A realtime and continuous assessment of cortisol in ISF using electrochemical impedance spectroscopy
Open accessAbstract
This study describes the functioning of a novel sensor to measure cortisol concentration in the interstitial fluid (ISF) of a human subject. ISF is extracted by means of vacuum pressure from micropores created on the stratum corneum layer of the skin. The pores are produced by focusing a near infrared laser on a layer of black dye material attached to the skin. The pores are viable for approximately three days after skin poration. Cortisol measurements are based on electrochemical impedance (EIS) technique. Gold microelectrode arrays functionalized with Dithiobis (succinimidyl propionate) self-assembled monolayer (SAM) have been used to fabricate an ultrasensitive, disposable, electrochemical cortisol immunosensor. The biosensor was successfully used for in-vitro measurement of cortisol in ISF. Tests in a laboratory setup show that the sensor exhibits a linear response to cortisol concentrations in the range 1 pm to 100 nM. A small pilot clinical study showed that in-vitro immunosensor readings, when compared with commercial evaluation using enzyme-linked immunoassay (ELISA) method, correlated well with cortisol levels in saliva and ISF. Further, circadian rhythm could be established between the subject's ISF and the saliva samples collected over 24 hours time-period. Cortisol levels in ISF were found reliably higher than in saliva. This Research establishes the feasibility of using impedance based biosensor architecture for a disposable, wearable cortisol detector. The projected commercial in-vivo real-time cortisol sensor device, besides being minimally invasive, will allow continuous ISF harvesting and cortisol monitoring over 24 hours even when the subject is asleep. Forthcoming, this sensor could be interfaced to a wireless health monitoring system that could transfer sensor data over existing wide-area networks such as the internet and a cellular phone network to enable real-time remote monitoring of subjects.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
ISF composition leads to reliable correlations between blood and ISF concentrations of many analytes, including electrolytes, metabolites, and proteins.
“This function, and the associated ISF composition, lead to reliable correlations between the blood and ISF concentrations of many analytes, including electrolytes (e.g., sodium, phosphate, magnesium, potassium or calcium), metabolites (e.g., glucose, alcohol, lactate or cortisol) and proteins65-69.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [69]
Cortisol has been measured in human blood, serum, urine, saliva, hair, interstitial fluid, and most recently eccrine sweat.
“Cortisol has been measured in human blood,106 serum,107 urine,107 saliva,107,108 hair,109 interstitial fluid,110 and most recently eccrine sweat.111”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [110]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Nighttime salivary cortisol: a useful test for the diagnosis of Cushing's syndrome.
Papanicolaou, D. A., Mullen, N., Kyrou, I., & Nieman, L. K. (2002) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2002-020534 · Cited by 241 · see citing works
Noninvasive Blood Glucose Monitoring
Klonoff, D. C. (1997) · Diabetes Care · doi:10.2337/diacare.20.3.433 · Cited by 209 · see citing works
Glucose measurement in patients with diabetes mellitus with dermal interstitial fluid
Bantle, J. P., & Thomas, W. (1997) · Journal of Laboratory and Clinical Medicine · doi:10.1016/s0022-2143(97)90044-5 · Cited by 126 · see citing works
Arya, S. K., Chornokur, G., Venugopal, M., & Bhansali, S. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.03.016 · Cited by 115 · see citing works
Self-citationLiu, D., Perdue, R. K., Sun, L., et al. (2004) · Langmuir · doi:10.1021/la049605p · Cited by 82 · see citing works
Krieger, D. T. (1975) · Journal of Steroid Biochemistry · doi:10.1016/0022-4731(75)90068-0 · Cited by 82 · see citing works
Antibody functionalized interdigitated μ-electrode (IDμE) based impedimetric cortisol biosensor
Arya, S. K., Chornokur, G., Venugopal, M., & Bhansali, S. (2010) · The Analyst · doi:10.1039/c0an00242a · Cited by 80 · see citing works
Self-citationVenugopal, M., Feuvrel, K. E., Mongin, D., Bambot, S., Faupel, M., Panangadan, A., Talukder, A., & Pidva, R. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2007.912544 · Cited by 76 · see citing works
Self-citationData‐Mining Analysis Suggests an Epigenetic Pathogenesis for Type 2 Diabetes
Wren, J. D., & Garner, H. R. (2005) · BioMed Research International · doi:10.1155/jbb.2005.104 · Cited by 76 · see citing works
Increased expression of 11β‐hydroxysteroid dehydrogenase type 1 in type 2 diabetic myotubes
Abdallah, B. M., Beck‐Nielsen, H., & Gaster, M. (2005) · European Journal of Clinical Investigation · doi:10.1111/j.1365-2362.2005.01552.x · Cited by 69 · see citing works
Gebhart, S. S. P., Faupel, M. L., Fowler, R. T., Kapsner, C., Lincoln, D., McGee, V., Pasqua, J. J., Steed, L., Wangsness, M., Xu, F., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503321827812 · Cited by 69 · see citing works
An immunoelectrochemical sensor for salivary cortisol measurement
Sun, K., Ramgir, N. S., & Bhansali, S. (2008) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2008.03.018 · Cited by 66 · see citing works
Self-citationSubcutaneous Glucose Concentration in Humans: Real estimation and continuous monitoring
Sternberg, F., Meyerhoff, C., Mennel, F. J., Bischof, F., & Pfeiffer, E. F. (1995) · Diabetes Care · doi:10.2337/diacare.18.9.1266 · Cited by 65 · see citing works
Gatti, R., Cappellin, E., Zecchin, B., Antonelli, G., Spinella, P., Mantero, F., & Palo, E. F. D. (2005) · Journal of Chromatography B · doi:10.1016/j.jchromb.2005.06.037 · Cited by 61 · see citing works
Stout, P. J., Racchini, J. R., & Hilgers, M. E. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.635 · Cited by 59 · see citing works
Real-time measurements of corticosteroids in conscious animals using an antibody-based electrode
Cook, C. J. (1997) · Nature Biotechnology · doi:10.1038/nbt0597-467 · Cited by 54 · see citing works
A fluorimetric assay for cortisol
Appel, D., Schmid, R. D., Drăgan, C.-A., Bureik, M., & Urlacher, V. B. (2005) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-005-0022-9 · Cited by 51 · see citing works
Stress, symptoms of depression and anxiety, and cortisol patterns in working parents
Kurina, L. M., Schneider, B., & Waite, L. J. (2004) · Stress and Health · doi:10.1002/smi.998 · Cited by 48 · see citing works
Fischer, U., Rebrin, K., von Woedtke, T., & Abel, P. (1994) · Hormone and Metabolic Research · doi:10.1055/s-2007-1001747 · Cited by 37 · see citing works
Immunosensors: The next generation
Turner, A. P. F. (1997) · Nature Biotechnology · doi:10.1038/nbt0597-421 · Cited by 30 · see citing works
A Golden Age of Clinical Chemistry: 1948–1960
Rosenfeld, L. (2000) · Clinical Chemistry · doi:10.1093/clinchem/46.10.1705 · Cited by 29 · see citing works
Analysis of cortisol with a flow displacement immunoassay
Kaptein, W. A., Zwaagstra, J. J., Venema, K., Ruiters, M. H. J., & Korf, J. (1997) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(97)00271-2 · Cited by 27 · see citing works
Detection of FITC-cortisol via Modulated Supraparticle Lighthouses
Petkus, M. M., McLauchlin, M. L., Vuppu, A. K., Rios, L., Garcia, A. A., & Hayes, M. A. (2006) · Analytical Chemistry · doi:10.1021/ac0512204 · Cited by 25 · see citing works
Continuous Glucose Monitoring: Long-Term Implantable Sensor Approach
Daniloff, G. Y. (1999) · Diabetes Technology & Therapeutics · doi:10.1089/152091599317170 · Cited by 19 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- [9]Bantle JP Thomas W Glucose measurement in patients with diabetes mellitus with dermal interstitial fluid J. Lab. Clin. Med 1997 130 436 441 9358083
- [23]Sun K Ramgir N Venugopal M Bhansali S Estimation of Cortisol level in interstitial fluid and saliva using electrochemical array 2008 22nd international conference, EUROSENSORS Dresden, Germany
Cited by
Sorted by citations
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373 · see citing works
Wearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315 · see citing works
Wearable sensors for monitoring the physiological and biochemical profile of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0150-9 · Cited by 537 · see citing works
Stress Biomarkers in Biological Fluids and Their Point-of-Use Detection
Steckl, A. J., & Ray, P. (2018) · ACS Sensors · doi:10.1021/acssensors.8b00726 · Cited by 319 · see citing works
Recent advances in cortisol sensing technologies for point-of-care application
Kaushik, A., Vasudev, A., Arya, S. K., Pasha, S. K., & Bhansali, S. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.09.060 · Cited by 316 · see citing works
Self-citationJia, M., Chew, W. M., Feinstein, Y., Skeath, P., & Sternberg, E. M. (2016) · The Analyst · doi:10.1039/c5an02387d · Cited by 113 · see citing works
Portable and wearable real-time stress monitoring: A critical review
Parlak, O. (2021) · Sensors and Actuators Reports · doi:10.1016/j.snr.2021.100036 · Cited by 60 · see citing works
- Authors
- Venugopal, M., Arya, S. K., Chornokur, G., & Bhansali, S.
- Affiliations
- Guided Therapeutics (United States) (US)
- University of South Florida (US)
- Corresponding author
- Venugopal, Manju
- Venue
- Sensors and Actuators A: Physical
- Publisher
- Elsevier BV
- Published
- 2011-12
- Volume
- 172
- Issue
- 1
- Pages
- 154-160
- ISSN
- 0924-4247 (ISSN-L)
- PMC
- PMC3234992
- OpenAlex
- W2021895870
- Cited by
- 100
- References
- 24
- Citations by year
- 2026: 6 · 2025: 11 · 2024: 12 · 2023: 10 · 2022: 6 · 2021: 7 · 2020: 9 · 2019: 9 · 2018: 7 · 2017: 2 · 2016: 6 · 2014: 4 · 2013: 8 · 2012: 3
- FWCI
- 3.35
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Sensors and Actuators A Physical (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- NIH (1R43MH085474-01)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Advanced biosensing and bioanalysis techniques
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Keywords
- cortisol measurement, interstitial fluid, electrochemical impedance spectroscopy, disposable biosensor, gold microelectrode array, near-infrared laser, circadian rhythm, salivary cortisol, wearable health monitoring, remote patient monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review