The effect of microgeometry, implant thickness and polyurethane chemistry on the foreign body response to subcutaneous implants
The DOI matches its Crossref record; 25 of the 28 references it cites were found in a registry or an index.
Abstract
We addressed the effect of implant thickness, implant porosity, and polyurethane (PU) chemistry on angiogenesis and on the foreign body response in rats. The following materials were implanted subcutaneously for 7 weeks then excised for histologic analysis: a solid PU; a solid polyurethane with silicone and polyethylene oxide (PU-S-PEO); porous expanded polytetrafluoroethylene (ePTFE); and porous polyvinyl alcohol sponge (PVA). Two thicknesses of PU-S-PEO were compared: 300 microns (thin) and 2000 microns (thick). Foreign body capsule (FBC) thickness was much less in PU-S-PEO implants than in PU implants. In addition, FBC were thinner in thin implants than in thick implants. FBC was much more dense in solid implants than in porous implants. As compared with solid implants, porous implants (PVA and ePTFE) led to a marked increase in the number of microvessels that developed adjacent to the implant, as observed both with hematoxylin/eosin staining and with an immunohistochemical anti-endothelial stain. We conclude that the polyethylene oxide and silicone moieties in PU reduce the thickness of the subsequent FBC. In addition, thin implants lead to a thin FBC. Porous implants (PVA and ePTFE) cause more angiogenesis than solid implants. These results may have implications for the measurement of blood-derived analytes by biosensors.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“This suggests that fibers with smaller diameters provided a thicker FBC formation.[ 124 ]”
From Unraveling and Harnessing the Immune Response at the Cell–Biomaterial Interface for Tissue Engineering Purposes (2024) · cites this paper as [124]
In the citing paper
“In macroscale, a similar response was observed for cylindrical polyurethane substrates provoking a larger FBR and thicker FBC in large diameter cylinders of 2 mm compared to 0.3 mm implants.”
From Unraveling and Harnessing the Immune Response at the Cell–Biomaterial Interface for Tissue Engineering Purposes (2024) · cites this paper as [124]
In the citing paper
“They found that increasing the size of implanted materials resulted in larger foreign body reactions, and the formation of a thicker layer of fibrosis around the implant23.”
From Size- and shape-dependent foreign body immune response to materials implanted in rodents and non-human primates (2015) · cites this paper as [23]
In the citing paper
“the physical properties of the implant (e.g., size) impact the FBR, and the large size of the stainless steel cage will influence the overall immune response to the materials.115”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [115]
In the citing paper
“the capsule thickness was ~20% thinner surrounding the smaller implant.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [115]
1 more claim about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Protein adsorption to poly(ethylene oxide) surfaces
Gombotz, W. R., Guanghui, W., Horbett, T. A., & Hoffman, A. S. (1991) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820251211 · Cited by 634
Neovascularization of synthetic membranes directed by membrane microarchitecture
Brauker, J. H., Carr‐Brendel, V. E., Martinson, L. A., Crudele, J., Johnston, W. D., & Johnson, R. C. (1995) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820291208 · Cited by 517 · cited 2×
“We were interested in counting only those vessels that were close to the implant membrane (‘‘close vascular structures’’ [11], so only those vessels that were within the rectangle (within 40 mm of the implant surface) were counted.”
BackgroundMethodPolyurethane Elastomer Biostability
Stokes, K., McVenes, R., & Anderson, J. M. (1995) · Journal of Biomaterials Applications · doi:10.1177/088532829500900402 · Cited by 366 · cited 1×
“Both PUs (PU-S-PEO and PU) used in our study were made with polycarbonate which has been shown to favor biocompatibility in that it limits chronic biodegradation and stress cracking [17,18].”
MethodEngineering the tissue which encapsulates subcutaneous implants. I. Diffusion properties
Sharkawy, A. A., Klitzman, B., Truskey, G. A., & Reichert, W. M. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19971205)37:3<401::aid-jbm11>3.0.co;2-e · Cited by 238
den Braber, E. T., de Ruijter, J. E., Ginsel, L. A., von Recum, A. F., & Jansen, J. A. (1996) · Biomaterials · doi:10.1016/0142-9612(96)00032-4 · Cited by 208
den Braber, E. T., de Ruijter, J. E., Smits, H. T. J., Ginsel, L. A., von Recum, A. F., & Jansen, J. A. (1996) · Biomaterials · doi:10.1016/0142-9612(96)85910-2 · Cited by 198 · cited 1×
“, has shown that specific geographic features in substrate microgeometry (for example, channel width) play an important role in the way that fibroblasts orient and attach to substrates [26,27].”
BackgroundQuantitative in-vivo studies on angiogenesis in a rat sponge model.
Andrade, S. P., Fan, T., & Lewis, G. P. (1987) · PubMed · Cited by 146
Sharkawy, A. A., Klitzman, B., Truskey, G. A., & Reichert, W. M. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19980615)40:4<586::aid-jbm10>3.0.co;2-e · Cited by 145
A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.
Updike, S. J., Shults, M. C., Gilligan, B. J., & Rhodes, R. K. (2000) · Diabetes Care · doi:10.2337/diacare.23.2.208 · Cited by 135
Tissue Response to Suture Materials Implanted Subcutaneously in a Rabbit Model
Setzen, G., & Williams, E. F. (1997) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-199712000-00023 · Cited by 117 · cited 1×
“Setzen and Williams examined the tissue response to sutures implanted subcutaneously in rabbits (diameter of approximately 200 mm) and found FBC thicknesses of 85–125 mm [22].”
Background
Research works printed but not linked: 1
- Sharkawy, 1998, Engineering the tissue which encapsulates subcutaneous implants. III. Effective tissue response times, J Biomed Mater Res, 40, 598, 10.1002/(sici)1097-4636(19980615)40:4<598::aid-jbm11>3.0.co;2-c · DOI · Paper page
Cited by
Zwitterionic hydrogels implanted in mice resist the foreign-body reaction
Zhang, L., Cao, Z., Bai, T., Carr, L., Ella-Menye, J.-R., Irvin, C., Ratner, B. D., & Jiang, S. (2013) · Nature Biotechnology · doi:10.1038/nbt.2580 · Cited by 1,035
Veiseh, O., Doloff, J. C., Ma, M., Vegas, A. J., Tam, H. H., Bader, A. R., Li, J., Langan, E., Wyckoff, J., Loo, W. S., et al. (2015) · Nature Materials · doi:10.1038/nmat4290 · Cited by 897
Macrophages, Foreign Body Giant Cells and Their Response to Implantable Biomaterials
Sheikh, Z., Brooks, P., Barzilay, O., Fine, N., & Glogauer, M. (2015) · Materials · doi:10.3390/ma8095269 · Cited by 701
“Also, fibrous capsule formation on a porous surfaces are often found to be thinner than that on dense solid implant surfaces [160,182].”
BackgroundThe pathology of the foreign body reaction against biomaterials
Klopfleisch, R., & Jung, F. (2017) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.35958 · Cited by 524
“Polyurethane including silicone and polyethylene oxide moieties showed a decreased encapsulation when compared to pure polyurethane material.(110) Furthermore, gelatin-based hydrogels or hydrogels containing poly(carboxybetaine methacrylate) (PCBMA) induce only minimal fibrous reactions when compared to for instance p…”
Sussman, E. M., Halpin, M. C., Muster, J., Moon, R. T., & Ratner, B. D. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-013-0933-0 · Cited by 416
Metadata
- Authors
- Ward, W. K., Slobodzian, E. P., Tiekotter, K. L., & Wood, M. D.
- Affiliations
- Legacy Health (US)
- Corresponding author
- Ward, W.Kenneth
- ORCID
- Wood, Michael D
- Venue
- Biomaterials
- Publisher
- Elsevier BV
- Published
- 2002-11
- Volume
- 23
- Issue
- 21
- Pages
- 4185-4192
- ISSN
- 0142-9612 (ISSN-L)
- OpenAlex
- W2051278851
- Cited by
- 176
- References
- 28
- Citations by year
- 2026: 2 · 2025: 10 · 2024: 8 · 2023: 11 · 2022: 9 · 2021: 15 · 2020: 11 · 2019: 7 · 2018: 7 · 2017: 7 · 2016: 7 · 2015: 10 · 2014: 4 · 2013: 10 · 2012: 7
- FWCI
- 4.49
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- Locations
- Biomaterials (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5r01dk055297-03)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01 DK55297)
- Topics
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Physical Sciences › Materials Science › Surfaces, Coatings and Films › Polymer Surface Interaction Studies
- Physical Sciences › Materials Science › Biomaterials › Hydrogels: synthesis, properties, applications
- Keywords
- foreign body reaction, subcutaneous implants, polyurethane, implant thickness, angiogenesis, poly(ethylene oxide), silicone, porous implants, expanded polytetrafluoroethylene, polyvinyl alcohol sponge, microgeometry, biosensors
- MeSH terms
- Animals, Biocompatible Materials, Foreign-Body Reaction, Male, Microscopy, Electron, Scanning, Polyethylene Glycols, Polytetrafluoroethylene, Polyurethanes, Polyvinyl Alcohol, Porosity, Rats, Sprague-Dawley, Neovascularization, Physiologic, Prostheses and Implants, Subcutaneous Tissue, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The effect of microgeometry, implant thickness and polyurethane chemistry on the foreign body response to subcutaneous implants. https://app.cochat.ai/papers/10.1016/s0142-9612(02)00160-6
- The paper
- Ward, W. K., Slobodzian, E. P., Tiekotter, K. L., & Wood, M. D. (2002). The effect of microgeometry, implant thickness and polyurethane chemistry on the foreign body response to subcutaneous implants. Biomaterials. https://doi.org/10.1016/s0142-9612(02)00160-6
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.