Deep brain stimulation and electromagnetic interference
Open accessThe DOI matches its Crossref record; 61 of the 61 references it cites were found in a registry or an index.
Abstract
Deep brain stimulation (DBS) has evolved into an approved and efficacious treatment for movement, obsessive-compulsive, and epilepsy disorders that are refractory to medical therapy, with current investigation into other disease conditions. However, there are unintentional and intentional sources of external electromagnetic interference (EMI) that can lead to either malfunctioning or damaged DBS devices, as well as injury to human tissue. Comprehensive studies and guidelines on such topics in the medical literature are scarce. Herein, we review the principles behind EMI, as well as the various potential sources of interference, both unintentional (e.g. stray EMI fields) and intentional (e.g. MRI scans, "brainjacking"). Additionally, we employ the Manufacturer and User Device Facility Experience (MAUDE) database to assess real-world instances of EMI (e.g., airport body scanners, magnetic resonance imaging (MRI), and electrosurgery) affecting DBS devices commonly implanted in the United States (US).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“For example, RF systems for charging may induce voltages in nearby data circuits, causing data loss or errors, while power harvesting in battery-less systems might disrupt high-frequency data signals (Bazaka and Jacob, 2012; Rahimpour et al., 2021; Won et al., 2021).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Rahimpour et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“For example, RF systems for charging may induce voltages in nearby data circuits, causing data loss or errors, while power harvesting in battery-less systems might disrupt high-frequency data signals (Bazaka and Jacob, 2012; Rahimpour et al., 2021; Won et al., 2021).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Rahimpour et al., 2021
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Deep Brain Stimulation for Treatment-Resistant Depression
Mayberg, H. S., Lozano, A. M., Voon, V., McNeely, H. E., Seminowicz, D., Hamani, C., Schwalb, J. M., & Kennedy, S. H. (2005) · Neuron · doi:10.1016/j.neuron.2005.02.014 · Cited by 4,059
A Randomized Trial of Deep-Brain Stimulation for Parkinson's Disease
Deuschl, G., Schade-Brittinger, C., Krack, P., Volkmann, J., Schäfer, H., Bötzel, K., Daniels, C., Deutschländer, A., Dillmann, U., Eisner, W., et al. (2006) · New England Journal of Medicine · doi:10.1056/nejmoa060281 · Cited by 2,944
Bilateral Deep-Brain Stimulation of the Globus Pallidus in Primary Generalized Dystonia
Vidailhet, M., Vercueil, L., Houeto, J.-L., Krystkowiak, P., Benabid, A.-L., Cornu, P., Lagrange, C., Tézenas du Montcel, S., Dormont, D., Grand, S., et al. (2005) · New England Journal of Medicine · doi:10.1056/nejmoa042187 · Cited by 1,094
Chronic Anterior Thalamus Stimulation for Intractable Epilepsy
Hodaie, M., Wennberg, R. A., Dostrovsky, J. O., & Lozano, A. M. (2002) · Epilepsia · doi:10.1046/j.1528-1157.2002.26001.x · Cited by 605
MR Procedures: Biologic Effects, Safety, and Patient Care
Shellock, F. G., & Crues, J. V. (2004) · Radiology · doi:10.1148/radiol.2323030830 · Cited by 563
Electrotherapy Explained: Principles and Practice
Robertson, V. J., Ward, A., Low, J. N., & Reed, A. M. (1991) · Cited by 334
Directional deep brain stimulation: an intraoperative double-blind pilot study
Pollo, C., Kaelin-Lang, A., Oertel, M. F., Stieglitz, L., Taub, E., Fuhr, P., Lozano, A. M., Raabe, A., & Schüpbach, M. (2014) · Brain · doi:10.1093/brain/awu102 · Cited by 327 · cited 1×
“Besides two cylindrical electrodes, the 8CH has two central contacts consisting of three segments each, which can be activated independently to concentrate stimulation in one direction, thereby theoretically reducing current needed to produce meaningful therapeutic effects by up to 43 % [11].”
BackgroundHenderson, J. M., Tkach, J., Phillips, M., Baker, K., Shellock, F. G., & Rezai, A. R. (2005) · Neurosurgery · doi:10.1227/01.neu.0000180810.16964.3e · Cited by 314 · cited 1×
“Examples of reported consequences of EMI include: (1) development of unexpected thalamic lesions, secondary to cardioversion [19]; (2) aphasia, dysarthria, and hemiparesis, resulting from a radiofrequency lesion produced by heating of a DBS electrode during MRI of the lumbar spine [20]; and even (3) persistent vegetat…”
BackgroundSecurity and privacy issues in implantable medical devices: A comprehensive survey
Camara, C., Peris-Lopez, P., & Tapiador, J. E. (2015) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2015.04.007 · Cited by 293
Interference with Cardiac Pacemakers by Cellular Telephones
Hayes, D. L., Wang, P. J., Reynolds, D. W., Estes, N. A. M., Griffith, J. L., Steffens, R. A., Carlo, G. L., Findlay, G. K., & Johnson, C. M. (1997) · New England Journal of Medicine · doi:10.1056/nejm199705223362101 · Cited by 282 · cited 1×
“These devices are known to be potential sources of interference with pacemakers and defibrillators; however, very little is known regarding interference with DBS [29,30].”
BackgroundElectrotherapy Explained: Principles and practice
Blanchard, M. (1991) · Physiotherapy · doi:10.1016/s0031-9406(10)61810-x · Cited by 214
Hacking the brain: brain–computer interfacing technology and the ethics of neurosecurity
Ienca, M., & Haselager, P. (2016) · Ethics and Information Technology · doi:10.1007/s10676-016-9398-9 · Cited by 195
Brainjacking: Implant Security Issues in Invasive Neuromodulation
Pycroft, L., Boccard, S. G., Owen, S. L. F., Stein, J. F., Fitzgerald, J. J., Green, A. L., & Aziz, T. Z. (2016) · World Neurosurgery · doi:10.1016/j.wneu.2016.05.010 · Cited by 177 · cited 1×
“Such attacks can be categorized as passive (“listening” for information during normal transmission) which may result in disruptions in authentication or loss of confidentiality, or active (send or modify messages to the IPG) which can alter therapy and side effect profile of stimulation [62].”
BackgroundDBS and diathermy interaction induces severe CNS damage
Nutt, J. G., Anderson, V. C., Peacock, J. H., Hammerstad, J. P., & Burchiel, K. J. (2001) · Neurology · doi:10.1212/wnl.56.10.1384 · Cited by 174 · cited 2×
“For example, case reports have described EMI causing DBS devices to inadvertently switch between ‘on’ and ‘off’ states, as well as permanent neurological damage [16].”
BackgroundDeep brain stimulation for essential tremor
Hubble, J. P., Busenbark, K. L., Wilkinson, S., Penn, R. D., Lyons, K., & Koller, W. C. (1996) · Neurology · doi:10.1212/wnl.46.4.1150 · Cited by 174
Beinart, R., & Nazarian, S. (2013) · Circulation · doi:10.1161/circulationaha.113.005697 · Cited by 173
Lipsman, N., Neimat, J. S., & Lozano, A. M. (2007) · Neurosurgery · doi:10.1227/01.neu.0000279719.75403.f7 · Cited by 151
Long-Term Results of Deep Brain Stimulation of the Anterior Cingulate Cortex for Neuropathic Pain
Boccard, S., Prangnell, S. J., Pycroft, L., Cheeran, B., Moir, L., Pereira, E. A. C., FitzGerald, J. J., Green, A. L., & Aziz, T. Z. (2017) · World Neurosurgery · doi:10.1016/j.wneu.2017.06.173 · Cited by 151
MAUDE (Manufacturer and User Facility Device Experience)
(2017) · Cited by 139
Loh, S. A., Carlson, G. A., Chang, E. I.-F., Huang, E. J., Palanker, D., & Gurtner, G. C. (2009) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0b013e3181bcee87 · Cited by 130 · cited 1×
“Indeed, the cutting surface of the PlasmaBlade operates at 40 °C and 100 °C, which is significantly lower than the 250–350 °C found in conventional electrocautery electrodes [48].”
BackgroundSecurity of implantable medical devices with wireless connections: The dangers of cyber-attacks
Pycroft, L., & Aziz, T. Z. (2018) · Expert Review of Medical Devices · doi:10.1080/17434440.2018.1483235 · Cited by 129
Deep Brain Stimulation in Movement and Psychiatric Disorders
Hardesty, D. E., & Sackeïm, H. A. (2006) · Biological Psychiatry · doi:10.1016/j.biopsych.2006.08.028 · Cited by 113
Neurosecurity: security and privacy for neural devices
Denning, T., Matsuoka, Y., & Kohno, T. (2009) · Neurosurgical Focus · doi:10.3171/2009.4.focus0985 · Cited by 112
Brainjacking in deep brain stimulation and autonomy
Pugh, J., Pycroft, L., Sandberg, A., Aziz, T. Z., & Savulescu, J. (2018) · Ethics and Information Technology · doi:10.1007/s10676-018-9466-4 · Cited by 111 · cited 1×
“Depending on the indication for DBS and the side effect profile, intentional third-party interference can theoretically undermine a patient’s autonomy [61].”
BackgroundSystems for deep brain stimulation: review of technical features
Amon, A., & Alesch, F. (2017) · Journal of Neural Transmission · doi:10.1007/s00702-017-1751-6 · Cited by 103
Mobile phone interference with medical equipment and its clinical relevance: a systematic review
Lawrentschuk, N., & Bolton, D. M. (2004) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.2004.tb06205.x · Cited by 99 · cited 1×
“Clinically relevant EMI with devices such as cell phones is uncommon with distances greater than 1 mm [22].”
BackgroundElectromagnetic Interference and Implanted Cardiac Devices: The Nonmedical Environment (Part I)
Misiri, J., Kusumoto, F. M., & Goldschlager, N. (2012) · Clinical Cardiology · doi:10.1002/clc.21998 · Cited by 91
Healey, J. S., Merchant, R. N., Simpson, C., Tang, T. K. K., Beardsall, M., Tung, S., Fraser, J. A., Long, L., van Vlymen, J. M., Manninen, P. H., et al. (2012) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2011.08.121 · Cited by 73
Electromagnetic Interference and Implanted Cardiac Devices: The Medical Environment (Part II)
Misiri, J., Kusumoto, F. M., & Goldschlager, N. (2012) · Clinical Cardiology · doi:10.1002/clc.21997 · Cited by 71
Clinically significant magnetic interference of implanted cardiac devices by portable headphones
Lee, S., Fu, K. E., Kohno, T., Ransford, B., & Maisel, W. H. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.07.003 · Cited by 67 · cited 1×
“Of those tested, 30 patients were shown to have clinically relevant interference [33].”
BackgroundClinical study of interference with cardiac pacemakers by a magnetic field at power line frequencies
TRIGANO, A. J., Blandeau, O., Souques, M., Gernez, J., & Magne, I. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2004.10.072 · Cited by 62 · cited 1×
“Other sources of electricity, such as high voltage power lines and transformers, have also been known to affect patients with other implantable devices, though not specifically DBS [36].”
BackgroundBaker, K. B., Nyenhuis, J. A., Hrdlicka, G., Rezai, A. R., Tkach, J. A., & Shellock, F. G. (2004) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20224 · Cited by 60 · cited 1×
“Testing of newer generation DBS systems with respect to these forces has shown that the magnitude of static field interactions has diminished to negligible levels [39].”
BackgroundGray, R. W., Bibens, W. T., & Shellock, F. G. (2005) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2005.07.005 · Cited by 57
Deep brain stimulation for treatment-resistant depression.
Taghva, A. S., Malone, D. A., & Rezai, A. R. (2013) · World Neurosurgery · doi:10.1016/j.wneu.2012.10.068 · Cited by 56
Cutaneous Electrosurgery in a Patient with a Deep Brain Stimulator
Weaver, J., Kim, S. J., Torres, A., & Jd (1999) · Dermatologic Surgery · doi:10.1046/j.1524-4725.1999.08167.x · Cited by 54 · cited 1×
“Indeed, monopolar electrocautery might provoke electrical shocks when the IPG is not turned off; therefore, it is recommended that the IPG voltage be set to zero to avoid EMI when monopolar electrocautery is employed [46].”
BackgroundThalamotomy Caused by Cardioversion in a Patient Treated with Deep Brain Stimulation
Yamamoto, T., Katayama, Y., Fukaya, C., Kurihara, J., Oshima, H., & Kasai, M. (2000) · Stereotactic and Functional Neurosurgery · doi:10.1159/000056466 · Cited by 53 · cited 1×
“Examples of reported consequences of EMI include: (1) development of unexpected thalamic lesions, secondary to cardioversion [19]; (2) aphasia, dysarthria, and hemiparesis, resulting from a radiofrequency lesion produced by heating of a DBS electrode during MRI of the lumbar spine [20]; and even (3) persistent vegetat…”
BackgroundInterference by cellular phones with permanent implanted pacemakers: an update*1
Hekmat, K., Salemink, B., Lauterbach, G., Schwinger, R. H. G., Südkamp, M., Weber, H. J., & Mehlhorn, U. (2004) · EP Europace · doi:10.1016/j.eupc.2004.03.010 · Cited by 53 · cited 1×
“These devices are known to be potential sources of interference with pacemakers and defibrillators; however, very little is known regarding interference with DBS [29,30].”
BackgroundWadasadawala, T., Pandey, A., Agarwal, J. P., Jalali, R., Laskar, S. G., Chowdhary, S., Budrukkar, A., Sarin, R., Deshpande, D., & Munshi, A. (2010) · Clinical Oncology · doi:10.1016/j.clon.2010.08.031 · Cited by 49 · cited 1×
“silicon dioxide insulator with excess electron-hole pairs, which may accumulate and create a net positive charge resulting in device malfunction [54].”
BackgroundVOUTSALATH, M. A., Bichakjian, C. K., Pelosi, F., BLUM, D., Johnson, T. M., & Farrehi, P. M. (2011) · Dermatologic Surgery · doi:10.1111/j.1524-4725.2011.02006.x · Cited by 48 · cited 1×
“However, the absence of a grounding pad may permit currents to disperse beyond the surgical site throughout the patient’s body, leading to potential EMI [45].”
BackgroundShrivastava, D., Abosch, A., Hughes, J. R., Goerke, U., DelaBarre, L. J., Visaria, R. K., Harel, N., & Vaughan, J. T. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/17/5651 · Cited by 45 · cited 1×
“In a prior study carried out by Shrivastava and colleagues in 2012, a cadaveric porcine model was designed to measure heat induction by MRI imaging [41].”
The Rate of Magnetic Resonance Imaging in Patients with Deep Brain Stimulation
Falowski, S. M., Safriel, Y. I., Ryan, M. P., & Hargens, L. M. (2016) · Stereotactic and Functional Neurosurgery · doi:10.1159/000444760 · Cited by 43
Letter to the Editor—Lifesaving therapy inhibition by phones containing magnets
Greenberg, J. C., Altawil, M., & Singh, G. (2021) · Heart Rhythm · doi:10.1016/j.hrthm.2020.12.032 · Cited by 41
Magnetic resonance safety testing of a newly‐developed fiber‐optic cardiac pacing lead
Greatbatch, W., Miller, V., & Shellock, F. G. (2002) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10129 · Cited by 37 · cited 1×
“Several factors, however, have been previously investigated to help mitigate and minimize interference, including the use of filtering systems used to filtering out the frequency bands of MRI, or the addition of fiber-optic components [66].”
BackgroundMartinelli, P. T., Schulze, K. E., & Nelson, B. R. (2004) · Dermatologic Surgery · doi:10.1111/j.1524-4725.2004.30308.x · Cited by 36
Kypta, A., Blessberger, H., SALEH, K., Hönig, S., Kammler, J., Neeser, K., & Steinwender, C. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12488 · Cited by 33 · cited 1×
“At the same time, replacement of implanted pacemakers and defibrillator generators using PlasmaBlade has been associated with reduced risk of lead damage, shorter hospital stays, and greater cost effectiveness when compared to scissors and conventional electrocautery [49].”
BackgroundFraix, V., Chabardès, S., Krainik, A., Seigneuret, É., Grand, S., Bas, J. L., Krack, P., Benabid, A., & Pollak, P. (2010) · Journal of neurosurgery · doi:10.3171/2010.1.jns09951 · Cited by 33 · cited 1×
“Reed switches were employed in early DBS system iterations to allow the patient or clinician to easily turn the device on or off using a handheld magnet [17]).”
MethodYu, S. S., Tope, W. D., & Grekin, R. C. (2005) · Dermatologic Surgery · doi:10.1111/j.1524-4725.2005.31808 · Cited by 33 · cited 2×
“Thermocautery helps achieve hemostasis by providing direct thermal energy at the metal tip, but without delivering any current to the patient [42].”
BackgroundMethodPaniccia, A., Rozner, M. A., Jones, E. L., Townsend, N. T., Varosy, P. D., Dunning, J. E., Girard, G., Weyer, C., Stiegmann, G. V., & Robinson, T. N. (2014) · The American Journal of Surgery · doi:10.1016/j.amjsurg.2014.09.011 · Cited by 30 · cited 1×
“However, a recent study on transvenous cardiac defibrillators implanted in anesthetized pigs showed that the PlasmaBlade in the coagulation mode actually transferred more EMI to the implanted device than conventional monopolar and bipolar devices [50].”
BackgroundSecuring Wireless Neurostimulators
Marin, E., Singelée, D., Yang, B., Volski, V. A., Vandenbosch, G. A. E., Nuttin, B., & Preneel, B. (2018) · ACM Conference on Data and Application Security and Privacy (CODASPY) · doi:10.1145/3176258.3176310 · Cited by 29
Czyżewski, A., Kotus, J., & Szwoch, G. (2019) · Remote Sensing · doi:10.3390/rs12010110 · Cited by 29 · cited 1×
“Remote traffic microwave sensors used to detect and measure traffic at the intersections of roadways can also be a cause of EMI by radiation interference [28].”
BackgroundElectromagnetic Environmental Influences on Implanted Deep Brain Stimulators
Blomstedt, P. C., Jabre, M., Bejjani, B.-P. W., & Koskinen, L. (2006) · Neuromodulation Technology at the Neural Interface · doi:10.1111/j.1525-1403.2006.00068.x · Cited by 28 · cited 3×
“For the patient with ET or PD benefiting from symptom relief or amelioration, EMI may result in recurrence of symptoms and increased disability, with potentially fatal outcomes [10].”
BackgroundMethodStandards for Medical Device Cybersecurity in 2018
Yuan, S., Fernando, A., & Klonoff, D. C. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818763634 · Cited by 26 · cited 1×
“Recent FDA guidelines may help reduce cybersecurity threats that are inherent to design features [65].”
BackgroundOptimization of Thermocautery in Excisional Dermatologic Surgery
LANE, J. E., O'Brien, E. M., & Kent, D. E. (2006) · Dermatologic Surgery · doi:10.1097/00042728-200605000-00010 · Cited by 20
Safety and interaction of patients with implantable cardiac defibrillators driving a hybrid vehicle
Tondato, F. J., Bazzell, J. L., Schwartz, L., Donald, B. W. M., Fisher, R., Anderson, S. S., Galindo, A. F. S., Dueck, A. C., & Scott, L. R. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2016.11.090 · Cited by 18 · cited 1×
“The increasing prevalence of electric and hybrid cars prompted a study by the Mayo Clinic to assess potential electromagnetic interference with cardiac defibrillators [34].”
BackgroundKainz, W., Alesch, F., & Chan, D. D. (2003) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-2-11 · Cited by 16 · cited 1×
“In a phantom study using the ITREL-III DBS model, cellular GSM phones with field strengths of 900 Hz and 1800 Hz were determined to cause neither inhibition nor shape changes to the stimulation pulse [31].”
MethodDeep Brain Stimulation for Treatment-Resistant Depression
Kaur, N., Chou, T., Corse, A. K., Arulpragasam, A. R., Deckersbach, T., & Evans, K. C. (2013) · Psychiatric Annals · doi:10.3928/00485713-20130806-04 · Cited by 16
Corzani, A., Ziacchi, M., Biffi, M., Allaria, L., Diemberger, I., Martignani, C., Bratten, T., Gardini, B., & Boriani, G. (2015) · Journal of Cardiovascular Medicine · doi:10.2459/jcm.0000000000000301 · Cited by 13
Art as Therapy for Parkinson's Disease
Wadeson, H. (2003) · Art Therapy · doi:10.1080/07421656.2003.10129636 · Cited by 12
Hybrid cars may interfere with implanted deep brain stimulators
Chen, C., Cole, W., & Bronte-Stewart, H. M. (2009) · Movement Disorders · doi:10.1002/mds.22739 · Cited by 3 · cited 1×
“Regardless, case reports do exist of hybrid vehicle interference with DBS devices [35].”
BackgroundA curious case of DBS radiofrequency programmer interference
Grewal, S. S., ReFaey, K., Grassle, A. L., Uitti, R. J., & Wharen, R. E. (2019) · npj Parkinson s Disease · doi:10.1038/s41531-019-0075-7 · Cited by 2 · cited 12×
“Each DBS device consists of three components: an implantable pulse generator (IPG), 4-channel (4CH) or 8-channel (8CH) leads implanted within the brain, and an extension cable connecting these two components.”
BackgroundInfluentialCardiac Devices and Electromagnetic Interference Revisited
Yu, S. S., Tope, W. D., & Grekin, R. C. (2005) · Dermatologic Surgery · doi:10.1097/00042728-200508000-00008
Research works printed but not linked: 3
- Arle J, Shils JL (Eds.), Essential Neuromodulation, Academic Press, 2011, pp. 325– 348. Apr 14. Chapter 15.
- SJM, Lead and Extension Kits for Deep Brain Stimulation Systems – Clinician’s Manual, 2017, pp. 1– 50 (null).
- Kainz W, Alesch F, Chan DD, Electromagnetic interference of GSM mobile phones with the implantable deep brain stimulator, ITREL-III, Biomed. Eng. Online 2 (2003) 11. 12773204 PMC156641 · DOI · Paper page
Cited by
A review on electromagnetic shielding magnesium alloys
Liu, L., Chen, X., & Pan, F. (2021) · Journal of Magnesium and Alloys · doi:10.1016/j.jma.2021.10.001 · Cited by 193
Zhang, T., Pan, N., Wang, Y., Liu, C., & Hu, S. (2021) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2021.749162 · Cited by 177
Advancements in 3D-printed architectures for electromagnetic interference shields
Zhang, G., Wang, H., Xie, W., Zhou, S., Nie, Z., Niwamanya, G., Zhao, Z., & Duan, H. (2024) · Journal of Materials Chemistry A · doi:10.1039/d3ta07181b · Cited by 61
Excitatory‐inhibitory modulation of transcranial focus ultrasound stimulation on human motor cortex
Zhang, T., Guo, B., Zuo, Z., Long, X., Hu, S., Li, S., Su, X., Wang, Y., & Liu, C. (2023) · CNS Neuroscience & Therapeutics · doi:10.1111/cns.14303 · Cited by 60
Li, J., Zhong, L., Wang, J., Feng, Z., Qu, Y. D., & Xu, R. (2024) · Journal of Magnesium and Alloys · doi:10.1016/j.jma.2024.03.013 · Cited by 44
Metadata
- Authors
- Rahimpour, S., Kiyani, M., Hodges, S. E., & Turner, D. A.
- Affiliations
- Duke University (US)
- Duke Medical Center (US)
- Corresponding author
- Rahimpour, Shervin
- Venue
- Clinical Neurology and Neurosurgery
- Publisher
- Elsevier BV
- Published
- 2021-04
- Volume
- 203
- Pages
- 106577
- ISSN
- 0303-8467 (ISSN-L)
- PMC
- PMC8081063
- OpenAlex
- W3131287423
- Cited by
- 63
- References
- 61
- Citations by year
- 2026: 11 · 2025: 12 · 2024: 13 · 2023: 17 · 2022: 4 · 2021: 5
- FWCI
- 1.25
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- doi.org/10.15223/policy-017 (stm-asf)
- doi.org/10.15223/policy-037 (stm-asf)
- doi.org/10.15223/policy-012 (stm-asf)
- doi.org/10.15223/policy-029 (stm-asf)
- doi.org/10.15223/policy-004 (stm-asf)
- Locations
- Clinical Neurology and Neurosurgery (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Neurological Disorders and Stroke (UH3NS103468, UH3NS129898)
- Topics
- Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
- Life Sciences › Neuroscience › Neurology › Transcranial Magnetic Stimulation Studies
- Social Sciences › Psychology › Clinical Psychology › Obsessive-Compulsive Spectrum Disorders
- Keywords
- deep brain stimulation, electromagnetic interference, MRI safety, electrosurgery, MAUDE database, movement disorders, epilepsy, obsessive-compulsive disorder
- MeSH terms
- Electromagnetic Fields, Electronics, Humans, Deep Brain Stimulation, Implantable Neurostimulators
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.