PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects

Open access
Kostrna, Jason, Oparina, Ekaterina, Palacios, Cristina, Rodriguez, Andres J., Pei, JunZhu, Ajmal, Ajmal, Ramella-Roman, Jessica C.·2026·Sensors

Abstract

Wearable devices are widely used for heart rate (HR) monitoring, yet their accuracy across varying BMI and Fitzpatrick skin tones remains uncertain. This study evaluated four wrist-worn devices (Apple, Fitbit, Samsung, Garmin) in 58 Hispanic adults with Fitzpatrick skin types III–V during a cycling protocol alternating moderate (64–76% HRmax) and vigorous (77–95% HRmax) intensities. Criterion HR was obtained via Polar H10 ECG, with agreement assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), mean bias, and intraclass correlation coefficients (ICC), complemented by Bland–Altman visualization. All devices showed small but systematic deviations from the ECG criterion (p < 0.001). Apple and Garmin exhibited the lowest MAE, whereas Fitbit and Samsung showed greater error. Within this cohort, greater dispersion of error around the mean bias line in the Bland–Altman plots was observed among participants with higher BMI and darker (Fitzpatrick V) skin tones, consistent with phenotype-linked agreement drift. These findings support inferences about commercial wrist-PPG devices in Hispanic adults with Fitzpatrick III–V and underscore the need for equity-focused algorithm refinement.

AI snapshot

Written from the paper’s full text.

TL;DR

In 58 Hispanic adults with Fitzpatrick III–V skin, four wrist PPG devices showed small but systematic deviations from ECG heart rate during recumbent cycling, with Apple and Garmin showing the lowest MAE. Greater error dispersion and bias/reliability drift appeared with higher BMI and darker skin tone, especially Fitzpatrick V, though these subgroup findings are provisional.

Key findings

  • All devices showed systematic deviation from the ECG criterion (p < 0.001). MAE was lowest for Apple (2.51 bpm) and Garmin (2.93 bpm), followed by Samsung (3.38 bpm) and Fitbit (3.89 bpm); MAPE values were 2–4%.
  • Fitzpatrick V participants had higher MAPE than III and IV: 4.48% [3.56, 5.50] vs 2.87% [2.31, 3.55] and 2.72% [2.39, 3.08], respectively. They also showed significantly lower bias than III and IV.
  • Higher BMI was associated with greater directional bias and reduced reliability: BMI > 30 vs < 25 differed for bias (p < 0.001, rrb = 0.416) and ICC (p < 0.001, rrb = −0.369), but not MAE or MAPE.
  • Higher body fat percentage correlated with greater bias (ρ = 0.300, p < 0.001) and lower ICC (ρ = −0.272, p < 0.001). Relative error declined modestly with increasing exercise intensity (√APE: b = −0.0094, p < 0.001, r = −0.283).

Methods

  • Single-site mixed factorial study of 58 self-identified Hispanic adults aged 18–50 with Fitzpatrick III–V (31 women, 27 men; mean age 23 years, SD = 5.91; mean BMI 29.73, SD = 7.43; Fitzpatrick III n = 30, IV n = 24, V n = 4).
  • Participants completed a recumbent cycling protocol with 5 min rest, alternating 2 min moderate (64–76% HRmax) and vigorous (77–95% HRmax) intervals, and 5 min recovery; criterion HR was obtained via Polar H10 ECG.
  • Agreement metrics were MAE, MAPE, signed bias, and ICC, with Bland–Altman plots; analyses used nonparametric tests, bootstrap CIs, Spearman correlations, linear regressions, and linear mixed-effects models.

Limitations

  • Stated: single-site study of self-identified Hispanic adults with Fitzpatrick III–V; Fitzpatrick I, II, and VI were not represented, and Fitzpatrick V was small (n = 4), so findings should not be generalized to those groups, non-Hispanic populations, free-living contexts, or pediatric/older adult populations without replication.
  • Stated: the recumbent, low-motion protocol approximates a best-case optical environment and may not reflect free-living wrist rotation, contact pressure, or ambient light; the trial was short and intensity-segmented, and firmware/algorithms were not manipulated and may have evolved during data collection.
  • Evident: BMI × Fitzpatrick and intensity × BF% × Fitzpatrick interactions were preregistered exploratory and underpowered, especially for Fitzpatrick V cells; the paper repeatedly flags these as hypothesis-generating rather than confirmatory.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.3390/s26102922

Cites

45 of 45 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

16 of 34 cited sources checked · 4 supported · 1 flagged · accuracy not yet measured

  • Premisecitation

    Pulse oximetry has documented skin-pigmentation-linked measurement errors.

    “Related work on pulse oximetry, which relies on the same optical principles, has also documented skin-pigmentation-linked measurement errors [14,15]”
    Quote in the paperSubstantial claim1 of 2 sources checked
    • cites [14] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
    • cites [15] Sjoding, 2020, Racial bias in pulse oximetry measurement, N. Engl. J. Med. source

      In the main text (page not provided), the study reports that pulse oximetry missed occult hypoxemia nearly three times as often in Black patients as in White patients. This documents a race-associated measurement error, but the study classified patients by self-identified race; it did not measure skin pigmentation or establish that pigmentation caused the difference. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
  • Premisecitation

    Boonya-Ananta et al. showed that with increasing melanin concentration, there was a consistent and measurable decline in PPG signal strength.

    “Boonya-Ananta et al. [13] showed that with increasing melanin concentration, there was a consistent and measurable decline in PPG signal strength.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
  • Premisecitation

    Wearables may skip readings entirely when quality drops below internal thresholds, effectively masking rather than mitigating error.

    “In some cases, wearables may even skip readings entirely when quality drops below internal thresholds, effectively masking rather than mitigating error [32].”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [32] Park, 2025, A PPG Signal Dataset Collected in Semi-Naturalistic Settings Using Galaxy Watch, Sci. Data source

      The Technical Validation section (Dataset Completeness Analysis; page unavailable) reports missing Galaxy Watch heart-rate measurements and says its internal status codes identify periods when the device deemed heart-rate calculations unreliable. It does not establish that readings were skipped because signal quality fell below an internal threshold, or characterize the missing readings as masking rather than mitigating error. (LLM, conf=0.85)

      Experimental — accuracy not yet measured
  • Premisecitation

    Failing to correct for PPG errors exacerbates existing healthcare disparities.

    “failing to correct for these errors exacerbates existing healthcare disparities [10,23,24,25,26]”
    Quote in the paperSubstantial claim1 of 3 sources checked
    • cites [10] Zonios, 2001, Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy, J. Investig. Dermatol. source
    • cites [23] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source

      The Discussion (page unavailable) says algorithms and healthcare decisions should account for wearable measurement error, and notes the potential for wearables to address existing healthcare disparities. It does not establish that failing to correct PPG errors exacerbates those disparities; the study found no statistically significant overall difference in heart-rate accuracy across skin tones. (LLM, conf=0.81)

      Experimental — accuracy not yet measured
    • cites [24] Costin, 2007, Human skin pigmentation: Melanocytes modulate skin color in response to stress, FASEB J. source
  • Premisecitation

    Boonya-Ananta et al. showed a 43% decrease in signal amplitude as BMI increased from 25 to 45 kg/m2.

    “Boonya-Ananta et al. [13] confirmed these findings using hybrid FEM and Monte Carlo models, showing a 43% decrease in signal amplitude as BMI increased from 25 to 45 kg/m2.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
  • Classified as probably not a substantial claim. 27
  • May combine two claims. 64
  • The quote only partly states this claim. 5
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductcitation

    A power analysis conducted prior to data collection indicated that a minimum of 52 participants would be needed to detect a medium effect size (f2(v) = 0.20) in a MANOVA with two groups and two dependent variables, assuming α = 0.05 and power (1 − β) = 0.80.

    “A power analysis conducted prior to data collection indicated that a minimum of 52 participants would be needed to detect a medium effect size (f2(v) = 0.20) in a MANOVA with two groups and two dependent variables (absolute and directional error), assuming α = 0.05 and power (1 − β) = 0.80 [23].”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [23] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source

      Split into 2 atoms — 0 confirmed, 1 partially supported, 1 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.

      Checked by a reasoning model — may be wrong.

      Experimental — accuracy not yet measured
  • May combine two claims.

    Premisecitation

    Obesity alters dermal parameters including increased dermal thickness, greater trans-epidermal water loss, reduced capillary density, and increased depth of arteries such as the radial artery.

    “Obesity alters several key dermal parameters: increased dermal thickness, greater trans-epidermal water loss (TEWL), reduced capillary density, and increased depth of arteries such as the radial artery.”
    Quote in the paperSubstantial claim1 of 2 sources checked
    • cites [16] Ndumele, 2016, Obesity and subtypes of incident cardiovascular disease, J. Am. Heart Assoc. source

      The Discussion addresses obesity-related cardiac remodeling and heart failure, not dermal changes. Searches for dermal thickness, trans-epidermal water loss, capillary density, and radial-artery depth found no passage in our copy addressing those claims. (LLM, conf=0.77)

      Experimental — accuracy not yet measured
    • cites [17] Yosipovitch, 2007, Obesity and the skin: Skin physiology and skin manifestations of obesity, J. Am. Acad. Dermatol. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Obesity is a known independent risk factor for multiple cardiovascular diseases.

    “Obesity is a known independent risk factor for multiple cardiovascular diseases [16,21,22]”
    Quote in the paperNot a substantial claim3 of 3 sources checked
    • cites [16] Ndumele, 2016, Obesity and subtypes of incident cardiovascular disease, J. Am. Heart Assoc. source

      The Results report that obesity remained associated with incident heart failure after adjustment for traditional cardiovascular risk factors. They found no significant adjusted association with coronary heart disease or stroke. The claim’s description of obesity as an established independent risk factor for multiple cardiovascular diseases therefore overstates this paper’s findings. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
    • cites [21] Wu, 2015, High blood pressure and all-cause and cardiovascular disease mortalities in community-dwelling older adults, Medicine source

      The Introduction identifies high blood pressure—not obesity—as a cardiovascular disease risk factor. The Results analyze BMI as a subgroup in the relationship between hypertension and CVD mortality, but do not establish obesity as an independent risk factor for multiple cardiovascular diseases (searched: obesity, obese, BMI, independent risk). (LLM, conf=0.76)

      Experimental — accuracy not yet measured
    • cites [22] Ogden, 2017, Prevalence of obesity among adults, by household income and education—United States, 2011–2014, MMWR Morb. Mortal. Wkly. Rep. source

      The Discussion mentions possible health risks associated with BMI, but does not identify cardiovascular diseases or describe obesity as an independent risk factor. Searches for cardiovascular disease, heart disease, and stroke found no passage addressing that claim; the paper instead analyzes obesity prevalence by income and education. (LLM, conf=0.74)

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Age-related changes, such as decreased vascular compliance, increased arterial stiffness, and thinning of the epidermis, reduce the pulsatility and clarity of the PPG waveform.

    “Age-related changes, such as decreased vascular compliance, increased arterial stiffness [18], and thinning of the epidermis [19], reduce the pulsatility and clarity of the PPG waveform.”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [18] Vlachopoulos, C., O’Rourke, M., and Nichols, W.W. (2011). McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles, CRC Press. source
    • cites [19] Derraik, J.G., Rademaker, M., Cutfield, W.S., Pinto, T.E., Tregurtha, S., Faherty, A., Peart, J.M., Drury, P.L., and Hofman, P.L. (2014). Effects of age, gender, BMI, and anatomical site on skin thickness in children and adults with diabetes. PLoS ONE, 9. source

      The Results (Effects of age, puberty, and gender; page unavailable) report that dermal thickness decreases with age in adults. The Methods say the dermal measurement includes a small contribution from the epidermis, but the paper does not establish that the epidermis itself thins or that thinning reduces PPG waveform pulsatility or clarity. The vascular changes are attached to a different citation. (LLM, conf=0.86)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Findingcitation

    Pooled comparisons by BMI category did not reach significance; however, several device-level contrasts indicated small directional biases with higher adiposity.

    “In contrast, pooled comparisons by BMI category did not reach significance; however, several device-level contrasts indicated small directional biases with higher adiposity (e.g., [12,13]).”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [12] Rodriguez, 2021, Monte Carlo analysis of optical heart rate sensors in commercial wearables: The effect of skin tone and obesity on the photoplethysmography (PPG) signal, Biomed. Opt. Express source
    • cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source

      The Results (Figures 6 and 8–9; page not provided) show that simulated PPG signals decline as modeled obesity increases, supporting the direction of the reported effect. The paper does not report pooled statistical comparisons by BMI category or device-level contrasts, and its modeled signal changes are not described as small device biases. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Findingown data

    Statistically reliable patterns tied to phenotype, particularly adiposity, and to a lesser extent skin pigmentation, were evident and align with optical and physiological mechanisms described in prior work.

    “Nonetheless, statistically reliable patterns tied to phenotype, particularly adiposity, and, to a lesser extent, skin pigmentation, were evident and align with the optical and physiological mechanisms described in prior work [13,43].”
    Quote in the paperSubstantial claim2 of 2 sources checked
    • cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
    • cites [43] Fine, J., Branan, K.L., Rodriguez, A.J., Boonya-Ananta, T., Ajmal Ramella-Roman, J.C., McShane, M.J., and Cote, G.L. (2021). Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors, 11. source
  • May combine two claims.

    Findingown data

    Apple and Garmin exhibited the lowest mean absolute error (MAE) among the four devices, whereas Fitbit and Samsung showed greater error.

    “Apple and Garmin exhibited the lowest MAE, whereas Fitbit and Samsung showed greater error.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    In a linear mixed-effects model, Fitzpatrick V participants had approximately 1.6 percentage points higher MAPE than types III-IV (p ≈ 0.05), while BMI was not a significant predictor.

    “The mixed-effects analysis indicated that Fitzpatrick V participants had approximately 1.6 percentage points higher MAPE than types III–IV (p ≈ 0.05) while BMI was not a significant predictor”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
  • May combine two claims.

    Findingown data

    Significant effects of BMI category were observed for ICC (U(n1 = 52, n2 = 88) = 3132, p < 0.001, rrb = −0.369), indicating that participants with higher BMI exhibited reduced measurement reliability.

    “and ICC, U (n1 = 52, n2 = 88) = 3132, p < 0.001, rrb = −0.369”
    Quote in the paperSubstantial claim
    • Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
  • May combine two claims.

    Findingown data

    Significant differences in bias were observed for the Fitbit device (U(n1 = 23, n2 = 13) = 81.0, p = 0.025, rrb = 0.458), with participants with BMI > 30 exhibiting greater directional error than participants with BMI < 25.

    “and Fitbit, U(n1 = 23, n2 = 13) = 81.0, p = 0.025, rrb = 0.458”
    Quote in the paperSubstantial claim
    • Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
  • May combine two claims. The quote only partly states this claim.

    Findingown data

    Significant differences in bias were observed for the Garmin device (U(n1 = 23, n2 = 13) = 63.0, p = 0.005, rrb = 0.579), with participants with BMI > 30 exhibiting greater directional error than participants with BMI < 25.

    “Garmin, U(n1 = 23, n2 = 13) = 63.0, p = 0.005, rrb = 0.579”
    Quote in the paperSubstantial claim
    • Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
  • May combine two claims. The quote only partly states this claim.

    Findingown data

    Significant effects of BMI category were observed for bias (U(n1 = 52, n2 = 88) = 1337, p < 0.001, rrb = 0.416), indicating that participants with higher BMI exhibited greater directional bias.

    “Significant effects of BMI were observed for bias, U(n1 = 52, n2 = 88) = 1337, p < 0.001, rrb = 0.416”
    Quote in the paperSubstantial claim
    • Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
  • May combine two claims. The quote only partly states this claim.

    Findingown data

    Significant differences in bias were observed for the Apple device (U(n1 = 22, n2 = 13) = 72.0, p = 0.016, rrb = 0.497), with participants with BMI > 30 exhibiting greater directional error than participants with BMI < 25.

    “Significant differences in bias were observed for the Apple, U(n1 = 22, n2 = 13) = 72.0, p = 0.016, rrb = 0.497”
    Quote in the paperSubstantial claim
    • Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
  • The quote only partly states this claim.

    Findingown data

    Across devices, participants with skin type III exhibited mean MAPE of 2.87% [2.31, 3.55].

    “participants with skin types III and IV exhibited mean MAPE of 2.87% [2.31, 3.55]”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
  • May combine two claims.

    Findingown data

    For the Garmin device, heart rate was not a significant predictor of bias, but was significantly associated with both √AE and √APE, suggesting modest improvements in accuracy with higher heart rate.

    “For Garmin device, HR was not a significant predictor of bias, but was significantly associated with both √AE and √APE, suggesting modest improvements in accuracy with higher HR.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • May combine two claims.

    Findingown data

    In the study sample, bias increased and ICC declined as body fat percentage rose.

    “The scatterplots in Figure 7 show that bias increased and ICC declined as BF% rose”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    No significant gender differences were found for MAE, MAPE, or bias (U = 3825.5, p = 0.437, rrb = −0.062; U = 3942.5, p = 0.648, rrb = −0.035; U = 3675.5, p = 0.228, rrb = −0.092).

    “No significant gender differences were found for MAE, U = 3825.5, p = 0.437, rrb= −0.062; MAPE, U = 3942.5, p = 0.648, rrb= −0.035, or bias, U = 3675.5, p = 0.228, rrb= −0.092.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    For the Samsung device, heart rate significantly predicted all three error metrics, indicating that directional, absolute, and relative errors decreased modestly with increasing heart rate.

    “For the Samsung device, HR significantly predicted all three error metrics, indicating that directional, absolute, and relative errors decreased modestly with increasing HR.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • May combine two claims.

    Findingown data

    Device-level gender comparisons did not yield any statistically significant differences in error metrics (ps > 0.05), although several comparisons showed small-to-moderate effect sizes.

    “Device-level gender comparisons did not yield any statistically significant differences in error metrics (ps > 0.05), although several comparisons showed small-to-moderate effect sizes.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    For the Apple device, heart rate did not significantly predict bias, but was significantly associated with √AE; no significant effect was observed for √APE.

    “For the Apple device, HR did not significantly predict bias, but was significantly associated with √AE; no significant effect was observed for √APE.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • May combine two claims.

    Findingown data

    The hierarchical models had marginal R2 values ranging from 0.011 to 0.113 and conditional R2 values ranging from 0.07 to 0.26.

    “Model fits indicated that marginal R2 values ranged from 0.011 to 0.113, reflecting the proportion of variance explained by the fixed effects alone (see Table 5). Conditional R2 values, which include both fixed and random effects, ranged from 0.07 to 0.26”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Gender had a significant effect on ICC (U(n1 = 96, n2 = 88) = 3104.5, p = 0.004, rrb = −0.230), indicating that female participants had significantly lower heart-rate reliability than male participants.

    “A significant effect of gender was observed for ICC, U(n1 = 96, n2 = 88) = 3104.5, p = 0.004, rrb= −0.230, indicating that female participants had significantly lower HR reliability than male participants.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    For the Fitbit device, heart rate significantly predicted all three error metrics, with particularly strong reductions in relative error.

    “For the Fitbit device, HR significantly predicted all three error metrics, with particularly strong reductions in relative error.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • May combine two claims.

    Findingown data

    The present study found that all devices deviated from the ECG criterion, systematic bias was small in magnitude, and Bland-Altman dispersion was generally narrow relative to criterion heart rate.

    “found that, although all devices deviated from the ECG criterion, systematic bias was small in magnitude and the Bland–Altman dispersion shown in Figure 4 and Figure 6 was generally narrow relative to criterion HR.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
  • May combine two claims.

    Findingown data

    Adiposity increases bias and decreases reliability in 3/4 devices.

    “adiposity increases bias and decreases reliability in 3/4 devices”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Regression analyses indicated that relative error decreased with increasing intensity across devices, although these slopes were small (<0.01 units per bpm) and varied by device.

    “Regression analyses indicated that the relative error decreased with increasing intensity across devices, although these slopes were small (<0.01 units per bpm) and varied by device.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • May combine two claims.

    Findingown data

    For the Samsung device, all three metrics exhibited significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone: bias b = −0.60, t(4799) = −3.40, p < 0.001; √AE b = 0.09, t(4795) = 3.66, p < 0.001; and √APE b = 0.07, t(4796) = 3.01, p = 0.003.

    “For the Samsung device, all three metrics exhibited significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone bias, b = −0.60, t(4799) = −3.40, p < 0.001; √(AE), b = 0.09, t(4795) = 3.66, p < 0.001; and √(APE), b = 0.07, t(4796) = 3.01, p = 0.003.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The mixed-effects models explained little variance via fixed effects, with far larger conditional R2 driven by participant-level random effects, indicating substantial individual differences not captured by the measured covariates.

    “Our mixed-effects models explained little variance via fixed effects, with far larger conditional R2 driven by participant-level random effects, indicating substantial individual differences not captured by the measured covariates.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    For the Apple device, significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone were observed for bias (b = 0.27, t(4947) = 2.69, p = 0.007) and √APE (b = −0.04, t(4968) = −2.22, p = 0.026); there was also a significant two-way interaction between Fitzpatrick and exercise intensity for √AE (b = −0.04, t(4963) = −2.96, p = 0.003).

    “For the Apple device, significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone were observed for both bias, b = 0.27, t(4947) = 2.69, p = 0.007) and √(APE) (b = −0.04, t(4968) = −2.22, p = 0.026. There was also a significant two-way interaction between Fitzpatrick and exercise intensity for √(AE), b = −0.04, t(4963) = −2.96, p = 0.003.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    For the Fitbit device, all three metrics indicated significant two-way interactions of exercise intensity and Fitzpatrick skin tone: bias b = 0.73, t(4957) = 5.16, p < 0.001; √AE b = −0.05, t(4950) = −2.61, p = 0.009; √APE b = −0.07, t(4950) = −3.51, p < 0.001.

    “For the Fitbit device, all three metrics bias, b = 0.73, t(4957) = 5.16, p < 0.001; √(AE), b = −0.05, t(4950) = −2.61, p = 0.009; √(APE): b = −0.07, t(4950) = −3.51, p < 0.001 indicated significant two-way interactions of exercise intensity and Fitzpatrick skin tone.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    For the Garmin device, all three metrics demonstrated significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone: bias b = 0.51, t(4935) = 4.46, p < 0.001; √AE b = −0.06, t(4962) = −3.07, p = 0.002; and √APE b = −0.07, t(4963) = −3.85, p < 0.001.

    “For the Garmin device, all three metrics demonstrated significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone bias, b = 0.51, t(4935) = 4.46, p < 0.001; √(AE), b = −0.06, t(4962) = −3.07, p = 0.002; and √(APE), b = −0.07, t(4963) = −3.85, p < 0.001.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    When devices were pooled, agreement degraded with darker pigmentation within the Fitzpatrick III-V range: Bland-Altman dispersion was broader at Fitzpatrick V and systematic bias was small but non-zero.

    “When devices were pooled, agreement degraded with darker pigmentation within the Fitzpatrick III–V range: the Bland–Altman dispersion shown in Figure 4 was broader at Fitzpatrick V and systematic bias was small but non-zero”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Device MAEs of approximately 2-4 bpm (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89) fall within the ±5 bpm/±10% tolerance typically used as a pass criterion in the ANSI/CTA-2065 consumer wearable testing standard, and MAPE values of 2-4% are comparable to or narrower than beat-to-beat variability across repeated ECG readings at rest.

    “the device MAEs of approximately 2–4 bpm in the present cohort (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89 bpm; Table 1) fall within the ±5 bpm/±10% tolerance typically used as a pass criterion in the ANSI/CTA-2065 consumer wearable testing standard, and the corresponding MAPE values of 2–4% (Table 1) are comparable to or narrower than the beat-to-beat variability observed across repeated”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Beat-to-beat variability across repeated ECG readings at rest is a valid benchmark for MAPE.
  • May combine two claims.

    Findingown data

    A minority of participants disproportionately occupied the error tails; in the sample, several of these cases had higher BMI and darker skin tones.

    “A minority of participants disproportionately occupied the error tails. In our sample, several of these cases had higher BMI and darker skin tones.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    In a single-site sample of Hispanic adults with Fitzpatrick skin types III-V, commercial wrist-PPG devices showed clinically acceptable agreement with ECG-derived heart rate during low-motion recumbent cycling, with a small-to-moderate systematic bias and greater error dispersion among participants with higher BMI and darker pigmentation.

    “In a single-site sample of Hispanic adults with Fitzpatrick skin types III–V, commercial wrist-PPG devices showed clinically acceptable agreement with ECG-derived HR during low-motion recumbent cycling, with a small-to-moderate systematic bias and greater error dispersion among participants with higher BMI and darker pigmentation.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
  • May combine two claims.

    Premisecitation

    Melanin absorbs light in the visible spectrum, especially at green wavelengths, reducing signal penetration into the blood-containing dermis and attenuating the AC component of the PPG signal.

    “increased melanin reduces signal penetration into the blood-containing dermis, attenuating the AC component of the PPG signal.”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites [10] Zonios, 2001, Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy, J. Investig. Dermatol. source
    • cites [11] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    Wrist assignment for each device (left vs. right) was randomized across participants.

    “Wrist assignment for each device (left vs. right) was randomized across participants.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationown data

    The models indicated substantial unaccounted for inter-individual variability, underscoring the importance of accounting for participant-level nesting in heart-rate error analyses.

    “suggesting substantial unaccounted for inter-individual variability and underscoring the importance of accounting for participant-level nesting in HR error analyses.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    When skin tone and obesity co-occur, the combined effect is more severe; Ajmal et al. noted a compounded degradation in signal quality when simulating a type VI skin tone with BMI > 40, with AC/DC ratio dropping more than 60%.

    “Ajmal et al. [12] noted a compounded degradation in signal quality when simulating a type VI skin tone with BMI > 40, particularly in devices with shallow optical penetration depth like the Fitbit Versa 2. The AC/DC ratio dropped more than 60%”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [12] Rodriguez, 2021, Monte Carlo analysis of optical heart rate sensors in commercial wearables: The effect of skin tone and obesity on the photoplethysmography (PPG) signal, Biomed. Opt. Express source
  • Classified as probably not a substantial claim. May combine two claims.

    Study design and conductreasoning

    Nonparametric statistical methods were prioritized to ensure robustness and interpretability and to accommodate the unequal cell sizes across Fitzpatrick groups.

    “Accordingly, nonparametric statistical methods were prioritized to ensure robustness and interpretability and to accommodate the unequal cell sizes across Fitzpatrick groups”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim.

    Study design and conductnone given

    For the Fitzpatrick subgroup analyses, device-participant combinations were treated as independent observations in nonparametric tests.

    “For the Fitzpatrick subgroup analyses, device–participant combinations were treated as independent observations in nonparametric tests.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim.

    Study design and conductreasoning

    The results were verified using a linear mixed-effects model with participant as a random effect.

    “because participants contributed up to four devices, we verified the results using a linear mixed-effects model with participant as a random effect.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    The heart-rate slopes were small (<0.01 units per bpm) and varied by device, so they are interpreted as modest, device-specific patterns rather than a generalizable improvement with intensity.

    “These slopes were small (<0.01 units per bpm) and varied by device, so we interpret them as modest, device-specific patterns rather than a generalizable improvement with intensity.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • May combine two claims.

    Interpretationown data

    The heart-rate slopes were statistically significant but very small in magnitude (e.g., b = −0.0094 for √APE, explaining ~8% of variance), so improvements in accuracy with higher heart rate should be considered modest.

    “Although the slopes were statistically significant, their magnitudes were very small (e.g., b = −0.0094 for √APE, explaining ~8% of variance), so improvements in accuracy with higher heart rate should be considered modest.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    Participants were randomly assigned to wrist-placement configurations for the wearable devices.

    “Participants were randomly assigned to wrist-placement configurations for the wearable devices”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    All data are provided at OSF.

    “All data are provided at OSF: https://osf.io/9pzr2/overview?view_only=575bd3c163464781a0b600aee6a4d6c6.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Sex differences in subcutaneous fat distribution may influence the light path length and scattering behavior.

    “Sex differences in subcutaneous fat distribution may also influence the light path length and scattering behavior [20].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [20] Freedman, 2013, A comparison of the Slaughter skinfold-thickness equations and BMI in predicting body fatness and cardiovascular disease risk factor levels in children, Am. J. Clin. Nutr. source
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductreasoning

    The recruitment target was increased to 58 individuals to mitigate potential data loss due to equipment malfunction or invalid sensor recordings.

    “To mitigate potential data loss due to equipment malfunction or invalid sensor recordings, the recruitment target was increased to 58 individuals.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    All procedures were preregistered on OSF.

    “All procedures were preregistered on OSF: https://osf.io/9pzr2/overview?view_only=575bd3c163464781a0b600aee6a4d6c6.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductown data

    Preliminary checks confirmed that wrist side did not influence outcomes (ps > 0.05).

    “Preliminary checks confirmed that wrist side did not influence outcomes (ps > 0.05).”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationown data

    Heart-rate-dependent reductions in error were device-specific: Samsung showed consistent, though modest, improvements across all error metrics; Apple showed more isolated effects; Garmin showed mixed effects; and Fitbit demonstrated the strongest heart-rate-related improvement in relative accuracy.

    “Collectively, these findings suggest that HR-dependent reductions in error were device-specific, with Samsung showing consistent, though modest, HR-related improvements across all error metrics, while Apple showed more isolated effects, Garmin showed mixed effects, and Fitbit demonstrated the strongest HR-related improvement in relative accuracy.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
  • Classified as probably not a substantial claim. May combine two claims.

    Study design and conductreasoning

    Interaction effects (BMI × Fitzpatrick; intensity × BF% × Fitzpatrick) are not confirmatory evidence of population-level moderation but are preregistered, exploratory patterns whose direction aligns with optical-physiological mechanisms hypothesized a priori.

    “we emphasize interaction effects (BMI × Fitzpatrick in Section 3.2; intensity × BF% × Fitzpatrick in Section 3.7) not as confirmatory evidence of population-level moderation but as preregistered, exploratory patterns whose direction of effect aligns with the optical–physiological mechanisms hypothesized a priori.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationreasoning

    In free-living use, wrist rotation, variable contact pressure, and ambient light will introduce additional noise and could amplify both overall error and phenotype-linked disparities.

    “In free-living use, wrist rotation, variable contact pressure, and ambient light will introduce additional noise and could amplify both overall error and phenotype-linked disparities.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Study design and conductnone given

    All wearable streams were synchronized to an ECG chest-strap criterion.

    “synchronized all wearable streams to an ECG chest-strap criterion”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
  • May combine two claims.

    Interpretationown data

    Garmin's heart-rate error was highly sensitive to the combined influence of adiposity, pigmentation, and exercise intensity; individuals with higher BF% and darker skin tones exhibited greater directional overestimation at higher intensities, but simultaneously experienced improvements in absolute and relative error.

    “These results suggest that Garmin’s HR error was highly sensitive to the combined influence of adiposity, pigmentation, and exercise intensity. Specifically, individuals with higher BF% and darker skin tones exhibited greater directional overestimation at higher intensities, but simultaneously experienced improvements in absolute and relative error”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Study design and conductreasoning

    This was a single-site study of self-identified Hispanic adults with Fitzpatrick skin types III-V; Fitzpatrick types I, II, and VI were not represented, and the Fitzpatrick V subgroup (n = 4) was small.

    “this was a single-site study of self-identified Hispanic adults with Fitzpatrick skin types III–V; Fitzpatrick types I, II, and VI were not represented, and the Fitzpatrick V subgroup (n = 4) was small.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Interpretationreasoning

    Findings should not be generalized to Fitzpatrick I-II or VI, to non-Hispanic populations, to free-living contexts, or to pediatric or older adult populations without external replication.

    “Findings should therefore not be generalized to Fitzpatrick I–II or VI, to non-Hispanic populations, to free-living contexts, or to pediatric or older adult populations without external replication.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Study design and conductreasoning

    The exclusion of Fitzpatrick I-II and VI was a pragmatic scope decision driven by power (an a priori MANOVA calculation targeting 52-58 participants precluded adequate cell sizes across all six Fitzpatrick types), a fixed budget, and a local volunteer pool concentrated in mid-range Fitzpatrick tones.

    “The exclusion of Fitzpatrick I–II and VI was a pragmatic scope decision driven by power (an a priori MANOVA calculation targeting 52–58 participants precluded adequate cell sizes across all six Fitzpatrick types), a fixed budget that capped participant incentives, and a local volunteer pool concentrated in mid-range Fitzpatrick tones”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Study design and conductown data

    Device-specific analyses yielded fewer skin-tone effects, and the cell size of the Fitzpatrick V subgroup was small, warranting cautious interpretation.

    “Device-specific analyses yielded fewer skin-tone effects, and the cell size of the Fitzpatrick V subgroup was small, warranting cautious interpretation.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    The clinical meaningfulness of these point estimates is dependent on intended use: acceptable for zone-based training and routine monitoring, potentially consequential for rigid clinical thresholds or long-duration energy balance calculations, and especially relevant where systematic bias clusters within phenotypic subgroups.

    “The clinical meaningfulness of these point estimates is therefore best read as dependent on intended use: acceptable for zone-based training and routine monitoring, potentially consequential for rigid clinical thresholds or long-duration energy balance calculations, and especially relevant where systematic bias clusters within phenotypic subgroups as observed here.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    The protocol minimized gross motion with recumbent cycling at a standardized cadence and arm stabilization.

    “We minimized gross motion with recumbent cycling at a standardized cadence and arm stabilization”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Interpretationreasoning

    A recumbent, low-motion context approximates a best-case optical environment; however, it may not benefit from accelerometry-informed algorithms.

    “A recumbent, low-motion context approximates a best-case optical environment; however, it may not benefit from accelerometry-informed algorithms.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Study design and conductown data

    Because the Fitzpatrick V stratum is underpowered, any interaction coefficient that depends on Fitzpatrick V cells is tentative and will only become confirmatory when replicated in larger, purposively stratified samples; the Bland-Altman visualization is leaned on as the more defensible evidentiary base within this cohort.

    “Because the Fitzpatrick V stratum is underpowered, any interaction coefficient that depends on Fitzpatrick V cells is necessarily tentative and will only become confirmatory when replicated in larger, purposively stratified samples; we flag these interactions accordingly and lean on the Bland–Altman visualization in Figure 4 and Figure 6 as the more defensible evidentiary base within this cohort.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Fitbit devices exhibited biased heart-rate estimation as a function of skin tone and intensity, but also demonstrated improvements in absolute and relative accuracy under higher-intensity conditions for individuals with higher pigmentation.

    “Together, these results suggest that while Fitbit devices exhibited biased HR estimation as a function of skin tone and intensity, they also demonstrated improvements in absolute and relative accuracy under higher-intensity conditions for individuals with higher pigmentation”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductnone given

    Nonparametric tests were pre-specified and suited to bounded, skewed error distributions.

    “pre-specified nonparametric tests suited to bounded, skewed error distributions”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning

    Commercial wearables work well on average under favorable conditions, while still exhibiting phenotypic sensitivity that can matter in specific contexts.

    “These observations reinforce that commercial wearables work well on average under favorable conditions, while still exhibiting phenotypic sensitivity that can matter in specific contexts.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Current commercial systems work well on average but remain vulnerable to errors in certain individuals, which are not random with respect to phenotypic features.

    “we highlight that current commercial systems work well on average but remain vulnerable to errors in certain individuals, which are not random with respect to phenotypic features.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Apple's heart-rate error was moderated by both skin tone and body composition, with combined effects of higher exercise intensity, BF%, and pigmentation tending to increase directional bias but reduce relative error.

    “Taken together, these results indicate that Apple’s HR error was moderated by both skin tone and body composition, with combined effects of higher exercise intensity, BF%, and pigmentation tending to increase directional bias, but reduce relative error”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Intensity-linked improvements are modest rather than decisive, but suggest that PPG systems may perform comparatively better when stroke volume and pulsatility increase.

    “We therefore interpret intensity-linked improvements as modest rather than decisive, but they do suggest that PPG systems may perform comparatively better when stroke volume and pulsatility increase [2,29].”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [2] Garet, 2005, Estimating relative physical workload using heart rate monitoring: A validation by whole-body indirect calorimetry, Eur. J. Appl. Physiol. source
    • cites [29] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
  • May combine two claims.

    Study design and conductown data

    Given modest subgroup sizes and collinearity between Fitzpatrick 5 and high BMI in these individuals, the observation that error-tail cases had higher BMI and darker skin tones is hypothesis-generating rather than confirmatory.

    “However, given the modest subgroup sizes and collinearity between Fitzpatrick 5 and high BMI in these individuals, this observation is hypothesis-generating rather than confirmatory.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Study design and conductnone given

    Mixed-effects models respected within-person nesting, allowing separation of systematic influences from participant-level heterogeneity without overfitting device noise.

    “We then complemented those tests with mixed-effects models that respected within-person nesting, allowing us to separate systematic influences from participant-level heterogeneity without overfitting device noise.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Samsung devices tended to underestimate heart rate during high-intensity exercise in individuals with both higher adiposity and darker skin; however, both absolute and relative error increased in these same individuals, indicating that precision and percentage accuracy deteriorated.

    “These findings suggest that Samsung devices tended to underestimate HR during high-intensity exercise in individuals with both higher adiposity and darker skin. However, both absolute and relative error increased in these same individuals, indicating that although the device’s bias trended downward, its precision and percentage accuracy deteriorated”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductreasoning

    The trial was short and intensity-segmented, so longer bouts may reveal thermal or vasomotor drift that the protocol could not capture.

    “The trial was short and intensity-segmented, so longer bouts may reveal thermal or vasomotor drift that our protocol could not capture.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Study design and conductown data

    Skin-tone representation was limited (Fitzpatrick V underpowered; type VI absent), which may limit the detectability of pigmentation effects.

    “Skin-tone representation was limited (Fitzpatrick V was underpowered; type VI was absent), which may limit the detectability of pigmentation effects.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim. May combine two claims. The quote only partly states this claim.

    Study design and conductown data

    The two- and three-way interaction terms involving Fitzpatrick skin tone are preregistered exploratory analyses and are flagged as hypothesis-generating rather than confirmatory due to small participant-level cell sizes (Fitzpatrick V n = 4).

    “we emphasize that these interactions are preregistered exploratory analyses”
    • Also stated in: § 3.7. Hierarchical Linear Modeling · plain
      “We therefore flag significant two- and three-way interaction coefficients involving Fitzpatrick skin tone as hypothesis-generating rather than confirmatory”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    These magnitudes are meaningfully smaller than the width of typical aerobic training zones and therefore unlikely to flip a user between zones during recreational training.

    “These magnitudes are meaningfully smaller than the width of typical aerobic training zones (e.g., moderate 64–76% vs. vigorous 77–95% HRmax) and are therefore unlikely to flip a user between zones during recreational training.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationnone given

    For practitioners and end-users, these devices are suitable for most training decisions, especially when corroborated by perceived exertion and performance cues.

    “Translationally, for practitioners and end-users, these devices are suitable for most training decisions, especially when corroborated by perceived exertion and performance cues.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationreasoning

    Although mechanisms driving the phenotypic effects (optical attenuation by melanin; altered dermal properties with higher adiposity) operate independently of self-identified ethnicity, similar directional patterns would be expected in non-Hispanic cohorts only where Fitzpatrick and BMI distributions overlap the study's, with magnitudes that may differ.

    “Although the mechanisms driving the phenotypic effects observed here (optical attenuation by melanin; altered dermal properties with higher adiposity) operate independently of self-identified ethnicity, similar directional patterns would be expected in non-Hispanic cohorts only where Fitzpatrick and BMI distributions overlap ours, with magnitudes that may differ”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationnone given

    Caution is warranted when applying rigid thresholds to individuals with higher adiposity and darker pigmentation.

    “However, caution is warranted when applying rigid thresholds (e.g., HR-defined clinical alerts) to individuals with higher adiposity and darker pigmentation.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductnone given

    The study did not manipulate firmware or algorithms, and versions may have evolved during data collection, constraining the specificity of device-level inferences.

    “Lastly, we did not manipulate firmware or algorithms, and versions may have evolved during data collection, constraining the specificity of device-level inferences.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationown data

    These findings support the use of wrist-PPG heart rate for most everyday training decisions within this cohort while indicating phenotype-linked agreement drift that warrants targeted evaluation in broader samples.

    “These findings support the use of wrist-PPG HR for most everyday training decisions within this cohort while indicating phenotype-linked agreement drift that warrants targeted evaluation in broader samples”
    Quote in the paperSubstantial claim
81 of 81
  • The paper relies on this without support.

    Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.

    “Self-reported Fitzpatrick type was used in final analyses.”
  • The paper relies on this without support.

    Commercial wearables differ substantially in how they process optical signals, producing device-specific error patterns.

    “Despite relying on the same underlying PPG technology, commercial wearables differ substantially in how they process optical signals, employing proprietary algorithms, sampling rates, sensor configurations, and noise-reduction strategies.”
  • The paper relies on this without support.

    Research on BMI’s role in PPG error and its interaction is nascent and warrants deeper investigation.

    “However, compared to skin tone, research on BMI’s role in PPG error and its interaction is nascent and warrants deeper investigation.”
  • The paper relies on this without support.

    BMI is a valid proxy for body composition/adiposity in this cohort.

    “To evaluate whether BMI category influenced device performance as originally hypothesized, Mann–Whitney U tests were conducted for each error metric (MAE, MAPE, bias, and ICC) comparing participants with BMI < 25 and those with BMI > 30.”
  • The paper relies on this without support.

    The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.

    “synchronized all wearable streams to an ECG chest-strap criterion”
  • The paper relies on this without support.

    The K5 metabolic system heart rate accurately indexes exercise intensity.

    “as indexed by K5 heart rate (HR)”
  • The paper relies on this without support.

    Beat-to-beat variability across repeated ECG readings at rest is a valid benchmark for MAPE.

    “comparable to or narrower than the beat-to-beat variability observed across repeated ECG readings at rest”
  • The paper relies on this without support.

    The sample of Hispanic adults with Fitzpatrick III–V is at elevated cardiometabolic risk.

    “a group at elevated cardiometabolic risk”
8 of 8

Statistics recomputed

Every reported test result, recomputed from its test statistic and degrees of freedom. A result shows as not adding up only once an independent check has confirmed it.

  • Consistentt(20149) = -2.8reported p = .005 · recomputed p = .005

    “t(20149) = −2.80, p = .005”

  • Consistentt(20149) = -15reported p < .001 · recomputed p < .001

    “t(20149) = −15.0, p < .001”

  • Consistentt(20149) = -43.6reported p < .001 · recomputed p < .001

    “t(20149) = −43.6, p < .001”

  • Consistentt(4947) = 2.69reported p = .007 · recomputed p = .007

    “t(4947) = 2.69, p = .007”

  • Consistentt(4968) = -2.22reported p = .026 · recomputed p = .026

    “t(4968) = −2.22, p = .026”

  • Consistentt(4963) = -2.96reported p = .003 · recomputed p = .003

    “t(4963) = −2.96, p = .003”

  • Consistentt(4935) = 4.46reported p < .001 · recomputed p < .001

    “t(4935) = 4.46, p < .001”

  • Consistentt(4962) = -3.07reported p = .002 · recomputed p = .002

    “t(4962) = −3.07, p = .002”

  • Consistentt(4963) = -3.85reported p < .001 · recomputed p < .001

    “t(4963) = −3.85, p < .001”

  • Consistentt(4957) = 5.16reported p < .001 · recomputed p < .001

    “t(4957) = 5.16, p < .001”

  • Consistentt(4950) = -2.61reported p = .009 · recomputed p = .009

    “t(4950) = −2.61, p = .009”

  • Consistentt(4950) = -3.51reported p < .001 · recomputed p < .001

    “t(4950) = −3.51, p < .001”

  • Consistentt(4799) = -3.4reported p < .001 · recomputed p < .001

    “t(4799) = −3.40, p < .001”

  • Consistentt(4795) = 3.66reported p < .001 · recomputed p < .001

    “t(4795) = 3.66, p < .001”

  • Consistentt(4796) = 3.01reported p = .003 · recomputed p = .003

    “t(4796) = 3.01, p = .003”

15 of 15

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • Discussion denies significant pooled BMI effects reported in Results

    Major

    The Results report significant pooled BMI-category effects for bias and ICC, whereas the Discussion says pooled BMI-category comparisons did not reach significance.

  • Reliability effect is attributed to three devices instead of one

    Minor

    The Discussion says adiposity decreases reliability in three of four devices. The device-level Results report a significant ICC difference only for Garmin; the other three devices merely trended toward significance.

  • The criterion HR source changes from Polar H10 to K5

    Minor

    The paper identifies the Polar H10 ECG as the criterion, but the data-analysis section identifies the K5 metabolic system as the criterion for the error metrics.

3 of 3

Cites

Flagged first, then by citations

Cited by

Still being checked.

Authors
Kostrna, J., Oparina, E., Palacios, C., Rodriguez, A. J., Pei, J., Ajmal, A., & Ramella-Roman, J. C.
Affiliations
  • Florida International University (US)
  • Department of Biomedical Engineering, Florida International University, Miami, FL 33199, USA
Corresponding author
Kostrna, Jason
Venue
Sensors
Publisher
MDPI AG
Published
2026-05-07
Volume
26
Issue
10
Pages
2922
ISSN
1424-8220 (ISSN-L)
OpenAlex
W7160488220
Cited by
2
References
45
Citations by year
2026: 2
FWCI
3.58
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • PubMed Central (submittedVersion, cc-by)
Funders
National Science Foundation (800022921)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
Keywords
heart rate monitoring, photoplethysmography, wearable devices, wrist-worn devices, body mass index, Hispanic adults, cycling exercise, device accuracy, algorithmic bias, health equity, Apple Watch, Garmin, Fitbit
MeSH terms
Wearable Electronic Devices, Adult, Body Composition, Electrocardiography, Female, Heart Rate, Hispanic or Latino, Humans, Male, Middle Aged, Skin Pigmentation, Photoplethysmography

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review