A lightweight magnetically shielded room with active shielding
Open accessAbstract
Magnetically shielded rooms (MSRs) use multiple layers of materials such as MuMetal to screen external magnetic fields that would otherwise interfere with high precision magnetic field measurements such as magnetoencephalography (MEG). Optically pumped magnetometers (OPMs) have enabled the development of wearable MEG systems which have the potential to provide a motion tolerant functional brain imaging system with high spatiotemporal resolution. Despite significant promise, OPMs impose stringent magnetic shielding requirements, operating around a zero magnetic field resonance within a dynamic range of ± 5 nT. MSRs developed for OPM-MEG must therefore effectively shield external sources and provide a low remnant magnetic field inside the enclosure. Existing MSRs optimised for OPM-MEG are expensive, heavy, and difficult to site. Electromagnetic coils are used to further cancel the remnant field inside the MSR enabling participant movements during OPM-MEG, but present coil systems are challenging to engineer and occupy space in the MSR limiting participant movements and negatively impacting patient experience. Here we present a lightweight MSR design (30% reduction in weight and 40–60% reduction in external dimensions compared to a standard OPM-optimised MSR) which takes significant steps towards addressing these barriers. We also designed a ‘window coil’ active shielding system, featuring a series of simple rectangular coils placed directly onto the walls of the MSR. By mapping the remnant magnetic field inside the MSR, and the magnetic field produced by the coils, we can identify optimal coil currents and cancel the remnant magnetic field over the central cubic metre to just |B|= 670 ± 160 pT. These advances reduce the cost, installation time and siting restrictions of MSRs which will be essential for the widespread deployment of OPM-MEG.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
OPM-MEG experiments involving participant motion require an OPM-optimised MSR with a very low remnant magnetic field and active magnetic shielding.
“These constraints mean that OPM-MEG experiments involving participant motion require an OPM-optimised MSR with a very low remnant magnetic field along with ‘active’ magnetic shielding”
From Naturalistic Hyperscanning with Wearable Magnetoencephalography (2023) · cites this paper as [38]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Hämäläinen, M., Hari, R., Ilmoniemi, R. J., Knuutila, J., & Lounasmaa, O. V. (1993) · Reviews of Modern Physics · doi:10.1103/revmodphys.65.413 · Cited by 4,684
Moving magnetoencephalography towards real-world applications with a wearable system
Boto, E., Holmes, N., Leggett, J., Roberts, G., Shah, V., Meyer, S. S., Muñoz, L. D., Mullinger, K. J., Tierney, T. M., Bestmann, S., et al. (2018) · Nature · doi:10.1038/nature26147 · Cited by 1,377
Self-citationHigh-Sensitivity Atomic Magnetometer Unaffected by Spin-Exchange Relaxation
Allred, J. C., Lyman, R. N., Kornack, T. W., & Romalis, M. V. (2002) · Physical Review Letters · doi:10.1103/physrevlett.89.130801 · Cited by 1,049
Magnetoencephalography for Brain Electrophysiology and Imaging
Baillet, S. (2017) · Nature Neuroscience · doi:10.1038/nn.4504 · Cited by 992
Magnetoencephalography: Evidence of Magnetic Fields Produced by Alpha-Rhythm Currents
Cohen, D. (1968) · Science · doi:10.1126/science.161.3843.784 · Cited by 750
A compact, high performance atomic magnetometer for biomedical applications
Shah, V. K., & Wakai, R. T. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/22/8153 · Cited by 322
Self-citationMulti-Channel Whole-Head OPM-MEG: Helmet Design and a Comparison with a Conventional System
Hill, R. M., Boto, E., Rea, M., Holmes, N., Leggett, J., Coles, L. A., Papastavrou, M., Everton, S. K., Hunt, B. A. E., Sims, D., et al. (2020) · NeuroImage · doi:10.1016/j.neuroimage.2020.116995 · Cited by 285
Self-citationOn-scalp MEG system utilizing an actively shielded array of optically-pumped magnetometers
Iivanainen, J., Zetter, R., Grön, M., Hakkarainen, K., & Parkkonen, L. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.03.022 · Cited by 269
Holmes, N., Leggett, J., Boto, E., Roberts, G., Hill, R. M., Tierney, T. M., Shah, V., Barnes, G. R., Brookes, M. J., & Bowtell, R. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2018.07.028 · Cited by 250
Self-citationDetection of very weak magnetic fields (10−9gauss) by 87Rb zero-field level crossing resonances
Dupont-Roc, J., Haroche, S., & Cohen-Tannoudji, C. N. (1969) · Physics Letters A · doi:10.1016/0375-9601(69)90480-0 · Cited by 180
Magnetoencephalography in Cognitive Neuroscience: A Primer
Groß, J. (2019) · Neuron · doi:10.1016/j.neuron.2019.07.001 · Cited by 177
Fully integrated, standalone zero field optically pumped magnetometer for biomagnetism
Shah, V., Osborne, J., Orton, J., & Alem, O. (2018) · doi:10.1117/12.2299197 · Cited by 173
Self-citationHolmes, N., Tierney, T. M., Leggett, J., Boto, E., Mellor, S., Roberts, G., Hill, R. M., Shah, V., Barnes, G. R., Brookes, M. J., et al. (2019) · Scientific Reports · doi:10.1038/s41598-019-50697-w · Cited by 163
Self-citationNon-Invasive Functional-Brain-Imaging with an OPM-based Magnetoencephalography System
Borna, A., Carter, T. R., Colombo, A. P., Jau, Y.-Y., McKay, J., Weisend, M., Taulu, S., Stephen, J. M., & Schwindt, P. D. D. (2020) · PLOS ONE · doi:10.1371/journal.pone.0227684 · Cited by 158
Towards OPM-MEG in a Virtual Reality Environment
Roberts, G., Holmes, N., Alexander, N., Boto, E., Leggett, J., Hill, R. M., Shah, V., Rea, M., Vaughan, R., Maguire, E. A., et al. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.06.010 · Cited by 128
Self-citationPrecision magnetic field modelling and control for wearable magnetoencephalography
Rea, M., Holmes, N., Hill, R. M., Boto, E., Leggett, J., Edwards, L. J., Woolger, D., Dawson, E., Shah, V., Osborne, J., et al. (2021) · NeuroImage · doi:10.1016/j.neuroimage.2021.118401 · Cited by 123
Self-citationPrinciples of quasistatic magnetic shielding with cylindrical and spherical shields
Hoburg, J. F. (1995) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/15.477342 · Cited by 109
A magnetically shielded room with ultra low residual field and gradient
Altarev, I., Babcock, E., howard, B. D., Burghoff, M., Chesnevskaya, S., Chupp, T. E., Degenkolb, S., Fan, I., Fierlinger, P., Frei, A., et al. (2014) · Review of Scientific Instruments · doi:10.1063/1.4886146 · Cited by 103
Self-citationSachdeva, N., Fan, I., Babcock, E., Burghoff, M., Chupp, T. E., Degenkolb, S., Fierlinger, P., Haude, S., Kraegeloh, E., Kilian, W., et al. (2019) · Physical Review Letters · doi:10.1103/physrevlett.123.143003 · Cited by 93
Self-citationJuchem, C., Umesh Rudrapatna, S., Nixon, T. W., & de Graaf, R. A. (2015) · NeuroImage · doi:10.1016/j.neuroimage.2014.11.011 · Cited by 74
Optimal Inverse Design of Magnetic Field Profiles in a Magnetically Shielded Cylinder
Packer, M., Hobson, P. J., Holmes, N., Leggett, J., Glover, P. M., Brookes, M. J., Bowtell, R. W., & Fromhold, T. M. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.14.054004 · Cited by 67
Self-citationMinimizing magnetic fields for precision experiments
Altarev, I., Fierlinger, P., Lins, T., Marino, M. G., Nießen, B., Petzoldt, G., Reisner, M., Stuiber, S., Sturm, M., Taggart Singh, J., et al. (2015) · Journal of Applied Physics · doi:10.1063/1.4922671 · Cited by 66
Self-citationOptimization of a Coil System for Generating Uniform Magnetic Fields inside a Cubic Magnetic Shield
Cao, Q., Pan, D., Li, J., Jin, Y., Sun, Z., Lin, S., Yang, G., & Li, L. (2018) · Energies · doi:10.3390/en11030608 · Cited by 57
Magnetic Field Mapping and Correction for Moving OP-MEG
Mellor, S., Tierney, T. M., O’Neill, G. C., Alexander, N., Seymour, R. A., Holmes, N., López, J. D., Hill, R. M., Boto, E., Rea, M., et al. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3100770 · Cited by 56
Self-citationVoigt, J., Knappe-Grüneberg, S., Schnabel, A., Körber, R., & Burghoff, M. (2013) · Metrology and Measurement Systems · doi:10.2478/mms-2013-0021 · Cited by 50
Magnetic field modeling with surface currents. Part II. Implementation and usage of bfieldtools
Zetter, R., Mäkinen, A. J., Iivanainen, J., Zevenhoven, K. C. J., Ilmoniemi, R. J., & Parkkonen, L. T. (2020) · Journal of Applied Physics · doi:10.1063/5.0016087 · Cited by 48
Multi-coil magnetic field modeling
Juchem, C., Green, D., & de Graaf, R. A. (2013) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2013.08.015 · Cited by 38
Evaluating motor cortical oscillations and age-related change in autism spectrum disorder
Gaetz, W. C., Rhodes, E., Bloy, L., Blaskey, L. G., Jackel, C. R., Brodkin, E. S., Waldman, A. T., Embick, D., Hall, S., & Roberts, T. P. L. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.116349 · Cited by 33
Magnetic field compensation coil design for magnetoencephalography
Kutschka, H., Doeller, C. F., Haueisen, J., & Maeß, B. (2021) · Scientific Reports · doi:10.1038/s41598-021-01894-z · Cited by 15
Magnetic Field Mapping and Correction for Moving OP-MEGPreprint
Mellor, S., Tierney, T. M., O’Neill, G. C., Alexander, N., Seymour, R. A., Holmes, N., López, J. D., Hill, R. M., Boto, E., Rea, M., et al. (2021) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2021.05.25.444975 · Cited by 13
Self-citationTikhonov Regularization and Constrained Quadratic Programming for Magnetic Coil Design Problems
Garda, B., & Galias, Z. (2014) · International Journal of Applied Mathematics and Computer Science · doi:10.2478/amcs-2014-0018 · Cited by 7
Flagged first, then by citations
Research works printed but not linked: 1
- Juhas, A., Pekari, N. & Toepfer, H. Magnetic field of rectangular current loop with sides parallel and perpendicular to the surface of high-permeability material. SJEE 11, 701–717 (2014).
Cited by
Sorted by citations
Measurement of Frontal Midline Theta Oscillations using OPM-MEG
Rhodes, N., Rea, M., Boto, E., Rier, L., Shah, V., Hill, R. M., Osborne, J., Doyle, C., Holmes, N., Coleman, S. C., et al. (2023) · NeuroImage · doi:10.1016/j.neuroimage.2023.120024 · Cited by 91
Self-citationHolmes, N., Rea, M., Hill, R. M., Leggett, J., Edwards, L. J., Hobson, P. J., Boto, E., Tierney, T. M., Rier, L., Rivero, G. R., et al. (2023) · NeuroImage · doi:10.1016/j.neuroimage.2023.120157 · Cited by 89
Self-citationOptically Pumped Magnetometers Recent Advances and Applications in Biomagnetism: A Review
Zhao, B., Li, L., Zhang, Y., Tang, J., Liu, Y., & Zhai, Y. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3297109 · Cited by 74
Magnetic shielding mechanism and structure design of composites at low frequency: A review
Xu, X., Liu, W., Huang, Y., Li, W., & Che, S. (2023) · Journal of Magnetism and Magnetic Materials · doi:10.1016/j.jmmm.2023.170509 · Cited by 72
Applications of OPM-MEG for translational neuroscience: a perspective
Brickwedde, M., Anders, P., Kühn, A. A., Lofredi, R., Holtkamp, M., Kaindl, A. M., Grent-‘t-Jong, T., Krüger, P., Sander, T. H., & Uhlhaas, P. J. (2024) · Translational Psychiatry · doi:10.1038/s41398-024-03047-y · Cited by 59
Schofield, H., Hill, R. M., Feys, O., Holmes, N., Osborne, J., Doyle, C., Bobela, D., Corvilain, P., Wens, V., Rier, L., et al. (2024) · Imaging Neuroscience · doi:10.1162/imag_a_00283 · Cited by 54
Self-citationNaturalistic Hyperscanning with Wearable Magnetoencephalography
Holmes, N., Rea, M., Hill, R. M., Boto, E., Leggett, J., Edwards, L. J., Rhodes, N., Shah, V., Osborne, J., Fromhold, T. M., et al. (2023) · Sensors · doi:10.3390/s23125454 · Cited by 52
Self-citationGutteling, T. P., Bonnefond, M., Clausner, T., Daligault, S., Romain, R., Mitryukovskiy, S. I., Fourcault, W., Josselin, V., Prado, M. L., Palacios-Laloy, A., et al. (2023) · Sensors · doi:10.3390/s23052801 · Cited by 52
Schofield, H., Boto, E., Shah, V., Hill, R. M., Osborne, J., Rea, M., Doyle, C., Holmes, N., Bowtell, R. W., Woolger, D., et al. (2022) · Contemporary Physics · doi:10.1080/00107514.2023.2182950 · Cited by 46
Self-citationAdaptive multipole models of optically pumped magnetometer data
Tierney, T. M., Seedat, Z. A., Pier, K. S., Mellor, S., & Barnes, G. R. (2024) · Human Brain Mapping · doi:10.1002/hbm.26596 · Cited by 39
Self-citationLiu, Y., Yang, J., Cao, F., Zhang, X., & Zheng, S. (2024) · Materials · doi:10.3390/ma17225469 · Cited by 32
Sun, J., Ren, J., Xu, X., Zhou, W., Qian, J., Wang, H., & Wang, H. (2024) · Measurement · doi:10.1016/j.measurement.2024.114594 · Cited by 31
Recording of Ictal Epileptic Activity Using on‐Scalp Magnetoencephalography
Feys, O., Corvilain, P., Hecke, A. V., Sculier, C., Rikir, E., Legros, B., Gaspard, N., Leurquin‐Sterk, G., Holmes, N., Brookes, M. J., et al. (2022) · Annals of Neurology · doi:10.1002/ana.26562 · Cited by 31
Self-citationFrom cryogenic to on‐scalp magnetoencephalography for the evaluation of paediatric epilepsy
Feys, O., & Tiège, X. D. (2023) · Developmental Medicine & Child Neurology · doi:10.1111/dmcn.15689 · Cited by 30
Simultaneous whole-head electrophysiological recordings using EEG and OPM-MEG
Seedat, Z. A., Pier, K. S., Holmes, N., Rea, M., Al-Hilaly, L., Tierney, T. M., Embury, C. M., Pardington, R., Mullinger, K. J., Cross, J. H., et al. (2024) · Imaging Neuroscience · doi:10.1162/imag_a_00179 · Cited by 29
Self-citationAnalysis and suppression of magnetic field interference for SERF atomic magnetometer
Jia, L., Li, J., Song, X., Qi, S., Suo, Y., Long, T., Dai, Y., & Ning, X. (2024) · Measurement · doi:10.1016/j.measurement.2024.114266 · Cited by 26
Determining sensor geometry and gain in a wearable MEG system
Hill, R. M., Rivero, G. R., Tyler, A. J., Schofield, H., Doyle, C., Osborne, J., Bobela, D., Rier, L., Gibson, J. M., Tanner, Z., et al. (2025) · Imaging Neuroscience · doi:10.1162/imag_a_00535 · Cited by 25
Self-citationDemagnetization Parameters Evaluation of Magnetic Shields Based on Anhysteretic Magnetization Curve
Yang, J., Shi, M., Zhang, X., Ma, Y., Liu, Y., Yuan, S., & Han, B. (2023) · Materials · doi:10.3390/ma16155238 · Cited by 23
Feys, O., Corvilain, P., Labyt, É., Mahmoudzadeh, M., Routier, L., Sculier, C., Holmes, N., Brookes, M. J., Goldman, S., Romain, R., et al. (2023) · Frontiers in Neuroscience · doi:10.3389/fnins.2023.1284262 · Cited by 23
Self-citationHardware-Based Interference Suppression Techniques for OPM-MEG: A Review
Jia, L., Li, J., Song, X., Ning, X., Yang, J., Qi, S., Long, T., Wu, Z., & Wang, R. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3509679 · Cited by 23
Feys, O., Wens, V., Depondt, C., Rikir, E., Gaspard, N., Paesschen, W. V., Aeby, A., Bodart, O., Carrette, E., Holmes, N., et al. (2025) · Epilepsia · doi:10.1111/epi.18439 · Cited by 22
Self-citationExperimental studies on the performance of magnetic shields under different magnetization conditions
Yang, J., Zhang, X., Shi, M., Yuan, S., Zhang, L., Wang, L., & Han, B. (2023) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/acc412 · Cited by 22
Tang, J., Qiu, S., Zhang, L., Sun, J., & Zhou, X. (2023) · Measurement · doi:10.1016/j.measurement.2023.113904 · Cited by 22
Suppression of Magnetic Noise and Field in Cubic Low-Noise Ferrite Magnetic Shields
Sun, B., Ma, D., Fang, X., Xue, Y., Lu, J., Chen, H., Zhang, M., Wei, H., Han, B., & Zhai, Y. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3385809 · Cited by 21
Xu, X., Liu, W., Zhao, Z., Huang, Y., & Zhu, T. (2024) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2024.115538 · Cited by 20
Xu, X., Liu, W., Zhao, Z., & Ye, F. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3395694 · Cited by 19
Wearable Magnetoencephalography in a Lightly Shielded Environment
Holmes, N., Leggett, J., Hill, R. M., Rier, L., Boto, E., Schofield, H., Hayward, T., Dawson, E., Woolger, D., Shah, V., et al. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3465654 · Cited by 19
Self-citationBenchtop Magnetic Shielding for Benchmarking Atomic Magnetometers
Hobson, P. J., Holmes, N., Patel, P., Styles, B., Chalmers, J., Morley, C., Davis, A., Packer, M., Smith, T. X., Raudonyte, S., et al. (2023) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2023.3293540 · Cited by 18
Self-citationXie, X., Zhou, X., Zhao, F., Wang, W., & Sun, J. (2023) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2023.3323996 · Cited by 17
Methods for Improving Movement Compatibility of Wearable OPM-MEG: A Review
Wang, C., Cao, F., Wang, W., Xu, W., Li, W., Gao, Z., & An, N. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3329043 · Cited by 16
Comprehensive analysis of the magnetic field closed-loop control system in SERF atomic magnetometer
Jia, L., Song, X., Li, J., Suo, Y., Long, T., Wu, Z., & Ning, X. (2024) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2024.115043 · Cited by 16
Yang, Z., Li, Y., Liu, X., Li, Y., Cui, P., Zheng, S., & Zhou, X. (2024) · IEEE Transactions on Automation Science and Engineering · doi:10.1109/tase.2024.3358542 · Cited by 16
Brazdzionis, J., Radwan, M. M., Thankam, F. G., Lal, M. R., Baron, D. A., Connett, D. A., Agrawal, D. K., & Miulli, D. E. (2023) · Cureus · doi:10.7759/cureus.42544 · Cited by 15
Chen, Y., Wen, T., Fang, Y., Han, B., & Liu, X. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3375985 · Cited by 13
Xu, X., Liu, W., Zhao, Z., & Wu, W. (2024) · Measurement · doi:10.1016/j.measurement.2024.115878 · Cited by 13
Zeidman, P., Ashburner, J., Barnes, G. R., Kowalczyk, O. S., Lambert, C., Vladimir Litvak, A., Nichols, T. E., Parr, T., Tierney, T. M., & Friston, K. J. (2025) · Cerebral Cortex · doi:10.1093/cercor/bhaf234 · Cited by 13
Self-citationWang, R., Yu-jie, M., Zhao, R., Ding, J., Li, L., Yang, Y., Wang, F., Cao, Z., Zhang, X., Lin, X., et al. (2025) · NeuroImage · doi:10.1016/j.neuroimage.2025.121403 · Cited by 12
Yeom, H. G., Kim, J. S., & Chung, C. K. (2023) · Scientific Data · doi:10.1038/s41597-023-02454-y · Cited by 12
Noise Reduction and Localization Accuracy in a Mobile Magnetoencephalography System
Bardouille, T., Smith, V., Vajda, E., Leslie, C. D., & Holmes, N. (2024) · Sensors · doi:10.3390/s24113503 · Cited by 11
Self-citationIn-situ magnetic fields monitoring and compensation for zero-field atomic magnetometers
Yan, Y., Zhang, K., Liu, Z., Liu, Z., Li, X., Zhou, Y., & Lu, J. (2024) · Measurement · doi:10.1016/j.measurement.2024.115410 · Cited by 10
The active magnetic compensation coil
Xu, X., & Liu, Y. (2024) · Review of Scientific Instruments · doi:10.1063/5.0186023 · Cited by 10
Feys, O., Wens, V., Corvilain, P., Ferez, M., Holmes, N., Brookes, M. J., & Tiège, X. D. (2023) · Epilepsia · doi:10.1111/epi.17806 · Cited by 10
Self-citationCompensation of complex environmental magnetic field disturbances for biomagnetic field measurements
Wang, H., Li, H., Sun, J., Liu, X., & Liu, X. (2025) · Measurement Science and Technology · doi:10.1088/1361-6501/adafd4 · Cited by 9
Brazdzionis, J., Marino, M. A., Siddiqi, I., & Miulli, D. E. (2023) · Cureus · doi:10.7759/cureus.45361 · Cited by 9
Xu, X., Liu, Y., Han, C., & Sun, X. H. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3428625 · Cited by 9
Zhao, J., Liu, J., Zhou, J., Zhang, L., & Sun, J. (2024) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/ad8241 · Cited by 8
Cui, P., Yang, Z., Li, Y., Zhang, L., & Liu, X. (2024) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2024.3508065 · Cited by 8
Xie, X., Zhou, X., Zhao, F., Wang, W., Yin, C., & Sun, J. (2025) · Measurement · doi:10.1016/j.measurement.2025.117052 · Cited by 8
Ohkubo, M. (2023) · Review of Scientific Instruments · doi:10.1063/5.0167372 · Cited by 8
Shi, M., Zhang, X., Yang, J., Ma, Y., Zhang, L., Tian, K., & Yuan, S. (2024) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/ad4b2c · Cited by 8
Lu, Y., Qiu, S., Sun, J., Zhang, L., & Tang, J. (2024) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/ad5aa7 · Cited by 7
Fast Degaussing Procedure for a Magnetically Shielded Room
Koss, P. A., Voigt, J., Rasser, R., & Schnabel, A. (2024) · Materials · doi:10.3390/ma17235877 · Cited by 7
Target-field design of surface permanent magnets
Hobson, P. J., Morley, C. K., Davis, A., Smith, T. X., & Fromhold, T. M. (2024) · Physical Review Applied · doi:10.1103/physrevapplied.22.034015 · Cited by 7
Tanner, Z., Rier, L., Fildes, J., Rivero, G., Schofield, H., Marani, C., Holmes, N., Hill, R. M., Shah, V., Doyle, C., et al. (2025) · Imaging Neuroscience · doi:10.1162/imag.a.10 · Cited by 7
Self-citationFedosov, N., Medvedeva, D., Shevtsov, O., & Ossadtchi, A. E. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/aded35 · Cited by 7
Magnetic field compensation method of dual-beam SERF atomic magnetometer under probing-axis bias
Song, X., Wu, D., Long, T., Wu, Z., & Jia, L. (2025) · Measurement · doi:10.1016/j.measurement.2025.118102 · Cited by 6
Affortit, P., Faye, A., Jones, D. H., Benson, E., Sine, B., Burridge, J., Ndoye, M. S., Barry, L., Moukouanga, D., Barnard, S., et al. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.11.09.622826 · Cited by 6
Sun, J., Gan, Z., Shi, Y., Zhou, W., Wang, K., & CHENG, J. (2025) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2025.3585053 · Cited by 5
Yao, J., Ding, T., & Che, L. (2025) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/adc0ac · Cited by 5
Zheng, S., Shi, Z., Zhang, H., Tian, K., & Zhao, F. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3558813 · Cited by 4
The suppression of magnetic noise in a magnetic shielding unit for atomic sensor applications
Karabinaoğlu, M. S., Günay, E., Çakır, B., Caklı, B. M., & Sevim, M. H. (2025) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2025.116833 · Cited by 4
Zhao, J., Zhou, X., & Sun, J. (2024) · Electronics · doi:10.3390/electronics13163163 · Cited by 4
Bias Calibration of Optically Pumped Magnetometers Based on Variable Sensitivity
Chen, J., Ye, C., Hou, X., Niu, Y., & Sun, L. (2025) · Sensors · doi:10.3390/s25020433 · Cited by 4
Study on the magnetic shielding properties of nanocrystalline strips under bending stress
Yuan, L., Xu, X., Wang, P., Huang, Y., & Cui, P. (2024) · Physica Scripta · doi:10.1088/1402-4896/ad302f · Cited by 4
Dual-coil active disturbance rejection control for compact MEG measurement systems
Wang, H., Meng, Y., Sun, J., Zhang, L., Zhang, H., & Zheng, S. (2026) · Control Engineering Practice · doi:10.1016/j.conengprac.2026.106833 · Cited by 3
Simulation Research on Low-Frequency Magnetic Noise in Fe-Based Nanocrystalline Magnetic Shields
Kang, S., Fan, W., Lu, J., & Quan, W. (2025) · Materials · doi:10.3390/ma18020330 · Cited by 3
Shangguan, S., Dou, S., Wen, T., Han, B., Liu, X., & Feng, R. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3555727 · Cited by 3
Cross-axis coupling suppression method based on TITO-SGPC for active magnetic compensation system
Yang, Z., Cui, P., Li, Y., Gong, A., & Zheng, S. (2025) · Measurement Science and Technology · doi:10.1088/1361-6501/addd3a · Cited by 3
Sun, J., Gan, Z., Zhou, W., Shi, Y., & Gong, X. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3584163 · Cited by 3
High-Precision Noise Suppression Method Based on IACAC-LSSVM for MEG Measurement
Yang, Z., Cui, P., Gong, A., Li, Y., & Zheng, S. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3588976 · Cited by 3
Wearable magnetoencephalography in an open environmentPreprint
Zheng, F., Li, C., Li, T., Yang, R., Xu, W., Cheng, H., Li, D., He, K., Yin, Y., Ru, X., et al. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.07.28.740707 · Cited by 3
Active Compensation for OPM-MEG Inside a Two-Layer Magnetically Shielded Room
Władziński, M., Jodko-Władzińska, A., & Sander, T. H. (2024) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-61625-9_6 · Cited by 2
Wu, H., Zhao, X., Cui, P., Zhang, H., Liu, J., & Wen, T. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3533616 · Cited by 2
Wang, H., Chen, Y., Li, H., Wang, X., Meng, Y., Sun, J., & Liu, X. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3545058 · Cited by 2
Sun, J., Wang, H., Ren, J., Chen, D., Wang, H., & Zhou, W. (2025) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/ae08ca · Cited by 2
Sun, J., Wang, H., Liu, X., Zhang, H., Li, H., Wang, Z., & Tian, G. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3619282 · Cited by 2
Shen, Z., Sai, T., Shen, P., Ye, J., Liu, M., & Wang, P. (2025) · Journal of Alloys and Compounds · doi:10.1016/j.jallcom.2025.185141 · Cited by 2
Active magnetic compensation control systems: a review
Xu, X., Zhang, X., Liu, Y., & Han, C. (2025) · Measurement · doi:10.1016/j.measurement.2025.120012 · Cited by 2
Yu, X., Zhang, D., Carey, T., Pöppel, E., Lau, E. F., & Sander, T. H. (2026) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2026.110149 · Cited by 2
Designing Optimal Loop, Saddle, and Ellipse-Based Magnetic Coils by Spherical Harmonic Mapping
Hobson, P. J., Hardwicke, N. L., Davis, A., Smith, T. X., Morley, C., Packer, M., Holmes, N., Weil, M. A., Brookes, M. J., Bowtell, R. W., et al. (2023) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2023.3284138 · Cited by 2
Self-citationAccurate magnetization modeling in multi-dimensional applications
Wang, Y., Yang, W., Liu, F., & Wang, Q. (2024) · Journal of Applied Physics · doi:10.1063/5.0173725 · Cited by 2
Wang, W., Tang, J., Zhao, F., Zhang, L., Zhai, Y., & Zhang, H. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3593510 · Cited by 2
Chen, Y., Shangguan, S., Liu, X., Deng, Y. Q., Song, P., & Han, B. (2025) · Measurement · doi:10.1016/j.measurement.2025.118996 · Cited by 2
Magnetoencephalography Without a Shielded RoomPreprint
Bezsudnova, Y., Alexander, N., Mellor, S., Mitryukovskiy, S. I., Romain, R., Palacios-Laloy, A., Barnes, G. R., Callaghan, M. F., & Tierney, T. M. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.08.06.743270 · Cited by 2
Self-citationMagnetic field interaction with high-permeability planar passive shields
Holmes, N., Leggett, J., Brookes, M. J., Edwards, L. J., Woolger, D., & Bowtell, R. W. (2025) · Journal of Applied Physics · doi:10.1063/5.0283999 · Cited by 1
Self-citationActive–Passive Coupling Analysis and Optimal Design of Internal and External Hybrid Coils in MSC
Liu, Y., Xu, X., & Han, C. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3641948 · Cited by 1
Neuroimaging and the Epilepsy Journey
Brinkmann, B. H. (2026) · Neuroimaging Clinics of North America · doi:10.1016/j.nic.2025.12.001 · Cited by 1
Yang, Z., Li, Y., Gong, A., Cui, P., & Zheng, S. (2026) · Measurement Science and Technology · doi:10.1088/1361-6501/ae411c · Cited by 1
Simulation and optimization of the Active Magnetic Shield of the n2EDM experiment
Ayres, N. J., Ban, G., Bison, G., Bodek, K., Бондар, В. М., Bouillaud, T., Caratsch, G. L., Chanel, E., Chen, W., Crawford, C., et al. (2026) · Journal of Instrumentation · doi:10.1088/1748-0221/21/04/p04029 · Cited by 1
Bailey, L. M., Knox, C., & Bardouille, T. (2026) · Sensors · doi:10.3390/s26103131 · Cited by 1
Cheng, Y., Huang, J., Luo, Y., & Lu, F. (2024) · Machines · doi:10.3390/machines12050317 · Cited by 1
Bi-planar magnetic stabilisation coils for an inertial sensor based on atom interferometry
Davis, A., Hobson, P. J., Smith, T. X., Morley, C. K., Sewell, H., Cotter, J. P., & Fromhold, T. M. (2024) · Physics Open · doi:10.1016/j.physo.2024.100227 · Cited by 1
A Design Method of High-Constrained and Anti-Disturbance Coil for Magnetic Shielding Rooms
Yuan, S., Shi, M., Zhang, X., Cui, P., Yang, J., Zhao, F., & Wang, L. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3551014 · Cited by 1
Peng, T., Liu, Z., Chen, J., Wang, S., & Ge, A. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3556216 · Cited by 1
Space magnetic field control based on active disturbance rejection generalized predictive control
Yang, Z., Cui, P., Li, Y., Jin, Y., & Zheng, S. (2025) · Journal of Applied Physics · doi:10.1063/5.0285495 · Cited by 1
Zhao, J., Liu, J., Zhang, L., & Sun, J. (2023) · SSRN Electronic Journal · doi:10.2139/ssrn.4511630
Exploring Directed Connectivity Patterns of OPM-MEG with Multivariate Transfer EntropyPreprint
Pedersen, M., Vashista, G., Henare, D., & Sowman, P. F. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.02.22.581475
Radiation and EMI shielding of 3D printed lightweight components for neuroimaging devices
Mirabedini, A., McCrowe, C., Welch, D., Deb, P., Pang, T. Y., Langenberg, F., Tzheng, S., Obruchkov, S. I., Davis, S. M., Donnan, G. A., et al. (2025) · Smart Materials in Manufacturing · doi:10.1016/j.smmf.2025.100085
Linearization of Magnetic Shielding Room Model based on Second-order Variable Phaser
Yang, Z., Cui, P., & Li, Y. (2025) · IEEE International Instrumentation and Measurement Technology Conference · doi:10.1109/i2mtc62753.2025.11079008
Hybrid MODWT-CSP Approach for Motor Imagery Classification in MEG-BCI Systems
Koç, G., Yousif, M. A. A., & Öztürk, M. (2025) · Signal Processing and Communications Applications Conference (SIU) · doi:10.1109/siu66497.2025.11111752
Metadata
- Authors
- Holmes, N., Rea, M., Chalmers, J., Leggett, J., Edwards, L. J., Nell, P., Pink, S., Patel, P., Wood, J., Murby, N., Woolger, D., Dawson, E., Mariani, C., Tierney, T. M., Mellor, S., O’Neill, G. C., Boto, E., Hill, R. M., Shah, V., Osborne, J., Pardington, R., Fierlinger, P., Barnes, G. R., Glover, P., Brookes, M. J., & Bowtell, R.
- Affiliations
- University of Nottingham (GB)
- Wellcome Centre for Human Neuroimaging (GB)
- QuSpin (United States) (US)
- Technical University of Munich (DE)
- Corresponding author
- Holmes, Niall
- ORCID
- Venue
- Scientific Reports
- Published
- 2022-08-09
- Volume
- 12
- Issue
- 1
- Pages
- 13561-13561
- ISSN
- 2045-2322 (ISSN-L)
- PMC
- PMC9363499
- OpenAlex
- W4291020559
- Cited by
- 121
- References
- 31
- Citations by year
- 2026: 24 · 2025: 39 · 2024: 34 · 2023: 22 · 2022: 2
- FWCI
- 9.95
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Scientific Reports (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Repository@Nottingham (University of Nottingham) (publishedVersion, other-oa)
- Repository@Nottingham (University of Nottingham) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- UCL Discovery (University College London) (submittedVersion)
- mediaTUM – the media and publications repository of the Technical University Munich (Technical University Munich) (submittedVersion, other-oa)
- Funders
- Engineering and Physical Sciences Research Council (EP/T001046/1)
- Innovate UK (104604)
- Wellcome Trust (203257/Z/16/Z)
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electromagnetic Compatibility and Measurements
- Physical Sciences › Engineering › Biomedical Engineering › Characterization and Applications of Magnetic Nanoparticles
- Keywords
- magnetically shielded room, active magnetic shielding, optically pumped magnetometers, mu-metal, electromagnetic coils
- MeSH terms
- Brain, Humans, Magnetic Resonance Imaging, Magnetic Resonance Spectroscopy, Magnetoencephalography, Functional Neuroimaging, Magnetic Fields
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review