In vivo two-photon microscopy reveals immediate microglial reaction to implantation of microelectrode through extension of processes
Open accessAbstract
OBJECTIVE: Penetrating cortical neural probe technologies allow investigators to record electrical signals in the brain. Implantation of probes results in acute tissue damage, and microglia density increases around implanted devices over weeks. However, the mechanisms underlying this encapsulation are not well understood in the acute temporal domain. The objective here was to evaluate dynamic microglial response to implanted probes using two-photon microscopy. APPROACH: Using two-photon in vivo microscopy, cortical microglia ∼200 µm below the surface of the visual cortex were imaged every minute in mice with green fluorescent protein-expressing microglia. MAIN RESULTS: Following probe insertion, nearby microglia immediately extended processes toward the probe at (1.6 ± 1.3) µm min(-1) during the first 30-45 min, but showed negligible cell body movement for the first 6 h. Six hours following probe insertion, microglia at distances <130.0 µm (p = 0.5) from the probe surface exhibit morphological characteristics of transitional stage (T-stage) activation, similar to the microglial response observed with laser-induced blood-brain barrier damage. T-stage morphology and microglia directionality indexes were developed to characterize microglial response to implanted probes. Evidence suggesting vascular reorganization after probe insertion and distant vessel damage was also observed hours after probe insertion. SIGNIFICANCE: A precise temporal understanding of the cellular response to microelectrode implantation will facilitate the search for molecular cues initiating and attenuating the reactive tissue response.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The implant’s size and volume directly correlate with the risk of intracerebral hemorrhage and ischemic injuries (Rosidi et al., 2011), with larger implants exerting more pressure, potentially leading to vascular compression and secondary injuries (Jaquins-Gerstl and Michael, 2009; Kozai et al., 2012).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Kozai et al., 2012
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
New diagnostic criteria for multiple sclerosis: Guidelines for research protocols
Poser, C. M., Paty, D. W., Scheinberg, L., McDonald, W. I., Davis, F. A., Ebers, G. C., Johnson, K. P., Sibley, W. A., Silberberg, D. H., & Tourtellotte, W. W. (1983) · Annals of Neurology · doi:10.1002/ana.410130302 · Cited by 7,384
Resting microglial cells are highly dynamic surveillants of brain parenchyma in vivo
Nimmerjahn, A., Kirchhoff, F., & Helmchen, F. (2005) · Science · doi:10.1126/science.1110647 · Cited by 5,827
Microglia: a sensor for pathological events in the CNS
Kreutzberg, G. W. (1996) · Trends in Neurosciences · doi:10.1016/0166-2236(96)10049-7 · Cited by 4,396
Deep tissue two-photon microscopy
Helmchen, F., & Denk, W. T. (2005) · Nature Methods · doi:10.1038/nmeth818 · Cited by 4,288
Synaptic Pruning by Microglia Is Necessary for Normal Brain Development
Paolicelli, R. C., Bolasco, G., Pagani, F., Maggi, L., Scianni, M., Panzanelli, P., Giustetto, M., Ferreira, T. A., Guiducci, E., Dumas, L., et al. (2011) · Science · doi:10.1126/science.1202529 · Cited by 4,244
Microglia Sculpt Postnatal Neural Circuits in an Activity and Complement-Dependent Manner
Schafer, D. P., Lehrman, E. K., Kautzman, A. G., Koyama, R., Mardinly, A. R., Yamasaki, R., Ransohoff, R. M., Greenberg, M. E., Barres, B. A., & Stevens, B. (2012) · Neuron · doi:10.1016/j.neuron.2012.03.026 · Cited by 4,089
ATP mediates rapid microglial response to local brain injury in vivo
Davalos, D., Grutzendler, J., Yang, G., Kim, J. V., Zuo, Y., Jung, S., Littman, D. R., Dustin, M. L., & Gan, W.-B. (2005) · Nature Neuroscience · doi:10.1038/nn1472 · Cited by 3,913
Nonlinear magic: multiphoton microscopy in the biosciences
Zipfel, W. R., Williams, R. M., & Webb, W. W. (2003) · Nature Biotechnology · doi:10.1038/nbt899 · Cited by 3,911
Microglia: Active sensor and versatile effector cells in the normal and pathologic brain
Hanisch, U.-K., & Kettenmann, H. (2007) · Nature Neuroscience · doi:10.1038/nn1997 · Cited by 3,676
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896
Research works printed but not linked: 4
- Betz A L, 1989, Cerebrovasc. Brain Metab. Rev., 1, 133
- Gilgunn P J Khilwani R Kozai T D Y Weber D J Cui X T Erdos G Ozdoganlar O B Fedder G K, 2012, 25th Int. Conf. on Micro Electro Mechanical Systems (MEMS), IEEE 2012, 56
- Kozai T D Y, 2010, J. Neural Eng., 7
- Nimmerjahn A, 2012, Cold Spring Harb. Protoc.
Cited by
Sorted by citations
Yang, J. C., Mun, J., Kwon, S. Y., Park, S., Bao, Z., & Park, S. (2019) · Advanced Materials · doi:10.1002/adma.201904765 · Cited by 1,641
Kozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721
Self-citationGlial responses to implanted electrodes in the brain
Salatino, J. W., Ludwig, K. A., Kozai, T. D. Y., & Purcell, E. K. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0154-1 · Cited by 642
Self-citationBrain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies
Kozai, T. D. Y., Jaquins-Gerstl, A. S., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn500256e · Cited by 622
Self-citationA Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389
Self-citationLewis, P. M., Ackland, H. M., Lowery, A. J., & Rosenfeld, J. V. (2015) · Brain Research · doi:10.1016/j.brainres.2014.11.020 · Cited by 228
Kozai, T. D. Y., Catt, K. A., Li, X., Gugel, Z. V., Olafsson, V. T., Vazquez, A. L., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.10.040 · Cited by 214
Self-citationState-of-the-art MEMS and microsystem tools for brain research
Seymour, J. P., Wu, F., Wise, K. D., & Yoon, E. (2017) · Microsystems & Nanoengineering · doi:10.1038/micronano.2016.66 · Cited by 221
Ultrasoft microwire neural electrodes improve chronic tissue integration
Du, Z. J., Kolarcik, C. L., Kozai, T. D. Y., Luebben, S. D., Sapp, S. A., Zheng, X. S., Nabity, J., & Cui, X. T. (2017) · Acta Biomaterialia · doi:10.1016/j.actbio.2017.02.010 · Cited by 214
Self-citationChronic in vivo stability assessment of carbon fiber microelectrode arrays
Patel, P. R., Zhang, H., Robbins, M. T., Nofar, J. B., Marshall, S. P., Kobylarek, M. J., Kozai, T. D. Y., Kotov, N. A., & Chestek, C. A. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/6/066002 · Cited by 204
Self-citationActively controlled release of Dexamethasone from neural microelectrodes in a chronic in vivo study
Boehler, C., Kleber, C., Martini, N., Xie, Y., Dryg, I., Stieglitz, T., Hofmann, U., & Asplund, M. (2017) · Biomaterials · doi:10.1016/j.biomaterials.2017.03.019 · Cited by 204
Chronic In Vivo Evaluation of PEDOT/CNT for Stable Neural Recordings
Kozai, T. D. Y., Catt, K. A., Du, Z. J., Na, K., Srivannavit, O., Haque, R., Seymour, J. P., Wise, K. D., Yoon, E., & Cui, X. T. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2445713 · Cited by 195
Self-citationKozai, T. D. Y., Gugel, Z., Li, X., Gilgunn, P. J., Khilwani, R., Ozdoganlar, O. B., Fedder, G. K., Weber, D. J., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.07.039 · Cited by 193
Self-citationAbiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants
Prasad, A., Xue, Q.-S., Dieme, R., Sankar, V., Mayrand, R. C., Nishida, T., Streit, W. J., & Sanchez, J. C. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00002 · Cited by 178
Patel, P. R., Na, K., Zhang, H., Kozai, T. D. Y., Kotov, N. A., Yoon, E., & Chestek, C. A. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/4/046009 · Cited by 165
Self-citationApollo, N. V., Maturana, M. I., Tong, W., Nayagam, D. A. X., Shivdasani, M. N., Foroughi, J., Wallace, G. G., Prawer, S., Ibbotson, M. R., & Garrett, D. J. (2015) · Advanced Functional Materials · doi:10.1002/adfm.201500110 · Cited by 156
Microglia inflict delayed brain injury after subarachnoid hemorrhage
Schneider, U. C., Davids, A., Brandenburg, S., Müller, A., Elke, A., Magrini, S., Atangana, E. N., Turkowski, K., Finger, T., Gutenberg, A., et al. (2015) · Acta Neuropathologica · doi:10.1007/s00401-015-1440-1 · Cited by 155
Kozai, T. D. Y., & Vazquez, A. L. (2015) · Journal of Materials Chemistry B · doi:10.1039/c5tb00108k · Cited by 154
Self-citationMechanical and Biological Interactions of Implants with the Brain and Their Impact on Implant Design
Prodanov, D., & Delbeke, J. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00011 · Cited by 152
Chronically Implanted Intracranial Electrodes: Tissue Reaction and Electrical Changes
Campbell, A., & Wu, C. (2018) · Micromachines · doi:10.3390/mi9090430 · Cited by 149
Kozai, T. D. Y., Du, Z. J., Gugel, Z. V., Smith, M. A., Chase, S. M., Bodily, L., Caparosa, E. M., Friedlander, R. M., & Cui, X. T. (2014) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2014.12.010 · Cited by 137
Self-citationGolabchi, A., Wu, B., Cao, B., Bettinger, C. J., & Cui, X. T. (2019) · Biomaterials · doi:10.1016/j.biomaterials.2019.119519 · Cited by 141
Self-citationIn Vivo Electrochemical Analysis of a PEDOT/MWCNT Neural Electrode Coating
Alba, N. A., Du, Z. J., Catt, K. A., Kozai, T. D. Y., & Cui, X. T. (2015) · Biosensors · doi:10.3390/bios5040618 · Cited by 137
Self-citationTranslational opportunities and challenges of invasive electrodes for neural interfaces
Shen, K., Chen, O., Edmunds, J. L., Piech, D. K., & Maharbiz, M. M. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01021-5 · Cited by 150
Kozai, T. D. Y., Li, X., Bodily, L. M., Caparosa, E. M., Zenonos, G. A., Carlisle, D. L., Friedlander, R. M., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.08.006 · Cited by 136
Self-citationMichelson, N. J., Vazquez, A. L., Eles, J. R., Salatino, J. W., Purcell, E. K., Williams, J. J., Cui, X. T., & Kozai, T. D. Y. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa9dae · Cited by 126
Self-citationFlexible fiber-based optoelectronics for neural interfaces
Park, S., Loke, G. Z. J., Fink, Y., & Anikeeva, P. (2019) · Chemical Society Reviews · doi:10.1039/c8cs00710a · Cited by 135
Kolarcik, C. L., Catt, K., Rost, E., Albrecht, I. N., Bourbeau, D., Du, Z., Kozai, T. D. Y., Luo, X., Weber, D. J., & Tracy Cui, X. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016008 · Cited by 126
Self-citationA Critical Review of Microelectrode Arrays and Strategies for Improving Neural Interfaces
Ferguson, M., Sharma, D., Ross, D., & Zhao, F. (2019) · Advanced Healthcare Materials · doi:10.1002/adhm.201900558 · Cited by 128
Eles, J. R., Vazquez, A. L., Snyder, N. R., Lagenaur, C. F., Murphy, M. C., Kozai, T. D. Y., & Cui, X. T. (2016) · Biomaterials · doi:10.1016/j.biomaterials.2016.10.054 · Cited by 117
Self-citationThe future of neuroscience: flexible and wireless implantable neural electronics
McGlynn, E., Nabaei, V., Ren, E., Galeote‐Checa, G., Das, R., Curia, G., & Heidari, H. (2021) · Advanced Science · doi:10.1002/advs.202002693 · Cited by 115
Lee, H. C., Ejserholm, F., Gaire, J., Currlin, S., Schouenborg, J., Wallman, L., Bengtsson, M., Park, K., & Otto, K. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa68f0 · Cited by 115
Dexamethasone retrodialysis attenuates microglial response to implanted probes in vivo
Kozai, T. D. Y., Jaquins‐Gerstl, A., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2016) · Biomaterials · doi:10.1016/j.biomaterials.2016.02.013 · Cited by 109
Self-citationFlexible and Implantable Microelectrodes for Chronically Stable Neural Interfaces
Shi, J., & Fang, Y. Z. (2018) · Advanced Materials · doi:10.1002/adma.201804895 · Cited by 104
Ultraflexible Neural Electrodes for Long-Lasting Intracortical Recording
He, F., Lycke, R., Ganji, M., Xie, C., & Luan, L. (2020) · iScience · doi:10.1016/j.isci.2020.101387 · Cited by 100
Recent advances in neural electrode–tissue interfaces
Woeppel, K., Yang, Q., & Cui, X. T. (2017) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2017.09.003 · Cited by 99
Self-citationNew Insights into Microglia–Neuron Interactions: A Neuron’s Perspective
Pósfai, B., Cserép, C., Orsolits, B., & Dénes, Á. (2019) · Neuroscience · doi:10.1016/j.neuroscience.2018.04.046 · Cited by 98
Electronic and Ionic Materials for Neurointerfaces
Ferro, M. D., & Melosh, N. A. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201704335 · Cited by 96
Li, P., Wang, L., Zhou, Y., Gan, Y., Zhu, W., Xia, Y., Jiang, X., Watkins, S. C., Vazquez, A. L., Thomson, A. W., et al. (2017) · Journal of the American Heart Association · doi:10.1161/jaha.117.006387 · Cited by 96
Self-citationDegenhart, A. D., Eles, J. R., Dum, R. P., Mischel, J., Smalianchuk, I., Endler, B., Ashmore, R. C., Tyler‐Kabara, E. C., Hatsopoulos, N. G., Wang, W., et al. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/4/046019 · Cited by 96
Self-citationIntracortical Recording Interfaces: Current Challenges to Chronic Recording Function
Gunasekera, B., Saxena, T. K., Bellamkonda, R. V., & Karumbaiah, L. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn5002864 · Cited by 96
Two-photon imaging of chronically implanted neural electrodes: Sealing methods and new insights
Kozai, T. D. Y., Eles, J. R., Vazquez, A. L., & Cui, X. T. (2015) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2015.10.007 · Cited by 88
Self-citationEles, J. R., Vazquez, A. L., Kozai, T. D. Y., & Cui, X. T. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.04.043 · Cited by 86
Self-citationDefining the Microglia Response during the Time Course of Chronic Neurodegeneration
Vincenti, J. E., Murphy, L., Grabert, K., McColl, B. W., Cancellotti, E., Freeman, T. C., & Manson, J. C. (2016) · Journal of Virology · doi:10.1128/jvi.02613-15 · Cited by 88
Elastomeric and soft conducting microwires for implantable neural interfaces
Kolarcik, C. L., Luebben, S. D., Sapp, S. A., Hanner, J., Snyder, N. R., Kozai, T. D. Y., Chang, E., Nabity, J., Nabity, S. T., Lagenaur, C. F., et al. (2015) · Soft Matter · doi:10.1039/c5sm00174a · Cited by 90
Self-citationCody, P. A., Eles, J. R., Lagenaur, C. F., Kozai, T. D. Y., & Cui, X. T. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.01.025 · Cited by 87
Self-citationTissue-Compliant Neural Implants from Microfabricated Carbon Nanotube Multilayer Composite
Zhang, H., Patel, P. R., Xie, Z., Swanson, S. D., Wang, X., & Kotov, N. A. (2013) · ACS Nano · doi:10.1021/nn402074y · Cited by 88
Miller, E. B., Zhang, P., Ching, K., Pugh, E. N., & Burns, M. E. (2019) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1903336116 · Cited by 85
Electrode Materials for Chronic Electrical Microstimulation
Zheng, X. S., Tan, C., Castagnola, E., & Cui, X. T. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100119 · Cited by 90
Self-citationBennett, C., Mohammed, F. S., Álvarez-Ciara, A., Nguyen, M. A., Dietrich, W. D., Rajguru, S. M., Streit, W. J., & Prasad, A. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.09.040 · Cited by 87
A Parylene Neural Probe Array for Multi-Region Deep Brain Recordings
Wang, X., Weltman, A., Xu, H., Slingsby-Smith, Z., Lecomte, A., Scholten, K., Song, D. B., & Meng, E. F. (2020) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2020.3000235 · Cited by 87
Michelson, N. J., Eles, J. R., Vazquez, A. L., Ludwig, K. A., & Kozai, T. D. Y. (2018) · Journal of Neuroscience Research · doi:10.1002/jnr.24370 · Cited by 85
Self-citationWellman, S. M., Li, L., Yaxiaer, Y., McNamara, I., & Kozai, T. D. Y. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00493 · Cited by 79
Self-citationMelatonin improves quality and longevity of chronic neural recording
Golabchi, A., Wu, B., Li, X., Carlisle, D. L., Kozai, T. D. Y., Friedlander, R. M., & Cui, X. T. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.07.026 · Cited by 74
Self-citationLee, Y., Shin, H., Lee, D., Choi, S., Cho, I., & Seo, J. (2021) · Advanced Science · doi:10.1002/advs.202100231 · Cited by 73
Neuroadhesive protein coating improves the chronic performance of neuroelectronics in mouse brain
Golabchi, A., Woeppel, K. M., Li, X., Lagenaur, C. F., & Cui, X. T. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112096 · Cited by 73
Self-citationIn vivo spatiotemporal dynamics of NG2 glia activity caused by neural electrode implantation
Wellman, S. M., & Kozai, T. D. Y. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.02.037 · Cited by 67
Self-citationA comparison of insertion methods for surgical placement of penetrating neural interfaces
Thielen, B., & Meng, E. F. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abf6f2 · Cited by 78
Welle, E. J., Patel, P. R., Woods, J. E., Petrossians, A., della Valle, E., Vega-Medina, A., Richie, J. M., Cai, D., Weiland, J. D., & Chestek, C. A. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab8343 · Cited by 68
Sohal, H. S., Clowry, G. J., Jackson, A., O'Neill, A. G., & Baker, S. N. (2016) · PLoS ONE · doi:10.1371/journal.pone.0165606 · Cited by 70
Yuede, C. M., Lee, H. J., Restivo, J. L., Davis, T. A., Hettinger, J. C., Wallace, C. E., Young, K. L., Hayne, M. R., Bu, G., Li, C.-Z., et al. (2016) · The Journal of Experimental Medicine · doi:10.1084/jem.20151428 · Cited by 65
Flexible and Implantable Polyimide Aptamer-Field-Effect Transistor Biosensors
Zhao, C., Man, T., Cao, Y. F., Weiss, P. S., Monbouquette, H. G., & Andrews, A. M. (2022) · ACS Sensors · doi:10.1021/acssensors.2c01909 · Cited by 61
Shoffstall, A. J., Paiz, J. E., Miller, D. M., Rial, G. M., Willis, M. T., Menendez, D. M., Hostler, S. R., & Capadona, J. R. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa9f32 · Cited by 60
Long-term intracortical microelectrode array performance in a human: a 5 year retrospective analysis
Colachis, S. C., Dunlap, C. F., Annetta, N. V., Tamrakar, S., Bockbrader, M. A., & Friedenberg, D. A. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac1add · Cited by 67
Alahi, M. E. E., Liu, Y., Xu, Z., Wang, H., Wu, T., & Mukhopadhyay, S. C. (2021) · Materials Today Communications · doi:10.1016/j.mtcomm.2021.102853 · Cited by 61
Aptamer-functionalized neural recording electrodes for the direct measurement of cocaine in vivo
Taylor, I. M., Du, Z. J., Bigelow, E. T., Eles, J. R., Horner, A. R., Catt, K. A., Weber, S. G., Jamieson, B. G., & Cui, X. T. (2017) · Journal of Materials Chemistry B · doi:10.1039/c7tb00095b · Cited by 53
Self-citationXie, Y., Martini, N., Hassler, C., Kirch, R. D., Stieglitz, T., Seifert, A., & Hofmann, U. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00034 · Cited by 52
Coating flexible probes with an ultra fast degrading polymer to aid in tissue insertion
Lo, M.-C., Wang, S., Singh, S., Damodaran, V. B., Kaplan, H. M., Kohn, J., Shreiber, D. I., & Zahn, J. D. (2015) · Biomedical Microdevices · doi:10.1007/s10544-015-9927-z · Cited by 57
Khilwani, R., Gilgunn, P. J., Kozai, T. D. Y., Ong, X. C., Korkmaz, E., Gunalan, P. K., Cui, X. T., Fedder, G. K., & Özdoğanlar, O. B. (2016) · Biomedical Microdevices · doi:10.1007/s10544-016-0125-4 · Cited by 51
Self-citationChen, K., Wellman, S. M., Yaxiaer, Y., Eles, J. R., & Kozai, T. D. Y. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.120526 · Cited by 46
Self-citationA guide towards long-term functional electrodes interfacing neuronal tissue
Renz, A. F., Reichmuth, A. M., Stauffer, F., Thompson‐Steckel, G., & Vörös, J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aae0c2 · Cited by 54
Stieger, K. C., Eles, J. R., Ludwig, K. A., & Kozai, T. D. Y. (2020) · Journal of Neuroscience Research · doi:10.1002/jnr.24676 · Cited by 50
Self-citationEngineering strategies towards overcoming bleeding and glial scar formation around neural probes
Otte, E., Vlachos, A., & Asplund, M. (2022) · Cell and Tissue Research · doi:10.1007/s00441-021-03567-9 · Cited by 54
In vivo Characterization of Amorphous Silicon Carbide As a Biomaterial for Chronic Neural Interfaces
Knaack, G. L., McHail, D. G., Borda, G., Koo, B., Peixoto, N., Cogan, S. F., Dumas, T. C., & Pancrazio, J. J. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00301 · Cited by 53
Autotaxin and lysophospholipids in rheumatoid arthritis.
Bourgoin, S. G., & Zhao, C. (2010) · PubMed · Cited by 49
In vivo spatiotemporal dynamics of astrocyte reactivity following neural electrode implantation
Savya, S. P., Li, F., Lam, S., Wellman, S. M., Stieger, K. C., Chen, K., Eles, J. R., & Kozai, T. D. Y. (2022) · Biomaterials · doi:10.1016/j.biomaterials.2022.121784 · Cited by 46
Self-citationSzymanski, L. J., Kellis, S., Liu, C. Y., Jones, K. T., Andersen, R. A., Commins, D., Lee, B., McCreery, D. B., & Miller, C. A. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac127e · Cited by 48
In Vivo Imaging of Microglia With Multiphoton Microscopy
Hierro‐Bujalance, C., Bacskai, B. J., & García-Alloza, M. (2018) · Frontiers in Aging Neuroscience · doi:10.3389/fnagi.2018.00218 · Cited by 50
Xu, H., Weltman, A., Scholten, K., Berger, T. W., Song, D. B., & Meng, E. F. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa9451 · Cited by 48
Bioactive Neuroelectronic Interfaces
Adewole, D. O., Serruya, M. D., Wolf, J. A., & Cullen, D. K. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00269 · Cited by 50
Wellman, S. M., Cambi, F., & Kozai, T. D. Y. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.08.046 · Cited by 43
Self-citationEles, J. R., & Kozai, T. D. Y. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.119767 · Cited by 44
Self-citationYang, Q., Wu, B., Eles, J. R., Vazquez, A. L., Kozai, T. D. Y., & Cui, X. T. (2020) · Advanced Biosystems · doi:10.1002/adbi.201900287 · Cited by 43
Self-citationLee, H. C., Gaire, J., Currlin, S., McDermott, M. D., Park, K., & Otto, K. J. (2017) · Frontiers in Neuroscience · doi:10.3389/fnins.2017.00513 · Cited by 40
Biohybrid neural interfaces: improving the biological integration of neural implants
Boulingre, M., Lara, R. P., & Green, R. A. (2023) · Chemical Communications · doi:10.1039/d3cc05006h · Cited by 41
Eles, J. R., Stieger, K. C., & Kozai, T. D. Y. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/abc29c · Cited by 35
Self-citationLi, F., Gallego, J., Tirko, N. N., Greaser, J., Bashe, D., Patel, R., Shaker, E., Valkenburg, G. E. V., Alsubhi, A. S., Wellman, S. M., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-49709-9 · Cited by 38
Self-citationBiocompatible reference electrodes to enhance chronic electrochemical signal fidelity in vivo
Seaton, B. T., & Heien, M. L. (2021) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-021-03640-w · Cited by 37
Potter–Baker, K. A., & Capadona, J. R. (2015) · ACS Macro Letters · doi:10.1021/mz500743a · Cited by 35
Gao, W., Yan, Z., Zhou, H., Xie, Y., Wang, H., Yang, J., Yu, J., Ni, C., Liu, P., Xie, M., et al. (2025) · Journal of Nanobiotechnology · doi:10.1186/s12951-025-03573-x · Cited by 38
Gilmour, A. D., Woolley, A. J., Poole-Warren, L. A., Thomson, C. E., & Green, R. A. (2016) · Biomaterials · doi:10.1016/j.biomaterials.2016.03.011 · Cited by 34
CHIME: CMOS-Hosted in vivo Microelectrodes for Massively Scalable Neuronal Recordings
Köllő, M., Rácz, R., Hanna, M.-E., Obaid, A., Angle, M. R., Wray, W., Kong, Y., Hierlemann, A. R., Müller, J., Melosh, N. A., et al. (2020) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.00834 · Cited by 31
High‐density implantable neural electrodes and chips for massive neural recordings
Wang, L., Suo, Y., Wang, J., Wang, X., Xue, K., An, J., Sun, X., Chen, Q., Tang, X., Zhao, Y., et al. (2024) · Brain‐X · doi:10.1002/brx2.65 · Cited by 31
Overcoming failure: improving acceptance and success of implanted neural interfaces
Dalrymple, A. N., Jones, S. T., Fallon, J. B., Shepherd, R. K., & Weber, D. J. (2025) · Bioelectronic Medicine · doi:10.1186/s42234-025-00168-7 · Cited by 38
Hammer, D. X., Lozzi, A., Abliz, E., Greenbaum, N., Agrawal, A., Krauthamer, V., & Welle, C. G. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.002823 · Cited by 32
Sommakia, S., Gaire, J., Rickus, J. L., & Otto, K. J. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00033 · Cited by 32
Insertional effect following electrode implantation: an underreported but important phenomenon
Hamani, C., Davidson, B., Lipsman, N., Abrahão, A., Nestor, S. M., Rabin, J. S., Giacobbe, P., Pagano, R. L., & Campos, A. C. P. (2024) · Brain Communications · doi:10.1093/braincomms/fcae093 · Cited by 32
Stocking, K. C., Vazquez, A. L., & Kozai, T. D. Y. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2889832 · Cited by 29
Self-citationUnderstanding the Role of Innate Immunity in the Response to Intracortical Microelectrodes
Hermann, J. K., & Capadona, J. R. (2018) · Critical Reviews in Biomedical Engineering · doi:10.1615/critrevbiomedeng.2018027166 · Cited by 31
Longitudinal neural and vascular structural dynamics produced by chronic microelectrode implantation
Welle, C. G., Gao, Y., Ye, M., Lozzi, A., Boretsky, A. R., Abliz, E., & Hammer, D. X. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.119831 · Cited by 27
Metadata
- Authors
- Kozai, T. D. Y., Vazquez, A. L., Weaver, C. L., Kim, S.-G., & Cui, X. T.
- Affiliations
- University of Pittsburgh (US)
- Corresponding author
- Kozai, Takashi D Yoshida
- Cui, X Tracy
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2012-10-17
- Volume
- 9
- Issue
- 6
- Pages
- 066001
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC3511663
- OpenAlex
- W2050826469
- Cited by
- 215
- References
- 85
- Citations by year
- 2026: 13 · 2025: 12 · 2024: 20 · 2023: 21 · 2022: 16 · 2021: 19 · 2020: 19 · 2019: 15 · 2018: 16 · 2017: 17 · 2016: 15 · 2015: 15 · 2014: 12 · 2013: 3 · 2012: 1
- FWCI
- 4.09
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institutes of Health
- National Institute of Neurological Disorders and Stroke (5R01NS44589, R01NS044589, R01NS062019, K01NS066131)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Neurology › Neuroinflammation and Neurodegeneration Mechanisms
- Health Sciences › Medicine › Clinical Neurology › Traumatic Brain Injury and Neurovascular Disturbances
- Keywords
- microglia, microglial activation, visual cortex, two-photon microscopy, neuroprosthetics
- MeSH terms
- Algorithms, Animals, Blood-Brain Barrier, Brain, Craniotomy, Data Interpretation, Statistical, Electrodes, Implanted, Image Processing, Computer-Assisted, Mice, Transgenic, Microelectrodes, Microscopy, Visual Cortex, Microglia, Photons, Green Fluorescent Proteins, Mice
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review