Spinal Pedicle Fixation: Reliability and Validity of Roentgenogram-Based Assessment and Surgical Factors on Successful Screw Placement
The DOI matches its Crossref record.
Abstract
The increased popularity of pedicle fixation prompted research to address two issues: the reliability and validity of roentgenograms as a technique for evaluating the success of pedicle fixation, and the effects of surgical factors on successful fixation. Thus, does approach—the point and angle of screw insertion, surgeon experience, practice, level of the spine involved, and screw size—effect success of pedicle fixation? Eight fresh thoracolumbar spines were harvested and cleaned of all soft tissues. Two surgeons, one more experienced in pedicle fixation than the other, used two pedicle fixation approaches (Weinstein and Roy-Camille) on both the left and right sides at levels T11–S1 of each specimen. All screws were placed under anteroposterior (AP) and lateral c-arm control. For specimens 1 to 3, 5.5 mm screws were used at T11–L1, and 7.0 mm screws at L2–S1. Unacceptable failure rates at L2 and L3 for the first three specimens resulted in a change of instrumentation for the remaining specimens, with 5.5 mm screws used at T11–L3 and 7.0 mm screws at L4–S1. When surgeons completed the fixations for a specimen, AP and lateral roentgenograms were taken and both surgeons independently evaluated the films to assess the success of each fixation. Failure was defined as evidence of any cortical perforation on any side of the pedicle in or outside of the spinal canal. After completing the roentgenogram evaluation, the specimen was transected in the midline, and the success of each pedicle fixation was evaluated by visual/tactile inspection. There were no disagreements between surgeons on the visual/tactile evaluations of the specimens. In contrast, inter-rater adjusted percent agreement for roentgenograms ratings was 74, judged to be less than satisfactory when considering that the ratings were on a 2-point scale (S or F). The overall failure rate was approximately 21% (26/124). Of the 26 failures, 92% represented cortical perforations within the spinal canal. Discrepancies between visual/tactile and roentgenogram-based evaluations were not encouraging. Occurrence of false-positive results was at rates of 8.1% and 6.5%, and false-negative results occurred at a rate of 14.5% amd 12.9% for the more and less experienced surgeons, respectively. In general, success rate was independent of surgical factors. Success was not significantly related to approach, surgeon experience, screw size, or spine level. There was, however, an appreciable practice effect, χ2 = 8.84, P < 0.003. Failure rates were 26.4% compared to 6.4% for specimens 1–4 and 5–8, respectively. Also of interest was a trend for success to be related to approach, depending on the region of the spine involved, F1,100=3.38, P<0.07. Examination of relative success rates indicated no significant differences between the Weinstein and Roy-Camille approach in the upper lumbar spine (T11–L2), but a trend toward greater success with the Weinstein approach in the lower lumbar spine (L3–S1); a 93.1% success rate for the Weinstein, compared with 78.6% for the Roy-Camille approach. Roentgenograms were found to produce unacceptably high rates of false-positive and false-negative evaluations. The lack of differences in success rates between the surgeons with different levels of experience in conjunction with a strong relationship between success rate and practice is consistent with poor roentgenogram evaluation. Surgeons cannot be expected to improve with experience when their tools (roentgenograms) do not allow accurate evaluation of their performance. However, significant improvement in success rate can be expected when accurate evaluation is provided. Unfortunately, this research exposes the inadequacies of current evaluation procedures without providing viable alternatives. The trend toward superior success rates with the Weinstein approach in the lower lumbar spine is particularly important since the approach is believed to have the added advantage of providing less interference with uninvolved adjacent motion segments. Such an advantage would be of little value if the success rate was not at least as good as demonstrated for other approaches.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Free Hand Pedicle Screw Placement in the Thoracic Spine: Is it Safe?
Kim, Y. J., Lenke, L. G., Bridwell, K. H., Cho, Y. S., & Riew, K. D. (2004) · Spine · doi:10.1097/01.brs.0000109983.12113.9b · Cited by 706
Complications Associated with Pedicle Screws*
LONSTEIN, J. E., DENIS, F., PERRA, J. H., PINTO, M. R., SMITH, M. D., & WINTER, R. B. (1999) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199911000-00003 · Cited by 542
Amiot, L.-P., Lang, K., Putzier, M., Zippel, H., & Labelle, H. (2000) · Spine · doi:10.1097/00007632-200003010-00012 · Cited by 450
Thomsen, K., Christensen, F. B., Eiskjær, S. P., Hansen, E. S., Fruensgaard, S., & Bünger, C. E. (1997) · Spine · doi:10.1097/00007632-199712150-00004 · Cited by 412
Rajasekaran, S., Vidyadhara, S., Ramesh, P., & Shetty, A. P. (2007) · Spine · doi:10.1097/01.brs.0000252094.64857.ab · Cited by 358
Accuracy of Computer-Assisted Pedicle Screw Placement: An In Vivo Computed Tomography Analysis
Schwarzenbach, O., Berlemann, U., Jost, B., Visarius, H., Arm, E., Langlotz, F., Nolte, L.-P., & Ozdoba, C. (1997) · Spine · doi:10.1097/00007632-199702150-00020 · Cited by 284
Laine, T., Schlenzka, D., Mäkitalo, K., Tallroth, K., Nolte, L.-P., & Visarius, H. (1997) · Spine · doi:10.1097/00007632-199706010-00018 · Cited by 277
Pedicle Screw Placement Using Image Guided Techniques
Merloz, P., Tonetti, J., Pittet, L., Coulomb, M., Lavalle??, S., & Sautot, P. (1998) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-199809000-00006 · Cited by 260
Application of frameless stereotaxy to pedicle screw fixation of the spine
Kalfas, I. H., Kormos, D. W., Murphy, M. A., McKenzie, R. L., Barnett, G. H., Bell, G. R., Steiner, C. P., Trimble, M. B., & Weisenberger, J. P. (1995) · Journal of Neurosurgery · doi:10.3171/jns.1995.83.4.0641 · Cited by 247
Stereotactic Navigation for Placement of Pedicle Screws in the Thoracic Spine
Youkilis, A. S., Quint, D. J., McGillicuddy, J. E., & Papadopoulos, S. M. (2001) · Neurosurgery · doi:10.1097/00006123-200104000-00015 · Cited by 232
Accuracy Requirements for Image-Guided Spinal Pedicle Screw Placement
Rampersaud, Y. R., Simon, D. A., & Foley, K. T. (2001) · Spine · doi:10.1097/00007632-200102150-00010 · Cited by 224
Image-guided spine surgery: state of the art and future directions
Tjardes, T., Shafizadeh, S., Rixen, D., Paffrath, T., Bouillon, B., Steinhausen, E. S., & Baethis, H. (2010) · European Spine Journal · doi:10.1007/s00586-009-1091-9 · Cited by 206
“PLIF. Especially the accuracy of pedicle screw placement has been lively discussed as misplacement rates of up to 30% in the lumbar spine and up to 55% in the thoracic spine have been reported [30, 107 , 110].”
BackgroundFoley, K. T., & Smith, M. M. (1996) · Neurosurgery Clinics of North America · doi:10.1016/s1042-3680(18)30385-1 · Cited by 195
Biomaterials in Spinal Implants: A Review
Warburton, A., Girdler, S. J., Mikhail, C. M., Ahn, A., & Cho, S. K. (2020) · Neurospine · doi:10.14245/ns.1938296.148 · Cited by 182
Clinical Accuracy of Fluoroscopic Computer-Assisted Pedicle Screw Fixation: A CT Analysis
Rampersaud, Y. R., Pik, J. H. T., Salonen, D., & Farooq, S. (2005) · Spine · doi:10.1097/01.brs.0000157490.65706.38 · Cited by 160
A new approach to computer-aided spine surgery: fluoroscopy-based surgical navigation
Nolte, L. P., Slomczykowski, M. A., Berlemann, U., Strauss, M. J., Hofstetter, R., Schlenzka, D., Laine, T., & Lund, T. (2000) · European Spine Journal · doi:10.1007/pl00010026 · Cited by 159
Metadata
- Authors
- WEINSTEIN, J. N., SPRATT, K. F., SPENGLER, D., BRICK, C., & REID, S.
- Venue
- Spine
- Published
- 1988-09
- Volume
- 13
- Issue
- 9
- Pages
- 1012-1018
- ISSN
- 0362-2436 (ISSN-L)
- OpenAlex
- W2029636776
- Cited by
- 532
- References
- 0
- FWCI
- 2.06
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.