Roadmap toward the 10 ps time-of-flight PET challenge
Open accessThe DOI matches its Crossref record; 81 of the 81 references it cites were found in a registry or an index.
Abstract
Since the seventies, positron emission tomography (PET) has become an invaluable medical molecular imaging modality with an unprecedented sensitivity at the picomolar level, especially for cancer diagnosis and the monitoring of its response to therapy. More recently, its combination with x-ray computed tomography (CT) or magnetic resonance (MR) has added high precision anatomic information in fused PET/CT and PET/MR images, thus compensating for the modest intrinsic spatial resolution of PET. Nevertheless, a number of medical challenges call for further improvements in PET sensitivity. These concern in particular new treatment opportunities in the context personalized (also called precision) medicine, such as the need to dynamically track a small number of cells in cancer immunotherapy or stem cells for tissue repair procedures. A better signal-to-noise ratio (SNR) in the image would allow detecting smaller size tumours together with a better staging of the patients, thus increasing the chances of putting cancer in complete remission. Moreover, there is an increasing demand for reducing the radioactive doses injected to the patients without impairing image quality. There are three ways to improve PET scanner sensitivity: improving detector efficiency, increasing geometrical acceptance of the imaging device and pushing the timing performance of the detectors. Currently, some pre-localization of the electron-positron annihilation along a line-of-response (LOR) given by the detection of a pair of annihilation photons is provided by the detection of the time difference between the two photons, also known as the time-of-flight (TOF) difference of the photons, whose accuracy is given by the coincidence time resolution (CTR). A CTR of about 10 picoseconds FWHM will ultimately allow to obtain a direct 3D volume representation of the activity distribution of a positron emitting radiopharmaceutical, at the millimetre level, thus introducing a quantum leap in PET imaging and quantification and fostering more frequent use of 11 C radiopharmaceuticals. The present roadmap article toward the advent of 10 ps TOF-PET addresses the status and current/future challenges along the development of TOF-PET with the objective to reach this mythic 10 ps frontier that will open the door to real-time volume imaging virtually without tomographic inversion. The medical impact and prospects to achieve this technological revolution from the detection and image reconstruction point-of-views, together with a few perspectives beyond the TOF-PET application are discussed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Cramming More Components Onto Integrated Circuits
Moore, G. E. (1998) · Proceedings of the IEEE · doi:10.1109/jproc.1998.658762 · Cited by 5,926
Properties and potential optoelectronic applications of lead halide perovskite nanocrystals
Kovalenko, M. V., Protesescu, L., & Bodnarchuk, M. I. (2017) · Science · doi:10.1126/science.aam7093 · Cited by 2,652 · cited 1×
“In this regard, a very interesting family of lead halide XPbBr3 perovskites colloidal nano-crystals has emerged due to their high defect tolerance (Kovalenko et al 2017), usually above band gap, and therefore bright luminescence.”
BackgroundLinewidth dependence of radiative exciton lifetimes in quantum wells
Feldmann, J., Peter, G., Göbel, E. O., Dawson, P., Moore, K., Foxon, C., & Elliott, R. J. (1987) · Physical Review Letters · doi:10.1103/physrevlett.59.2337 · Cited by 1,097
Cherry, S. R., Jones, T., Karp, J. S., Qi, J., Moses, W. W., & Badawi, R. D. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.184028 · Cited by 763 · cited 1×
“Extending the axial length of the scanner to 2 m is the goal of the EXPLORER project, proposed and led by UC Davis (https://explorer.ucdavis.edu/, Cherry et al 2017) and funded by the NIH in the USA.”
MethodFirst Human Imaging Studies with the EXPLORER Total-Body PET Scanner*
Badawi, R. D., Shi, H., Hu, P., Chen, S., Xu, T., Price, P. M., Ding, Y., Spencer, B. A., Nardo, L., Liu, W., et al. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.119.226498 · Cited by 722
Timepix3: a 65K channel hybrid pixel readout chip with simultaneous ToA/ToT and sparse readout
Poikela, T., Plosila, J., Westerlund, T., Campbell, M., Gaspari, M. D., Llopart, X., Gromov, V, Kluit, R., Beuzekom, M. V., Zappon, F., et al. (2014) · Journal of Instrumentation · doi:10.1088/1748-0221/9/05/c05013 · Cited by 559
Cryo-CMOS Circuits and Systems for Quantum Computing Applications
Patra, B., Incandela, R. M., van Dijk, J. P. G., Homulle, H. A. R., Song, L., Shahmohammadi, M., Staszewski, R. B., Vladimirescu, A., Babaie, M., Sebastiano, F., et al. (2018) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2017.2737549 · Cited by 551
Self-citationContinuous-wave biexciton lasing at room temperature using solution-processed quantum wells
Grim, J. Q., Christodoulou, S., Di Stasio, F., Krahne, R., Cingolani, R., Manna, L., & Moreels, I. (2014) · Nature Nanotechnology · doi:10.1038/nnano.2014.213 · Cited by 532
Performance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484 · cited 3×
“…DOI measurement achieving a few mm FWHM (Yoshida et al 2017), with total axial fieldof-view reaching 2 m (Badawi et al 2019) or with TOF-PET scintillators and associated electronics for fast coincidences, which amounts a CTR of 215 ps FWHM with a 26.3 cm axial field-of-view (van Sluis et al 2019).”
BackgroundMethodInfluentialCharacterization of PET/CT images using texture analysis: the past, the present… any future?
Hatt, M., Tixier, F., Pierce, L., Kinahan, P. E., Le Rest, C. C., & Visvikis, D. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-016-3427-0 · Cited by 475
Self-citationRecent developments in time-of-flight PET
Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0138-3 · Cited by 368
Self-citationLiquid xenon detectors for particle physics and astrophysics
Aprile, E., & Doke, T. (2010) · Reviews of Modern Physics · doi:10.1103/revmodphys.82.2053 · Cited by 317
The digital silicon photomultiplier — Principle of operation and intrinsic detector performance
Frach, T., Prescher, G., Degenhardt, C., de Gruyter, R., Schmitz, A., & Ballizany, R. (2009) · 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC) · doi:10.1109/nssmic.2009.5402143 · Cited by 317
Pushing the Limits in Time-of-Flight PET Imaging
Lecoq, P. (2017) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2017.2756674 · Cited by 302
Self-citationRowland, C. E., Fedin, I., Zhang, H., Gray, S. K., Govorov, A. O., Talapin, D. V., & Schaller, R. D. (2015) · Nature Materials · doi:10.1038/nmat4231 · Cited by 288
Time-of-flight positron emission tomography: status relative to conventional PET.
Budinger, T. F. (1983) · PubMed · Cited by 283
Dynamic whole-body PET imaging: principles, potentials and applications
Rahmim, A., Lodge, M. A., Karakatsanis, N. A., Panin, V. Y., Zhou, Y., McMillan, A., Cho, S., Zaidi, H., Casey, M. E., & Wahl, R. L. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4153-6 · Cited by 273
Gundacker, S., Martinez Turtos, R., Kratochwil, N., Pots, R. H., Paganoni, M., Lecoq, P., & Auffray, E. (2020) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ab63b4 · Cited by 265
Self-citationSimultaneous Reconstruction of Activity and Attenuation in Time-of-Flight PET
Rezaei, A., Defrise, M., Bal, G., Michel, C., Conti, M., Watson, C., & Nuyts, J. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2212719 · Cited by 261
Self-citationTotal-body imaging: Transforming the role of positron emission tomography
Cherry, S. R., Badawi, R. D., Karp, J. S., Moses, W. W., Price, P., & Jones, T. (2017) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf6169 · Cited by 251
Time-of-flight PET data determine the attenuation sinogram up to a constant
Defrise, M., Rezaei, A., & Nuyts, J. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/4/885 · Cited by 234
Self-citationEffects of $\hbox {Ca}^{2+}$ Co-Doping on the Scintillation Properties of LSO:Ce
Spurrier, M. A., Szupryczynski, P., Yang, K., Carey, A. A., & Melcher, C. L. (2008) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2007.913486 · Cited by 227
A Matheematical Model for Positron-Emission Tomography Systems Having Time-of-Flight Measurements
Snyder, D. L., Thomas, L. J., & Ter-Pogossian, M. M. (1981) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1981.4332168 · Cited by 217
Levin, C. S., Maramraju, S. H., Khalighi, M. M., Deller, T. W., Delso, G., & Jansen, F. (2016) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2016.2537811 · Cited by 198
Online proton therapy monitoring: clinical test of a Silicon-photodetector-based in-beam PET
Ferrero, V., Fiorina, E., Morrocchi, M., Pennazio, F., Baroni, G., Battistoni, G., Belcari, N., Camarlinghi, N., Ciocca, M., Del Guerra, A., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-22325-6 · Cited by 197
Self-citationApoptosis-detecting radioligands: current state of the art and future perspectives
Lahorte, C. M. M., Vanderheyden, J.-L., Steinmetz, N., Van de Wiele, C., Dierckx, R. A., & Slegers, G. (2004) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-004-1555-4 · Cited by 196
Simultaneous reconstruction of activity and attenuation in Time-of-Flight PET
Rezaei, A., Nuyts, J., Defrise, M., Bal, G., Michel, C., Conti, M., & Watson, C. (2011) · 2011 IEEE Nuclear Science Symposium Conference Record · doi:10.1109/nssmic.2011.6153883 · Cited by 185
Self-citationTomitani, T. (1981) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1981.4335769 · Cited by 181
The new opportunities for high time resolution clinical TOF PET
Conti, M., & Bendriem, B. (2019) · Clinical and Translational Imaging · doi:10.1007/s40336-019-00316-5 · Cited by 180
Advances in time-of-flight PET
Surti, S., & Karp, J. S. (2016) · Physica Medica · doi:10.1016/j.ejmp.2015.12.007 · Cited by 173
Self-citationGundacker, S., Auffray, E., Pauwels, K., & Lecoq, P. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/7/2802 · Cited by 173
Self-citationTransparent Ultra-High-Loading Quantum Dot/Polymer Nanocomposite Monolith for Gamma Scintillation
Liu, C., Zhou, L., Hajagos, T. J., Kishpaugh, D., Chen, D. Y., & Pei, Q. (2017) · ACS Nano · doi:10.1021/acsnano.7b02923 · Cited by 168
Francesco, A. D., Bugalho, R., Oliveira, L. B., Pacher, L., Rivetti, A., Rolo, M. D. D. R., Silva, J. C., Silva, R., & Varela, J. (2016) · Journal of Instrumentation · doi:10.1088/1748-0221/11/03/c03042 · Cited by 142
Self-citationLee, M.-J., Ximenes, A. R., Padmanabhan, P., Wang, T.-J., Huang, K.-C., Yamashita, Y., Yaung, D.-N., & Edoardo Charbon, A. (2018) · IEEE Journal of Selected Topics in Quantum Electronics · doi:10.1109/jstqe.2018.2827669 · Cited by 133
PET imaging of synaptic density: A new tool for investigation of neuropsychiatric diseases
Cai, Z., Li, S., Matuskey, D., Nabulsi, N., & Huang, Y. (2019) · Neuroscience Letters · doi:10.1016/j.neulet.2018.07.038 · Cited by 131 · cited 1×
“…diseases could also benefit from ultra-low-dose imaging, such as monitoring response to therapy from major depressive or bi-polar disorders, or even screening for schizophrenia or neurodegenerative diseases using upcoming specific probes (Sloan et al 2019, Hazlett et al 2019, Cai et al 2019).”
BackgroundDesign considerations for a limited angle, dedicated breast, TOF PET scanner
Surti, S., & Karp, J. S. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/11/010 · Cited by 129
Self-citationCates, J. W., Gundacker, S., Auffray, E., Lecoq, P. R., & Levin, C. S. (2018) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aadbcd · Cited by 123
Self-citationAttenuation correction in emission tomography using the emission data—A review
Berker, Y., & Li, Y. (2016) · Medical Physics · doi:10.1118/1.4938264 · Cited by 110
Using convolutional neural networks to estimate time-of-flight from PET detector waveforms
Berg, E., & Cherry, S. R. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa9dc5 · Cited by 108
Meirelles, G. D. S. P., Erdi, Y. E., Nehmeh, S. A., Squire, O. D., Larson, S. M., Humm, J. L., & Schöder, H. M. (2007) · Journal of Nuclear Medicine · doi:10.2967/jnumed.106.038034 · Cited by 103
Advances in digital SiPMs and their application in biomedical imaging
Schaart, D. R., Edoardo Charbon, A., Frach, T., & Schulz, V. (2015) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2015.10.078 · Cited by 101
Can Transient Phenomena Help Improving Time Resolution in Scintillators?
Lecoq, P. R., Korzhik, M., & Vasiliev, A. (2013) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2013.2282232 · Cited by 100
Self-citationOn the use of CdSe scintillating nanoplatelets as time taggers for high-energy gamma detection
Turtos, R. M., Gundacker, S., Omelkov, S. I., Mahler, B., Khan, A. H., Saaring, J., Meng, Z., Васильев, А. Н., Dujardin, C., Kirm, M., et al. (2019) · npj 2D Materials and Applications · doi:10.1038/s41699-019-0120-8 · Cited by 100
Self-citationCoincidence time resolution of 30 ps FWHM using a pair of Cherenkov-radiator-integrated MCP-PMTs
Ota, R., Nakajima, K., Ogawa, I., Tamagawa, Y., Shimoi, H., Suyama, M., & Hasegawa, T. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab0fce · Cited by 97
Comparison of image quality and lesion detection between digital and analog PET/CT
López-Mora, D. A., Flotats, A., Fuentes-Ocampo, F., Camacho, V., Fernández, A., Ruiz, A., Duch, J., Sizova, M., Domènech, A., Estorch, M., et al. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-019-4260-z · Cited by 94 · cited 1×
“Furthermore, the digital system detected more lesions in 27% of the patients with all additional lesions measuring 10 mm (Lopez-Mora et al 2019).”
BackgroundBismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PET
Kwon, S. I., Gola, A. G., Ferri, A., Piemonte, C., & Cherry, S. R. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/18/l38 · Cited by 94
Direct time-of-flight for quantitative, real-time in-beam PET: a concept and feasibility study
Crespo, P., Shakirin, G., Fiedler, F., Enghardt, W., & Wagner, A. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/23/002 · Cited by 91
Organ-Dedicated Molecular Imaging Systems
González, A. J., Sánchez, F., & Benlloch, J. (2018) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2018.2846745 · Cited by 89
Self-citationFast reconstruction of 3D time-of-flight PET data by axial rebinning and transverse mashing
Vandenberghe, S., Daube-Witherspoon, M. E., Lewitt, R. M., & Karp, J. S. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/6/017 · Cited by 87
Self-citationThe Promise of Molecular Imaging in the Study and Treatment of Infectious Diseases
Jain, S. K. (2017) · Molecular Imaging and Biology · doi:10.1007/s11307-017-1055-0 · Cited by 77
Vision 20/20: Positron emission tomography in radiation therapy planning, delivery, and monitoring
Parodi, K. (2015) · Medical Physics · doi:10.1118/1.4935869 · Cited by 73
Self-citationHigher SNR PET image prediction using a deep learning model and MRI image
Liu, C., & Qi, J. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab0dc0 · Cited by 73
Parametric Imaging With PET and SPECT
Gallezot, J., Lu, Y., Naganawa, M., & Carson, R. E. (2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2019.2908633 · Cited by 73
Design study of anin situPET scanner for use in proton beam therapy
Surti, S., Zou, W., Daube-Witherspoon, M. E., McDonough, J. E., & Karp, J. S. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/9/002 · Cited by 63
Self-citationNoise Performance of a Regulated Cascode Transimpedance Amplifier for Radiation Detectors
Oliveira, L. B., Leitao, C. M., & Silva, M. M. (2012) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2011.2180449 · Cited by 60
Beam-on imaging of short-lived positron emitters during proton therapy
Buitenhuis, H. J. T., Diblen, F., Brzeziński, K., Brandenburg, S., & Dendooven, P. G. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa6b8c · Cited by 56
Endocannabinoid signaling in psychiatric disorders: a review of positron emission tomography studies
Sloan, M. E., Grant, C. W., Gowin, J. L., Ramchandani, V. A., & Foll, B. L. (2018) · Acta Pharmacologica Sinica · doi:10.1038/s41401-018-0081-z · Cited by 54
FDG PET radiomics: a review of the methodological aspects
Lovinfosse, P., Visvikis, D., Hustinx, R., & Hatt, M. (2018) · Clinical and Translational Imaging · doi:10.1007/s40336-018-0292-9 · Cited by 51
Self-citationJacobson, O., & Chen, X. S. (2013) · Pharmacological Reviews · doi:10.1124/pr.113.007625 · Cited by 50
Towards time-of-flight PET with a semiconductor detector
Ariño‐Estrada, G., Mitchell, G. S., Kwon, S. I., Du, J., Kim, H., Cirignano, L. J., Shah, K. S., & Cherry, S. R. (2018) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aaaa4e · Cited by 47
Imaging by Injection of Accelerated Radioactive Particle Beams
Llacer, J., Chatterjee, A., Alpen, E. L., Saunders, W., Andreae, S. W., & Jackson, H. C. (1984) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.1984.4307660 · Cited by 44
TOF PET offset calibration from clinical data
Werner, M. E., & Karp, J. S. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/12/4031 · Cited by 43
Time-of-flight positron emission tomography using liquid xenon scintillation
Doke, T., Kikuchi, J., & Nishikido, F. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.07.067 · Cited by 36
Measurements of timing resolution of ultra-fast silicon detectors with the SAMPIC waveform digitizer
Breton, D., Cacqueray, V. D., Delagnes, E., et al. (2016) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2016.08.019 · Cited by 36
Self-citationSparse Detector Configuration in SiPM Digital Photon Counting PET: a Feasibility Study
Zhang, J., Knopp, M. I., & Knopp, M. V. (2018) · Molecular Imaging and Biology · doi:10.1007/s11307-018-1250-7 · Cited by 36
Image-Based Modeling of PSF Deformation With Application to Limited Angle PET Data
Matej, S., Li, Y., Panetta, J. V., Karp, J. S., & Surti, S. (2016) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2016.2607019 · Cited by 28
Self-citationThe Potential Role of Total Body PET Imaging in Assessment of Atherosclerosis
Schmall, J. P., Karp, J. S., & Alavi, A. A. (2019) · PET Clinics · doi:10.1016/j.cpet.2018.12.007 · Cited by 25
Carson, R. E., & Kuo, P. H. (2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2019.2912129 · Cited by 24
Nolet, F., Lemaire, W., Dubois, F., Roy, N., Carrier, S., Samson, A., Charlebois, S. A., Fontaine, R., & Pratte, J. ‐. (2019) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2019.162891 · Cited by 24
Self-citationSPECT/CT Imaging of Mycobacterium tuberculosis Infection with [125I]anti-C3d mAb
Foss, C. A., Kulik, L. N., Ordoñez, A. A., Jain, S. K., Holers, V. M., Thurman, J. M., & Pomper, M. G. (2018) · Molecular Imaging and Biology · doi:10.1007/s11307-018-1228-5 · Cited by 23
New Developments in Imaging Cell-Based Therapy
Martínez, O., Sosabowski, J. K., Maher, J. F., & Papa, S. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.213348 · Cited by 21
Delso, G., Khalighi, M., ter Voert, E. E. G. W., Barbosa, F. D. G., Sekine, T., Hüllner, M. W., & Veit‐Haibach, P. (2016) · Radiology · doi:10.1148/radiol.2016152509 · Cited by 19
Hazlett, E. A., Vaccaro, D. H., Haznedar, M. M., & Goldstein, K. E. (2018) · Psychiatry Research · doi:10.1016/j.psychres.2018.12.030 · Cited by 19
Kuo, F.-W., Babaie, M., Chen, H.-N. R., et al. (2018) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2018.2855972 · Cited by 17
Simulation study to improve the performance of a whole-body PbF 2 Cherenkov TOF-PET scanner
Consuegra, D., Korpar, S., Križan, P., Pestotnik, R., Razdevšek, G., & Dolenec, R. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab6f97 · Cited by 17
Self-citationSystem characteristics of simulated limited angle TOF PET
Vandenberghe, S., & Lemahieu, I. A. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.10.139 · Cited by 14
Self-citationReal Time Artificial Neural Network FPGA Implementation for Triple Coincidences Recovery in PET
Geoffroy, C., Michaud, J.-B., Tétrault, M., Clerk-Lamalice, J., Brunet, C.-A., Lecomte, R., & Fontaine, R. (2015) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2015.2432754 · Cited by 13
High-Resolution Brain 18F-FDG Images Provided by Fully Digital PET
Salvadori, J., Perrin, M., Marie, P., Imbert, L., & Verger, A. (2019) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000002483 · Cited by 11
Consistency equations in native detector coordinates and timing calibration for time-of-flight PET
Li, Y. (2018) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/aaf756 · Cited by 10
Imaging in Neurodegeneration: Movement Disorders
Sossi, V., Cheng, J.-C. K., & Klyuzhin, I. S. (2018) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2018.2871760 · Cited by 6
Dedicated Molecular Imaging Systems for Human Neurological Studies
Hutton, B. F., Yamaya, T., & Furenlid, L. R. (2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2019.2912128 · Cited by 5
Cited by
Esmaeil Zadeh, I., Chang, J., Los, J. W. N., Gyger, S., Elshaari, A. W., Steinhauer, S., Dorenbos, S. N., & Zwiller, V. (2021) · Applied Physics Letters · doi:10.1063/5.0045990 · Cited by 331
Advanced Monte Carlo simulations of emission tomography imaging systems with GATE
Sarrut, D., Bała, M., Bardiès, M., Bert, J., Chauvin, M., Chatzipapas, K., Dupont, M., Etxebeste, A., M Fanchon, L., Jan, S., et al. (2021) · Physics in Medicine & Biology · doi:10.1088/1361-6560/abf276 · Cited by 204
Self-citationPositronium imaging with the novel multiphoton PET scanner
Moskal, P., Dulski, K., Chug, N., Curceanu, C., Czerwiński, E., Dadgar, M., Gajewski, J., Gajos, A., Grudzień, G., Hiesmayr, B. C., et al. (2021) · Science Advances · doi:10.1126/sciadv.abh4394 · Cited by 180
Update on latest advances in time-of-flight PET
Surti, S., & Karp, J. S. (2020) · Physica Medica · doi:10.1016/j.ejmp.2020.10.031 · Cited by 96
“50 ps, and approach values closer to 10 ps [22].”
BackgroundSelf-citation
Metadata
- Authors
- Lecoq, P., Morel, C., Prior, J. O., Visvikis, D., Gundacker, S., Auffray, E., Križan, P., Turtos, R. M., Thers, D., Charbon, E., Varela, J., de La Taille, C., Rivetti, A., Breton, D., Pratte, J.-F., Nuyts, J., Surti, S., Vandenberghe, S., Marsden, P., Parodi, K., Benlloch, J. M., & Benoit, M.
- Venue
- Physics in Medicine & Biology
- Publisher
- IOP Publishing
- Published
- 2020-10-20
- Volume
- 65
- Issue
- 21
- Pages
- 21RM01
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W3027275779
- Cited by
- 258
- References
- 81
- FWCI
- 15.35
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.