Effectiveness of Computer-Aided Detection in Community Mammography Practice
Open accessThe DOI matches its Crossref record; 35 of the 35 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Computer-aided detection (CAD) is applied during screening mammography for millions of US women annually, although it is uncertain whether CAD improves breast cancer detection when used by community radiologists. METHODS: We investigated the association between CAD use during film-screen screening mammography and specificity, sensitivity, positive predictive value, cancer detection rates, and prognostic characteristics of breast cancers (stage, size, and node involvement). Records from 684 956 women who received more than 1.6 million film-screen mammograms at Breast Cancer Surveillance Consortium facilities in seven states in the United States from 1998 to 2006 were analyzed. We used random-effects logistic regression to estimate associations between CAD and specificity (true-negative examinations among women without breast cancer), sensitivity (true-positive examinations among women with breast cancer diagnosed within 1 year of mammography), and positive predictive value (breast cancer diagnosed after positive mammograms) while adjusting for mammography registry, patient age, time since previous mammography, breast density, use of hormone replacement therapy, and year of examination (1998-2002 vs 2003-2006). All statistical tests were two-sided. RESULTS: Of 90 total facilities, 25 (27.8%) adopted CAD and used it for an average of 27.5 study months. In adjusted analyses, CAD use was associated with statistically significantly lower specificity (OR = 0.87, 95% confidence interval [CI] = 0.85 to 0.89, P < .001) and positive predictive value (OR = 0.89, 95% CI = 0.80 to 0.99, P = .03). A non-statistically significant increase in overall sensitivity with CAD (OR = 1.06, 95% CI = 0.84 to 1.33, P = .62) was attributed to increased sensitivity for ductal carcinoma in situ (OR = 1.55, 95% CI = 0.83 to 2.91; P = .17), although sensitivity for invasive cancer was similar with or without CAD (OR = 0.96, 95% CI = 0.75 to 1.24; P = .77). CAD was not associated with higher breast cancer detection rates or more favorable stage, size, or lymph node status of invasive breast cancer. CONCLUSION: CAD use during film-screen screening mammography in the United States is associated with decreased specificity but not with improvement in the detection rate or prognostic characteristics of invasive breast cancer.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(2017) · Springer eBooks · doi:10.1007/978-3-319-40618-3 · Cited by 17,242
Percentage Points for a Generalized ESD Many-Outlier Procedure
Rosner, B. (1983) · Technometrics · doi:10.1080/00401706.1983.10487848 · Cited by 1,268 · cited 1×
“We assessed the potential impact of outliers by creating scat-terplots and computing extreme studentized deviates of pre-to post-CAD changes in specificity and sensitivity within facilities (35), which suggested that one or both measures were outlying within four facilities.”
MethodCarney, P. A., Miglioretti, D. L., Yankaskas, B. C., Kerlikowske, K., Rosenberg, R., Rutter, C. M., Geller, B. M., Abraham, L. A., Taplin, S. H., Dignan, M., et al. (2003) · Annals of Internal Medicine · doi:10.7326/0003-4819-138-3-200302040-00008 · Cited by 1,142 · cited 1×
“We included patient variables associated with interpretive accuracy including age, breast density, and time since previous mammogram (27,28).”
MethodSelf-citationNational Use of Postmenopausal Hormone Therapy: Annual Trends and Response to Recent Evidence
Hersh, A. L., Stefanick, M. L., & Stafford, R. S. (2004) · JAMA · doi:10.1001/jama.291.1.47 · Cited by 983 · cited 1×
“In the Women’s Health Initiative, women randomly assigned to HT had nearly two-thirds greater cumulative breast biopsy risk (29), and reduced HT use after publication of the Women’s Health Initiative results in July 2002 may largely explain the observed reduction in biopsy recommendations (32).”
BackgroundThe Decrease in Breast-Cancer Incidence in 2003 in the United States
Ravdin, P. M., Cronin, K. A., Howlader, N., Berg, C. D., Chlebowski, R. T., Feuer, E. J., Edwards, B. K., & Berry, D. A. (2007) · New England Journal of Medicine · doi:10.1056/nejmsr070105 · Cited by 982 · cited 1×
“Nishikawa and Pesce (33) argue that studies with matched designs are the most accurate means of assessing CAD clinical impacts (ie, in which outcomes are compared before and after CAD application on the same mammogram) (34).”
BackgroundFreer, T. W., & Ulissey, M. J. (2001) · Radiology · doi:10.1148/radiol.2203001282 · Cited by 751
Influence of Computer-Aided Detection on Performance of Screening Mammography
Fenton, J. J., Taplin, S. H., Carney, P. A., Abraham, L., Sickles, E. A., D'Orsi, C., Berns, E. A., Cutter, G., Hendrick, R. E., Barlow, W. E., et al. (2007) · New England Journal of Medicine · doi:10.1056/nejmoa066099 · Cited by 619 · cited 8×
“In a study of CAD implementation from 1998 to 2002 within the national Breast Cancer Surveillance Consortium (BCSC), CAD was associated with decreased specificity and positive predictive value (PPV1), and increased recall and biopsy rates, but no difference in cancer detection rates was observed (10).”
BackgroundMethodResultInfluentialSelf-citationPotential Contribution of Computer-aided Detection to the Sensitivity of Screening Mammography
Warren Burhenne, L. J., Wood, S. A., D'Orsi, C. J., Feig, S. A., Kopans, D. B., O'Shaughnessy, K. F., Sickles, E. A., Tabar, L., Vyborny, C. J., & Castellino, R. A. (2000) · Radiology · doi:10.1148/radiology.215.2.r00ma15554 · Cited by 554 · cited 1×
“However, small retrospective studies suggest that the performance impacts of CAD are similar when used in digital (43–46) and filmscreen environments (8,11,19).”
BackgroundSelf-citationBreast Cancer after Use of Estrogen plus Progestin in Postmenopausal Women
Chlebowski, R. T., Kuller, L. H., Prentice, R. L., Stefanick, M. L., Manson, J. E., Gass, M., Aragaki, A. K., Ockene, J. K., Lane, D. S., Sarto, G. E., et al. (2009) · New England Journal of Medicine · doi:10.1056/nejmoa0807684 · Cited by 472
Birdwell, R. L., Ikeda, D. M., O’Shaughnessy, K. F., & Sickles, E. A. (2001) · Radiology · doi:10.1148/radiology.219.1.r01ap16192 · Cited by 428 · cited 2×
“However, small retrospective studies suggest that the performance impacts of CAD are similar when used in digital (43–46) and filmscreen environments (8,11,19).”
BackgroundPerformance Benchmarks for Screening Mammography
Rosenberg, R. D., Yankaskas, B. C., Abraham, L. A., Sickles, E. A., Lehman, C. D., Geller, B. M., Carney, P. A., Kerlikowske, K., Buist, D. S. M., Weaver, D. L., et al. (2006) · Radiology · doi:10.1148/radiol.2411051504 · Cited by 336 · cited 3×
“CAD is now applied to the large majority of screening mammograms in the United States with annual direct Medicare costs exceeding $30 million (2).”
BackgroundSelf-citationErnster, V. L., Barclay, J., Kerlikowske, K., Wilkie, H., & Ballard-Barbash, R. (2000) · Archives of Internal Medicine · doi:10.1001/archinte.160.7.953 · Cited by 298 · cited 1×
“Although our analyses may lack sufficient power to exclude a small benefit of CAD in terms of invasive breast cancer detection, the principal contribution of CAD may be increased detection for DCIS—a precancerous lesion with an illdefined long-term prognosis (39).”
BackgroundSingle Reading with Computer-Aided Detection for Screening Mammography
Gilbert, F. J., Astley, S. M., Gillan, M. G. C., Agbaje, O. F., Wallis, M. G., James, J., Boggis, C. R. M., & Duffy, S. W. (2008) · New England Journal of Medicine · doi:10.1056/nejmoa0803545 · Cited by 296 · cited 1×
“Variability in the impact of CAD on recall rates was also observed in a meta-analysis (17) and across three sites in a trial comparing mammogram interpretation by a single radiologist with CAD versus interpretation by two radiologists (double reading) (42).”
BackgroundScreening Mammograms: Interpretation with Computer-aided Detection—Prospective Evaluation
Morton, M. J., Whaley, D. H., Brandt, K. R., & Amrami, K. K. (2006) · Radiology · doi:10.1148/radiol.2392042121 · Cited by 218 · cited 1×
“However, matched studies typically impose the restriction that radiologists can only upgrade final BI-RADS assessments after viewing CAD output, resulting in the recall of women who would not have been recalled in the absence of CAD (11,13,36,37).”
BackgroundComputer-aided Detection with Screening Mammography in a University Hospital Setting
Birdwell, R. L., Bandodkar, P., & Ikeda, D. M. (2005) · Radiology · doi:10.1148/radiol.2362040864 · Cited by 187
Dean, J. C., & Ilvento, C. C. (2006) · American Journal of Roentgenology · doi:10.2214/ajr.05.0111 · Cited by 163 · cited 1×
“However, matched studies typically impose the restriction that radiologists can only upgrade final BI-RADS assessments after viewing CAD output, resulting in the recall of women who would not have been recalled in the absence of CAD (11,13,36,37).”
BackgroundChapter 7: The Wisconsin Breast Cancer Epidemiology Simulation Model
Fryback, D. G., Stout, N. K., Rosenberg, M. A., Trentham-Dietz, A., Kuruchittham, V., & Remington, P. L. (2006) · JNCI Monographs · doi:10.1093/jncimonographs/lgj007 · Cited by 163
Taylor, P. M., & POTTS, H. W. W. (2008) · European Journal of Cancer · doi:10.1016/j.ejca.2008.02.016 · Cited by 160 · cited 3×
“The current findings are consistent with a meta-analysis that suggests a modest increase in recall rates with CAD with little or no impact on cancer detection rates (17).”
BackgroundResultKerlikowske, K. M., Miglioretti, D. L., Buist, D. S. M., Walker, R. L., & Carney, P. A. (2007) · JNCI Journal of the National Cancer Institute · doi:10.1093/jnci/djm111 · Cited by 156 · cited 2×
“We also specified whether patients were using hormone therapy (HT) because HT has been associated with increased risk of abnormal mammography, breast biopsy, and breast cancer (29–31).”
BackgroundSelf-citationRosenberg, J., Chia, Y. L., & Plevritis, S. K. (2005) · Breast Cancer Research and Treatment · doi:10.1007/s10549-004-1470-1 · Cited by 152 · cited 1×
“On the basis of cancer characteristics associated with increased mortality (7) and decreased survival (25,26), we defined tumors (including both DCIS and invasive cancer) as advanced if they were designated stage II–IV, if tumors measured greater than 15 mm in diameter, or if lymph node involvement was observed.”
BackgroundGromet, M. (2008) · American Journal of Roentgenology · doi:10.2214/ajr.07.2812 · Cited by 144
Autier, P., Héry, C., Haukka, J. K., Boniol, M., & Byrnes, G. B. (2009) · Journal of Clinical Oncology · doi:10.1200/jco.2009.22.7041 · Cited by 128 · cited 3×
“These findings raise concerns that CAD, as currently implemented in clinical practice, may have little or no impact on breast cancer mortality, which may depend on earlier detection of invasive breast cancer (6,7).”
BackgroundProspective Assessment of Computer-Aided Detection in Interpretation of Screening Mammography
Ko, J. M., Nicholas, M. J., Mendel, J. B., & Slanetz, P. J. (2006) · American Journal of Roentgenology · doi:10.2214/ajr.05.1582 · Cited by 121 · cited 1×
“However, matched studies typically impose the restriction that radiologists can only upgrade final BI-RADS assessments after viewing CAD output, resulting in the recall of women who would not have been recalled in the absence of CAD (11,13,36,37).”
BackgroundDuffy, S. W., Tabár, L. K., Viták, B., Day, N., Smith, R. A., Chen, H. H. T., & Yen, M. F. A. (2003) · European Journal of Cancer · doi:10.1016/s0959-8049(03)00259-4 · Cited by 108 · cited 1×
“These findings raise concerns that CAD, as currently implemented in clinical practice, may have little or no impact on breast cancer mortality, which may depend on earlier detection of invasive breast cancer (6,7).”
BackgroundEstrogen Plus Progestin and Breast Cancer Detection by Means of Mammography and Breast Biopsy
Chlebowski, R. T. (2008) · Archives of Internal Medicine · doi:10.1001/archinternmed.2007.123 · Cited by 107 · cited 1×
“In the Women’s Health Initiative, women randomly assigned to HT had nearly two-thirds greater cumulative breast biopsy risk (29), and reduced HT use after publication of the Women’s Health Initiative results in July 2002 may largely explain the observed reduction in biopsy recommendations (32).”
BackgroundHow Widely Is Computer-Aided Detection Used in Screening and Diagnostic Mammography?
Rao, V. M., Levin, D. C., Parker, L., Cavanaugh, B. C., Frangos, A. J., & Sunshine, J. H. (2010) · Journal of the American College of Radiology · doi:10.1016/j.jacr.2010.05.019 · Cited by 103 · cited 2×
“Congress subsequently amended the Medicare statute to include supplemental coverage for CAD use, and CAD is now applied on nearly three of four screening mammograms performed on American women (2).”
BackgroundMethodCan Computer-aided Detection Be Detrimental to Mammographic Interpretation?
Philpotts, L. E. (2009) · Radiology · doi:10.1148/radiol.2531090689 · Cited by 97 · cited 2×
“In contrast, this study assesses the effectiveness of CAD use in actual everyday practice conditions, in which radiologists with variable experience and expertise may use CAD in a nonstandardized idiosyncratic fashion (5).”
BackgroundAssociation between Mammography Timing and Measures of Screening Performance in the United States
Yankaskas, B. C., Taplin, S. H., Ichikawa, L., Geller, B. M., Rosenberg, R. D., Carney, P. A., Kerlikowske, K. M., Ballard-Barbash, R. R., Cutter, G., & Barlow, W. E. (2005) · Radiology · doi:10.1148/radiol.2342040048 · Cited by 95 · cited 1×
“We included patient variables associated with interpretive accuracy including age, breast density, and time since previous mammogram (27,28).”
MethodSelf-citationYang, S. K., Moon, W. K., Cho, N., Park, J. S., Cha, J. H., Kim, S. M., Kim, S. J., & Im, J.-G. (2007) · Radiology · doi:10.1148/radiol.2441060756 · Cited by 78
The Preponderance of Evidence Supports Computer-aided Detection for Screening Mammography
Birdwell, R. L. (2009) · Radiology · doi:10.1148/radiol.2531090611 · Cited by 63 · cited 1×
“Despite broad acceptance and use, it is unclear if the benefits of CAD during screening mammography outweigh its potential risks and costs (4,5).”
BackgroundNoble, M., Bruening, W., Uhl, S., & Schoelles, K. M. (2008) · Archives of Gynecology and Obstetrics · doi:10.1007/s00404-008-0841-y · Cited by 52
Taplin, S. H., Rutter, C. M., & Lehman, C. D. (2006) · American Journal of Roentgenology · doi:10.2214/ajr.05.0940 · Cited by 52
Self-citationSkaane, P., Kshirsagar, A., Stapleton, S. J., Young, K., & Castellino, R. A. (2007) · American Journal of Roentgenology · doi:10.2214/ajr.05.2207 · Cited by 52
Alberdi, E., Povyakalo, A. A., Strigini, L., Ayton, P., Hartswood, M., Procter, R., & Slack, R. (2005) · The British Journal of Radiology · doi:10.1259/bjr/37646417 · Cited by 48 · cited 1×
“In contrast, CAD is less likely to mark invasive cancers presenting as uncalcified masses (8,9) and may therefore have little or no impact on the detection of higher-risk invasive breast cancer (14,15).”
BackgroundBreast Cancer Detection Rate: Designing Imaging Trials to Demonstrate Improvements
Jiang, Y., Miglioretti, D. L., Metz, C. E., & Schmidt, R. A. (2007) · Radiology · doi:10.1148/radiol.2432060253 · Cited by 35 · cited 1×
“These outcomes may be impossible to assess with adequate statistical power within matched studies or even randomized trials (38).”
Background
Cited by
Large scale deep learning for computer aided detection of mammographic lesions
Kooi, T., Litjens, G., van Ginneken, B., Gubern-Mérida, A., Sánchez, C. I., Mann, R., den Heeten, A., & Karssemeijer, N. (2017) · Medical Image Analysis · doi:10.1016/j.media.2016.07.007 · Cited by 965
Deep Learning in Medical Image Analysis
Chan, H.-P., Samala, R. K., Hadjiiski, L. M., & Zhou, C. (2020) · Advances in Experimental Medicine and Biology · doi:10.1007/978-3-030-33128-3_1 · Cited by 747
Detecting and classifying lesions in mammograms with Deep Learning
Ribli, D., Horváth, A., Unger, Z., Pollner, P., & Csabai, I. (2018) · Scientific Reports · doi:10.1038/s41598-018-22437-z · Cited by 696
Diagnostic Accuracy of Digital Screening Mammography With and Without Computer-Aided Detection
Lehman, C. D., Wellman, R. D., Buist, D. S. M., Kerlikowske, K., Tosteson, A. N. A., & Miglioretti, D. L. (2015) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2015.5231 · Cited by 678
Kim, H.-E., Kim, H. H., Han, B.-K., Kim, K. H., Han, K., Nam, H., Lee, E. H., & Kim, E.-K. (2020) · The Lancet Digital Health · doi:10.1016/s2589-7500(20)30003-0 · Cited by 525
Metadata
- Authors
- Fenton, J. J., Abraham, L., Taplin, S. H., Geller, B. M., Carney, P. A., D'Orsi, C., Elmore, J. G., Barlow, W. E., & for the Breast Cancer Surveillance Consortium
- Venue
- JNCI Journal of the National Cancer Institute
- Publisher
- Oxford University Press (OUP)
- Published
- 2011-07-27
- Volume
- 103
- Issue
- 15
- Pages
- 1152-1161
- ISSN
- 0027-8874 (ISSN-L)
- 1460-2105
- OpenAlex
- W2095972041
- Cited by
- 168
- References
- 35
- FWCI
- 10.64
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.