The optimal controller delay for myoelectric prostheses
Open accessAbstract
A tradeoff exists when considering the delay created by multifunctional prosthesis controllers. Large controller delays maximize the amount of time available for EMG signal collection and analysis (and thus maximize classification accuracy); however, large delays also degrade prosthesis performance by decreasing the responsiveness of the prosthesis. To elucidate an "optimal controller delay" twenty able-bodied subjects performed the Box and Block Test using a device called PHABS (prosthetic hand for able bodied subjects). Tests were conducted with seven different levels of controller delay ranging from nearly 0-300 ms and with two different artificial hand speeds. Based on repeted measures ANOVA analysis and a linear mixed effects model, the optimal controller delay was found to range between approximately 100 ms for fast prehensors and 125 ms for slower prehensors. Furthermore, the linear mixed effects model shows that there is a linear degradation in performance with increasing delay.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Longer latencies occur with more nodes, which can be detrimental to real-time control applications such as myoelectric prosthesis control or brain-computer interfaces.
“Longer latencies occur, which can be detrimental to real-time control applications such as myoelectric prosthesis control [43] or brain–computer interfaces.”
From Comparison between Two Time Synchronization and Data Alignment Methods for Multi-Channel Wearable Biosensor Systems Using BLE Protocol (2023) · cites this paper as [43]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A robust, real-time control scheme for multifunction myoelectric control
Englehart, K., & Hudgins, B. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.813539 · Cited by 1,743 · see citing works
Voluntary finger movement in man: Cerebral potentials and theory
Deecke, L., Gr�zinger, B., & Kornhuber, H. H. (1976) · Biological Cybernetics · doi:10.1007/bf00336013 · Cited by 757 · see citing works
EMG feature evaluation for movement control of upper extremity prostheses
Zardoshti-Kermani, M., Wheeler, B. C., Badie, K., & Hashemi, R. M. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481972 · Cited by 477 · see citing works
Graupe, D., Salahi, J., & Kohn, K. H. (1982) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(82)90021-8 · Cited by 205 · cited 1× · see citing works
“In an early work, they stated that delays of greater than 200 ms are unacceptable [19] and in a later paper they stated that delays of 100–300 ms are satisfactory [20].”
Background
Flagged first, then by citations
Research works printed but not linked: 33
- [1][1].Englehart K, Hudgins B, and Parker PA, “A wavelet-based continuous classification scheme for multifunction myoelectric control,” IEEE Trans. Biomed. Eng, vol. 48, no. 3, pp. 302–311, March. 2001. [DOI] [PubMed] [Google Scholar]
- [2][2].Du S and Vuskovic MI, “Temporal vs. spectral approach to feature extraction from prehensile EMG signals,” in Proc. IEEE Int. Conf. Inf. Reuse Integration, November. 8-10, 2004, pp. 344–350. [Google Scholar]
- [3][3].Yamada M, Niwa N, and Uchiyama A, “Evaluation of a multifunctional hand prosthesis system using EMG controlled animation,” IEEE Trans. Biomed. Eng, vol. 30, pp. 759–763, 1983. [DOI] [PubMed] [Google Scholar]
- [4][4].Hogan N, “A review of the methods of processing EMG for use as a proportional control signal,” Biomed. Eng, vol. 11, pp. 81–86, 1976. [PubMed] [Google Scholar]
- [5][5].Kyberd PJ, Holland OE, Chappell PH, Smith S, Tregidgo R, Bagwell PJ, and Snaith M, “MARCUS: A two degree of freedom hand prosthesis with hierarchical grip control,” IEEE Trans. Rehabil. Eng, vol. 3, no. 1, pp. 70–76, March. 1995. [Google Scholar]
- [6][6].Conklin JE, “Linearity of the tracking performance function,” Perceptual Motor Skills, vol. 9, pp. 387–391, 1959. [Google Scholar]
- [7][7].Ferrell WR, “Remote manipulation with transmission delay,” IEEE Trans. Human Factors Electronics, vol. HFE6, pp. 24–32, 1965. [Google Scholar]
- [8][8].Ferrell WR and Sheridan TB, “Supervisory Control of Remote Manipulation,” IEEE Spectrum, vol. 4, pp. 81–88, 1967. [Google Scholar]
- [9][9].MacKenzie I and Ware C, “Lag as a determinant of human performance in interactive systems,” in Proc. ACM Conf. Human Factors Comput. Syst.—INTERCHI ’ 93, 1993, pp. 488–493. [Google Scholar]
- [10][10].Sheridan TB and Ferrell WR, “Remote manipulative control with transmission delay,” IEEE Trans. Human Factors Eng, vol. HFE4, pp. 25–29, September. 1963. [Google Scholar]
- [11][11].Smith KU, Smith WM, and Smith MF, Delayed Sensory Feedback and Behavior. Philadelphia, PA: W.B. Saunders and Company, 1962. [Google Scholar]
- [12][12].Smith WM, “Feedback: Real-time delayed vision of one’s own tracking behavior,” Science, vol. 176, pp. 939–940, 1972. [DOI] [PubMed] [Google Scholar]
- [13][13].Smith WM and Bowen KF, “The effects of delayed and displaced visual feedback on motor control,” J. Mot. Behav, vol. 12, pp. 91–101, 1980. [DOI] [PubMed] [Google Scholar]
- [14][14].Held R, Efstathiou A, and Greene M, “Adaptation to displaced and delayed visual feedback from the hand,” J. Exp. Psychol, vol. 72, pp. 887–891, 1966. [Google Scholar]
- [15][15].Paciga JE, Richard PD, and Scott RN, “Error rate in five-state myoelectric control systems,” Med. Biol. Eng. Comput, vol. 18, pp. 287–290, 1980. [DOI] [PubMed] [Google Scholar]
- [16][16].Childress DS and Weir RF, “Control of limb prostheses,” in Atlas of Amputations and Limb Deficiencies: Surgical, Prosthetic, and Rehabilitation Principles, Smith DG, Michael JW, and B. J. H., Eds., 3rd ed. Rosemont, IL: J. H. Bowker, 2004, pp. 173–195. [Google Scholar]
- [18][18].Englehart K, Hudgins B, and Chan A, “Continuous multifunction myoelectric control using pattern recognition,” Technol. Disability, vol. 15, pp. 95–103, 2003. [Google Scholar]
- [20][20].Graupe D, Salahi J, and Zhang DS, “Stochastic analysis of myoelectric temporal signatures for multifunctional single-site activation of prostheses and orthoses,” J. Biomed. Eng, vol. 7, pp. 18–29, 1985. [DOI] [PubMed] [Google Scholar]
- [21][21].Hefftner G, Zucchini W, and Jaros GG, “The electromyogram (EMG) as a control signal for functional neuromuscular stimulation—Part I: Autoregressive modeling as a means of EMG signature discrimination,” IEEE Trans. Biomed. Eng, vol. 35, no. 4, pp. 230–237, April. 1988. [DOI] [PubMed] [Google Scholar]
- [22][22].Chu JK, Moon I, and Mun M, “A real-time EMG pattern recognition based on linear-nonlinear feature projection for multifunction myoelectric hand,” presented at the IEEE 9th Int. Conf. Rehabilitation Robotics (ICORR), Chicago, IL, 2005. [Google Scholar]
- [23][23].Myobock Arm Components Catalog. Duderstadt, Germany: Otto Bock Healthcare, 2004/2005. [Google Scholar]
- [24][24].Heckathorne CW, “Components for electric-powered systems,” in Atlas of Amputations and Limb Deficiencies, Smith DG, Michael JW, and Bowker JH, Eds., 3rd ed. Rosemont, IL: American Academy of Orthopaedic Surgeons, 2004, pp. 145–171. [Google Scholar]
- [25][25].Childress DS, “An approach to powered grasp,” in Advances in External Control of Human ExtremitiesFourth International Symposium on External Control of Human Extremities, Gavrilovic MM and Wilson AB Jr., Eds., Belgrade, Yugoslavia, 1973, pp. 159–167. [Google Scholar]
- [26][26].Weir RF, “Design of artificial arms and hands for prosthetic applications,” in Standard Handbook of Biomedical Engineering and Design, Kutz M, Ed. New York: McGraw–Hill, 2003, ch. 32, pp. 32.1–32.61. [Google Scholar]
- [28][28].Goodkin DE, Hertsgaard D, and Seminary J, “Upper extremity function in multiple sclerosis: Improving assessment sensitivity with box- and-block and nine-hole peg tests,” Arch. Phys. Med. Rehabil, vol. 69, pp. 850–854, 1988. [PubMed] [Google Scholar]
- [29][29].Desrosiers J, Bravo G, Hebert R, Dutil E, and Mercier L, “Validation of the Box and Block Test as a measure of dexterity of elderly people: Reliability, validity, and norms studies,” Arch. Phys. Med. Rehabil, vol. 75, pp. 751–755, 1994. [PubMed] [Google Scholar]
- [30][30].Mathiowetz V, Volland G, Kashman N, and Weber K, “Adult norms for the Box and Block Test of manual dexterity,” Am. J. Occup. Ther, vol. 39, pp. 386–391, 1985. [DOI] [PubMed] [Google Scholar]
- [31][31].Goodman G and Bazyk S, “The effects of a short thumb opponens splint on hand function in cerebral palsy: A single-subject study,” Am. J. Occup. Ther, vol. 45, pp. 726–731, 1991. [DOI] [PubMed] [Google Scholar]
- [32][32].Goodkin DE, Rudick RA, VanderBrug Medendorp S, Daughtry MM, Schwetz KM, Fischer J, and Van Dyke C, “Low-dose (7.5 mg) oral methotrexate reduces the rate of progression in chronic progressive multiple sclerosis,” Ann. Neurol, vol. 37, pp. 30–40, 1995. [DOI] [PubMed] [Google Scholar]
- [33][33].Cauraugh JH and Kim SB, “Chronic stroke motor recovery: Duration of active neuromuscular stimulation,” J. Neurol. Sci, vol. 215, pp. 13–19, 2003. [DOI] [PubMed] [Google Scholar]
- [34][34].Carey JR, Kimberley TJ, Lewis SM, Auerbach EJ, Dorsey L, Rundquist P, and Ugurbil K, “Analysis of fMRI and finger tracking training in subjects with chronic stroke,” Brain, vol. 125, pp. 773–788, 2002. [DOI] [PubMed] [Google Scholar]
- [35][35].Cauraugh J, Light K, Kim S, Thigpen M, and Behrman A, “Chronic motor dysfunction after stroke: Recovering wrist and finger extension by electromyography-triggered neuromuscular stimulation,” Stroke, vol. 31, pp. 1360–1364, 2000. [DOI] [PubMed] [Google Scholar]
- [36][36].Park E and Meek SG, “Adaptive filtering of the electromyographic signal for prosthetic control and force estimation,” IEEE Trans. Biomed. Eng, vol. 42, no. 10, pp. 1048–1052, October. 1995. [DOI] [PubMed] [Google Scholar]
Cited by
Sorted by citations
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001 · see citing works
Quantifying Pattern Recognition—Based Myoelectric Control of Multifunctional Transradial Prostheses
Li, G., Schultz, A. E., & Kuiken, T. A. (2010) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2039619 · Cited by 443 · see citing works
A survey of sensor fusion methods in wearable robotics
Novak, D., & Riener, R. (2015) · Robotics and Autonomous Systems · doi:10.1016/j.robot.2014.08.012 · Cited by 252 · see citing works
Control of prosthetic hands via the peripheral nervous system
Ciancio, A. L., Cordella, F., Barone, R., Romeo, R. A., Bellingegni, A. D., Sacchetti, R., Davalli, A., Di Pino, G., Ranieri, F., Di Lazzaro, V., et al. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00116 · Cited by 131 · see citing works
Raspopovic, S., Carpaneto, J., Udina, E., Navarro, X., & Micera, S. (2010) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-7-17 · Cited by 81 · see citing works
Betthauser, J. L., Krall, J. T., Bannowsky, S. G., Levay, G., Kaliki, R. R., Fifer, M. S., & Thakor, N. V. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2019.2943309 · Cited by 73 · see citing works
Fully Implantable Multi-Channel Measurement System for Acquisition of Muscle Activity
Lewis, S., Russold, M., Dietl, H., Ruff, R., Audi, J. M. C., Hoffmann, K.-P., Abu-Saleh, L., Schroeder, D., Krautschneider, W. H., Westendorff, S., et al. (2013) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2013.2253992 · Cited by 52 · see citing works
Material considerations for peripheral nerve interfacing
Kim, Y.-T., & Romero-Ortega, M. I. (2012) · MRS Bulletin · doi:10.1557/mrs.2012.99 · Cited by 47 · see citing works
Lukyanenko, P., Dewald, H. A., Lambrecht, J., Kirsch, R. F., Tyler, D. J., & Williams, M. R. (2021) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-021-00833-3 · Cited by 27 · see citing works
Ultrasound Features of Skeletal Muscle Can Predict Kinematics of Upcoming Lower-Limb Motion
Jahanandish, M. H., Rabe, K. G., Fey, N. P., & Hoyt, K. (2021) · Annals of Biomedical Engineering · doi:10.1007/s10439-020-02617-7 · Cited by 23 · see citing works
Wang, H., Li, J., McDonald, B. E., Farrell, T. R., Huang, X., & Clancy, E. A. (2023) · Sensors · doi:10.3390/s23052465 · Cited by 18 · see citing works
Self-citation
- Authors
- Farrell, T. R., & Weir, R. F.
- Affiliations
- Northwestern University (US)
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2007-03
- Volume
- 15
- Issue
- 1
- Pages
- 111-118
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- PMC
- PMC8173529
- OpenAlex
- W2125585124
- Cited by
- 438
- References
- 4
- Citations by year
- 2026: 27 · 2025: 26 · 2024: 30 · 2023: 33 · 2022: 30 · 2021: 40 · 2020: 32 · 2019: 35 · 2018: 32 · 2017: 31 · 2016: 14 · 2015: 19 · 2014: 29 · 2013: 10 · 2012: 11
- FWCI
- 5.74
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Biomedical Imaging and Bioengineering (R01EB001672, 1 R01EB01672-01)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Physical Sciences › Engineering › Biomedical Engineering › Prosthetics and Rehabilitation Robotics
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- myoelectric prosthesis, EMG signal processing, Box and Block Test, phaB, EMG signal classification
- MeSH terms
- Artificial Limbs, Electric Stimulation Therapy, Feedback, Hand, Humans, Prosthesis Design, Therapy, Computer-Assisted, Muscle, Skeletal, Hand Strength, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review