Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
Background: It is uncertain whether skin pigmentation affects pulse oximeter accuracy at low HbO2 saturation. Methods: The accuracy of finger pulse oximeters during stable, plateau levels of arterial oxygen saturation (Sao2) between 60 and 100% were evaluated in 11 subjects with darkly pigmented skin and in 10 with light skin pigmentation. Oximeters tested were the Nellcor N-595 with the OxiMax-A probe (Nellcor Inc., Pleasanton, CA), the Novametrix 513 (Novametrix Inc., Wallingford, CT), and the Nonin Onyx (Nonin Inc., Plymouth, MN). Semisupine subjects breathed air-nitrogen-carbon dioxide mixtures through a mouthpiece. A computer used end-tidal oxygen and carbon dioxide concentrations determined by mass spectrometry to estimate breath-by-breath Sao2, from which an operator adjusted inspired gas to rapidly achieve 2- to 3-min stable plateaus of desaturation. Comparisons of oxygen saturation measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM3) were used in a multivariate model to determine the interrelation between saturation, skin pigmentation, and oximeter bias (Spo2 - Sao2). Results: At 60-70% Sao2, Spo2 (mean of three oximeters) overestimated Sao2 (bias +/- SD) by 3.56 +/- 2.45% (n = 29) in darkly pigmented subjects, compared with 0.37 +/- 3.20% (n = 58) in lightly pigmented subjects (P < 0.0001). The SD of bias was not greater with dark than light skin. The dark-light skin differences at 60-70% Sao2 were 2.35% (Nonin), 3.38% (Novametrix), and 4.30% (Nellcor). Skin pigment-related differences were significant with Nonin below 70% Sao2, with Novametrix below 90%, and with Nellcor at all ranges. Pigment-related bias increased approximately in proportion to desaturation. Conclusions: The three tested pulse oximeters overestimated arterial oxygen saturation during hypoxia in dark-skinned individuals.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Related work on pulse oximetry, which relies on the same optical principles, has also documented skin-pigmentation-linked measurement errors [14,15]”
From PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects (2026) · cites this paper as [14]
In the citing paper
“Notably, darker skin pigmentation may influence PPG-based SpO2 and PR readings [7-10].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [10]
In the citing paper
“Measurements of the pulse oximeter are known to be affected by the patient’s skin color, as the device systematically overestimates oxygen saturation levels in nonwhite patients.”
From Addressing bias in big data and AI for health care: A call for open science (2021) · cites this paper as [2]
In the citing paper
“Accuracy errors in pulse oximeter readings have been associated with SpO2 overestimations, especially at low saturation conditions [23,24,73,93].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [23]
In the citing paper
“The effect of skin pigmentation in pulse oximeter readings under hypoxia conditions was tested in 11 dark-skinned male subjects and 10 light-skinned males, who were made to breathe an air mixture (nitrogen-carbon monoxide). The three pulse oximeters used showed overestimation in dark-skinned individuals that increases linearly as SpO2 levels decreases [23].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [23]
3 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 333 · cited 1×
“Major known variability is caused by anemia, light scattering, venous and tissue pulsation by mechanical force from nearby arteries, pulsatile variations in tissue thickness in the light path other than in the arteries, nail polish, and skin pigment.(3) Therefore, pulse oximeter design is often based on empirically de…”
BackgroundSelf-citationJubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296
Errors in 14 pulse oximeters during profound hypoxia
Severinghaus, J. W., Naifeh, K. H., & Koh, S. O. (1989) · Journal of Clinical Monitoring · doi:10.1007/bf01617877 · Cited by 217
Self-citationEffects of Intravenously Administered Dyes on Pulse Oximetry Readings
SCHELLER, M. S., UNGER, R. J., & KELNER, M. J. (1986) · Anesthesiology · doi:10.1097/00000542-198611000-00023 · Cited by 194 · cited 1×
“Intravenously injected dyes cause transient errors.(13) Lee et al.”
BackgroundRALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11501.x · Cited by 192 · cited 2×
“%) in black patients.4–7 Model simulations of errors due to various pigments were reviewed by Ralston et al.8 Cote et al.9 reported that nail polish and ink on skin surface can cause errors, a finding confirmed anecdotally by others from fingerprinting ink,10 henna,11 and meconium.12 Intravenously injected dyes cause…”
The Effect of Nail Polish on Pulse Oximetry
Cot??, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 172 · cited 2×
“%) in black patients.4–7 Model simulations of errors due to various pigments were reviewed by Ralston et al.8 Cote et al.9 reported that nail polish and ink on skin surface can cause errors, a finding confirmed anecdotally by others from fingerprinting ink,10 henna,11 and meconium.12 Intravenously injected dyes cause…”
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50
Cited by
Items 401–434 of 434
Shi, C., Goodall, M., Dumville, J. C., Hill, J. E., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · Figshare · doi:10.6084/m9.figshare.20495994
An ECG- and PPG-Based Wearable Atrial Fibrillation Detection Device: Signal Acquisition
Kohabir, A. (., & Smit, A. (. (2021) · Research Repository (Delft University of Technology)
Wearable multi-wavelength photoplethysmography deep learning heart rate estimation
Ray, D. (2024) · e-space (Manchester Metropolitan University) · Cited by 1
Monte Carlo investigations of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R. (2025) · City Research Online (City University London)
Purvez, A. (2026) · Cureus · doi:10.7759/cureus.101995
“Skin pigmentation / dyshemoglobins Systematic bias in SpO ₂ under some conditions may complicate interpretation [8].”
A systems framework for improving clinical trial recruitment: A literature-informed approach
Zimmerman, B. E., Moellner, A., & Safdar, N. (2026) · Journal of Clinical and Translational Science · doi:10.1017/cts.2026.10238
“For example, a study of pulse oximeter devices revealed that patients with darker skin tones were almost three times more likely to have undetected hypoxia compared to patients with lighter skin tones [6], hindering accurate decision-making regarding a patient ’ s treatment [7].”
Skin tone and clinical dataset from a prospective trial on acute care patients
Hao, S., de Matos, J. C. R. G., Dempsey, K., Alwakeel, M., Houghtaling, J., Hong, C., Gichoya, J. W., Kibbe, W., Pencina, M. J., Cox, C. E., et al. (2026) · Scientific Data · doi:10.1038/s41597-025-06457-9
“Studies also suggest that conventional calibration methods for pulse oximeters may not account for physiological variations across different skin tones, underscoring the need for tailored approaches 13–15 .”
Perioperative outcomes and health disparities: Transforming healthcare
Ehie, O. A., Wood, A. W., Chang, O. L. B., Langnas, E. M., Vu, T.-G. T., Lilaonitkul, M., & Gropper, M. A. (2026) · Journal of the National Medical Association · doi:10.1016/j.jnma.2026.01.003
de Barros, K. B., Pinheiro, J. S., Sampaio, M. S., da Cruz, L. B., Basile-Filho, A., Guirro, R. R. D. J., & Bachmann, L. (2026) · Physiological Measurement · doi:10.1088/1361-6579/ae622c
“(Bickler et al 2005, Fawzy et al 2022, Sjolding et al 2020).”
Low-cost point-of-care wireless multimodal wearable to monitor vitals and tissue oxygen saturation
Nguyen, T. K., George, T. M., Veluvolu, V. D., Azar, T. S., Frankenberg, A., Park, J., Park, J., Martin, W. S., Khaksari, K., Shadgan, B., et al. (2026) · Journal of Physics Photonics · doi:10.1088/2515-7647/ae637f
Spectral emission profile and wavelength tolerances affect pulse oximeter performancePreprint
Reiser, M. F., Breidenassel, A., & Amft, O. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.04.20.719559
Ortiz-Ruiz, G., Garay-Fernández, M., Tuta-Quintero, E., Bastidas, A. R., Lara, A., Márquez, A. M., Aponte, C., Guevara, J., & Güezguan-Pérez, J. A. (2026) · Advances in respiratory medicine · doi:10.3390/arm94030028
“However, its accuracy may be influenced by factors affecting pulse oximetry performance, including hyperbilirubinemia, dark skin pigmentation, and the use of high-dose vasopressors [9,12].”
Pilot Study: Using Reflectance as a Proxy for Skin Pigmentation to Assess Pulse Oximeter Accuracy
Branstad, A., Buttar, S., Kabir, R., Byrapuneni, S., Eid, M., Sandozi, A., BELTRAND, E., Evans, M. D., Joy, B. F., Goertzen, L., et al. (2026) · Hospital Pediatrics · doi:10.1542/hpeds.2025-008471
Perioperative outcomes: Transforming healthcare
Ehie, O. A., Wood, A. W., Chang, O. L. B., Langnas, E. M., Vu, T.-G. T., Lilaonitkul, M., & Gropper, M. A. (2026) · Journal of the National Medical Association · doi:10.1016/j.jnma.2026.05.008
Evaluation of the accuracy of pulse oximeters in real clinical settings
Leeudomwong, T., & Sanponpute, T. (2026) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ae826b
“Numerous studies have compared device performance across manufacturers [10, 11], assessed accuracy under specific clinical conditions [8, 9], and investigated bias related to skin pigmentation [12].”
Villavicencio, R. E. A., Panata, S. A. T., & Benavides, J. A. C. (2026) · Revista Científica Multidisciplinar G-nerando · doi:10.66473/rcmg.v7i2.1274
Zhang, H., & Jetley, G. (2026) · Production and Operations Management · doi:10.1177/10591478261474395
Moon, S. I., & Kwon, N. K. (2026) · Electronics · doi:10.3390/electronics15153477
Watson, J. N., Prough, D. S., & Giarracco, D. J. (2026) · IEEE Access · doi:10.1109/access.2026.3723237
Pulse Ox+: An Exploratory Model for Melanin-Corrected Pulse Oximetry
Herman, M., Knezevic, I., Lorenzo, J., Lu, J., Reilly, H., Chandrashekaran, P., Sridhar, M., & Maheswaran, B. (2026) · ASEE Northeast Section Conference (ASEE-NE) · doi:10.18260/1-2--58595
Suresha, C. H. (2026) · International Journal of Modern Science and Research Technology · doi:10.67317/ijmsrt26aug031
Spectral Feature-Driven Explainable Framework for Hypoxemia Detection Using Photoplethysmography
Meghana, P., Mohammad, F. H., & Adithya, T. S. H. S. (2026) · IETE Technical Review · doi:10.1080/02564602.2026.2716356
Mechanical Contact Conditions in Wearable Reflectance Photoplethysmography: Scoping Review
Yang, C., Xie, J., He, Z., Li, J., & Liu, C. (2026) · JMIR mhealth and uhealth · doi:10.2196/99333
Sanjeev, O. P., Mani, U. A., Singh, T. K., & Kumari, P. (2026) · Emergency Care Journal · doi:10.4081/ecj.2026.15247
Lee, W., 신재연, Song, M., Song, M.-K., Kim, K. M., Noh, G., & Cho, S. P. (2026) · Sensors · doi:10.3390/s26175676
“Skin-pigmentation-related bias is particularly relevant to pulse oximetry; previous studies have reported greater SpO 2 –SaO 2 errors, overestimation during hypoxemia, and a higher prevalence of occult hypoxemia in individuals with darker skin pigmentation [62–64].”
Ţiţoc, O.-A., & Tudose, D. (2026) · Results in Engineering · doi:10.1016/j.rineng.2026.113072
Peripheral Perfusion: The Missing Physiological Variable in ICU Pulse Oximetry Accuracy Studies
Sjödin, C., Lindstrand, F. O., Karlsson, N., & Widing, H. (2026) · Nursing in Critical Care · doi:10.1111/nicc.70708
Osorio-Sanchez, L., May, J. M., & Kyriacou, P. A. (2026) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-032-39674-7_17
Feitoza, L. F. B. B. (2024) · K-State Research Exchange (Kansas State University)
Supporting mental health and wellness of bioengineering students
Miller, I. (2022) · Illinois Digital Environment for Access to Learning and Scholarship (University of Illinois at Urbana-Champaign)
Zhang, Y., Gao, L., Zhu, W., Chen, W., Liu, Y., Chen, H., 王艳玲, Li, Y., Huang, F., Tang, J., et al. (2026) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-026-01500-5
Gao, X., Wang, J., Xiong, L., Hu, B., Jiang, C., Li, J., Chen, S., & Zhang, C. (2026) · Frontiers in Medicine · doi:10.3389/fmed.2026.1935082
Shu, X., & Wen, X. (2026) · Biosensors · doi:10.3390/bios16100565
Effects of Skin Pigmentation on Pulse Oximetry: A Pilot Study
Cotton, S. A., Dahdoul, T., Best, N. C., Guo, Y., Lee, J.-A., Malhotra, A., & McGuire, W. C. (2026) · American Journal of Critical Care · doi:10.4037/ajcc2026253
Metadata
- Authors
- Bickler, P. E., Feiner, J. R., & Severinghaus, J. W.
- Affiliations
- Professor.
- Associate Professor.
- Professor Emeritus.
- ORCID
- Bickler, Philip E.
- Venue
- Anesthesiology
- Published
- 2005-04-01
- Volume
- 102
- Issue
- 4
- Pages
- 715-719
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2142600751
- Semantic Scholar
- edacc3866d8c23322d36e03c8162e9837d9416f8
- Cited by
- 427
- References
- 16
- Citations by year
- 2026: 35 · 2025: 46 · 2024: 49 · 2023: 54 · 2022: 86 · 2021: 36 · 2020: 20 · 2019: 12 · 2018: 7 · 2017: 5 · 2016: 10 · 2015: 6 · 2014: 4 · 2013: 10 · 2012: 10
- FWCI
- 4.93
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Anesthesiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, skin pigmentation, dark skin, arterial oxygen saturation, low oxygen saturation, hypoxia, pulse oximeter accuracy, desaturation, end-tidal oxygen
- MeSH terms
- Adult, Black or African American, Female, Humans, Male, Middle Aged, Oximetry, Oxygen, Oxyhemoglobins, Skin Pigmentation, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.