Evaluation of 2D multiband EPI imaging for high-resolution, whole-brain, task-based fMRI studies at 3T: Sensitivity and slice leakage artifacts
Open accessThe DOI matches its Crossref record; 28 of the 28 references it cites were found in a registry or an index.
Abstract
Functional magnetic resonance imaging (fMRI) studies that require high-resolution whole-brain coverage have long scan times that are primarily driven by the large number of thin slices acquired. Two-dimensional multiband echo-planar imaging (EPI) sequences accelerate the data acquisition along the slice direction and therefore represent an attractive approach to such studies by improving the temporal resolution without sacrificing spatial resolution. In this work, a 2D multiband EPI sequence was optimized for 1.5mm isotropic whole-brain acquisitions at 3T with 10 healthy volunteers imaged while performing simultaneous visual and motor tasks. The performance of the sequence was evaluated in terms of BOLD sensitivity and false-positive activation at multiband (MB) factors of 1, 2, 4, and 6, combined with in-plane GRAPPA acceleration of 2× (GRAPPA 2), and the two reconstruction approaches of Slice-GRAPPA and Split Slice-GRAPPA. Sensitivity results demonstrate significant gains in temporal signal-to-noise ratio (tSNR) and t-score statistics for MB 2, 4, and 6 compared to MB 1. The MB factor for optimal sensitivity varied depending on anatomical location and reconstruction method. When using Slice-GRAPPA reconstruction, evidence of false-positive activation due to signal leakage between simultaneously excited slices was seen in one instance, 35 instances, and 70 instances over the ten volunteers for the respective accelerations of MB 2×GRAPPA 2, MB 4×GRAPPA 2, and MB 6×GRAPPA 2. The use of Split Slice-GRAPPA reconstruction suppressed the prevalence of false positives significantly, to 1 instance, 5 instances, and 5 instances for the same respective acceleration factors. Imaging protocols using an acceleration factor of MB 2×GRAPPA 2 can be confidently used for high-resolution whole-brain imaging to improve BOLD sensitivity with very low probability for false-positive activation due to slice leakage. Imaging protocols using higher acceleration factors (MB 3 or MB 4×GRAPPA 2) can likely provide even greater gains in sensitivity but should be carefully optimized to minimize the possibility of false activations.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Ashburner, J., & Friston, K. J. (2005) · NeuroImage · doi:10.1016/j.neuroimage.2005.02.018 · Cited by 7,507 · cited 3×
“The ROIs were defined using the Anatomy Toolbox in SPM8 (Eickhoff et al., 2005).”
A new SPM toolbox for combining probabilistic cytoarchitectonic maps and functional imaging data
Eickhoff, S. B., Stephan, K. E., Mohlberg, H., Grefkes, C., Fink, G. R., Amunts, K., & Zilles, K. (2005) · NeuroImage · doi:10.1016/j.neuroimage.2004.12.034 · Cited by 4,033 · cited 2×
“The ROIs were defined using the Anatomy Toolbox in SPM8 (Eickhoff et al., 2005).”
MethodStatistical Parametric Mapping: The Analysis of Functional Brain Images
Penny, W. D., Friston, K. J., Ashburner, J., Kiebel, S. J., & Thomas, N. (2007) · UCL Discovery (University College London) · Cited by 3,672
Moeller, S., Yacoub, E., Olman, C. A., Auerbach, E., Strupp, J., Harel, N., & Uğurbil, K. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22361 · Cited by 1,586 · cited 6×
“Contents lists available at ScienceDirect NeuroImage j ourna l homepage: www.e lsev ie r .com/ locate /yn img Full Length Article Evaluation of 2Dmultiband EPI imaging for high-resolution, whole-brain, task-based fMRI studies at 3T: Sensitivity and slice leakage artifacts Nick Todd a,⁎, Steen Moeller b, Edward J. Auer…”
BackgroundMethodInfluentialSelf-citationResting-state fMRI in the human connectome project
Smith, S. M., Beckmann, C. F., Andersson, J., Auerbach, E. J., Bijsterbosch, J., Douaud, G., Duff, E., Feinberg, D. A., Griffanti, L., Harms, M. P., et al. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.05.039 · Cited by 1,502 · cited 1×
“…have used amultiband 2D EPI sequence for functionalMRI (fMRI) studies to demonstrate that the faster scan time can improve statistical power (Feinberg et al., 2010; Smith et al., 2013) and better sample signal fluctuations due to physiology (Tong and Frederick, 2014; Tong et al., 2014).”
MethodSelf-citationMultiplexed echo planar imaging for sub-second whole brain FMRI and fast diffusion imaging
Feinberg, D. A., Moeller, S., Smith, S. M., Auerbach, E., Ramanna, S., Glasser, M. F., Miller, K. L., Ugurbil, K., & Yacoub, E. (2010) · PLoS ONE · doi:10.1371/journal.pone.0015710 · Cited by 1,296 · cited 2×
“…have used amultiband 2D EPI sequence for functionalMRI (fMRI) studies to demonstrate that the faster scan time can improve statistical power (Feinberg et al., 2010; Smith et al., 2013) and better sample signal fluctuations due to physiology (Tong and Frederick, 2014; Tong et al., 2014).”
MethodSelf-citationSetsompop, K., Gagoski, B. A., Polimeni, J. R., Witzel, T., Wedeen, V. J., & Wald, L. L. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23097 · Cited by 1,285 · cited 8×
“Contents lists available at ScienceDirect NeuroImage j ourna l homepage: www.e lsev ie r .com/ locate /yn img Full Length Article Evaluation of 2Dmultiband EPI imaging for high-resolution, whole-brain, task-based fMRI studies at 3T: Sensitivity and slice leakage artifacts Nick Todd a,⁎, Steen Moeller b, Edward J. Auer…”
BackgroundMethodInfluentialClassical and Bayesian Inference in Neuroimaging: Applications
Friston, K. J., Glaser, D. E., Henson, R. N. A., Kiebel, S., Phillips, C., & Ashburner, J. (2002) · NeuroImage · doi:10.1006/nimg.2002.1091 · Cited by 719 · cited 9×
“To determine if a falsepositive activation occurred at an aliased location, the voxel with the largest t-score value for all activation clusters defined by SPM were considered as “seed” voxels.”
BackgroundMethodInfluentialBreuer, F. A., Blaimer, M., Heidemann, R. M., Mueller, M. F., Griswold, M. A., & Jakob, P. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20401 · Cited by 633 · cited 3×
“The next major advance was the development of the blipped-CAIPI approach that allowed for controlled aliasing of the simultaneously excited slices to be used with the EPI acquisition (Breuer et al., 2005; Feinberg and Setsompop, 2013; Moeller et al., 2010; Setsompop et al., 2012).”
BackgroundMethodEvaluation of slice accelerations using multiband echo planar imaging at 3T
Xu, J., Moeller, S., Auerbach, E. J., Strupp, J., Smith, S. M., Feinberg, D. A., Yacoub, E., & Uğurbil, K. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.07.055 · Cited by 514 · cited 5×
“For the case ofmultiband2DEPI, artifacts tend tomanifest as signal leaking from one slice into another simultaneously excited slice (Xu et al., 2013).”
BackgroundMethodInfluentialSelf-citationUltra-fast MRI of the human brain with simultaneous multi-slice imaging
Feinberg, D. A., & Setsompop, K. (2013) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2013.02.002 · Cited by 433 · cited 5×
“Contents lists available at ScienceDirect NeuroImage j ourna l homepage: www.e lsev ie r .com/ locate /yn img Full Length Article Evaluation of 2Dmultiband EPI imaging for high-resolution, whole-brain, task-based fMRI studies at 3T: Sensitivity and slice leakage artifacts Nick Todd a,⁎, Steen Moeller b, Edward J. Auer…”
BackgroundMethodInfluentialDeichmann, R., Schwarzbauer, C., & Turner, R. (2004) · NeuroImage · doi:10.1016/j.neuroimage.2003.09.062 · Cited by 391 · cited 1×
“…times of 10.0 ms, and 12.46 ms), and a high-resolution T1-weighted anatomical image was acquired (1.0 mm isotropic resolution, parameter details in (Deichmann et al., 2004)). fMRI data processing The data sets from the four different multiband factors and two different reconstruction types were…”
BackgroundInfluentialThree dimensional echo-planar imaging at 7 Tesla
Poser, B. A., Koopmans, P. J., Witzel, T., Wald, L. L., & Barth, M. (2010) · NeuroImage · doi:10.1016/j.neuroimage.2010.01.108 · Cited by 284 · cited 1×
“…temporal resolution, and ongoing work is investigating the merits of multi-shot 3D gradient echo EPI, multi-slab echo-volumar imaging, and 3D balanced steadystate free precession approaches (Lutti et al., 2013; Poser et al., 2010; Posse et al., 2012; Swisher et al., 2012; Tijssen et al., 2011).”
BackgroundInterslice leakage artifact reduction technique for simultaneous multislice acquisitions
Cauley, S. F., Polimeni, J. R., Bhat, H., Wald, L. L., & Setsompop, K. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24898 · Cited by 255 · cited 19×
“GRAPPA reconstruction, and 441, 1,428, and 2,068 for the MB 2, MB 4, and MB 6 cases with Split Slice-GRAPPA reconstruction.”
BackgroundMethodInfluentialThe rapid development of high speed, resolution and precision in fMRI
Feinberg, D. A., & Yacoub, E. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2012.01.049 · Cited by 119 · cited 2×
“Many studies have considered the question of how temporal resolution affects sensitivity to the blood oxygen leveldependent (BOLD) signal (Feinberg and Yacoub, 2012; Lin et al., 2012; Neggers et al., 2008; Posse et al., 2012), and recently several investigators have used amultiband 2D EPI sequence…”
MethodInfluentialSelf-citationNeggers, S. F. W., Hermans, E. J., & Ramsey, N. F. (2008) · NMR in Biomedicine · doi:10.1002/nbm.1235 · Cited by 118 · cited 2×
“Many studies have considered the question of how temporal resolution affects sensitivity to the blood oxygen leveldependent (BOLD) signal (Feinberg and Yacoub, 2012; Lin et al., 2012; Neggers et al., 2008; Posse et al., 2012), and recently several investigators have used amultiband 2D EPI sequence for functionalMRI (f…”
MethodPosse, S., Ackley, E., Mutihac, R., Rick, J., Shane, M., Murray-Krezan, C., Zaitsev, M., & Speck, O. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2012.02.059 · Cited by 89 · cited 4×
“Published by Elsevier Inc a b s t r a c t a r t i c l e i n f o”
BackgroundMethodInfluentialSimultaneous echo refocusing in EPI
Feinberg, D. A., Reese, T. G., & Wedeen, V. J. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10227 · Cited by 89 · cited 2×
“Before the advent of multiband imaging, several approaches existed for accelerating 2D acquisitions in the slice direction (Bishop and Plewes, 1991; Crooks et al., 1982; Feinberg et al., 2002; Loenneker et al., 1996) but none became widely adopted.”
BackgroundLutti, A., Thomas, D. L., Hutton, C., & Weiskopf, N. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24398 · Cited by 87 · cited 2×
“…temporal resolution, and ongoing work is investigating the merits of multi-shot 3D gradient echo EPI, multi-slab echo-volumar imaging, and 3D balanced steadystate free precession approaches (Lutti et al., 2013; Poser et al., 2010; Posse et al., 2012; Swisher et al., 2012; Tijssen et al., 2011).”
BackgroundSelf-citationTracking cerebral blood flow in BOLD fMRI using recursively generated regressors
Tong, Y., & Frederick, B. B. (2014) · Human Brain Mapping · doi:10.1002/hbm.22564 · Cited by 69 · cited 2×
“, 2013) and better sample signal fluctuations due to physiology (Tong and Frederick, 2014; Tong et al., 2014).”
BackgroundMethodSENSE and simultaneous multislice imaging
Zahneisen, B., Ernst, T., & Poser, B. A. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25519 · Cited by 58 · cited 1×
“These methods include SENSE/GRAPPA, 3D-GRAPPA, and SENSE (Breuer et al., 2005; Setsompop et al., 2012; Zahneisen et al., 2014; Zhu et al., 2012).”
MethodTong, Y., Hocke, L. M., & Frederick, B. B. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25041 · Cited by 41 · cited 2×
“, 2013) and better sample signal fluctuations due to physiology (Tong and Frederick, 2014; Tong et al., 2014).”
BackgroundMethodLoenneker, T., Hennel, F., & Hennig, J. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910350613 · Cited by 39 · cited 2×
“Before the advent of multiband imaging, several approaches existed for accelerating 2D acquisitions in the slice direction (Bishop and Plewes, 1991; Crooks et al., 1982; Feinberg et al., 2002; Loenneker et al., 1996) but none became widely adopted.”
BackgroundUltrafast inverse imaging techniques for fMRI
Lin, F., Tsai, K. W., Chu, Y.-H., Witzel, T., Nummenmaa, A. R., Raij, T., Ahveninen, J., Kuo, W., & Belliveau, J. W. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2012.01.072 · Cited by 39 · cited 2×
“…considered the question of how temporal resolution affects sensitivity to the blood oxygen leveldependent (BOLD) signal (Feinberg and Yacoub, 2012; Lin et al., 2012; Neggers et al., 2008; Posse et al., 2012), and recently several investigators have used amultiband 2D EPI sequence for…”
MethodInfluentialSchulz, J., Siegert, T., Bazin, P., Maclaren, J., Herbst, M., Zaitsev, M., & Turner, R. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.08.006 · Cited by 37 · cited 2×
“that are inwhitematter could be rejected as false (Schulz et al., 2014).”
BackgroundSwisher, J. D., Sexton, J. A., Gatenby, J. C., Gore, J. C., & Tong, F. (2012) · PLoS ONE · doi:10.1371/journal.pone.0034626 · Cited by 23 · cited 2×
“…temporal resolution, and ongoing work is investigating the merits of multi-shot 3D gradient echo EPI, multi-slab echo-volumar imaging, and 3D balanced steadystate free precession approaches (Lutti et al., 2013; Poser et al., 2010; Posse et al., 2012; Swisher et al., 2012; Tijssen et al., 2011).”
BackgroundTijssen, R. H. N., Okell, T. W., & Miller, K. L. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2011.05.070 · Cited by 14 · cited 2×
“…temporal resolution, and ongoing work is investigating the merits of multi-shot 3D gradient echo EPI, multi-slab echo-volumar imaging, and 3D balanced steadystate free precession approaches (Lutti et al., 2013; Poser et al., 2010; Posse et al., 2012; Swisher et al., 2012; Tijssen et al., 2011).”
BackgroundLarkman (2001) · doi:10.1002/1522-2586(200102)13:2<313::aid-jmri1045>3.0.co;2-w · cited 2×
“…ourna l homepage: www.e lsev ie r .com/ locate /yn img Full Length Article Evaluation of 2Dmultiband EPI imaging for high-resolution, whole-brain, task-based fMRI studies at 3T: Sensitivity and slice leakage artifacts Nick Todd a,⁎, Steen Moeller b, Edward J. Auerbach b, Essa Yacoub b,…”
Method
Cited by
Advances in resting state fMRI acquisitions for functional connectomics
Raimondo, L., Oliveira, Ĺ. A. F., Heij, J., Priovoulos, N., Kundu, P., Leoni, R. F., & van der Zwaag, W. (2021) · NeuroImage · doi:10.1016/j.neuroimage.2021.118503 · Cited by 180
Wisse, L. E. M., Daugherty, A. M., Olsen, R. K., Berron, D., Carr, V. A., Stark, C. E. L., Amaral, R. S. C., Amunts, K., Augustinack, J. C., Bender, A. R., et al. (2017) · Hippocampus · doi:10.1002/hipo.22671 · Cited by 167
“With new technological advancements at 3T and increasing availability of high-resolution 7T imaging (Feinberg and Yacoub, 2012; Todd et al., 2016), it will become increasingly feasible to functionally separate these subregions.”
BackgroundAccurate autocorrelation modeling substantially improves fMRI reliability
Olszowy, W., Aston, J., Rua, C., & Williams, G. B. (2019) · Nature Communications · doi:10.1038/s41467-019-09230-w · Cited by 164
“SPM has an alternative approach: FAST, but we know of only two studies which have used it (Todd et al., 2016; Bollmann et al., 2018).”
MethodCortical tau deposition follows patterns of entorhinal functional connectivity in aging
Adams, J. N., Maass, A., Harrison, T. M., Baker, S. L., & Jagust, W. J. (2019) · eLife · doi:10.7554/elife.49132 · Cited by 160
Metadata
- Authors
- Todd, N., Moeller, S., Auerbach, E. J., Yacoub, E., Flandin, G., & Weiskopf, N.
- Venue
- NeuroImage
- Publisher
- Elsevier BV
- Published
- 2016-01
- Volume
- 124
- Issue
- Pt A
- Pages
- 32-42
- ISSN
- 1053-8119 (ISSN-L)
- OpenAlex
- W1810655782
- Cited by
- 173
- References
- 28
- FWCI
- 11.09
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.