Sources of inaccuracy in the measurement of adult patients’ resting blood pressure in clinical settings: a systematic review
Open accessThe DOI matches its Crossref record; 386 of the 390 references it cites were found in a registry or an index, and 1 are not checked yet.
Abstract
BACKGROUND: To interpret blood pressure (BP) data appropriately, healthcare providers need to be knowledgeable of the factors that can potentially impact the accuracy of BP measurement and contribute to variability between measurements. METHODS: A systematic review of studies quantifying BP measurement inaccuracy. Medline and CINAHL databases were searched for empirical articles and systematic reviews published up to June 2015. Empirical articles were included if they reported a study that was relevant to the measurement of adult patients' resting BP at the upper arm in a clinical setting (e.g. ward or office); identified a specific source of inaccuracy; and quantified its effect. Reference lists and reviews were searched for additional articles. RESULTS: A total of 328 empirical studies were included. They investigated 29 potential sources of inaccuracy, categorized as relating to the patient, device, procedure or observer. Significant directional effects were found for 27; however, for some, the effects were inconsistent in direction. Compared with true resting BP, significant effects of individual sources ranged from -23.6 to +33 mmHg SBP and -14 to +23 mmHg DBP. CONCLUSION: A single BP value outside the expected range should be interpreted with caution and not taken as a definitive indicator of clinical deterioration. Where a measurement is abnormally high or low, further measurements should be taken and averaged. Wherever possible, BP values should be recorded graphically within ranges. This may reduce the impact of sources of inaccuracy and reduce the scope for misinterpretations based on small, likely erroneous or misleading, changes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“On a beat-to-beat basis, BP levels can vary markedly (from –24 mmHg to 33 mmHg) in response to usual extrinsic and behavioural factors4”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [4]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“the highly variable nature of office and out-of-office BP readings has been widely acknowledged”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [4]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Standardization is essential because there are many aspects of BP measurement, including the patient’s physical condition, clinic environment, and equipment that can readily result in inaccurate and unreliable values.23,25,26”
From International Consensus on Standardized Clinic Blood Pressure Measurement – A Call to Action (2023) · cites this paper as [25]
In the citing paper
“Even when a validated automated device is used, the accuracy of BP measurement may be influenced by many factors. 19 – 21”
From Hypertension Canada guideline for the diagnosis and treatment of hypertension in adults in primary care (2025) · cites this paper as [19]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Surviving Sepsis Campaign: International Guidelines for Management of Severe Sepsis and Septic Shock
Dellinger, R. P., Levy, M. M., Rhodes, A., Annane, D., Gerlach, H., Opal, S. M., Sevransky, J. E., Sprung, C. L., Douglas, I. S., Jaeschke, R., et al. (2013) · Critical Care Medicine · doi:10.1097/ccm.0b013e31827e83af · Cited by 9,694
The Surviving Sepsis Campaign Guidelines Committee including The Pediatric Subgroup*, Dellinger, R. P., Levy, M. M., Rhodes, A., Annane, D., Gerlach, H., Opal, S. M., Sevransky, J. E., Sprung, C. L., Douglas, I. S., et al. (2013) · Intensive Care Medicine · doi:10.1007/s00134-012-2769-8 · Cited by 7,290
Intravenous Fluids and Acute Kidney Injury
Ding, X., Cheng, Z., & Qian, Q. (2017) · Blood Purification · doi:10.1159/000452702 · Cited by 6,827
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220 · cited 1×
“Therefore, when interpreting the accuracy of a device, both the mean bias and the width of the limits of agreement should be considered [165,166].”
BackgroundPickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051 · cited 4×
“Generally, this change is assumed to involve an increase in both SBP and DBP from normal daytime values [4].”
BackgroundMethodInfluentialHuman blood pressure determination by sphygmomanometry.
Perloff, D., Grim, C., Flack, J., Frohlich, E. D., Hill, M., McDonald, M., & Morgenstern, B. Z. (1993) · Circulation · doi:10.1161/01.cir.88.5.2460 · Cited by 1,591
Williams, B., Poulter, N. R., Brown, M. J., Davis, M., McInnes, G. T., Potter, J. F., Sever, P. S., & McG Thom, S. (2004) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001683 · Cited by 1,272
Effects of Caffeine on Plasma Renin Activity, Catecholamines and Blood Pressure
Robertson, D., Frölich, J. C., Carr, R. K., Watson, J. T., Hollifield, J. W., Shand, D. G., & Oates, J. A. (1978) · New England Journal of Medicine · doi:10.1056/nejm197801262980403 · Cited by 757
Cited by
Items 101–200 of 422
Hypertension: The most common chronic health problem in Spain. A call to action
Banegas, J. R., & Gijón‐Conde, T. (2022) · Hipertensión y Riesgo Vascular · doi:10.1016/j.hipert.2022.03.004 · Cited by 17
Resonance sonomanometry for noninvasive, continuous monitoring of blood pressure
Jimenez, R., Yurk, D., Dell, S. J., Rutledge, A. C., Fu, M., Dempsey, W. P., Abu‐Mostafa, Y. S., Rajagopal, A., & Rajagopal, A. B. (2024) · PNAS Nexus · doi:10.1093/pnasnexus/pgae252 · Cited by 16
Digit Preference in Office Blood Pressure Measurements, United States 2015–2019
Foti, K. E., Appel, L. J., Matsushita, K., Coresh, J., Alexander, G. C., & Selvin, E. (2020) · American Journal of Hypertension · doi:10.1093/ajh/hpaa196 · Cited by 16
Boonyasai, R. T., McCannon, E. L., & Landavaso, J. E. (2019) · Current Hypertension Reports · doi:10.1007/s11906-019-0936-9 · Cited by 15
Medical Telemonitoring for the Management of Hypertension in Older Patients in Japan
Fujiwara, T., Sheppard, J. P., Hoshide, S., Kario, K., & McManus, R. J. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20032227 · Cited by 14
Elzeky, M. E. H., & Shahine, N. F. M. (2022) · BMC Nursing · doi:10.1186/s12912-022-01137-0 · Cited by 14
Mancusi, C., Bisogni, V., Maloberti, A., Manzi, M. V., Visco, V., Biolcati, M., Giani, V., Spannella, F., Monticone, S., Saladini, F., et al. (2022) · Blood Pressure · doi:10.1080/08037051.2022.2137461 · Cited by 14
van Putten, L. D., & Bamford, K. E. (2023) · IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops (CVPRW) · doi:10.1109/cvprw59228.2023.00634 · Cited by 14
Burger, P. M., Monpellier, V. M., Deden, L. N., Kooiman, L., Liem, R. S. L., Hazebroek, E. J., Janssen, I. M. C., & Westerink, J. (2020) · Surgery for Obesity and Related Diseases · doi:10.1016/j.soard.2020.07.011 · Cited by 14
Ebinger, J. E., Driver, M. P., Huang, T. Y., Magraner, J., Botting, P. G., Wang, M., Chen, P., Bello, N. A., Ouyang, D., Theurer, J., et al. (2024) · Scientific Reports · doi:10.1038/s41598-024-52406-8 · Cited by 14
Picone, D. S., Schultz, M. G., Armstrong, M. K., Black, J. A., Bos, W. J. W., Chen, C., Cheng, H., Crémer, A., Dwyer, N., Hughes, A. D., et al. (2021) · Hypertension · doi:10.1161/hypertensionaha.120.16109 · Cited by 14
Webb, N., Orme, M. E., Witkowski, M., Nakanishi, R., & Langer, J. (2018) · Diabetes Therapy · doi:10.1007/s13300-018-0397-1 · Cited by 13
Andreadis, E. A., Geladari, C. V., & Angelopoulos, E. T. (2019) · Journal of Clinical Hypertension · doi:10.1111/jch.13748 · Cited by 13
Standardized home blood pressure monitoring: Rationale behind the 722 protocol
Lin, H., Pan, H., Chen, C., Cheng, H., Chia, Y. C., Sogunuru, G. P., Tay, J. C., Turana, Y., Verma, N., Kario, K., et al. (2022) · Journal of Clinical Hypertension · doi:10.1111/jch.14549 · Cited by 13
Optimum waveform envelopes and amplitude ratios in oscillometric blood pressure estimation
Alvarado Alvarez, M., Padwal, R., Ringrose, J., Jalali, A., & Hiebert, W. (2021) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000485 · Cited by 12
The determinants of complication trajectories in American Indians with type 2 diabetes
Reynolds, E. L., Akıncı, G., Banerjee, M., Looker, H. C., Patterson, A., Nelson, R. G., Feldman, E. L., & Callaghan, B. C. (2021) · JCI Insight · doi:10.1172/jci.insight.146849 · Cited by 12
Chapman, N., Jayasinghe, S., Moore, M. N., Picone, D. S., Schultz, M. G., Jose, M., McCallum, R. W., Armstrong, M. K., Peng, X., Marwick, T. H., et al. (2024) · Hypertension Research · doi:10.1038/s41440-024-01841-1 · Cited by 12
Farahani, S., Farahani, I., Burckhardt, B. B., Schwender, H. R., & Laeer, S. (2020) · Pharmacy · doi:10.3390/pharmacy8040217 · Cited by 12
Clark, C. E., Thomas, D., Llewellyn, D. J., Ferrucci, L., Bandinelli, S., & Campbell, J. L. (2020) · British Journal of General Practice · doi:10.3399/bjgp20x709589 · Cited by 12
Rodrigues, J. A. L., Yamane, A. C., Gonçalves, T. C. P., Kalva‐Filho, C. A., Papoti, M., & Júnior, C. R. B. (2019) · IJC Heart & Vasculature · doi:10.1016/j.ijcha.2019.100378 · Cited by 12
Kumar, D., Singh, S., Kashyap, V., & Singh, S. K. (2020) · Journal of Family Medicine and Primary Care · doi:10.4103/jfmpc.jfmpc_796_19 · Cited by 11
Cuff Size Variation Across Manufacturers of Home Blood Pressure Devices: A Current Patient Dilemma
Shahi, S., Jackson, S. L., Streeter, T. E., He, S., & Wall, H. K. (2023) · American Journal of Hypertension · doi:10.1093/ajh/hpad060 · Cited by 11
Hypertension Detection based on Machine Learning
Marin, I., & Goga, N. (2019) · doi:10.1145/3352700.3352723 · Cited by 11
Calleja-Romero, A., Vicente‐Rodríguez, G., & Garatachea, N. (2021) · Journal of Sports Sciences · doi:10.1080/02640414.2021.1986276 · Cited by 10
Fink, J. T., Magnan, E. M., Johnson, H. M., Bednarz, L., Allen, G. O., Greenlee, R. T., Bolt, D. M., & Smith, M. A. (2018) · High Blood Pressure & Cardiovascular Prevention · doi:10.1007/s40292-018-0284-x · Cited by 10
Masrouri, S., Tabnak, P., Chevli, P. A., Mirzai, S., Slipczuk, L. N., Vasu, S., Lima, J. A. C., & Shapiro, M. D. (2025) · European Journal of Preventive Cardiology · doi:10.1093/eurjpc/zwaf163 · Cited by 10
An Ongoing Challenge: Why Do Primary Care Providers Struggle to Adhere to Blood Pressure Guidelines?
Mitsnefes, M. M., & Bolling, C. F. (2021) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2021.12.005 · Cited by 10
Automated blood pressure measuring devices: how are they clinically validated for accuracy?
Ringrose, J. S., & Padwal, R. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00761-2 · Cited by 9
“, antecedent rest, effects of cold/heat, taking during the measurement, ambient noise, full bladder, smoking, poor technique, inappropriate cuff application or size, incorrect positioning, arrhythmia) [7].”
BackgroundWilkes, M., Kramer, A., Pugmire, J., Pilkington, C., Zaniello, B., & Zahradka, N. (2024) · Journal of Medical Internet Research · doi:10.2196/58888 · Cited by 9
Telemonitoring and remote counseling in hypertensive patients. Looking for new ways to do old jobs
Васильевич, И. М., Звартау, Н. Э., Emelyanov, I. V., & Конради, А. О. (2019) · Arterial’naya Gipertenziya (Arterial Hypertension) · doi:10.18705/1607-419x-2019-25-4-337-356 · Cited by 9
Blood pressure targets in chronic kidney disease: still no consensus
Lucas, B. J., & Taal, M. W. (2023) · Current Opinion in Nephrology & Hypertension · doi:10.1097/mnh.0000000000000920 · Cited by 9
Goupil, R., Tsuyuki, R. T., Santesso, N. A., Terenzi, K. A., Habert, J., Cheng, G., Gysel, S. C., Bruneau, J., Leung, A. A. C., Campbell, N. R. C., et al. (2025) · Canadian Family Physician · doi:10.46747/cfp.20250616001 · Cited by 9
The 2017 U.S. Hypertension Guidelines: What Is Important for Older Adults?
Cushman, W. C., & Johnson, K. (2018) · Journal of the American Geriatrics Society · doi:10.1111/jgs.15395 · Cited by 9
Clinical Validation of a Soft Wireless Continuous Blood Pressure Sensor During Surgery
Chou, E., Cheung, S. Y. C., Maxwell, H., Pham, N. K., Khine, M., & Rinehart, J. B. (2021) · Frontiers in Digital Health · doi:10.3389/fdgth.2021.696606 · Cited by 9
Katigbak, C., & Fontenot, H. B. (2018) · Nursing for Women s Health · doi:10.1016/j.nwh.2018.06.003 · Cited by 9
Nagaraju, S. P., Shenoy, S. V., Rao, I. R., Bhojaraja, M. V., Rangaswamy, D., & Prabhu, R. A. (2022) · International Journal of Nephrology and Renovascular Disease · doi:10.2147/ijnrd.s343582 · Cited by 9
Remote Blood Pressure Monitoring: A Comprehensive Review
Berlot, A., Berman, J., Borad, A., Mason, T. D., Petriceks, A. H., & Juraschek, S. P. (2025) · American Journal of Hypertension · doi:10.1093/ajh/hpaf097 · Cited by 9
Billups, S. J., Saseen, J. J., Griend, J. P. V., & Schilling, L. M. (2018) · Journal of Clinical Hypertension · doi:10.1111/jch.13345 · Cited by 9
Zachwieja, J., Neyman‐Bartkowiak, A., Rabiega, A., Wojciechowska, M., Barabasz, M., Musielak, A., Silska‐Dittmar, M., & Ostalska‐Nowicka, D. (2020) · Clinical and Experimental Hypertension · doi:10.1080/10641963.2020.1714642 · Cited by 9
Blood pressure measurement technique in clinical practice in the NHS Greater Glasgow and Clyde
Setjiadi, D., Geddes, C., & Delles, C. (2024) · Journal of Human Hypertension · doi:10.1038/s41371-024-00984-5 · Cited by 9
Muldoon, M. F., Kronish, I. M., & Shimbo, D. (2018) · Circulation Cardiovascular Quality and Outcomes · doi:10.1161/circoutcomes.117.004543 · Cited by 9
German, C. A., Elfassy, T., Singleton, M. J., Rodríguez, C. J., Ambrosius, W. T., & Yeboah, J. (2021) · American Journal of Hypertension · doi:10.1093/ajh/hpab059 · Cited by 9
Xu, J., Lin, J., Gan, H., He, Q., Wang, W., & Liu, Y. (2025) · Frontiers in Pharmacology · doi:10.3389/fphar.2025.1585061 · Cited by 8
The Development of an Educational Video on Blood Pressure Measurement for Pharmacy Students
Farahani, S., Farahani, I., Burckhardt, B. B., Monser, K., & Laeer, S. (2021) · Advances in Medical Education and Practice · doi:10.2147/amep.s302728 · Cited by 9
Ji, W., Chizuk, H. M., Leddy, J. J., Sisto, S. A., & Haider, M. N. (2024) · Physiological Reports · doi:10.14814/phy2.70114 · Cited by 8
The effect of arterial stiffness on cuff-based blood pressure measurement
Ma, Y., Chen, Y., Ma, Y., & Feng, X. (2021) · Extreme Mechanics Letters · doi:10.1016/j.eml.2021.101298 · Cited by 8
Padwal, R. S., Campbell, N. R. C., Schutte, A. E., Olsen, M. H., Delles, C., Etyang, A. O., Cruickshank, J. K., Stergiou, G. S., Rakotz, M., Wozniak, G., et al. (2020) · Revista Panamericana de Salud Pública · doi:10.26633/rpsp.2020.88 · Cited by 8
Der hypertensive Notfall: Empfehlungen zum initialen medikamentösen Therapiemanagement
Strauß, M., Leischik, R., Jehn, U., Padberg, J.-S., Pistulli, R., Kümpers, P., & Reinecke, H. (2020) · Medizinische Klinik - Intensivmedizin und Notfallmedizin · doi:10.1007/s00063-020-00728-6 · Cited by 8
Implementation of Video Blood Pressure Visits in the Veterans Health Administration
Carrico, M., Frosch, C., Craig, K., Carter, M. D., Falk, J., Guerrero, S., Huang, L. L., Kossoudji, A., Michelson, T. R., Miller, P., et al. (2023) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2023.0269 · Cited by 8
Omar, H., Moamen, A., Abdelhamid, B. M., Rady, A., & Farouk, I. (2023) · Anaesthesia Pain & Intensive Care · doi:10.35975/apic.v27i4.2255 · Cited by 8
Horsley, K. J., Ramsay, J. O., Ditto, B., & Costa, D. D. (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002995 · Cited by 8
Schutte, A. E., & Gnanenthiran, S. R. (2021) · Hypertension · doi:10.1161/hypertensionaha.121.18156 · Cited by 7
Genome mining yields putative disease-associated ROMK variants with distinct defects
(2023) · doi:10.1371/journal.pgen.1011051 · Cited by 7
Arterial hypertension—clinical trials update 2024
Tokcan, M., Lauder, L., Götzinger, F., Böhm, M., & Mahfoud, F. (2024) · Hypertension Research · doi:10.1038/s41440-024-01900-7 · Cited by 7
Fonseca‐Reyes, S., Fonseca-Cortés, K., Coca, A., Romero‐Velarde, E., & Molina, J. J. P. (2022) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000629 · Cited by 7
Perceptions of pharmacists on the quality of automated blood pressure devices: a national survey
Picone, D. S., Peterson, G. M., Jackson, S., Campbell, N. R. C., Delles, C., Olsen, M. H., Padwal, R. S., Schutte, A. E., & Sharman, J. E. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00670-4 · Cited by 7
Hanevold, C. D., & Brady, T. M. (2024) · Current Hypertension Reports · doi:10.1007/s11906-024-01298-0 · Cited by 7
Kadowaki, S., Kadowaki, T., Hozawa, A., Fujiyoshi, A., Hisamatsu, T., Satoh, A., Arima, H., Tanaka, S., Torii, S., Kondo, K., et al. (2020) · Hypertension Research · doi:10.1038/s41440-020-00533-w · Cited by 7
Alpert, B. S., Schwartz, J. E., Shapiro, M., & Wexler, R. (2023) · EClinicalMedicine · doi:10.1016/j.eclinm.2023.102219 · Cited by 7
Relations between white coat effect of blood pressure and arterial stiffness
Cao, R., Yue, J., Gao, T., Sun, G., & Yang, X. (2022) · Journal of Clinical Hypertension · doi:10.1111/jch.14573 · Cited by 7
Comparability of oscillometric to simultaneous auscultatory blood pressure measurement in children
Ringrose, J., Alabbas, A., Jalali, A., Khinda, H., Morgan, C., Yiu, V., Alexander, R. T., & Padwal, R. S. (2019) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000367 · Cited by 7
There Is No 1-Size-Fits-All to Blood Pressure Measurement—Cuff Size Matters
Lalika, M., Juraschek, S. P., & Brewer, L. C. (2023) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2023.3277 · Cited by 7
Genome mining yields new disease-associated ROMK variants with distinct defects
Nguyen, N. H., Sarangi, S., McChesney, E. M., Sheng, S., Porter, A. W., Kleyman, T. R., Pitluk, Z. W., & Brodsky, J. L. (2023) · doi:10.1101/2023.05.05.539609 · Cited by 7
Edward, A., Kagaruki, G. B., Manase, F., Appel, L. J., & Matsushita, K. (2022) · BMC Health Services Research · doi:10.1186/s12913-022-08064-5 · Cited by 6
Vazeou, A., Birkebæk, N., Iotova, V., Cherubini, V., Piccini, B., Biester, T., Stipančić, G., Jefferies, C., Maffeis, C., Stergiou, G. S., et al. (2020) · Pediatric Diabetes · doi:10.1111/pedi.13114 · Cited by 6
Blood Pressure Measurement—Following a Standardized Protocol Is What Matters
Muntner, P. (2022) · American Journal of Hypertension · doi:10.1093/ajh/hpac047 · Cited by 6
Treatment of hypertension in older people
Wing, L. M. H., & Gabb, G. M. (2018) · Journal of Pharmacy Practice and Research · doi:10.1002/jppr.1417 · Cited by 6
Farahani, S., Farahani, I., Deters, M. A., Schwender, H. R., Burckhardt, B. B., & Laeer, S. (2021) · Healthcare · doi:10.3390/healthcare9070822 · Cited by 6
Michaud, A., Lamarre-Cliché, M., & Cloutier, L. (2019) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000382 · Cited by 6
Alidadi, M., Mohebbi, M., & Forouzanfar, M. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3455100 · Cited by 6
Is auscultation an issue when validating 24-h blood pressure monitoring devices?
Wuerzner, G., Vybornova, A., Wurzner-Ghajarzadeh, A., Polychronopoulou, E., Fallet, S., Solà, J., & Alpert, B. S. (2020) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000462 · Cited by 6
Вячеславовна, П. Е., Fedorova, E. Y., & Gorbunov, V. M. (2022) · CARDIOVASCULAR THERAPY AND PREVENTION · doi:10.15829/1728-8800-2022-3263 · Cited by 5
The Effect of Chlorhexidine Mouthwash on Blood Pressure: A Systematic Review and Meta‐Analysis
Toonen, L. S. J., Swaaij, B. W. M. V., Vagevuur, M. L., van der Weijden, G. A., Timmerman, M. F., & Slot, D. E. (2026) · International Journal of Dental Hygiene · doi:10.1111/idh.70035 · Cited by 5
Evidence-Based, Streamlined Approach to Measure Blood Pressure in Primary Care Settings
Foti, K. E., Moran, A. E., Matsushita, K., Appel, L. J., Juraschek, S. P., Pathni, A. K., Banigbe, B., Dessie, G. A., Belbase, B., Bhuiyan, M. R., et al. (2025) · Hypertension · doi:10.1161/hypertensionaha.125.24527 · Cited by 5
Shaver, N., Beck, A., Bennett, A., Wilson, B. J., Garritty, C., Subnath, M., Grad, R. M., Persaud, N., Thériault, G., Flemming, J. A., et al. (2024) · Systematic Reviews · doi:10.1186/s13643-023-02392-1 · Cited by 5
Deragon, V., Baati, Y., Trudelle, L., Merabtine, A., Chapman, N., Picone, D. S., & Goupil, R. (2025) · Journal of the American Heart Association · doi:10.1161/jaha.125.042797 · Cited by 5
Noninvasive pediatric blood pressure assessment: exploring the clinicians’ perspective
Stewart, N. P., Quinlan, C., Best, S., & Mynard, J. P. (2024) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000693 · Cited by 5
Smith, K. C., Gaziano, T. A., Mushlin, A. I., Cutler, D., Menzies, N. A., & Pandya, A. (2025) · Annals of Internal Medicine · doi:10.7326/annals-25-00560 · Cited by 5
Assessment of Blood Pressure: Techniques and Implications From Clinical Trials
Waguespack, D. R., & Dwyer, J. P. (2019) · Advances in Chronic Kidney Disease · doi:10.1053/j.ackd.2019.02.002 · Cited by 5
Markuleva, M. V., Сергеевич, Г. М., Gerashchenko, S. M., Накибович, Х. Р., & Ivshin, I. V. (2022) · Sensors · doi:10.3390/s22114229 · Cited by 5
Implications of Inaccurate Blood Pressure Measurement on Hypertension Prevalence
Leung, A. A. C., Hiremath, S. S., Williams, J. V. A., & Tsuyuki, R. T. (2024) · CJC Open · doi:10.1016/j.cjco.2024.10.011 · Cited by 5
Inauen, C., Boss, J. M., Mira, K., Luft, A. R., Kulcsár, Z., Willms, J. F., Bögli, S. Y., & Keller, E. (2022) · Neurocritical Care · doi:10.1007/s12028-022-01519-x · Cited by 5
Self-reported practices of doctors and nurses for the measurement of blood pressure
Kırhan, İ., Kır, S., & Dilek, M. (2020) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000482 · Cited by 5
Chandel, T., Miranda, V., Lowe, A., & Lee, T. C. (2024) · Technologies · doi:10.3390/technologies12040044 · Cited by 5
Wan, T.-X., Wu, Y., Wu, Y., Hu, W., & Su, H. (2021) · Hypertension Research · doi:10.1038/s41440-020-00595-w · Cited by 5
Salazar, M. R., Espeche, W. G., Stavile, R. N., Balbín, E., Sisnieguez, B. C. L., Sisnieguez, C. E. L., March, C. E., Cor, S., Acero, I. E., & Carbajal, H. A. (2018) · Journal of Human Hypertension · doi:10.1038/s41371-018-0057-y · Cited by 5
Vischer, A. S., & Burkard, T. (2023) · Bulletin of pioneering researches of medical and clinical science · doi:10.51847/ihm1rqlmqu · Cited by 4
A Modulation Layer to Increase Neural Network Robustness Against Data Quality Issues
Abdelhack, M., Zhang, J., Tripathi, S., Fritz, B. A., Felsky, D., Avidan, M. S., Chen, Y., & King, C. R. (2023) · Transactions on machine learning research · doi:10.48550/arxiv.2107.08574 · Cited by 4
The Need for Accurate Data on Blood Pressure Measurement in the Dental Office
Elias, M. F., & Goodell, A. L. (2020) · American Journal of Hypertension · doi:10.1093/ajh/hpaa023 · Cited by 4
Makita, L. M., Kleimmann, R. D. F., de Oliveira, R. R., Neves, H. A. F., Corso, A. M. S., Alves, V. S., Kojima, G. S. A., Koppanatham, A., Kowacs, P. A., & Piovesan, É. J. (2025) · Headache The Journal of Head and Face Pain · doi:10.1111/head.14921 · Cited by 4
Calibration‐free blood pressure estimation based on a convolutional neural network
Cho, J.-W., Shin, H., & Choi, A. (2023) · Psychophysiology · doi:10.1111/psyp.14480 · Cited by 4
Parker, R. A., Padfield, P. L., Hanley, J., Pinnock, H., Kennedy, J., Stoddart, A., Hammersley, V., Sheikh, A., & McKinstry, B. (2021) · BMC Medical Research Methodology · doi:10.1186/s12874-021-01219-8 · Cited by 4
Are blood-pressure targets needed to prevent cardiovascular disease?
Schutte, A. E. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00595-4 · Cited by 4
Navigating the Complexities of Cancer Treatment-Induced Hypertension
Arriola‐Montenegro, J., Roth, J., & Suarez, M. L. G. (2025) · Journal of Cardiovascular Development and Disease · doi:10.3390/jcdd12060235 · Cited by 4
Association of blood pressure variability during acute care hospitalization and incident dementia
Ebinger, J. E., Driver, M. P., Botting, P. G., Wang, M., Cheng, S. C., & Tan, Z. S. (2023) · Frontiers in Neurology · doi:10.3389/fneur.2023.1085885 · Cited by 4
Blood pressure measurement practices in children and adolescents within primary care setting
Čeponytė, K., Ažukaitis, K., & Jankauskienė, A. (2025) · Frontiers in Pediatrics · doi:10.3389/fped.2025.1571419 · Cited by 4
Lu, H.-W., & Hsu, H. (2022) · International Journal of General Medicine · doi:10.2147/ijgm.s392915 · Cited by 4
Systolic BP Amplification: Systematic Review and Individual Participant Meta-Analysis
Picone, D. S., Bui, T. V., Schultz, M. G., Otáhal, P., Hughes, A. D., Black, J. A., Westerhof, B. E., Chen, C.-H., Liang, F., Pucci, G., et al. (2025) · Hypertension · doi:10.1161/hypertensionaha.124.24483 · Cited by 4
Shin, Y., Chaung, S., & Kim, H. (2019) · Journal of Korean Academy of Fundamentals of Nursing · doi:10.7739/jkafn.2019.26.4.282 · Cited by 4
Vazeou, A., Tittel, S. R., Birkebæk, N., Kordonouri, O., Iotova, V., Piccini, B., Saboo, B., Lyckå, A. P., Seget, S., Maahs, D. M., et al. (2022) · Diabetes Care · doi:10.2337/dc21-2472 · Cited by 4
Metadata
- Authors
- Kallioinen, N., Hill, A., Horswill, M. S., Ward, H. E., & Watson, M. O.
- Affiliations
- The University of Queensland (AU)
- Metro North Hospital and Health Service (AU)
- Corresponding author
- Hill, Andrew
- Venue
- Journal of Hypertension
- Published
- 2017-03
- Volume
- 35
- Issue
- 3
- Pages
- 421-441
- ISSN
- 0263-6352 (ISSN-L)
- PMC
- PMC5278896
- OpenAlex
- W2563521117
- Semantic Scholar
- 17b4fd73a6b155571b2a3a3aee7dbbd6bc78909a
- Cited by
- 357
- References
- 390
- Citations by year
- 2026: 34 · 2025: 46 · 2024: 45 · 2023: 45 · 2022: 45 · 2021: 46 · 2020: 41 · 2019: 28 · 2018: 21 · 2017: 6
- FWCI
- 3.19
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Journal of Hypertension (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Funders
- University of Queensland
- Metro North Hospital and Health Service
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- blood pressure measurement, measurement error, resting blood pressure, upper arm blood pressure, measurement variability, clinical deterioration, blood pressure monitoring, measurement error sources, blood pressure interpretation
- MeSH terms
- Alcohol Drinking, Arm, Urinary Bladder, Blood Pressure, Blood Pressure Determination, Caffeine, Cold Temperature, Eating, Humans, Hypertension, Nicotine, Rest, Reproducibility of Results, White Coat Hypertension, Tobacco Use
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.