Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact foot-pad electrodes
Abstract
Conventional single frequency bioimpedance analysis (BIA) systems require technician placement of arm and leg gel electrodes, a suitable location for recumbent measurements, and a separate measurement of body weight. The aim of this study was to evaluate a new single frequency 50 KHz leg-to-leg bioimpedance analysis (BIA) system combined with a digital scale that employs stainless steel pressure-contact foot pad electrodes for standing impedance and body weight measurements. Healthy adults were evaluated for 1) electrode validity and 2) potential for body component estimation. Pressure-contact foot-pad electrode measured impedance was highly correlated with (N = 9, r = 0.99, P < 0.001) impedance measured using conventional gel electrodes applied to the plantar surface of both lower extremities; mean (+/-SD) impedance was systematically higher by about 15 ohms for pressure contact electrodes (526 +/- 56 ohms vs 511 +/- 59 ohms; P < 0.001). Second, the relationship between stature-adjusted leg-to-leg impedance (H2/Z) measured by the new system and two body composition components (total body water by 3H2O dilution (N = 144); and fat-free body mass, by underwater weighing and dual x-ray absorptiometry (N = 231)) was modeled using multiple regression analysis. Correlation coefficients for H2/Z alone versus body composition components were lower for leg-to-leg BIA than for arm-to-leg BIA; correlation coefficients and SEEs became similar for the leg-to-leg and arm-to-leg BIA systems with addition of three covariates (age, gender, and waist/hip circumference ratio) to regression models. The leg-to-leg pressure contact electrode BIA system has overall performance characteristics for impedance measurement and body composition analysis similar to conventional arm-to-leg gel electrode BIA and offers the advantage of increased speed and ease of measurement.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Assessment of fat-free mass using bioelectrical impedance measurements of the human body
Lukaski, H., Johnson, P., Bolonchuk, W., & Lykken, G. (1985) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/41.4.810 · Cited by 1,896
Estimation of total body water by bioelectrical impedance analysis
Kushner, R., & Schoeller, D. (1986) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/44.3.417 · Cited by 772
Bioelectrical Impedance Analysis: A Review of Principles and Applications
Kushner, R. F. (1992) · Journal of the American College of Nutrition · doi:10.1080/07315724.1992.12098245 · Cited by 625
A simplified method for determination of residual lung volumes.
Wilmore, J. H. (1969) · Journal of Applied Physiology · doi:10.1152/jappl.1969.27.1.96 · Cited by 533
Estimation of human body composition by electrical impedance methods: a comparative study
Segal, K. R., Gutin, B., Presta, E., Wang, J., & Van Itallie, T. B. (1985) · Journal of Applied Physiology · doi:10.1152/jappl.1985.58.5.1565 · Cited by 464
Why bioelectrical impedance analysis should be used for estimating adiposity.
Houtkooper, L., Lohman, T., Going, S., & Howell, W. (1996) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.436s · Cited by 361
Heymsfield, S., Lichtman, S., Baumgartner, R., Wang, J., Kamen, Y., Aliprantis, A., & Pierson, R. (1990) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/52.1.52 · Cited by 280
Self-citationEstimation of body composition from bioelectric impedance of body segments
Baumgartner, R. N., Chumlea, W. C., & Roche, A. F. (1989) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/50.2.221 · Cited by 201
Bioelectrical Impedance and Body Composition: Present Status and Future Directions
Chumlea, W. C., & Guo, S. S. (2009) · Nutrition Reviews · doi:10.1111/j.1753-4887.1994.tb01404.x · Cited by 194
Cited by
Khalil, S., Mohktar, M., & Ibrahim, F. (2014) · Sensors · doi:10.3390/s140610895 · Cited by 712
Assessment methods in human body composition
Lee, S. Y., & Gallagher, D. (2008) · Current Opinion in Clinical Nutrition & Metabolic Care · doi:10.1097/mco.0b013e32830b5f23 · Cited by 694
Self-citationNew bioimpedance analysis system: improved phenotyping with whole-body analysis
Pietrobelli, A., Rubiano, F., St-Onge, M.-P., & Heymsfield, S. B. (2004) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1601993 · Cited by 379
Self-citationJebb, S. A., Cole, T. J., Doman, D., Murgatroyd, P. R., & Prentice, A. M. (2000) · British Journal of Nutrition · doi:10.1017/s0007114500000155 · Cited by 378
Fatty liver in non‐alcoholic non‐overweight Japanese adults: Incidence and clinical characteristics
Omagari, K., Kadokawa, Y., Masuda, J., Egawa, I., Sawa, T., Hazama, H., Ohba, K., Isomoto, H., Mizuta, Y., Hayashida, K., et al. (2002) · Journal of Gastroenterology and Hepatology · doi:10.1046/j.1440-1746.2002.02846.x · Cited by 275
Metadata
- Authors
- NU??EZ, C., GALLAGHER, D., VISSER, M., PI-SUNYER, F. X., WANG, Z., & HEYMSFIELD, S. B.
- Venue
- Medicine & Science in Sports & Exercise
- Published
- 1997-04
- Volume
- 29
- Issue
- 4
- Pages
- 524-531
- ISSN
- 0195-9131 (ISSN-L)
- OpenAlex
- W2021830033
- Cited by
- 281
- References
- 17
- FWCI
- 3.44
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact foot-pad electrodes. https://app.cochat.ai/papers/10.1097/00005768-199704000-00015
- The paper
- NU??EZ, C., GALLAGHER, D., VISSER, M., PI-SUNYER, F. X., WANG, Z., & HEYMSFIELD, S. B. (1997). Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact foot-pad electrodes. Medicine & Science in Sports & Exercise. https://doi.org/10.1097/00005768-199704000-00015
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.