Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
Open accessAbstract
OBJECTIVE: Brain-computer interfaces (BCIs) using chronically implanted intracortical microelectrode arrays (MEAs) have the potential to restore lost function to people with disabilities if they work reliably for years. Current sensors fail to provide reliably useful signals over extended periods of time for reasons that are not clear. This study reports a comprehensive retrospective analysis from a large set of implants of a single type of intracortical MEA in a single species, with a common set of measures in order to evaluate failure modes. APPROACH: Since 1996, 78 silicon MEAs were implanted in 27 monkeys (Macaca mulatta). We used two approaches to find reasons for sensor failure. First, we classified the time course leading up to complete recording failure as acute (abrupt) or chronic (progressive). Second, we evaluated the quality of electrode recordings over time based on signal features and electrode impedance. Failure modes were divided into four categories: biological, material, mechanical, and unknown. MAIN RESULTS: Recording duration ranged from 0 to 2104 days (5.75 years), with a mean of 387 days and a median of 182 days (n = 78). Sixty-two arrays failed completely with a mean time to failure of 332 days (median = 133 days) while nine array experiments were electively terminated for experimental reasons (mean = 486 days). Seven remained active at the close of this study (mean = 753 days). Most failures (56%) occurred within a year of implantation, with acute mechanical failures the most common class (48%), largely because of connector issues (83%). Among grossly observable biological failures (24%), a progressive meningeal reaction that separated the array from the parenchyma was most prevalent (14.5%). In the absence of acute interruptions, electrode recordings showed a slow progressive decline in spike amplitude, noise amplitude, and number of viable channels that predicts complete signal loss by about eight years. Impedance measurements showed systematic early increases, which did not appear to affect recording quality, followed by a slow decline over years. The combination of slowly falling impedance and signal quality in these arrays indicates that insulating material failure is the most significant factor. SIGNIFICANCE: This is the first long-term failure mode analysis of an emerging BCI technology in a large series of non-human primates. The classification system introduced here may be used to standardize how neuroprosthetic failure modes are evaluated. The results demonstrate the potential for these arrays to record for many years, but achieving reliable sensors will require replacing connectors with implantable wireless systems, controlling the meningeal reaction, and improving insulation materials. These results will focus future research in order to create clinical neuroprosthetic sensors, as well as valuable research tools, that are able to safely provide reliable neural signals for over a decade.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Previous demonstrations of intracortical insertion of MEAs showed injury and neuronal loss at the implantation location99, as well as signal instabilities contributing to signal loss and performance decline over time100,101.”
From The speech neuroprosthesis (2024) · cites this paper as [100]
In the citing paper
“In spite of the formidable results of using UEA/USEA technologies described above, most penetrating microelectrode arrays currently have maximum in vivo lifetimes of several months or a few years (Suner et al 2005, Prasad et al 2012, Barrese et al 2013).”
From Clinical applications of penetrating neural interfaces and Utah Electrode Array technologies (2016) · cites this paper as Barrese et al 2013
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Neuronal ensemble control of prosthetic devices by a human with tetraplegia
Hochberg, L. R., Serruya, M. D., Friehs, G. M., Mukand, J. A., Saleh, M., Caplan, A. H., Branner, A., Chen, D., Penn, R. D., & Donoghue, J. P. (2006) · Nature · doi:10.1038/nature04970 · Cited by 3,295
Self-citationReach and grasp by people with tetraplegia using a neurally controlled robotic arm
Hochberg, L. R., Bacher, D., Jarosiewicz, B., Masse, N. Y., Simeral, J. D., Vogel, J., Haddadin, S., Liu, J., Cash, S. S., van der Smagt, P., et al. (2012) · Nature · doi:10.1038/nature11076 · Cited by 2,719
Self-citationResponse of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896
Biran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893
Brain responses to micro-machined silicon devices
Szarowski, D. H., Andersen, M. D., Retterer, S. T., Spence, A. J., Isaacson, M. S., Craighead, H. G., Turner, J. N., & Shain, W. G. (2003) · Brain Research · doi:10.1016/s0006-8993(03)03023-3 · Cited by 811
Chronic recording capability of the Utah Intracortical Electrode Array in cat sensory cortex
Rousche, P. J., & Normann, R. A. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00031-4 · Cited by 646
Cerebral Astrocyte Response to Micromachined Silicon Implants
Turner, J. N., Shain, W. G., Szarowski, D. H., Andersen, M., Martins, S., Isaacson, M. S., & Craighead, H. G. (1999) · Experimental Neurology · doi:10.1006/exnr.1998.6983 · Cited by 603
Single-neuron dynamics in human focal epilepsy
Truccolo, W., Donoghue, J., Hochberg, L. R., Eskandar, E. N., Madsen, J. R., Anderson, W. S., Brown, E. N., Halgren, E., & Cash, S. S. (2011) · Nature Neuroscience · doi:10.1038/nn.2782 · Cited by 528
Self-citationFactors influencing the biocompatibility of insertable silicon microshafts in cerebral cortex
Edell, D. J., Toi, V. V., McNeil, V. M., & Clark, L. D. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.141202 · Cited by 481
McConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480
Suner, S., Fellows, M. R., Vargas-Irwin, C., Nakata, G. K., & Donoghue, J. P. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.857687 · Cited by 415 · cited 2×
“The day-to-day variability in the number of signals typically recorded on these arrays supports this view, as previously reported (Suner et al 2005, Chestek et al 2011).”
Self-citationKim, S.-P., Simeral, J. D., Hochberg, L. R., Donoghue, J. P., & Black, M. J. (2008) · Journal of Neural Engineering · doi:10.1088/1741-2560/5/4/010 · Cited by 407
Self-citationComplex impedance spectroscopy for monitoring tissue responses to inserted neural implants
Williams, J. C., Hippensteel, J. A., Dilgen, J., Shain, W. G., & Kipke, D. R. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/4/007 · Cited by 402
A glass/silicon composite intracortical electrode array
Jones, K. E., Campbell, P. K., & Normann, R. A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368134 · Cited by 354
Chestek, C. A., Gilja, V., Nuyujukian, P., Foster, J. D., Fan, J. M., Kaufman, M. T., Churchland, M. M., Rivera-Alvidrez, Z., Cunningham, J. P., Ryu, S. I., et al. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/4/045005 · Cited by 350
Bjornsson, C. S., Oh, S., Al‐Kofahi, Y., Lim, Y.-J., Smith, K. L., Turner, J. N., De, S., Roysam, B., Shain, W. G., & Kim, S. J. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/3/002 · Cited by 330
An implantable wireless neural interface for recording cortical circuit dynamics in moving primates
Borton, D. A., Yin, M., Aceros, J. C., & Nurmikko, A. V. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/2/026010 · Cited by 326
Zhong, Y., & Bellamkonda, R. V. (2007) · Brain Research · doi:10.1016/j.brainres.2007.02.024 · Cited by 326
A method for pneumatically inserting an array of penetrating electrodes into cortical tissue
Rousche, P. J., & Normann, R. A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368133 · Cited by 305
Biomechanical analysis of silicon microelectrode-induced strain in the brain
Lee, H., Bellamkonda, R. V., Sun, W., & Levenston, M. E. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/4/003 · Cited by 281
Neuronal Interactions Improve Cortical Population Coding of Movement Direction
Maynard, E. M., Hatsopoulos, N. G., Ojakangas, C. L., Acuna, B. D., Sanes, J. N., Normann, R. A., & Donoghue, J. P. (1999) · Journal of Neuroscience · doi:10.1523/jneurosci.19-18-08083.1999 · Cited by 274
Self-citationLiu, X., McCreery, D. B., Carter, R. R., Bullara, L. A., Yuen, T. G. H., & Agnew, W. F. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.788468 · Cited by 272
The impact of chronic blood–brain barrier breach on intracortical electrode function
Saxena, T. K., Karumbaiah, L., Gaupp, E. A., Patkar, R., Patil, K., Betancur, M. I., Stanley, G. B., & Bellamkonda, R. V. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.03.007 · Cited by 270
Characterization of parylene C as an encapsulation material for implanted neural prostheses
Hassler, C., von Metzen, R. P., Ruther, P., & Stieglitz, T. (2010) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.31584 · Cited by 263
Parylene as a Chronically Stable, Reproducible Microelectrode Insulator
Loeb, G. E., Bak, M. J., Salcman, M., & Schmidt, E. M. (1977) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1977.326115 · Cited by 258
A quantitative overview of biophysical forces impinging on neural function
Mueller, J. K., & Tyler, W. J. (2014) · Physical Biology · doi:10.1088/1478-3975/11/5/051001 · Cited by 225
Long-term implants of Parylene-C coated microelectrodes
Schmidt, E., Mcintosh, J. S., & Bak, M. (2006) · Medical and Biological Engineering and Computing · doi:10.1007/bf02441836 · Cited by 216
Bansal, A. K., Truccolo, W., Vargas-Irwin, C. E., & Donoghue, J. P. (2011) · Journal of Neurophysiology · doi:10.1152/jn.00781.2011 · Cited by 206
Self-citationCorrosion of tungsten microelectrodes used in neural recording applications
Patrick, E. E., Orazem, M. E., Sanchez, J. C., & Nishida, T. (2011) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2011.03.012 · Cited by 168
Quantifying long-term microelectrode array functionality using chronic in vivo impedance testing
Prasad, A., & Sanchez, J. C. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/2/026028 · Cited by 164
The astrocyte/meningeal cell interface – a barrier to successful nerve regeneration?
Shearer, M. C., & Fawcett, J. W. (2001) · Cell and Tissue Research · doi:10.1007/s004410100384 · Cited by 162
Griffith, R. W., & Humphrey, D. R. (2006) · Neuroscience Letters · doi:10.1016/j.neulet.2006.07.018 · Cited by 147
In Vivo Electrical Impedance Spectroscopy of Tissue Reaction to Microelectrode Arrays
Mercanzini, A., Colin, P., Bensadoun, J., Bertsch, A., & Renaud, P. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2018457 · Cited by 130
Potter, K. A., Buck, A. C., Self, W. K., Callanan, M. E., Sunil, S., & Capadona, J. R. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.05.035 · Cited by 129
A technique to prevent dural adhesions to chronically implanted microelectrode arrays
Maynard, E. M., Fernández, E., & Normann, R. A. (2000) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(00)00159-x · Cited by 102
Passivation and corrosion of microelectrode arrays
Schmitt, G., Schultze, J. W., Faßbender, F. J., Buß, G., Luth, H., & Schöning, M. J. (1999) · Electrochimica Acta · doi:10.1016/s0013-4686(99)00094-8 · Cited by 95
Tissue response to potential neuroprosthetic materials implanted subdurally
Yuen, T. G. H., Agnew, W. F., & Bullara, L. A. (1987) · Biomaterials · doi:10.1016/0142-9612(87)90103-7 · Cited by 92
Impedance Characterization of Microarray Recording Electrodes in Vitro
Merrill, D. R., & Tresco, P. A. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.856245 · Cited by 74
Histopathological Changes Produced by Implanted Electrodes in Cat Brains
Collias, J. C., & Manuelidis, E. E. (1957) · Journal of neurosurgery · doi:10.3171/jns.1957.14.3.0302 · Cited by 68
The ultrastructure of the sheath around chronically implanted electrodes in brain
Schultz, R. L., & Willey, T. J. (1976) · Journal of Neurocytology · doi:10.1007/bf01181577 · Cited by 63
Aspects of silicone rubber as an encapsulant for neurological prostheses
Donaldson, P. E. K. (1991) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446293 · Cited by 53
Impedance characterization of microarray recording electrodes in vitro
Merrill, D. R., & Tresco, P. A. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1404210 · Cited by 3
Research works printed but not linked: 2
- McConnell G C, 2009, J. Neural Eng., 6
- Prasad A, 2012, J. Neural Eng., 9
Cited by
Sorted by citations
Yang, J. C., Mun, J., Kwon, S. Y., Park, S., Bao, Z., & Park, S. (2019) · Advanced Materials · doi:10.1002/adma.201904765 · Cited by 1,641
Monitoring of Vital Signs with Flexible and Wearable Medical Devices
Khan, Y., Ostfeld, A. E., Lochner, C. M., Pierre, A., & Arias, A. C. (2016) · Advanced Materials · doi:10.1002/adma.201504366 · Cited by 1,408
Electronic dura mater for long-term multimodal neural interfaces
Minev, I. R., Musienko, P., Hirsch, A., Barraud, Q., Wenger, N., Moraud, E. M., Gandar, J., Capogrosso, M., Milekovic, T., Asboth, L., et al. (2015) · Science · doi:10.1126/science.1260318 · Cited by 1,082
Materials and technologies for soft implantable neuroprostheses
Lacour, S. P., Courtine, G., & Guck, J. (2016) · Nature Reviews Materials · doi:10.1038/natrevmats.2016.63 · Cited by 738
Neural recording and modulation technologies
Chen, R., Canales, A., & Anikeeva, P. (2017) · Nature Reviews Materials · doi:10.1038/natrevmats.2016.93 · Cited by 655
Metadata
- Authors
- Barrese, J. C., Rao, N., Paroo, K., Triebwasser, C., Vargas-Irwin, C., Franquemont, L., & Donoghue, J. P.
- Affiliations
- Brown University (US)
- Corresponding author
- Barrese, James C
- Rao, Naveen
- Paroo, Kaivon
- Triebwasser, Corey
- Vargas-Irwin, Carlos
- Franquemont, Lachlan
- Donoghue, John P
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2013-11-12
- Volume
- 10
- Issue
- 6
- Pages
- 066014
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC4868924
- OpenAlex
- W1999169980
- Cited by
- 600
- References
- 42
- Citations by year
- 2026: 39 · 2025: 40 · 2024: 43 · 2023: 49 · 2022: 38 · 2021: 63 · 2020: 68 · 2019: 53 · 2018: 67 · 2017: 47 · 2016: 34 · 2015: 28 · 2014: 28 · 2013: 1
- FWCI
- 14.88
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Neurological Disorders and Stroke (R01NS025074)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Analog and Mixed-Signal Circuit Design
- Keywords
- intracortical microelectrode arrays, failure mode analysis, brain-computer interface, chronic implantation, non-human primates, rhesus monkeys, silicon microelectrodes, electrode impedance, meningeal reaction, connector failure, neuroprosthetics, chronic neural recording, recording duration
- MeSH terms
- Animals, Cerebral Cortex, Electrodes, Implanted, Female, Macaca mulatta, Male, Microelectrodes, Silicon, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review