A magnetically actuated microcatheter with soft rotatable tip for enhanced endovascular access and treatment efficiency
Open accessThe DOI matches its Crossref record; 80 of the 80 references it cites were found in a registry or an index.
Abstract
Endovascular interventions require fast access to affected regions, followed by effective treatment. Catheterizations are effective approaches for treating vascular diseases; however, they face challenges in accessibility, efficiency, and invasiveness in narrow, tortuous vascular systems. This study presents a submillimeter magnetically actuated soft rotatable-tipped microcatheter (MSRM) designed to access small blood vessels and provide efficient, minimally invasive therapeutic interventions for blood clot treatment. The MSRM’s rotatable tip design enhances accessibility and navigation speed through a rotation-assisted active steering strategy. Improved blood clot treatment efficiency is achieved through the MSRM’s multifunctionality: It can accelerate drug-blood clot interactions, mechanically break down blood clots, and retrieve clot debris. The low invasiveness is attributed to the soft material design and conservative actuation strategy. The performance of the MSRM is validated in both in vitro phantom studies and in vivo rabbit models, and the invasiveness is evaluated using a human placenta model.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures






![Fig. 7. Invasiveness evaluation in human placenta blood vessels. (A) Setup of the invasiveness evaluation test. We choose four segments from the superficial blood vessels of the human placenta. In segment I, the proposed microcatheter contacts the blood vessel’s inner wall and is rotational actuated for 30 min. In segment II, the proposed microcatheter contacts the blood vessel’s inner wall and is repeatedly inserted 100 times. In segment III, a commercial guidewire is inserted into the blood vessel with tip contact on the vessel wall 100 times. Segment IV is blank control. (B to E) Example images of the histological analysis that show the physical impact on the endothelial cell layer and the inner vessel wall. The black arrows show the cell layer detachments, and the dashed line shows the torn and missing cell layer [rotational actuation and contact frictional insertion (commercial)].](https://figures.cochat.ai/paper-figures/ca12f82f4a7643ddbe1ba23aa17d2627/a-magnetically-actuated-microcatheter-with-soft-rotatable-tip-for-enhanced-figure-7.jpg)
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Soft robots, steerable catheters, and magnetic microbots allow non-traumatic vascular navigation, targeted drug or clot therapies, and on-demand embolization [153,155,156,157,158,159,160] with MRI-guided magnetic navigation enhancing visualization and safety [17,155,156].”
From The Heart’s Electromagnetic Field in Emotions, Empathy and Human Connection: Biosensor-Derived Insights into Heart–Brain Axis Mechanisms and a Basis for Novel BioMagnetoTherapies (2026) · cites this paper as [17]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Saver, J. L. (2006) · Stroke · doi:10.1161/01.str.0000196957.55928.ab · Cited by 2,181
Small vessel disease: mechanisms and clinical implications
Wardlaw, J. M., Smith, C., & Dichgans, M. (2019) · The Lancet Neurology · doi:10.1016/s1474-4422(19)30079-1 · Cited by 1,777
De Luca, G., Suryapranata, H., Ottervanger, J. P., & Antman, E. M. (2004) · Circulation · doi:10.1161/01.cir.0000121424.76486.20 · Cited by 1,542
Ferromagnetic soft continuum robots
Kim, Y., Parada, G. A., Liu, S., & Zhao, X. (2019) · Science Robotics · doi:10.1126/scirobotics.aax7329 · Cited by 1,312
Saver, J. L., Chapot, R., Agid, R., Hassan, A. E., Jadhav, A. P., Liebeskind, D. S., Lobotesis, K., Meila, D., Meyer, L., Raphaeli, G., et al. (2020) · Stroke · doi:10.1161/strokeaha.120.028956 · Cited by 456
Telerobotic neurovascular interventions with magnetic manipulation
Kim, Y., Genevriere, E., Harker, P., Choe, J., Balicki, M., Regenhardt, R. W., Vranic, J. E., Dmytriw, A. A., Patel, A. B., & Zhao, X. (2022) · Science Robotics · doi:10.1126/scirobotics.abg9907 · Cited by 422
Current and Emerging Robot-Assisted Endovascular Catheterization Technologies: A Review
Rafii-Tari, H., Payne, C. J., & Yang, G.-Z. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-013-0946-8 · Cited by 288
Thermal diffusion in liquid mixtures and polymer solutions
Wiegand, S. (2004) · Journal of Physics: Condensed Matter · doi:10.1088/0953-8984/16/10/r02 · Cited by 280
Sperm Micromotors for Cargo Delivery through Flowing Blood
Xu, H., Medina-Sánchez, M., Maitz, M. F., Werner, C., & Schmidt, O. G. (2020) · ACS Nano · doi:10.1021/acsnano.9b07851 · Cited by 264
Ferromagnetic soft catheter robots for minimally invasive bioprinting
Zhou, C., Yang, Y., Wang, J., Wu, Q., Gu, Z., Zhou, Y., Liu, X., Yang, Y., Tang, H., Ling, Q., et al. (2021) · Nature Communications · doi:10.1038/s41467-021-25386-w · Cited by 242
Self-citationYang, X., Shang, W., Lu, H., Liu, Y., Yang, L., Tan, R., Wu, X., & Shen, Y. (2020) · Science Robotics · doi:10.1126/scirobotics.abc8191 · Cited by 210
Self-citationMagnetic control of continuum devices
Edelmann, J., Petruska, A. J., & Nelson, B. J. (2017) · The International Journal of Robotics Research · doi:10.1177/0278364916683443 · Cited by 194
Dexterous helical magnetic robot for improved endovascular access
Dreyfus, R., Boehler, Q., Lyttle, S., Gruber, P., Lussi, J., Chautems, C., Gervasoni, S., Berberat, J., Seibold, D., Ochsenbein-Kölble, N., et al. (2024) · Science Robotics · doi:10.1126/scirobotics.adh0298 · Cited by 185
Magnetically Actuated Continuum Medical Robots: A Review
Yang, Z., Yang, H., Cao, Y., Cui, Y., & Zhang, L. (2023) · Advanced Intelligent Systems · doi:10.1002/aisy.202200416 · Cited by 165
Self-citationMagnetic helical micro-/nanomachines: Recent progress and perspective
Dong, Y., Wang, L., Iacovacci, V., Wang, X., Zhang, L., & Nelson, B. J. (2022) · Matter · doi:10.1016/j.matt.2021.10.010 · Cited by 160
Self-citationHu, J., Huang, S., Zhu, L., Huang, W., Zhao, Y., Jin, K., & ZhuGe, Q. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b09423 · Cited by 159
Mao, L., Yang, P., Tian, C., Shen, X., Wang, F., Zhang, H., Meng, X., & Xie, H. (2024) · Nature Communications · doi:10.1038/s41467-024-48058-x · Cited by 150
Patient-Specific Magnetic Catheters for Atraumatic Autonomous Endoscopy
Pittiglio, G., Lloyd, P., da Veiga, T., Onaizah, O., Pompili, C., Chandler, J. H., & Valdastri, P. (2022) · Soft Robotics · doi:10.1089/soro.2021.0090 · Cited by 144
Adaptive wireless millirobotic locomotion into distal vasculature
Wang, T., Ugurlu, H., Yan, Y., Li, M., Li, M., Wild, A., Yıldız, E., Schneider, M., Sheehan, D., Hu, W., et al. (2022) · Nature Communications · doi:10.1038/s41467-022-32059-9 · Cited by 143
Contact and friction between catheter and blood vessel
TAKASHIMA, K., Shimomura, R., Kitou, T., Terada, H., Yoshinaka, K., & IKEUCHI, K. (2006) · Tribology International · doi:10.1016/j.triboint.2005.10.010 · Cited by 134
Magnetically Actuated Medical Robots: An in vivo Perspective
Nelson, B. J., Gervasoni, S., Chiu, P. W. Y., Zhang, L., & Zemmar, A. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3165713 · Cited by 125
Self-citationFlow driven robotic navigation of microengineered endovascular probes
Pancaldi, L., Dirix, P., Fanelli, A., Lima, A. M., Stergiopulos, N., Mosimann, P. J., Ghezzi, D., & Sakar, M. S. (2020) · Nature Communications · doi:10.1038/s41467-020-20195-z · Cited by 117
Piskarev, Y., Shintake, J., Chautems, C., Lussi, J., Boehler, Q., Nelson, B. J., & Floreano, D. (2022) · Advanced Functional Materials · doi:10.1002/adfm.202107662 · Cited by 109
Blood clot fracture properties are dependent on red blood cell and fibrin content
Fereidoonnezhad, B., Dwivedi, A., Johnson, S., McCarthy, R., & McGarry, P. (2021) · Acta Biomaterialia · doi:10.1016/j.actbio.2021.03.052 · Cited by 104
A Magnetic Continuum Robot With Multi-Mode Control Using Opposite-Magnetized Magnets
Lin, D., Jiao, N., Wang, Z., & Liu, L. (2021) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2021.3061376 · Cited by 97
Flow driven robotic navigation of microengineered endovascular probes
Pancaldi, L., Dirix, P., Fanelli, A., Lima, A. M., Stergiopulos, N., Mosimann, P. J., Ghezzi, D., & Sakar, M. S. (2020) · RePEc: Research Papers in Economics · Cited by 95
tPA-anchored nanorobots for in vivo arterial recanalization at submillimeter-scale segments
Wang, B., Wang, Q., Chan, K. F., Ning, Z., Wang, Q., Ji, F., Yang, H., Jiang, S., Zhang, Z., Ip, B. Y. M., et al. (2024) · Science Advances · doi:10.1126/sciadv.adk8970 · Cited by 95
Self-citationLee, S., Lee, S., Kim, S., Yoon, C.-H., Park, H.-J., Kim, J.-Y., & Choi, H. (2018) · Scientific Reports · doi:10.1038/s41598-018-22110-5 · Cited by 92
Direct activation of platelets by addition of CaCl2 leads coagulation of platelet-rich plasma
Toyoda, T., Isobe, K., Tsujino, T., Koyata, Y., Ohyagi, F., Watanabe, T., Nakamura, M., Kitamura, Y., Okudera, H., Nakata, K., et al. (2018) · International Journal of Implant Dentistry · doi:10.1186/s40729-018-0134-6 · Cited by 90
A Multiscale Assembly and Packaging System for Manufacturing of Complex Micro-Nano Devices
Das, A. N., Murthy, R., Popa, D. O., & Stephanou, H. E. (2011) · IEEE Transactions on Automation Science and Engineering · doi:10.1109/tase.2011.2173570 · Cited by 89
Settecase, F., Sussman, M. S., Wilson, M. W., Hetts, S. W., Arenson, R. L., Malba, V., Bernhardt, A. F., Kucharczyk, W., & Roberts, T. P. L. (2007) · Medical Physics · doi:10.1118/1.2750963 · Cited by 79
Leclerc, J., Zhao, H., Bao, D., & Becker, A. T. (2020) · IEEE Transactions on Robotics · doi:10.1109/tro.2020.2988636 · Cited by 78
Mfoumou, E., Julien Tripette, A., Blostein, M. D., & Cloutier, G. (2013) · Thrombosis Research · doi:10.1016/j.thromres.2013.11.001 · Cited by 67
Maisano, F., Vanermen, H. K., Seeburger, J., Mack, M. J., Falk, V., Denti, P., Taramasso, M., & Alfieri, O. (2012) · European Journal of Cardio-Thoracic Surgery · doi:10.1093/ejcts/ezs069 · Cited by 65
Lin, D., Wang, J., Jiao, N., Wang, Z., & Liu, L. (2021) · Advanced Intelligent Systems · doi:10.1002/aisy.202000211 · Cited by 65
Ultrasound-Guided Wired Magnetic Microrobot With Active Steering and Ejectable Tip
Yang, Z., Yang, L., Zhang, M., Xia, N., & Zhang, L. (2022) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2022.3153809 · Cited by 62
Self-citationNguyen, K. T., Kim, S., Min, H.-K., Hoang, M. C., Go, G., Kang, B., Kim, J., Choi, E., Hong, A., Park, J., et al. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3046513 · Cited by 60
Elasticity and fracture strain of whole blood clots.
Riha, P., Wang, X., Liao, R., & Stoltz, J. F. (1999) · PubMed · Cited by 58
Modeling magnetic soft continuum robot in nonuniform magnetic fields via energy minimization
Li, J., & Wang, L. (2024) · International Journal of Mechanical Sciences · doi:10.1016/j.ijmecsci.2024.109688 · Cited by 57
Self-citationDiverse thrombus composition in thrombectomy stroke patients with longer time to recanalization
Abbasi, M., Larco, J. A., Mereuta, M. O., Liu, Y., Fitzgerald, S. T., Dai, D., Kadirvel, R., Savastano, L. E., Kallmes, D. F., & Brinjikji, W. (2021) · Thrombosis Research · doi:10.1016/j.thromres.2021.11.018 · Cited by 52
A Flexible Catheter System for Ultrasound-Guided Magnetic Projectile Delivery
Sikorski, J., Heunis, C. M., Obeid, R., Venkiteswaran, V. K., & Misra, S. (2021) · IEEE Transactions on Robotics · doi:10.1109/tro.2021.3123865 · Cited by 51
Yang, Z., Yang, L., Zhang, M., Zhang, C., Yu, S. C. H., & Zhang, L. (2021) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2021.3121267 · Cited by 48
Self-citationImpact of aspiration catheter size on clinical outcomes in aspiration thrombectomy
Schartz, D. A., Ellens, N. R., Kohli, G. S., Rahmani, R., Akkipeddi, S. M. K., Colby, G. P., Hui, F. K., Bhalla, T., Mattingly, T. K., & Bender, M. T. (2022) · Journal of NeuroInterventional Surgery · doi:10.1136/jnis-2022-019246 · Cited by 48
Yuan, S., Xu, C., Cui, B., Zhang, T., Liang, B., Yuan, W., & Ren, H. (2024) · Nature Communications · doi:10.1038/s41467-024-51633-x · Cited by 48
A Novel Wireless 5-D Electromagnetic Tracking System Based on Nine-Channel Sinusoidal Signals
Yang, W., Zhang, C., Dai, H., Hu, C., & Xia, X. (2020) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2020.3011732 · Cited by 47
Wang, L., Zhao, Y., Zheng, B., Huo, Y., Fan, Y., Ma, D., Gu, Y., & Wang, P. (2023) · Science Advances · doi:10.1126/sciadv.adg8600 · Cited by 46
Self-citationLee, W., Nam, J., Kim, J.-Y., Jung, E., Kim, N., & Jang, G. (2020) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2020.2965480 · Cited by 44
Maegerlein, C., Prothmann, S., Lucia, K. E., Zimmer, C., Friedrich, B., & Kaesmacher, J. (2017) · CardioVascular and Interventional Radiology · doi:10.1007/s00270-017-1614-4 · Cited by 43
Tang, X., Manamanchaiyaporn, L., Zhou, Q., Huang, C., Li, L., Li, Z., Wang, L., Wang, J., Ren, L., Xu, T., et al. (2022) · Small · doi:10.1002/smll.202202848 · Cited by 43
Self-citationSimultaneous Localization and Actuation Using Electromagnetic Navigation Systems
von Arx, D., Fischer, C., Torlakcik, H., Pané, S., Nelson, B. J., & Boehler, Q. (2023) · IEEE Transactions on Robotics · doi:10.1109/tro.2023.3340324 · Cited by 41
Reprogrammable Soft Swimmers for Minimally Invasive Thrombus Extraction
Pozhitkova, A. V., Kladko, D. V., Винник, Д. А., Taskaev, S. V., & Vinogradov, V. V. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c04745 · Cited by 39
Magnetically assisted soft milli-tools for occluded lumen morphology detection
Yan, Y., Wang, T., Zhang, R., Liu, Y., Hu, W., & Sitti, M. (2023) · Science Advances · doi:10.1126/sciadv.adi3979 · Cited by 38
Zhang, T., Li, G., Ren, H., Yang, L., Yang, X., Tan, R., Tang, Y., Guo, D., Zhao, H., Shang, W., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-55199-6 · Cited by 38
Self-citationZhang, B., Wu, H., Kim, H., Welch, P. J., Cornett, A. M., Stocker, G. E., Nogueira, R. G., Kim, J., Owens, G. E., Dayton, P. A., et al. (2023) · Research · doi:10.34133/research.0048 · Cited by 36
Kahilogullari, G., Ugur, H. C., Comert, A., Tekdemir, I., & Kanpolat, Y. (2012) · Journal of neurosurgery · doi:10.3171/2012.1.jns111013 · Cited by 35 · cited 1×
“To assess the accessibility of the current microcath-eter design and two proposed downsized alternatives, we analyzed anatomical vessel data from the MCA (table S4) ( 59 ).”
Separable and Recombinable Magnetic Robot for Robotic Endovascular Intervention
Sa, J., Park, J., Jung, E., Kim, N., Lee, D., Bae, S., Lee, Y., & Jang, G. (2023) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2023.3243801 · Cited by 33
Rapid Blood Clot Removal via Remote Delamination and Magnetization of Clot Debris
Wang, Q., Wang, B., Chan, K. F., Song, X., Wang, Q., Ji, F., Su, L., Ip, B. Y. M., Ko, H., Chiu, P. W. Y., et al. (2025) · Advanced Science · doi:10.1002/advs.202415305 · Cited by 30
Self-citationDesign and Evaluation of a Magnetic Rotablation Catheter for Arterial Stenosis
Heunis, C. M., Behrendt, K. J., Hekman, E. E. G., Moers, C., de Vries, J. P. P. M., & Misra, S. (2021) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2021.3092608 · Cited by 29
Fiber Bragg Grating-Based Sensor System for Sensing the Shape of Flexible Needles
Zhang, L., Li, C., Dong, H., Liu, X., Sun, T. S., Grattan, K. T. V., & Zhao, J. (2022) · Measurement · doi:10.1016/j.measurement.2022.112251 · Cited by 29
Self-citationMR signal change in venous thrombus relates organizing process and thrombolytic response in rabbit
Kuroiwa, Y., Yamashita, A., Miyati, T., Furukoji, E., Takahashi, M., Azuma, T., Sugimura, H., Asanuma, T., Tamura, S., Kawai, K., et al. (2011) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2011.04.015 · Cited by 25
Interactive Multi‐Stage Robotic Positioner for Intra‐Operative MRI‐Guided Stereotactic Neurosurgery
He, Z., Dai, J., Ho, J., Tong, H., Wang, X., Fang, G., Liang, L., Cheung, C., Guo, Z., Chang, H., et al. (2023) · Advanced Science · doi:10.1002/advs.202305495 · Cited by 23
Mechanical thrombectomy of distal cerebral vessel occlusions of the anterior circulation
Sepp, D., Petzsche, M. R. H., Zarth, T., Wunderlich, S., Ikenberg, B. D., Maegerlein, C., Zimmer, C., Berndt, M. T., Boeckh‐Behrens, T., & Kirschke, J. S. (2023) · Scientific Reports · doi:10.1038/s41598-023-32634-0 · Cited by 21
Magnetically Guided Microcatheter for Targeted Injection of Magnetic Particle Swarms
Torlakcik, H., Sevim, S., Alves, P., Mattmann, M., Llacer‐Wintle, J., Pinto, M., Moreira, R., Flouris, A. D., Landers, F. C., Chen, X., et al. (2024) · Advanced Science · doi:10.1002/advs.202404061 · Cited by 21
Friction injury of the central vein caused by catheter for hemodialysis: an in vitro study
Wang, Z., Wang, K., & Xu, Y. (2024) · Scientific Reports · doi:10.1038/s41598-024-56485-5 · Cited by 19
Magnetic Micro-Driller System for Nasolacrimal Duct Recanalization
Yang, H., Yang, Z., Jin, D., Su, L., Chan, K. F., Chong, K. K. L., Pang, C. P., & Zhang, L. (2022) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2022.3182105 · Cited by 18
Self-citationA Fast Soft Robotic Laser Sweeping System Using Data-Driven Modeling Approach
Wang, K., Wang, X., Ho, J., Fang, G., Zhu, B., Xie, R., Liu, Y., Au, K. W. S., Chan, J. Y. K., & Kwok, K. (2023) · IEEE Transactions on Robotics · doi:10.1109/tro.2023.3262118 · Cited by 17
The Columbus steerable guidewire in neurointerventions: early clinical experience and applications
von Hessling, A., del Castillo, T. R., Karwacki, G. M., & Roos, J. E. (2021) · Journal of NeuroInterventional Surgery · doi:10.1136/neurintsurg-2021-017296 · Cited by 16
Introduction of a Rabbit Carotid Artery Model for Sonothrombolysis Research
Hölscher, T., Fisher, D. J., Ahadi, G., & Voie, A. H. (2012) · Translational Stroke Research · doi:10.1007/s12975-012-0194-5 · Cited by 15
Endovascular Microrobotics for Neurointervention
Santiago-Dieppa, D. R., & Friend, J. (2024) · Annual Review of Control, Robotics, and Autonomous Systems · doi:10.1146/annurev-control-060523-010720 · Cited by 12 · cited 1×
“First, the absence of active steerability compromises procedural precision, leading to prolonged operation times, suboptimal device placement, and poorer patient outcomes ( 6 – 11 ).”
Screw-Tip Soft Magnetically Steerable Needles
Sperry, A. J., Schwehr, T. J., Pinegar, E. K., Richards, O. B., Rolston, J. D., Alexander, M. D., Coats, B., Abbott, J. J., & Kuntz, A. (2023) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2023.3265721 · Cited by 12
Steerable Sheath Technology in the Ablation of Atrial Fibrillation
Joseph, J. P., Wong, K., Ginks, M., Bashir, Y., Betts, T. R., & Rajappan, K. (2014) · Recent Advances in Cardiovascular Drug Discovery (Formerly Recent Patents on Cardiovascular Drug Discovery) · doi:10.2174/1574890109666140228154948 · Cited by 10
Chu, Y., Liu, H., Xing, P., Lou, G., & Wu, C. (2011) · Laboratory Animals · doi:10.1258/la.2011.011022 · Cited by 8
Berger, M. C., Simgen, A., Dietrich, P., & Naziri, W. (2025) · Neurointervention · doi:10.5469/neuroint.2024.00500 · Cited by 8
High‐speed submillimeter magnetic‐driven rotational scanning side‐imaging OCT probe
Zhang, C., Yang, F., Wang, F., Tan, Q., He, H., Zhang, S., Wu, J., & Tu, S. (2023) · Journal of Biophotonics · doi:10.1002/jbio.202300106 · Cited by 7
Novel mitral clipping technique overcoming extreme atrial dilatation
Tiroch, K. A., Vorpahl, M., & Seyfarth, M. (2013) · Catheterization and Cardiovascular Interventions · doi:10.1002/ccd.25269 · Cited by 6
Cardiovascular catheter stiffness – a static measurement approach
Brandt-Wunderlich, C., Grabow, N., Schmitz, K., Siewert, S., & Schmidt, W. (2021) · Current Directions in Biomedical Engineering · doi:10.1515/cdbme-2021-2184 · Cited by 5
Zhang, B., Wu, H., Kim, H., et al. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.11.02.514936 · Cited by 3
Villalba, L., Deen, R., Tonson-Older, B., & Costello, C. (2024) · Journal of Vascular Surgery Venous and Lymphatic Disorders · doi:10.1016/j.jvsv.2024.101899 · Cited by 3
An Echogenic Clot Method for Thrombolysis Monitoring in Thrombotic Stroke Models
Carter, D. E. (2023) · Medical Research Archives · doi:10.18103/mra.v11i3.3702 · Cited by 2
Risks of microcatheter injections in acute stroke treatment
Gupta, R. G. (2009) · Nature Reviews Neurology · doi:10.1038/nrneurol.2009.29 · Cited by 1
Cited by
Micro/Nanorobots for Combating Brain Disorders: Challenges, Advances, and Perspectives
Zhang, Q., & Sun, D. B. (2025) · Advanced Science · doi:10.1002/advs.202516592 · Cited by 15
Bio-inspired Magnetic Helical Miniature Robots: Mechanisms, Control and Biomedical Applications
Zhu, A., Li, Y., Zheng, Y., & Yang, L. (2025) · Journal of Bionic Engineering · doi:10.1007/s42235-025-00795-5 · Cited by 13
Self-citationMagnetic slippery microcatheter with artificial cilia for low-friction interventions
Yang, S., Shen, J., He, J., Xia, Z., Zhang, Y., & Wang, B. (2025) · Science Advances · doi:10.1126/sciadv.adw9926 · Cited by 10
Biophysics-informed design of biohybrid microrobots
Quan, X., Sun, B., Song, X., & Zhang, L. (2026) · Nature Reviews Bioengineering · doi:10.1038/s44222-026-00416-8 · Cited by 10
Self-citationStable magnetic soft structures
Wang, H., Zhang, Y., Liu, X., Li, H., Chen, X., Cao, Z.-Q., Luo, Q., Ren, Z., & Hu, W. (2025) · Science Advances · doi:10.1126/sciadv.adz4952 · Cited by 5
Self-citation
Metadata
- Authors
- Zhang, M., Yang, L., Yang, H., Su, L., Xue, J., Wang, Q., Hao, B., Jiang, Y., Chan, K. F., Sung, J. J. Y., Ko, H., Liu, X., Wang, L., Ip, B. Y. M., Leung, T. W. H., & Zhang, L.
- Affiliations
- Chinese University of Hong Kong (HK)
- Hong Kong Polytechnic University (HK)
- Nanyang Technological University (SG)
- University of Science and Technology of China (CN)
- Corresponding author
- Yang, Lidong
- Ip, Bonaventure Yiu Ming
- Leung, Thomas Wai Hong
- Zhang, Li
- ORCID
- Venue
- Science Advances
- Published
- 2025-06-20
- Volume
- 11
- Issue
- 25
- Pages
- eadv1682-eadv1682
- ISSN
- 2375-2548 (ISSN-L)
- PMC
- PMC13108821
- OpenAlex
- W4411472631
- Cited by
- 36
- References
- 79
- Citations by year
- 2026: 26 · 2025: 10
- FWCI
- 11.27
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- Science Advances (publishedVersion, cc-by)
- PubMed (publishedVersion)
- PubMed Central (publishedVersion, cc-by)
- DR-NTU (Nanyang Technological University) (submittedVersion)
- PolyU Institutional Research Archive (Hong Kong Polytechnic University) (submittedVersion, cc-by)
- Topics
- Physical Sciences › Physics and Astronomy › Condensed Matter Physics › Micro and Nano Robotics
- Physical Sciences › Engineering › Biomedical Engineering › Soft Robotics and Applications
- Physical Sciences › Engineering › Mechanical Engineering › Advanced Materials and Mechanics
- Keywords
- endovascular access, thrombolysis, clot retrieval, active steering, soft robotics, microcatheterization, rabbit model
- MeSH terms
- Animals, Equipment Design, Female, Humans, Pregnancy, Rabbits, Thrombosis, Phantoms, Imaging, Endovascular Procedures, Catheters
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.