Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
Background: It is uncertain whether skin pigmentation affects pulse oximeter accuracy at low HbO2 saturation. Methods: The accuracy of finger pulse oximeters during stable, plateau levels of arterial oxygen saturation (Sao2) between 60 and 100% were evaluated in 11 subjects with darkly pigmented skin and in 10 with light skin pigmentation. Oximeters tested were the Nellcor N-595 with the OxiMax-A probe (Nellcor Inc., Pleasanton, CA), the Novametrix 513 (Novametrix Inc., Wallingford, CT), and the Nonin Onyx (Nonin Inc., Plymouth, MN). Semisupine subjects breathed air-nitrogen-carbon dioxide mixtures through a mouthpiece. A computer used end-tidal oxygen and carbon dioxide concentrations determined by mass spectrometry to estimate breath-by-breath Sao2, from which an operator adjusted inspired gas to rapidly achieve 2- to 3-min stable plateaus of desaturation. Comparisons of oxygen saturation measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM3) were used in a multivariate model to determine the interrelation between saturation, skin pigmentation, and oximeter bias (Spo2 - Sao2). Results: At 60-70% Sao2, Spo2 (mean of three oximeters) overestimated Sao2 (bias +/- SD) by 3.56 +/- 2.45% (n = 29) in darkly pigmented subjects, compared with 0.37 +/- 3.20% (n = 58) in lightly pigmented subjects (P < 0.0001). The SD of bias was not greater with dark than light skin. The dark-light skin differences at 60-70% Sao2 were 2.35% (Nonin), 3.38% (Novametrix), and 4.30% (Nellcor). Skin pigment-related differences were significant with Nonin below 70% Sao2, with Novametrix below 90%, and with Nellcor at all ranges. Pigment-related bias increased approximately in proportion to desaturation. Conclusions: The three tested pulse oximeters overestimated arterial oxygen saturation during hypoxia in dark-skinned individuals.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Related work on pulse oximetry, which relies on the same optical principles, has also documented skin-pigmentation-linked measurement errors [14,15]”
From PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects (2026) · cites this paper as [14]
In the citing paper
“Notably, darker skin pigmentation may influence PPG-based SpO2 and PR readings [7-10].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [10]
In the citing paper
“Measurements of the pulse oximeter are known to be affected by the patient’s skin color, as the device systematically overestimates oxygen saturation levels in nonwhite patients.”
From Addressing bias in big data and AI for health care: A call for open science (2021) · cites this paper as [2]
In the citing paper
“Accuracy errors in pulse oximeter readings have been associated with SpO2 overestimations, especially at low saturation conditions [23,24,73,93].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [23]
In the citing paper
“The effect of skin pigmentation in pulse oximeter readings under hypoxia conditions was tested in 11 dark-skinned male subjects and 10 light-skinned males, who were made to breathe an air mixture (nitrogen-carbon monoxide). The three pulse oximeters used showed overestimation in dark-skinned individuals that increases linearly as SpO2 levels decreases [23].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [23]
3 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 333 · cited 1×
“Major known variability is caused by anemia, light scattering, venous and tissue pulsation by mechanical force from nearby arteries, pulsatile variations in tissue thickness in the light path other than in the arteries, nail polish, and skin pigment.(3) Therefore, pulse oximeter design is often based on empirically de…”
BackgroundSelf-citationJubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296
Errors in 14 pulse oximeters during profound hypoxia
Severinghaus, J. W., Naifeh, K. H., & Koh, S. O. (1989) · Journal of Clinical Monitoring · doi:10.1007/bf01617877 · Cited by 217
Self-citationEffects of Intravenously Administered Dyes on Pulse Oximetry Readings
SCHELLER, M. S., UNGER, R. J., & KELNER, M. J. (1986) · Anesthesiology · doi:10.1097/00000542-198611000-00023 · Cited by 194 · cited 1×
“Intravenously injected dyes cause transient errors.(13) Lee et al.”
BackgroundRALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11501.x · Cited by 192 · cited 2×
“%) in black patients.4–7 Model simulations of errors due to various pigments were reviewed by Ralston et al.8 Cote et al.9 reported that nail polish and ink on skin surface can cause errors, a finding confirmed anecdotally by others from fingerprinting ink,10 henna,11 and meconium.12 Intravenously injected dyes cause…”
The Effect of Nail Polish on Pulse Oximetry
Cot??, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 172 · cited 2×
“%) in black patients.4–7 Model simulations of errors due to various pigments were reviewed by Ralston et al.8 Cote et al.9 reported that nail polish and ink on skin surface can cause errors, a finding confirmed anecdotally by others from fingerprinting ink,10 henna,11 and meconium.12 Intravenously injected dyes cause…”
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50
Cited by
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114
Addressing bias in big data and AI for health care: A call for open science
Norori, N., Hu, Q., Aellen, F. M., Faraci, F. D., & Tzovara, A. (2021) · Patterns · doi:10.1016/j.patter.2021.100347 · Cited by 903
“Measurements of the pulse oximeter are known to be affected by the patient’s skin color, as the device systematically overestimates oxygen saturation levels in nonwhite patients.(2) As a result, Black patients are”
BackgroundBTS guideline for emergency oxygen use in adult patients
O'Driscoll, B. R., Howard, L. S., Davison, A. G., & on behalf of the British Thoracic Society (2008) · Thorax · doi:10.1136/thx.2008.102947 · Cited by 729
BTS guideline for oxygen use in adults in healthcare and emergency settings
O'Driscoll, B. R., Howard, L. S., Earis, J., & Mak, V (2017) · Thorax · doi:10.1136/thoraxjnl-2016-209729 · Cited by 679
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673
Human Identification Using Finger Images
Kumar, A., & Zhou, Y. (2012) · IEEE Transactions on Image Processing · doi:10.1109/tip.2011.2171697 · Cited by 624
“[31-32].”
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400
“This study and a previous study (3) have found that”
BackgroundResultSelf-citationMahle, W. T., Newburger, J. W., Matherne, G. P., Smith, F. C., Hoke, T. R., Koppel, R., Gidding, S. S., Beekman, R. H., Grosse, S. D., & on behalf of the American Heart Association Congenital Heart Defects Committee of the Council on Cardiovascular Disease in the Young, C. O. C. N. (2009) · Pediatrics · doi:10.1542/peds.2009-1397 · Cited by 346
Fawzy, A., Wu, T. D., Wang, K., Robinson, M. L., Farha, J., Bradke, A., Golden, S. H., Xu, Y., & Garibaldi, B. T. (2022) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2022.1906 · Cited by 326
“3%) had a total of 32 282 concurrently measured SpO2 and SaO2 levels during their hospital encounter, with a median (IQR) of 10 (2-30) measurements per patient.”
BackgroundMahle, W. T., Newburger, J. W., Matherne, G. P., Smith, F. C., Hoke, T. R., Koppel, R., Gidding, S. S., Beekman, R. H., & Grosse, S. D. (2009) · Circulation · doi:10.1161/circulationaha.109.192576 · Cited by 286
Luks, A. M., & Swenson, E. R. (2020) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202005-418fr · Cited by 263
Wong, A.-K. I., Charpignon, M., Kim, H., Josef, C., de Hond, A. A. H., Fojas, J. J., Tabaie, A., Liu, X., Mireles-Cabodevila, E., Carvalho, L., et al. (2021) · JAMA Network Open · doi:10.1001/jamanetworkopen.2021.31674 · Cited by 268
Jarzyna, D., Jungquist, C. R., Pasero, C., Willens, J. S., Nisbet, A., Oakes, L., Dempsey, S. J., Santangelo, D., & Polomano, R. C. (2011) · Pain Management Nursing · doi:10.1016/j.pmn.2011.06.008 · Cited by 253
New principle for measuring arterial blood oxygenation, enabling motion-robust remote monitoring
van Gastel, M., Stuijk, S., & de Haan, G. (2016) · Scientific Reports · doi:10.1038/srep38609 · Cited by 253
Vashist, S., Schneider, E., & Luong, J. (2014) · Diagnostics · doi:10.3390/diagnostics4030104 · Cited by 244
“[36], has more than two decades of experience in the manufacturing of pulse oximeters with the desired clinical accuracy [87–89].”
BackgroundThe 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 187
Pulse Oximetry in Pediatric Practice
Fouzas, S., Priftis, K. N., & Anthracopoulos, M. B. (2011) · Pediatrics · doi:10.1542/peds.2011-0271 · Cited by 219
Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers
Bickler, P. E., Feiner, J. R., & Rollins, M. D. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e318297d763 · Cited by 222
Self-citationBrown, S. M., Grissom, C. K., Moss, M., Rice, T. W., Schoenfeld, D., Hou, P. C., Thompson, B. T., & Brower, R. G. (2016) · CHEST · doi:10.1016/j.chest.2016.01.003 · Cited by 205
Does “ AI ” stand for augmenting inequality in the era of covid-19 healthcare?
Leslie, D., Mazumder, A., Peppin, A., Wolters, M. K., & Hagerty, A. (2021) · BMJ · doi:10.1136/bmj.n304 · Cited by 177
Camera Measurement of Physiological Vital Signs
McDuff, D. (2023) · ACM Computing Surveys · doi:10.1145/3558518 · Cited by 159
“Evidence of biases in SpO2 measurement with skin type is prevalent, with three monitors tested overestimating oxygen saturation in darker skin types in adults [16] and infants [155].”
BackgroundAn introduction to digital determinants of health
Chidambaram, S., Jain, B., Jain, U., Mwavu, R., Baru, R., Thomas, B., Greaves, F., Jayakumar, S., Jain, P., Rojo, M., et al. (2024) · PLOS Digital Health · doi:10.1371/journal.pdig.0000346 · Cited by 175
Chen, W., Janz, D. R., Shaver, C. M., Bernard, G. R., Bastarache, J. A., & Ware, L. B. (2015) · Chest · doi:10.1378/chest.15-0169 · Cited by 171
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158
“75%) to be at unclear risk of bias for all four domains or high risk of bias for at least one domain,[34,38,44,47,48,53,54,56-61,63] and the remaining 18 (56.”
BackgroundInfluentialRemote management of covid-19 using home pulse oximetry and virtual ward support
Greenhalgh, T., Knight, M., Inada-Kim, M., Fulop, N. J., Leach, J., & Vindrola-Padros, C. (2021) · BMJ · doi:10.1136/bmj.n677 · Cited by 156
Noncontact Monitoring of Blood Oxygen Saturation Using Camera and Dual-Wavelength Imaging System
Shao, D., Liu, C., Tsow, F., Yang, Y., Du, Z., Iriya, R., Yu, H., & Tao, N. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2481896 · Cited by 150
“pigmentation on fingers, for whom traditional pulse oximetry may otherwise lead to inaccurate measurements [8].”
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155
“oxygen saturation levels were over-estimated during hypoxia for patients with higher skin melanin content [46].”
BackgroundHenry, N. R., Hanson, A. C., Schulte, P. J., Warner, N. S., Manento, M. N., Weister, T. J., & Warner, M. A. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005394 · Cited by 133
“Given the physiological and biomechanical underpinnings of pulse oximetry, observed racial differences in hypoxemia detection are most plausibly related to variations in skin pigmentation, which may impact light absorption and scattering in the epidermis, resulting in differential transmission to the photodetector, as…”
BackgroundAccuracy of Pulse Oximetry in Children
Ross, P. A., Newth, C. J. L., & Khemani, R. G. (2014) · Pediatrics · doi:10.1542/peds.2013-1760 · Cited by 129
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124
A New Antispoofing Approach for Biometric Devices
Reddy, P. V., Kumar, A., Rahman, S. M. K., & Mundra, T. S. (2008) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.2003432 · Cited by 123
Brown, S. M., Duggal, A., Hou, P. C., Tidswell, M., Khan, A., Exline, M. C., Park, P. K., Schoenfeld, D., Liu, M., Grissom, C. K., et al. (2017) · Critical Care Medicine · doi:10.1097/ccm.0000000000002514 · Cited by 122
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120
“Studies of adults have reported mixed findings; some have identified overestimation of oxygen saturation by 2–10% in participants with greater skin pigmentation [41, 42].”
BackgroundPulse Oximetry at High Altitude
Luks, A. M., & Swenson, E. R. (2011) · High Altitude Medicine & Biology · doi:10.1089/ham.2011.0013 · Cited by 120
“These accuracy problems may be exacerbated in dark-skinned individuals traveling at high altitude; it has been reported that several pulse oximeter models overestimate arterial oxygen saturation with exposure of such individuals to hypoxia (Bickler et al., 2005; Feiner et al. 2007).”
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119
Valbuena, V. S. M., Barbaro, R. P., Claar, D., Valley, T. S., Dickson, R. P., Gay, S. E., Sjoding, M. W., & Iwashyna, T. J. (2022) · Chest · doi:10.1016/j.chest.2021.09.025 · Cited by 119
Respiratory Monitoring: Current State of the Art and Future Roads
Costanzo, I., Sen, D., Rhein, L., & Güler, Ü. (2020) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2020.3036330 · Cited by 119
Four Types of Pulse Oximeters Accurately Detect Hypoxia during Low Perfusion and Motion
Louie, A., Feiner, J. R., Bickler, P. E., Rhodes, L., Bernstein, M., & Lucero, J. (2018) · Anesthesiology · doi:10.1097/aln.0000000000002002 · Cited by 106
Self-citationLipnick, M. S., Feiner, J. R., Au, P., Bernstein, M., & Bickler, P. E. (2016) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001300 · Cited by 100
Self-citationPulse oximetry in the oesophagus
Kyriacou, P. A. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/1/r01 · Cited by 90
“Skin pigmentation has shown variable effects on pulse oximetry (Volgyesi and Spahr-Schopfer 1991, Bickler et al 2005).”
The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2016.11.043 · Cited by 96
“Studies of adults have demonstrated that pulse oximetry oxygen saturation (SpO2) consistently overestimates measured SaO2 in individuals with hypoxemia who have dark skin.(2,3) The impact of skin pigmentation on the accuracy of pulse oximetry in infants with hypoxemia is unknown.”
BackgroundPulse Oximeter Performance, Racial Inequity, and the Work Ahead
Okunlola, O. E., Lipnick, M. S., Batchelder, P. B., Bernstein, M., Feiner, J. R., & Bickler, P. E. (2022) · Respiratory Care · doi:10.4187/respcare.09795 · Cited by 96
“The first is that a small positive bias in oximeter readings still exists in individuals with dark skin pigmentation, similarly observed by Bickler and Feiner in 2005 and 2007.(4,6) The second is that the paucity of data points within the red rectangle in the Clinimark laboratory data, as compared to the number of dat…”
BackgroundResultSelf-citationRacial Disparity in Oxygen Saturation Measurements by Pulse Oximetry: Evidence and Implications
Jamali, H., Castillo, L. T., Morgan, C. C., Coult, J., Muhammad, J. L., Osobamiro, O. O., Parsons, E. C., & Adamson, R. (2022) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202203-270cme · Cited by 95
Monitoring in the Intensive Care
Kipnis, E., Ramsingh, D., Bhargava, M., Dincer, E., Cannesson, M., Broccard, A., Vallet, B., Bendjelid, K., & Thibault, R. (2012) · Critical Care Research and Practice · doi:10.1155/2012/473507 · Cited by 87
“False readings are also seen in severe anemia (Hb < 5 g/dL) [128], in presence of excessive skin pigmentation [115, 129], nail polish [119], or dyes.”
Martín, D., Johns, C., Sorrell, L., Healy, E., Phull, M., Olusanya, O., Peters, M. J., & Fabes, J. (2024) · British Journal of Anaesthesia · doi:10.1016/j.bja.2024.01.023 · Cited by 96
“56 Similarly, a bias of 2.4 e 3.6% at an SaO 2 range of 70 e 80% for Nellcor N-595 and Nonin 9700 pulse oximeters with adhesive probes and 4.5 e 4.9% was reported at an SaO 2 of 60% and 70%.”
BackgroundInfluentialThe Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude
Dünnwald, T., Kienast, R., Niederseer, D., & Burtscher, M. (2021) · Sensors · doi:10.3390/s21041263 · Cited by 82
“Skin pigmentation Pulse oximeters are possibly less accurate during hypoxia in dark-skinned individuals at lower saturation (<80%) resulting in overestimations [46,47].”
BackgroundAjmal, Boonya-Ananta, T., Rodriguez, A. J., Du Le, V. N., & Ramella-Roman, J. C. (2021) · Biomedical Optics Express · doi:10.1364/boe.439893 · Cited by 79
Taylor, S. P., Bundy, H. E., Smith, W. M., Skavroneck, S., Taylor, B., & Kowalkowski, M. A. (2020) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202009-1164oc · Cited by 79
Laboratory Testing in the Intensive Care Unit
Ezzie, M. E., Aberegg, S. K., & O’Brien, J. M. (2007) · Critical Care Clinics · doi:10.1016/j.ccc.2007.07.005 · Cited by 79
Ikeda, K., MacLeod, D. B., Grocott, H. P., Moretti, E. W., Ames, W. A., & Vacchiano, C. (2014) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000000463 · Cited by 76
Gudelunas, M. K., Lipnick, M. S., Hendrickson, C. M., Vanderburg, S. B., Okunlola, B., Auchus, I., Feiner, J., & Bickler, P. E. (2023) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000006755 · Cited by 69
“3,4 Since documentation of missed hypoxemia diagnosis due to melanin was documented by Bickler et al 5 and Feiner et al, 6 the FDA has required inclusion of subjects with darkly pigmented skin in validation studies.”
MethodSelf-citationCrooks, C. J., West, J., Morling, J. R., Simmonds, M., Juurlink, I., Briggs, S., Cruickshank, S., Hammond-Pears, S., Shaw, D., Card, T. R., et al. (2022) · European Respiratory Journal · doi:10.1183/13993003.03246-2021 · Cited by 68
“3% higher than the true value in black individuals compared to white individuals [7].”
NORMATIVE VALUES AND PREDICTORS OF RETINAL OXYGEN SATURATION
Jani, P. D., Mwanza, J.-C., Billow, K. B., Waters, A. M., Moyer, S. S., & Garg, S. (2013) · Retina · doi:10.1097/iae.0b013e3182979e7b · Cited by 68
Noninvasive hemoglobin sensing and imaging: optical tools for disease diagnosis
Taylor-Williams, M., Spicer, G. L. C., Bale, G. M., & Bohndiek, S. E. (2022) · Journal of Biomedical Optics · doi:10.1117/1.jbo.27.8.080901 · Cited by 74
“Skin pigmentation leads to an overestimation of arterial oxygen saturation in dark-skinned individuals, which could seriously impact medical decision making and long-term outcomes.(158) These limitations merit increased attention in research and development given the potential for longterm and widespread use of pulse…”
BackgroundHarskamp, R. E., Bekker, L., Himmelreich, J. C. L., Clercq, L. D., Karregat, E. P. M., Sleeswijk, M. E., & Lucassen, W. A. M. (2021) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2021-000939 · Cited by 70
Pu, L. J., Shen, Y., Lu, L., Zhang, R., Zhang, Q., & Shen, W. F. (2012) · Cardiovascular Diabetology · doi:10.1186/1475-2840-11-110 · Cited by 63
“These observations support a notion that the causes of high bias does include skin effect [33,34], and when SaO2 needs to be determined with a high degree of accuracy, arterial blood gases are recommended in type 2 diabetic patients with poor glycemic control.”
BackgroundMaterialized Oppression in Medical Tools and Technologies
Liao, S., & Carbonell, V. (2022) · The American Journal of Bioethics · doi:10.1080/15265161.2022.2044543 · Cited by 67
“The degree of the bias varies proportionately with skin tone, with the darkest skin tones getting the least accurate results (Bickler, Feiner, and Severinghaus 2005; Cecil et al. 1988; Feiner, Severinghaus, and Bickler 2007; MoranThomas 2020).”
BackgroundSudat, S. E. K., Wesson, P., Rhoads, K. F., Brown, S., Aboelata, N., Pressman, A. R., Mani, A., & Azar, K. M. J. (2022) · American Journal of Epidemiology · doi:10.1093/aje/kwac164 · Cited by 68
“(1, 2) The CDC noted the potential for under-detection of hypoxemia in darker-skinned people with COVID-19, citing data from several studies.(3-9) Despite this caution, an oxygen saturation measured by pulse oximetry of less than 94% remains a key determinant of severe disease.”
BackgroundSelf-reported Race/Ethnicity and Intraoperative Occult Hypoxemia: A Retrospective Cohort Study
Burnett, G. W., Stannard, B., Wax, D. B., Lin, H.-M., Pyram-Vincent, C., DeMaria, S., & Levin, M. A. (2022) · Anesthesiology · doi:10.1097/aln.0000000000004153 · Cited by 57
“These findings demonstrated Sao 2 and Spo 2 discrepancies in Black patients at much higher Spo 2 values than previously described.(14,15) These are important and concerning findings, but may not be applicable to a population of patients undergoing anesthetic care.”
BackgroundResultIn-Home Cardiovascular Monitoring System for Heart Failure: Comparative Study
Conn, N. J., Schwarz, K. Q., & Borkholder, D. A. (2019) · JMIR mhealth and uhealth · doi:10.2196/12419 · Cited by 63
Review on remote heart rate measurements using photoplethysmography
Lee, R. J., Sivakumar, S., & Lim, K. H. (2023) · Multimedia Tools and Applications · doi:10.1007/s11042-023-16794-9 · Cited by 63
“From the hardware side, the contact devices for collecting ground truth label data can exhibit biases [84].”
Evaluating consumer and clinical sleep technologies: an American Academy of Sleep Medicine update
Schutte-Rodin, S., Deak, M. C., Khosla, S., Goldstein, C. A., Yurcheshen, M. E., Chiang, A. A.-P., Gault, D., Kern, J., O’Hearn, D. J., Ryals, S. M., et al. (2021) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.9580 · Cited by 60
“66,67 79 Similarly, products utilizing PPG may have variable accuracy in different conditions and populations.”
The Democratization of Artificial Intelligence: Net Politics in the Era of Learning Algorithms
Sudmann, A. (2019) · AI critique · doi:10.14361/9783839447192 · Cited by 59
Diagnosis and Management of Acute Respiratory Failure
Lagina, M., & Valley, T. S. (2024) · Critical Care Clinics · doi:10.1016/j.ccc.2024.01.002 · Cited by 65
Accuracy of Carboxyhemoglobin Detection by Pulse CO-Oximetry During Hypoxemia
Feiner, J., Rollins, M. D., Sall, J. W., Eilers, H., Au, P., & Bickler, P. E. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31828610a0 · Cited by 52
Self-citationMEMS-Based Pulse Wave Sensor Utilizing a Piezoresistive Cantilever
Nguyen, T., Mizuki, Y., Tsukagoshi, T., Takahata, T., Ichiki, M., & Shimoyama, I. (2020) · Sensors · doi:10.3390/s20041052 · Cited by 54
“However, this method has several limitations including high power consumption compared with the force sensing-based method and the non-negligible effect of the skin tone and skin pigmentation on the measurement result [12,13].”
MethodMelanometry for objective evaluation of skin pigmentation in pulse oximetry studies
Vasudevan, S., Vogt, W. C., Weininger, S., & Pfefer, T. J. (2024) · Communications Medicine · doi:10.1038/s43856-024-00550-7 · Cited by 58
“Other subjective classi fi cation methods adopted in literature include self-reported skin color 114 – 116 , racial/ethnic classi fi cations 1 – 5, 35 – 55 , Although subjective methodshave been used forcategorizationof skin pigmentation levels for many years and the aforementioned methods provide a…”
BackgroundMethodFawzy, A., Wu, T. D., Wang, K., Sands, K. E., Fisher, A. M., Arnold Egloff, S. A., DellaVolpe, J. D., Iwashyna, T. J., Xu, Y., & Garibaldi, B. T. (2023) · JAMA Network Open · doi:10.1001/jamanetworkopen.2023.30856 · Cited by 50
Afshari, A. A., Saager, R. B., Burgos, D. E., Vogt, W. C., Wang, J., Mendoza, G., Weininger, S., Sung, K., Durkin, A. J., & Pfefer, T. J. (2022) · Biomedical Optics Express · doi:10.1364/boe.454020 · Cited by 58
Maley, J. H., Alba, G. A., Barry, J. T., Bartels, M. N., Fleming, T. K., Oleson, C. V., Rydberg, L., Sampsel, S., Silver, J. K., Sipes, S., et al. (2022) · PM&R · doi:10.1002/pmrj.12744 · Cited by 55
Sanz, F., Dean, N. C., Dickerson, J., Jones, B. E., Knox, D., Fernández‐Fabrellas, E., Chiner, E., Briones, M. L., Cervera, Á., Aguar, M. C., et al. (2015) · Respirology · doi:10.1111/resp.12560 · Cited by 49
Pulse Oximetry: The Working Principle, Signal Formation, and Applications
Leppänen, T., Kainulainen, S., Korkalainen, H., Sillanmäki, S., Kulkas, A., Töyräs, J., & Nikkonen, S. (2022) · Advances in experimental medicine and biology · doi:10.1007/978-3-031-06413-5_12 · Cited by 57
Thille, A. W., Balen, F., Carteaux, G., Chouihed, T., Frat, J. P., Girault, C., L’Her, E., Marjanovic, N. S., Nay, M.-A., Ray, P., et al. (2024) · Annals of Intensive Care · doi:10.1186/s13613-024-01389-w · Cited by 52
“However, technical problems and measurement biases, such as low peripheral blood flow, anemia, and skin pigmentation can significantly compromise the accuracy of SpO 2 measurements [32–34].”
BackgroundGehring, H., Dümbgen, L., Peterlein, M., Hagelberg, S., & DIBBELT, L. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000268713.58174.cc · Cited by 47
“The Bland–Altman procedure for error analysis and presentation is useful for the presentation of pulse oximeter accuracy data (11).”
BackgroundMethodWiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K. (2022) · Anaesthesia · doi:10.1111/anae.15581 · Cited by 45
“Previous work has suggested that bias is increased at lower oxygen saturation levels (≤90%), especially in patients who have greater degrees of skin pigmentation [6, 7].”
BackgroundLeeb, G., Auchus, I., Law, T. J., Bickler, P. E., Feiner, J., Hashi, S., Monk, E. P., Igaga, E. N., Bernstein, M., Chou, Y., et al. (2024) · EBioMedicine · doi:10.1016/j.ebiom.2024.105051 · Cited by 48
Self-citationAccuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients
Blanchet, M.-A., Mercier, G., Delobel, A., Nayet, E., Bouchard, P.-A., Simard, S., L’Her, E., Branson, R. D., & Lellouche, F. (2023) · Respiratory Care · doi:10.4187/respcare.10582 · Cited by 47
“20,26,36 Oximeters potentially overestimate S aO 2 in patients with dark-skin pigmentation, which can explain the difference in the bias found in Singh et al 21 study and in our study.”
InfluentialSkin Assessment in Patients with Dark Skin Tone
Pusey‐Reid, E., Quinn, L., Samost, M. E., & Reidy, P. A. (2023) · AJN American Journal of Nursing · doi:10.1097/01.naj.0000921800.61980.7e · Cited by 46
Valdez-Lowe, C., Ghareeb, S. A., & Artinian, N. T. (2009) · AJN American Journal of Nursing · doi:10.1097/01.naj.0000352474.55746.81 · Cited by 72
Pulse oximetry in family practice: indications and clinical observations in patients with COPD
Schermer, T. R., Leenders, J., Veen, H. I. ’., van den Bosch, W., Wissink, A., Smeele, I. J. M., & Chavannes, N. H. (2009) · Family Practice · doi:10.1093/fampra/cmp063 · Cited by 43
Putting explainable AI in context: institutional explanations for medical AI
Theunissen, M., & Browning, J. (2022) · Ethics and Information Technology · doi:10.1007/s10676-022-09649-8 · Cited by 37
“The technology was tested out in a predominantly white community (Bickler, 2005), and a study published in late 2020 (Sjoding et al., 2020) high-lighted significant variability depending on skin color with the measurement of the oxygen saturation levels: darker skinned individuals were typically…”
BackgroundRemote Photoplethysmography for Heart Rate and Blood Oxygenation Measurement: A Review
Pirzada, P., Wilde, A., & Harris‐Birtill, D. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3405414 · Cited by 48
“Another study reported an increased error rate of 3.56% ± 2.45% among darker skin tone participants compared with white skin tone participants with an error rate of 0.37% ± 3.20% SpO 2 [57] , an issue that has more recently been researched [58], [59], [60].”
InfluentialA survey of skin tone assessment in prospective research
Weir, V. R., Dempsey, K., Gichoya, J. W., Rotemberg, V. M., & Wong, A.-K. I. (2024) · npj Digital Medicine · doi:10.1038/s41746-024-01176-8 · Cited by 48
Prospective validation of smartphone-based heart rate and respiratory rate measurement algorithms
Bae, S., Borac, S., Emre, Y., Wang, J., Wu, J., Kashyap, M., Kang, S., Chen, L., Moran, M., Cannon, J., et al. (2022) · Communications Medicine · doi:10.1038/s43856-022-00102-x · Cited by 46
“Because the demand for remote triage, diagnosis and monitoring is burgeoning in the wake of the COVID-19 pandemic, there is increased attention being paid to accurate remote physical examination.(3,23,24) There are several limitations in this work.”
BackgroundAccuracy of a portable pulse oximeter in monitoring hypoxemic infants with cyanotic heart disease
Harris, B. U., Stewart, S. E., Verma, A., Hoen, H. M., Stein, M. L., Wright, G. E., & Ramamoorthy, C. (2019) · Cardiology in the Young · doi:10.1017/s1047951119001355 · Cited by 39
Lin, S., Lin, H.-A., Pan, Y.-H., & Hou, S. (2022) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2022-0656 · Cited by 40
In vivo validation of cerebral near-infrared spectroscopy: a review
la Cour, A., Greisen, G. O., & Hyttel-Sørensen, S. (2018) · Neurophotonics · doi:10.1117/1.nph.5.4.040901 · Cited by 39
Enabling Continuous Wearable Reflectance Pulse Oximetry at the Sternum
Chan, M., Ganti, V. G., Heller, J. A., Abdallah, C. A., Etemadi, M., & Inan, O. T. (2021) · Biosensors · doi:10.3390/bios11120521 · Cited by 38
“Note that in [35,36], it was reported that melanin content leads to SpO 2 2 content affects the accuracy of the chest-based pulse oximetry at various SaO levels.”
BackgroundChesley, C. F., Lane-Fall, M. B., Panchanadam, V., Harhay, M. O., Wani, A. A., Mikkelsen, M. E., & Fuchs, B. D. (2022) · Respiratory Care · doi:10.4187/respcare.09769 · Cited by 41
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 53
“Notably, darker skin pigmentation may influence PPG-based SpO 2 and PR readings [7-10].”
Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches
Jiang, Y., Spies, C., Magin, J., Bhosai, S. J., Snyder, L., & Dunn, J. (2023) · PLOS Digital Health · doi:10.1371/journal.pdig.0000296 · Cited by 49
Smith, R., & Hofmeyr, R. (2019) · South African Medical Journal · doi:10.7196/samj.2019.v109i3.13633 · Cited by 38
Racial effects on Masimo pulse oximetry: a laboratory study
Barker, S. J., & Wilson, W. C. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00927-w · Cited by 36
“The possible effect of skin color on the accuracy of oxygen saturation measured by pulse oximetry (SpO2) has been a topic of discussion for many years [1, 2].”
BackgroundEffects of skin tone on photoacoustic imaging and oximetry
Else, T. R., Hacker, L., Gröhl, J., Bunce, E. V., Tao, R., & Bohndiek, S. E. (2023) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s1.s11506 · Cited by 49
Wrist-Sensor Pulse Oximeter Enables Prolonged Patient Monitoring in Chronic Lung Diseases
Guber, A., Shochet, G. E., Kohn, S., & Shitrit, D. (2019) · Journal of Medical Systems · doi:10.1007/s10916-019-1317-2 · Cited by 35
Culnan, D. M., Craft-Coffman, B., Bitz, G. H., Capek, K. D., Tu, Y., Lineaweaver, W. C., & Kuhlmann-Capek, M. J. (2018) · Annals of Plastic Surgery · doi:10.1097/sap.0000000000001351 · Cited by 47
Accuracy and Reliability of Pulse O2 Saturation Measured by a Wrist-worn Oximeter
Hermand, É., Coll, C., Richalet, J., & Lhuissier, F. (2021) · International Journal of Sports Medicine · doi:10.1055/a-1337-2790 · Cited by 35
Diagnostic room-air pulse oximetry: effects of smoking, race, and sex
Witting, M. D., & Scharf, S. M. (2008) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2007.04.002 · Cited by 34
Racial and skin color mediated disparities in pulse oximetry in infants and young children
Sharma, M., Brown, A. W., Powell, N. M., Rajaram, N., Tong, L., Mourani, P. M., & Schootman, M. (2024) · Paediatric Respiratory Reviews · doi:10.1016/j.prrv.2023.12.006 · Cited by 33
Metadata
- Authors
- Bickler, P. E., Feiner, J. R., & Severinghaus, J. W.
- Affiliations
- Professor.
- Associate Professor.
- Professor Emeritus.
- ORCID
- Bickler, Philip E.
- Venue
- Anesthesiology
- Published
- 2005-04-01
- Volume
- 102
- Issue
- 4
- Pages
- 715-719
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2142600751
- Cited by
- 427
- References
- 16
- Citations by year
- 2026: 35 · 2025: 46 · 2024: 49 · 2023: 54 · 2022: 86 · 2021: 36 · 2020: 20 · 2019: 12 · 2018: 7 · 2017: 5 · 2016: 10 · 2015: 6 · 2014: 4 · 2013: 10 · 2012: 10
- FWCI
- 4.93
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Anesthesiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, skin pigmentation, dark skin, arterial oxygen saturation, low oxygen saturation, hypoxia, pulse oximeter accuracy, desaturation, end-tidal oxygen
- MeSH terms
- Adult, Black or African American, Female, Humans, Male, Middle Aged, Oximetry, Oxygen, Oxyhemoglobins, Skin Pigmentation, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.