Practice Parameters for the Use of Autotitrating Continuous Positive Airway Pressure Devices for Titrating Pressures and Treating Adult Patients with Obstructive Sleep Apnea Syndrome: An Update for 2007
Open accessAbstract
These practice parameters are an update of the previously published recommendations regarding the use of autotitrating positive airway pressure (APAP) devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome. Continuous positive airway pressure (CPAP) at an effective setting verified by attended polysomnography is a standard treatment for obstructive sleep apnea (OSA). APAP devices change the treatment pressure based on feedback from various patient measures such as airflow, pressure fluctuations, or measures of airway resistance. These devices may aid in the pressure titration process, address possible changes in pressure requirements throughout a given night and from night to night, aid in treatment of OSA when attended CPAP titration has not or cannot be accomplished, or improve patient comfort. A task force of the Standards of Practice Committee of the American Academy of Sleep Medicine has reviewed the literature published since the 2002 practice parameter on the use of APAP. Current recommendations follow: (1) APAP devices are not recommended to diagnose OSA; (2) patients with congestive heart failure, patients with significant lung disease such as chronic obstructive pulmonary disease; patients expected to have nocturnal arterial oxyhemoglobin desaturation due to conditions other than OSA (e.g., obesity hypoventilation syndrome); patients who do not snore (either naturally or as a result of palate surgery); and patients who have central sleep apnea syndromes are not currently candidates for APAP titration or treatment; (3) APAP devices are not currently recommended for split-night titration; (4) certain APAP devices may be used during attended titration with polysomnography to identify a single pressure for use with standard CPAP for treatment of moderate to severe OSA; (5) certain APAP devices may be initiated and used in the self-adjusting mode for unattended treatment of patients with moderate to severe OSA without significant comorbidities (CHF, COPD, central sleep apnea syndromes, or hypoventilation syndromes); (6) certain APAP devices may be used in an unattended way to determine a fixed CPAP treatment pressure for patients with moderate to severe OSA without significant comorbidities (CHF, COPD, central sleep apnea syndromes, or hypoventilation syndromes); (7) patients being treated with fixed CPAP on the basis of APAP titration or being treated with APAP must have close clinical follow-up to determine treatment effectiveness and safety; and (8) a reevaluation and, if necessary, a standard attended CPAP titration should be performed if symptoms do not resolve or the APAP treatment otherwise appears to lack efficacy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Kushida, C. A., Littner, M. R., Morgenthaler, T., Alessi, C. A., Bailey, D., Coleman, J., Friedman, L., Hirshkowitz, M., Kapen, S., Kramer, M., et al. (2005) · Sleep · doi:10.1093/sleep/28.4.499 · Cited by 2,054
Self-citationGrading Strength of Recommendations and Quality of Evidence in Clinical Guidelines
Guyatt, G., Gutterman, D., Baumann, M. H., Addrizzo-Harris, D., Hylek, E. M., Phillips, B., Raskob, G., Lewis, S. Z., & Schünemann, H. (2006) · Chest · doi:10.1378/chest.129.1.174 · Cited by 1,220
Littner, M. R., Kushida, C., Wise, M., G. Davila, D., Morgenthaler, T., Lee-Chiong, T., Hirshkowitz, M., Loube, D. L., Bailey, D., Berry, R. B., et al. (2005) · Sleep · doi:10.1093/sleep/28.1.113 · Cited by 1,142
Self-citationKushida, C. A., Littner, M. R., Hirshkowitz, M., Morgenthaler, T. I., Alessi, C. A., Bailey, D., Boehlecke, B., Brown, T. M., Coleman, J., Friedman, L., et al. (2006) · Sleep · doi:10.1093/sleep/29.3.375 · Cited by 775
Self-citationEvaluation of Positive Airway Pressure Treatment for Sleep Related Breathing Disorders in Adults
Gay, P., Weaver, T., Loube, D., & Iber, C. (2006) · Sleep · doi:10.1093/sleep/29.3.381 · Cited by 614
Alternative Methods of Titrating Continuous Positive Airway Pressure: A Large Multicenter Study
Masa, J. F., Jiménez, A., Durán, J., Capote, F., Monasterio, C., Mayos, M., Terán, J., Hernández, L., Barbé, F., Maimó, A., et al. (2004) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200312-1787oc · Cited by 328
Practice Parameters for Using Polysomnography to Evaluate Insomnia: An Update
Littner, M., Hirshkowitz, M., Kramer, M., Kapen, S., Anderson, W. M., Bailey, D., Berry, R. B., Davila, D., Johnson, S., Kushida, C., et al. (2003) · Sleep · doi:10.1093/sleep/26.6.754 · Cited by 306
Ayas, N. T., Patel, S. R., Malhotra, A., Schulzer, M., Malhotra, M., Jung, D., Fleetham, J., & White, D. P. (2004) · Sleep · doi:10.1093/sleep/27.2.249 · Cited by 255
Ryan, C. F., Lowe, L. A., Li, D., & Fleetham, J. A. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.4.939 · Cited by 247
A manual for assessing health practices & designing practice policies : the explicit approach
Eddy, D. M. (1992) · Medical Entomology and Zoology · Cited by 218
Littner, M., Hirshkowitz, M., Davila, D., Anderson, W. M., Kushida, C. A., Woodson, B. T., Johnson, S. F., & Wise, M. S. (2002) · Sleep · doi:10.1093/sleep/25.2.143 · Cited by 183
Comparison between Automatic and Fixed Positive Airway Pressure Therapy in the Home
Massie, C. A., McArdle, N., Hart, R. W., Schmidt-Nowara, W. W., Lankford, A., Hudgel, D. W., Gordon, N., & Douglas, N. J. (2003) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200201-022oc · Cited by 177
Autoadjusting CPAP Therapy Based on Impedance Efficacy, Compliance and Acceptance
RANDERATH, W. J., SCHRAEDER, O., GALETKE, W., FELDMEYER, F., & RÜHLE, K.-H. (2001) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.163.3.2006168 · Cited by 143
Massie, C. A., & Hart, R. W. (2003) · Chest · doi:10.1378/chest.123.4.1112 · Cited by 131
Rules of evidence and clinical recommendations for the management of patients.
Sackett, D. L. (1993) · PubMed · Cited by 130
Comparison of Manual and Automatic CPAP Titration in Patients With Sleep Apnea/Hypopnea Syndrome
Lloberes, P., Ballester, E., Montserrat, J. M., Botifoll, E., Ramirez, A., Reolid, A., Gistau, C., & Rodriguez‐Roisin, R. (1996) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.154.6.8970366 · Cited by 117
Hukins, C. (2004) · Sleep · doi:10.1093/sleep/27.8.1512 · Cited by 100
West, S. D. (2006) · Thorax · doi:10.1136/thx.2005.046300 · Cited by 93
Stroke and treatment with nasal CPAP
Palombini, L., & Guilleminault, C. (2006) · European Journal of Neurology · doi:10.1111/j.1468-1331.2006.01169.x · Cited by 87
Accuracy of an Unattended Home CPAP Titration in the Treatment of Obstructive Sleep Apnea
Sériès, F. (2000) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.162.1.9908023 · Cited by 85
Senn, O., Brack, T., Matthews, F., Russi, E. W., & Bloch, K. E. (2003) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200304-542oc · Cited by 73
Efficacy and Cost of Home-Initiated Auto-nCPAP versus Conventional nCPAP
Planès, C., D'Ortho, M.-P., Foucher, A., Berkani, M., Leroux, K., Essalhi, M., Delclaux, C., Quera-Salva, M.-A., & Lofaso, F. (2003) · Sleep · doi:10.1093/sleep/26.2.156 · Cited by 68
Noseda, A., Kempenaers, C., Kerkhofs, M., Braun, S., Linkowski, P., & Jann, E. (2004) · Chest · doi:10.1378/chest.126.1.31 · Cited by 63
Nussbaumer, Y., Bloch, K. E., Genser, T., & Thurnheer, R. (2006) · Chest · doi:10.1378/chest.129.3.638 · Cited by 63
The Effect of Nightly Nasal CPAP Treatment on Underlying Obstructive Sleep Apnea and Pharyngeal Size
Collop, N. A., Block, A. J., & Hellard, D. W. (1991) · CHEST Journal · doi:10.1378/chest.99.4.855 · Cited by 58
Stammnitz, A., Jerrentrup, A., Penzel, T., Peter, J. H., Vogelmeier, C. F., & Becker, H. F. (2004) · European Respiratory Journal · doi:10.1183/09031936.04.00074304 · Cited by 55
Berkani, M., Lofaso, F., CHOUAID, C., d’Ortho, M., Theret, D., Grillier-Lanoir, V., HARF, A., & HOUSSET, B. (1998) · European Respiratory Journal · doi:10.1183/09031936.98.12040759 · Cited by 53
Unattended Home Diagnosis and Treatment of Obstructive Sleep Apnea Without Polysomnography
Fletcher, E. C., Stich, J., & Yang, K. L. (2000) · Archives of Family Medicine · doi:10.1001/archfami.9.2.168 · Cited by 43
Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea
Nolan, G. M., Ryan, S., O’Connor, T. M., & McNicholas, W. T. (2006) · European Respiratory Journal · doi:10.1183/09031936.06.00127205 · Cited by 42
Kessler, R., Weitzenblum, E., Chaouat, A., Iamandi, C., & Alliotte, T. (2003) · CHEST Journal · doi:10.1378/chest.123.3.704 · Cited by 42
Preference for fixed or automatic CPAP in patients with obstructive sleep apnea syndrome
MARRONE, O. (2004) · Sleep Medicine · doi:10.1016/j.sleep.2003.09.011 · Cited by 41
Mortimore, I. L., Kochhar, P., & Douglas, N. J. (1996) · Thorax · doi:10.1136/thx.51.2.190 · Cited by 39
Lloberes, P., Rodríguez, B., Roca, A. P., Sagalés, M., de la Calzada, M. D., Giménez, S., Romero, O., & Sampol, G. (2004) · Respiratory Medicine · doi:10.1016/j.rmed.2003.12.014 · Cited by 38
Efficacy of Flow- vs Impedance-Guided Autoadjustable Continuous Positive Airway Pressure
Pevernagie, D., Proot, P. M., Hertegonne, K. B. C., Neyens, M. C., Hoornaert, K., & PAUWELS, R. A. (2004) · CHEST Journal · doi:10.1378/chest.126.1.25 · Cited by 38
Randerath, W. J., Galetke, W., & Ruhle, K.-H. (2003) · PubMed · Cited by 34
Effects of fixed compared to automatic CPAP on sleep in Obstructive Sleep Apnoea Syndrome
Resta, O., Carratù, P., Depalo, A., Giliberti, T., Ardito, M., Marrone, O., & Insalaco, G. (2004) · Monaldi Archives for Chest Disease · doi:10.4081/monaldi.2004.694 · Cited by 23
Krieger, J. (1999) · Sleep Medicine Reviews · doi:10.1016/s1087-0792(99)90022-9 · Cited by 21
Hertegonne, K. B. C., Proot, P. M., PAUWELS, R. A., & Pevernagie, D. (2003) · Respiratory Medicine · doi:10.1016/s0954-6111(03)00113-6 · Cited by 19
Cited by
Adult Obstructive Sleep Apnea Task Force of the American Academy of Sleep Medicine, Epstein, L. J., Kristo, D., Strollo, P. J., Friedman, N., Malhotra, A., Patil, S. P., Ramar, K., Rogers, R., Schwab, R. J., et al. (2009) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.27497 · Cited by 3,360
Collop, N. A., Anderson, W. M., Boehlecke, B., Claman, D., Goldberg, R., Gottlieb, D. J., Hudgel, D., Sateia, M., Schwab, R., & Portable Monitoring Task Force of the American Academy of Sleep Medicine (2007) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.27032 · Cited by 1,550
Self-citationPatil, S. P., Ayappa, I. A., Caples, S. M., Kimoff, R. J., Patel, S. R., & Harrod, C. G. (2019) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.7640 · Cited by 881
Trends in CPAP adherence over twenty years of data collection: A flattened curve
Rotenberg, B. W., Murariu, D., & Pang, K. P. (2016) · Journal of Otolaryngology - Head & Neck Surgery · doi:10.1186/s40463-016-0156-0 · Cited by 742
Patil, S. P., Ayappa, I. A., Caples, S. M., Kimoff, R. J., Patel, S. R., & Harrod, C. G. (2019) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.7638 · Cited by 721
Metadata
- Authors
- Morgenthaler, T. I., Aurora, R. N., Brown, T., Zak, R., Alessi, C., Boehlecke, B., Chesson, A. L., Friedman, L., Kapur, V., Maganti, R., Owens, J., Pancer, J., Swick, T. J., & Standards of Practice Committee of the AASM
- Venue
- Sleep
- Publisher
- Oxford University Press (OUP)
- Published
- 2008-01-01
- Volume
- 31
- Issue
- 1
- Pages
- 141-147
- ISSN
- 0161-8105 (ISSN-L)
- 1550-9109
- OpenAlex
- W2278194140
- Cited by
- 334
- References
- 39
- FWCI
- 17.69
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.