Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Open accessAbstract
OBJECTIVE: Brain-computer interfaces (BCIs) using chronically implanted intracortical microelectrode arrays (MEAs) have the potential to restore lost function to people with disabilities if they work reliably for years. Current sensors fail to provide reliably useful signals over extended periods of time for reasons that are not clear. This study reports a comprehensive retrospective analysis from a large set of implants of a single type of intracortical MEA in a single species, with a common set of measures in order to evaluate failure modes. APPROACH: Since 1996, 78 silicon MEAs were implanted in 27 monkeys (Macaca mulatta). We used two approaches to find reasons for sensor failure. First, we classified the time course leading up to complete recording failure as acute (abrupt) or chronic (progressive). Second, we evaluated the quality of electrode recordings over time based on signal features and electrode impedance. Failure modes were divided into four categories: biological, material, mechanical, and unknown. MAIN RESULTS: Recording duration ranged from 0 to 2104 days (5.75 years), with a mean of 387 days and a median of 182 days (n = 78). Sixty-two arrays failed completely with a mean time to failure of 332 days (median = 133 days) while nine array experiments were electively terminated for experimental reasons (mean = 486 days). Seven remained active at the close of this study (mean = 753 days). Most failures (56%) occurred within a year of implantation, with acute mechanical failures the most common class (48%), largely because of connector issues (83%). Among grossly observable biological failures (24%), a progressive meningeal reaction that separated the array from the parenchyma was most prevalent (14.5%). In the absence of acute interruptions, electrode recordings showed a slow progressive decline in spike amplitude, noise amplitude, and number of viable channels that predicts complete signal loss by about eight years. Impedance measurements showed systematic early increases, which did not appear to affect recording quality, followed by a slow decline over years. The combination of slowly falling impedance and signal quality in these arrays indicates that insulating material failure is the most significant factor. SIGNIFICANCE: This is the first long-term failure mode analysis of an emerging BCI technology in a large series of non-human primates. The classification system introduced here may be used to standardize how neuroprosthetic failure modes are evaluated. The results demonstrate the potential for these arrays to record for many years, but achieving reliable sensors will require replacing connectors with implantable wireless systems, controlling the meningeal reaction, and improving insulation materials. These results will focus future research in order to create clinical neuroprosthetic sensors, as well as valuable research tools, that are able to safely provide reliable neural signals for over a decade.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Previous demonstrations of intracortical insertion of MEAs showed injury and neuronal loss at the implantation location99, as well as signal instabilities contributing to signal loss and performance decline over time100,101.”
From The speech neuroprosthesis (2024) · cites this paper as [100]
In the citing paper
“In spite of the formidable results of using UEA/USEA technologies described above, most penetrating microelectrode arrays currently have maximum in vivo lifetimes of several months or a few years (Suner et al 2005, Prasad et al 2012, Barrese et al 2013).”
From Clinical applications of penetrating neural interfaces and Utah Electrode Array technologies (2016) · cites this paper as Barrese et al 2013
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Neuronal ensemble control of prosthetic devices by a human with tetraplegia
Hochberg, L. R., Serruya, M. D., Friehs, G. M., Mukand, J. A., Saleh, M., Caplan, A. H., Branner, A., Chen, D., Penn, R. D., & Donoghue, J. P. (2006) · Nature · doi:10.1038/nature04970 · Cited by 3,295
Self-citationReach and grasp by people with tetraplegia using a neurally controlled robotic arm
Hochberg, L. R., Bacher, D., Jarosiewicz, B., Masse, N. Y., Simeral, J. D., Vogel, J., Haddadin, S., Liu, J., Cash, S. S., van der Smagt, P., et al. (2012) · Nature · doi:10.1038/nature11076 · Cited by 2,719
Self-citationResponse of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896
Biran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893
Brain responses to micro-machined silicon devices
Szarowski, D. H., Andersen, M. D., Retterer, S. T., Spence, A. J., Isaacson, M. S., Craighead, H. G., Turner, J. N., & Shain, W. G. (2003) · Brain Research · doi:10.1016/s0006-8993(03)03023-3 · Cited by 811
Chronic recording capability of the Utah Intracortical Electrode Array in cat sensory cortex
Rousche, P. J., & Normann, R. A. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00031-4 · Cited by 646
Cerebral Astrocyte Response to Micromachined Silicon Implants
Turner, J. N., Shain, W. G., Szarowski, D. H., Andersen, M., Martins, S., Isaacson, M. S., & Craighead, H. G. (1999) · Experimental Neurology · doi:10.1006/exnr.1998.6983 · Cited by 603
Single-neuron dynamics in human focal epilepsy
Truccolo, W., Donoghue, J., Hochberg, L. R., Eskandar, E. N., Madsen, J. R., Anderson, W. S., Brown, E. N., Halgren, E., & Cash, S. S. (2011) · Nature Neuroscience · doi:10.1038/nn.2782 · Cited by 528
Self-citationFactors influencing the biocompatibility of insertable silicon microshafts in cerebral cortex
Edell, D. J., Toi, V. V., McNeil, V. M., & Clark, L. D. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.141202 · Cited by 481
McConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480
Research works printed but not linked: 2
- McConnell G C, 2009, J. Neural Eng., 6
- Prasad A, 2012, J. Neural Eng., 9
Cited by
Sorted by citations
Items 401–500 of 599
Khajehzadeh, M., Geramifard, N., Abbott, J., Wu, Y., Kumara, S. T., Thai, T. T. D., Hernandez-Reynoso, A. G., Joshi‐Imre, A., Blum, K. P., Miller, L. E., et al. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/addb79 · Cited by 7
Yang, S., Wang, H., Lo, Y., Lai, H.-Y., Chen, K., Lan, Y.-H., Kao, C.-C., Chou, C., Lin, S.-H., Huang, J.-W., et al. (2020) · Frontiers in Computational Neuroscience · doi:10.3389/fncom.2020.00022 · Cited by 6
Kim, Y., Mueller, N. N., Schwartzman, W. E., Sarno, D., Wynder, R., Hoeferlin, G. F., Gisser, K., Capadona, J. R., & Hess‐Dunning, A. (2023) · Micromachines · doi:10.3390/mi14051015 · Cited by 6
Drakopoulou, S., Varkevisser, F., Sohail, L., Aqamolaei, M., Costa, T. L., & Spyropoulos, G. D. (2023) · Frontiers in Electronics · doi:10.3389/felec.2023.1250655 · Cited by 8
Song, S., Druschel, L. N., Kasthuri, N. M., Wang, J., Conard, J. H., Chan, E. R., Acharya, A. P., & Capadona, J. R. (2024) · Scientific Reports · doi:10.1038/s41598-024-68017-2 · Cited by 6
A mobile embedded platform for high performance neural signal computation and communication
Heelan, C., Komar, J., Vargas-Irwin, C. E., Simeral, J. D., & Nurmikko, A. V. (2015) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2015.7348356 · Cited by 6
Self-citationMicrofabrication, assembly, and hermetic packaging of mm-sized free-floating neural probes
Yeon, P., Gonzalez, J. L., Zia, M., Rajan, S. K., May, G. S., Bakir, M. S., & Ghovanloo, M. (2017) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2017.8325214 · Cited by 6
Insulation of Carbon Nanotube Yarn Electrodes for Intrafascicular Neural Stimulation and Recording
Su, J., Zhang, X., Li, M. N., Gao, T., Wang, R., Chai, X. Y., Zhang, D. G., Zhang, X. H., & Sui, X. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8716984 · Cited by 6
Evaluation of Pneumatic Insertion Stability of Utah Slanted Electrode Arrays in Rat Sciatic Nerve
Thomas, W. M., Leber, M., Crew, J., & Warren, D. J. (2022) · 2022 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871237 · Cited by 5
Magnetic Implantable Devices and Materials for the Brain
Chen, X., Wu, D., Li, K., & Han, M. (2025) · Small Methods · doi:10.1002/smtd.202501460 · Cited by 5
Dhawan, V., Wu, B., Narayanan, S., Shi, D., Patel, Y., & Cui, X. T. (2025) · Biomaterials · doi:10.1016/j.biomaterials.2025.123929 · Cited by 6
Li, J., Fan, K., Ma, X., Wang, C., Min, J., Chen, Y., Song, Y., Xu, Y., & Gao, W. F. (2026) · Nature Materials · doi:10.1038/s41563-026-02624-4 · Cited by 5
Cortelli, G., Cramer, T., Patruno, L., Fraboni, B., & de Miranda, S. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202300931 · Cited by 6
High-density 3D pyramid-shaped microelectrode arrays for brain-machine interface applications
Ghane-Motlagh, B., & Sawan, M. (2014) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2014.6981738 · Cited by 5
Gueli, C., Martens, J., Eickenscheidt, M., & Stieglitz, T. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630832 · Cited by 5
Vasko, J. L., Aume, L., Tamrakar, S., Colachis, S. C., Dunlap, C. F., Rich, A., Meyers, E., Gabrieli, D. J., & Friedenberg, D. A. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.858377 · Cited by 5
Unsupervised Neural Manifold Alignment for Stable Decoding of Movement from Cortical Signals
Ganjali, M., Mehridehnavi, A., Rakhshani, S., & Khorasani, A. (2023) · International Journal of Neural Systems · doi:10.1142/s0129065724500060 · Cited by 6
Effects of micromachining on anti-oxidant elution from a mechanically-adaptive polymer
Mueller, N. N., Kim, Y., Ocoko, M. Y. M., Dernelle, P., Kale, I. P., Patwa, S., Hermoso, A. C., Chirra, D., Capadona, J. R., & Hess‐Dunning, A. (2024) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ad27f7 · Cited by 5
Darlot, F., Villard, P., Salam, L. A., Rousseau, L., & Piret, G. (2024) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2024.1408088 · Cited by 6
Biological and Mechanical Limitations for Chronic Fast‐Scan Cyclic Voltammetry Sensor Design
Perillo, M. L., Gupta, B., Saxena, A., Veltri, A. P., Li, W., Siegenthaler, J. R., & Purcell, E. K. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202401808 · Cited by 7
Foldable 3D opto-electro array for optogenetic neuromodulation and physiology recording
Gong, Y., Liu, X., Jiang, Z., Weber, A. J., & Li, W. (2025) · Microsystems & Nanoengineering · doi:10.1038/s41378-024-00842-x · Cited by 5
Prospects for a Robust Cortical Recording Interface
Shoffstall, A. J., & Capadona, J. R. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00028-0 · Cited by 5
3D Reconstruction of the Intracortical Volume Around a Hybrid Microelectrode Array
Nambiar, A., Nolta, N. F., & Han, M. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00393 · Cited by 5
Cavuto, M. L., & Constandinou, T. G. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717128 · Cited by 5
Ereifej, E. S., Li, Y., Goss-Varley, M., Kim, Y., Meade, S. M., Chen, K., Rayyan, J., Feng, H., Dona, K. R., McMahon, J. A., et al. (2020) · Micromachines · doi:10.3390/mi11090838 · Cited by 5
Babaroud, N. B., Dekker, R. H., Holk, O., Tiringer, U., Taheri, P., Horváth, D., Nánási, T., Ulbert, I., Serdijn, W. A., & Giagka, V. (2021) · European Microelectronics and Packaging Conference · doi:10.23919/empc53418.2021.9584961 · Cited by 5
Implantable computer-controlled adaptive multielectrode positioning system
Ferrea, E., Suriya-Arunroj, L., Hoehl, D., Thomas, U., & Gail, A. (2017) · Journal of Neurophysiology · doi:10.1152/jn.00504.2017 · Cited by 5
Kim, T., Zhong, Y., & Giszter, S. F. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2911912 · Cited by 5
Insights From Dynamic Neuro-Immune Imaging on Murine Immune Responses to CNS Damage
Dorand, R. D., Benson, B. L., Huang, L. F., Petrošiūtė, A., & Huang, A. Y.-C. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00737 · Cited by 5
Recessed Traces for Planarized Passivation of Chronic Neural Microelectrodes
Nolta, N. F., Ghelich, P., & Han, M. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8857171 · Cited by 5
Micro-coil probes for magnetic intracortical neural stimulation: Trade-offs in materials and design
Thyagarajan, K., Lujan, R. A., Wang, Q., Lu, J., Kor, S., Kakimoto, B., Chang, N., & Bert, J. A. (2021) · APL Materials · doi:10.1063/5.0023486 · Cited by 5
Donoghue, J. P. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00025-5 · Cited by 6
Self-citationOrbán, G., Meszéna, D., Tasnády, K. R., Rózsa, B. J., Ulbert, I., & Márton, G. (2019) · PLoS ONE · doi:10.1371/journal.pone.0221510 · Cited by 5
Rapid Detection of Inactive Channels during Multi-unit Intracranial Recordings
Premchand, B., Toe, K. K., Wang, C. C., Libedinsky, C., Ang, K. K., & So, R. Q. (2019) · IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI ...) · doi:10.1109/bhi.2019.8834654 · Cited by 5
Benchtop Carbon Fiber Microelectrode Array Fabrication ToolkitPreprint
Richie, J., Patel, P. R., Welle, E. J., Dong, T., Chen, L., Shih, A. J., & Chestek, C. A. (2021) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2021.03.22.436422 · Cited by 5
Gregory, B. A., Thompson, C. H., Salatino, J. W., Railing, M. J., Zimmerman, A. F., Gupta, B., Williams, K., Beatty, J. A., Cox, C. L., & Purcell, E. K. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.09.15.507997 · Cited by 5
Design and Microfabrication Considerations for Reliable Flexible Intracortical Implants
Sohal, H. S., Vassilevski, K., Jackson, A., Baker, S. N., & O’Neill, A. (2016) · Frontiers in Mechanical Engineering · doi:10.3389/fmech.2016.00005 · Cited by 4
Yi, D., Landry, N., Blake, S., Baron, J., & Chen, L. (2024) · Journal of Manufacturing Science and Engineering · doi:10.1115/1.4065693 · Cited by 4
Nguyen, C. K., Geramifard, N., Wu, Y., Bhatt, M., Joshi‐Imre, A., Negi, S., & Cogan, S. F. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae2956 · Cited by 4
Yang, S., Huang, C.-J., & Huang, J.-S. (2024) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2024.108208 · Cited by 4
Kao, J. C., Ryu, S. I., & Shenoy, K. V. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318548 · Cited by 4
Badi, M., Borgognon, S., O’Doherty, J. E., & Shokur, S. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00008-3 · Cited by 4
Electrochemical Properties of Sputtered Ruthenium Oxide Neural Stimulation and Recording Electrodes
Chakraborty, B. (2023) · Electrochem · doi:10.3390/electrochem4030023 · Cited by 5
Li, L., Menendez, D. M., Hartzler, A., Pogharian, A., Zaorski, B., Chen, A., Palen, J., Traylor, B., Quill, E., Pawlowski, C. L., et al. (2024) · Biomaterials · doi:10.1016/j.biomaterials.2025.123082 · Cited by 4
Fabrication of all-silicon carbide neural interfaces
Diaz-Botia, C. A., Luna, L. E., Chamanzar, M., Carraro, C., Sabes, P. N., Maboudian, R., & Maharbiz, M. M. (2017) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2017.8008319 · Cited by 4
Ahmed, T., Tasneem, N., & Walker, R. M. (2021) · IEEE International New Circuits and Systems Conference (NEWCAS) · doi:10.1109/newcas50681.2021.9462757 · Cited by 4
Novel method of fabricating self-dissolvable and freely floating neural array
Leber, M., Bhandari, R., Solzbacher, F., & Negi, S. (2017) · International Conference on Solid-State Sensors, Actuators and Microsystems · doi:10.1109/transducers.2017.7994400 · Cited by 4
Implantable Device Fabrication and Packaging
Stieglitz, T. (2021) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_102-1 · Cited by 4
High-Density Implantable Microelectrode Arrays for Brain-Machine Interface Applications
Ghane-Motlagh, B., & Sawan, M. (2014) · Advances in science and technology · doi:10.4028/www.scientific.net/ast.96.95 · Cited by 4
Invasive Brain–Computer Interfaces for Functional Restoration
Ajiboye, A. B., & Kirsch, R. F. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00027-9 · Cited by 7
Phase relationship between micro-electrocorticography and cortical neurons
Richner, T. J., Brodnick, S. K., Thongpang, S., Sandberg, A. A., Krugner-Higby, L. A., & Williams, J. C. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab335b · Cited by 4
The Human Body: A Special Environment
Clément, C. (2019) · Springer eBooks · doi:10.1007/978-3-030-27852-6_4 · Cited by 4
The ‘Brussels 4’: essential requirements for implantable brain–computer interface user autonomy
Oxley, T. J., Deo, D. R., Cernera, S., Sawyer, A., Putrino, D. F., Ramsey, N. F., & Fry, A. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ada0e6 · Cited by 6
Burkhart, G., Grabinski, S. E., Wang, J., Glusauskas, G., Thompson, A., Lee, A., Zhu, Z., Donskey, C. J., von Recum, H. A., Zhang, L., et al. (2025) · Biomaterials · doi:10.1016/j.biomaterials.2025.123834 · Cited by 3
Chan, K. S., Guo, Y., Duan, Q., Wang, X., Hao, Y., Wu, S., & Dai, X. (2026) · ACS Nano · doi:10.1021/acsnano.6c05215 · Cited by 3
Micropillar Electrode Array: From Metal to Dielectric Interface
Branch, B. A., Schei, J. L., Gupta, G., Dattelbaum, A. M., Petsev, D. N., & George, J. S. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2425305 · Cited by 3
Li, L., Hartzler, A., Menendez, D. M., Zhang, J., Chen, A., Lam, D., Traylor, B., Quill, E., Nethery, D. E., Hoeferlin, G. F., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-63583-z · Cited by 3
High-Speed Communication Up to 600 Mbps Over FDA-Cleared implantable Wirelines
Ahmed, T., Tasneem, N., & Walker, R. M. (2018) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2018.8584733 · Cited by 3
Recent developments and future perspectives on neuroelectronic devices
Rinklin, P., & Wolfrum, B. L. (2021) · e-Neuroforum · doi:10.1515/nf-2021-0019 · Cited by 3
Engineering Materials for Neurotechnology
Ghezzi, D. (2023) · Advanced Engineering Materials · doi:10.1002/adem.202201412 · Cited by 2
Bouton, C. E. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00029-2 · Cited by 3
Technical Challenges of Active Implantable Medical Devices for Neurotechnology
Morales, J. M. H., & Clément, C. (2018) · IEEE CPMT Symposium Japan (ICSJ) · doi:10.1109/icsj.2018.8602885 · Cited by 3
Restoring the sense of touch with electrical stimulation of the nerve and brain
Callier, T., & Bensmaı̈a, S. J. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00010-1 · Cited by 3
Tools for Surface Treatment of Silicon Planar Intracortical Microelectrodes
Krebs, O. K., Mittal, G., Ramani, S., Zhang, J., Shoffstall, A. J., Cogan, S. F., Pancrazio, J. J., & Capadona, J. R. (2022) · Journal of Visualized Experiments · doi:10.3791/63500 · Cited by 3
McDermott, M. D., Olczak, K. P., & Otto, K. J. (2017) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2017.8008336 · Cited by 3
Extracting Semantic-Dynamic Features for Long-Term Stable Brain Computer Interface
Fang, T., Zheng, Q., Qi, Y., & Pan, G. (2023) · Proceedings of the AAAI Conference on Artificial Intelligence · doi:10.1609/aaai.v37i5.25738 · Cited by 2
In Vitro and In Vivo Biocompatibility Testing of Silicon Carbide for Neural Interfaces
Knaack, G. L. (2014) · doi:10.13021/mars/7344 · Cited by 3
Rotermund, D., Pistor, J., Hoeffmann, J., Schellenberg, T., Boll, D., Tolstosheeva, E., Gauck, D., Stemmann, H., Peters-Drolshagen, D., Kreiter, A. K., et al. (2016) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/036855 · Cited by 3
Human psychophysics of direct cortical stimulation of somatosensory cortex
Cronin, J. A. (2018) · ResearchWorks at the University of Washington (University of Washington) · Cited by 3
Clustering with Fast, Automated and Reproducible assessment applied to longitudinal neural trackingPreprint
Zhu, H., Li, X., Sun, L., He, F., Zhao, Z., Luan, L., Tran, N. M., & Xie, C. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2003.08533 · Cited by 3
Topological supramolecular network enabled highly conductive and stretchable organic bioelectronicsPreprint
Jiang, Y., Zhang, Z., Wang, Y.-X., Li, D., Coen, C., Hwaun, E., Chen, G., Wu, H., Zhong, D., Niu, S., et al. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.01.16.476423 · Cited by 4
Riggins, T. E., Whitsitt, Q. A., Saxena, A., Hunter, E., Hunt, B., Thompson, C. H., Moore, M. G., & Purcell, E. K. (2023) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2023.01.06.522870 · Cited by 3
Distributed Neural Interfaces: Challenges and Trends in Scaling Implantable Technology
Szostak, K. M., Feng, P., Mazza, F., & Constandinou, T. G. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_11 · Cited by 2
Wang, S., Yan, X., Jiao, X., & Yang, H. (2024) · Materials · doi:10.3390/ma17102334 · Cited by 2
Forrest, A. M., Kunigk, N. G., Collinger, J. L., Gaunt, R. A., Chen, X., Geest, J. P. V., & Kozai, T. D. Y. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae1bda · Cited by 2
Mirkiani, S., O’Sullivan, C. L., Arefadib, A., Tyreman, N., Sari, K., Wilson, D., Tawakol, O., Troyk, P. R., Fox, R., & Mushahwar, V. K. (2026) · Scientific Reports · doi:10.1038/s41598-026-42212-9 · Cited by 2
Yi, D., Thorson, J., Kwankam, N., & Chen, L. (2026) · Frontiers in Neuroscience · doi:10.3389/fnins.2026.1774302 · Cited by 2
Jiang, Z., Zou, G., Ju, W., Yu, J., Guo, C., Zhu, M. J., Vaibavi, S. R., Li, W. M., Ackers-Johnson, M. A., Vidyasekar, P., et al. (2026) · Science Advances · doi:10.1126/sciadv.aed7125 · Cited by 2
Aquatic Animals Operating at High Reynolds Numbers
Fish, F. E. (2020) · Cambridge University Press eBooks · doi:10.1017/9781139058995.011 · Cited by 2
Colachis, S. C., Ganzer, P. D., Friedenberg, D. A., & Sharma, G. (2021) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_119-1 · Cited by 2
Roy, J., Sarah, U. T., Lissorgues, G., Français, O., Rezgui, A., Poulichet, P., Takhedmit, H., Scorsone, E., & Rousseau, L. (2024) · Sensors · doi:10.3390/s24113619 · Cited by 2
Noninvasive EEG Recordings from Freely Moving Piglets
de Camp, N. V., Dietze, S., Klaßen, M., & Bergeler, J. (2018) · Journal of Visualized Experiments · doi:10.3791/58226 · Cited by 2
Parylene neural probe with embedded CMOS multiplexing amplifier
Forssell, M., & Fedder, G. K. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512938 · Cited by 2
Simultaneous impedance measurements of the Utah electrodes array : A finite element method analysis
della Valle, E., & Weiland, J. D. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717172 · Cited by 2
Thermoset Polymers for Bioelectronic Interfaces - Engineering of Thermomechanical Properties
Duran-Martinez, A. C., Hosseini, S., Nero, D. D., Joshi‐Imre, A., Voit, W. E., & Ecker, M. (2019) · IEEE Electronic Components and Technology Conference (ECTC) · doi:10.1109/ectc.2019.00195 · Cited by 2
Gozzi, N., & Valle, G. (2023) · Elsevier eBooks · doi:10.1016/b978-0-443-18498-7.00014-4 · Cited by 2
A Multiplexed Electrochemical Measurement System for Characterization of Implanted Electrodes
Gardner, A. T., Strathman, H. J., & Walker, R. M. (2020) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas45731.2020.9180496 · Cited by 2
NOVEL DESIGN STRATEGIES TO REDUCE THE FOREIGN BODY RESPONSE TO CENTRAL NERVOUS SYSTEM IMPLANTS
Skousen, J. L. (2012) · doi:10.13140/2.1.3896.4485 · Cited by 2
Design of Ceramic Packages for Acoustically Coupled Implantable Medical DevicesPreprint
Shen, K., & Maharbiz, M. M. (2019) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/836312 · Cited by 2
High-Stability Polyimide-based Flexible Electrodes with IrO x to Interface the Mouse Vagus NervePreprint
Sun, T., Falcone, J., Crosfield, C., Lopez, M. F., Peragine, J., Modi, R., Sharma, R., Baker, B., Anderson, G., Savani, S., et al. (2021) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2021.05.10.442867 · Cited by 2
Material damage to multielectrode arrays after electrolytic lesioning is insignificantPreprint
Tor, A., Clarke, S. E., Bray, I. E., & Nuyujukian, P. (2025) · eLife · doi:10.1101/2025.03.26.645429 · Cited by 3
Letner, J. G., Lam, J. L. W., Copenhaver, M. G., Barrow, M., Patel, P. R., Richie, J. M., Lee, J., Kim, H.-S., Cai, D., Weiland, J. D., et al. (2026) · Frontiers in Neuroscience · doi:10.3389/fnins.2026.1848862 · Cited by 1
Rethink the motor cortical control via the experiment-analysis-model flywheel: an overview
Jiang, H., Bu, X., Shi, Q., Zheng, Z., Zhu, D., Tang, H., & Chen, Y. (2025) · Med-X · doi:10.1007/s44258-025-00059-1 · Cited by 1
Gardetto, A., Atkins, D. J., Cannoletta, G., Zappatore, G. A., & Carrabba, A. S. (2025) · Prosthesis · doi:10.3390/prosthesis7050111 · Cited by 1
Friend, Not Foe: Lowered Tissue Reactivity to Long‐Term Polyimide Implants
Orlemann, C., Santis, L. M. D., Neering, P., Boehler, C., Sharma, K., Aarts, A. A. A., Holzhammer, T., van Daal, R. J. J., Ruther, P., Asplund, M., et al. (2026) · Advanced Science · doi:10.64898/2026.02.06.703281 · Cited by 3
Meyer, L. M., & Zamani, M. (2026) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae6dfd · Cited by 1
A roadmap to navigate the future of neural engineering
Grill, W. M., Chestek, C. A., Wang, Y., Aberra, A. S., Alain Destexhe, A., Chan, R. H. M., Lu, B., Hanein, Y., Robinson, J. T., Whitmire, C., et al. (2026) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae78c2 · Cited by 2
Nanostructured coatings on soft-polymer based neural probes for addressing neuroinflammation
Ocoko, M. Y. M., Kasthuri, N. M., Lugo, I., Hanzlicek, B., Wang, J., Duncan, J., Wang, H., Capadona, J. R., Hamedani, H. A., & Hess‐Dunning, A. (2026) · Acta Biomaterialia · doi:10.1016/j.actbio.2026.06.052 · Cited by 1
Chen, Q., Huang, N., Tong, C., Liu, P., Que, S., Xue, H., Xiao, X., & Song, E. (2026) · npj Biosensing · doi:10.1038/s44328-026-00109-7 · Cited by 2
Burkhart, G., Wang, J. J., Gao, Y., Glusauskas, G., Thompson, A., Newman, L., Pancrazio, J. J., von Recum, H. A., Hess‐Dunning, A., Hamedani, H. A., et al. (2026) · Journal of Materials Chemistry B · doi:10.1039/d5tb02785c · Cited by 1
Metadata
- Authors
- Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P.
- Affiliations
- Brown University (US)
- Corresponding author
- Barrese, James C
- Rao, Naveen
- Paroo, Kaivon
- Triebwasser, Corey
- Vargas-Irwin, Carlos
- Franquemont, Lachlan
- Donoghue, John P
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2013-11-12
- Volume
- 10
- Issue
- 6
- Pages
- 066014
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC4868924
- OpenAlex
- W1999169980
- Cited by
- 600
- References
- 42
- Citations by year
- 2026: 39 · 2025: 40 · 2024: 43 · 2023: 49 · 2022: 38 · 2021: 63 · 2020: 68 · 2019: 53 · 2018: 67 · 2017: 47 · 2016: 34 · 2015: 28 · 2014: 28 · 2013: 1
- FWCI
- 14.88
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Neurological Disorders and Stroke (R01NS025074)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Analog and Mixed-Signal Circuit Design
- Keywords
- intracortical microelectrode arrays, failure mode analysis, brain-computer interface, chronic implantation, non-human primates, rhesus monkeys, silicon microelectrodes, electrode impedance, meningeal reaction, connector failure, neuroprosthetics, chronic neural recording, recording duration
- MeSH terms
- Animals, Cerebral Cortex, Electrodes, Implanted, Female, Macaca mulatta, Male, Microelectrodes, Silicon, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review