Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities
Open accessThe DOI matches its Crossref record; 13 of the 13 references it cites were found in a registry or an index.
Abstract
The accuracy of wrist worn heart rate monitors based on photoplethysmography (PPG) is not fully clinically accepted. Therefore, we aimed to validate heart rate measurements of a commercially available PPG heart rate monitor, i.e. the Garmin Forerunner® 225. Twelve healthy volunteers (six women; mean age: 28 years) performed a treadmill protocol consisting of: five minutes sitting, five minutes standing, 10 minutes walking at 4 km/h, 10 minutes walking at a gradient of 5% and intensity of 4–6 metabolic equivalents (METs), 10 minutes walking at a gradient of 8% and intensity of seven METs or more. Walking speeds were individually determined. Walking bouts were separated by a standardised five minute rest period. Heart rate was measured as the average of the last three minutes standing and of each walking bout. A three lead patch-based electrocardiogram (ECG; Zensor®) was used as criterion method. Statistical analyses included Pearson’s correlation (r), paired t-tests, root mean squared error (RMSE) and Bland?Altman plots. The mean values per three minutes of every condition did not differ significantly between the Garmin Forerunner® 225 and the Zensor®. RMSE was 3.01 beats per minute (bpm) or 2.89%. The Bland–Altman bias was 1.57 bpm. Limits of agreement (LoA) were wide, ranging from 32.53 to 29.40 bpm. However, Pearson’s r ranged from 0.650 to 0.868 suggesting moderate to strong validity. Generally, mean heart rates, r values, RMSE and the Bland–Altman bias indicated good overall agreement in this sample of healthy adults, but wide LoA are making it difficult to trust individual measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Garber, C. E., Blissmer, B., Deschenes, M. R., Franklin, B. A., Lamonte, M. J., Lee, I.-M., Nieman, D. C., & Swain, D. P. (2011) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e318213fefb · Cited by 9,319
2011 Compendium of Physical Activities: A Second Update of Codes and MET Values
AINSWORTH, B. E., HASKELL, W. L., HERRMANN, S. D., MECKES, N., BASSETT, D. R., TUDOR-LOCKE, C., GREER, J. L., VEZINA, J., WHITT-GLOVER, M. C., & LEON, A. S. (2011) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e31821ece12 · Cited by 6,317
Another look at measures of forecast accuracy
Hyndman, R. J., & Koehler, A. B. (2006) · International Journal of Forecasting · doi:10.1016/j.ijforecast.2006.03.001 · Cited by 5,757
Comparison of cardioprotective benefits of vigorous versus moderate intensity aerobic exercise
Swain, D. P., & Franklin, B. A. (2006) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2005.07.130 · Cited by 660 · cited 1×
“Since exercise intensity is the most important determinant for health benefits of exercise [11] and heart rate remains the preferred method to prescribe and monitor training intensity [12], these heart rate monitors need to be valid.”
Assessment of Physical Activity and Energy Expenditure: An Overview of Objective Measures
Hills, A. P., Mokhtar, N., & Byrne, N. M. (2014) · Frontiers in Nutrition · doi:10.3389/fnut.2014.00005 · Cited by 575
Accuracy of wrist-worn heart rate monitors
Wang, R., Blackburn, G., Desai, M., Phelan, D., Gillinov, L., Houghtaling, P., & Gillinov, M. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2016.3340 · Cited by 386
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · cited 2×
“LoA provided by the Bland–Altman statistics are discussed, but rarely affect the final conclusion [6,10,21,22].”
Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M., & Lee, J.-M. (2016) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000106 · Cited by 276 · cited 1×
“Nevertheless, even in studies with larger sample sizes, LoA tend to be wide depending on the exercise intensity and on the heart rate monitor tested [10,22,28].”
The advantages of wearable green reflected photoplethysmography
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9506-z · Cited by 209 · cited 1×
“Another external factor influencing PPG sensitivity is sweat [24].”
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203 · cited 1×
“quality of PPG data is sensitive to external factors such as large movements [23].”
Cited by
Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., et al. (2020) · JMIR mHealth and uHealth · doi:10.2196/18694 · Cited by 715
“A total of 29 studies (266 comparisons) examined wearable device heart rate measurements compared with reference standard criterion measures, including electrocardiography [22,23,38-40,43,44,54,61,62,67-70,142,162,176], Polar brand chest straps [20,21,24-28,58,63,71,163], and pulse oximetry [66], in controlled setting…”
BackgroundHenriksen, A., Haugen Mikalsen, M., Woldaregay, A. Z., Muzny, M., Hartvigsen, G., Hopstock, L. A., & Grimsgaard, S. (2018) · Journal of Medical Internet Research · doi:10.2196/jmir.9157 · Cited by 542
“In 22 of the studies, one or more Garmin devices were used [32,33,46,49,50,62,77-92].”
MethodFalter, M., Budts, W., Goetschalckx, K., Cornelissen, V., & Buys, R. (2019) · JMIR mHealth and uHealth · doi:10.2196/11889 · Cited by 142
“Validation studies have been done to evaluate the accuracy of HR, EE, and other measurements in healthy subjects for a variety of fitness trackers [4-16].”
Self-citationZhang, Y., Weaver, R. G., Armstrong, B., Burkart, S., Zhang, S., & Beets, M. W. (2020) · Journal of Sports Sciences · doi:10.1080/02640414.2020.1767348 · Cited by 88
Rajala, S., Lindholm, H., & Taipalus, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac7ac · Cited by 70
“For instance, Stahl et al (2016) and Claes et al (2017) have validated commercial PPG heart rate monitors and found them to be accurate.”
Background
Metadata
- Authors
- Claes, J., Buys, R., Avila, A., Finlay, D., Kennedy, A., Guldenring, D., Budts, W., & Cornelissen, V.
- Venue
- Journal of Medical Engineering & Technology
- Publisher
- Informa UK Limited
- Published
- 2017-07-04
- Volume
- 41
- Issue
- 6
- Pages
- 480-485
- ISSN
- 0309-1902 (ISSN-L)
- 1464-522X
- OpenAlex
- W2725379293
- Cited by
- 49
- References
- 13
- FWCI
- 1.98
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities. https://app.cochat.ai/papers/10.1080/03091902.2017.1333166
- The paper
- Claes, J., Buys, R., Avila, A., Finlay, D., Kennedy, A., Guldenring, D., Budts, W., & Cornelissen, V. (2017). Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities. Journal of Medical Engineering & Technology. https://doi.org/10.1080/03091902.2017.1333166
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.