The DOI matches its Crossref record; 43 of the 43 references it cites were found in a registry or an index.
Needle insertion into soft tissue: A survey
Abstract
Needle insertion in soft tissue has attracted considerable attention in recent years due to its application in minimally invasive percutaneous procedures such as biopsies and brachytherapy. This paper presents a survey of the current state of research on needle insertion in soft tissue. It examines the topic from several aspects, e.g. modeling needle insertion forces, modeling tissue deformation and needle deflection during insertion, robot-assisted needle insertion, and the effect of different trajectories on tissue deformation. All studies show that the axial force of a needle during insertion in soft tissue is the summation of different forces distributed along the needle shaft such as stiffness force, frictional force and cutting force. Some studies have modeled these forces. The force data in some procedures is used for identifying tissue layers as the needle is inserted or for path planning. Needle deflection and tissue deformation are major problems for accurate needle insertion and attempts have been made to model them. Using current models several insertion techniques have been developed which are briefly reviewed in this paper.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Zienkiewicz, O. C. (1989) · Cited by 14,802
To Err Is Human: Building a Safer Health System
(2000) · National Academies Press eBooks · doi:10.17226/9728 · Cited by 14,044
To Err is Human: Building a Safer Health System
Hendee, W. R. (2001) · Journal of Vascular and Interventional Radiology · doi:10.1016/s1051-0443(01)70072-3 · Cited by 10,103
Biomechanics. Mechanical Properties of Living Tissues
Fung, Y. C., & Skalak, R. (1982) · Journal of Applied Mechanics · doi:10.1115/1.3162171 · Cited by 6,025
Nonlinear programming: theory and algorithms
(1993) · Computers & Operations Research · doi:10.1016/0305-0548(93)90042-h · Cited by 5,779
Barry Issenberg, S., Mcgaghie, W. C., Petrusa, E. R., Lee Gordon, D., & Scalese, R. J. (2005) · Medical Teacher · doi:10.1080/01421590500046924 · Cited by 3,966
Biomechanics: Mechanical Properties of Living Tissues
Fung, Y. C., & Skalak, R. (1981) · Journal of Biomechanical Engineering · doi:10.1115/1.3138285 · Cited by 2,769
Elastic Moduli of Breast and Prostate Tissues under Compression
Krouskop, T. A., Wheeler, T. M., Kallel, F., Garra, B. S., & Hall, T. (1998) · Ultrasonic Imaging · doi:10.1177/016173469802000403 · Cited by 1,703
Nonlinear programming: Theory and algorithms
Plastria, F. (1994) · European Journal of Operational Research · doi:10.1016/0377-2217(94)90039-6 · Cited by 1,467
Medical robotics in computer-integrated surgery
Taylor, R. H., & Stoianovici, D. (2003) · IEEE Transactions on Robotics and Automation · doi:10.1109/tra.2003.817058 · Cited by 997
Force Modeling for Needle Insertion Into Soft Tissue
Okamura, A. M., Simone, C., & O'Leary, M. D. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.831542 · Cited by 830
Biomechanics: Mechanical Properties of Living Tissues, 2nd ed.
Fung, Y. C., & Cowin, S. C. (1994) · Journal of Applied Mechanics · doi:10.1115/1.2901550 · Cited by 745
Nonholonomic Modeling of Needle Steering
Webster, R. J., Cowan, N. J., Chirikjian, G., & Okamura, A. M. (2006) · Springer Tracts in Advanced Robotics · doi:10.1007/11552246_4 · Cited by 527
Needle insertion modeling and simulation
DiMaio, S. P., & Salcudean, S. E. (2003) · IEEE Transactions on Robotics and Automation · doi:10.1109/tra.2003.817044 · Cited by 518
Tissue response to mechanical vibrations for “sonoelasticity imaging”
Parker, K. J., Huang, S. R., Musulin, R. A., & Lerner, R. M. (1990) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(90)90003-u · Cited by 478
Finite element modeling in surgery simulation
Bro-Nielsen, M. (1998) · Proceedings of the IEEE · doi:10.1109/5.662874 · Cited by 401
The role of force feedback in surgery: analysis of blunt dissection
Wagner, C. R., Stylopoulos, N., & Howe, R. D. (2005) · Proceedings 10th Symposium on Haptic Interfaces for Virtual Environment and Teleoperator Systems. HAPTICS 2002 · doi:10.1109/haptic.2002.998943 · Cited by 379
Needle Steering and Motion Planning in Soft Tissues
DiMaio, S., & Salcudean, S. E. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.846734 · Cited by 263
A survey of perceptual feedback issues in dexterous telemanipulation. II. Finger touch feedback
Shimoga, K. B. (2002) · Proceedings of IEEE Virtual Reality Annual International Symposium · doi:10.1109/vrais.1993.380769 · Cited by 245
State of the art in surgical robotics: Clinical applications and technology challenges
Cleary, K., & Nguyen, C. (2001) · Computer Aided Surgery · doi:10.1002/igs.10019 · Cited by 238
Modeling of needle insertion forces for robot-assisted percutaneous therapy
Simone, C., & Okamura, A. M. (2003) · Proceedings 2002 IEEE International Conference on Robotics and Automation (Cat. No.02CH37292) · doi:10.1109/robot.2002.1014848 · Cited by 200
Robot-assisted 3D-TRUS guided prostate brachytherapy: System integration and validation
Wei, Z., Wan, G., Gardi, L., Mills, G., Downey, D., & Fenster, A. (2004) · Medical Physics · doi:10.1118/1.1645680 · Cited by 185
Measuring in vivo animal soft tissue properties for haptic modeling in surgical simulation.
Brouwer, I., Ustin, J. S., Bentley, L., Sherman, A., Dhruv, N. T., & Tendick, F. (2001) · PubMed · Cited by 180
System for Robotically Assisted Prostate Biopsy and Therapy with Intraoperative CT Guidance
Fichtinger, G., DeWeese, T. L., Patriciu, A., Tanacs, A., Mazilu, D., Anderson, J. H., Masamune, K., Taylor, R. H., & Stoianovici, D. (2002) · Academic Radiology · doi:10.1016/s1076-6332(03)80297-0 · Cited by 158
The Inaccessible or Undrainable Abscess: How to Drain It
Maher, M. M., Gervais, D. A., Kalra, M. K., Lucey, B. C., Sahani, D. V., Arellano, R. S., Hahn, P. F., & Müeller, P. R. (2004) · Radiographics · doi:10.1148/rg.243035100 · Cited by 145
A novel ultrasound indentation system for measuring biomechanical properties of in vivo soft tissue
Han, L., Noble, J. A., & Burcher, M. (2003) · Ultrasound in Medicine & Biology · doi:10.1016/s0301-5629(02)00776-7 · Cited by 128
Dosimetric effects of needle divergence in prostate seed implant using and radioactive seeds
Nath, S. K., Chen, Z., Yue, N. J., Trumpore, S., & Peschel, R. E. (2000) · Medical Physics · doi:10.1118/1.598971 · Cited by 119
Haptic interfaces for virtual environment and teleoperator systems
Adelstein, B. D., & Colgate, J. E. (1995) · Cited by 116
Needle Steering and Model-Based Trajectory Planning
DiMaio, S., & Salcudean, S. E. (2003) · Lecture notes in computer science · doi:10.1007/978-3-540-39899-8_5 · Cited by 112
Image-guided surgery: From X-rays to Virtual Reality
Peters, T. M. (2001) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255840008907997 · Cited by 112
System for MR Image–guided Prostate Interventions: Canine Study
Susil, R. C., Krieger, A., Derbyshire, J. A., et al. (2003) · Radiology · doi:10.1148/radiol.2283020911 · Cited by 79
Sensorless planning for medical needle insertion procedures
Alterovitz, R., Goldberg, K., Pouliot, J., et al. (2004) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2003.1249671 · Cited by 77
3D modeling of the human upper limb including the biomechanics of joints, muscles and soft tissues
Maurel, W. (1999) · Infoscience (Ecole Polytechnique Fédérale de Lausanne) · doi:10.5075/epfl-thesis-1906 · Cited by 74
A robotic system for blood sampling
Zivanovic, A., & Davies, B. (2000) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/4233.826854 · Cited by 73
Schemes for the identification of tissue types and boundaries at the tool point for surgical needles
Brett, P. N., Harrison, A. J., & Thomas, T. A. (2000) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/4233.826856 · Cited by 70
A computational technique for interactive needle insertions in 3D nonlinear material
Nienhuys, H.-W., & van der Stappen, A. F. (2004) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.2004.1308127 · Cited by 57
Automatic surgical tools for penetrating flexible tissues
Brett, P. N., Fraser, C., Hennigan, M., Griffiths, M. V., & Kamel, Y. N. (1995) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.391778 · Cited by 53
Rizun, P., McBeth, P. B., Louw, D. F., & Sutherland, G. R. (2004) · Surgical Innovation · doi:10.1177/107155170401100206 · Cited by 46
Predicting respiratory motion for active canceling during percutaneous needle insertion
Riviere, C. N., Thakral, A., Iordachita, I. I., Mitroi, G., & Stoianovici, D. (2005) · 2001 Conference Proceedings of the 23rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2001.1019580 · Cited by 40
Force models for needle insertion created from measured needle puncture data.
Holton, L. (2001) · PubMed · Cited by 37
Projection-Based Needle Segmentation in 3D Ultrasound Images
Ding, M., & Fenster, A. (2003) · Lecture notes in computer science · doi:10.1007/978-3-540-39903-2_40 · Cited by 14
A haptic guidance tool for CT-directed percutaneous interventions
Hagmann, E., Rouiller, P., Helmer, P., Grange, S., & Baur, C. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1403786 · Cited by 9
Cited by
Continuum Robots for Medical Applications: A Survey
Burgner-Kahrs, J., Rucker, D. C., & Choset, H. (2015) · IEEE Transactions on Robotics · doi:10.1109/tro.2015.2489500 · Cited by 1,557
Microneedle, bio-microneedle and bio-inspired microneedle: A review
Ma, G., & Wu, C. (2017) · Journal of Controlled Release · doi:10.1016/j.jconrel.2017.02.011 · Cited by 477
Mechanics of Flexible Needles Robotically Steered through Soft Tissue
Misra, S., Reed, K. B., Schafer, B. W., Ramesh, K. T., & Okamura, A. M. (2010) · The International Journal of Robotics Research · doi:10.1177/0278364910369714 · Cited by 314
Three-Dimensional Needle Shape Reconstruction Using an Array of Fiber Bragg Grating Sensors
Roesthuis, R. J., Kemp, M., van den Dobbelsteen, J. J., & Misra, S. (2014) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2013.2269836 · Cited by 281
DLR MiroSurge: a versatile system for research in endoscopic telesurgery
Hagn, U., Konietschke, R., Tobergte, A., Nickl, M., Jörg, S., Kübler, B., Passig, G., Gröger, M., Fröhlich, F., Seibold, U., et al. (2010) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-009-0372-4 · Cited by 270
Metadata
- Authors
- Abolhassani, N., Patel, R., & Moallem, M.
- Venue
- Medical Engineering & Physics
- Publisher
- IOP Publishing
- Published
- 2006-08-28
- Volume
- 29
- Issue
- 4
- Pages
- 413-431
- ISSN
- 1350-4533 (ISSN-L)
- 1873-4030
- OpenAlex
- W2090415875
- Cited by
- 704
- References
- 43
- FWCI
- 12.19
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.