Closed-loop control in insulin pumps for type-1 diabetes mellitus: safety and efficacy
Open accessThe DOI matches its Crossref record; 27 of the 27 references it cites were found in a registry or an index.
Abstract
INTRODUCTION: Type 1 diabetes is a lifelong disease with high management burden. The majority of people with type 1 diabetes fail to achieve glycemic targets. Algorithm-driven automated insulin delivery (closed-loop) systems aim to address these challenges. This review provides an overview of commercial and emerging closed-loop systems. AREAS COVERED: We review safety and efficacy of commercial and emerging hybrid closed-loop systems. A literature search was conducted and clinical trials using day-and-night closed-loop systems during free-living conditions were used to report on safety data. We comment on efficacy where robust randomized controlled trial data for a particular system are available. We highlight similarities and differences between commercial systems. EXPERT OPINION: Study data shows that hybrid closed-loop systems are safe and effective, consistently improving glycemic control when compared to standard therapy. While a fully closed-loop system with minimal burden remains the end-goal, these hybrid closed-loop systems have transformative potential in diabetes care.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,237 · cited 1×
“Based on consensus statements on the reporting of CGM data, the most common outcome measures of glycaemic control are (27, 34): mean sensor glucose; time in the normoglycaemic range defined as between 3.”
BackgroundSelf-citationInternational consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069 · cited 4×
“Sensor accuracy is measured through the mean absolute relative difference, which is the average of the absolute error between CGM and matched reference values (27).”
BackgroundInfluentialSelf-citationMiller, K. M., Foster, N. C., Beck, R. W., Bergenstal, R. M., DuBose, S. N., DiMeglio, L. A., Maahs, D. M., & Tamborlane, W. V. (2015) · Diabetes Care · doi:10.2337/dc15-0078 · Cited by 1,254 · cited 2×
“More than half of people with type 1 diabetes in the USA use pump therapy (4).”
BackgroundSix-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Raghinaru, D., Lum, J. W., Buckingham, B. A., Kudva, Y. C., Laffel, L. M., Levy, C. J., Pinsker, J. E., Wadwa, R. P., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1907863 · Cited by 1,030 · cited 2×
“6 more hours per day spent in the target range (49).”
BackgroundMethodValidation of Time in Range as an Outcome Measure for Diabetes Clinical Trials
Beck, R. W., Bergenstal, R. M., Riddlesworth, T. D., Kollman, C., Li, Z., Brown, A. S., & Close, K. L. (2019) · Diabetes Care · doi:10.2337/dc18-1444 · Cited by 793 · cited 1×
“It is now suggested to use CGM metrics alongside HbA1c when assessing glycaemic control (32, 33).”
MethodThreshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652
Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes
Bergenstal, R. M., Garg, S., Weinzimer, S. A., Buckingham, B. A., Bode, B. W., Tamborlane, W. V., & Kaufman, F. R. (2016) · JAMA · doi:10.1001/jama.2016.11708 · Cited by 644 · cited 1×
“There were no episodes of severe hypoglycaemia or DKA over 12,389 patient days, confirming the safety of the system in the home setting (36, 37) (Table 2).”
BackgroundGarg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585 · cited 2×
“There were no episodes of severe hypoglycaemia or DKA over 12,389 patient days, confirming the safety of the system in the home setting (36, 37) (Table 2).”
BackgroundMethodThe fallacy of average: how using HbA1c alone to assess glycemic control can be misleading
Beck, R. W., Connor, C. G., Mullen, D. M., Wesley, D. M., & Bergenstal, R. M. (2017) · Diabetes Care · doi:10.2337/dc17-0636 · Cited by 477 · cited 3×
“It is now suggested to use CGM metrics alongside HbA1c when assessing glycaemic control (32, 33).”
BackgroundMethodWeisman, A., Bai, J.-W., Cardinez, M., Kramer, C. K., & Perkins, B. A. (2017) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(17)30167-5 · Cited by 463 · cited 2×
“In a systematic review by Weisman et al, PID algorithms had less improved time in target compared to model predictive control (MPC) and fuzzy logic algorithms (22).”
BackgroundMethodHome Use of an Artificial Beta Cell in Type 1 Diabetes
Thabit, H., Tauschmann, M., Allen, J. M., Leelarathna, L., Hartnell, S., Wilinska, M. E., Acerini, C. L., Dellweg, S., Benesch, C., Heinemann, L., et al. (2015) · New England Journal of Medicine · doi:10.1056/nejmoa1509351 · Cited by 450 · cited 1×
“Thabit et al conducted a twelve week open-label randomised crossover study comparing the closed-loop system to sensor augmented pump therapy in 33 adults (55), while Tauschmann et al reported a similarly designed study enrolling 86 participants aged 6 years and older with suboptimal glycaemic control (56).”
BackgroundSelf-citationTauschmann, M., Thabit, H., Bally, L., Allen, J. M., Hartnell, S., Wilinska, M. E., Ruan, Y., Sibayan, J., Kollman, C., Cheng, P., et al. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31947-0 · Cited by 440 · cited 2×
“In the study by Tauschmann et al, there were no severe hypoglycaemia events and one DKA event in the closed-loop group due to infusion set failure (56).”
BackgroundSelf-citationBekiari, E., Kitsios, K., Thabit, H., Tauschmann, M., Athanasiadou, E., Karagiannis, T., Haidich, A.-B., Hovorka, R., & Tsapas, A. (2018) · BMJ · doi:10.1136/bmj.k1310 · Cited by 437 · cited 1×
“These systems have demonstrated efficacy and safety in the home setting in children, adolescents, adults and pregnant women (21, 22).”
BackgroundSelf-citationMisso, M. L., Egberts, K. J., Page, M., O'Connor, D., & Shaw, J. (2010) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd005103.pub2 · Cited by 406
Forlenza, G. P., Li, Z., Buckingham, B. A., Pinsker, J. E., Cengiz, E., Wadwa, R. P., Ekhlaspour, L., Church, M. M., Weinzimer, S. A., Jost, E., et al. (2018) · Diabetes Care · doi:10.2337/dc18-0771 · Cited by 246
Hovorka, R., Kumareswaran, K., Harris, J., Allen, J. M., Elleri, D., Xing, D., Kollman, C., Nodale, M., Murphy, H. R., Dunger, D. B., et al. (2011) · BMJ · doi:10.1136/bmj.d1855 · Cited by 244
Self-citationComing of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 223 · cited 1×
“1 Proportional, integral, derivative controllers—The proportional, integral, derivative (PID) controller is an algorithm treating to target by directing insulin doses based on the difference from target glucose at the current point of time (proportional), the rate of change in measured glucose level over time (derivat…”
BackgroundSelf-citationBenhamou, P.-Y., Franc, S., Reznik, Y., Thivolet, C., Schaepelynck, P., Renard, E., Guerci, B., Chaillous, L., Lukas-Croisier, C., Jeandidier, N., et al. (2019) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30003-2 · Cited by 217
One year clinical experience of the first commercial hybrid closed-loop system
Lal, R. A., Basina, M., Maahs, D. M., Hood, K., Buckingham, B., & Wilson, D. M. (2019) · Diabetes Care · doi:10.2337/dc19-0855 · Cited by 215
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0143 · Cited by 197
Clarke, W. L., Anderson, S., Breton, M., Patek, S., Kashmer, L., & Kovatchev, B. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300506 · Cited by 185
Breton, M. D., Cherñavvsky, D. R., Forlenza, G. P., DeBoer, M. D., Robic, J., Wadwa, R. P., Messer, L. H., Kovatchev, B. P., & Maahs, D. M. (2017) · Diabetes Care · doi:10.2337/dc17-0883 · Cited by 179
Kovatchev, B., Cheng, P., Anderson, S. M., Pinsker, J. E., Boscari, F., Buckingham, B. A., Doyle, F. J., Hood, K. K., Brown, S. A., Breton, M. D., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0333 · Cited by 141
Multinational Home Use of Closed-Loop Control Is Safe and Effective
Anderson, S. M., Raghinaru, D., Pinsker, J. E., Boscari, F., Renard, E., Buckingham, B. A., Nimri, R., Doyle, F. J., Brown, S. A., Keith-Hynes, P., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2468 · Cited by 120 · cited 1×
“The system was first tested for safety in the unmonitored home setting in an uncontrolled study of 30 adults who used the algorithm for a two week period (45), which was extended to six months for 14 participants (46).”
MethodGingras, V., Taleb, N., Roy‐Fleming, A., Legault, L., & Rabasa‐Lhoret, R. (2018) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.13052 · Cited by 113 · cited 3×
“Other limitations are its dosage cap and potential side-effects such as nausea and increased risk of hypoglycaemia with pre-meal administration (86).”
BackgroundInfluentialSherr, J. L., Buckingham, B. A., Forlenza, G. P., Galderisi, A., Ekhlaspour, L., Wadwa, R. P., Carria, L., Hsu, L., Berget, C., Peyser, T. A., et al. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0286 · Cited by 92 · cited 1×
“The single-arm study included 36 participants, aged 6 to 65 years and was preceded by a 7 day standard outpatient therapy phase (68).”
MethodRandomized Controlled Trial of Mobile Closed-Loop Control
Kovatchev, B., Anderson, S. M., Raghinaru, D., Kudva, Y. C., Laffel, L. M., Levy, C., Pinsker, J. E., Wadwa, R. P., Buckingham, B., Doyle, F. J., et al. (2020) · Diabetes Care · doi:10.2337/dc19-1310 · Cited by 60 · cited 3×
“In the three month parallel group study there were no DKA events and one severe hypoglycaemia event not related to device malfunction (48).”
BackgroundMethod
Cited by
Collyns, O. J., Meier, R. A., Betts, Z. L., Chan, D. S. H., Frampton, C., Frewen, C. M., Hewapathirana, N. M., Jones, S. D., Roy, A., Grosman, B., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2250 · Cited by 276
“Cambridge system (CamAPS FX), which have shown comparable outcomes in clinical trials (11,12).”
Chen, S., Fan, S., Qiao, Z., Wu, Z., Lin, B., Li, Z., Riegler, M. A., Wong, M. Y. H., Opheim, A., Korostynska, O., et al. (2025) · Advanced Materials · doi:10.1002/adma.202500412 · Cited by 164
Engineering islets from stem cells for advanced therapies of diabetes
Siehler, J., Blöchinger, A. K., Meier, M., & Lickert, H. (2021) · Nature Reviews Drug Discovery · doi:10.1038/s41573-021-00262-w · Cited by 125
Matejko, B., Juza, A., Kieć-Wilk, B., Cyranka, K., Krzyżowska, S., Chen, X., Cohen, O., Da Silva, J., Malecki, M. T., & Klupa, T. (2022) · Diabetes Care · doi:10.2337/dc22-0470 · Cited by 72
“Hybrid closed-loop (HCL) insulin delivery technology automatically increases, decreases, and suspends insulin delivery based on real-time continuous glucose monitoring (CGM) data (5,6).”
Background
Metadata
- Authors
- Fuchs, J., & Hovorka, R.
- Venue
- Expert Review of Medical Devices
- Publisher
- Informa UK Limited
- Published
- 2020-07-02
- Volume
- 17
- Issue
- 7
- Pages
- 707-720
- ISSN
- 1743-4440 (ISSN-L)
- 1745-2422
- OpenAlex
- W3035785430
- Cited by
- 82
- References
- 27
- FWCI
- 8.76
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.