Fifteen years experience with finger arterial pressure monitoring: assessment of the technology
Abstract
We review the Finapres technology, embodied in several TNO-prototypes and in the Ohmeda 2300 and 2300e Finapres NIBP. Finapres is an acronym for FINger Arterial PRESsure, the device delivers a continuous finger arterial pressure waveform. Many papers report on the accuracy of the device in comparison with intra-arterial or with noninvasive but intermittent blood pressure measurements. We compiled the results of 43 such papers and found systolic, diastolic and mean accuracies, in this order, ranging from -48 to 30 mmHg, from -20 to 18 mmHg, and from -13 to 25 mmHg. Weighted for the number of subjects included pooled accuracies were -0.8 (SD 11.9), -1.6 (8.3) and -1.6 (7.6) mmHg respectively. Subdividing the pooled group according to criteria such as reference blood pressure, place of application, and prototype or commercial device we found no significant differences in mean differences or SD. Measurement at the finger allows uninterrupted recordings of long duration. The transmission of the pressure pulse along the arm arteries, however, causes distortion of the pulse waveform and depression of the mean blood pressure level. These effects can be reduced by appropriate filtering, and upper arm 'return-to-flow' calibration to bring accuracy and precision within AAMI limits. For the assessment of beat-to-beat changes in blood pressure and assessment of blood pressure variability Finapres proved a reliable alternative for invasive measurements when mean and diastolic pressures are concerned. Differences in systolic pressure are larger and reach statistical significance but are not of clinical relevance. Finger arteries are affected by contraction and dilatation in relation to psychological and physical (heat, cold, blood loss, orthostasis) stress. Effects of these phenomena are reduced by the built-in Physiocal algorithm. However, full smooth muscle contraction should be avoided in the awake patient by comforting the patient, and covering the hand. Arterial state can be monitored by observing the behaviour of the Physiocal algorithm. We conclude that Finapres accuracy and precision usually suffice for reliable tracking of changes in blood pressure. Diagnostic accuracy may be achieved with future application of corrective measures.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Computation of aortic flow from pressure in humans using a nonlinear, three-element model
Wesseling, K. H., Jansen, J. R., Settels, J. J., & Schreuder, J. J. (1993) · Journal of Applied Physiology · doi:10.1152/jappl.1993.74.5.2566 · Cited by 1,137
Parati, G., Casadei, R., Groppelli, A., Di Rienzo, M., & Mancia, G. (1989) · Hypertension · doi:10.1161/01.hyp.13.6.647 · Cited by 899
Tilt table testing for assessing syncope
Benditt, D. G., Ferguson, D. W., Grubb, B. P., Kapoor, W. N., Kugler, J., Lerman, B. B., Maloney, J. D., Ravielle, A., Ross, B., Sutton, R., et al. (1996) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(96)00236-7 · Cited by 568
Photoelectric measurement of blood pressure, volume and flow in the finger
Peňáz, J. (1973) · Medical Entomology and Zoology · Cited by 529
Sympathetic Activation and Loss of Reflex Sympathetic Control in Mild Congestive Heart Failure
Grassi, G., Seravalle, G., Cattaneo, B. M., Lanfranchi, A., Vailati, S., Giannattasio, C., Del Bo, A., Sala, C., Bolla, G. B., Pozzi, M., et al. (1995) · Circulation · doi:10.1161/01.cir.92.11.3206 · Cited by 432
Reconstruction of Brachial Artery Pressure From Noninvasive Finger Pressure Measurements
Bos, W. J. W., van Goudoever, J., van Montfrans, G. A., van den Meiracker, A. H., & Wesseling, K. H. (1996) · Circulation · doi:10.1161/01.cir.94.8.1870 · Cited by 365
Imholz, B. P. M., Settels, J. J., van der Meiracker, A. H., Wesseling, K. H., & Wieling, W. (1990) · Cardiovascular Research · doi:10.1093/cvr/24.3.214 · Cited by 344
Self-citationIMHOLZ, B. P. M., MONTFRANS, G. A. V., SETTELS, J. J., HOEVEN, G. M. A. V. D., KAREMAKER, J. M., & WIELING, W. (1988) · Cardiovascular Research · doi:10.1093/cvr/22.6.390 · Cited by 275
Self-citationFeasibility of ambulatory, continuous 24-hour finger arterial pressure recording.
Imholz, B. P., Langewouters, G. J., van Montfrans, G. A., Parati, G., van Goudoever, J., Wesseling, K. H., Wieling, W., & Mancia, G. (1993) · Hypertension · doi:10.1161/01.hyp.21.1.65 · Cited by 259
Self-citationPhysiocal, calibrating finger vascular physiology for finapres
Wesseling, K. H., Wit, B. D., van der Hoeven, G. M. A., Goudoever, J. V., & Settels, J. J. (1995) · TU/e Research Portal · Cited by 229
Cited by
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
European Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206
Guidelines on management (diagnosis and treatment) of syncope
Brignole, M. (2001) · European Heart Journal · doi:10.1053/euhj.2001.2739 · Cited by 951
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
Metadata
- Authors
- Imholz, B.
- Venue
- Cardiovascular Research
- Publisher
- Oxford University Press (OUP)
- Published
- 1998-06
- Volume
- 38
- Issue
- 3
- Pages
- 605-616
- ISSN
- 0008-6363 (ISSN-L)
- OpenAlex
- W2147476476
- Cited by
- 834
- References
- 61
- FWCI
- 19.54
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Fifteen years experience with finger arterial pressure monitoring: assessment of the technology. https://app.cochat.ai/papers/10.1016/s0008-6363(98)00067-4
- The paper
- Imholz, B. (1998). Fifteen years experience with finger arterial pressure monitoring: assessment of the technology. Cardiovascular Research. https://doi.org/10.1016/s0008-6363(98)00067-4
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.