The influence of adipose tissue on spatially resolved near-infrared spectroscopy derived skeletal muscle oxygenation: the extent of the problem
Open accessThe DOI matches its Crossref record; 46 of the 46 references it cites were found in a registry or an index.
Abstract
Objective . Near-infrared spectroscopy (NIRS) measurements of tissue oxygen saturation (StO 2 ) are useful for the assessment of skeletal muscle perfusion and function during exercise, however, they are influenced by overlying skin and adipose tissue. This study explored the extent and nature of the influence of adipose tissue thickness (ATT) on StO 2 . Approach . NIR spatially resolved spectroscopy (SRS) derived oxygenation was measured on vastus lateralis in 56 patients with chronic heart failure (CHF) and 20 healthy control (HC) subjects during rest and moderate intensity exercise with simultaneous assessment of oxygen uptake kinetics ( τ ). In vitro measurements were performed on a flow cell with a blood mixture with full oxygen saturation (100%), which was gradually decreased to 0% by adding sodium metabisulfite. Experiments were repeated with 2 mm increments of porcine fat layer between the NIRS device and flow cell up to 14 mm. Main results . Lower ATT, higher τ , and CHF were independently associated with lower in vivo StO 2 in multiple regression analysis, whereas age and gender showed no independent relationship. With greater ATT, in vitro StO 2 was reduced from 100% to 74% for fully oxygenated blood and increased from 0% to 68% for deoxygenated blood. Significance . This study shows that ATT independently confounds NIR-SRS derived StO 2 by overestimating actual skeletal muscle oxygenation and by decreasing its sensitivity for deoxygenation. Because physiological properties (e.g. presence of disease and slowing of τ ) also influence NIR-SRS, a correction based on optical properties is needed to interpret calculated values as absolute StO 2 .
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 Relation between adipose tissue thickness (ATT) and (a) resting tissue saturation index (TSIbaseline), (b) TSI at the end of moderate-intensity constant work rate exercise (TSIend-exercise), (c) response of TSI to moderate intensity exercise ( Δ TSIexercise), and (d) response of deoxygenated hemoglobin concentration to moderate intensity exercise ( Δ [HHb]exercise) for healthy control (HC) subjects and chronic heart failure (CHF) patients.](https://figures.cochat.ai/paper-figures/e29cf078739e468cb4329b7dede794ab/the-influence-of-adipose-tissue-on-spatially-resolved-near-infrared-spectroscopy-figure-2.png)


Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, some studies explored other regions during cycling which could affect the determination of the threshold17,47, although the rectus femoris is also a power output producer in this area where there could be a higher proportion of adipose tissue52 or because its neuromuscular activation is not affected by the increase in workload during the test (e.g., gastrocnemius)53.”
From Reliability of threshold determination using portable muscle oxygenation monitors during exercise testing: a systematic review and meta-analysis (2023) · cites this paper as [52]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A new method for detecting anaerobic threshold by gas exchange
Beaver, W. L., Wasserman, K., & Whipp, B. J. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.60.6.2020 · Cited by 4,103 · cited 1×
“The gas exchange threshold (GET) was determined by the mean of the independent assessments of two experienced physicians using the V-slope method (Beaver et al 1986).”
Fletcher, G. F., Ades, P. A., Kligfield, P., Arena, R., Balady, G. J., Bittner, V. A., Coke, L. A., Fleg, J. L., Forman, D. E., Gerber, T. C., et al. (2013) · Circulation · doi:10.1161/cir.0b013e31829b5b44 · Cited by 2,235 · cited 1×
“The maximal exercise protocol started with a 1 min resting period and 4 min of unloaded pedaling before work rate started to increase with an individualized ramp rate aiming to reach exhaustion within 8 – 12 min (Fletcher et al 2013).”
Poole, D. C., & Jones, A. M. (2012) · Comprehensive Physiology · doi:10.1002/cphy.c100072 · Cited by 514 · cited 1×
“…muscle regions found by Koga et al (2015a), a greater proportion of oxidative muscle fibers in subjects with faster oxygen uptake kinetics can also explain the relation between the degree of measured deoxygenation and speed of pulmonary oxygen uptake kinetics in this study (Poole and Jones 2012).”
Anthropometric measurements in the elderly: age and gender differences
Perissinotto, E., Pisent, C., Sergi, G., Grigoletto, F., Enzi, G., & ILSA Working Group (2002) · British Journal of Nutrition · doi:10.1079/bjn2001487 · Cited by 462 · cited 1×
“However, all of these StO 2 -related parameters may in turn be associated with ATT, which has previously been demonstrated for age and gender (Perissinotto et al 2002).”
Ferrari, M., Muthalib, M., & Quaresima, V. (2011) · Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences · doi:10.1098/rsta.2011.0230 · Cited by 450 · cited 2×
“…(TD) and frequency-domain (FD) instruments that allow quantification of the actual tissue optical properties to calculate absolute oxygenated and deoxygenated hemoglobin concentrations, the signal-to-noise ratio and temporal resolution are not as high as for the SRS technique (Ferrari et al 2011).”
Adipose tissue thickness affects in vivo quantitative near-IR spectroscopy in human skeletal muscle
VAN BEEKVELT, M. C. P., BORGHUIS, M. S., VAN ENGELEN, B. G. M., WEVERS, R. A., & COLIER, W. N. J. M. (2001) · Clinical Science · doi:10.1042/cs1010021 · Cited by 369 · cited 4×
“…al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt et al 2001, Koga et al 2011, Ohmae et al 2014, van der Zwaard et al 2016).”
InfluentialGrassi, B., & Quaresima, V. (2016) · Journal of Biomedical Optics · doi:10.1117/1.jbo.21.9.091313 · Cited by 345 · cited 1×
“Deeper muscle regions commonly have a greater proportion of oxidative fibers compared to superficial regions, leading to a greater degree of oxidative energy metabolism and a concomitantly greater sensitivity for vascular control (Grassi and Quaresima 2016).”
BackgroundGoossens, G. H., Bizzarri, A., Venteclef, N., Essers, Y., Cleutjens, J. P., Konings, E., Jocken, J. W. E., Čajlaković, M., Ribitsch, V., Clément, K., et al. (2011) · Circulation · doi:10.1161/circulationaha.111.027813 · Cited by 324 · cited 1×
“Second, Goossens et al (2011) demonstrated that subjects with greater ATT have even higher hemoglobin oxygenation in adipose tissue than lean subjects, at least in fasting conditions, amplifying the aforementioned effect.”
Ryan, T. E., Erickson, M. L., Brizendine, J. T., Young, H.-J., & McCully, K. K. (2012) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00319.2012 · Cited by 233 · cited 1×
“Conventional methods to correct for ATT are based on the assumption of constant tissue properties among measured subjects (e.g. hemoglobin content, oxygen consumption, and oxygen saturation) (Niwayama et al 2000, Ryan et al 2012, Bowen et al 2013).”
Richardson, T. E., Kindig, C. A., Musch, T. I., & Poole, D. C. (2003) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00308.2003 · Cited by 131 · cited 1×
“…present results of a lower TSI baseline for CHF, and an independent association of CHF with StO 2 , may be explained on the basis of pathophysiological consequences of the disease, as for example reduced skin blood flow (Andreassen et al 1998), and compromised vascularity (Richardson et al 2003).”
Homma, S., Fukunaga, T., & Kagaya, A. (1996) · Journal of Biomedical Optics · doi:10.1117/12.252417 · Cited by 123 · cited 1×
“…distinctly increasing resting oxygenation values (Geraskin et al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt…”
Dynamics of muscle microcirculatory and blood–myocyte O2flux during contractions
Poole, D. C., Copp, S. W., Hirai, D. M., & Musch, T. I. (2011) · Acta Physiologica · doi:10.1111/j.1748-1716.2010.02246.x · Cited by 105 · cited 1×
“…fact that resting oxygenation was lower in their CHF patients (57.2 ± 7.2% versus 67.5 ± 4.9% in our study) could have led to a greater sensitivity of oxygen uptake kinetics to resting oxygenation, slowing oxygen uptake kinetics in a more O 2 availability/ delivery dependent way (Poole et al 2011).”
Influentialvan der Zwaard, S., Jaspers, R. T., Blokland, I. J., Achterberg, C., Visser, J. M., den Uil, A. R., Hofmijster, M. J., Levels, K., Noordhof, D. A., de Haan, A., et al. (2016) · PLOS ONE · doi:10.1371/journal.pone.0162914 · Cited by 95 · cited 1×
“…al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt et al 2001, Koga et al 2011, Ohmae et al 2014, van der Zwaard et al 2016).”
Niwayama, M., Lin, L., Shao, J., Kudo, N., & Yamamoto, K. (2000) · Review of Scientific Instruments · doi:10.1063/1.1322578 · Cited by 94 · cited 4×
“…al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt et al 2001, Koga et al 2011, Ohmae et al 2014, van der Zwaard et al 2016).”
InfluentialKoga, S., Poole, D. C., Fukuoka, Y., Ferreira, L. F., Kondo, N., Ohmae, E., & Barstow, T. J. (2011) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00101.2011 · Cited by 92 · cited 2×
“…al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt et al 2001, Koga et al 2011, Ohmae et al 2014, van der Zwaard et al 2016).”
InfluentialBowen, T. S., Rossiter, H. B., Benson, A. P., Amano, T., Kondo, N., Kowalchuk, J. M., & Koga, S. (2013) · Experimental Physiology · doi:10.1113/expphysiol.2013.073270 · Cited by 66 · cited 1×
“Conventional methods to correct for ATT are based on the assumption of constant tissue properties among measured subjects (e.g. hemoglobin content, oxygen consumption, and oxygen saturation) (Niwayama et al 2000, Ryan et al 2012, Bowen et al 2013).”
Sperandio, P. A., Borghi-Silva, A., Barroco, A., Nery, L. E., Almeida, D. R., & Neder, J. A. (2009) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00596.2009 · Cited by 66 · cited 1×
“…can be used in the assessment of the contribution of O 2 delivery and utilization to limitations of oxidative metabolism in skeletal muscle during exercise (Ferrari et al 2011), for instance in patients with chronic heart failure (CHF) (Sperandio et al 2009, Bowen et al 2012, Niemeijer et al 2016).”
Tissue hemoglobin index: a non-invasive optical measure of total tissue hemoglobin
Myers, D. E., McGraw, M. K., George, M., Mulier, K. E., & Beilman, G. J. (2009) · Critical Care · doi:10.1186/cc8000 · Cited by 66 · cited 3×
“Moreover, tissue composition may be variable between subjects due to differences in hemoglobin content (Myers et al 2009, Adami et al 2015).”
Messere, A., & Roatta, S. (2013) · Physiological Reports · doi:10.1002/phy2.179 · Cited by 63 · cited 2×
“Messere and Roatta (2013) reported that MBL-based blood volume indices (i.e. total hemoglobin) were more affected by cutaneous warming than those based on SRS (i.e. tissue hemoglobin index), while sensitivity to exercise-induced changes of blood volume was similar between methodologies.”
Koga, S., Barstow, T. J., Okushima, D., Rossiter, H. B., Kondo, N., Ohmae, E., & Poole, D. C. (2015) · Journal of Applied Physiology · doi:10.1152/japplphysiol.01003.2014 · Cited by 60 · cited 3×
“However, recent work by Koga et al (2015b) using NIR-TRS revealed µ ′ s varies during exercise and whole body heating.”
Kemps, H. M. C., De Vries, W. R., Hoogeveen, A. R., Zonderland, M. L., Thijssen, E. J. M., & Schep, G. (2007) · European Journal of Applied Physiology · doi:10.1007/s00421-007-0398-7 · Cited by 54 · cited 3×
“Gas exchange parameters for the constant work rate exercise test were averaged into 10 s sampling intervals after removal of outliers (values > 3 SDs from the local mean were omitted) (Kemps et al 2007).”
InfluentialSelf-citationFeng, W., Haishu, D., Fenghua, T., Jun, Z., Qing, X., & Xianwu, T. (2001) · Physiological Measurement · doi:10.1088/0967-3334/22/1/323 · Cited by 53 · cited 1×
“…al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt et al 2001, Koga et al 2011, Ohmae et al 2014, van der Zwaard et al 2016).”
Hunter, R. J., Patterson, M. S., Farrell, T. J., & Hayward, J. E. (2002) · Physics in Medicine and Biology · doi:10.1088/0031-9155/47/2/302 · Cited by 48 · cited 1×
“…an intralipid emulsion with a concentration of 20% were added to produce a 300 ml mixture with an intralipid concentration of 1%, and optical properties approximating absorption and reduced scattering coefficients of perfused muscle tissue (Lin et al 2000, Hunter et al 2002, Di Ninni et al 2012).”
A validated model of oxygen uptake and circulatory dynamic interactions at exercise onset in humans
Benson, A. P., Grassi, B., & Rossiter, H. B. (2013) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00184.2013 · Cited by 47 · cited 1×
“The first 20 s of the V ˙ O 2 response after onset were omitted from kinetics analysis, since it is assumed that the rapid increase of V ˙ O 2 during this period (cardiodynamic phase or phase I) represents increased pulmonary blood flow rather than microvascular gas exchange (Benson et al 2013).”
Yang, Y., Soyemi, O. O., Landry, M. R., & Soller, B. R. (2005) · Optics Express · doi:10.1364/opex.13.001570 · Cited by 47 · cited 1×
“In contrast to suggestions by other authors (Yang et al 2005), the in vitro experiments show that even a minimal ATT of 1 mm is expected to influence StO 2 measurements to a great degree.”
Abozguia, K., Phan, T. T., Shivu, G. N., Maher, A. R., Ahmed, I., Wagenmakers, A. J. M., & Frenneaux, M. P. (2008) · European Journal of Heart Failure · doi:10.1016/j.ejheart.2008.05.009 · Cited by 44 · cited 1×
“…has been found for considerable intersubject variability of tissue hemoglobin content (Myers et al 2009, Adami et al 2015), for reduced skeletal muscle oxygen consumption in CHF (Abozguia et al 2008), and for a relation between resting StO 2 and oxygen uptake kinetics in CHF (Bowen et al 2012).”
Peltonen, J. E., Koponen, A. S., Pullinen, K., Hägglund, H., Aho, J. M., Kyröläinen, H., & Tikkanen, H. O. (2012) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2012.04.002 · Cited by 40 · cited 1×
“Furthermore, they may reflect pathophysiological changes related to certain diseases (e.g. CHF, diabetes, etc) (Bowen et al 2012, Peltonen et al 2012), and their subsequent consequences on exercise capacity (e.g. slowed oxygen uptake kinetics).”
Koga, S., Poole, D. C., Kondo, N., Oue, A., Ohmae, E., & Barstow, T. J. (2015) · European Journal of Applied Physiology · doi:10.1007/s00421-014-3019-2 · Cited by 39 · cited 3×
“However, recent work by Koga et al (2015b) using NIR-TRS revealed µ ′ s varies during exercise and whole body heating.”
Bowen, T. S., Cannon, D. T., Murgatroyd, S. R., Birch, K. M., Witte, K. K., & Rossiter, H. B. (2012) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00779.2011 · Cited by 39 · cited 5×
“However, while there was a relation between TSI end-exercise and oxygen kinetics, the relation between resting StO 2 and τ V ˙ O onset 2 reported by Bowen et al (2012) in their group of CHF patients could not be confirmed.”
InfluentialSpee, R. F., Niemeijer, V. M., Wijn, P. F., Doevendans, P. A., & Kemps, H. M. (2016) · European Journal of Preventive Cardiology · doi:10.1177/2047487316661615 · Cited by 37
Self-citationFat emulsions as diffusive reference standards for tissue simulating phantoms?
Ninni, P. D., Bérubé-Lauzière, Y., Mercatelli, L., Sani, E., & Martelli, F. (2012) · Applied Optics · doi:10.1364/ao.51.007176 · Cited by 37 · cited 1×
“…an intralipid emulsion with a concentration of 20% were added to produce a 300 ml mixture with an intralipid concentration of 1%, and optical properties approximating absorption and reduced scattering coefficients of perfused muscle tissue (Lin et al 2000, Hunter et al 2002, Di Ninni et al 2012).”
Niemeijer, V. M., Spee, R. F., Jansen, J. P., Buskermolen, A. B. C., van Dijk, T., Wijn, P. F. F., & Kemps, H. M. C. (2017) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12269 · Cited by 33 · cited 1×
“Absolute values of TSI during moderate-intensity exercise testing were calculated as the average of the last minute of the resting phase (TSI baseline ), and the average of the last minute of the exercise phase (TSI end-exercise ), as described previously (Niemeijer et al 2017).”
Self-citationGeraskin, D., Boeth, H., & Kohl‐Bareis, M. (2009) · Journal of Biomedical Optics · doi:10.1117/1.3184425 · Cited by 33 · cited 5×
“Indeed, the thickness of the adipose tissue layer has been shown to confound NIRS measurements, distinctly increasing resting oxygenation values (Geraskin et al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for…”
InfluentialOhmae, E., Nishio, S., Oda, M., Suzuki, H., Suzuki, T., Ohashi, K., Koga, S., Yamashita, Y., & Watanabe, H. (2014) · Journal of Biomedical Optics · doi:10.1117/1.jbo.19.6.067005 · Cited by 33 · cited 2×
“…al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT (Niwayama et al 2000, Feng et al 2001, van Beekvelt et al 2001, Koga et al 2011, Ohmae et al 2014, van der Zwaard et al 2016).”
Peltonen, J. E., Hägglund, H., Koskela-Koivisto, T., Koponen, A. S., Aho, J. M., Rissanen, A., Shoemaker, J. K., Tiitinen, A. E., & Tikkanen, H. O. (2013) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2013.05.014 · Cited by 32 · cited 1×
“The latter associations are considered to mirror physiological differences between, for example, younger and older subjects (Takagi et al 2013), and males and females (Cooper et al 2010, Peltonen et al 2013, Takagi et al 2016).”
Costes, F., Denis, C., Roche, F., Prieur, F., ENJOLRAS, F., & Barthélemy, J. (1999) · Archives of Physiology and Biochemistry · doi:10.1076/apab.107.2.159.4343 · Cited by 30 · cited 3×
“Furthermore, measurements performed by Costes et al (1999) suggest a relation between aging and resting oxygenation.”
Niwayama, M., Suzuki, H., Yamashita, T., & Yasuda, Y. (2012) · The Journal of Japanese College of Angiology · doi:10.7133/jca.52.211 · Cited by 28 · cited 2×
“However, this is deemed an unsatisfactory approach when dealing with heterogeneous groups of subjects (van Beekvelt et al 2001, Niwayama et al 2012), as for example CHF patients.”
Kime, R., Fujioka, M., Osawa, T., Takagi, S., Niwayama, M., Kaneko, Y., Osada, T., Murase, N., & Katsumura, T. (2013) · Advances in experimental medicine and biology · doi:10.1007/978-1-4614-7411-1_23 · Cited by 28 · cited 2×
“While this observation indicates that TSI is more sensitive to skeletal muscle deoxygenation than [HHb], it simultaneously demonstrates a lower sensitivity of TSI for increases of skin and adipose tissue oxygenation due to skin blood flow changes (Kime et al 2013).”
Adami, A., Koga, S., Kondo, N., Cannon, D. T., Kowalchuk, J. M., Amano, T., & Rossiter, H. B. (2015) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00918.2014 · Cited by 27 · cited 2×
“Moreover, tissue composition may be variable between subjects due to differences in hemoglobin content (Myers et al 2009, Adami et al 2015).”
Shiga, T., Lin, L., Takahashi, M., Yamamoto, K., Niwayama, M., & Kudo, N. (2000) · Frontiers of Medical & Biological Engineering · doi:10.1163/15685570052061531 · Cited by 27 · cited 3×
“…an intralipid emulsion with a concentration of 20% were added to produce a 300 ml mixture with an intralipid concentration of 1%, and optical properties approximating absorption and reduced scattering coefficients of perfused muscle tissue (Lin et al 2000, Hunter et al 2002, Di Ninni et al 2012).”
Cooper, C. E., Penfold, S.-M., Elwell, C. E., & Angus, C. (2010) · Advances in Experimental Medicine and Biology · doi:10.1007/978-1-4419-1241-1_25 · Cited by 21 · cited 5×
“While SRS-NIR derived StO 2 is confounded by ATT, it is also influenced by age, gender, disease state and oxygen uptake kinetics (Costes et al 1999, Geraskin et al 2009, Cooper et al 2010, Bowen et al 2012, Grieger et al 2013).”
InfluentialNiemeijer, V. M., Spee, R. F., Schoots, T., Wijn, P. F. F., & Kemps, H. (2016) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00474.2016 · Cited by 18 · cited 1×
“…can be used in the assessment of the contribution of O 2 delivery and utilization to limitations of oxidative metabolism in skeletal muscle during exercise (Ferrari et al 2011), for instance in patients with chronic heart failure (CHF) (Sperandio et al 2009, Bowen et al 2012, Niemeijer et al 2016).”
Self-citationAndreassen, A. K., Kirkebøen, K. A., Gullestad, L. L., Simonsen, S., & Kvernebo, K. (1998) · International Journal of Cardiology · doi:10.1016/s0167-5273(98)00103-x · Cited by 18
Analysis of NIRS-Based Muscle Oxygenation Parameters by Inclusion of Adipose Tissue Thickness
Grieger, S., Geraskin, D., Steimers, A., & Kohl‐Bareis, M. (2013) · Advances in experimental medicine and biology · doi:10.1007/978-1-4614-7411-1_18 · Cited by 17 · cited 4×
“…tissue layer has been shown to confound NIRS measurements, distinctly increasing resting oxygenation values (Geraskin et al 2009, Cooper et al 2010, Grieger et al 2013), reducing optical density (Homma et al 1996, Lin et al 2000), and decreasing sensitivity for deoxygenation with greater ATT…”
InfluentialAging Affects Spatial Distribution of Leg Muscle Oxygen Saturation During Ramp Cycling Exercise
Takagi, S., Kime, R., Murase, N., Watanabe, T., Osada, T., Niwayama, M., & Katsumura, T. (2013) · Advances in experimental medicine and biology · doi:10.1007/978-1-4614-7411-1_22 · Cited by 9 · cited 1×
“The latter associations are considered to mirror physiological differences between, for example, younger and older subjects (Takagi et al 2013), and males and females (Cooper et al 2010, Peltonen et al 2013, Takagi et al 2016).”
Takagi, S., Kime, R., Niwayama, M., Osada, T., Murase, N., Sakamoto, S., & Katsumura, T. (2016) · Advances in experimental medicine and biology · doi:10.1007/978-1-4939-3023-4_7 · Cited by 4 · cited 1×
“The latter associations are considered to mirror physiological differences between, for example, younger and older subjects (Takagi et al 2013), and males and females (Cooper et al 2010, Peltonen et al 2013, Takagi et al 2016).”
Cited by
Muscle Oximetry in Sports Science: A Systematic Review
Perrey, S., & Ferrari, M. (2018) · Sports Medicine · doi:10.1007/s40279-017-0820-1 · Cited by 220
“[86] confirmed that adipose tissue thickness independently confounds by either overestimating actual skeletal muscle oxygenation or decreasing its sensitivity for deoxygenation.”
BackgroundHamaoka, T., & McCully, K. K. (2019) · The Journal of Physiological Sciences · doi:10.1007/s12576-019-00697-2 · Cited by 92
Lutz, J., Memmert, D., Raabe, D., Dornberger, R., & Donath, L. (2019) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17010059 · Cited by 114
“Another issue is the adipose tissue thickness influences the reflection of infrared lights [83].”
BackgroundSendra-Pérez, C., Sanchez-Jimenez, J. L., Marzano-Felisatti, J. M., Encarnación-Martínez, A., Salvador-Palmer, R., & Priego-Quesada, J. I. (2023) · Scientific Reports · doi:10.1038/s41598-023-39651-z · Cited by 52
Keller, J. L., & Kennedy, K. G. (2021) · European Journal of Applied Physiology · doi:10.1007/s00421-021-04810-5 · Cited by 36
Metadata
- Authors
- Niemeijer, V. M., Jansen, J. P., van Dijk, T., Spee, R. F., Meijer, E. J., Kemps, H. M. C., & Wijn, P. F. F.
- Affiliations
- Máxima Medisch Centrum (NL)
- Eindhoven University of Technology (NL)
- Corresponding author
- Niemeijer, Victor M
- Meijer, Eduard J
- Venue
- Physiological Measurement
- Publisher
- IOP Publishing
- Published
- 2017-02-16
- Volume
- 38
- Issue
- 3
- Pages
- 539-554
- ISSN
- 0967-3334 (ISSN-L)
- 1361-6579
- OpenAlex
- W2584254667
- Semantic Scholar
- 7dc7f6574734882068cf9811e33e1573501bbe1b
- Cited by
- 61
- References
- 46
- Citations by year
- 2026: 9 · 2025: 7 · 2024: 7 · 2023: 10 · 2022: 12 · 2021: 3 · 2020: 2 · 2019: 6 · 2018: 3 · 2017: 2
- FWCI
- 2.74
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- creativecommons.org/licenses/by/3.0/ (vor)
- Locations
- Physiological Measurement (publishedVersion, cc-by)
- PubMed (publishedVersion)
- TU/e Research Portal (submittedVersion, other-oa)
- TU/e Research Portal (submittedVersion)
- Funders
- Foundation ‘Friends of the Heart’, Eindhoven, The Netherlands
- Scientific Foundation of Máxima Medical Centre, Veldhoven, The Netherlands
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Keywords
- adipose tissue thickness, skeletal muscle oxygenation, tissue oxygen saturation, vastus lateralis, chronic heart failure
- MeSH terms
- Adipose Tissue, Adult, Aged, Aged, 80 and over, Chronic Disease, Female, Heart Failure, Humans, Male, Middle Aged, Oxygen, Case-Control Studies, Muscle, Skeletal, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.