Repeatability of In Vivo Parafoveal Cone Density and Spacing Measurements
Open accessThe DOI matches its Crossref record; 32 of the 32 references it cites were found in a registry or an index.
Abstract
PURPOSE: To assess the repeatability and measurement error associated with cone density and nearest neighbor distance (NND) estimates in images of the parafoveal cone mosaic obtained with an adaptive optics scanning light ophthalmoscope (AOSLO). METHODS: Twenty-one participants with no known ocular pathology were recruited. Four retinal locations, approximately 0.65° eccentricity from the center of fixation, were imaged 10 times in randomized order with an AOSLO. Cone coordinates in each image were identified using an automated algorithm (with or without manual correction) from which cone density and NND were calculated. Owing to naturally occurring fixational instability, the 10 images recorded from a given location did not overlap entirely. We thus analyzed each image set both before and after alignment. RESULTS: Automated estimates of cone density on the unaligned image sets showed a coefficient of repeatability of 11,769 cones/mm(2) (17.1%). The primary reason for this variability appears to be fixational instability, as aligning the 10 images to include the exact same retinal area results in an improved repeatability of 4358 cones/mm(2) (6.4%) using completely automated cone identification software. Repeatability improved further by manually identifying cones missed by the automated algorithm, with a coefficient of repeatability of 1967 cones/mm(2) (2.7%). NND showed improved repeatability and was generally insensitive to the undersampling by the automated algorithm. CONCLUSIONS: As our data were collected in a young, healthy population, this likely represents a best-case estimate for corresponding measurements in patients with retinal disease. Similar studies need to be carried out on other imaging systems (including those using different imaging modalities, wavefront correction technology, and/or image analysis software), as repeatability would be expected to be highly sensitive to initial image quality and the performance of cone identification algorithms. Separate studies addressing intersession repeatability and interobserver reliability are also needed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Human photoreceptor topography
Curcio, C. A., Sloan, K. R., Kalina, R. E., & Hendrickson, A. E. (1990) · Journal of Comparative Neurology · doi:10.1002/cne.902920402 · Cited by 3,167 · cited 2×
“It is known from a number of studies that this is where cone density is changing most rapidly.(17,18,32) One would expect that in the periphery, where cone density is more uniform, that the repeatability would be affected less by fixational instability and that there may be less of a difference between the automated a…”
BackgroundSupernormal vision and high-resolution retinal imaging through adaptive optics
Liang, J., Williams, D. R., & Miller, D. T. (1997) · Journal of the Optical Society of America A · doi:10.1364/josaa.14.002884 · Cited by 1,454 · cited 10×
“T op r ig ht 69 ,9 87 8, 32 9 11 ,7 71 [1 1, 34 3– 12 ,1 99 ] 16 .”
InfluentialThe arrangement of the three cone classes in the living human eye
Roorda, A., & Williams, D. R. (1999) · Nature · doi:10.1038/17383 · Cited by 982
Statistics Notes: Measurement error proportional to the mean
Bland, J. M., & Altman, D. G. (1996) · BMJ · doi:10.1136/bmj.313.7049.106 · Cited by 543
Organization of the human trichromatic cone mosaic
Hofer, H., Carroll, J., Neitz, J., Neitz, M., & Williams, D. R. (2005) · The Journal of Neuroscience · doi:10.1523/jneurosci.2414-05.2005 · Cited by 364 · cited 5×
“T op le ft 70 ,9 28 3, 27 7 4, 63 1 [4 ,4 62 –4 ,8 00 ] 6.”
InfluentialSelf-citationDubra, A., Sulai, Y. N., Norris, J. L., Cooper, R. F., Dubis, A. M., Williams, D. R., & Carroll, J. (2011) · Biomedical Optics Express · doi:10.1364/boe.2.001864 · Cited by 352 · cited 2×
“A previously described adaptive optics scanning light ophthalmoscope (AOSLO) was used to image the parafoveal cone mosaic of the right eye.(28,29) The wavelength of the super luminescent diode used for retinal imaging was 775 nm, subtending a field of view of 0.”
BackgroundMethodSelf-citationTalcott, K. E., Ratnam, K., Sundquist, S. M., Lucero, A. S., Lujan, B. J., Tao, W., Porco, T. C., Roorda, A., & Duncan, J. L. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-6479 · Cited by 276 · cited 2×
“A recent study by Talcott et al.(24) performed a repeated-measures analysis of cone spacing in three normal eyes and found no significant change in cone spacing over time periods ranging from 16 to 53 months.”
MethodVariation of Cone Photoreceptor Packing Density with Retinal Eccentricity and Age
Song, H., Chui, T. Y. P., Zhong, Z., Elsner, A. E., & Burns, S. A. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.11-7199 · Cited by 267 · cited 4×
“These were chosen based on those reported previously by other groups.(16,18) Interestingly, as the sampling window size decreased, we observed a decrease in the repeatability and an increase in the measurement error for both cone density and NND, although there was some variability in the effect.”
BackgroundMethodInfluentialRodieck, R. W. (1991) · Visual Neuroscience · doi:10.1017/s095252380001049x · Cited by 262 · cited 1×
“, missed cones) than our estimates of cone density (Table 3), it has been shown previously that measures of cone spacing based on an exclusion radius are even less sensitive to undersampling.(34,35) Such insensitivity could be viewed as either an advantage or disadvantage.”
BackgroundHigh-Resolution Imaging with Adaptive Optics in Patients with Inherited Retinal Degeneration
Duncan, J. L., Zhang, Y., Gandhi, J., Nakanishi, C., Othman, M., Branham, K. E. H., Swaroop, A., & Roorda, A. (2007) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.06-1422 · Cited by 262 · cited 2×
“B ot to m le ft 67 ,4 51 10 ,3 87 14 ,6 79 [1 4, 14 5– 15 ,2 13 ] 21 .”
The relationship between visual resolution and cone spacing in the human fovea
Rossi, E. A., & Roorda, A. (2010) · Nature Neuroscience · doi:10.1038/nn.2465 · Cited by 255 · cited 3×
“B ot to m le ft 67 ,4 51 10 ,3 87 14 ,6 79 [1 4, 14 5– 15 ,2 13 ] 21 .”
Automated identification of cone photoreceptors in adaptive optics retinal images
Li, K. Y., & Roorda, A. (2007) · Journal of the Optical Society of America A · doi:10.1364/josaa.24.001358 · Cited by 222 · cited 2×
“This is a modified version of the previously described algorithm of Li and Roorda.(12) This algorithm first ap-”
MethodIntersubject Variability of Foveal Cone Photoreceptor Density in Relation to Eye Length
Li, K. Y., Tiruveedhula, P. K., & Roorda, A. (2010) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.10-5499 · Cited by 204 · cited 1×
“reported a range from about 64,000 cones/mm(2) to 98,000 cones/mm(2) at a comparable eccentricity across 18 subjects.(17)”
Optical fiber properties of individual human cones
Roorda, A., & Williams, D. R. (2002) · Journal of Vision · doi:10.1167/2.5.4 · Cited by 203 · cited 1×
“The higher transverse resolution provided by AO makes it possible to examine features of the photoreceptor mosaic such as the spatial arrangement of the different spectral types of cone within the mosaic,(3,4) temporal reflectance changes of individual cones and rods, and even the orientation tuning of individual cone…”
BackgroundFunctional photoreceptor loss revealed with adaptive optics: An alternate cause of color blindness
Carroll, J., Neitz, M., Hofer, H., Neitz, J., & Williams, D. R. (2004) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0401440101 · Cited by 191
Self-citationIndividual Variations in Human Cone Photoreceptor Packing Density: Variations with Refractive Error
Chui, T. Y. P., Song, H., & Burns, S. A. (2008) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.08-2135 · Cited by 191 · cited 1×
“These were chosen based on those reported previously by other groups.(16,18) Interestingly, as the sampling window size decreased, we observed a decrease in the repeatability and an increase in the measurement error for both cone density and NND, although there was some variability in the effect.”
MethodChoi, S. S., Doble, N. P., Hardy, J. L., Jones, S. M., Keltner, J. L., Olivier, S. S., & Werner, J. S. (2006) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.05-0997 · Cited by 177
The Reflectance of Single Cones in the Living Human Eye
Pallikaris, A., Williams, D. R., & Hofer, H. (2003) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.03-0094 · Cited by 176 · cited 6×
“B ot to m r ig ht 70 ,0 19 3, 34 8 4, 73 2 [4 ,5 60 –4 ,9 04 ] 6.”
InfluentialIn vivo functional imaging of human cone photoreceptors
Jonnal, R. S., Rha, J., Zhang, Y., Cense, B., Gao, W., & Miller, D. T. (2007) · Optics Express · doi:10.1364/oe.15.016141 · Cited by 169 · cited 4×
“B ot to m le ft 70 ,7 41 1, 82 6 2, 58 1 [2 ,4 87 –2 ,6 75 ] 3.”
InfluentialPhotoreceptor Structure and Function in Patients with Congenital Achromatopsia
Genead, M. A., Fishman, G. A., Rha, J., Dubis, A. M., Bonci, D. M. O., Dubra, A., Stone, E. M., Neitz, M., & Carroll, J. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.11-7762 · Cited by 166
Self-citationSpatial properties of retinal mosaics: An empirical evaluation of some existing measures
Cook, J. E. (1996) · Visual Neuroscience · doi:10.1017/s0952523800007094 · Cited by 166 · cited 1×
“, missed cones) than our estimates of cone density (Table 3), it has been shown previously that measures of cone spacing based on an exclusion radius are even less sensitive to undersampling.(34,35) Such insensitivity could be viewed as either an advantage or disadvantage.”
BackgroundMerino, D., Duncan, J. L., Tiruveedhula, P. K., & Roorda, A. (2011) · Biomedical Optics Express · doi:10.1364/boe.2.002189 · Cited by 162 · cited 1×
“As has been shown recently, it is now possible to image the rod mosaic.(2,29,33) Unlike the cone mosaic, which appears to reach an asymptotic density beyond about 5 mm, rod density changes throughout the retina, first increasing sharply moving away from the fovea and then decreasing beyond the rod-rim.”
BackgroundAdaptive-optics imaging of human cone photoreceptor distribution
Chui, T. Y., Song, H., & Burns, S. A. (2008) · Journal of the Optical Society of America A · doi:10.1364/josaa.25.003021 · Cited by 159 · cited 3×
“B ot to m le ft 70 ,5 75 1, 30 5 1, 84 4 [1 ,7 77 –1 ,9 11 ] 2.”
Imaging outer segment renewal in living human cone photoreceptors
Jonnal, R. S., Besecker, J., Derby, J. C., Kocaoglu, O. P., Cense, B., Gao, W., Wang, Q., & Miller, D. T. (2010) · Optics Express · doi:10.1364/oe.18.005257 · Cited by 147 · cited 3×
“B ot to m r ig ht 70 ,2 04 1, 38 7 1, 96 0 [1 ,8 89 –2 ,0 31 ] 2.”
Cone Photoreceptor Abnormalities Correlate with Vision Loss in Patients with Stargardt Disease
Chen, Y., Ratnam, K., Sundquist, S. M., Lujan, B. J., Ayyagari, R., Gudiseva, V. H., Roorda, A., & Duncan, J. L. (2011) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.10-6538 · Cited by 130
Imaging retinal mosaics in the living eye
Rossi, E. A., Chung, M. E., Dubra, A., Hunter, J. J., Merigan, W. H., & Williams, D. R. (2011) · Eye · doi:10.1038/eye.2010.221 · Cited by 113
Self-citation(2018) · Encyclopedia of Image Processing · doi:10.1201/9781351032742-11 · Cited by 101
Adaptive optics retinal imaging reveals S-cone dystrophy in tritan color vision deficiency
Baraas, R. C., Carroll, J., Gunther, K. L., Chung, M., Williams, D. R., Foster, D. H., & Neitz, M. (2007) · Journal of the Optical Society of America A · doi:10.1364/josaa.24.001438 · Cited by 100 · cited 6×
“B ot to m r ig ht 65 ,8 04 9, 12 5 12 ,8 96 [1 2, 42 7– 13 ,3 65 ] 19 .”
InfluentialSelf-citationCone photoreceptor mosaic disruption associated with Cys203Arg mutation in the M-cone opsin
Carroll, J., Baraas, R. C., Wagner-Schuman, M., Rha, J., Siebe, C. A., Sloan, C., Tait, D. M., Thompson, S., Morgan, J. I. W., Neitz, J., et al. (2009) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0910128106 · Cited by 59 · cited 1×
“T op le ft 73 ,6 15 11 ,1 64 15 ,7 77 [1 5, 20 3– 16 ,3 51 ] 21 .”
Self-citationCarroll, J., Rossi, E. A., Porter, J., Neitz, J., Roorda, A., Williams, D. R., & Neitz, M. (2010) · Vision Research · doi:10.1016/j.visres.2010.07.009 · Cited by 49 · cited 1×
“B ot to m le ft 67 ,3 68 4, 34 8 6, 14 5 [5 ,9 21 –6 ,3 69 ] 9.”
Self-citationRha, J., Dubis, A. M., Wagner-Schuman, M. L., Tait, D. M., Godara, P., Schroeder, B., Stepien, K. E., & Carroll, J. (2009) · Advances in experimental medicine and biology · doi:10.1007/978-1-4419-1399-9_35 · Cited by 32 · cited 2×
“In a single patient with a red-green color vision defect, Rha et al.(25) observed a 3.”
BackgroundInfluentialSelf-citationBoretsky, A. R., Motamedi, M., Bell, B. A., & van Kuijk, F. (2011) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.10-7033 · Cited by 10
Cited by
Pharmaceutical and biomedical analysis of cannabinoids: A critical review
Citti, C., Braghiroli, D., Vandelli, M. A., & Cannazza, G. (2018) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2017.06.003 · Cited by 249
Metadata
- Authors
- Garrioch, R., Langlo, C., Dubis, A. M., Cooper, R. F., Dubra, A., & Carroll, J.
- Venue
- Optometry and Vision Science
- Publisher
- Wiley
- Published
- 2012-05
- Volume
- 89
- Issue
- 5
- Pages
- 632-643
- ISSN
- 1040-5488 (ISSN-L)
- OpenAlex
- W2012737567
- Cited by
- 165
- References
- 32
- FWCI
- 9.12
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.