACM Transactions on Sensor Networks · doi:10.1145/3441626 · Cited by 55
A practical guidance for assessments of sedentary behavior at work: A PEROSH initiative
Holtermann, A., Schellewald, V., Mathiassen, S. E., Gupta, N., Pinder, A., Punakallio, A., Veiersted, K. B., Weber, B., Takala, E., Draicchio, F., et al. (2017) · Applied Ergonomics · doi:10.1016/j.apergo.2017.03.012 · Cited by 56
Enabling the assessment of trauma-induced hemorrhage via smart wearable systems
Zia, J. S., Kimball, J. P., Rolfes, C., Hahn, J., & Inan, O. T. (2020) · Science Advances · doi:10.1126/sciadv.abb1708 · Cited by 51
Atrial fibrillation screening with photo-plethysmography through a smartphone camera
Verbrugge, F. H., Proesmans, T., Vijgen, J. M., Müllens, W., Rivero‐Ayerza, M., Herendael, H. V., Vandervoort, P. M., & Nuyens, D. (2019) · EP Europace · doi:10.1093/europace/euz119 · Cited by 50
Hirotsu, C., Betta, M., Bernardi, G., Marques-Vidal, P. M., Vollenweider, P., Waeber, G., Pichot, V., Roche, F., Siclari, F., Haba‐Rubio, J., et al. (2020) · SLEEP · doi:10.1093/sleep/zsz322 · Cited by 50
Warth, M., Keßler, J., Hillecke, T. K., & Bardenheuer, H. J. (2016) · Journal of Pain and Symptom Management · doi:10.1016/j.jpainsymman.2016.01.008 · Cited by 48
A novel approach for IoT based wearable health monitoring and messaging system
Manas, M., Sinha, A., Sharma, S., & Mahboob, M. R. (2018) · Journal of Ambient Intelligence and Humanized Computing · doi:10.1007/s12652-018-1101-z · Cited by 47
Recent advances in flexible and wearable OLEDs for biomedical applications: a review
Cho, E. H., Kim, Y., Sim, J., Yeon, H., Baek, S., Jeong, S. M., Lee, J., Jeon, Y., & Choi, K. C. (2025) · Materials Horizons · doi:10.1039/d5mh00742a · Cited by 47
Advances in photoplethysmography: beyond arterial oxygen saturation
Bartels, K., & Thiele, R. H. (2015) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/s12630-015-0458-0 · Cited by 46
The Impact of Heart Rate Variability Monitoring on Preventing Severe Cardiovascular Events
Turcu, A.-M., Ilie, A. C., Ștefăniu, R., Țăranu, S. M., Sandu, I. A., Alexa‐Stratulat, T., Pîslaru, A. I., & Alexa, I. D. (2023) · Diagnostics · doi:10.3390/diagnostics13142382 · Cited by 46
Oxygen saturation measurements in telemonitoring of patients with COPD: a systematic review
Buekers, J., Boever, P. D., Vaes, A. W., Aerts, J., Wouters, E. F. �., Spruit, M. A., & Theunis, J. (2017) · Expert Review of Respiratory Medicine · doi:10.1080/17476348.2018.1417842 · Cited by 43
Are Activity Wrist-Worn Devices Accurate for Determining Heart Rate during Intense Exercise?
Martín-Escudero, P., Cabanas, A. M., Dotor-Castilla, M. L., Galindo-Canales, M., Miguel‐Tobal, F., Fernández, C., Fuentes, M. E., & Giannetti, R. (2023) · Bioengineering · doi:10.3390/bioengineering10020254 · Cited by 41
Effects of Contact Pressure in Reflectance Photoplethysmography in an In Vitro Tissue-Vessel Phantom
May, J. M., Mejía‐Mejía, E., Nomoni, M., Budidha, K., Choi, C., & Kyriacou, P. A. (2021) · Sensors · doi:10.3390/s21248421 · Cited by 44
Susana, E., Ramli, K., Purnamasari, P. D., & Apriantoro, N. H. (2023) · Information · doi:10.3390/info14030145 · Cited by 45
Subaşı, A. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-817444-9.00002-7 · Cited by 37
A robust low-cost adaptive filtering technique for phonocardiogram signal denoising
Pauline, S. H., & Dhanalakshmi, S. (2022) · Signal Processing · doi:10.1016/j.sigpro.2022.108688 · Cited by 37
Lin, Q., Song, S., Wegberg, R. V., Sijbers, W., Biswas, D., Konijnenburg, M., Vanhoof, C., Tavernier, F., & van Helleputte, N. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3130470 · Cited by 30
Pulse Oximetry: The Working Principle, Signal Formation, and Applications
Leppänen, T., Kainulainen, S., Korkalainen, H., Sillanmäki, S., Kulkas, A., Töyräs, J., & Nikkonen, S. (2022) · Advances in experimental medicine and biology · doi:10.1007/978-3-031-06413-5_12 · Cited by 46
Development of a Wearable In-Ear PPG System for Continuous Monitoring
Pedrana, A., Comotti, D., Re, V., & Traversi, G. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3008479 · Cited by 35
A Systematic Review and IoMT Based Big Data Framework for COVID-19 Prevention and Detection
Hamid, S., Bawany, N. Z., Sodhro, A. H., Lakhan, A. R., & Ahmed, S. (2022) · Electronics · doi:10.3390/electronics11172777 · Cited by 35
Chu, J., Yang, W.-T., Lu, W.-R., Chang, Y.-T., Hsieh, T.-H., & Yang, F.-L. (2021) · Sensors · doi:10.3390/s21237815 · Cited by 34
Genetic determination of the vascular reactions in humans in response to the diving reflex
Baranova, T. I., Berlov, D. N., Glotov, O. S., Korf, E. A., Minigalin, A. D., Mitrofanova, A. V., Ahmetov, I. I., & Glotov, A. S. (2016) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00080.2016 · Cited by 34
Central Sleep Apnea Detection by Means of Finger Photoplethysmography
Massie, F., Vits, S., Khachatryan, A., Pee, B. V., Verbraecken, J. A., & Bergmann, J. (2023) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2023.3236393 · Cited by 34
Garrett, A., Kim, B., Sie, E. J., Gurel, N. Z., Marsili, F., Boas, D. A., & Roblyer, D. M. (2023) · Biomedical Optics Express · doi:10.1364/boe.482740 · Cited by 32
Evaluation of Remote Photoplethysmography Measurement Conditions toward Telemedicine Applications
Tohma, A., Nishikawa, M., Hashimoto, T., Yamazaki, Y., & Sun, G. (2021) · Sensors · doi:10.3390/s21248357 · Cited by 31
Ninety years of pulse oximetry: history, current status, and outlook
Quaresima, V., Ferrari, M., & Scholkmann, F. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33307 · Cited by 29
Kumaran, B., & Watson, T. (2018) · Electromagnetic Biology and Medicine · doi:10.1080/15368378.2017.1422260 · Cited by 28
Gautam, Y., & Jebelli, H. (2024) · Automation in Construction · doi:10.1016/j.autcon.2024.105537 · Cited by 27
Non-Contact Photoplethysmogram and Instantaneous Heart Rate Estimation from Infrared Face Video
Martinez, N., Bertrán, M., Sapiro, G., & Wu, H. (2019) · Proceedings - International Conference on Image Processing · doi:10.1109/icip.2019.8803109 · Cited by 27
TransPhys: Transformer-based unsupervised contrastive learning for remote heart rate measurement
Wang, R., Sun, H.-M., Hao, R.-R., Pan, A., & Jia, R. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105058 · Cited by 27
Warth, M., Koehler, F., Weber, M., Bardenheuer, H. J., Ditzen, B., & Keßler, J. (2019) · BMC Palliative Care · doi:10.1186/s12904-019-0397-6 · Cited by 27
Noncontact Cardiac Parameters Estimation Using Radar Acoustics for Healthcare IoT
Rong, Y., Lenz, I., & Bliss, D. W. (2023) · IEEE Internet of Things Journal · doi:10.1109/jiot.2023.3317670 · Cited by 26
Advances in Photoplethysmography for Personalized Cardiovascular Monitoring
Kim, S., Xiao, X., & Chen, J. (2022) · Biosensors · doi:10.3390/bios12100863 · Cited by 25
Ripollés‐Melchor, J., Casans‐Francés, R., Espinosa, Á. V., Gurumeta, A. A., Feldheiser, A., López‐Timoneda, F., & Vecino, J. M. C. (2016) · Revista Española de Anestesiología y Reanimación (English Edition) · doi:10.1016/j.redare.2016.03.001 · Cited by 25
Cugmas, B., Štruc, E., & Spīgulis, J. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aaf433 · Cited by 24
Lee, H.-T., Shim, M., Liu, X., Cheon, H.-R., Kim, S., Han, C. H., & Hwang, H.-J. (2025) · Biomedical Engineering Letters · doi:10.1007/s13534-025-00469-5 · Cited by 29
Long, N. M. H., & Chung, W. (2022) · IEEE photonics journal · doi:10.1109/jphot.2022.3153506 · Cited by 24
Res-BiANet: A Hybrid Deep Learning Model for Arrhythmia Detection Based on PPG Signal
Wu, Y., Tang, Q., Zhan, W., Li, S., & Chen, Z. (2024) · Electronics · doi:10.3390/electronics13030665 · Cited by 25
Kim, S., Gäbel, C., Aguilar‐Raab, C., Hillecke, T. K., & Warth, M. (2018) · The Arts in Psychotherapy · doi:10.1016/j.aip.2018.06.002 · Cited by 27
On the Personalized Learning Space in Educational Metaverse Based on Heart Rate Signal
Zhao, Z., Zhao, B., Ji, Z., & Liang, Z. (2022) · International Journal of Information and Communication Technology Education · doi:10.4018/ijicte.314565 · Cited by 22
Intelligent sensing devices and systems for personalized mental health
Xing, Y., Yang, Y., Yang, K., Lu, A., Xing, L., Mackie, K., & Guo, F. (2025) · Med-X · doi:10.1007/s44258-025-00057-3 · Cited by 25
Mo, W., Mohan, R., Li, W., Zhang, X., Sellke, E. W., Fan, W., DiMaio, J. M., & Thatcher, J. E. (2015) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2080699 · Cited by 21
Choi, B., Yim, J. Y., Shin, H., & Noh, G. (2021) · Journal of Medical Internet Research · doi:10.2196/23920 · Cited by 22
Polak, A. G., Klich, B., Saganowski, S., Prucnal, M. A., & Kazienko, P. (2022) · Sensors · doi:10.3390/s22187047 · Cited by 21
SensCon: Embedding Physiological Sensing into Virtual Reality Controllers
Chiossi, F., Kosch, T., Menghini, L., Villa, S., & Mayer, S. (2023) · Proceedings of the ACM on Human-Computer Interaction · doi:10.1145/3604270 · Cited by 21
A supervised learning approach for the robust detection of heart beat in plethysmographic data
Grisan, E., Cantisani, G., Tarroni, G., Yoon, S. K., & Rossi, M. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319716 · Cited by 21
Wang, D., Hu, Q., & Yang, C. (2022) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2022.105654 · Cited by 20
Costa, T. B. D. S., Dias, F. M., Cárdenas, D. A. C., de Toledo, M. A. F., de Lima, D. M., Krieger, J. E., & Gutierrez, M. A. (2023) · IEEE Access · doi:10.1109/access.2023.3284458 · Cited by 21
Li, M., Song, S., Wang, D., Yang, T., Zheng, Y., Lin, Q., Zheng, F., Huang, K., Tan, Z., & Zhao, M. (2023) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2023.3335068 · Cited by 20
Noninvasive In Vivo Estimation of Blood-Glucose Concentration by Monte Carlo Simulation
Haque, C. A., Hossain, S., Kwon, T.-H., & Kim, K. (2021) · Sensors · doi:10.3390/s21144918 · Cited by 19
Identification of pyridine analogs as new predator-derived kairomones
Brechbühl, J., Moine, F., Nenniger-Tosato, M., Sporkert, F., & Broillet, M.-C. (2015) · Frontiers in Neuroscience · doi:10.3389/fnins.2015.00253 · Cited by 19
Pal, R., Rudas, Á., Kim, S., Chiang, J. N., Barney, A. M., & Cannesson, M. P. (2024) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2024.108283 · Cited by 22
DeepHeart: Accurate Heart Rate Estimation from PPG Signals Based on Deep Learning
Chang, X., Li, G., Tu, L., Xing, G., & Tian, H. (2019) · IEEE International Conference on Mobile Ad-Hoc and Sensor Systems (MASS) · doi:10.1109/mass.2019.00051 · Cited by 18
Prada, E. J. A., & García, J. F. C. (2024) · Sensors · doi:10.3390/s24227193 · Cited by 18
Schoenmakers, M., Saygin, M., Sikora, M. T., Vaessen, T., Noordzij, M. L., & de Geus, E. J. C. (2025) · Behavior Research Methods · doi:10.3758/s13428-025-02685-4 · Cited by 19
Song, S., Lin, Q., Hoof, C. A. V., & Helleputte, N. V. (2020) · IEEE Transactions on Circuits & Systems II Express Briefs · doi:10.1109/tcsii.2020.3002952 · Cited by 16
Are Wearable ECG Devices Ready for Hospital at Home Application?
Medina, J., Silva, R. D. S., & Holgado-Terriza, J. A. (2025) · Sensors · doi:10.3390/s25102982 · Cited by 15
Ripollés‐Melchor, J., Casans‐Francés, R., Espinosa, Á. V., Gurumeta, A. A., Feldheiser, A., López‐Timoneda, F., & Vecino, J. M. C. (2016) · Revista Española de Anestesiología y Reanimación · doi:10.1016/j.redar.2015.07.009 · Cited by 15
Saha, S., Jindal, K., Shakti, D., Tewary, S., & Sardana, V. (2022) · Expert Systems · doi:10.1111/exsy.12958 · Cited by 17
Ajtay, B. E., Béres, S., & Hejjel, L. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.104033 · Cited by 18
Tong, Y., Huang, Z., Zhang, Z., Yin, M., Shan, G., Wu, J., & Qin, F. (2023) · Measurement · doi:10.1016/j.measurement.2023.113605 · Cited by 14
Comparison of multiple cardiac signal acquisition technologies for heart rate variability analysis
Charlier, P., Cabon, M., Herman, C., Benouna, F., Logier, R., Houfflin‐Debarge, V., Jeanne, M., & Jonckheere, J. D. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00382-0 · Cited by 14
Human–Machine Interaction in Driving Assistant Systems for Semi-Autonomous Driving Vehicles
Lee, H.-G., Kang, D.-H., & Kim, D. (2021) · Electronics · doi:10.3390/electronics10192405 · Cited by 15
Quantification of the Phenomena Affecting Reflective Arterial Photoplethysmography
Rovas, G., Bikia, V., & Stergiopulos, N. (2023) · Bioengineering · doi:10.3390/bioengineering10040460 · Cited by 14
A Smart Health System: Monitoring Comatose Patient’s Physiological Conditions Remotely
Erlina, T., Marreza, R. S., & Putri, R. E. (2018) · International Conference on Information Technology Systems and Innovation · doi:10.1109/icitsi.2018.8696094 · Cited by 13
Liu, G., Hu, X., Wang, E., Zhou, G., Cai, J., & Zhang, S. (2019) · Computational and Mathematical Methods in Medicine · doi:10.1155/2019/5363712 · Cited by 13
Heart Rate assessment by means of a novel approach applied to signals of different nature
Cosoli, G., Casacanditella, L., Tomasini, E. P., & Scalise, L. (2017) · Journal of Physics Conference Series · doi:10.1088/1742-6596/778/1/012001 · Cited by 13
Huang, Y., & Deng, Y. (2022) · Applied Sciences · doi:10.3390/app12126007 · Cited by 12
Ceugniez, M., Devanne, H., & Hermand, É. (2024) · JMIR mhealth and uhealth · doi:10.2196/54871 · Cited by 13
Prucnal, M. A., Polak, A. G., & Kazienko, P. (2024) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2024.107336 · Cited by 12
Comparison of Different Methods to Estimate Blood Oxygen Saturation using PPG
Haque, C. A., Hossain, S., Kwon, T.-H., & Kim, K. (2021) · International Conference on Information and Communication Technology Convergence (ICTC) · doi:10.1109/ictc52510.2021.9621142 · Cited by 12
Prada, E. J. A., & Valencia, R. M. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.104335 · Cited by 12
Using cardiac output monitoring to guide perioperative haemodynamic therapy
McGuinness, S., & Parke, R. L. (2015) · Current Opinion in Critical Care · doi:10.1097/mcc.0000000000000212 · Cited by 13
Athaya, T., & Choi, S. (2021) · Journal of Sensors · doi:10.1155/2021/9925033 · Cited by 12
Metadata
- Authors
- Alian, A. A., & Shelley, K. H.
- Affiliations
- Yale University (US)
- Corresponding author
- Shelley, Kirk H.
- Venue
- Best Practice & Research Clinical Anaesthesiology
- Publisher
- Elsevier BV
- Published
- 2014-12
- Volume
- 28
- Issue
- 4
- Pages
- 395-406
- ISSN
- 1521-6896 (ISSN-L)
- 1753-3740
- OpenAlex
- W4245509891
- Cited by
- 289
- References
- 83
- Citations by year
- 2026: 18 · 2025: 45 · 2024: 36 · 2023: 44 · 2022: 37 · 2021: 25 · 2020: 20 · 2019: 21 · 2018: 17 · 2017: 8 · 2016: 7 · 2015: 9
- FWCI
- 4.15
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- Best practice & research. Clinical anaesthesiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, pulse oximetry, arrhythmia monitoring, oxygen saturation, heart rate, hypovolemia, early goal-directed therapy, fluid management, vasopressors, inotropic agents
- MeSH terms
- Hemodynamics, Humans, Oximetry, Photoplethysmography, Hypovolemia
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review