Blood pressure monitoring: theory and practice. European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability Teaching Course Proceedings
The DOI matches its Crossref record; 57 of the 57 references it cites were found in a registry or an index.
Abstract
The European Society of Hypertension (ESH) Working Group on Blood Pressure (BP) Monitoring and Cardiovascular Variability organized a Teaching Course on 'Blood Pressure Monitoring: Theory and Practice' during the 2017 ESH Meeting in Milan, Italy. This course performed by 11 international BP monitoring experts covered key topics of BP monitoring, including office BP measurement, ambulatory BP monitoring, home BP monitoring, ambulatory versus home BP, white-coat and masked hypertension, cuff use, and BP variability. This article presents a summary of the proceedings of the ESH BP Monitoring Teaching Course, including essential information, practical issues, and recommendations on the clinical application of BP monitoring methods, aiming to the optimal management of patients with suspected or diagnosed hypertension.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and identify the white coat and masked hypertension phenomena (see below).”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [17]
In the citing paper
“more closely associated with hypertension-induced organ damage and the risk of cardiovascular events”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [17]
In the citing paper
“and 10%-15% have masked hypertension.”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [17]
In the citing paper
“Masked hypertension: These patients are at similar risk of cardiovascular events as sustained hypertensives.”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [17]
In the citing paper
“About 10%-30% of subjects attending clinics due to high BP have white coat hypertension”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [17]
4 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Rothwell, P. M., Howard, S. C., Dolan, E., O'Brien, E., Dobson, J. E., Dahlöf, B., Sever, P. S., & Poulter, N. R. (2010) · The Lancet · doi:10.1016/s0140-6736(10)60308-x · Cited by 1,784
Self-citationO'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationEuropean Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
Self-citationKario, K., Pickering, T. G., Umeda, Y., Hoshide, S., Hoshide, Y., Morinari, M., Murata, M., Kuroda, T., Schwartz, J. E., & Shimada, K. (2003) · Circulation · doi:10.1161/01.cir.0000056521.67546.aa · Cited by 1,250
Self-citationEuropean Society of Hypertension practice guidelines for ambulatory blood pressure monitoring
Parati, G., Stergiou, G., O’Brien, E., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2014) · Journal of Hypertension · doi:10.1097/hjh.0000000000000221 · Cited by 1,021
Self-citationAssessment and management of blood-pressure variability
Parati, G., Ochoa, J. E., Lombardi, C., & Bilo, G. (2013) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2013.1 · Cited by 853
Self-citationParati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837
Self-citationde la Sierra, A., Segura, J., Banegas, J. R., Gorostidi, M., de la Cruz, J. J., Armario, P., Oliveras, A., & Ruilope, L. M. (2011) · Hypertension · doi:10.1161/hypertensionaha.110.168948 · Cited by 809
Webb, A. J., Fischer, U., Mehta, Z., & Rothwell, P. M. (2010) · The Lancet · doi:10.1016/s0140-6736(10)60235-8 · Cited by 729
Pickering, T. G., Miller, N. H., Ogedegbe, G., Krakoff, L. R., Artinian, N. T., & Goff, D. (2008) · Hypertension · doi:10.1161/hypertensionaha.107.189010 · Cited by 668
Blood pressure control by home monitoring: meta-analysis of randomised trials
Cappuccio, F. P., Kerry, S. M., Forbes, L., & Donald, A. (2004) · BMJ · doi:10.1136/bmj.38121.684410.ae · Cited by 509
Leung, A. A., Nerenberg, K., Daskalopoulou, S. S., McBrien, K., Zarnke, K. B., Dasgupta, K., Cloutier, L., Gelfer, M., Lamarre-Cliche, M., Milot, A., et al. (2016) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2016.02.066 · Cited by 461
Piper, M. A., Evans, C. V., Burda, B. U., Margolis, K. L., O'Connor, E., & Whitlock, E. P. (2015) · Annals of Internal Medicine · doi:10.7326/m14-1539 · Cited by 380
Mancia, G., Bombelli, M., Facchetti, R., Madotto, F., Corrao, G., Trevano, F. Q., Grassi, G., & Sega, R. (2007) · Hypertension · doi:10.1161/hypertensionaha.107.088708 · Cited by 376
Self-citationPierdomenico, S. D., & Cuccurullo, F. (2011) · American Journal of Hypertension · doi:10.1038/ajh.2010.203 · Cited by 349
Stergiou, G. S., Asayama, K., Thijs, L., Kollias, A., Niiranen, T. J., Hozawa, A., Boggia, J., Johansson, J. K., Ohkubo, T., Tsuji, I., et al. (2014) · Hypertension · doi:10.1161/hypertensionaha.113.02741 · Cited by 304
Self-citationSchillaci, G., Bilo, G., Pucci, G., Laurent, S., Macquin-Mavier, I., Boutouyrie, P., Battista, F., Settimi, L., Desamericq, G., Dolbeau, G., et al. (2012) · Hypertension · doi:10.1161/hypertensionaha.112.197491 · Cited by 285
Self-citationHigh prevalence of masked uncontrolled hypertension in people with treated hypertension
Banegas, J. R., Ruilope, L. M., de la Sierra, A., de la Cruz, J. J., Gorostidi, M., Segura, J., Martell, N., Garcia-Puig, J., Deanfield, J., & Williams, B. (2014) · European Heart Journal · doi:10.1093/eurheartj/ehu016 · Cited by 223
Parati, G., Omboni, S., Rizzoni, D., Agabiti-Rosei, E., & Manciaa, G. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816110-00016 · Cited by 198
Self-citationAmbulatory Blood Pressure Measurement: What Is the International Consensus?
O’Brien, E., Parati, G., & Stergiou, G. (2013) · Hypertension · doi:10.1161/hypertensionaha.113.02148 · Cited by 183
Self-citationHome Blood Pressure Monitoring in the Diagnosis and Treatment of Hypertension: A Systematic Review
Stergiou, G. S., & Bliziotis, I. A. (2011) · American Journal of Hypertension · doi:10.1038/ajh.2010.194 · Cited by 182
Self-citationThe Cardiovascular Risk of White-Coat Hypertension
Franklin, S. S., Thijs, L., Asayama, K., Li, Y., Hansen, T. W., Boggia, J. G., Jacobs, L., Zhang, Z., Kikuya, M., Björklund‐Bodegârd, K., et al. (2016) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.08.035 · Cited by 170
Self-citationParati, G., Ochoa, J. E., Salvi, P., Lombardi, C., & Bilo, G. (2013) · Diabetes Care · doi:10.2337/dcs13-2043 · Cited by 158
Self-citationSTERGIOU, G., SALGAMI, E., TZAMOURANIS, D., & ROUSSIAS, L. (2005) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2005.01.003 · Cited by 157
Self-citationAsayama, K., Thijs, L., Li, Y., Gu, Y.-M., Hara, A., Liu, Y.-P., Zhang, Z., Wei, F.-F., Lujambio, I., Mena, L. J., et al. (2014) · Hypertension · doi:10.1161/hypertensionaha.114.03614 · Cited by 157
Self-citationMancia, G., Bombelli, M., Brambilla, G., Facchetti, R., Sega, R., Toso, E., & Grassi, G. (2013) · Hypertension · doi:10.1161/hypertensionaha.111.00690 · Cited by 151
Self-citationStergiou, G. S., Parati, G., Vlachopoulos, C., Achimastos, A., Andreadis, E., Asmar, R., Avolio, A., Benetos, A., Bilo, G., Boubouchairopoulou, N., et al. (2016) · Journal of Hypertension · doi:10.1097/hjh.0000000000000969 · Cited by 148
Self-citationOffice, Home, and Ambulatory Blood Pressures as Predictors of Cardiovascular Risk
Niiranen, T. J., Mäki, J., Puukka, P., Karanko, H., & Jula, A. M. (2014) · Hypertension · doi:10.1161/hypertensionaha.114.03292 · Cited by 143
Home Blood Pressure Variability as Cardiovascular Risk Factor in the Population of Ohasama
Asayama, K., Kikuya, M., Schutte, R., Thijs, L., Hosaka, M., Satoh, M., Hara, A., Obara, T., Inoue, R., Metoki, H., et al. (2013) · Hypertension · doi:10.1161/hypertensionaha.111.00138 · Cited by 131
Review: a century of confusion; which bladder for accurate blood pressure measurement?
O'Brien, E. T. (1996) · PubMed · Cited by 106
Self-citationA Short History of Automated Office Blood Pressure – 15 Years to SPRINT
Myers, M. G. (2016) · The Journal of Clinical Hypertension · doi:10.1111/jch.12820 · Cited by 95
Shimbo, D., Abdalla, M., Falzon, L., Townsend, R. R., & Muntner, P. (2016) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2015.12.013 · Cited by 94
Home Blood Pressure as a Cardiovascular Outcome Predictor: It’s Time to Take This Method Seriously
Stergiou, G. S., Siontis, K. C. M., & Ioannidis, J. P. A. (2010) · Hypertension · doi:10.1161/hypertensionaha.110.150771 · Cited by 84
Self-citationCuspidi, C., Rescaldani, M., Tadic, M., Sala, C., Grassi, G., & Mancia, G. (2015) · Journal of Hypertension · doi:10.1097/hjh.0000000000000416 · Cited by 84
Self-citationKario, K., Hoshide, S., Haimoto, H., Yamagiwa, K., Uchiba, K., Nagasaka, S., Yano, Y., Eguchi, K., Matsui, Y., Shimizu, M., et al. (2015) · The Journal of Clinical Hypertension · doi:10.1111/jch.12500 · Cited by 76
Self-citationBlood pressure measurement in very obese patients: a challenging problem
Palatini, P., & Parati, G. (2011) · Journal of Hypertension · doi:10.1097/hjh.0b013e3283435b65 · Cited by 72
Self-citationO’Brien, E. (2011) · Journal of Internal Medicine · doi:10.1111/j.1365-2796.2011.02356.x · Cited by 71
Self-citationRequirements for professional office blood pressure monitors
Stergiou, G. S., Parati, G., Asmar, R., & O’Brien, E. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834fcfa5 · Cited by 71
Self-citationNasothimiou, E. G., Tzamouranis, D., Rarra, V., Roussias, L. G., & Stergiou, G. S. (2012) · Hypertension Research · doi:10.1038/hr.2012.19 · Cited by 69
Self-citationPatients’ preference for ambulatory versus home blood pressure monitoring
Nasothimiou, E. G., Karpettas, N., Dafni, M. G., & Stergiou, G. S. (2014) · Journal of Human Hypertension · doi:10.1038/jhh.2013.104 · Cited by 66
Self-citationBonso, E., Saladini, F., Zanier, A., Benetti, E., Dorigatti, F., & Palatini, P. (2010) · Hypertension Research · doi:10.1038/hr.2010.146 · Cited by 63
Self-citationHome Blood Pressure Monitoring: Primary Role in Hypertension Management
Stergiou, G. S., Κόλλιας, Α., Zeniodi, M., Karpettas, N., & Ntineri, A. (2014) · Current Hypertension Reports · doi:10.1007/s11906-014-0462-8 · Cited by 61
Self-citationRectangular cuffs may overestimate blood pressure in individuals with large conical arms
Palatini, P., Benetti, E., Fania, C., Malipiero, G., & Saladini, F. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834f98a0 · Cited by 60
Self-citationStergiou, G. S., Nasothimiou, E. G., Destounis, A., Poulidakis, E., Evagelou, I., & Tzamouranis, D. (2012) · American Journal of Hypertension · doi:10.1038/ajh.2012.82 · Cited by 54
Self-citationClinical characteristics of isolated clinic hypertension
Vinyoles, E., Felip, À., Pujol, E., de la Sierra, A., Durà, R., del Rey, R. H., Sobrino, J., Gorostidi, M., de la Figuera, M., Segura, J., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e3282f3150b · Cited by 53
Parati, G., & Schumacher, H. (2013) · Hypertension Research · doi:10.1038/hr.2013.145 · Cited by 48
Self-citationTatasciore, A., Zimarino, M., Tommasi, R., Renda, G., Schillaci, G., Parati, G., & Caterina, R. D. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328361e4a6 · Cited by 47
Self-citationReporting bias: Achilles' heel of home blood pressure monitoring
Myers, M. G., & Stergiou, G. S. (2014) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2014.02.001 · Cited by 44
Self-citationParati, G., Dolan, E., Ley, L., & Schumacher, H. (2014) · Journal of Hypertension · doi:10.1097/hjh.0000000000000169 · Cited by 38
Self-citationStergiou, G. S., Tzamouranis, D., Nasothimiou, E. G., & Protogerou, A. D. (2008) · Journal of Human Hypertension · doi:10.1038/jhh.2008.70 · Cited by 33
Self-citationTechniques for Self‐Measurement of Blood Pressure: Limitations and Needs for Future Research
Palatini, P., & Frick, G. N. (2012) · Journal of Clinical Hypertension · doi:10.1111/j.1751-7176.2011.00586.x · Cited by 30
Self-citationComparison of the Conical Cuff and the Standard Rectangular Cuffs
MAXWELL, G. F., PRUIJT, J. F. M., & ARNTZENIUS, A. C. (1985) · International Journal of Epidemiology · doi:10.1093/ije/14.3.468 · Cited by 30
Impact of Miscuffing During Home Blood Pressure Measurement on the Prevalence of Masked Hypertension
Mourad, J., Lopez‐Sublet, M., Bahous, S. A., Villeneuve, F., Jaboureck, O., Dourmap, C., Denolle, T., Fourcade, J., & Baguet, J.-P. (2013) · American Journal of Hypertension · doi:10.1093/ajh/hpt084 · Cited by 24
Viera, A. J., Tuttle, L. A., Voora, R. A., & Olsson, E. (2015) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000147 · Cited by 16
Azaki, A., Diab, R., Harb, A., Asmar, R., & Chahine, M. N. (2017) · Vascular Health and Risk Management · doi:10.2147/vhrm.s126285 · Cited by 14
Self-citationStergiou, G. S., Ntineri, A., & Κόλλιας, Α. (2014) · Hypertension · doi:10.1161/hypertensionaha.114.04076 · Cited by 13
Self-citationWhy Is It That We Continue to Deny Patients Ambulatory Blood Pressure Monitoring?
O'Brien, E. T. (2016) · Hypertension · doi:10.1161/hypertensionaha.115.06777 · Cited by 10
Self-citation
Cited by
2020 International Society of Hypertension Global Hypertension Practice Guidelines
Unger, T., Borghi, C., Charchar, F., Khan, N. A., Poulter, N. R., Prabhakaran, D., Ramirez, A., Schlaich, M., Stergiou, G. S., Tomaszewski, M., et al. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.15026 · Cited by 4,132
Self-citation2020 International Society of Hypertension global hypertension practice guidelines
Unger, T., Borghi, C., Charchar, F., Khan, N. A., Poulter, N. R., Prabhakaran, D., Ramirez, A., Schlaich, M., Stergiou, G. S., Tomaszewski, M., et al. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002453 · Cited by 1,283
“The diagnosis might be made on a single visit, if BP is ≥180/110 mm Hg and there is evidence of cardiovascular disease (CVD).(1,2,17,18) • The recommended patient management according to office BP levels is presented in Table 4.”
InfluentialSelf-citation2024 Clinical practice guidelines for Hypertension in adults
Kobalava, Z. D., Konradi, A. O., Nedogoda, S. V., Shlyakhto, E. V., Arutyunov, G. P., Baranova, E. I., Barbarash, O. L., Bobkova, N. V., Boytsov, S. A., Bubnova, M. G., et al. (2024) · Russian Journal of Cardiology · doi:10.15829/1560-4071-2024-6117 · Cited by 211
Tully, P. J., Yano, Y., Launer, L. J., Kario, K., Nagai, M., Mooijaart, S. P., Claassen, J. A. H. R., Lattanzi, S., Vincent, A. D., Tzourio, C., et al. (2020) · Journal of the American Heart Association · doi:10.1161/jaha.119.013841 · Cited by 119
Self-citationSystemic Hypertension at High Altitude
Narvaez-Guerra, O. D., Herrera-Enriquez, K. B., Medina‐Lezama, J., & Chirinos, J. A. (2018) · Hypertension · doi:10.1161/hypertensionaha.118.11140 · Cited by 91
Metadata
- Authors
- Stergiou, G. S., Palatini, P., Asmar, R., Bilo, G., de la Sierra, A., Head, G., Kario, K., Mihailidou, A., Wang, J., Mancia, G., O’Brien, E., & Parati, G.
- Affiliations
- National and Kapodistrian University of Athens (GR)
- University of Padua (IT)
- IRCCS Istituto Auxologico Italiano (IT)
- Jichi Medical University (JP)
- Royal North Shore Hospital (AU)
- University of Milano-Bicocca (IT)
- Corresponding author
- Stergiou, George S.
- ORCID
- Venue
- Blood Pressure Monitoring
- Published
- 2018-02
- Volume
- 23
- Issue
- 1
- Pages
- 1-8
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2768135819
- Semantic Scholar
- 044a65ac1218ac9c5cf2c6fe3834d9af3a2506eb
- Cited by
- 57
- References
- 57
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 6 · 2023: 6 · 2022: 11 · 2021: 5 · 2020: 10 · 2019: 8 · 2018: 8
- FWCI
- 5.71
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Blood Pressure Monitoring (publishedVersion)
- PubMed (publishedVersion)
- BOA (University of Milano-Bicocca) (submittedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- hypertension, ambulatory blood pressure monitoring, home blood pressure monitoring, office blood pressure measurement, white coat hypertension, masked hypertension, blood pressure variability, antihypertensive therapy
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Europe, Humans, Hypertension, Italy, Prognosis, Societies, Medical, Blood Pressure Monitoring, Ambulatory, White Coat Hypertension, Masked Hypertension
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.