PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects
Open accessAbstract
Wearable devices are widely used for heart rate (HR) monitoring, yet their accuracy across varying BMI and Fitzpatrick skin tones remains uncertain. This study evaluated four wrist-worn devices (Apple, Fitbit, Samsung, Garmin) in 58 Hispanic adults with Fitzpatrick skin types III–V during a cycling protocol alternating moderate (64–76% HRmax) and vigorous (77–95% HRmax) intensities. Criterion HR was obtained via Polar H10 ECG, with agreement assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), mean bias, and intraclass correlation coefficients (ICC), complemented by Bland–Altman visualization. All devices showed small but systematic deviations from the ECG criterion (p < 0.001). Apple and Garmin exhibited the lowest MAE, whereas Fitbit and Samsung showed greater error. Within this cohort, greater dispersion of error around the mean bias line in the Bland–Altman plots was observed among participants with higher BMI and darker (Fitzpatrick V) skin tones, consistent with phenotype-linked agreement drift. These findings support inferences about commercial wrist-PPG devices in Hispanic adults with Fitzpatrick III–V and underscore the need for equity-focused algorithm refinement.
AI snapshot
Written from the paper’s full text.
TL;DR
In 58 Hispanic adults with Fitzpatrick III–V skin, four wrist PPG devices showed small but systematic deviations from ECG heart rate during recumbent cycling, with Apple and Garmin showing the lowest MAE. Greater error dispersion and bias/reliability drift appeared with higher BMI and darker skin tone, especially Fitzpatrick V, though these subgroup findings are provisional.
Key findings
- All devices showed systematic deviation from the ECG criterion (p < 0.001). MAE was lowest for Apple (2.51 bpm) and Garmin (2.93 bpm), followed by Samsung (3.38 bpm) and Fitbit (3.89 bpm); MAPE values were 2–4%.
- Fitzpatrick V participants had higher MAPE than III and IV: 4.48% [3.56, 5.50] vs 2.87% [2.31, 3.55] and 2.72% [2.39, 3.08], respectively. They also showed significantly lower bias than III and IV.
- Higher BMI was associated with greater directional bias and reduced reliability: BMI > 30 vs < 25 differed for bias (p < 0.001, rrb = 0.416) and ICC (p < 0.001, rrb = −0.369), but not MAE or MAPE.
- Higher body fat percentage correlated with greater bias (ρ = 0.300, p < 0.001) and lower ICC (ρ = −0.272, p < 0.001). Relative error declined modestly with increasing exercise intensity (√APE: b = −0.0094, p < 0.001, r = −0.283).
Methods
- Single-site mixed factorial study of 58 self-identified Hispanic adults aged 18–50 with Fitzpatrick III–V (31 women, 27 men; mean age 23 years, SD = 5.91; mean BMI 29.73, SD = 7.43; Fitzpatrick III n = 30, IV n = 24, V n = 4).
- Participants completed a recumbent cycling protocol with 5 min rest, alternating 2 min moderate (64–76% HRmax) and vigorous (77–95% HRmax) intervals, and 5 min recovery; criterion HR was obtained via Polar H10 ECG.
- Agreement metrics were MAE, MAPE, signed bias, and ICC, with Bland–Altman plots; analyses used nonparametric tests, bootstrap CIs, Spearman correlations, linear regressions, and linear mixed-effects models.
Limitations
- Stated: single-site study of self-identified Hispanic adults with Fitzpatrick III–V; Fitzpatrick I, II, and VI were not represented, and Fitzpatrick V was small (n = 4), so findings should not be generalized to those groups, non-Hispanic populations, free-living contexts, or pediatric/older adult populations without replication.
- Stated: the recumbent, low-motion protocol approximates a best-case optical environment and may not reflect free-living wrist rotation, contact pressure, or ambient light; the trial was short and intensity-segmented, and firmware/algorithms were not manipulated and may have evolved during data collection.
- Evident: BMI × Fitzpatrick and intensity × BF% × Fitzpatrick interactions were preregistered exploratory and underpowered, especially for Fitzpatrick V cells; the paper repeatedly flags these as hypothesis-generating rather than confirmatory.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
16 of 34 cited sources checked · 4 supported · 1 flagged · accuracy not yet measured
- Premisecitation
Pulse oximetry has documented skin-pigmentation-linked measurement errors.
“Related work on pulse oximetry, which relies on the same optical principles, has also documented skin-pigmentation-linked measurement errors [14,15]”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [14] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
- cites [15] Sjoding, 2020, Racial bias in pulse oximetry measurement, N. Engl. J. Med. source
In the main text (page not provided), the study reports that pulse oximetry missed occult hypoxemia nearly three times as often in Black patients as in White patients. This documents a race-associated measurement error, but the study classified patients by self-identified race; it did not measure skin pigmentation or establish that pigmentation caused the difference. (LLM, conf=0.84)
Experimental — accuracy not yet measured
- Premisecitation
Boonya-Ananta et al. showed that with increasing melanin concentration, there was a consistent and measurable decline in PPG signal strength.
“Boonya-Ananta et al. [13] showed that with increasing melanin concentration, there was a consistent and measurable decline in PPG signal strength.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
- Premisecitation
Wearables may skip readings entirely when quality drops below internal thresholds, effectively masking rather than mitigating error.
“In some cases, wearables may even skip readings entirely when quality drops below internal thresholds, effectively masking rather than mitigating error [32].”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [32] Park, 2025, A PPG Signal Dataset Collected in Semi-Naturalistic Settings Using Galaxy Watch, Sci. Data source
The Technical Validation section (Dataset Completeness Analysis; page unavailable) reports missing Galaxy Watch heart-rate measurements and says its internal status codes identify periods when the device deemed heart-rate calculations unreliable. It does not establish that readings were skipped because signal quality fell below an internal threshold, or characterize the missing readings as masking rather than mitigating error. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Premisecitation
Failing to correct for PPG errors exacerbates existing healthcare disparities.
“failing to correct for these errors exacerbates existing healthcare disparities [10,23,24,25,26]”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [10] Zonios, 2001, Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy, J. Investig. Dermatol. source
- cites [23] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
The Discussion (page unavailable) says algorithms and healthcare decisions should account for wearable measurement error, and notes the potential for wearables to address existing healthcare disparities. It does not establish that failing to correct PPG errors exacerbates those disparities; the study found no statistically significant overall difference in heart-rate accuracy across skin tones. (LLM, conf=0.81)
Experimental — accuracy not yet measured - cites [24] Costin, 2007, Human skin pigmentation: Melanocytes modulate skin color in response to stress, FASEB J. source
- Premisecitation
Boonya-Ananta et al. showed a 43% decrease in signal amplitude as BMI increased from 25 to 45 kg/m2.
“Boonya-Ananta et al. [13] confirmed these findings using hybrid FEM and Monte Carlo models, showing a 43% decrease in signal amplitude as BMI increased from 25 to 45 kg/m2.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
- Premisecitation
Studies using recumbent cycling have demonstrated that errors in HR measurement persist and increase with rising intensity, even with reduced limb movement.
“Studies using recumbent cycling have demonstrated that errors in HR measurement not only persist but actually increase with rising intensity, even with reduced limb movement [28].”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [28] Hung, S.H., Serwa, K., Rosenthal, G., and Eng, J.J. (2025). Validity of heart rate measurements in wrist-based monitors across skin tones during exercise. PLoS ONE, 20. source
The Study procedures and Discussion sections say the researchers used a recumbent bike to reduce arm movement and still observed HR measurement errors as exercise intensity increased. However, the Discussion specifies that the increase occurred predominantly in participants with darker skin tones, not lighter skin tones; the citing sentence omits that qualification. Page numbers are not provided in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measured
- Interpretationcitation
This pattern is consistent with optical absorption by melanin in the green band and reduced pulsatile content reaching the photodiode.
“consistent with optical absorption by melanin in the green band and reduced pulsatile content reaching the photodiode [10,11,28].”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [10] Zonios, 2001, Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy, J. Investig. Dermatol. source
- cites [11] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
- cites [28] Hung, S.H., Serwa, K., Rosenthal, G., and Eng, J.J. (2025). Validity of heart rate measurements in wrist-based monitors across skin tones during exercise. PLoS ONE, 20. source
The Introduction (page unavailable) says melanin absorbs green light and reduces reflection, potentially affecting PPG signal accuracy. The Discussion (page unavailable) suggests this may impair detection of pulsatile signal changes, but does not establish that reduced pulsatile content reaches the photodiode; it says the mechanism requires further research. (LLM, conf=0.84)
Experimental — accuracy not yet measured
- Findingown data
All devices showed significantly reduced intraclass correlation coefficients (ICCs) compared to perfect agreement.
“Additionally, all devices showed significantly reduced ICCs compared to perfect agreement, indicating some degree of reliability loss (see Table 1).”
Quote in the paperNot a substantial claim - Findingown data
Participants with Fitzpatrick skin type V exhibited significantly greater MAPE than those with skin type IV (U(n1 = 11, n2 = 81) = 205, p = 0.003, rrb = 0.607).
“and skin type IV, U(n1 = 11, n2 = 81) = 205, p = 0.003, rrb = 0.607”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
No significant main effect of Fitzpatrick skin tone was observed for MAE (H(2, n = 194) = 5.89, p = 0.053, η2 = 0.175).
“while MAE, H(2, n = 194) = 5.89, p = 0.053, η2 = 0.175”
Quote in the paperNot a substantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Apple and Garmin exhibited the lowest overall bias values among the devices.
“Among devices, Apple and Garmin exhibited the lowest overall MAE and bias values.”
Quote in the paperSubstantial claim - Findingown data
No significant device-specific interactions were observed between BMI group and Fitzpatrick skin tone (all ps > 0.37).
“Across all models, no significant interactions were observed between BMI group and Fitzpatrick skin tone (all ps > 0.37”
Quote in the paperNot a substantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Rests on an assumption: Commercial wearables differ substantially in how they process optical signals, producing device-specific error patterns.
- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
A significant main effect of Fitzpatrick skin tone was observed for bias (H(2, n = 194) = 10.03, p = 0.007, η2 = 0.263).
“and bias, H(2, n = 194) = 10.03, p = 0.007, η2 = 0.263”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Participants with Fitzpatrick skin type V exhibited significantly lower bias than those with Fitzpatrick skin type IV (U(n1 = 11, n2 = 81) = 684, p = 0.004, rrb = 0.557).
“and Fitzpatrick IV, U(n1 = 11, n2 = 81) = 684, p = 0.004, rrb = 0.557”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
No significant interaction was found between BMI category and Fitzpatrick skin tone for MAE (H(5, n = 140) = 1.85, p = 0.108).
“No significant interaction effects were found for MAE, H(5, n = 140) = 1.85, p = 0.108”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
No significant main effect of Fitzpatrick skin tone was observed for ICC (H(2, n = 194) = 3.03, p = 0.220, η2 = 0.102).
“and ICC, H(2, n = 194) = 3.03, p = 0.220, η2 = 0.102 did not reach statistical significance”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Device mean absolute percentage error (MAPE) values were 2-4%.
“corresponding MAPE values of 2–4% (Table 1)”
Quote in the paperSubstantial claim - Findingown data
A significant interaction was observed between BMI category and Fitzpatrick skin tone for ICC (H(5, n = 140) = 3.32, p = 0.007).
“and ICC, H(5, n = 140) = 3.32, p = 0.007”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
A significant interaction was observed between BMI category and Fitzpatrick skin tone for bias (H(5, n = 140) = 4.80, p = 0.001).
“A significant interaction was observed for bias, H(5, n = 140) = 4.80, p = 0.001”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
All four wrist-worn devices showed small but systematic deviations from the ECG criterion heart rate (p < 0.001).
“All devices showed small but systematic deviations from the ECG criterion (p < 0.001).”
Quote in the paperSubstantial claim - Findingown data
No significant interaction was found between BMI category and Fitzpatrick skin tone for MAPE (H(5, n = 140) = 2.05, p = 0.075).
“or MAPE, H(5, n = 140) = 2.05, p = 0.075”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
A significant main effect of Fitzpatrick skin tone was observed for MAPE (H(2, n = 194) = 7.97, p = 0.019, η2 = 0.215).
“Significant main effects of skin tone were observed for MAPE, H(2, n = 194) = 7.97, p = 0.019, η2 = 0.215”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Participants with Fitzpatrick skin type V exhibited significantly greater MAPE than those with skin type III (U(n1 = 11, n2 = 102) = 292, p = 0.009, rrb = 0.511).
“who exhibited significantly greater MAPE than those with skin type III, U(n1 = 11, n2 = 102) = 292, p = 0.009, rrb = 0.511”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Device mean absolute errors (MAEs) were approximately 2-4 bpm (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89 bpm).
“device MAEs of approximately 2–4 bpm (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89 bpm; Table 1)”
Quote in the paperSubstantial claim - Findingown data
Participants with Fitzpatrick skin type V exhibited significantly lower bias than those with Fitzpatrick skin type III (U(n1 = 11, n2 = 102) = 870, p = 0.003, rrb = 0.517).
“as well as significantly lower bias than both Fitzpatrick III, U(n1 = 11, n2 = 102) = 870, p = 0.003, rrb = 0.517”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
All other device-metric comparisons for skin tone yielded non-significant results (ps > 0.10).
“All other device-metric comparisons yielded non-significant results (ps > 0.10).”
Quote in the paperNot a substantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Fitbit trended toward significance for ICC by BMI (U(n1 = 23, n2 = 13) = 208, p = 0.056, rrb = −0.391).
“U(n1 = 23, n2 = 13) = 208, p = 0.056, rrb = −0.391”
Quote in the paperNot a substantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
No significant correlation was found between BF% and MAPE (ρ(188) = 0.015, p = 0.840, ρ2 < 0.001).
“or MAPE, ρ (188) = 0.015, p = 0.840, ρ2 < 0.001”
Quote in the paperSubstantial claim - Findingown data
All other device-level BMI comparisons were non-significant (ps > 0.10).
“All other comparisons were non-significant (ps > 0.10”
Quote in the paperNot a substantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
All other device-metric BF% correlations were non-significant (ps > 0.05).
“All other device–metric correlations were non-significant (ps > 0.05”
Quote in the paperNot a substantial claim - Findingown data
Apple bias trended toward significance by skin tone (H(2, n = 49) = 4.28, p = 0.118, η2 = 0.078).
“while Apple bias trended toward significance, H (2, n = 49) = 4.28, p = 0.118, η2 = 0.078”
Quote in the paperNot a substantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Across devices, participants with skin type V exhibited mean MAPE of 4.48% [3.56, 5.50].
“whereas type V exhibited 4.48% [3.56, 5.50]”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
The Samsung bias effect was primarily driven by participants with Fitzpatrick skin type V, who showed significantly greater directional error than those with skin type III (U(n1 = 2, n2 = 24) = 46, p = 0.025, rrb = −0.917).
“who showed significantly greater directional error than those with skin type III, U(n1 = 2, n2 = 24) = 46, p = 0.025, rrb = −0.917”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Significant monotonic association was observed between BF% and ICC (ρ(188) = −0.272, p < 0.001, ρ2 = 0.074), indicating that higher BF% was associated with reduced reliability of heart rate measurement.
“and ICC, ρ (188) = −0.272, p < 0.001, ρ2 = 0.074”
Quote in the paperSubstantial claim - Findingown data
No significant differences by BMI category were observed for MAPE (U(n1 = 52, n2 = 88) = 2253, p = 0.882, rrb = 0.015).
“or MAPE, U(n1 = 52, n2 = 88) = 2253, p = 0.882, rrb = 0.015”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
Significant positive correlation was observed between BF% and bias for the Apple device (ρ = 0.309, p = 0.032).
“Significant positive correlations were observed for bias in the Apple, ρ = 0.309, p = 0.032”
Quote in the paperSubstantial claim - Findingown data
Across devices, participants with skin type IV exhibited mean MAPE of 2.72% [2.39, 3.08].
“and 2.72% [2.39, 3.08]”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Significant positive correlation was observed between BF% and bias for the Garmin device (ρ = 0.358, p = 0.013).
“Garmin, ρ = 0.358, p = 0.013”
Quote in the paperSubstantial claim - Findingown data
The Samsung bias effect was primarily driven by participants with Fitzpatrick skin type V, who showed significantly greater directional error than those with skin type IV (U(n1 = 2, n2 = 21) = 41, p = 0.016, rrb = −0.952).
“and skin type IV, U(n1 = 2, n2 = 21) = 41, p = 0.016, rrb = −0.952”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Samsung trended toward significance for ICC by BMI (U(n1 = 23, n2 = 13) = 176.0, p = 0.094, rrb = −0.354).
“and U(n1 = 23, n2 = 13) = 176.0, p = 0.094, rrb = −0.354”
Quote in the paperNot a substantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
No significant correlation was found between BF% and MAE (ρ(188) = 0.110, p = 0.129, ρ2 = 0.012).
“No significant correlations were found between BF% and MAE, ρ (188) = 0.110, p = 0.129, ρ2 = 0.012”
Quote in the paperSubstantial claim - Findingown data
No significant differences by BMI category were observed for MAE (U(n1 = 52, n2 = 88) = 1986, p = 0.194, rrb = 0.132).
“No significant differences were observed for MAE, U(n1 = 52, n2 = 88) = 1986, p = 0.194, rrb = 0.132”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
Apple trended toward significance for ICC by BMI (U(n1 = 22, n2 = 13) = 200, p = 0.054, rrb = −0.399).
“the Apple, Fitbit, and Samsung devices trended toward significance for ICC, U(n1 = 22, n2 = 13) = 200, p = 0.054, rrb = −0.399”
Quote in the paperNot a substantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
Garmin showed a significant difference in ICC by BMI (U(n1 = 23, n2 = 13) = 212.0, p = 0.041, rrb = −0.418).
“Garmin also showed a significant difference in ICC, U(n1 = 23, n2 = 13) = 212.0, p = 0.041, rrb = −0.418”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Findingown data
A significant main effect of skin tone was observed for Samsung bias (H(2, n = 47) = 6.64, p = 0.036, η2 = 0.106).
“A significant main effect of skin tone was observed for Samsung bias, H (2, n = 47) = 6.64, p = 0.036, η2 = 0.106”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Findingown data
Significant negative correlation was observed between BF% and ICC for the Apple device (ρ = −0.319, p = 0.027).
“A significant negative correlation was also observed between BF% and ICC for the Apple device, ρ = −0.319, p = 0.027”
Quote in the paperSubstantial claim - Findingown data
Significant positive correlation was observed between BF% and bias for the Samsung device (ρ = 0.316, p = 0.033).
“and Samsung, ρ = 0.316, p = 0.033”
Quote in the paperSubstantial claim - Findingown data
Significant monotonic association was observed between BF% and bias (ρ(188) = 0.300, p < 0.001, ρ2 = 0.090), indicating that higher BF% was associated with greater directional error.
“Significant monotonic associations were observed for bias, ρ (188) = 0.300, p < 0.001, ρ2 = 0.090”
Quote in the paperSubstantial claim - Findingown data
Age had a small but statistically significant negative correlation with ICC (ρ(184) = −0.207, p = 0.005, ρ2 = 0.043), indicating that older participants tended to have slightly lower agreement with the criterion measure.
“A small but statistically significant correlation was found for ICC, ρ(184) = −0.207, p = 0.005, ρ2 = 0.043, indicating that older participants tended to have slightly lower agreement with the criterion measure.”
Quote in the paperNot a substantial claim - Findingown data
Age had no significant associations with MAE, MAPE, or bias (ρ(184) = −0.097, p = 0.191; ρ(184) = −0.046, p = 0.528; ρ(184) = −0.098, p = 0.188).
“No significant associations were observed for MAE, ρ(184) = −0.097, p = 0.191, ρ2 = 0.009; MAPE, ρ(184) = −0.046, p = 0.528, ρ2 = 0.002; or bias, ρ(184) = −0.098, p = 0.188, ρ2 = 0.010.”
Quote in the paperNot a substantial claim - Findingown data
Heart rate had a stronger negative relationship with square root-transformed absolute percentage error (b = −0.0094, t(20149) = −43.6, p < 0.001, r = −0.283), indicating relative error significantly declined with increasing heart rate.
“A stronger relationship emerged between HR and √APE, b = −0.0094, t(20149) = −43.6, p < 0.001, r = −0.283, indicating that relative error significantly declined with increasing HR.”
Quote in the paperNot a substantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
- Findingown data
Heart rate had a statistically significant but negligible effect on bias (b = −0.0044, t(20149) = −2.80, p = 0.005, r = −0.020), indicating a very slight tendency for directional error to decrease as heart rate increased.
“A statistically significant, albeit negligible, effect of HR on bias was observed, b = −0.0044, t(20149) = −2.80, p = 0.005, r = −0.020, indicating a very slight tendency for directional error to decrease as heart rate increased.”
Quote in the paperNot a substantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
- Findingown data
Heart rate had a small but significant negative association with square root-transformed absolute error (b = −0.0035, t(20149) = −15.0, p < 0.001, r = −0.105), suggesting absolute error modestly decreased at higher intensities.
“A small but significant negative association was also found between HR and √AE, b = −0.0035, t(20149) = −15.0, p < 0.001, r = −0.105, suggesting that absolute error modestly decreased at higher intensities.”
Quote in the paperNot a substantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
- Findingown data
Device-level correlations showed no significant associations between age and any error metric (all ps > 0.05).
“Device-level correlations showed no significant associations between age and any error metric (all ps > 0.05)”
Quote in the paperNot a substantial claim - Findingown data
Relative accuracy tends to improve with intensity.
“relative accuracy tends to improve with intensity”
Quote in the paperSubstantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
- Findingown data
Pigmentation exerts a smaller but directionally consistent effect when power is pooled.
“pigmentation exerts a smaller but directionally consistent effect when power is pooled”
Quote in the paperSubstantial claim - Interpretationown data
Combinations of high BMI and darker skin tone are uniquely associated with deviations in heart rate estimation, particularly in bias and reliability.
“These findings suggest that combinations of high BMI and darker skin tone are uniquely associated with deviations in heart rate estimation, particularly in bias and reliability”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Interpretationown data
Device MAEs fell within the ±5 bpm/±10% tolerance used as a pass criterion in the ANSI/CTA-2065 consumer wearable testing standard.
“device MAEs of approximately 2–4 bpm (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89 bpm; Table 1) fall within the ±5 bpm/±10% tolerance used as a pass criterion in the ANSI/CTA-2065 consumer wearable testing standard”
Quote in the paperNot a substantial claim - Premisecitation
Ajmal et al. quantified up to a 15% drop in PPG signal quality (AC/DC ratio) between Fitzpatrick type I and VI skin.
“Ajmal et al. [12] quantified this impact using Monte Carlo simulations, reporting up to a 15% drop in signal quality (AC/DC ratio) between Fitzpatrick type I and VI skin.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [12] Rodriguez, 2021, Monte Carlo analysis of optical heart rate sensors in commercial wearables: The effect of skin tone and obesity on the photoplethysmography (PPG) signal, Biomed. Opt. Express source
- Interpretationown data
BMI category effects are interpreted as agreement drift visible in the Bland-Altman plots, with greater dispersion of error around the mean bias line among participants with BMI ≥ 30 compared with BMI < 25.
“BMI category effects are interpreted here as agreement drift visible in the Bland–Altman plots of Figure 6, which color code individual observations by BMI group and show greater dispersion of error around the mean bias line among participants with BMI ≥ 30 compared with BMI < 25.”
Quote in the paperNot a substantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
- Interpretationown data
The skin-tone effect on MAPE persists after accounting for BMI.
“confirming that the skin-tone effect persists after accounting for BMI”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
- Study design and conductreasoning
A stratified sampling strategy based on BMI and skin tone was used to ensure representation across key phenotypic characteristics relevant to wearable-sensor performance.
“To ensure representation across key phenotypic characteristics relevant to wearable-sensor performance, a stratified sampling strategy was used based on BMI and skin tone.”
Quote in the paperSubstantial claim - Interpretationown data
The observed reduction in ICC with age may reflect a small general trend rather than device-specific variation.
“suggesting that the observed reduction in ICC with age may reflect a small general trend rather than device-specific variation.”
Quote in the paperNot a substantial claim - Premisecitation
Ajmal et al. observed over a 60% drop in AC/DC signal strength between normal-weight and morbidly obese individuals.
“Ajmal et al. [12] observed over a 60% drop in AC/DC signal strength between normal-weight and morbidly obese individuals.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [12] Rodriguez, 2021, Monte Carlo analysis of optical heart rate sensors in commercial wearables: The effect of skin tone and obesity on the photoplethysmography (PPG) signal, Biomed. Opt. Express source
- Premisecitation
The Polar H10 has shown strong agreement with clinical ECG instrumentation in prior validation studies.
“The H10 has shown strong agreement with clinical ECG instrumentation in prior validation studies [36].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] Schweizer, 2019, RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise, Eur. J. Appl. Physiol. source
- Premisecitation
Exercise intensity alone can compromise PPG accuracy.
“Even when movement is minimized, as it is on a stationary cycle, exercise intensity alone can compromise PPG accuracy.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [27] Fine, J., Branan, K.L., Rodriguez, A.J., Boonya-Ananta, T., Ajmal Ramella-Roman, J.C., McShane, M.J., and Coté, G.L. (2025). Optical considerations for wearable photoplethysmography during high-intensity exercise. Biosensors, 15.
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
By focusing on a Hispanic cohort stratified on BMI and Fitzpatrick types III-V, the study provides population-relevant estimates in a group at elevated cardiometabolic risk and historically underrepresented in validation datasets.
“First, by focusing on a Hispanic cohort stratified on BMI and Fitzpatrick types III–V, we provide population-relevant estimates in a group at elevated cardiometabolic risk and historically underrepresented in validation datasets.”
Quote in the paperSubstantial claim- Rests on an assumption: The sample of Hispanic adults with Fitzpatrick III–V is at elevated cardiometabolic risk.
- Interpretationown data
Person-specific factors, beyond BMI, BF%, and skin tone, account for much of the residual error.
“This pattern suggests that person-specific factors, beyond BMI, BF%, and skin tone, account for much of the residual error.”
Quote in the paperSubstantial claim
- Classified as probably not a substantial claim. 27
- May combine two claims. 64
- The quote only partly states this claim. 5
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationA power analysis conducted prior to data collection indicated that a minimum of 52 participants would be needed to detect a medium effect size (f2(v) = 0.20) in a MANOVA with two groups and two dependent variables, assuming α = 0.05 and power (1 − β) = 0.80.
“A power analysis conducted prior to data collection indicated that a minimum of 52 participants would be needed to detect a medium effect size (f2(v) = 0.20) in a MANOVA with two groups and two dependent variables (absolute and directional error), assuming α = 0.05 and power (1 − β) = 0.80 [23].”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [23] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Split into 2 atoms — 0 confirmed, 1 partially supported, 1 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationObesity alters dermal parameters including increased dermal thickness, greater trans-epidermal water loss, reduced capillary density, and increased depth of arteries such as the radial artery.
“Obesity alters several key dermal parameters: increased dermal thickness, greater trans-epidermal water loss (TEWL), reduced capillary density, and increased depth of arteries such as the radial artery.”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [16] Ndumele, 2016, Obesity and subtypes of incident cardiovascular disease, J. Am. Heart Assoc. source
The Discussion addresses obesity-related cardiac remodeling and heart failure, not dermal changes. Searches for dermal thickness, trans-epidermal water loss, capillary density, and radial-artery depth found no passage in our copy addressing those claims. (LLM, conf=0.77)
Experimental — accuracy not yet measured - cites [17] Yosipovitch, 2007, Obesity and the skin: Skin physiology and skin manifestations of obesity, J. Am. Acad. Dermatol. source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationObesity is a known independent risk factor for multiple cardiovascular diseases.
“Obesity is a known independent risk factor for multiple cardiovascular diseases [16,21,22]”
Quote in the paperNot a substantial claim3 of 3 sources checked- cites [16] Ndumele, 2016, Obesity and subtypes of incident cardiovascular disease, J. Am. Heart Assoc. source
The Results report that obesity remained associated with incident heart failure after adjustment for traditional cardiovascular risk factors. They found no significant adjusted association with coronary heart disease or stroke. The claim’s description of obesity as an established independent risk factor for multiple cardiovascular diseases therefore overstates this paper’s findings. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [21] Wu, 2015, High blood pressure and all-cause and cardiovascular disease mortalities in community-dwelling older adults, Medicine source
The Introduction identifies high blood pressure—not obesity—as a cardiovascular disease risk factor. The Results analyze BMI as a subgroup in the relationship between hypertension and CVD mortality, but do not establish obesity as an independent risk factor for multiple cardiovascular diseases (searched: obesity, obese, BMI, independent risk). (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [22] Ogden, 2017, Prevalence of obesity among adults, by household income and education—United States, 2011–2014, MMWR Morb. Mortal. Wkly. Rep. source
The Discussion mentions possible health risks associated with BMI, but does not identify cardiovascular diseases or describe obesity as an independent risk factor. Searches for cardiovascular disease, heart disease, and stroke found no passage addressing that claim; the paper instead analyzes obesity prevalence by income and education. (LLM, conf=0.74)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationAge-related changes, such as decreased vascular compliance, increased arterial stiffness, and thinning of the epidermis, reduce the pulsatility and clarity of the PPG waveform.
“Age-related changes, such as decreased vascular compliance, increased arterial stiffness [18], and thinning of the epidermis [19], reduce the pulsatility and clarity of the PPG waveform.”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [18] Vlachopoulos, C., O’Rourke, M., and Nichols, W.W. (2011). McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles, CRC Press. source
- cites [19] Derraik, J.G., Rademaker, M., Cutfield, W.S., Pinto, T.E., Tregurtha, S., Faherty, A., Peart, J.M., Drury, P.L., and Hofman, P.L. (2014). Effects of age, gender, BMI, and anatomical site on skin thickness in children and adults with diabetes. PLoS ONE, 9. source
The Results (Effects of age, puberty, and gender; page unavailable) report that dermal thickness decreases with age in adults. The Methods say the dermal measurement includes a small contribution from the epidermis, but the paper does not establish that the epidermis itself thins or that thinning reduces PPG waveform pulsatility or clarity. The vascular changes are attached to a different citation. (LLM, conf=0.86)
Experimental — accuracy not yet measured
May combine two claims.
FindingcitationPooled comparisons by BMI category did not reach significance; however, several device-level contrasts indicated small directional biases with higher adiposity.
“In contrast, pooled comparisons by BMI category did not reach significance; however, several device-level contrasts indicated small directional biases with higher adiposity (e.g., [12,13]).”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [12] Rodriguez, 2021, Monte Carlo analysis of optical heart rate sensors in commercial wearables: The effect of skin tone and obesity on the photoplethysmography (PPG) signal, Biomed. Opt. Express source
- cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
The Results (Figures 6 and 8–9; page not provided) show that simulated PPG signals decline as modeled obesity increases, supporting the direction of the reported effect. The paper does not report pooled statistical comparisons by BMI category or device-level contrasts, and its modeled signal changes are not described as small device biases. (LLM, conf=0.84)
Experimental — accuracy not yet measured
May combine two claims.
Findingown dataStatistically reliable patterns tied to phenotype, particularly adiposity, and to a lesser extent skin pigmentation, were evident and align with optical and physiological mechanisms described in prior work.
“Nonetheless, statistically reliable patterns tied to phenotype, particularly adiposity, and, to a lesser extent, skin pigmentation, were evident and align with the optical and physiological mechanisms described in prior work [13,43].”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [13] Rodriguez, 2021, Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI), Sci. Rep. source
- cites [43] Fine, J., Branan, K.L., Rodriguez, A.J., Boonya-Ananta, T., Ajmal Ramella-Roman, J.C., McShane, M.J., and Cote, G.L. (2021). Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors, 11. source
May combine two claims.
Findingown dataApple and Garmin exhibited the lowest mean absolute error (MAE) among the four devices, whereas Fitbit and Samsung showed greater error.
“Apple and Garmin exhibited the lowest MAE, whereas Fitbit and Samsung showed greater error.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn a linear mixed-effects model, Fitzpatrick V participants had approximately 1.6 percentage points higher MAPE than types III-IV (p ≈ 0.05), while BMI was not a significant predictor.
“The mixed-effects analysis indicated that Fitzpatrick V participants had approximately 1.6 percentage points higher MAPE than types III–IV (p ≈ 0.05) while BMI was not a significant predictor”
Quote in the paperSubstantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
May combine two claims.
Findingown dataSignificant effects of BMI category were observed for ICC (U(n1 = 52, n2 = 88) = 3132, p < 0.001, rrb = −0.369), indicating that participants with higher BMI exhibited reduced measurement reliability.
“and ICC, U (n1 = 52, n2 = 88) = 3132, p < 0.001, rrb = −0.369”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
May combine two claims.
Findingown dataSignificant differences in bias were observed for the Fitbit device (U(n1 = 23, n2 = 13) = 81.0, p = 0.025, rrb = 0.458), with participants with BMI > 30 exhibiting greater directional error than participants with BMI < 25.
“and Fitbit, U(n1 = 23, n2 = 13) = 81.0, p = 0.025, rrb = 0.458”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
May combine two claims. The quote only partly states this claim.
Findingown dataSignificant differences in bias were observed for the Garmin device (U(n1 = 23, n2 = 13) = 63.0, p = 0.005, rrb = 0.579), with participants with BMI > 30 exhibiting greater directional error than participants with BMI < 25.
“Garmin, U(n1 = 23, n2 = 13) = 63.0, p = 0.005, rrb = 0.579”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
May combine two claims. The quote only partly states this claim.
Findingown dataSignificant effects of BMI category were observed for bias (U(n1 = 52, n2 = 88) = 1337, p < 0.001, rrb = 0.416), indicating that participants with higher BMI exhibited greater directional bias.
“Significant effects of BMI were observed for bias, U(n1 = 52, n2 = 88) = 1337, p < 0.001, rrb = 0.416”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
May combine two claims. The quote only partly states this claim.
Findingown dataSignificant differences in bias were observed for the Apple device (U(n1 = 22, n2 = 13) = 72.0, p = 0.016, rrb = 0.497), with participants with BMI > 30 exhibiting greater directional error than participants with BMI < 25.
“Significant differences in bias were observed for the Apple, U(n1 = 22, n2 = 13) = 72.0, p = 0.016, rrb = 0.497”
Quote in the paperSubstantial claim- Rests on an assumption: BMI is a valid proxy for body composition/adiposity in this cohort.
The quote only partly states this claim.
Findingown dataAcross devices, participants with skin type III exhibited mean MAPE of 2.87% [2.31, 3.55].
“participants with skin types III and IV exhibited mean MAPE of 2.87% [2.31, 3.55]”
Quote in the paperNot a substantial claim- Rests on an assumption: Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
May combine two claims.
Findingown dataFor the Garmin device, heart rate was not a significant predictor of bias, but was significantly associated with both √AE and √APE, suggesting modest improvements in accuracy with higher heart rate.
“For Garmin device, HR was not a significant predictor of bias, but was significantly associated with both √AE and √APE, suggesting modest improvements in accuracy with higher HR.”
Quote in the paperSubstantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
May combine two claims.
Findingown dataIn the study sample, bias increased and ICC declined as body fat percentage rose.
“The scatterplots in Figure 7 show that bias increased and ICC declined as BF% rose”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataNo significant gender differences were found for MAE, MAPE, or bias (U = 3825.5, p = 0.437, rrb = −0.062; U = 3942.5, p = 0.648, rrb = −0.035; U = 3675.5, p = 0.228, rrb = −0.092).
“No significant gender differences were found for MAE, U = 3825.5, p = 0.437, rrb= −0.062; MAPE, U = 3942.5, p = 0.648, rrb= −0.035, or bias, U = 3675.5, p = 0.228, rrb= −0.092.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor the Samsung device, heart rate significantly predicted all three error metrics, indicating that directional, absolute, and relative errors decreased modestly with increasing heart rate.
“For the Samsung device, HR significantly predicted all three error metrics, indicating that directional, absolute, and relative errors decreased modestly with increasing HR.”
Quote in the paperNot a substantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
May combine two claims.
Findingown dataDevice-level gender comparisons did not yield any statistically significant differences in error metrics (ps > 0.05), although several comparisons showed small-to-moderate effect sizes.
“Device-level gender comparisons did not yield any statistically significant differences in error metrics (ps > 0.05), although several comparisons showed small-to-moderate effect sizes.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor the Apple device, heart rate did not significantly predict bias, but was significantly associated with √AE; no significant effect was observed for √APE.
“For the Apple device, HR did not significantly predict bias, but was significantly associated with √AE; no significant effect was observed for √APE.”
Quote in the paperSubstantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
May combine two claims.
Findingown dataThe hierarchical models had marginal R2 values ranging from 0.011 to 0.113 and conditional R2 values ranging from 0.07 to 0.26.
“Model fits indicated that marginal R2 values ranged from 0.011 to 0.113, reflecting the proportion of variance explained by the fixed effects alone (see Table 5). Conditional R2 values, which include both fixed and random effects, ranged from 0.07 to 0.26”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataGender had a significant effect on ICC (U(n1 = 96, n2 = 88) = 3104.5, p = 0.004, rrb = −0.230), indicating that female participants had significantly lower heart-rate reliability than male participants.
“A significant effect of gender was observed for ICC, U(n1 = 96, n2 = 88) = 3104.5, p = 0.004, rrb= −0.230, indicating that female participants had significantly lower HR reliability than male participants.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor the Fitbit device, heart rate significantly predicted all three error metrics, with particularly strong reductions in relative error.
“For the Fitbit device, HR significantly predicted all three error metrics, with particularly strong reductions in relative error.”
Quote in the paperNot a substantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
May combine two claims.
Findingown dataThe present study found that all devices deviated from the ECG criterion, systematic bias was small in magnitude, and Bland-Altman dispersion was generally narrow relative to criterion heart rate.
“found that, although all devices deviated from the ECG criterion, systematic bias was small in magnitude and the Bland–Altman dispersion shown in Figure 4 and Figure 6 was generally narrow relative to criterion HR.”
Quote in the paperSubstantial claim- Rests on an assumption: The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
May combine two claims.
Findingown dataAdiposity increases bias and decreases reliability in 3/4 devices.
“adiposity increases bias and decreases reliability in 3/4 devices”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataRegression analyses indicated that relative error decreased with increasing intensity across devices, although these slopes were small (<0.01 units per bpm) and varied by device.
“Regression analyses indicated that the relative error decreased with increasing intensity across devices, although these slopes were small (<0.01 units per bpm) and varied by device.”
Quote in the paperNot a substantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
May combine two claims.
Findingown dataFor the Samsung device, all three metrics exhibited significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone: bias b = −0.60, t(4799) = −3.40, p < 0.001; √AE b = 0.09, t(4795) = 3.66, p < 0.001; and √APE b = 0.07, t(4796) = 3.01, p = 0.003.
“For the Samsung device, all three metrics exhibited significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone bias, b = −0.60, t(4799) = −3.40, p < 0.001; √(AE), b = 0.09, t(4795) = 3.66, p < 0.001; and √(APE), b = 0.07, t(4796) = 3.01, p = 0.003.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe mixed-effects models explained little variance via fixed effects, with far larger conditional R2 driven by participant-level random effects, indicating substantial individual differences not captured by the measured covariates.
“Our mixed-effects models explained little variance via fixed effects, with far larger conditional R2 driven by participant-level random effects, indicating substantial individual differences not captured by the measured covariates.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor the Apple device, significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone were observed for bias (b = 0.27, t(4947) = 2.69, p = 0.007) and √APE (b = −0.04, t(4968) = −2.22, p = 0.026); there was also a significant two-way interaction between Fitzpatrick and exercise intensity for √AE (b = −0.04, t(4963) = −2.96, p = 0.003).
“For the Apple device, significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone were observed for both bias, b = 0.27, t(4947) = 2.69, p = 0.007) and √(APE) (b = −0.04, t(4968) = −2.22, p = 0.026. There was also a significant two-way interaction between Fitzpatrick and exercise intensity for √(AE), b = −0.04, t(4963) = −2.96, p = 0.003.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor the Fitbit device, all three metrics indicated significant two-way interactions of exercise intensity and Fitzpatrick skin tone: bias b = 0.73, t(4957) = 5.16, p < 0.001; √AE b = −0.05, t(4950) = −2.61, p = 0.009; √APE b = −0.07, t(4950) = −3.51, p < 0.001.
“For the Fitbit device, all three metrics bias, b = 0.73, t(4957) = 5.16, p < 0.001; √(AE), b = −0.05, t(4950) = −2.61, p = 0.009; √(APE): b = −0.07, t(4950) = −3.51, p < 0.001 indicated significant two-way interactions of exercise intensity and Fitzpatrick skin tone.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor the Garmin device, all three metrics demonstrated significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone: bias b = 0.51, t(4935) = 4.46, p < 0.001; √AE b = −0.06, t(4962) = −3.07, p = 0.002; and √APE b = −0.07, t(4963) = −3.85, p < 0.001.
“For the Garmin device, all three metrics demonstrated significant three-way interactions between exercise intensity, BF%, and Fitzpatrick skin tone bias, b = 0.51, t(4935) = 4.46, p < 0.001; √(AE), b = −0.06, t(4962) = −3.07, p = 0.002; and √(APE), b = −0.07, t(4963) = −3.85, p < 0.001.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataWhen devices were pooled, agreement degraded with darker pigmentation within the Fitzpatrick III-V range: Bland-Altman dispersion was broader at Fitzpatrick V and systematic bias was small but non-zero.
“When devices were pooled, agreement degraded with darker pigmentation within the Fitzpatrick III–V range: the Bland–Altman dispersion shown in Figure 4 was broader at Fitzpatrick V and systematic bias was small but non-zero”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataDevice MAEs of approximately 2-4 bpm (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89) fall within the ±5 bpm/±10% tolerance typically used as a pass criterion in the ANSI/CTA-2065 consumer wearable testing standard, and MAPE values of 2-4% are comparable to or narrower than beat-to-beat variability across repeated ECG readings at rest.
“the device MAEs of approximately 2–4 bpm in the present cohort (Apple 2.51, Garmin 2.93, Samsung 3.38, Fitbit 3.89 bpm; Table 1) fall within the ±5 bpm/±10% tolerance typically used as a pass criterion in the ANSI/CTA-2065 consumer wearable testing standard, and the corresponding MAPE values of 2–4% (Table 1) are comparable to or narrower than the beat-to-beat variability observed across repeated”
Quote in the paperSubstantial claim- Rests on an assumption: Beat-to-beat variability across repeated ECG readings at rest is a valid benchmark for MAPE.
May combine two claims.
Findingown dataA minority of participants disproportionately occupied the error tails; in the sample, several of these cases had higher BMI and darker skin tones.
“A minority of participants disproportionately occupied the error tails. In our sample, several of these cases had higher BMI and darker skin tones.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn a single-site sample of Hispanic adults with Fitzpatrick skin types III-V, commercial wrist-PPG devices showed clinically acceptable agreement with ECG-derived heart rate during low-motion recumbent cycling, with a small-to-moderate systematic bias and greater error dispersion among participants with higher BMI and darker pigmentation.
“In a single-site sample of Hispanic adults with Fitzpatrick skin types III–V, commercial wrist-PPG devices showed clinically acceptable agreement with ECG-derived HR during low-motion recumbent cycling, with a small-to-moderate systematic bias and greater error dispersion among participants with higher BMI and darker pigmentation.”
Quote in the paperSubstantial claim- Rests on an assumption: The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
May combine two claims.
PremisecitationMelanin absorbs light in the visible spectrum, especially at green wavelengths, reducing signal penetration into the blood-containing dermis and attenuating the AC component of the PPG signal.
“increased melanin reduces signal penetration into the blood-containing dermis, attenuating the AC component of the PPG signal.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [10] Zonios, 2001, Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy, J. Investig. Dermatol. source
- cites [11] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenWrist assignment for each device (left vs. right) was randomized across participants.
“Wrist assignment for each device (left vs. right) was randomized across participants.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThe models indicated substantial unaccounted for inter-individual variability, underscoring the importance of accounting for participant-level nesting in heart-rate error analyses.
“suggesting substantial unaccounted for inter-individual variability and underscoring the importance of accounting for participant-level nesting in HR error analyses.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationWhen skin tone and obesity co-occur, the combined effect is more severe; Ajmal et al. noted a compounded degradation in signal quality when simulating a type VI skin tone with BMI > 40, with AC/DC ratio dropping more than 60%.
“Ajmal et al. [12] noted a compounded degradation in signal quality when simulating a type VI skin tone with BMI > 40, particularly in devices with shallow optical penetration depth like the Fitbit Versa 2. The AC/DC ratio dropped more than 60%”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [12] Rodriguez, 2021, Monte Carlo analysis of optical heart rate sensors in commercial wearables: The effect of skin tone and obesity on the photoplethysmography (PPG) signal, Biomed. Opt. Express source
Classified as probably not a substantial claim. May combine two claims.
Study design and conductreasoningNonparametric statistical methods were prioritized to ensure robustness and interpretability and to accommodate the unequal cell sizes across Fitzpatrick groups.
“Accordingly, nonparametric statistical methods were prioritized to ensure robustness and interpretability and to accommodate the unequal cell sizes across Fitzpatrick groups”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductnone givenFor the Fitzpatrick subgroup analyses, device-participant combinations were treated as independent observations in nonparametric tests.
“For the Fitzpatrick subgroup analyses, device–participant combinations were treated as independent observations in nonparametric tests.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductreasoningThe results were verified using a linear mixed-effects model with participant as a random effect.
“because participants contributed up to four devices, we verified the results using a linear mixed-effects model with participant as a random effect.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe heart-rate slopes were small (<0.01 units per bpm) and varied by device, so they are interpreted as modest, device-specific patterns rather than a generalizable improvement with intensity.
“These slopes were small (<0.01 units per bpm) and varied by device, so we interpret them as modest, device-specific patterns rather than a generalizable improvement with intensity.”
Quote in the paperSubstantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
May combine two claims.
Interpretationown dataThe heart-rate slopes were statistically significant but very small in magnitude (e.g., b = −0.0094 for √APE, explaining ~8% of variance), so improvements in accuracy with higher heart rate should be considered modest.
“Although the slopes were statistically significant, their magnitudes were very small (e.g., b = −0.0094 for √APE, explaining ~8% of variance), so improvements in accuracy with higher heart rate should be considered modest.”
Quote in the paperSubstantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenParticipants were randomly assigned to wrist-placement configurations for the wearable devices.
“Participants were randomly assigned to wrist-placement configurations for the wearable devices”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenAll data are provided at OSF.
“All data are provided at OSF: https://osf.io/9pzr2/overview?view_only=575bd3c163464781a0b600aee6a4d6c6.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSex differences in subcutaneous fat distribution may influence the light path length and scattering behavior.
“Sex differences in subcutaneous fat distribution may also influence the light path length and scattering behavior [20].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20] Freedman, 2013, A comparison of the Slaughter skinfold-thickness equations and BMI in predicting body fatness and cardiovascular disease risk factor levels in children, Am. J. Clin. Nutr. source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningThe recruitment target was increased to 58 individuals to mitigate potential data loss due to equipment malfunction or invalid sensor recordings.
“To mitigate potential data loss due to equipment malfunction or invalid sensor recordings, the recruitment target was increased to 58 individuals.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenAll procedures were preregistered on OSF.
“All procedures were preregistered on OSF: https://osf.io/9pzr2/overview?view_only=575bd3c163464781a0b600aee6a4d6c6.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataPreliminary checks confirmed that wrist side did not influence outcomes (ps > 0.05).
“Preliminary checks confirmed that wrist side did not influence outcomes (ps > 0.05).”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataHeart-rate-dependent reductions in error were device-specific: Samsung showed consistent, though modest, improvements across all error metrics; Apple showed more isolated effects; Garmin showed mixed effects; and Fitbit demonstrated the strongest heart-rate-related improvement in relative accuracy.
“Collectively, these findings suggest that HR-dependent reductions in error were device-specific, with Samsung showing consistent, though modest, HR-related improvements across all error metrics, while Apple showed more isolated effects, Garmin showed mixed effects, and Fitbit demonstrated the strongest HR-related improvement in relative accuracy.”
Quote in the paperSubstantial claim- Rests on an assumption: The K5 metabolic system heart rate accurately indexes exercise intensity.
Classified as probably not a substantial claim. May combine two claims.
Study design and conductreasoningInteraction effects (BMI × Fitzpatrick; intensity × BF% × Fitzpatrick) are not confirmatory evidence of population-level moderation but are preregistered, exploratory patterns whose direction aligns with optical-physiological mechanisms hypothesized a priori.
“we emphasize interaction effects (BMI × Fitzpatrick in Section 3.2; intensity × BF% × Fitzpatrick in Section 3.7) not as confirmatory evidence of population-level moderation but as preregistered, exploratory patterns whose direction of effect aligns with the optical–physiological mechanisms hypothesized a priori.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
InterpretationreasoningIn free-living use, wrist rotation, variable contact pressure, and ambient light will introduce additional noise and could amplify both overall error and phenotype-linked disparities.
“In free-living use, wrist rotation, variable contact pressure, and ambient light will introduce additional noise and could amplify both overall error and phenotype-linked disparities.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductnone givenAll wearable streams were synchronized to an ECG chest-strap criterion.
“synchronized all wearable streams to an ECG chest-strap criterion”
Quote in the paperSubstantial claim- Rests on an assumption: The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
May combine two claims.
Interpretationown dataGarmin's heart-rate error was highly sensitive to the combined influence of adiposity, pigmentation, and exercise intensity; individuals with higher BF% and darker skin tones exhibited greater directional overestimation at higher intensities, but simultaneously experienced improvements in absolute and relative error.
“These results suggest that Garmin’s HR error was highly sensitive to the combined influence of adiposity, pigmentation, and exercise intensity. Specifically, individuals with higher BF% and darker skin tones exhibited greater directional overestimation at higher intensities, but simultaneously experienced improvements in absolute and relative error”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductreasoningThis was a single-site study of self-identified Hispanic adults with Fitzpatrick skin types III-V; Fitzpatrick types I, II, and VI were not represented, and the Fitzpatrick V subgroup (n = 4) was small.
“this was a single-site study of self-identified Hispanic adults with Fitzpatrick skin types III–V; Fitzpatrick types I, II, and VI were not represented, and the Fitzpatrick V subgroup (n = 4) was small.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
InterpretationreasoningFindings should not be generalized to Fitzpatrick I-II or VI, to non-Hispanic populations, to free-living contexts, or to pediatric or older adult populations without external replication.
“Findings should therefore not be generalized to Fitzpatrick I–II or VI, to non-Hispanic populations, to free-living contexts, or to pediatric or older adult populations without external replication.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductreasoningThe exclusion of Fitzpatrick I-II and VI was a pragmatic scope decision driven by power (an a priori MANOVA calculation targeting 52-58 participants precluded adequate cell sizes across all six Fitzpatrick types), a fixed budget, and a local volunteer pool concentrated in mid-range Fitzpatrick tones.
“The exclusion of Fitzpatrick I–II and VI was a pragmatic scope decision driven by power (an a priori MANOVA calculation targeting 52–58 participants precluded adequate cell sizes across all six Fitzpatrick types), a fixed budget that capped participant incentives, and a local volunteer pool concentrated in mid-range Fitzpatrick tones”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataDevice-specific analyses yielded fewer skin-tone effects, and the cell size of the Fitzpatrick V subgroup was small, warranting cautious interpretation.
“Device-specific analyses yielded fewer skin-tone effects, and the cell size of the Fitzpatrick V subgroup was small, warranting cautious interpretation.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe clinical meaningfulness of these point estimates is dependent on intended use: acceptable for zone-based training and routine monitoring, potentially consequential for rigid clinical thresholds or long-duration energy balance calculations, and especially relevant where systematic bias clusters within phenotypic subgroups.
“The clinical meaningfulness of these point estimates is therefore best read as dependent on intended use: acceptable for zone-based training and routine monitoring, potentially consequential for rigid clinical thresholds or long-duration energy balance calculations, and especially relevant where systematic bias clusters within phenotypic subgroups as observed here.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe protocol minimized gross motion with recumbent cycling at a standardized cadence and arm stabilization.
“We minimized gross motion with recumbent cycling at a standardized cadence and arm stabilization”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
InterpretationreasoningA recumbent, low-motion context approximates a best-case optical environment; however, it may not benefit from accelerometry-informed algorithms.
“A recumbent, low-motion context approximates a best-case optical environment; however, it may not benefit from accelerometry-informed algorithms.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataBecause the Fitzpatrick V stratum is underpowered, any interaction coefficient that depends on Fitzpatrick V cells is tentative and will only become confirmatory when replicated in larger, purposively stratified samples; the Bland-Altman visualization is leaned on as the more defensible evidentiary base within this cohort.
“Because the Fitzpatrick V stratum is underpowered, any interaction coefficient that depends on Fitzpatrick V cells is necessarily tentative and will only become confirmatory when replicated in larger, purposively stratified samples; we flag these interactions accordingly and lean on the Bland–Altman visualization in Figure 4 and Figure 6 as the more defensible evidentiary base within this cohort.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataFitbit devices exhibited biased heart-rate estimation as a function of skin tone and intensity, but also demonstrated improvements in absolute and relative accuracy under higher-intensity conditions for individuals with higher pigmentation.
“Together, these results suggest that while Fitbit devices exhibited biased HR estimation as a function of skin tone and intensity, they also demonstrated improvements in absolute and relative accuracy under higher-intensity conditions for individuals with higher pigmentation”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenNonparametric tests were pre-specified and suited to bounded, skewed error distributions.
“pre-specified nonparametric tests suited to bounded, skewed error distributions”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningCommercial wearables work well on average under favorable conditions, while still exhibiting phenotypic sensitivity that can matter in specific contexts.
“These observations reinforce that commercial wearables work well on average under favorable conditions, while still exhibiting phenotypic sensitivity that can matter in specific contexts.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataCurrent commercial systems work well on average but remain vulnerable to errors in certain individuals, which are not random with respect to phenotypic features.
“we highlight that current commercial systems work well on average but remain vulnerable to errors in certain individuals, which are not random with respect to phenotypic features.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataApple's heart-rate error was moderated by both skin tone and body composition, with combined effects of higher exercise intensity, BF%, and pigmentation tending to increase directional bias but reduce relative error.
“Taken together, these results indicate that Apple’s HR error was moderated by both skin tone and body composition, with combined effects of higher exercise intensity, BF%, and pigmentation tending to increase directional bias, but reduce relative error”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataIntensity-linked improvements are modest rather than decisive, but suggest that PPG systems may perform comparatively better when stroke volume and pulsatility increase.
“We therefore interpret intensity-linked improvements as modest rather than decisive, but they do suggest that PPG systems may perform comparatively better when stroke volume and pulsatility increase [2,29].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [2] Garet, 2005, Estimating relative physical workload using heart rate monitoring: A validation by whole-body indirect calorimetry, Eur. J. Appl. Physiol. source
- cites [29] Maeda, 2011, Relationship between measurement site and motion artifacts in wearable reflected photoplethysmography, J. Med. Syst. source
May combine two claims.
Study design and conductown dataGiven modest subgroup sizes and collinearity between Fitzpatrick 5 and high BMI in these individuals, the observation that error-tail cases had higher BMI and darker skin tones is hypothesis-generating rather than confirmatory.
“However, given the modest subgroup sizes and collinearity between Fitzpatrick 5 and high BMI in these individuals, this observation is hypothesis-generating rather than confirmatory.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenMixed-effects models respected within-person nesting, allowing separation of systematic influences from participant-level heterogeneity without overfitting device noise.
“We then complemented those tests with mixed-effects models that respected within-person nesting, allowing us to separate systematic influences from participant-level heterogeneity without overfitting device noise.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataSamsung devices tended to underestimate heart rate during high-intensity exercise in individuals with both higher adiposity and darker skin; however, both absolute and relative error increased in these same individuals, indicating that precision and percentage accuracy deteriorated.
“These findings suggest that Samsung devices tended to underestimate HR during high-intensity exercise in individuals with both higher adiposity and darker skin. However, both absolute and relative error increased in these same individuals, indicating that although the device’s bias trended downward, its precision and percentage accuracy deteriorated”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductreasoningThe trial was short and intensity-segmented, so longer bouts may reveal thermal or vasomotor drift that the protocol could not capture.
“The trial was short and intensity-segmented, so longer bouts may reveal thermal or vasomotor drift that our protocol could not capture.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataSkin-tone representation was limited (Fitzpatrick V underpowered; type VI absent), which may limit the detectability of pigmentation effects.
“Skin-tone representation was limited (Fitzpatrick V was underpowered; type VI was absent), which may limit the detectability of pigmentation effects.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims. The quote only partly states this claim.
Study design and conductown dataThe two- and three-way interaction terms involving Fitzpatrick skin tone are preregistered exploratory analyses and are flagged as hypothesis-generating rather than confirmatory due to small participant-level cell sizes (Fitzpatrick V n = 4).
“we emphasize that these interactions are preregistered exploratory analyses”
- Also stated in: § 3.7. Hierarchical Linear Modeling · plain
“We therefore flag significant two- and three-way interaction coefficients involving Fitzpatrick skin tone as hypothesis-generating rather than confirmatory”
Quote in the paperSubstantial claim- Also stated in: § 3.7. Hierarchical Linear Modeling · plain
May combine two claims.
InterpretationreasoningThese magnitudes are meaningfully smaller than the width of typical aerobic training zones and therefore unlikely to flip a user between zones during recreational training.
“These magnitudes are meaningfully smaller than the width of typical aerobic training zones (e.g., moderate 64–76% vs. vigorous 77–95% HRmax) and are therefore unlikely to flip a user between zones during recreational training.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable).
Interpretationnone givenFor practitioners and end-users, these devices are suitable for most training decisions, especially when corroborated by perceived exertion and performance cues.
“Translationally, for practitioners and end-users, these devices are suitable for most training decisions, especially when corroborated by perceived exertion and performance cues.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningAlthough mechanisms driving the phenotypic effects (optical attenuation by melanin; altered dermal properties with higher adiposity) operate independently of self-identified ethnicity, similar directional patterns would be expected in non-Hispanic cohorts only where Fitzpatrick and BMI distributions overlap the study's, with magnitudes that may differ.
“Although the mechanisms driving the phenotypic effects observed here (optical attenuation by melanin; altered dermal properties with higher adiposity) operate independently of self-identified ethnicity, similar directional patterns would be expected in non-Hispanic cohorts only where Fitzpatrick and BMI distributions overlap ours, with magnitudes that may differ”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
Interpretationnone givenCaution is warranted when applying rigid thresholds to individuals with higher adiposity and darker pigmentation.
“However, caution is warranted when applying rigid thresholds (e.g., HR-defined clinical alerts) to individuals with higher adiposity and darker pigmentation.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenThe study did not manipulate firmware or algorithms, and versions may have evolved during data collection, constraining the specificity of device-level inferences.
“Lastly, we did not manipulate firmware or algorithms, and versions may have evolved during data collection, constraining the specificity of device-level inferences.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThese findings support the use of wrist-PPG heart rate for most everyday training decisions within this cohort while indicating phenotype-linked agreement drift that warrants targeted evaluation in broader samples.
“These findings support the use of wrist-PPG HR for most everyday training decisions within this cohort while indicating phenotype-linked agreement drift that warrants targeted evaluation in broader samples”
Quote in the paperSubstantial claim
The paper relies on this without support.
Self-reported Fitzpatrick skin type accurately classifies skin tone for the analyses.
“Self-reported Fitzpatrick type was used in final analyses.”
The paper relies on this without support.
Commercial wearables differ substantially in how they process optical signals, producing device-specific error patterns.
“Despite relying on the same underlying PPG technology, commercial wearables differ substantially in how they process optical signals, employing proprietary algorithms, sampling rates, sensor configurations, and noise-reduction strategies.”
The paper relies on this without support.
Research on BMI’s role in PPG error and its interaction is nascent and warrants deeper investigation.
“However, compared to skin tone, research on BMI’s role in PPG error and its interaction is nascent and warrants deeper investigation.”
The paper relies on this without support.
BMI is a valid proxy for body composition/adiposity in this cohort.
“To evaluate whether BMI category influenced device performance as originally hypothesized, Mann–Whitney U tests were conducted for each error metric (MAE, MAPE, bias, and ICC) comparing participants with BMI < 25 and those with BMI > 30.”
The paper relies on this without support.
The ECG chest-strap criterion provides valid heart rate measurement for agreement analysis.
“synchronized all wearable streams to an ECG chest-strap criterion”
The paper relies on this without support.
The K5 metabolic system heart rate accurately indexes exercise intensity.
“as indexed by K5 heart rate (HR)”
The paper relies on this without support.
Beat-to-beat variability across repeated ECG readings at rest is a valid benchmark for MAPE.
“comparable to or narrower than the beat-to-beat variability observed across repeated ECG readings at rest”
The paper relies on this without support.
The sample of Hispanic adults with Fitzpatrick III–V is at elevated cardiometabolic risk.
“a group at elevated cardiometabolic risk”
Statistics recomputed
Every reported test result, recomputed from its test statistic and degrees of freedom. A result shows as not adding up only once an independent check has confirmed it.
- Consistentt(20149) = -2.8reported p = .005 · recomputed p = .005
“t(20149) = −2.80, p = .005”
- Consistentt(20149) = -15reported p < .001 · recomputed p < .001
“t(20149) = −15.0, p < .001”
- Consistentt(20149) = -43.6reported p < .001 · recomputed p < .001
“t(20149) = −43.6, p < .001”
- Consistentt(4947) = 2.69reported p = .007 · recomputed p = .007
“t(4947) = 2.69, p = .007”
- Consistentt(4968) = -2.22reported p = .026 · recomputed p = .026
“t(4968) = −2.22, p = .026”
- Consistentt(4963) = -2.96reported p = .003 · recomputed p = .003
“t(4963) = −2.96, p = .003”
- Consistentt(4935) = 4.46reported p < .001 · recomputed p < .001
“t(4935) = 4.46, p < .001”
- Consistentt(4962) = -3.07reported p = .002 · recomputed p = .002
“t(4962) = −3.07, p = .002”
- Consistentt(4963) = -3.85reported p < .001 · recomputed p < .001
“t(4963) = −3.85, p < .001”
- Consistentt(4957) = 5.16reported p < .001 · recomputed p < .001
“t(4957) = 5.16, p < .001”
- Consistentt(4950) = -2.61reported p = .009 · recomputed p = .009
“t(4950) = −2.61, p = .009”
- Consistentt(4950) = -3.51reported p < .001 · recomputed p < .001
“t(4950) = −3.51, p < .001”
- Consistentt(4799) = -3.4reported p < .001 · recomputed p < .001
“t(4799) = −3.40, p < .001”
- Consistentt(4795) = 3.66reported p < .001 · recomputed p < .001
“t(4795) = 3.66, p < .001”
- Consistentt(4796) = 3.01reported p = .003 · recomputed p = .003
“t(4796) = 3.01, p = .003”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Discussion denies significant pooled BMI effects reported in Results
MajorThe Results report significant pooled BMI-category effects for bias and ICC, whereas the Discussion says pooled BMI-category comparisons did not reach significance.
“In contrast, pooled comparisons by BMI category did not reach significance”
“Significant effects of BMI were observed for bias, U(n1 = 52, n2 = 88) = 1337, p < 0.001, rrb = 0.416, and ICC, U (n1 = 52, n2 = 88) = 3132, p < 0.001, rrb = −0.369”
Reliability effect is attributed to three devices instead of one
MinorThe Discussion says adiposity decreases reliability in three of four devices. The device-level Results report a significant ICC difference only for Garmin; the other three devices merely trended toward significance.
“adiposity increases bias and decreases reliability in 3/4 devices”
“Garmin also showed a significant difference in ICC, U(n1 = 23, n2 = 13) = 212.0, p = 0.041, rrb = −0.418, and the Apple, Fitbit, and Samsung devices trended toward significance for ICC”
The criterion HR source changes from Polar H10 to K5
MinorThe paper identifies the Polar H10 ECG as the criterion, but the data-analysis section identifies the K5 metabolic system as the criterion for the error metrics.
“Criterion HR was obtained via Polar H10 ECG”
“Four primary error metrics were derived for each device: Mean Absolute Error (MAE), Mean Absolute Percentage Error (MAPE), signed Bias, and Intraclass Correlation Coefficient (ICC) with the criterion HR (K5 metabolic system).”
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 29,639 · see citing works
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · see citing works
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339 · see citing works
Understanding bland altman analysis
Giavarina, D. (2015) · Biochemia Medica · doi:10.11613/bm.2015.015 · Cited by 4,022 · see citing works
Age-predicted maximal heart rate revisited
Tanaka, H., Monahan, K. D., & Seals, D. R. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01054-8 · Cited by 3,775 · cited 1× · see citing works
“Target intensity zones were defined as 64–76% and 77–95% of age-predicted maximal heart rate, respectively, using the Tanaka equation (208–0.7 × age) [38].”
InfluentialGeneralized equations for predicting body density of men
Jackson, A. S., & Pollock, M. L. (1978) · British Journal Of Nutrition · doi:10.1079/bjn19780152 · Cited by 3,205 · see citing works
McDonald's Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles
Vlachopoulos, C., O'Rourke, M., & Nichols, W. W. (2011) · doi:10.1201/b13568 · Cited by 1,613 · see citing works
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · see citing works
Human skin pigmentation: Melanocytes modulate skin color in response to stress
Costin, G.-E., & Hearing, V. J. (2007) · The FASEB Journal · doi:10.1096/fj.06-6649rev · Cited by 976 · see citing works
Smart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · cited 3× · see citing works
“As wearable devices become central to health monitoring, failing to correct for these errors exacerbates existing healthcare disparities [10,23–26].”
InfluentialWearables and the Medical Revolution
Dunn, J., Runge, R., & Snyder, M. (2018) · Personalized Medicine · doi:10.2217/pme-2018-0044 · Cited by 654 · see citing works
Shcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625 · see citing works
Obesity and the skin: Skin physiology and skin manifestations of obesity
Yosipovitch, G., DeVore, A., & Dawn, A. (2007) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2006.12.004 · Cited by 524 · cited 1× · see citing works
“Ndumele et al. [16] and Yosipovitch et al. [17] have documented reduced dermal blood flow and altered skin hydration in obese individuals, both of which compromise the optical contrast needed for accurate PPG detection.”
Zonios, G., Bykowski, J., & Kollias, N. (2001) · Journal of Investigative Dermatology · doi:10.1046/j.0022-202x.2001.01577.x · Cited by 477 · see citing works
Prevalence of obesity among adults, by household income and education—United States, 2011-2014.
Ogden, C. L., Fakhouri, T. H., Carroll, M. D., Hales, C. M., Fryar, C. D., Li, X., & Freedman, D. S. (2017) · MMWR. Morbidity and Mortality Weekly Report · doi:10.15585/mmwr.mm6650a1 · Cited by 456 · see citing works
RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438 · see citing works
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
Warburton, D. E. R., Jamnik, V. K., Bredin, S. S. D., & Gledhill, N. (2011) · The Health & Fitness Journal of Canada · doi:10.14288/hfjc.v4i2.103 · Cited by 375 · see citing works
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Self-citationNelson, B. W., & Allen, N. B. (2019) · JMIR mHealth and uHealth · doi:10.2196/10828 · Cited by 310 · see citing works
Obesity and subtypes of incident cardiovascular disease
Ndumele, C. E., Matsushita, K., Lazo, M., Bello, N., Blumenthal, R. S., Gerstenblith, G., Nambi, V., Ballantyne, C. M., Solomon, S. D., Selvin, E., et al. (2016) · Journal of the American Heart Association · doi:10.1161/jaha.116.003921 · Cited by 259 · see citing works
Kim, K. B., & Baek, H. J. (2023) · Electronics · doi:10.3390/electronics12132923 · Cited by 238 · see citing works
Wu, C.-Y., Hu, H.-Y., Chou, Y.-J., Huang, N., Chou, Y.-C., & Li, C.-P. (2015) · Medicine · doi:10.1097/md.0000000000002160 · Cited by 207 · see citing works
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203 · see citing works
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162 · see citing works
Freedman, D. S., Horlick, M., & Berenson, G. S. (2013) · The American Journal of Clinical Nutrition · doi:10.3945/ajcn.113.065961 · Cited by 152 · see citing works
Assessment of the Fitbit Charge 2 for monitoring heart rate
Benedetto, S., Caldato, C., Bazzan, E., Greenwood, D. C., Pensabene, V., & Actis, P. (2018) · PLOS ONE · doi:10.1371/journal.pone.0192691 · Cited by 149 · see citing works
Accuracy of wearable heart rate monitors in cardiac rehabilitation
Etiwy, M., Akhrass, Z., Gillinov, L., Alashi, A., Wang, R., Blackburn, G., Gillinov, S. M., Phelan, D., Gillinov, A. M., Houghtaling, P. L., et al. (2019) · Cardiovascular Diagnosis and Therapy · doi:10.21037/cdt.2019.04.08 · Cited by 143 · see citing works
Limiting racial disparities and bias for wearable devices in health science research
Colvonen, P. J., DeYoung, P. N., Bosompra, N.-O. A., & Owens, R. L. (2020) · Sleep · doi:10.1093/sleep/zsaa159 · Cited by 137 · see citing works
Spectroscopic Characteristics of Human Melanin In Vivo
Kollias, N., & Baqer, A. (1985) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12275011 · Cited by 135 · see citing works
Derraik, J. G. B., Rademaker, M., Cutfield, W. S., Pinto, T. E., Tregurtha, S., Faherty, A., Peart, J. M., Drury, P. L., & Hofman, P. L. (2014) · PLoS ONE · doi:10.1371/journal.pone.0086637 · Cited by 129 · see citing works
Scardulla, F., Cosoli, G., Spinsante, S., Poli, A., Iadarola, G., Pernice, R., Busacca, A., Pasta, S., Scalise, L., & D’Acquisto, L. (2023) · Measurement · doi:10.1016/j.measurement.2023.113150 · Cited by 120 · see citing works
Ajmal, Boonya-Ananta, T., Rodriguez, A. J., Du Le, V. N., & Ramella-Roman, J. C. (2021) · Biomedical Optics Express · doi:10.1364/boe.439893 · Cited by 79 · see citing works
Self-citationBoonya-ananta, T., Rodriguez, A. J., Ajmal, A., Du Le, V. N., Hansen, A. K., Hutcheson, J. D., & Ramella-Roman, J. C. (2021) · Scientific Reports · doi:10.1038/s41598-021-82124-4 · Cited by 70 · cited 5× · see citing works
“Nonetheless, statistically reliable patterns tied to phenotype, particularly adiposity, and, to a lesser extent, skin pigmentation, were evident and align with the optical and physiological mechanisms described in prior work [13,43].”
InfluentialSelf-citationGaret, M., Boudet, G., Montaurier, C., Vermorel, M., Coudert, J., & Chamoux, A. (2005) · European Journal of Applied Physiology · doi:10.1007/s00421-004-1228-9 · Cited by 67 · see citing works
Ludwig, M., Hoffmann, K., Endler, S., Asteroth, A., & Wiemeyer, J. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.00778 · Cited by 55 · see citing works
Wearable photoplethysmography devices
Marozas, V., & Charlton, P. H. (2021) · Zenodo (CERN European Organization for Nuclear Research) · doi:10.5281/zenodo.4601548 · Cited by 45 · see citing works
Investigating the impact of contact pressure on photoplethysmograms
Sirkiä, J., Panula, T., & Kaisti, M. (2024) · Biomedical Engineering Advances · doi:10.1016/j.bea.2024.100123 · Cited by 22 · see citing works
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
A PPG Signal Dataset Collected in Semi-Naturalistic Settings Using Galaxy Watch
Park, S., Zheng, D., & Lee, U. (2025) · Scientific Data · doi:10.1038/s41597-025-05152-z · Cited by 12 · see citing works
Wearable photoplethysmography devices
Charlton, P. H., & Marozas, V. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00011-6 · Cited by 10 · see citing works
Practical usefulness of heart rate monitoring in physical exercise
Povea, C., & Cabrera, A. (2018) · Revista Colombiana de Cardiología · doi:10.1016/j.rccar.2018.05.004 · Cited by 6 · see citing works
Flagged first, then by citations
Cited by
Still being checked.
- Authors
- Kostrna, J., Oparina, E., Palacios, C., Rodriguez, A. J., Pei, J., Ajmal, A., & Ramella-Roman, J. C.
- Affiliations
- Florida International University (US)
- Department of Biomedical Engineering, Florida International University, Miami, FL 33199, USA
- Corresponding author
- Kostrna, Jason
- ORCID
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2026-05-07
- Volume
- 26
- Issue
- 10
- Pages
- 2922
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC13210717
- OpenAlex
- W7160488220
- Semantic Scholar
- 484520d23dea6e92962e8e041de6aa9a074c11ba
- Cited by
- 2
- References
- 45
- Citations by year
- 2026: 2
- FWCI
- 3.58
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- PubMed Central (submittedVersion, cc-by)
- Funders
- National Science Foundation (800022921)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- heart rate monitoring, photoplethysmography, wearable devices, wrist-worn devices, body mass index, Hispanic adults, cycling exercise, device accuracy, algorithmic bias, health equity, Apple Watch, Garmin, Fitbit
- MeSH terms
- Wearable Electronic Devices, Adult, Body Composition, Electrocardiography, Female, Heart Rate, Hispanic or Latino, Humans, Male, Middle Aged, Skin Pigmentation, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review