How well do activity monitors estimate energy expenditure? A systematic review and meta-analysis of the validity of current technologies
Open accessAbstract
Objective: To determine the accuracy of wrist and arm-worn activity monitors’ estimates of energy expenditure (EE). Data sources: SportDISCUS (EBSCOHost), PubMed, MEDLINE (Ovid), PsycINFO (EBSCOHost), Embase (Ovid) and CINAHL (EBSCOHost). Design: A random effects meta-analysis was performed to evaluate the difference in EE estimates between activity monitors and criterion measurements. Moderator analyses were conducted to determine the benefit of additional sensors and to compare the accuracy of devices used for research purposes with commercially available devices. Eligibility criteria: We included studies validating EE estimates from wrist-worn or arm-worn activity monitors against criterion measures (indirect calorimetry, room calorimeters and doubly labelled water) in healthy adult populations. Results: 60 studies (104 effect sizes) were included in the meta-analysis. Devices showed variable accuracy depending on activity type. Large and significant heterogeneity was observed for many devices (I 2 >75%). Combining heart rate or heat sensing technology with accelerometry decreased the error in most activity types. Research-grade devices were statistically more accurate for comparisons of total EE but less accurate than commercial devices during ambulatory activity and sedentary tasks. Conclusions: EE estimates from wrist and arm-worn devices differ in accuracy depending on activity type. Addition of physiological sensors improves estimates of EE, and research-grade devices are superior for total EE. These data highlight the need to improve estimates of EE from wearable devices, and one way this can be achieved is with the addition of heart rate to accelerometry. PROSPEROregistration number: CRD42018085016.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Downs, S. H., & Black, N. (1998) · Journal of Epidemiology and Community Health · doi:10.1136/jech.52.6.377 · Cited by 8,252
Distribution Theory for Glass's Estimator of Effect Size and Related Estimators
Hedges, L. V. (1981) · Journal of Educational Statistics · doi:10.2307/1164588 · Cited by 5,695
Evaluating non-randomised intervention studies
Deeks, J., Dinnes, J., D'Amico, R., Sowden, A., Sakarovitch, C., Song, F., Petticrew, M., & Altman, D. (2003) · Health Technology Assessment · doi:10.3310/hta7270 · Cited by 2,907
A Re-Evaluation of Random-Effects Meta-Analysis
Higgins, J. P. T., Thompson, S. G., & Spiegelhalter, D. J. (2009) · Journal of the Royal Statistical Society Series A: Statistics in Society · doi:10.1111/j.1467-985x.2008.00552.x · Cited by 2,569
Systematic review of the validity and reliability of consumer-wearable activity trackers
Evenson, K. R., Goto, M. M., & Furberg, R. D. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0314-1 · Cited by 1,232
Shcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625
Energy balance measurement: when something is not better than nothing
Dhurandhar, N. V., Schoeller, D., Brown, A. W., Heymsfield, S. B., Thomas, D., Sørensen, T. I. A., Speakman, J. R., Jeansonne, M., Allison, D. B., & the Energy Balance Measurement Working Group (2015) · International Journal of Obesity · doi:10.1038/ijo.2014.199 · Cited by 567
Fitbit®: An accurate and reliable device for wireless physical activity tracking
Diaz, K. M., Krupka, D. J., Chang, M. J., Peacock, J., Ma, Y., Goldsmith, J., Schwartz, J. E., & Davidson, K. W. (2015) · International Journal of Cardiology · doi:10.1016/j.ijcard.2015.03.038 · Cited by 405
Validity of Consumer-Based Physical Activity Monitors
LEE, J.-M., KIM, Y., & WELK, G. J. (2014) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000287 · Cited by 401
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336
Research works printed but not linked: 78
- 10.1136/bmj.k7 · DOI
- Agha, 2017, The rising prevalence of obesity, Int J Surg Oncol, 2, e17, 10.1097/IJ9.0000000000000017 · DOI · Paper page
- 10.3945/an.115.008755 · DOI · Paper page
- 10.1063/1.354123 · doi:10.1063/1.354123
- DeLany, 2012, Measurement of energy expenditure, Pediatr Blood Cancer, 58, 129, 10.1002/pbc.23369 · DOI
- Hills, 2014, Assessment of physical activity and energy expenditure: an overview of objective measures, Front Nutr, 1, 1, 10.3389/fnut.2014.00005 · DOI · Paper page
- 10.1089/152091503322641088 · DOI
- Diaz, 2016, Validation of the Fitbit One® for physical activity measurement at an upper torso attachment site, BMC Res Notes, 9, 213, 10.1186/s13104-016-2020-8 · DOI · Paper page
- 10.1038/ejcn.2014.51 · DOI
- 10.3389/fphys.2017.00788 · DOI · Paper page
- 10.1001/jama.2017.19163 · DOI · Paper page
- Nelson, 2016, Validity of consumer-based physical activity monitors for specific activity types, Med Sci Sports Exerc, 48, 1619, 10.1249/MSS.0000000000000933 · DOI · Paper page
- 10.1186/1479-5868-9-84 · DOI · Paper page
- 10.1161/JAHA.116.003231 · DOI · Paper page
- Cohen J . Statistical power analysis for the behavioral sciences. Massachusetts: Academic Press, 1977.
- 10.1002/sim.1186 · DOI · Paper page
- 10.1136/bmj.315.7109.629 · DOI · Paper page
- Lopez, 2018, Validation of sensewear armband in children, adolescents, and adults, Scand J Med Sci Sports, 28, 487, 10.1111/sms.12920 · doi:10.1111/sms.12920
- Reeve, 2014, Accuracy of the sensewear armband mini and the bodymedia FIT in resistance training, J Sci Med Sport, 17, 630, 10.1016/j.jsams.2013.08.007 · DOI
- Machač, 2013, Validation of physical activity monitors in individuals with diabetes: energy expenditure estimation by the multisensor sensewear armband Pro3 and the step counter Omron HJ-720 against indirect calorimetry during walking, Diabetes Technol Ther, 15, 413, 10.1089/dia.2012.0235 · DOI
- Alsubheen, 2016, Accuracy of the vivofit activity tracker, J Med Eng Technol, 40, 298, 10.1080/03091902.2016.1193238 · DOI
- Bai, 2018, Comparative evaluation of heart rate-based monitors: apple watch vs fitbit charge HR, J Sports Sci, 36, 1734, 10.1080/02640414.2017.1412235 · DOI
- Benito, 2012, Validation of the sensewear armband in circuit resistance training with different loads, Eur J Appl Physiol, 112, 3155, 10.1007/s00421-011-2269-5 · DOI
- 10.1136/bjsm.2008.048868 · DOI · Paper page
- 10.1111/j.1600-0412.2011.01172.x · DOI
- Bhammar, 2016, Validity of sensewear® Armband v5.2 and v2.2 for estimating energy expenditure, J Sports Sci, 34, 1830, 10.1080/02640414.2016.1140220 · DOI
- Brazeau, 2011, Accuracy of the SenseWear Armband™ during ergocycling, Int J Sports Med, 32, 761, 10.1055/s-0031-1279768 · doi:10.1055/s-0031-1279768
- Brazeau, 2014, Accuracy of energy expenditure estimation by activity monitors differs with ethnicity, Int J Sports Med, 35, 847, 10.1055/s-0034-1371837 · doi:10.1055/s-0034-1371837
- 10.1016/j.jsams.2014.11.001 · DOI
- 10.1136/bjsm.2007.045575 · DOI
- 10.1186/s12966-014-0119-7 · DOI
- 10.1038/ejcn.2014.241 · DOI
- Casiraghi, 2013, Energy expenditure evaluation in humans and non-human primates by sensewear armband. Validation of energy expenditure evaluation by SenseWear Armband by direct comparison with indirect calorimetry, PLoS One, 8, 10.1371/journal.pone.0073651 · DOI
- 10.1249/MSS.0b013e3181fc7162 · DOI · Paper page
- Correa, 2016, Evaluation of the ability of three physical activity monitors to predict weight change and estimate energy expenditure, Appl Physiol Nutr Metab, 41, 758, 10.1139/apnm-2015-0461 · DOI
- Dondzila, 2016, Comparative accuracy of fitness tracking modalities in quantifying energy expenditure, J Med Eng Technol, 40, 325, 10.1080/03091902.2016.1197978 · DOI
- Erdogan, 2010, Accuracy of the polar S810i(TM) heart rate monitor and the sensewear pro armband(TM) to estimate energy expenditure of indoor rowing exercise in overweight and obese individuals, J Sports Sci Med, 9, 508
- 10.1249/01.MSS.0000128144.91337.38 · DOI · Paper page
- Gastin, 2018, Validity of the ActiGraph GT3X+ and BodyMedia SenseWear Armband to estimate energy expenditure during physical activity and sport, J Sci Med Sport, 21, 291, 10.1016/j.jsams.2017.07.022 · DOI
- Heiermann, 2011, Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetry, Gerontology, 57, 473, 10.1159/000322109 · DOI
- Imboden, 2018, Comparison of four Fitbit and Jawbone activity monitors with a research-grade ActiGraph accelerometer for estimating physical activity and energy expenditure, Br J Sports Med, 52, bjsports-2016-096990, 10.1136/bjsports-2016-096990 · DOI
- Jakicic, 2004, Evaluation of the sensewear pro armband to assess energy expenditure during exercise, Med Sci Sports Exerc, 36, 897, 10.1249/01.MSS.0000126805.32659.43 · DOI · Paper page
- 10.1249/MSS.0b013e3181e0b3ff · DOI · Paper page
- Kim, 2015, Criterion validity of competing accelerometry-based activity monitoring devices, Med Sci Sports Exerc, 47, 2456, 10.1249/MSS.0000000000000691 · DOI
- 10.1249/01.MSS.0000132379.09364.F8 · DOI · Paper page
- Koehler, 2011, Assessing energy expenditure in male endurance athletes: validity of the sensewear armband, Med Sci Sports Exerc, 43, 1328, 10.1249/MSS.0b013e31820750f5 · DOI
- Lee, 2011, Accuracy of an infrared LED device to measure heart rate and energy expenditure during rest and exercise, J Sports Sci, 29, 1645, 10.1080/02640414.2011.609899 · DOI
- 10.1093/gerona/glr101. · DOI
- Martien, 2015, Energy expenditure in institutionalized older adults: validation of sensewear mini, Med Sci Sports Exerc, 47, 1265, 10.1249/MSS.0000000000000529 · DOI
- 10.1016/j.ypmed.2012.02.013 · DOI
- Melanson, 2009, A novel approach for measuring energy expenditure in free-living humans, Conf Proc IEEE Eng Med Biol Soc, 2009, 6873
- Montoye, 2017, Comparative accuracy of a wrist-worn activity tracker and a smart shirt for physical activity assessment, Meas Phys Educ Exerc Sci, 21, 201, 10.1080/1091367X.2017.1331166 · DOI
- Murakami, 2016, Accuracy of wearable devices for estimating total energy expenditure, JAMA Intern Med, 176, 702, 10.1001/jamainternmed.2016.0152 · DOI · Paper page
- Papazoglou, 2006, Evaluation of a multisensor armband in estimating energy expenditure in obese individuals, Obesity, 14, 2217, 10.1038/oby.2006.260 · DOI
- Price, 2017, Validation of the fitbit one, garmin vivofit and jawbone UP activity tracker in estimation of energy expenditure during treadmill walking and running, J Med Eng Technol, 41, 208, 10.1080/03091902.2016.1253795 · DOI
- Reece, 2015, Validation of the sensewear armband as a measure of sedentary behavior and light Activity, J Phys Act Health, 12, 1229, 10.1123/jpah.2014-0136 · DOI
- Rousset, 2015, Comparison of total energy expenditure assessed by two devices in controlled and free-living conditions, Eur J Sport Sci, 15, 391, 10.1080/17461391.2014.949309 · DOI
- Ryan, 2013, Measurement of energy expenditure by activity monitors, Phys Ther Rev, 18, 239, 10.1179/1743288X13Y.0000000063 · DOI
- Slinde, 2013, Energy expenditure by multisensor armband in overweight and obese lactating women validated by doubly labeled water, Obesity, 21, 2231, 10.1002/oby.20363 · DOI
- 10.1249/MSS.0b013e31825ce76f · DOI
- Stackpool, 2013, The accuracy of various activity trackers in estimating steps taken and energy expenditure, J Fit Res, 53, 1689
- St-Onge, 2007, Evaluation of a portable device to measure daily energy expenditure in free-living adults, Am J Clin Nutr, 85, 742, 10.1093/ajcn/85.3.742 · DOI · Paper page
- Soric, 2012, Validation of the sensewear armband during recreational in-line skating, Eur J Appl Physiol, 112, 1183, 10.1007/s00421-011-2045-6 · DOI
- Tucker, 2015, Validity and reliability of Nike + Fuelband for estimating physical activity energy expenditure, BMC Sports Sci Med Rehabil, 7, 14, 10.1186/s13102-015-0008-7 · DOI
- 10.1186/1756-0500-5-187 · DOI
- 10.1519/JSC.0b013e3182a1f836 · DOI
- Van Hoye, 2015, Validation of the sensewear armband in different ambient temperatures, J Sports Sci, 33, 1007, 10.1080/02640414.2014.981846 · DOI
- Vernillo, 2015, Validity of the sensewear armband to assess energy expenditure in graded walking, J Phys Act Health, 12, 178, 10.1123/jpah.2013-0437 · DOI
- Wahl, 2017, Criterion-validity of commercially available physical activity tracker to estimate step count, covered distance and energy expenditure during sports conditions, Front Physiol, 8, 725, 10.3389/fphys.2017.00725 · DOI · Paper page
- 10.3945/ajcn.115.111070 · DOI
- Pollard, 2017, Gender differences in walking (for leisure, transport and in total) across adult life: a systematic review, BMC Public Health, 17, 341, 10.1186/s12889-017-4253-4 · doi:10.1186/s12889-017-4253-4
- Schneller, 2015, Validation of five minimally obstructive methods to estimate physical activity energy expenditure in young adults in semi-standardized settings, Sensors, 15, 6133, 10.3390/s150306133 · DOI
- Welk, 2007, Field validation of the MTI actigraph and bodymedia armband monitor using the IDEEA monitor, Obesity, 15, 918, 10.1038/oby.2007.624 · DOI · Paper page
- Montoye, 2016, Comparison of activity type classification accuracy from accelerometers worn on the hip, wrists, and thigh in young, apparently healthy adults, Meas Phys Educ Exerc Sci, 20, 173, 10.1080/1091367X.2016.1192038 · DOI
- Plasqui, 2017, Smart approaches for assessing free-living energy expenditure following identification of types of physical activity, Obes Rev, 18 Suppl 1, 50, 10.1111/obr.12506 · DOI
- 10.1152/japplphysiol.00092.2006 · DOI · Paper page
- 10.1093/ije/dyn204 · DOI · Paper page
- Kipp, 2018, Calculating metabolic energy expenditure across a wide range of exercise intensities: the equation matters, Appl Physiol Nutr Metab, 43, 639, 10.1139/apnm-2017-0781 · DOI
Cited by
Exploring the Role of Wearable Technology in Sport Kinematics and Kinetics: A Systematic Review
Adesida, Y., Papi, E., & McGregor, A. H. (2019) · Sensors · doi:10.3390/s19071597 · Cited by 288
Stellingwerff, T., Heikura, I. A., Meeusen, R., Bermon, S., Seiler, S., Mountjoy, M. L., & Burke, L. M. (2021) · Sports Medicine · doi:10.1007/s40279-021-01491-0 · Cited by 171
Doherty, C., Baldwin, M., Keogh, A., Caulfield, B., & Argent, R. (2024) · Sports Medicine · doi:10.1007/s40279-024-02077-2 · Cited by 133
Low energy availability in female athletes: from the lab to the field
Heikura, I. A., Stellingwerff, T., & Areta, J. L. (2022) · European Journal of Sport Science · doi:10.1080/17461391.2021.1915391 · Cited by 84
“In general, HR monitors or power meters (cycling, Heikura et al., 2019) are likely to be superior to the use of metabolic equivalents (METs, Ainsworth, 2000), but even these have the potential for errors in the magnitude of ∼100 to 600 kcal/d (O’Driscoll, 2020).”
BackgroundTuricchi, J., O'Driscoll, R., Finlayson, G., Beaulieu, K., Deighton, K., & Stubbs, R. J. (2019) · Obesity Reviews · doi:10.1111/obr.12849 · Cited by 37
Self-citation
Metadata
- Authors
- O’Driscoll, R., Turicchi, J., Beaulieu, K., Scott, S., Matu, J., Deighton, K., Finlayson, G., & Stubbs, J.
- Venue
- British Journal of Sports Medicine
- Publisher
- BMJ
- Published
- 2018-09-07
- Volume
- 54
- Issue
- 6
- Pages
- 332-340
- ISSN
- 0306-3674 (ISSN-L)
- 1473-0480
- OpenAlex
- W2889930586
- Cited by
- 188
- References
- 19
- FWCI
- 2.55
- Type
- journal article
- Language
- en
- Open access
- hybrid
- Funders
- University of Leeds
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: How well do activity monitors estimate energy expenditure? A systematic review and meta-analysis of the validity of current technologies. https://app.cochat.ai/papers/10.1136/bjsports-2018-099643
- The paper
- O’Driscoll, R., Turicchi, J., Beaulieu, K., Scott, S., Matu, J., Deighton, K., Finlayson, G., & Stubbs, J. (2020). How well do activity monitors estimate energy expenditure? A systematic review and meta-analysis of the validity of current technologies. British Journal of Sports Medicine. https://doi.org/10.1136/bjsports-2018-099643
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.