Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups
Open accessAbstract
This study aims to establish local diagnostic reference levels (DRLs) for contrast-enhanced computed tomography (CECT) abdomen and pelvis scans in adult patients within the Mangalore region, emphasizing variations in radiation dose metrics across body mass index (BMI) categories.
Overview
CoChat Summary
## TL;DR
This prospective multicenter study of 500 adults undergoing multiphase contrast-enhanced CT abdomen/pelvis scans in Mangalore established local DRLs using the 75th percentile of CTDIvol and DLP. Radiation dose increased significantly with BMI across all phases, suggesting dose optimization should be tailored to body habitus.
## Key findings
- The study analyzed 500 abdomen and pelvis CECT examinations across five health care facilities; patients were 18–60 years old.
- The local DRLs for the plain phase were CTDIvol 8.30 mGy and DLP 460.50 mGy·cm at the 75th percentile.
- Dose metrics rose with BMI: in the plain phase, mean CTDIvol increased from 7.42 ± 1.41 mGy in underweight patients to 8.41 ± 1.21 mGy in overweight patients; DLP ranged from 400.00 to 450.55 mGy·cm.
- BMI showed significant positive correlations with dose descriptors: CTDIvol r = 0.333–0.429 and DLP r = 0.338–0.486 across phases, all p < 0.001.
## Methods
- Design: Prospective multicenter cross-sectional study in the Mangalore region.
- Sample/data: 500 adult patients (18–60 years) undergoing multiphase CECT abdomen/pelvis; excluded bariatric patients and those undergoing single-slice CT scans.
- Analysis: Recorded CTDIvol and DLP for plain, arterial, and portovenous phases; used descriptive statistics, ANOVA with Bonferroni post hoc, Kruskal–Wallis with post hoc tests, and Pearson correlation.
## Limitations
- The paper states that excluding bariatric patients and single-slice CT systems may limit applicability to all patient populations and older technologies.
- The study is limited to a single region (Mangalore), so the authors say national-level DRLs need further research.
- Evident from the text: the analysis focuses on CT dose descriptors only and does not report clinical outcomes or image quality measures.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures



Verification
Claims the paper makes
1 Valid · 6 Not valid · 5 Not checked · 43 Own results and interpretations
At least one cited source doesn’t support this claim or points to a different paper, and none confirmed it — see the verdicts below.
In the paper
“demonstrating lower radiation doses than national and international benchmarks.”
- Also stated in: § Discussion · boosted
“The DRLs derived in this study (CTDIvol: 8.3 mGy, DLP: 460.5 mGy·cm) are notably lower than most international and national DRLs reported in the literature.”
§ Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups > Abstract
What the sources say
- Disputedcites [9] sourceExperimental — accuracy not yet measured
The Results (page unavailable) say the study’s head DRL was comparable to the University of California study, while its chest, abdomen, and multiphase-neck DRLs were considerably higher—not lower. The Conclusion says a national CT-dose benchmark for India was anticipated in the future, rather than reporting that these DRLs fell below one. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
- Could not verifycites [14] sourceExperimental — accuracy not yet measured
The Introduction discusses DRLs generally, while the Results (section 3.2) compare estimated standard doses among scanners at the study facility. Searches for national and international benchmarks found no passage establishing study DRLs or comparing its doses with those benchmarks. (LLM, conf=0.76)
- Not checked — no full text of the sourcecites [10] source
boostedn = 500- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 2 assumptions not shown here.
- Also stated in: § Discussion · boosted
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“Among Indian studies, the present DRL values are lower than those from Kerala (CTDIvol: 10.6 mGy, DLP: 509.3 mGy·cm) and South India (CTDIvol: 12 mGy, DLP: 550 mGy·cm). 10 12”
§ Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups > Discussion
What the sources say
- Not checked — no full text of the sourcecites [10] source
- Could not verifycites [12] sourceExperimental — accuracy not yet measured
The Results and Discussion report DRLs for Addis Ababa and compare them with values from several other countries, but not Kerala or South India. Searches for Kerala, South India, and the quoted CTDIvol and DLP values found no passage addressing the cited comparison. The citing paper’s own “present DRL values” cannot be assessed from this source. (LLM, conf=0.76)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“Among Indian studies, the present DRL values are lower than those from Kerala (CTDIvol: 10.6 mGy, DLP: 509.3 mGy·cm) and South India (CTDIvol: 12 mGy, DLP: 550 mGy·cm). 10 12”
§ Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups > Discussion
What the sources say
- Could not verifycites [12] sourceExperimental — accuracy not yet measured
The Results and Discussion report CT dose values for Addis Ababa and compare them with values from several other countries (Tables 4–5; page unavailable). Searches for South India and the claimed 12 mGy CTDIvol and 550 mGy·cm DLP found no passage reporting those values for a South India study. (LLM, conf=0.77)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“studies conducted in Portugal (CTDIvol: 18 mGy, DLP: 800 mGy·cm) and Japan (CTDIvol: 18.2 mGy, DLP: 870.9 mGy·cm) reflect significantly higher dose levels. 9 14”
§ Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups > Discussion
What the sources say
- Could not verifycites [14] sourceExperimental — accuracy not yet measured
The Results (§3.2; page unavailable) discuss CTDIvol and DLP and refer to standard-dose values in Table 3, but the passages in our copy do not give the claimed pair of values. Searches for “18.2” and “870.9” found no passage reporting those figures for Japan. (LLM, conf=0.72)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“studies conducted in Portugal (CTDIvol: 18 mGy, DLP: 800 mGy·cm) and Japan (CTDIvol: 18.2 mGy, DLP: 870.9 mGy·cm) reflect significantly higher dose levels. 9 14”
§ Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups > Discussion
What the sources say
- Could not verifycites [9] sourceExperimental — accuracy not yet measured
The Results section says the authors compared their Kerala dose data with data from other countries, but the passages in our copy do not identify Portugal or report the attributed CTDIvol of 18 mGy and DLP of 800 mGy·cm. Searches for Portugal and the specific values found no passage establishing them; the table contents are not present in the retrieved text. (LLM, conf=0.76)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
29 more claims
Create a free account to see all 55 claims and every source verdict.
- The quote only partly states this claim. 3
- Probably not a substantial claim. 7
- It may state two separate claims as one. 13
Citations
Cites
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
Vañó, E., Miller, D. L., Martin, C. J., Rehani, M. M., Kang, K., Rosenstein, M., Ortiz-López, P., Mattsson, S., Padovani, R., & Rogers, A. (2017) · Annals of the ICRP · doi:10.1177/0146645317717209 · Cited by 1,043
Radiation dose reduction in computed tomography: techniques and future perspective
Yu, L., Liu, X., Leng, S., Kofler, J. M., Ramirez-Giraldo, J. C., Qu, M., Christner, J., Fletcher, J. G., & McCollough, C. H. (2009) · Imaging in Medicine · doi:10.2217/iim.09.5 · Cited by 395 · cited 1×
“These methods preserve optimum image quality with dose reduction.”
Establishment of CT diagnostic reference levels in Ireland
Foley, S. J., McEntee, M. F., & Rainford, L. A. (2012) · The British Journal of Radiology · doi:10.1259/bjr/15839549 · Cited by 186
Updated Australian diagnostic reference levels for adult CT
Lee, K. L., Beveridge, T., Sanagou, M., & Thomas, P. (2020) · Journal of Medical Radiation Sciences · doi:10.1002/jmrs.372 · Cited by 75
Roch, P., Célier, D., Dessaud, C., et al. (2017) · European Journal of Radiology · doi:10.1016/j.ejrad.2017.11.002 · Cited by 70
The establishment of computed tomography diagnostic reference levels in Portugal
Santos, J., Foley, S. J., Paulo, G., McEntee, M. F., & Rainford, L. A. (2013) · Radiation Protection Dosimetry · doi:10.1093/rpd/nct226 · Cited by 69
Matsunaga, Y., Chida, K., Kondo, Y., KOBAYASHI, K. P., Kobayashi, M., Minami, K., Suzuki, S., & Asada, Y. (2018) · British Journal of Radiology · doi:10.1259/bjr.20180290 · Cited by 66
National diagnostic reference levels: What they are, why we need them and what’s next
Thomas, P. (2022) · Journal of Medical Imaging and Radiation Oncology · doi:10.1111/1754-9485.13375 · Cited by 27
Erem, G., Ameda, F., Otike, C., Olwit, W., Mubuuke, A. G., Schandorf, C., Kisolo, A., & Kawooya, M. G. (2022) · BMC Medical Imaging · doi:10.1186/s12880-022-00838-x · Cited by 26 · cited 1×
“5,6 DRLs are typically determined through surveys of patient doses collected from standard procedures performed across multiple facilities within a country or region.”
Sohrabi, M., Parsi, M., & Mianji, F. (2018) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncx252 · Cited by 18
Cited by
Jaseemudheen, M. M., Raghuraja, U., Shetty, L., & Sri Krishna, U. (2026) · Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging · doi:10.1055/s-0046-1815910 · Cited by 1
Self-citationLeung, K. H., Lee, Y. L., Ng, T. M., & Lam, T. P. (2026) · Journal of medical imaging and radiation sciences · doi:10.1016/j.jmir.2026.102563
Metadata
- Authors
- M., J. M., U., R., U., S. K., Shetty, L., Dabholkar, A., & Ramachandraiah, C.
- Affiliations
- Department of Radiodiagnosis and Imaging, Nitte (Deemed to be university), K.S. Hegde Medical Academy, Mangalore, Karnataka, India
- K S Hegde Medical Academy (IN)
- Dr. D.Y. Patil Vidyapeeth, Pune (IN)
- Department of MIT, A.J. Institute of Medical Sciences, Mangalore, Karnataka, India
- ORCID
- Venue
- Indian Journal of Radiology and Imaging
- Publisher
- Georg Thieme Verlag KG
- Published
- 2025-04-01
- Volume
- 36
- Issue
- 01
- Pages
- 040-047
- ISSN
- 0970-2016 (ISSN-L)
- 0971-3026
- 1998-3808
- PMC
- PMC12788930
- OpenAlex
- W4409051350
- Cited by
- 2
- References
- 15
- Citations by year
- 2026: 2
- FWCI
- 1.14
- Type
- journal article
- Language
- en
- Open access
- diamond
- doi.org
- Licence
- Locations
- Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiology practices and education
- Keywords
- diagnostic reference levels, CT radiation dose, body mass index, contrast-enhanced CT, volumetric CT dose index, dose-length product, dose optimization, patient safety, Mangalore
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups. https://app.cochat.ai/papers/10.1055/s-0045-1806868
- The paper
- M., J. M., U., R., U., S. K., Shetty, L., Dabholkar, A., & Ramachandraiah, C. (2026). Establishing Diagnostic Reference Level for Adult Abdomen and Pelvis CT Scans: A Benchmarking Study across BMI Groups. Indian Journal of Radiology and Imaging. https://doi.org/10.1055/s-0045-1806868
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.