Cuffless blood pressure measuring devices: review and statement by the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability
Open accessThe DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
Background: Many cuffless blood pressure (BP) measuring devices are currently on the market claiming that they provide accurate BP measurements. These technologies have considerable potential to improve the awareness, treatment, and management of hypertension. However, recent guidelines by the European Society of Hypertension do not recommend cuffless devices for the diagnosis and management of hypertension. Objective: This statement by the European Society of Hypertension Working Group on BP Monitoring and Cardiovascular Variability presents the types of cuffless BP technologies, issues in their validation, and recommendations for clinical practice. Statements: Cuffless BP monitors constitute a wide and heterogeneous group of novel technologies and devices with different intended uses. Cuffless BP devices have specific accuracy issues, which render the established validation protocols for cuff BP devices inadequate for their validation. In 2014, the Institute of Electrical and Electronics Engineers published a standard for the validation of cuffless BP devices, and the International Organization for Standardization is currently developing another standard. The validation of cuffless devices should address issues related to the need of individual cuff calibration, the stability of measurements post calibration, the ability to track BP changes, and the implementation of machine learning technology. Clinical field investigations may also be considered and issues regarding the clinical implementation of cuffless BP readings should be investigated. Conclusion: Cuffless BP devices have considerable potential for changing the diagnosis and management of hypertension. However, fundamental questions regarding their accuracy, performance, and implementation need to be carefully addressed before they can be recommended for clinical use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Vlachopoulos, C., Aznaouridis, K., O'Rourke, M. F., Safar, M. E., Baou, K., & Stefanadis, C. (2010) · European Heart Journal · doi:10.1093/eurheartj/ehq024 · Cited by 1,363
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · cited 1×
“Copyright © 2022 Wolters Kluwer preejectionperiod variations and smoothmuscle contraction mediated by varying sympathetic nervous activity and local mechanisms in the distal arm arteries [49].”
BackgroundStergiou, G. S., Palatini, P., Parati, G., O’Brien, E., Januszewicz, A., Lurbe, E., Persu, A., Mancia, G., Kreutz, R., & on behalf of the European Society of Hypertension Council and the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002843 · Cited by 859
Self-citationCuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621
Stergiou, G. S., Alpert, B., Mieke, S., Asmar, R., Atkins, N., Eckert, S., Frick, G., Friedman, B., Graßl, T., Ichikawa, T., et al. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001634 · Cited by 278
Self-citationAccuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277
Self-citationStergiou, G. S., Palatini, P., Asmar, R., Ioannidis, J. P., Kollias, A., Lacy, P., McManus, R. J., Myers, M. G., Parati, G., Shennan, A., et al. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002039 · Cited by 218 · cited 1×
“Copyright © 2022 Wolters Kluwer Bland–Altman scatterplots have been widely used to assess the agreement between test and reference BP devices [69].”
MethodSelf-citationEvaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals
Mukkamala, R., Yavarimanesh, M., Natarajan, K., Hahn, J.-O., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S. (2021) · Hypertension · doi:10.1161/hypertensionaha.121.17747 · Cited by 211
Self-citationAllen, J., & Murray, A. (2003) · Physiological Measurement · doi:10.1088/0967-3334/24/2/306 · Cited by 202
Self-citationCuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180
Self-citation
Cited by
Mancia, G., Kreutz, R., Brunström, M., Burnier, M., Grassi, G., Januszewicz, A., Muiesan, M. L., Tsioufis, K., Agabiti-Rosei, E., Algharably, E. A. E., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003480 · Cited by 3,473
Self-citation2024 ESC Guidelines for the management of elevated blood pressure and hypertension
McEvoy, J. W., McCarthy, C. P., Bruno, R. M., Brouwers, S., Canavan, M. D., Ceconi, C., Christodorescu, R. M., Daskalopoulou, S. S., Ferro, C. J., Gerdts, E., et al. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae178 · Cited by 2,298
Self-citationWriting Committee Members*, Jones, D. W., Ferdinand, K. C., Taler, S. J., Johnson, H. M., Shimbo, D., Abdalla, M., Altieri, M. M., Bansal, N., Bello, N. A., et al. (2025) · Hypertension · doi:10.1161/hyp.0000000000000249 · Cited by 310
Jones, D. W., Ferdinand, K. C., Taler, S. J., Johnson, H. M., Shimbo, D., Abdalla, M., Altieri, M. M., Bansal, N., Bello, N. A., Bress, A. P., et al. (2025) · JACC · doi:10.1016/j.jacc.2025.05.007 · Cited by 280
Kario, K., Tomitani, N., Wang, T.-D., Park, S., Li, Y., Shin, J., Tsoi, K., Chen, C.-H., Cheng, H.-M., Siddique, S., et al. (2023) · Hypertension Research · doi:10.1038/s41440-023-01397-6 · Cited by 40
Self-citation
Metadata
- Authors
- Stergiou, G. S., Mukkamala, R., Avolio, A., Kyriakoulis, K. G., Mieke, S., Murray, A., Parati, G., Schutte, A. E., Sharman, J. E., Asmar, R., McManus, R. J., Asayama, K., De La Sierra, A., Head, G., Kario, K., Kollias, A., Myers, M., Niiranen, T., Ohkubo, T., Wang, J., Wuerzner, G., O’Brien, E., Kreutz, R., & Palatini, P.
- Venue
- Journal of Hypertension
- Published
- 2022-06-16
- Volume
- 40
- Issue
- 8
- Pages
- 1449-1460
- ISSN
- 0263-6352 (ISSN-L)
- 1473-5598
- OpenAlex
- W4283030380
- Cited by
- 227
- References
- 16
- FWCI
- 11.81
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Cuffless blood pressure measuring devices: review and statement by the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability. https://app.cochat.ai/papers/10.1097/hjh.0000000000003224
- The paper
- Stergiou, G. S., Mukkamala, R., Avolio, A., Kyriakoulis, K. G., Mieke, S., Murray, A., Parati, G., Schutte, A. E., Sharman, J. E., Asmar, R., McManus, R. J., Asayama, K., De La Sierra, A., Head, G., Kario, K., Kollias, A., Myers, M., Niiranen, T., Ohkubo, T., . . . Palatini, P. (2022). Cuffless blood pressure measuring devices: review and statement by the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability. Journal of Hypertension. https://doi.org/10.1097/hjh.0000000000003224
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.