Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Open accessAbstract
OBJECTIVE: To determine whether pulse oximeter (PO) accuracy and signal quality are affected by level of skin pigmentation. METHODS: Observational study in a community hospital ED. Consecutive adult patients undergoing arterial blood gas determination were enrolled into the study. Skin pigmentation was determined by comparison with standardized color swatches under controlled lighting; assigned values were used to stratify patients into 3 groups (light, intermediate, and dark) using predetermined criteria. Simultaneous with arterial blood sampling, staff recorded PO reading of O2 saturation using the Nellcor D-25 oximeter. PO values were compared with criterion standard values measured using a 4-wavelength spectrophotometer or co-oximeter. PO signal quality also was recorded. Bias (the mean difference between PO and co-oximeter-measured values of hemoglobin saturation) and precision (the standard deviation of the bias) were calculated. Groups were compared using one-way ANOVA, Bartlett's test for variances, and chi2 test. RESULTS: O2 saturation data were obtained for 284 patients. Bias values did not differ between the 3 skin pigment groups (p = 0.79). Precision was of borderline significance (p = 0.05), but there was no dose-response relation between skin pigmentation and precision. Study personnel reported suboptimal PO function most often among patients in the dark group (p = 0.003), but this finding was of no clinical significance. PO signal failure was rare (<1% of all patients). CONCLUSIONS: Although several prior studies suggest the contrary, this study found that skin pigmentation does not affect the bias or precision of pulse oximetry. Furthermore, skin pigmentation has no clinically significant effect on PO signal quality.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The skin pigmentation effect on pulse oximetry accuracy has long been studied, reporting some contradictory conclusions.
“The skin pigmentation effect on pulse oximetry accuracy has long been studied reporting some contradictory conclusions.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [47]
Some older studies did not identify a considerable bias related to skin pigmentation, suggesting that dark skin pigmentation does not affect the quality of the signal.
“Conversely, some older studies did not identify a considerable bias related to skin pigmentation [44,45,46,56] suggesting that dark skin pigmentation does not affect the quality of the signal [47].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [47]
Objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [47]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
Yelderman, M., & New, W. (1983) · Anesthesiology · doi:10.1097/00000542-198310000-00015 · Cited by 488 · see citing works
Effects of Methemoglobinemia on Pulse Oximetry and Mixed Venous Oximetry
Barker, S. J., Tremper, K. K., & Hyatt, J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00021 · Cited by 384 · see citing works
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 337 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
The Effect of Carbon Monoxide Inhalation on Pulse Oximetry and Transcutaneous Po 2
Barker, S. J., & Tremper, K. K. (1987) · Anesthesiology · doi:10.1097/00000542-198705000-00014 · Cited by 282 · see citing works
The Effect of Nail Polish on Pulse Oximetry
COT, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 171 · see citing works
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145 · see citing works
The Accuracy and Response Characteristics of a Simplified Ear Oximeter
Rebuck, A. S., Chapman, K. R., & D’Urzo, A. D. (1983) · CHEST Journal · doi:10.1378/chest.83.6.860 · Cited by 127 · see citing works
Statistical Methodology: II. Reliability and Validity Assessment in Study Design, Part B
Karros, D. J. (1997) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1997.tb03723.x · Cited by 118 · see citing works
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108 · see citing works
Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.
Ries, A. L., Farrow, J. T., & Clausen, J. L. (1985) · PubMed · doi:10.1164/arrd.1985.132.3.685 · Cited by 106 · see citing works
Pulse Oximeter Desaturation Due to Methemoglobinemia
Eisenkraft, J. B. (1988) · Anesthesiology · doi:10.1097/00000542-198802000-00018 · Cited by 106 · see citing works
Ear oximetry: accuracy and practicability in the assessment of arterial oxygenation.
Saunders, N. A., Powles, A. C. P., & Rebuck, A. S. (1976) · PubMed · doi:10.1164/arrd.1976.113.6.745 · Cited by 105 · see citing works
Pulse Oximetry Is Accurate in Acute Anemia From Hemorrhage
Jay, G. D., Hughes, L., & Renzi, F. P. (1994) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(94)70158-x · Cited by 59 · see citing works
Self-citationStatistical Methodology: II. Reliability and Validity Assessment in Study Design, Part A
Karras, D. J. (1997) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1997.tb03646.x · Cited by 57 · see citing works
Pulse oximetry versus measurement of transcutaneous oxygen
New, W. (1985) · The Journal of Clinical Monitoring · doi:10.1007/bf02832200 · Cited by 54 · see citing works
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50 · see citing works
Clinical Practice of Emergency Medicine
Evans, R. J. (1996) · Emergency Medicine Journal · doi:10.1136/emj.13.5.368-b · Cited by 48 · see citing works
Agreement between Noninvasive Oximetric Values for Oxygen Saturation
Cahan, C., Decker, M. J., Hoekje, P. L., & Strohl, K. P. (1990) · CHEST Journal · doi:10.1378/chest.97.4.814 · Cited by 47 · see citing works
Noninvasive oximetry in pigmented patients.
Wang, Y. T., & Poh, S. C. (1985) · PubMed · Cited by 10 · see citing works
Tobin, M. J. (1990) · JAMA · doi:10.1001/jama.264.2.244 · Cited by 9 · see citing works
Flagged first, then by citations
Research works printed but not linked: 5
- Hess D, Kacmarek RM. Techniques and devices for moni- toring oxygenation. Respir Care. 1993; 38:646-71.
- Barker SJ, Tremper KK, Hyatt JH. Effect of methemoglo- binemia on pulse oximetry and mixed venous oximetry. Anes- thesiology. 1989; 70:112-7.
- New W. Pulse oximetry. J Clin Monit. 1985; 1:126-31.
- Young K. A study of the accuracy of pulse oximetry among varying skin pigmentation populations [abstract]. Am J Crit Care. 1995; 4: 256.
- Saunders NA, Powles ACP, Rebuck AS. E a r oximetry: ac- curacy and practicability in the assessment of arterial oxygen- ation. Am Rev Respir Dis. 1976; 113:745-9.
Cited by
Sorted by citations
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120 · see citing works
The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2016.11.043 · Cited by 96 · see citing works
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K. (2022) · Anaesthesia · doi:10.1111/anae.15581 · Cited by 45 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“It was calculated as , where precision is the SD of bias [42,43].”
Haxha, S., Nwibor, C., Ali, M., Sakel, M., Saunders, K., Dyo, V., & Nabakooza, S. (2024) · Sensors · doi:10.3390/s24113301 · Cited by 12 · see citing works
A Novel Photoplethysmography Sensor for Vital Signs Monitoring from the Human Trachea
May, J. M., Phillips, J. P., Fitchat, T., Ramaswamy, S., Snidvongs, S., & Kyriacou, P. A. (2019) · Biosensors · doi:10.3390/bios9040119 · Cited by 12 · see citing works
The Big Consequences of Small Discrepancies: Why Racial Differences in Pulse Oximetry Errors Matter
Holder, A. L., & Wong, A.-K. I. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005447 · Cited by 13 · see citing works
- Authors
- Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C.
- Affiliations
- Harvard University (US)
- Lowell General Hospital (US)
- Corresponding author
- Adler, Jonathan N.
- ORCID
- Adler, Jonathan N.
- Venue
- Academic Emergency Medicine
- Publisher
- Wiley
- Published
- 1998-10
- Volume
- 5
- Issue
- 10
- Pages
- 965-970
- ISSN
- 1069-6563 (ISSN-L)
- 1553-2712
- OpenAlex
- W2053872280
- Cited by
- 102
- References
- 23
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 16 · 2023: 7 · 2022: 33 · 2021: 8 · 2020: 3 · 2019: 4 · 2018: 1 · 2016: 2 · 2014: 3 · 2013: 2
- FWCI
- 0.00
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Academic Emergency Medicine (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5k14hl003533-04)
- National Heart, Lung, and Blood Institute (HL03533)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, skin pigmentation, oxygen saturation, measurement accuracy, signal quality, arterial blood gases, emergency department, hemoglobin oxygen saturation, measurement bias, measurement precision
- MeSH terms
- Adult, Aged, Emergency Service, Hospital, Female, Humans, Male, Middle Aged, Oximetry, Skin Pigmentation, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review