The first study comparing a wearable watch‐type blood pressure monitor with a conventional ambulatory blood pressure monitor on in‐office and out‐of‐office settings

Open access
Kario, Kazuomi, Shimbo, Daichi, Tomitani, Naoko, Kanegae, Hiroshi, Schwartz, Joseph E., Williams, Bryan·2020·The Journal of Clinical Hypertension

Abstract

Wearable blood pressure (BP) monitoring devices which measure BP levels accurately both in and out of the office are valuable for hypertension management using digital technology. The authors have conducted the first comparison study of BPs measured by a recently developed wrist‐worn watch‐type oscillometric BP monitoring (WBPM) device, the “HeartGuide,” versus BPs measured by an ambulatory BP monitoring (ABPM) device, A&D TM‐2441, in the office (total of 4 readings alternately measured in the sitting position) and outside the office (30‐minutes interval measurements during daytime) in 50 consecutive patients (mean age 66.1 ± 10.8 years). The 2 BP monitoring devices were simultaneously worn on the same non‐dominant arm throughout the monitoring period. The mean difference (±SD) in systolic BPs (average of 2 readings) between WBPM and ABPM was 0.8 ± 12.8 mm Hg (P = .564) in the office and 3.2 ± 17.0 mm Hg (P < .001) outside the office. The proportion of differences that were within ±10 mm Hg was 58.7% in the office and 47.2% outside the office. In a mixed‐effects model analysis, the temporal trend in the difference between the out‐of‐office BPs measured by the two devices was not statistically significant. In conclusion, the difference between the WBPM and ABPM device was acceptable both in and out of the office.

Integrity signals

Registry

Crossref · 10.1111/jch.13799

Cites

21 of 21 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • In office-based setting, the mean difference in systolic blood pressure between the HeartGuide wristwatch and an ambulatory BP device was 0.8 ± 12.8 mmHg.

    “The mean difference (±s.d.) in systolic BP between both groups was 0.8 ± 12.8 mmHg in the office-based setting and 3.2 ± 17.0 mmHg in the ambulatory setting25.”
  • In ambulatory setting, the mean difference in systolic blood pressure between the HeartGuide wristwatch and an ambulatory BP device was 3.2 ± 17.0 mmHg.

    “The mean difference (±s.d.) in systolic BP between both groups was 0.8 ± 12.8 mmHg in the office-based setting and 3.2 ± 17.0 mmHg in the ambulatory setting25.”

Cites

1 retracted

Flagged first, then by citations

1 of the 22 references couldn’t be matched to a known paper.

Cited by

Sorted by citations

Authors
Kario, K., Shimbo, D., Tomitani, N., Kanegae, H., Schwartz, J. E., & Williams, B.
Affiliations
  • Jichi Medical University (JP)
  • Columbia University Irving Medical Center (US)
  • Stony Brook University (US)
  • University College London (GB)
Corresponding author
Kario, Kazuomi
Venue
The Journal of Clinical Hypertension
Publisher
Wiley
Published
2020-01-19
Volume
22
Issue
2
Pages
135-141
ISSN
  • 1524-6175 (ISSN-L)
  • 1751-7176
OpenAlex
W2999897056
Cited by
112
References
22
Citations by year
2026: 11 · 2025: 12 · 2024: 17 · 2023: 22 · 2022: 23 · 2021: 12 · 2020: 15
FWCI
11.14
Type
journal article
Language
en
Open access
Locations
  • Journal of Clinical Hypertension (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Funders
Omron Healthcare Co., Ltd.
Topics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
Keywords
wearable blood pressure monitor, ambulatory blood pressure monitoring, device agreement, out-of-office blood pressure measurement, antihypertensive therapy
MeSH terms
Wearable Electronic Devices, Aged, Blood Pressure, Humans, Hypertension, Middle Aged, Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review