A dual-respiration chamber system with automated calibration
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Schoffelen, Paul F. M., Klaas R. Westerterp, Wim H. M. Saris, and Foppe Ten Hoor. A dual-respiration chamber system with automated calibration. J. Appl. Physiol. 83(6): 2064–2072, 1997.—This study characterizes respiration chambers with fully automated calibration. The system consists of two 14-m 3 pull-type chambers. Care was taken to provide a friendly environment for the subjects, with the possibility of social contact during the experiment. Gas analysis was automated to correct for analyzer drift and barometric pressure variations and to provide ease of use. Methods used for checking the system’s performance are described. The gas-analysis repeatability was within 0.002%. Results of alcohol combustion (50–350 ml/min CO 2) show an accuracy of 0.5 ± 2.0 (SD) % for O 2 consumption and −0.3 ± 1.6% for CO 2 production for 2- to 24-h experiments. It is concluded that response time is not the main factor with respect to the smallest practical measurement interval (duration); volume, mixing, gas-analysis accuracy, and levels of O 2 consumption and CO 2 production are at least equally important. The smallest practical interval was 15–25 min, as also found with most chamber systems described in the literature. We chose to standardize 0.5 h as the minimum measurement interval.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Determinants of 24-hour energy expenditure in man. Methods and results using a respiratory chamber.
Ravussin, E., Lillioja, S., Anderson, T. E., Christin, L., & Bogardus, C. (1986) · Journal of Clinical Investigation · doi:10.1172/jci112749 · Cited by 1,166 · cited 3×
“process (8, 12, 15), and the other is the (90–99%)”
BackgroundMethod
Cited by
Timmers, S., Konings, E., Bilet, L., Houtkooper, R. H., van de Weijer, T., Goossens, G. H., Hoeks, J., van der Krieken, S., Ryu, D., Kersten, S., et al. (2011) · Cell Metabolism · doi:10.1016/j.cmet.2011.10.002 · Cited by 1,320
van Can, J., Sloth, B., Jensen, C. B., Flint, A., Blaak, E. E., & Saris, W. H. M. (2014) · International Journal of Obesity · doi:10.1038/ijo.2013.162 · Cited by 537
Self-citationDirks, M. L., Wall, B. T., van de Valk, B., Holloway, T. M., Holloway, G. P., Chabowski, A., Goossens, G. H., & van Loon, L. J. C. (2016) · Diabetes · doi:10.2337/db15-1661 · Cited by 391
“Dietary Intake During the screening visit, resting energy expenditure was measured by indirect calorimetry using an open-circuit ventilated hood system (Omnical; Maastricht University, Maastricht, the Netherlands) (25).”
MethodLejeune, M. P., Westerterp, K. R., Adam, T. C., Luscombe-Marsh, N. D., & Westerterp-Plantenga, M. S. (2006) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/83.1.89 · Cited by 363
“The gas samples to be measured were selected by a computer that also stored and processed the data (27).”
MethodSelf-citationReijnders, D., Goossens, G. H., Hermes, G. D. A., Neis, E. P. J. G., van der Beek, C. M., Most, J., Holst, J. J., Lenaerts, K., Kootte, R. S., Nieuwdorp, M., et al. (2016) · Cell Metabolism · doi:10.1016/j.cmet.2016.06.016 · Cited by 348
Westerterp‐Plantenga, M. S., Lejeune, M. P. G. M., & Kovacs, E. M. R. (2005) · Obesity Research · doi:10.1038/oby.2005.142 · Cited by 328
High protein intake sustains weight maintenance after body weight loss in humans
Westerterp-Plantenga, M. S., Lejeune, M. P. G. M., Nijs, I., van Ooijen, M., & Kovacs, E. M. R. (2004) · International Journal of Obesity · doi:10.1038/sj.ijo.0802461 · Cited by 326
Diet induced thermogenesis measured over 24h in a respiration chamber: effect of diet composition
Westerterp, K., Wilson, S., & Rolland, V (1999) · International Journal of Obesity · doi:10.1038/sj.ijo.0800810 · Cited by 251
Self-citationWeekly Subcutaneous Pegylated Recombinant Native Human Leptin (PEG-OB) Administration in Obese Men
Hukshorn, C. J., Saris, W. H. M., Westerterp-Plantenga, M. S., Farid, A. R., Smith, F. J., & Campfield, L. A. (2000) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jcem.85.11.6955 · Cited by 220
Self-citationDemonstration of a day-night rhythm in human skeletal muscle oxidative capacity
van Moorsel, D., Hansen, J., Havekes, B., Scheer, F. A. J. L., Jörgensen, J. A., Hoeks, J., Schrauwen-Hinderling, V. B., Duez, H., Lefebvre, P., Schaper, N. C., et al. (2016) · Molecular Metabolism · doi:10.1016/j.molmet.2016.06.012 · Cited by 218
Additional Protein Intake Limits Weight Regain after Weight Loss in Humans
Lejeune, M. P. G. M., Kovacs, E. M. R., & Westerterp-Plantenga, M. S. (2005) · British Journal of Nutrition · doi:10.1079/bjn20041305 · Cited by 211
“Gas analyses were performed by a paramagnetic oxygen analyser (type 500A, Servomex, Crowborough, Sussex, UK) and an IR carbon dioxide analyser (type 500A, Servomex, Crowborough, Sussex, UK), similar to the analysis system described by Schoffelen et al. (1997).”
MethodJanssen, X., Cliff, D. P., Reilly, J. J., Hinkley, T., Jones, R. A., Batterham, M., Ekelund, U., Brage, S., & Okely, A. D. (2013) · PLoS ONE · doi:10.1371/journal.pone.0079124 · Cited by 190
“Due to the calorimeter sampling frequency and the time lag that exists when measuring EE in large volumes, it was not possible to measure EE in time blocks shorter than 10 min [13].”
BackgroundMethodLarsen, T. M., Toubro, S., van Baak, M. A., Gottesdiener, K. M., Larson, P., Saris, W. H., & Astrup, A. (2002) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/76.4.780 · Cited by 179
Self-citationGluconeogenesis and energy expenditure after a high-protein, carbohydrate-free diet
Veldhorst, M. A., Westerterp-Plantenga, M. S., & Westerterp, K. R. (2009) · The American Journal of Clinical Nutrition · doi:10.3945/ajcn.2009.27834 · Cited by 170
Self-citationHoeks, J., van Herpen, N. A., Mensink, M., Moonen-Kornips, E., van Beurden, D., Hesselink, M. K. C., & Schrauwen, P. (2010) · Diabetes · doi:10.2337/db10-0519 · Cited by 156
Wijers, S. L. J., Schrauwen, P., Saris, W. H. M., & van Marken Lichtenbelt, W. D. (2008) · PLoS ONE · doi:10.1371/journal.pone.0001777 · Cited by 135
Self-citationIndividual Thermogenic Responses to Mild Cold and Overfeeding Are Closely Related
Wijers, S. L. J., Saris, W. H. M., & van Marken Lichtenbelt, W. D. (2007) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2007-1065 · Cited by 99
Self-citationde Ligt, M., Bruls, Y. M. H., Hansen, J., Habets, M.-F., Havekes, B., Nascimento, E. B. M., Moonen-Kornips, E., Schaart, G., Schrauwen-Hinderling, V. B., van Marken Lichtenbelt, W., et al. (2018) · Molecular Metabolism · doi:10.1016/j.molmet.2018.04.004 · Cited by 79
Hochstenbach-Waelen, A., Veldhorst, M. A., Nieuwenhuizen, A. G., Westerterp-Plantenga, M. S., & Westerterp, K. R. (2009) · The American Journal of Clinical Nutrition · doi:10.3945/ajcn.2008.26917 · Cited by 68
Self-citationWesterterp-Plantenga, M. S., Lejeune, M. P. G. M., Smeets, A. J. P. G., & Luscombe-Marsh, N. D. (2009) · Physiology & Behavior · doi:10.1016/j.physbeh.2009.03.010 · Cited by 61
Veldhorst, M. A. B., Westerterp, K. R., van Vught, A. J. A. H., & Westerterp-Plantenga, M. S. (2010) · British Journal of Nutrition · doi:10.1017/s0007114510002060 · Cited by 50
Self-citationNguyen, T., de Jonge, L., Smith, S. R., & Bray, G. A. (2003) · Medical and Biological Engineering and Computing · doi:10.1007/bf02345320 · Cited by 47
The influence of a moderate temperature drift on thermal physiology and perception
Ivanova, Y. M., Pallubinsky, H., Kramer, R., & van Marken Lichtenbelt, W. (2021) · Physiology & Behavior · doi:10.1016/j.physbeh.2020.113257 · Cited by 45
Hochstenbach-Waelen, A., Westerterp-Plantenga, M. S., Veldhorst, M. A. B., & Westerterp, K. R. (2009) · The Journal of Nutrition · doi:10.3945/jn.109.110403 · Cited by 38
Self-citationDrummen, M., Tischmann, L., Gatta-Cherifi, B., Fogelholm, M., Raben, A., Adam, T. C., & Westerterp-Plantenga, M. S. (2020) · The Journal of Nutrition · doi:10.1093/jn/nxz281 · Cited by 34
“O2 consumption and CO2 production were continuously measured by open-circuit ventilated indirect calorimetry (28).”
MethodInfluentialPohlmeier, A. M., Phy, J. L., Watkins, P., Boylan, M., Spallholz, J., Harris, K. S., & Cooper, J. A. (2014) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2014-0073 · Cited by 34
Leptin and energy restriction induced adaptation in energy expenditure
Camps, S. G. J. A., Verhoef, S. P. M., & Westerterp, K. R. (2015) · Metabolism · doi:10.1016/j.metabol.2015.06.016 · Cited by 22
Self-citation
Metadata
- Authors
- Schoffelen, P. F. M., Westerterp, K. R., Saris, W. H. M., & Ten Hoor, F.
- Venue
- Journal of Applied Physiology
- Publisher
- American Physiological Society
- Published
- 1997-12-01
- Volume
- 83
- Issue
- 6
- Pages
- 2064-2072
- ISSN
- 0161-7567 (ISSN-L)
- 8750-7587
- 1522-1601
- OpenAlex
- W2168684133
- Cited by
- 252
- References
- 1
- FWCI
- 4.62
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.