Minimally interactive placenta segmentation from three-dimensional ultrasound images
Open accessAbstract
Purpose: Placental size in early pregnancy has been associated with important clinical outcomes, including fetal growth. However, extraction of placental size from three-dimensional ultrasound (3DUS) requires time-consuming interactive segmentation methods and is prone to user variability. We propose a semiautomated segmentation technique that requires minimal user input to robustly measure placental volume from 3DUS images. Approach: For semiautomated segmentation, a single, central 2D slice was manually annotated to initialize an automated multi-atlas label fusion (MALF) algorithm. The dataset consisted of 47 3DUS volumes obtained at 11 to 14 weeks in singleton pregnancies (28 anterior and 19 posterior). Twenty-six of these subjects were imaged twice within the same session. Dice overlap and surface distance were used to quantify the automated segmentation accuracy compared to expert manual segmentations. The mean placental volume measurements obtained by our method and VOCAL (virtual organ computer-aided analysis), a leading commercial semiautomated method, were compared to the manual reference set. The test–retest reliability was also assessed. Results: The overlap between our automated segmentation and manual (mean Dice: 0.824 ± 0.061, median: 0.831) was within the range reported by other methods requiring extensive manual input. The average surface distance was 1.66 ± 0.96 mm. The correlation coefficient between test–retest volumes was r = 0.88, and the intraclass correlation was ICC ( 1 ) = 0.86. Conclusions: MALF is a promising method that can allow accurate and reliable segmentation of the placenta with minimal user interaction. Further refinement of this technique may allow for placental biometry to be incorporated into clinical pregnancy surveillance.
AI snapshot
Written from the paper’s full text.
TL;DR
A minimally interactive multi-atlas label fusion method segments the placenta from first-trimester 3D ultrasound using only a single manually annotated 2D slice as input. It achieved Dice overlap of 0.824 ± 0.061 with expert manual segmentations and test–retest reliability of r = 0.88 / ICC(1) = 0.86.
Key findings
- Overall Dice overlap between automated and manual segmentations was 0.824 ± 0.061 (median 0.831); anterior placentas: 0.838 ± 0.063 (median 0.852), posterior placentas: 0.803 ± 0.055 (median 0.821).
- Mean surface distance was 1.66 ± 0.96 mm overall, 1.61 ± 1.17 mm for anterior and 1.73 ± 0.53 mm for posterior placentas.
- Mean placental volume from the proposed method was 79.0 ± 18.2 cc, not significantly different from manual reference (80.7 ± 30.5 cc; p > 0.54), whereas VOCAL gave 69.7 ± 29.7 cc, significantly smaller than manual (p < 1 × 10⁻⁷).
- Test–retest volume correlation was r = 0.88 and ICC(1) = 0.86 (95% CI 0.72–0.94); the method took <2 min per image on average.
Methods
- Dataset: 47 3DUS volumes from singleton pregnancies at 11–14 weeks (28 anterior, 19 posterior placentas); 26 subjects were imaged twice in the same session.
- Each target image was initialized by manual segmentation of one central 2D slice, propagated to a 3D slab, then segmented via multi-atlas registration and joint label fusion, repeated with bootstrapping and postprocessing.
- Accuracy was evaluated with leave-one-out atlases against expert manual segmentations using Dice overlap, surface distance, and volume comparison to VOCAL; test–retest reliability used correlation and ICC.
Limitations
- The paper states that the cohort was limited to singleton first-trimester pregnancies; later stages and nonsingletons remain future work.
- The paper notes that minimal user input may still be prohibitive for large population studies, and that full automation remains a goal.
- The paper reports lower posterior performance and attributes it partly to fewer posterior atlases; it also shows sensitivity to manual input extent, with uneven undersegmentation or a thin centerline leading to missing parts or inadequate initialization.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 34 cited sources checked · 0 supported · 0 flagged
- Findingown data§ Abstract
The correlation coefficient between test-retest placental volumes was r = 0.88.
“The correlation coefficient between test -retest volumes was r ¼ 0.88 , and the intraclass correlation was ICC ð 1 Þ ¼ 0.86 .”
Quote in the paperSubstantial claim- Rests on an assumption: The two 3DUS acquisitions from the same subject within the same session represent the same underlying placental volume, so their volume difference reflects measurement reliability rather than biological change.
- Findingown data§ 3.1 Accuracy Experiment
The mean Dice overlap in the posterior placentas in the accuracy experiment was 0.803 ± 0.055 (N = 19, median 0.821).
“and 0.803 0.055 in the posterior placentas ( N ¼ 19 , median: 0.821).”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
The median Dice overlap in the accuracy experiment was 0.831.
“The median Dice overlap was 0.831.”
Quote in the paperSubstantial claim - Findingown data§ Abstract
The intraclass correlation between test-retest placental volumes was ICC(1) = 0.86.
“The correlation coefficient between test -retest volumes was r ¼ 0.88 , and the intraclass correlation was ICC ð 1 Þ ¼ 0.86 .”
Quote in the paperSubstantial claim- Rests on an assumption: The two 3DUS acquisitions from the same subject within the same session represent the same underlying placental volume, so their volume difference reflects measurement reliability rather than biological change.
- Findingown data§ 3.1 Accuracy Experiment
The mean Dice overlap in the anterior placentas in the accuracy experiment was 0.838 ± 0.063 (N = 28, median 0.852).
“The mean Dice overlap was 0.838 0.063 in the anterior placentas ( N ¼ 28 , median: 0.852)”
Quote in the paperSubstantial claim
- Findingown data§ Abstract
The average surface distance between automated and manual segmentation was 1.66 ± 0.96 mm.
“The average surface distance was 1.66 0.96 mm .”
Quote in the paperSubstantial claim- Rests on an assumption: Expert manual segmentations are accurate enough to serve as the reference standard for evaluating automated segmentation accuracy.
- Findingown data§ 2.3.4 Multi-atlas joint label fusion
The JLF segmentation sometimes erroneously includes a small portion of amniotic fluid due to registration errors.
“In particular, sometimes the JLF segmentation erroneously includes a small portion of the amniotic fluid due to registration errors.”
Quote in the paperNot a substantial claim - Findingown data§ 3.1 Accuracy Experiment
The overall Dice overlap ratio between the minimally interactive segmentation method and expert manual segmentations in the accuracy experiment was 0.824 ± 0.061 (mean ± standard deviation, N = 47).
“The overall Dice overlap ratio was 0.824 0.061 (mean ± standard deviation and N ¼ 47 ).”
- Also stated in: § 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods · plain
“Our mean Dice overlap in this study is 0.82.”
Quote in the paperSubstantial claim - Also stated in: § 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods · plain
- Findingown data§ Abstract
The mean Dice overlap between automated segmentation and manual segmentation was 0.824 ± 0.061, with median 0.831.
“The overlap between our automated segmentation and manual (mean Dice: 0.824 0.061 , median: 0.831) was within the range reported by other methods requiring extensive manual input.”
Quote in the paperSubstantial claim- Rests on an assumption: Expert manual segmentations are accurate enough to serve as the reference standard for evaluating automated segmentation accuracy.
- Findingown data§ 3.1 Accuracy Experiment
The mean surface distance for the posterior placentas in the accuracy experiment was 1.73 ± 0.53 mm (N = 19).
“and 1.73 0.53 mm for the posterior placentas ( N ¼ 19 ).”
Quote in the paperNot a substantial claim - Findingown data§ 3.1 Accuracy Experiment
The Dice overlap for the 3D slab stage was 0.699 ± 0.062 (median 0.698).
“The Dice overlap was 0.699 0.062 for the 3D slab (median: 0.698).”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
The mean surface distance for the entire population in the accuracy experiment was 1.66 ± 0.96 mm (N = 47).
“The mean surface distance was 1.66 0.96 mm for the entire population ( N ¼ 47 ).”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
The mean surface distance for the anterior placentas in the accuracy experiment was 1.61 ± 1.17 mm (N = 28).
“The mean surface distance was 1.61 1.17 mm for the anterior placentas ( N ¼ 28 )”
Quote in the paperNot a substantial claim - Findingown data§ 3.2 Test -Retest Reliability Experiment
The correlation coefficient between the volume estimates obtained from the two different 3DUS images acquired from each patient was r = 0.88 (N = 26).
“The correlation coefficient was r ¼ 0.88 between the volume estimates obtained from the two different 3DUS images acquired from each patient ( N ¼ 26 ).”
Quote in the paperSubstantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
A thin centerline is completely inadequate for initialization.
“whereas a thin centerline is completely inadequate for initialization.”
Quote in the paperNot a substantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
An uneven segmentation leads to missing parts in the final segmentation.
“An uneven segmentation leads to missing parts in the final segmentation”
Quote in the paperNot a substantial claim - Findingown data§ 3.2 Test -Retest Reliability Experiment
No statistically significant differences were observed between the two sets of measurements (p > 0.10 in paired t-test).
“No statistically significant differences were observed between the two sets of measurements ( p > 0.10 in paired t -test).”
Quote in the paperSubstantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
In the test-retest reliability experiment, a small difference in Dice overlap performance between the anterior and posterior groups reached statistical significance (p < 0.03) in a two-sample equal variance t-test with one tail.
“Furthermore, in the test -retest reliability experiment [Fig. 6(b)], a small difference in Dice overlap performance was again observed between the anterior and posterior groups. As in the accuracy experiment, this difference reached statistical significance ( p < 0.03 ) in two-sample equal variance t -test with one tail.”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
The difference between the initial JLF estimate and the final segmentation was statistically significant (p < 0.003).
“The difference between the initial JLF estimate and the final segmentation was also statistically significant ( p < 0.003 ).”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
No statistically significant differences were found between the proposed method and manual volume measurements (p > 0.54) in a paired t-test.
“No statistically significant differences were found between the two measurements ( p > 0.54 ) in a paired t -test.”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
The difference between the 3D slab and the initial JLF estimate was highly statistically significant (p < 1 × 10^-16) in a one-tailed paired t-test.
“The difference between the 3D slab and the initial JLF estimate was highly statistically significant ( p < 1 × 10 -16 ) in a one-tailed paired t -test.”
Quote in the paperSubstantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
In the accuracy experiment, a small difference in Dice overlap performance between the anterior and posterior groups reached statistical significance (p < 0.03) in a two-sample equal variance t-test with one tail.
“In the accuracy experiment, a small difference in Dice overlap performance was observed between the anterior and posterior groups [Fig. 3(a)], which reached statistical significance ( p < 0.03 ) in two-sample equal variance t -test with one tail.”
Quote in the paperNot a substantial claim - Findingown data§ 3.1 Accuracy Experiment
The placental volume obtained from the expert manual segmentations was 80.7 ± 30.5 cc (mean ± standard deviation).
“The placental volume obtained from the expert manual segmentations was 80.7 30.5 cc (mean ± standard deviation)”
Quote in the paperSubstantial claim - Findingown data§ 3.2 Test -Retest Reliability Experiment
The ICC(1) was 0.86 (95% confidence interval: 0.72 < ICC < 0.94).
“The ICC was ICC ð 1 Þ ¼ 0.86 (95% confidence interval: 0.72 < ICC < 0.94 ).”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
After the first round of MALF, the Dice overlap was 0.812 ± 0.061 (median 0.823).
“After the first round of MALF, the Dice overlap was 0.812 0.061 (median: 0.823).”
Quote in the paperSubstantial claim - Findingown data§ 3.2 Test -Retest Reliability Experiment
The Dice overlap in the anterior placentas in the test-retest reliability experiment was 0.835 ± 0.066 (N = 14, median 0.850).
“The Dice overlap was 0.835 0.066 in the anterior placentas ( N ¼ 14 , median: 0.850)”
Quote in the paperSubstantial claim - Findingown data§ 3.2 Test -Retest Reliability Experiment
The overall Dice overlap ratio in the test-retest reliability experiment was 0.812 ± 0.064 (mean ± standard deviation, N = 26, median 0.818).
“The overall Dice overlap ratio was 0.812 0.064 (mean ± standard deviation, N ¼ 26 , median: 0.818).”
Quote in the paperSubstantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
The atlas set in the test-retest reliability experiment contained only 7 posterior placentas compared to 14 anterior placentas.
“We note that the atlas set in this experiment only contained 7 posterior placentas compared to 14 anterior placentas.”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
After bootstrapping, the final Dice overlap was 0.824 ± 0.061 (median 0.831).
“After bootstrapping, the final Dice overlap was 0.824 0.061 (median: 0.831).”
Quote in the paperSubstantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
An even undersegmentation also yields a comparable result, with some localized defects.
“An even undersegmentation also yields a comparable result, with some localized defects.”
Quote in the paperNot a substantial claim - Findingown data§ 4.2 Performance on Anterior versus Posterior Placentas
Initialization with a bounding box and with an oversegmentation lead to practically the same result as using the true 2D segmentation as input.
“Initialization with a bounding box and with an oversegmentation lead to practically the same result as using the true 2D segmentation as input.”
Quote in the paperSubstantial claim - Findingnone given§ 3.1 Accuracy Experiment
For both typical subjects, the very thin, high curvature extremities of the placenta and the regions of tissue that appear hypoechoic are challenging to fully capture with the method.
“Nevertheless, we note that for both subjects, the very thin, high curvature extremities of the placenta as well as the regions of the tissue that appear hypoechoic are challenging to fully capture with our method.”
Quote in the paperNot a substantial claim - Findingown data§ 3.2 Test -Retest Reliability Experiment
The Dice overlap in the posterior placentas in the test-retest reliability experiment was 0.784 ± 0.052 (N = 12, median 0.800).
“and 0.784 0.052 in the posterior placentas ( N ¼ 12 , median: 0.800).”
Quote in the paperSubstantial claim - Findingown data§ 3.1 Accuracy Experiment
The placental volume obtained from the proposed method was 79.0 ± 18.2 cc.
“compared to 79.0 18.2 cc for our proposed method.”
Quote in the paperSubstantial claim - Findingown data§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The performance deterioration from the accuracy experiment to the reliability experiment is relatively small (0.824 ± 0.061 in the accuracy experiment compared to 0.812 ± 0.064 in the reliability experiment).
“Nevertheless, we note that the performance deterioration is relatively small ( 0.824 0.061 in the accuracy experiment, compared to 0.812 0.064 in the reliability experiment).”
Quote in the paperSubstantial claim - Findingreasoning§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
In the test-retest reliability experiment, the Dice overlap performances are lower than in the accuracy experiment.
“In the test -retest reliability experiment, the Dice overlap performances are lower than the accuracy experiment [Fig. 6(b)].”
Quote in the paperNot a substantial claim - Findingown data§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The proposed method takes less than 2 minutes per image on average, including the time to load the 3DUS volume, select a slice toward the middle of the placenta, and manually trace the placenta on that slice.
“This takes < 2 min per image on average, including the time to load the 3DUS volume, select a slice toward the middle of the placenta, and manually trace the placenta on that slice.”
Quote in the paperSubstantial claim - Premisereasoning§ 1 Introduction
Multi-atlas segmentation is particularly relevant for placental imaging where maternal habitus and fetal shadowing artifact can obscure the boundaries between the placenta and surrounding tissue.
“This technique is particularly relevant for placental imaging where maternal habitus and fetal shadowing artifact can obscure the boundaries between the placenta and surrounding tissue.”
Quote in the paperSubstantial claim - Interpretationown data§ Abstract
The overlap between automated segmentation and manual segmentation was within the range reported by other methods requiring extensive manual input.
“The overlap between our automated segmentation and manual (mean Dice: 0.824 0.061 , median: 0.831) was within the range reported by other methods requiring extensive manual input.”
Quote in the paperSubstantial claim- Rests on an assumption: Expert manual segmentations are accurate enough to serve as the reference standard for evaluating automated segmentation accuracy.
- Interpretationown data§ Abstract
MALF is a promising method that can allow accurate and reliable segmentation of the placenta with minimal user interaction.
“MALF is a promising method that can allow accurate and reliable segmentation of the placenta with minimal user interaction.”
Quote in the paperSubstantial claim- Rests on an assumption: Expert manual segmentations are accurate enough to serve as the reference standard for evaluating automated segmentation accuracy.
- Rests on an assumption: The two 3DUS acquisitions from the same subject within the same session represent the same underlying placental volume, so their volume difference reflects measurement reliability rather than biological change.
- Premisereasoning§ 1 Introduction
Quantification of the placenta from 3DUS images is difficult to automate, especially in early pregnancy when the appearance of the placenta and uterine tissue are not well differentiated.
“However, quantification of the placenta from 3DUS images is a difficult task to automate. This is especially true in early pregnancy when the appearance of the placenta and uterine tissue are not well differentiated.”
Quote in the paperSubstantial claim - Premisereasoning§ 2.3.4 Multi-atlas joint label fusion
The performance of MALF is largely determined by the accuracy of atlas-target registrations.
“The performance of MALF is largely determined by the accuracy of atlas-target registrations.”
Quote in the paperSubstantial claim - Premisereasoning§ 1 Introduction
Weak contrast and noise in 3DUS images make it difficult to accurately identify the placenta boundary.
“This weak contrast, combined with the noisy nature of 3DUS images, makes it difficult to accurately identify the placenta boundary.”
Quote in the paperSubstantial claim - Premisecitation§ 1 Introduction
Placental size and shape are significantly associated with perinatal outcomes and fetal programming.
“Researchers have observed that placental size and shape are significantly associated with perinatal outcomes and fetal programming. 1 -3”
Quote in the paperSubstantial claim0 of 3 sources checked - Interpretationreasoning§ 4.3 Sensitivity to Manual Input
Even if the manual mask contains about half as many pixels as the true placenta in the initial 2D slice, this may be relatively well tolerated by the algorithm if these pixels are evenly distributed across the placenta.
“Even if the manual mask contains about half as many pixels as the true placenta in the initial 2D slice, this may be relatively well tolerated by the algorithm if these pixels are evenly distributed across the placenta.”
Quote in the paperNot a substantial claim - Interpretationreasoning§ 4.3 Sensitivity to Manual Input
The extent of the manual input can be quite important, perhaps more so than the boundary placement.
“However, we note that the extent of the manual input can be quite important, perhaps more so than the boundary placement.”
Quote in the paperNot a substantial claim - Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The interactive random walker method proposed by Stevenson et al. requires extensive user input in 3D.
“but requires extensive user input in 3D”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Stevenson et al. 9
Not checked in the free run.
- Interpretationown data§ 5 Conclusion
The method is not very sensitive to imperfect manual input.
“We have also shown that our method is not very sensitive to imperfect manual input.”
Quote in the paperSubstantial claim - Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The fully automated MALF method proposed by Oguz et al. is limited to anterior placentas only.
“but is limited to anterior placentas only.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Oguz et al. 10
Not checked in the free run.
- Interpretationreasoning§ 4.3 Sensitivity to Manual Input
The precise boundaries of the manual input do not directly contribute to the final outcome.
“As such, the precise boundaries of the manual input do not directly contribute to the final outcome.”
Quote in the paperSubstantial claim - Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The interactive random walker method proposed by Stevenson et al. reaches a Dice overlap of 0.86.
“The interactive random walker method proposed by Stevenson et al. 9 reaches a Dice overlap of 0.86”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Stevenson et al. 9
Not checked in the free run.
- Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The fully automated deep learning method proposed by Looney et al. is applicable to both anterior and posterior placentas.
“The fully automated deep learning method proposed by Looney et al. 11 is applicable to both anterior and posterior placentas”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Looney et al. 11
Not checked in the free run.
- Study design and conductnone given§ 5 Conclusion
The entire cohort in this study were singleton pregnancies within the first trimester.
“An important limitation of this study is that the entire cohort were singleton pregnancies within the first trimester.”
Quote in the paperSubstantial claim - Interpretationreasoning§ 4.2 Performance on Anterior versus Posterior Placentas
The difference in performance between anterior and posterior placentas is likely caused by the more challenging nature of the segmentation task in posterior placentas, where fetal shadowing is common.
“This difference in performance is likely caused by the more challenging nature of the segmentation task in posterior placentas, where fetal shadowing is common.”
Quote in the paperNot a substantial claim- Rests on an assumption: Fetal shadowing is common in posterior placentas.
- Interpretationreasoning§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The extensive user input and sensitivity to initialization of the Stevenson et al. method may be especially prohibitive for clinical adoption.
“this may be especially prohibitive for clinical adoption.”
Quote in the paperSubstantial claim - Interpretationreasoning§ 4.3 Sensitivity to Manual Input
If only the left half of the placenta is included in the manual input, it is quite unlikely that the algorithm will be able to recover the right half.
“However, if, for example, only the left half of the placenta is included, it is quite unlikely that the algorithm will be able to recover the right half.”
Quote in the paperNot a substantial claim - Interpretationown data§ 3.1 Accuracy Experiment
The overall agreement between the method's results and the expert manual segmentation is very good in typical subjects.
“We note that the overall agreement between our results and the expert manual segmentation is very good in these typical subjects.”
Quote in the paperSubstantial claim - Interpretationown data§ 4.2 Performance on Anterior versus Posterior Placentas
The test-retest results further support the argument that increasing the number of posterior placentas in the library might improve the performance in posterior placenta segmentations.
“This further supports the argument that increasing the number of posterior placentas in the library might improve the performance in the posterior placenta segmentations.”
Quote in the paperNot a substantial claim - Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
An extension of the Looney et al. method reported by Looney et al. 13 achieves a mean Dice overlap of 0.81.
“An extension of the method by Looney et al., 11 again reported by Looney et al., 13 achieves a mean Dice overlap of 0.81.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Looney et al., 13
Not checked in the free run.
- Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The fully automated deep learning method proposed by Looney et al. has rather limited accuracy, with a median Dice overlap of 0.73 reported.
“but its accuracy is rather limited, with a median Dice overlap of 0.73 reported.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Looney et al. 11
Not checked in the free run.
- Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The fully automated MALF method proposed by Oguz et al. achieves a Dice overlap of 0.83.
“The fully automated MALF method proposed by Oguz et al. 10 achieves 0.83”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Oguz et al. 10
Not checked in the free run.
- Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The interactive random walker method proposed by Stevenson et al. is sensitive to subtle variations in initialization.
“and is sensitive to subtle variations in initialization”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Stevenson et al. 9
Not checked in the free run.
- Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The Looney et al. study evaluated the accuracy of the method against the results of the random walker method rather than against expert manual segmentations.
“An additional shortcoming of this study is that the accuracy of the method is evaluated against the results of the random walker method, 9 rather than to expert manual segmentations.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites Looney et al. 11
Not checked in the free run.
- cites random walker method, 9
Not checked in the free run.
- Interpretationreasoning§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The proposed method offers a solution for salvaging images where fully automated methods may fail for a variety of reasons (e.g., poor image quality or unusual anatomy that is not well represented by the training set).
“our proposed method offers a solution for salvaging images where fully automated methods may fail for a variety of reasons (e.g., poor image quality or unusual anatomy that is not well represented by the training set).”
Quote in the paperSubstantial claim- Rests on an assumption: No fully automated method will work 100% of the time.
- Interpretationreasoning§ 4.3 Sensitivity to Manual Input
If the manual input segmentation is too small, this can be problematic since such small features are prone to disappear during the slice-to-slice propagation step, due to interpolation.
“Additionally, we note that if the manual input segmentation is too small, this can be problematic since such small features are prone to disappear during the slice-to-slice propagation step, due to interpolation.”
Quote in the paperNot a substantial claim - Interpretationown data§ 3.1 Accuracy Experiment
The results illustrate the gain in performance provided by each of the three stages in the algorithm.
“These results illustrate the gain in the performance provided by each of the three stages in our algorithm.”
Quote in the paperSubstantial claim - Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The Yang et al. 12 method has limited performance for placenta segmentation, with a Dice overlap of 0.64 reported.
“However, it has limited performance for placenta segmentation (Dice overlap of 0.64 reported).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Yang et al., 12
Not checked in the free run.
- Interpretationown data§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
In contrast to other methods, the proposed method achieves high accuracy with minimal user input.
“In contrast, our method achieves high accuracy with minimal user input.”
Quote in the paperSubstantial claim - Interpretationreasoning§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation Methods
The lower Dice overlap in the test-retest reliability experiment is likely due to the much smaller library of 3D atlas images used in that experiment.
“This is likely due to the much smaller library of 3D atlas images here.”
Quote in the paperNot a substantial claim
- May combine two claims. 18
- Classified as probably not a substantial claim. 13
May combine two claims.
Findingown data§ 2.3.3 Interactive initialization based on manual input on a single 2D sliceThe 3D slab from propagating a single manually segmented 2D slice provides a robust estimation of the 3D shape and position of the placenta but its exact boundaries and extent are not yet identified at this stage.
“The overlay with the manual 3D segmentation (shown in green) shows that this procedure provides a robust estimation of the 3D shape and position of the placenta but its exact boundaries and extent are not yet identified at this stage.”
Quote in the paperNot a substantial claim- Rests on an assumption: Expert manual segmentations are accurate enough to serve as the reference standard for evaluating automated segmentation accuracy.
May combine two claims.
Findingown data§ 3.1 Accuracy ExperimentFor VOCAL, the mean volume was 69.7 ± 29.7 cc, which was significantly smaller than the manual volume (p < 1 × 10^-7, paired t-test).
“However, for VOCAL, the mean volume was 69.7 29.7 cc , which was significantly smaller than the manual volume ( p < 1 × 10 -7 , paired t -test).”
Quote in the paperSubstantial claimMay combine two claims.
Premisecitation§ 1 IntroductionUsing a library of atlases manually created by experts allows the automated segmentation approach to incorporate knowledge on the appearance and shape of the placenta in a diverse set of subjects and leads to highly robust segmentation results in new subjects.
“Using a library of atlases manually created by experts allows the automated segmentation approach to incorporate knowledge on the appearance and shape of the placenta in a diverse set of subjects and leads to highly robust segmentation results in new subjects.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ 1 IntroductionMulti-atlas segmentation has been applied successfully to many challenging segmentation problems such as identifying hippocampal subfields in MRI, the spinal cord, and the mitral valve in preoperative 3D echocardiography.
“This approach has been applied successfully to many challenging segmentation problems 15 such as identifying hippocampal subfields in magnetic resonance images (MRI) of the brain, 17,18 the spinal cord, 19 and the mitral valve in preoperative 3D echocardiography. 20”
Quote in the paperNot a substantial claim0 of 4 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone given§ 1 IntroductionUterine contractions can dramatically affect the shape of the placenta.
“Uterine contractions can also dramatically affect the shape of the placenta.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductown data§ 2.2 Expert Manual SegmentationEach of the 47 subjects was manually segmented under the supervision of a clinical expert with over 10 years of experience in prenatal ultrasound imaging, who provided guidance, final inspection, and approval.
“Each of the 47 subjects were manually segmented under the supervision of our clinical expert, who has over 10 years of experience in prenatal ultrasound imaging and who has segmented hundreds of placentas for other research endeavors. The clinical expert provided guidance for the manual segmentation training and performed final inspection and approval for each segmentation.”
Quote in the paperSubstantial claimMay combine two claims.
Premisecitation§ 1 IntroductionCurrent techniques for quantifying placental volume and morphology from 3DUS images require substantial manual input, which can be cumbersome and poorly standardized.
“As a result of these challenges, current techniques for quantifying placental volume and morphology from 3DUS images require substantial amount of manual input from the user, which can be cumbersome and poorly standardized.”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
Premisereasoning§ 1 IntroductionThe shape and relative location of the placenta are highly variable, which can make automatic detection of the placenta location difficult.
“Additionally, the shape and relative location of the placenta are highly variable. In particular, placentas can be positioned anteriorly, posteriorly, or laterally to the amniotic sac. This variability can make it difficult to even detect the location of the placenta automatically, much less determine its precise boundaries.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone given§ 2.1 DataThe study protocol was approved by the University of Pennsylvania Institutional Review Board and informed consent was obtained from all subjects.
“The study protocol was approved by the University of Pennsylvania Institutional Review Board (IRB# 811129). Informed consent was obtained from all subjects.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ 1 IntroductionA first fully automated algorithm based on multi-atlas segmentation techniques was proposed by Oguz et al.
“A first fully automated algorithm based on multi-atlas segmentation techniques was proposed by Oguz et al., 10”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Oguz, 2016, Fully automated placenta segmentation from 3D ultrasound images
Not checked in the free run.
May combine two claims.
Premisecitation§ 1 IntroductionThe previously demonstrated fully automated multi-atlas framework was limited to anterior-only placentas, and its generalization to arbitrarily positioned placentas is problematic due to initialization problems from variability in location and pose.
“Although we previously demonstrated the applicability of this fully automated framework in a limited dataset of anterior-only placentas, 10 the generalization of this technique to arbitrarily positioned placentas is problematic due to initialization problems arising from the wide variability in location and pose of the placenta.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [10] Oguz, 2016, Fully automated placenta segmentation from 3D ultrasound images
Not checked in the free run.
May combine two claims.
Premisereasoning§ 2.3.4 Multi-atlas joint label fusionAccurate image registration is of paramount importance to MALF success, and this can be problematic for placental imaging because the shape and position of the placenta within the imaging window are highly variable.
“Since the MALF is a registration-based method, accurate image registration is of paramount importance to its success. This can be problematic for our application because the shape and position of the placenta within the imaging window are highly variable.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ 1 IntroductionTwo-dimensional and three-dimensional ultrasound have been used to study in utero placental size and morphology early in pregnancy in relation to birthweight and preeclampsia.
“In particular, two-dimensional and three-dimensional ultrasound (3DUS) have been used to study in utero placental size and morphology early in pregnancy in relation to birthweight 4 and preeclampsia. 5,6”
Quote in the paperNot a substantial claim0 of 3 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation§ 1 IntroductionAlternative methods based on deep learning have been proposed for placental segmentation.
“and a few alternative methods based on the recently popular deep learning methods have been proposed. 11 -14”
Quote in the paperNot a substantial claim0 of 4 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone given§ 1 IntroductionPosterior placentas are often particularly challenging to segment due to shadowing from the fetus.
“Furthermore, posterior placentas are often particularly challenging to segment due to shadowing from the fetus.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Premisenone given§ 2.3.4 Multi-atlas joint label fusionThe placenta appears bright in 3DUS images, whereas amniotic fluid appears dark.
“The placenta appears bright in the 3DUS images, whereas the fluid appears dark.”
Quote in the paperSubstantial claimMay combine two claims.
Premisereasoning§ 1 IntroductionRecently developed fully automated methods are becoming increasingly capable for population-based studies but are currently far from guaranteeing adequate performance in every subject; difficult-to-segment images continue to pose a significant challenge.
“Although these recently developed methods are becoming increasingly capable for fully automated population-based studies, they are currently far from guaranteeing adequate performance in every subject; as such, difficult-to-segment images such as those with a high amount of noise or unusually shaped placenta continue to pose a significant challenge.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone given§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation MethodsSuch failure cases may form a relatively small portion of a large dataset and can arguably be excluded from a population study, although selection bias may be a concern.
“Such failure cases may form a relatively small portion of a large dataset and can arguably be excluded from a population study (although selection bias may be a concern).”
Quote in the paperNot a substantial claimMay combine two claims.
Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation MethodsThe Oguz et al. 14 method combines deep learning techniques and the multi-atlas approach, unlike the currently presented method which only uses the multi-atlas component.
“This method 14 combines deep learning techniques and the multiatlas approach, unlike the currently presented method which only uses the multi-atlas component.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationreasoning§ 4.2 Performance on Anterior versus Posterior PlacentasAn additional contributor to the performance difference may be the relatively small number of posterior subjects in the atlas library; increasing the number of posterior placentas in the library might provide a better representation of shape and appearance distribution and help further minimize the performance difference.
“An additional contributor to the performance difference may be the relatively small number of posterior subjects in our atlas library; increasing the number of posterior placentas in the library might provide a better representation of shape and appearance distribution and help further minimize the performance difference.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown data§ 5 ConclusionThe presented minimally interactive placenta segmentation method achieves excellent performance while only requiring user input on a single 2D slice of a 3DUS volume.
“We presented a minimally interactive placenta segmentation method that achieves excellent performance while only requiring user input on a single 2D slice of a 3DUS volume.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown data§ 4.3 Sensitivity to Manual InputThese results show that precise manual input is not strictly required, and that it may be better to err on the side of oversegmentation for initializing this method.
“These results show that precise manual input is not strictly required, and that it may be better to err on the side of oversegmentation for initializing this method.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisereasoning§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation MethodsThe Looney et al. 13 method is also evaluated against random walker results rather than expert manual segmentations.
“Like the previous version, 11 this method is also evaluated against random walker results rather than expert manual segmentations.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Looney et al., 13
Not checked in the free run.
May combine two claims.
Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation MethodsAn extension of the multi-atlas approach reported by Oguz et al. 14 makes the method applicable to both anterior and posterior placentas and achieves a mean Dice overlap of 0.86.
“An extension of the multi-atlas approach, 10 reported by Oguz et al., 14 makes the method applicable to both anterior and posterior placentas and achieves a mean Dice overlap of 0.86.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Oguz et al., 14
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisecitation§ 4.1 Segmentation Accuracy and Comparison to Other Placenta Segmentation MethodsAnother fully automated deep learning method, proposed by Yang et al. 12, is a general-purpose segmentation technique that is applicable to many different organs.
“Another fully automated deep learning method, proposed by Yang et al., 12 is a generalpurpose segmentation technique that is applicable to many different organs.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Yang et al., 12
Not checked in the free run.
The paper relies on this without support.
Expert manual segmentations are accurate enough to serve as the reference standard for evaluating automated segmentation accuracy.
“Dice overlap and surface distance were used to quantify the automated segmentation accuracy compared to expert manual segmentations.”
“Each of the 47 subjects were manually segmented under the supervision of our clinical expert, who has over 10 years of experience in prenatal ultrasound imaging and who has segmented hundreds of placentas for other research endeavors.”
The paper relies on this without support.
The two 3DUS acquisitions from the same subject within the same session represent the same underlying placental volume, so their volume difference reflects measurement reliability rather than biological change.
“26 of these subjects (14 anterior and 12 posterior) were imaged twice within the same session; the patients were allowed to move and the ultrasound probe position was changed between the acquisition of the two images.”
The paper relies on this without support.
Later stages of pregnancy are challenging to image with 3DUS as the placenta becomes too big for the imaging window.
“Later stages of pregnancy are challenging to image with 3DUS as the placenta becomes too big for the imaging window.”
The paper relies on this without support.
No fully automated method will work 100% of the time.
“it is well known that no fully automated method will work 100% of the time”
The paper relies on this without support.
Fetal shadowing is common in posterior placentas.
“where fetal shadowing is common”
The paper relies on this without support.
The dataset used to evaluate the Oguz et al. 10 method is a subset of the one used in the current manuscript.
“We note that the dataset used to evaluate this method is a subset of the one used in the current manuscript.”
The paper relies on this without support.
The dataset used to evaluate the Oguz et al. 14 method is the same as the one used in the current manuscript.
“We note that the dataset used to evaluate this method is the same as the one used in the current manuscript.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
There have been semi-automated and fully automated methods proposed for placental segmentation from 3D ultrasound, with varying results.
“There have been some semi-automated and fully automated methods proposed for the segmentation task, with varying results [10,11,12,13,14,15,16,17].”
From Fully Automated Placental Volume Quantification From<scp>3D</scp>Ultrasound for Prediction of Small‐for‐Gestational‐Age Infants (2022) · cites this paper as [15]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Reference 20 is described as addressing a different valve
MinorThe text cites reference 20 for segmentation of the mitral valve, whereas the cited paper's title identifies the tricuspid aortic valve.
“the mitral valve in preoperative 3D echocardiography. 20”
“20. A. M. Pouch et al., ' Automated segmentation and geometrical modeling of the tricuspid aortic valve in 3D echocardiographic images, '”
Random-walker study is attributed to a different lead author
MinorThe discussion attributes reference 9's random-walker method to Stevenson et al., but the paper's reference 9 lists Collins et al. as its authors.
“The interactive random walker method proposed by Stevenson et al. 9 reaches a Dice overlap of 0.86”
“9. L. Collins et al., ' 3D ultrasound segmentation of the placenta using the random Walker algorithm: reliability and agreement, '”
Cites
Yushkevich, P. A., Piven, J., Hazlett, H. C., Smith, R. G., Ho, S., Gee, J. C., & Gerig, G. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2006.01.015 · Cited by 8,868 · see citing works
Self-citationFetal and placental size and risk of hypertension in adult life.
Barker, D. J., Bull, A. R., Osmond, C., & Simmonds, S. J. (1990) · BMJ · doi:10.1136/bmj.301.6746.259 · Cited by 1,728 · see citing works
Multi-Atlas Segmentation with Joint Label Fusion
Wang, H., Suh, J. W., Das, S. R., Pluta, J. B., Craige, C., & Yushkevich, P. A. (2013) · IEEE Transactions on Pattern Analysis and Machine Intelligence · doi:10.1109/tpami.2012.143 · Cited by 796 · see citing works
Self-citationMulti-atlas segmentation of biomedical images: A survey
Iglesias, J. E., & Sabuncu, M. R. (2015) · Medical Image Analysis · doi:10.1016/j.media.2015.06.012 · Cited by 667 · see citing works
Wang, H., & Yushkevich, P. A. (2013) · Frontiers in Neuroinformatics · doi:10.3389/fninf.2013.00027 · Cited by 218 · see citing works
Self-citationWang, H., Das, S. R., Suh, J. W., Altinay, M., Pluta, J., Craige, C., Avants, B., & Yushkevich, P. A. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2011.01.006 · Cited by 177 · see citing works
Self-citationLooney, P., Stevenson, G. N., Nicolaides, K. H., Plasencia, W., Molloholli, M., Natsis, S., & Collins, S. L. (2018) · JCI Insight · doi:10.1172/jci.insight.120178 · Cited by 111 · see citing works
In Vivo Quantification of Placental Insufficiency by BOLD MRI: A Human Study
Luo, J., Türk, E. A., Bibbo, C., Gagoski, B. A., Roberts, D. J., Vangel, M. G., Tempany, C. M., Barnewolt, C. E., Estroff, J. A., Palanisamy, A., et al. (2017) · Scientific Reports · doi:10.1038/s41598-017-03450-0 · Cited by 86 · see citing works
Schwartz, N., Sammel, M. D., Leite, R., & Parry, S. (2014) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2014.02.033 · Cited by 80 · see citing works
Self-citationBiswas, S., & Ghosh, S. K. (2008) · Early Human Development · doi:10.1016/j.earlhumdev.2007.09.017 · Cited by 76 · see citing works
Groupwise multi-atlas segmentation of the spinal cord’s internal structure
Asman, A. J., Bryan, F. W., Smith, S. A., Reich, D. S., & Landman, B. A. (2014) · Medical Image Analysis · doi:10.1016/j.media.2014.01.003 · Cited by 56 · see citing works
Schwartz, N., Wang, E., & Parry, S. (2012) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.11136 · Cited by 49 · see citing works
Self-citationAutomatic 3D ultrasound segmentation of the first trimester placenta using deep learning
Looney, P., Stevenson, G. N., Nicolaides, K. H., Plasencia, W., Molloholli, M., Natsis, S., & Collins, S. L. (2017) · 2017 IEEE 14th International Symposium on Biomedical Imaging (ISBI 2017) · doi:10.1109/isbi.2017.7950519 · Cited by 47 · see citing works
Stevenson, G. N., Collins, S. L., Ding, J., Impey, L., & Noble, J. A. (2015) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2015.07.021 · Cited by 46 · see citing works
Collins, S. L., Stevenson, G. N., Noble, J. A., & Impey, L. (2013) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2012.09.003 · Cited by 39 · see citing works
A Multi-Atlas-Based Segmentation Framework for Prostate Brachytherapy
Nouranian, S., Mahdavi, S. S., Spadinger, I., Morris, W. J., Salcudean, S. E., & Abolmaesumi, P. (2014) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2014.2371823 · Cited by 35 · see citing works
Gross placental measures and childhood growth
Baptiste-Roberts, K., Salafia, C. M., Nicholson, W. K., Duggan, A., Wang, N.-Y., & Brancati, F. L. (2009) · The Journal of Maternal-Fetal & Neonatal Medicine · doi:10.1080/14767050802415728 · Cited by 30 · see citing works
Pouch, A. M., Wang, H., Takabe, M., Jackson, B. M., Sehgal, C. M., Gorman, J. H., Gorman, R. C., & Yushkevich, P. A. (2013) · Lecture Notes in Computer Science · doi:10.1007/978-3-642-40811-3_61 · Cited by 25 · see citing works
Self-citation(2017) · Lecture Notes in Computer Science · doi:10.1007/978-3-319-66182-7 · Cited by 23 · see citing works
Combining Deep Learning and Multi-atlas Label Fusion for Automated Placenta Segmentation from 3DUS
Oguz, B. U., Wang, J., Yushkevich, N., Pouch, A., Gee, J., Yushkevich, P. A., Schwartz, N., & Oguz, I. (2018) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-00807-9_14 · Cited by 13 · see citing works
Self-citation366: Placental volume measurements early in pregnancy predict adverse perinatal outcomes
Schwartz, N., Coletta, J., Srinivas, S., Pessel, C., Timor, I. E., Parry, S., & Salafia, C. (2009) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2009.10.381 · Cited by 3 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Zimmer, V. A., Gómez, A., Skelton, E., Wright, R., Wheeler, G., Deng, S., Ghavami, N., Lloyd, K., Matthew, J., Kainz, B., et al. (2022) · Medical Image Analysis · doi:10.1016/j.media.2022.102639 · Cited by 34 · see citing works
The human placenta project: Funded studies, imaging technologies, and future directions
Herrera, C. L., Kim, M. J., Do, Q. N., Owen, D. M., Fei, B., Twickler, D. M., & Spong, C. Y. (2023) · Placenta · doi:10.1016/j.placenta.2023.08.067 · Cited by 15 · see citing works
Li, J., Shi, Z., Zhu, J., Liu, J., Qiu, L., Song, Y., Wang, L., Li, Y., Liu, Y., Zhang, D., et al. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105680 · Cited by 14 · see citing works
A Multi-task Approach Using Positional Information for Ultrasound Placenta Segmentation
Zimmer, V. A., Gómez, A., Skelton, E., Ghavami, N., Wright, R., Li, L., Matthew, J., Hajnal, J. V., & Schnabel, J. A. (2020) · Lecture notes in computer science · doi:10.1007/978-3-030-60334-2_26 · Cited by 13 · see citing works
Schwartz, N., Oguz, I., Wang, J., Pouch, A., Yushkevich, N., Parameshwaran, S., Gee, J., Yushkevich, P., & Oguz, B. (2022) · Journal of Ultrasound in Medicine · doi:10.1002/jum.15835 · Cited by 11 · see citing works
Self-citationJittou, A., Fazazy, K. E., & Riffi, J. (2025) · Visual Computing for Industry Biomedicine and Art · doi:10.1186/s42492-025-00197-8 · Cited by 6 · see citing works
Charting normative reference values and Z-scores for MRI-derived in vivo placental growth
Jacobwitz, M., Ngwa, J. S., Kapse, K. J., Limperopoulos, C., & Andescavage, N. N. (2025) · Pediatric Radiology · doi:10.1007/s00247-025-06469-y · Cited by 3 · see citing works
Sagberg, K., Lie, T., Peterson, H. F., Hillestad, V., Eskild, A., & Bø, L. E. (2025) · Minimally Invasive Therapy & Allied Technologies · doi:10.1080/13645706.2025.2449699 · Cited by 3 · see citing works
Li, F., Wang, J., Hong, Y., Yang, X., Jiang, J., Wang, Z., Zeng, S., Fan, Y., Lyu, G., Li, P., et al. (2025) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2025.109110 · Cited by 2 · see citing works
Self-citationGiraldo, S. M., Torres-Valencia, N. A., Torres-Valencia, C. A., & Restrepo, H. F. (2023) · Journal of Clinical Ultrasound · doi:10.1002/jcu.23604 · Cited by 1 · see citing works
A Review of Tools Applied in Processing of Medical Images
Haldorai, A., & Anandakumar, S. (2021) · Journal of Biomedical and Sustainable Healthcare Applications · doi:10.53759/0088/jbsha202101006 · Cited by 1 · see citing works
Interactive Segmentation Model for Placenta Segmentation from 3D Ultrasound Images
Li, H., Oguz, B. U., Arenas, G. A., Yao, X., Wang, J., Pouch, A. M., Byram, B. C., Schwartz, N., & Oğuz, İ. (2024) · Lecture notes in computer science · doi:10.1007/978-3-031-73647-6_13
Self-citationAutomatic Segmentation of Human Placenta from 3D Multimodal Ultrasound Data
Singh, S., Stevenson, G. N., Mian, B., Welsh, A. W., & Sowmya, A. (2024) · International Conference on Digital Image Computing: Techniques and Applications (DICTA) · doi:10.1109/dicta63115.2024.00045
- Authors
- Oguz, I., Yushkevich, N., Pouch, A. M., Oguz, B. U., Wang, J., Parameshwaran, S., Gee, J. C., Yushkevich, P. A., & Schwartz, N.
- Affiliations
- Vanderbilt Univ.
- University of Pennsylvania (US)
- Venue
- Journal of Medical Imaging
- Publisher
- SPIE-Intl Soc Optical Eng
- Published
- 2020-02-22
- Volume
- 7
- Issue
- 01
- Pages
- 1-14
- ISSN
- 2329-4302 (ISSN-L)
- 2329-4310
- PMC
- PMC7035878
- OpenAlex
- W3007685270
- Semantic Scholar
- e5e5f2e2296ee786b678b16c766e0c028ca65b54
- Cited by
- 13
- References
- 21
- Citations by year
- 2025: 4 · 2024: 2 · 2023: 3 · 2022: 1 · 2021: 2 · 2020: 1
- FWCI
- 0.68
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of medical imaging (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Nvidia Corporation
- National Institute of Child Health and Human Development (U01 HD087180)
- National Institute of Biomedical Imaging and Bioengineering (R01 EB017255)
- National Institute of Neurological Disorders and Stroke (R01 NS094456)
- Topics
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Health Sciences › Medicine › Health Informatics › Artificial Intelligence in Healthcare and Education
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Keywords
- placenta segmentation, placental volume, 3D ultrasound, semi-automatic segmentation, multi-atlas label fusion, early pregnancy, test-retest reliability, fetal growth, VOCAL
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-10.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-10.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review